Radhakrishnan B*, Kurup P G G**, Ramakrishnan G***, Chandralingam S****

Similar documents
Evaluation of Phantom Equivalent Materials in Polychromatic Diagnostic X-Ray Beam

Journal of Chemical and Pharmaceutical Research, 2012, 4(1): Research Article

Journal of Chemical and Pharmaceutical Research, 2012, 4(9): Research Article

RESULTS AND DISCUSSION ON EFFECT OF COLLIMATOR SIZE ON THE ATTENUATION OF GAMMA RAY IN PURE ELEMENTS

Journal of Chemical and Pharmaceutical Research, 2012, 4(9): Research Article

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

Evaluation of gamma shielding parameters of bismuth oxide filled polymer composites

INTERACTIONS OF RADIATION WITH MATTER

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

X-ray Interaction with Matter

Rad T 290 Worksheet 2

1Meherun Nahar, Sazzad, 3Abdus Sattar Mollah and 4Mir Mohammad Akramuzzaman Correspondence Address INTRODUCTION MATERIALS AND METHODS Mathematical

PHYS 3650L - Modern Physics Laboratory

ATTENUATION STUDIES ON DRY AND HYDRATED CROSS-LINKED HYDROPHILIC COPOLYMER MATERIALS AT 8.02 TO kev USING X-RAY FLUORESCENT SOURCES

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE

Journal of Chemical and Pharmaceutical Research

Journal of Chemical and Pharmaceutical Research

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Physics of Radiotherapy. Lecture II: Interaction of Ionizing Radiation With Matter

Comparative Analysis of Nuclear Cross Sections in Monte Carlo Methods for Medical Physics Applications

JRPR. A Study of Shielding Properties of X-ray and Gamma in Barium Compounds. Original Research. Introduction

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

Basic physics Questions

Project Memorandum. N N o. = e (ρx)(µ/ρ) (1)

M [scale units/s] of the system

A comprehensive study on energy absorption and exposure buildup factors for some Vitamins and Tissue Equivalent Materials

Introduction. X-Ray Production and Quality. Fluorescence Yield. Fluorescence X-Rays. Initiating event. Initiating event 3/18/2011

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

International Journal of Scientific Research and Reviews

Geant4 Monte Carlo code application in photon interaction parameter of composite materials and comparison with XCOM and experimental data

Comparison of some human tissues and some commonly used thermoluminescent dosimeters for photon energy absorption

Chapter 2: Interactions of Radiation with Matter

Mass attenuation coefficients of beta particles in elements

Shielding of Ionising Radiation with the Dosimetry & Shielding Module

Gamma Radiation Absorption Characteristics of Concrete with Components of Different Type Materials

Measurement of atomic number and mass attenuation coefficient in magnesium ferrite

Mass Attenuation Coefficients of Lead Nitrate Dilute Solutions for 13.5 cm Thickness at 1173 and 1333 kev

Determination of Photon Ambient Dose Buildup Factors for Radiological Applications for Points and Plaque Source Configurations Using MCNP5

City University of Hong Kong

Current and Recent ICRU Activities in Radiation Protection Dosimetry and Measurements

Neutron Shielding Properties Of Concrete With Boron And Boron Containing Mineral

Chapter VIII: Photographic films

Low-energy broad-beam photon shielding data for constituents of concrete

Radiation Detection and Measurement

Chapter 1 Introduction of Radiation and Radioactivity

Monte Carlo simulation for the estimation of iron in human whole blood and comparison with experimental data

STANDARD WATER PHANTOM BACKSCATTER FACTORS FOR MEDIUM ENERGY X-RAYS

Efficiencies of Some Spherical Ion Chambers in Continuous and Pulsed Radiation: A Numerical Evaluation

APPLIED RADIATION PHYSICS

X-ray Spectroscopy. Danny Bennett and Maeve Madigan. October 12, 2015

X-ray Spectroscopy. c David-Alexander Robinson & Pádraig Ó Conbhuí. 14th March 2011

Dependence Of Gamma Ray Attenuation On Concentration Of Manganese (II) Chloride Solution

11/19/2014. Chapter 3: Interaction of Radiation with Matter in Radiology and Nuclear Medicine. Nuclide Families. Family Nuclides with Same: Example

International Journal of Scientific & Engineering Research Volume 9, Issue 1, January ISSN

Experimental and Theoretical Investigation of Gamma Attenuation of Building Materials

Theoretical analysis of comparative patient skin dose and exposure technique approaches in planar radiography

EEE4101F / EEE4103F Radiation Interactions & Detection

Georgia Institute of Technology. Radiation Detection & Protection (Day 3)

Radiation Dose, Biology & Risk

DETERMINATION OF X-RAY TOTAL ATTENUATION COEFFICIENT IN Zr, Ag, In FOR ENERGY RANGE BETWEEN kev

Simulation Modeling in Dosimetry

Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co

Calibration of a pencil ionization chamber with and without preamplifier

B E A M Q U A L I T Y S P E C I F I C A T I O N I N K I L O-V O L T A G E R A D I O T H E R A P Y

Spectral Filtering for Improving Quality of Material Discrimination Using Dual Energy X-rays

Chapter 3. Effective atomic numbers and electron densities for some scintillators and plasticizers

THE ACTIVITY CALIBRATOR

Evaluation of Gamma-Ray Attenuation Parameters for Some Materials

Attenuation of Radiation in Matter. Attenuation of gamma particles

ISSN: International Journal Of Core Engineering & Management (IJCEM) Volume 2, Issue 6, September 2015

Measurements of K- shell production cross-section and fluorescence yield for Y element

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440

Study of Photon Interaction Parameters for Some Oxide Glasses as Gamma Ray Shielding Materials

Linear attenuation coefficient calculation for both pure silicon (Si) and silicone supported with lead

CHAPTER 2 INTERACTION OF RADIATION WITH MATTER

ABSOLUTE AIR-KERMA MEASUREMENT IN A SYNCHROTRON LIGHT BEAM BY IONIZATION FREE-AIR CHAMBER

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays

An analytic review of the different methods for evaluating the Z-effective of the series of glass [Li 2 O B 2 O 3 SiO 2 MnO]

PHITS calculation of the radiation field in HIMAC BIO

Interactions with Matter Photons, Electrons and Neutrons

Journal of Chemical and Pharmaceutical Research

Radiation Dosimetry. Electron interactions with matter. Important processes in radiotherapy. Contents. Alun Beddoe

Quantitative determination of the effect of the harmonic component in. monochromatised synchrotron X-ray beam experiments

Progress in Nuclear Science and Technology, Volume 6,

Determination of the shielding power of different materials against gamma radiation

PERSPECTIVES OF PERSONNEL EXTERNAL DOSIMETRY AT STANFORD LINEAR ACCELERATOR CENTER

Bureau International des Poids et Mesures

Interactions of Radiation with Matter

Determination of the activity of radionuclides

Study on Mass Attenuation Coefficients and Optical Properties for Boro-Tellurite Glass doped with Barium

MASS ATTENUATION COEFFICIENT OF LEAD

LECTURE 4 PRINCIPLE OF IMAGE FORMATION KAMARUL AMIN BIN ABDULLAH

Initial Certification

Evaluation of the water equivalence of solid phantoms using gamma ray transmission measurements

Interaction of charged particles and photons with matter

FXA UNIT G485 Module X-Rays. Candidates should be able to : I = I 0 e -μx

Heuijin Lim, Manwoo Lee, Jungyu Yi, Sang Koo Kang, Me Young Kim, Dong Hyeok Jeong

Physics of Radiography

Journal of Applicable Chemistry

Gamma-ray shielding of concretes including magnetite in different rate

Transcription:

Photon Interaction Cross Section of Materials in Heterogeneous Energy Spectrum of Medical Diagnostic X-Ray Beam Radhakrishnan B*, Kurup P G G**, Ramakrishnan G***, Chandralingam S**** *Senior Medical Physicist, Apollo Speciality Hospital, Chennai [Part-time Research Scholar, Jawaharlal Nehru Technological University, Hyderabad] **Chief Medical Physicist, Apollo Speciality Hospital, Chennai ***Associate Professor of Radiation Physics, Govt Medical College, Kottayam, Kerala ****Professor of Physics, Jawaharlal Nehru Technological University, Hyderabad ABSTRACT Experimental determination of mass attenuation coefficients for mono-energetic x-ray and Gaa rays has been reported by various investigators. In this study, an attempt is made to evaluate the mass attenuation coefficients (µ/ρ) and total photon interaction cross section (σ tot ) values of few samples in a heterogeneous x-ray beam from a medical diagnostic x-ray unit, operated at a peak voltage of 85 KVp. Sheets of aluminium, copper, brass and tissue equivalent plastic (virtual water) have been employed. By comparing the experimental values of mass attenuation coefficients (µ/ρ) with those values published in the literature, the prominent energy of interacting photons are estimated to be 36.8 KeV, 48.7 KeV, 47.8 KeV and 35.5 KeV for aluminium, copper, brass and virtual water respectively. From these experimental µ/ρ values, the total photon interaction cross section values (σ tot ) are calculated in the units of barn/atom. These experimental σ tot values are in good agreement with the theoretical values deduced in the literature with percentage deviation of only +0.03% and +0.09% for aluminium and copper (elements) respectively at the prominent energy of interaction. The percentage deviation is -1.12% for Brass (alloy) and +5.77% for virtual water (mixture). The results are discussed. Keywords - Heterogeneous x-ray energy spectrum, mass attenuation coefficient, medical diagnostic x-ray unit, mono-energetic beam, photon interaction cross section 1. Introduction The photon interaction cross sections of various materials due to mono-energetic photons have been studied by various investigators. Hubbell [1,2,3,4] has done extensive work to deduce the photon interaction cross section values for various elements, compounds and mixtures and these values have been published in National Standard Reference Data of National Bureau of Standards, Washington. McMaster, etal [5] had compiled the x-ray properties of various materials. These values have been calculated theoretically based on Klein-Nishana equation by these investigators. R. Nowotny [6] and Akintude A Okunade, etal [7] have developed computer programs to deduce the theoretical values of mass attenuation coefficients of various materials based on the data compiled by Hubbell [8] and Boone, etal [9]. NIST [10] has published a computer program, known as XCOM, to determine the photon interaction cross section of various materials for a wide range of mono-energetic energies. Attempts have been done by Nathuram, etal [11], Kerur, etal [12], Kaginelli, etal [13] and few other investigators to determine the mass attenuation coefficients (µ/ρ) and photon interaction cross sections of few materials experimentally with monoenergetic radiation from radioactive isotopes and characteristic x-ray radiation from variable energy x-ray source and reported that their results are in good agreement with these theoretical values. X-rays, being the 1708 P a g e

most coon diagnostic tool used on human population, experimental determination of the photon interaction cross sections of various materials due to the heterogeneous beam of diagnostic x-ray beams is worth to be explored. This study explores the experimental determination of mass attenuation coefficients (µ/ρ) of few samples in the medical diagnostic x-ray beam and correlation of these data with the published theoretical values. An estimation of prominent energy of the interacting beam is also done correlating the experimental mass attenuation coefficients (µ/ρ) with the theoretical values. 2. Materials & Methods Gary D, etal [14] has described the method of calibration of x-ray units. In the present study, X-ray generators, which have undergone quality assurance tests, are employed. The sample elements used were copper, brass, aluminum and virtual water of varying thicknesses in pure form. The virtual water is composed of 77.5% of Carbon, 10%of Hydrogen, 5% of Oxygen, 3.5% of Nitrogen, 2.5%of Fluorine and 1.5% Calcium. The sample sheets were maintained at a specific distance from the focus and the incident intensity was measured using an ion chamber detector. The transmitted intensity was measured with the same exposure factors. The thicknesses were gradually increased and HVL values were measured. The results were tabulated and plots were obtained. 3. Experimental Set-up Fig.1 Experimental set up of measurements The experimental set up is shown in Fig.1. The measurements were done on a 500mA medical diagnostic x-ray unit (Siemens/Heliophos D). A calibrated ion chamber based detector is placed on the table top so that the detector position is around 55 cm from the focus of the x-ray tube. A stand is fabricated to place the absorbers in such a way that a distance of around 30 cm is maintained between the focus and the absorbers and a minimum of 20 cm is maintained between the absorbers and the detector to avoid scattered photons and secondary electrons reaching the detector. 1709 P a g e

The x-ray unit was initially subjected to a set of quality assurance test procedures before being employed for the present study. RADCHECK PLUS [Fluke/model 06-526], which is capable of integrating the x-ray output, is used for the purpose of measuring the integrated exposure. Exposure factors of 85 KVp and 16 mas were chosen using a collimator field size which covers the detector area sufficiently. Exposure was given without any absorber placed above the detector. The reading shown by the detector was noted as incident intensity (I ). An absorber of known thickness was placed on the stand and the exposure reading was noted again as transmitted intensity (I). The thicknesses of the absorbers were varied and the readings were noted. The above procedure was repeated for other absorbers as well. The sample sheets of Aluminium, Copper and Brass of thicknesses varying from 0.5 to 3.5 at the increment of 0.5 and virtual water of thicknesses varying from 0.5cm to 3.5cm with the increment of 0.5cm were used in this study. Plots of intensity (exposure) Vs thickness are obtained for each sample material and Half value Layer (HVL) for each one, which is the thickness required to reduce the transmitted intensity of radiation to half of its original intensity was obtained from these plots. These plots for all the absorbers used are depicted in Fig (2) and Fig (3). The mass attenuation coefficients (µ/ρ) of each material were calculated by the following equations, I = I e (µ/ρ) (ρx) ---------- (1) and µ = 0.693/HVL ----------- (2) where µ - Linear attenuation coefficient in cm -1 µ/ρ - Mass attenuation coefficient in cm 2 /gm ρx - Thickness of the absorber in gm/cm 2 The total photon interaction cross section has been calculated as σ tot = (µ/ρ)a/n A -----------(3) where A - relative atomic mass of the target element N A - Avagadro s number 4. Results & Discussion Fig. 2 and fig. 3 shows the variation of transmitted intensity/exposure with increase in absorber thickness. The experimental µ/ρ values of the materials due to the heterogeneous energy spectrum from the diagnostic x-ray beam were evaluated using (2) after obtaining HVL thicknesses from these plots for all samples. The average energy of the unfiltered x-ray beam is empirically considered as one-third of the KVp applied and this average energy is expressed in KeV [15,16,17]. Since 85 KVp was the kilovoltage applied to the x-ray tube in this study, the empirical average energy of the unfiltered x-ray beam is approximately 28 KeV. However, due to the elimination of low energy components of the x-ray beam by the inherent and additional filtration provided by the x-ray tube, the practical average energy would be much higher. 1710 P a g e

Exposure (mgy) Intensity (mgy) Radhakrishnan B, Kurup P G G, Ramakrishnan G, Chandralingam S/ International Journal Thickness Vs Intensity 3 2.5 2 1.5 1 0.5 Alumini um Copper Brass 0 0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.5 5.5 Absorber thickness fig 2. Thickness Vs intensity plot for Aluminium, Copper and Brass Intensity Vs Exposure for Tissue equivalent plastic (Virtual water) 3 2.5 2 1.5 7 85 KVp 1 0.5 70 KVp 0 0 0.5 1 1.5 2 2.5 3 3.5 4 Thickness of absorber (cm) fig 3. Thickness Vs intensity plot for virtual water 1711 P a g e

The theoretical mass attenuation coefficient (µ/ρ) values of the samples used were determined using the XMudat computer program [6] developed by R. Nowotny based on the data published by Hubbell [8] and Boone [9]. The variation of µ/ρ for a wide range of energy ranging from 1 KeV to 10 MeV for monoenergetic photons for these absorbers as derived from this XMudat program is shown in Fig (4). These values were compared with the experimental mass attenuation coefficient (µ/ρ) values. The results indicate that the prominent energy of the interacting x-ray photons reaching the detector after suffering attenuation is different for different materials. This prominent energy predictably depends largely on the atomic number of the materials because of the predominance of photo electric effect in the diagnostic x-ray energy range and the photo electric effect is directly proportional to Z 3[17]. In the case of aluminium (Z=13) and virtual water (Z=7.4), the prominent energy of interaction was found to be 36.8 KeV and 35.5 KeV respectively which indicate that the average energy of the filtered x-ray beam is not much affected by the attenuation due to these two samples. However, for copper (Z=29) and brass (Z=30), the prominent energy of interaction turned out to be 48.7 KeV and 47.8 KeV respectively indicating the removal of some more low energy components of the incident x-ray beam. These results are shown in the Table (1). This table also shows the percentage of deviation of experimental mass attenuation coefficient (µ/ρ) values from the theoretical values had we considered the average energy of the unfiltered x-ray beam, ie, 28 KeV for all these interactions. The percentage of deviation ranges from 17% for aluminium to 37.4% for copper and 34.46% for brass while for virtual water the percentage of deviation is 24.47%. fig. 4 Theoretical variation of Photon Mass attenuation coefficient with energy from 1 Kev to 10 MeV (Ref: Xmudat program based on Hubbell & Boone) 1712 P a g e

Table 1: Mass Attenuation coefficients (Experimental & theoretical): Absorber material µ/ρ (experimental) (cm 2 /gm) Aluminium 0.685 Copper 2.813 Brass 3.066 Tissue equivalent Plastic (virtual water) 0.286 * Energy of the x- ray spectrum close to the µ/ρ (experimental) values 36.8 KeV 48.7 KeV 47.8 KeV 35.5 KeV µ/ρ (theoretical) (cm 2 /gm) for empirical average energy (28 KeV) % Deviation of µ/ρ values at empirical average energy (28 KeV) 0.568 17.00 4.490 37.40 4.678 34.46 0.378 24.47 Xmudat program based on Hubbell & Boone The total photon interaction cross-sections (σ tot ) of these samples were calculated using eq (3) from the experimental µ/ρ values. The photoelectric effect is the predominant interaction mode in the diagnostic x-ray energy range used in the study. σ tot values thus calculated were compared against the theoretical σ tot values which were estimated through the XCOM computer program developed by the National Institute of Standards and Technology (NIST) [10], United States. The σ tot values including a small cross section of Rayleigh scattering obtained through this program agree closely with the experimental σ tot values in the case of aluminium and copper (elements) with a percentage deviation of only +0.03% and +0.09%. The percentage deviation is -1.12% for brass (alloy) and it is +5.77% for virtual water (mixture). These results are shown in the Table (2). Table 2: Total photon interaction cross-section values (Experimemtal & theoretical): Absorber material Aluminium Copper σ tot (experimental) (barns/atom) 30.68 296.78 σ tot (theoretical) (barns/atom) % Deviation at prominent energy σ tot (theoretical) (barns/atom) at empirical average energy of x-ray (28 KeV) % Deviation at empirical average energy (28 KeV) 30.67 +0.03 25.43 20.70 296.50 +0.09 473.80 37.40 Brass 323.50 327.17-1.12 493.63 34.46 Tissue equivalent Plastic (virtual water) 5.60 5.29 +5.77 7.40 24.46 * XCOM program developed by NIST based Hubbell & Sieltzer S M) 1713 P a g e

However, σ tot values estimated based on the theoretical µ/ρ values at the empirical unfiltered average x-ray energy of 28 KeV deviates from 20.7% for aluminium to 37.4% for copper. For virtual water, the percentage of deviation is 24.46% and for brass it is 34.46%. These deviations demonstrate that the practical average energy of the x-ray beam is much more than 28 KeV due to the inherent and additional filtration of the x-ray tube. The prominent energy of interaction for each set of samples is still higher than this filtered average energy for the reasons described earlier. 5. Conclusion The comparison of experimental values of mass attenuation coefficients (µ/ρ) and total photon interaction cross section (σ tot ) due to diagnostic x-ray energy spectrum of heterogeneous nature reveals that the experimental and theoretical values for the empirical average x-ray energy do not agree accurately with each other. This is due to the fact that the actual average energy of the filtered x-ray beam is more than the empirical average energy. The prominent energy of the photons which interacted with these absorbers could be deduced from the experimental mass attenuation coefficients (µ/ρ) which is higher than the predicted average energy of the heterogeneous beam of x-ray. Therefore the radiation shielding requirements for radiation safety considerations and the estimated absorbed dose by various anatomical structures during diagnostic procedures based on the empirical average energy of a heterogeneous beam of x-ray spectrum will not be accurate and knowledge of the prominent energy of interaction among the entire spectrum of the beam will give more accurate estimation. References 1. Hubbell, J.H. and Berger, M.J., Sections 4.1 and 4.2 in Jaeger, R.G. (ed.): Engineering Compendium on Radiation Shielding (IAEA, Vienna), Vol. 1, Ch. 4, pp. 167-202, Springer, Berlin (1968). 2. Hubbell, J.H., Photon Cross Sections, Attenuation Coefficients and Energy Absorption Coefficients from 10 kev to 100 GeV, Natl. Stand. Ref. Data Ser. 29 (1969). 3. Hubbell, J. H., Review and history of photon cross section calculations, Physics in Medicine and Biology, 2006, vol 51, R45-R262 4. Hubbell, J.H., Photon mass attenuation and energy absorption coefficients from 1Kev to 20 Mev, Int. Journal of Appl. Radiat. Isot. 1982, vol 33, p1269-90 5. McMaster, W.H., Del Grande, N.K.,Mallett, J.H, and Hubbell, J.H., Compilation of X-ray Cross Sections, Lawrence Livermore Lab., Report UCRL-50174, (1969) 6. R. Nowotny, Photon attenuation data on PC, XMuDat: Version 1.0.1 of August 1998, IAEA-NDS- 195, Docementation series of the Nuclear Data Services of the International Atomic Energy Agency 7. Akintude A Okunade etal, Parameters and computer software for the evaluation of mass attenuation and mass energy absorption coefficients for body tissues and substitutes, Journal of Medical Physics, Vol32, p 124-132 8. Hubbell J H, Seiltzer S M, NISTIR 5632,1995 9. Boone A M, Chavez A E, Medical Physics, Vol 23, 1996 10. http://physics.nist.gov/ Physics reference data/xcom 1714 P a g e

11. R. Nathuram, I. S. Sundar Rao and M. K. Mehta, Photoelectric cross sections for 6-20 Kev photons in beryllium, carbon, magnesium, aluminium, silicon, copper, silver and lead, Phys. Rev. A 37, 1988, p4978-4981 12. Kerur B R, Manjula V T, Lagare M T, Anil Kumar S, Mass Attenuation coefficient of saccharides for x-rays in the energy range from8 KeV 32 KeV Radiation Measurements, 2009, vol 44, p63-67 13. Kaginelli, S B, Rajeshwari T, Shravanabasappa, Kerur B R, Anil S Kumar, Effective atomic number and electron density of dosimetric material, Journal of Medical Physics, Vol 34, No.3, 2009, p176-179 14. Gary D, Fullertone and Royce Gragg, Calibration of medium energy x-ray units (20 to 300 KV), (Handbook of Medical Physics, vol I, CRC Press, 1984), p103-117 15. Marco Aurelio de Sousa Lacerda etal, The methodology for evaluating half-value layer and its influence on the diagnostic radiology, Radiologia Brasileira, vol 40, 2007 16. ICRU report 17, Radiation Dosimetry : X-rays Generated at potentials of 5 to 150KV (1970) 17. Khan, F. M, The Physics of Radiotherapy, (Lippincott Williams & Willins, Fourth Edition, 2010) 1715 P a g e