Complete Materials Deformulation Using TGA-IR

Similar documents
Introduction to Fourier Transform Infrared Spectroscopy

Introduction to Fourier Transform Infrared Spectroscopy

The Raman Spectroscopy of Graphene and the Determination of Layer Thickness

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

MetWorks Metabolite Identification Software

XPS Surface Characterization of Disposable Laboratory Gloves and the Transfer of Glove Components to Other Surfaces

Fast, Effective XPS Point Analysis of Metal Components

Hyphenated Spectroscopy Techniques

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

for XPS surface analysis

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

of mass spectrometry

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

HYPHENATED TECHNOLOGY GUIDE

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Evolved gas analysis by simultaneous thermogravimetric differential thermal analysis-fourier transformation infrared spectroscopy (TG-DTA-FTIR)

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

HYPHENATED TECHNOLOGY GUIDE

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Drugs of Abuse by Direct Sample Probe Using Single Quadrupole GC/MS

Thermo Scientific K-Alpha + XPS Spectrometer. Fast, powerful and accessible chemical analysis for surface and thin film characterization

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

Thermo Scientific Pesticide Explorer Collection. Start-to-finish. workflows for pesticide analysis

Bath Salts and Cannabinoids Analyzed by GC-IR

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Elemental Analysis: CHNS/O characterization of polymers and plastics

HILIC Method Development in a Few Simple Steps

Probabilistic Approach for Material Verification and Identification in Pharmaceutical Applications

Performance and Control of the Agilent Nano Indenter DCM

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Collected Applications Thermal Analysis EVOLVED GAS ANALYSIS

Finnigan TC/EA. High Temperature Conversion Elemental Analyzer

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Thermo Scientific FLASH HT Plus Elemental Analyzer for IRMS

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

This page intentionally left blank

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Monitoring Protein PEGylation with Ion Exchange Chromatography

Analysis of a Verapamil Microsomal Incubation using Metabolite ID and Mass Frontier TM

How to Use Kinetic Modeling Analysis To Predict Profile Part Failure

Thermogravimetric Analysis Advanced Techniques for Better Materials Characterisation

DETERMINATION OF THE DECOMPOSITION PRODUCTS OF CALCIUM OXALATE USING THERMAL GRAVIMETRY AND INFRARED SPECTROMETRY

Verification of Pharmaceutical Raw Materials Using the Spectrum Two N FT-NIR Spectrometer

High Pressure DSC Differential Scanning Calorimeter

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

TGA Thermal Gravimetric Analysis

IN QUALITATIVE ANALYSIS,

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Yin and yang in chemistry education: the complementary nature of FTIR and NMR spectroscopies

DSC AND TG/DTA AS PROBLEM-SOLVING TOOLS: CHARACTERIZATION OF PHARMACEUTICAL COMPOUNDS

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

TA Instruments NEW AR-G2 INTRODUCING THE WORLD S FIRST MAGNETIC BEARING RHEOMETER! The World Leader in Thermal Analysis and Rheology Systems

Thermo Scientific GasBench II

Modulated DSC Paper #8 Use Of Quasi-isothermal Mode for Improved Understanding of Structure Change

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

Educational experiment package Volume 1. Molecular spectroscopy

Prima PRO Process Mass Spectrometer

High-Resolution Accurate-Mass (HRAM) Phthalate Screening using Direct Analysis in Real Time (DART) Ambient Ionization

Thermal Methods of Analysis Theory, General Techniques and Applications. Prof. Tarek A. Fayed

APPLICATIONS OF THERMAL ANALYSIS IN POLYMER AND COMPOSITES CHARACTERIZATION. Wei Xie TA Instruments

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES

TOPEM the new advanced multi-frequency TMDSC technique

Battle the heat in your lab

ARGUS VI. Static Vacuum Mass Spectrometer. Static Vacuum ARGUS VI. Multicollection Low Volume Precision

Vinyl Ester Resins. Process description. General composition. Cure temperature. Main applications. Application Guide. Thermoset Chemicals

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Keysight Technologies Measuring Substrate-Independent Young s Modulus of Low-k Films by Instrumented Indentation. Application Note

EM 27/SUN Series. Innovation with Integrity. For Atmospheric Measurements FT-IR

Keysight Technologies Measurement Uncertainty of VNA Based TDR/TDT Measurement. Application Note

Quick Guide QUICK GUIDE. Activity 1: Determine the Reaction Rate in the Presence or Absence of an Enzyme

Human C-reactive protein ELISA using Sword Peroxidase Reagents

High Power Factors and Contaminants in Transformer Oil

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

It is the purpose of this paper to demonstrate that the detection limit is directly related to the sample size, for both TG and TG/FTIR.

Total analysis with DSC, TMA and TGA-EGA

Positive and Nondestructive Identification of Acrylic-Based Coatings

Thermal Analysis Premium

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Reference Standards Page 156 For calibrating your spectrometer. PIKECalc Page 159 For FTIR sampling computations

Keysight Technologies Instrumented Indentation Testing with the Keysight Nano Indenter G200. Application Note

Keysight Technologies Oxygen-Free High-Resolution Electrochemical SPM. Application Note

viridis Columns Bringing a New Dimension to SFC Combining state-of-the-art media manufacturing with industry leading column technology, Viridis

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

IR-LC Deposition and Detection System. Polymer Deformulation and Additive Analysis by a Single GPC-IR Run

Transcription:

Application Note: 51694 Complete Materials Deformulation Using TGA-IR Michael Bradley, Thermo Fisher Scientific, Madison, WI, USA Key Words Feedstock Supplies Finished Products Multi-component Search OMNIC Series Software OMNIC Specta Software Polymers Rubbers TGA-IR Thermal Gravimetric Analysis Vapor Phase Spectral Library Introduction Key applications in analytical laboratories often involve failure analysis and competitive deformulation. For example, black rubber gaskets from two suppliers stating the same production process behave differently one seals, the other fails or two expanded polymer foams exhibit very different mechanical behavior. These situations call for deformulation, and a qualitative and quantitative comparison of the results, to determine the source of the differences. Thermal Gravimetric Analysis (TGA) is a powerful deformulation tool common in material analysis. Basically, a small amount of sample placed on a pan is inserted into a furnace, with the pan hanging on a sensitive microbalance. A slow temperature ramp is applied with the sample bathed in an inert purge gas. The increasing temperature forces various components of the material into the vapor phase, resulting in weight loss. TGA provides excellent quantitative information about weight losses as a function of temperature, but does not identify the vapors being driven off. Coupling the TGA to a Fourier Transform infrared (FT-IR) spectrometer adds considerable utility to the method. The vapors released from the material pass through a gas cell inserted in the FT-IR, where the spectrum is obtained. Spectral searching provides the last link identification of the vapors. The weight loss data from the TGA, the rapid response and high sensitivity of the spectrometer, and spectral searching combine to provide potent analytical insights into the chemistry of the sample. Traditionally, the one difficulty with the infrared portion of the analysis has been the simultaneous evolution of gases. The TGA does not discriminate between simple and complex gas evolutions, but the instantaneous IR spectra are summations of all the gases being driven off at that time. This complexity normally requires some skill to untangle. The need to use multiple search regions to selectively extract individual components and the generally unfamiliar look of gas phase spectra to many investigators contribute to the potential for missed critical answers. As shown below, this difficulty is largely surmounted with the powerful Thermo Scientific OMNIC Specta spectroscopy software, which performs multi-component searches at a single click. We will examine three different scenarios, based on industrial situations with finished products and feedstock supplies. The first involves an epoxy, examining the components of the mixture prior to and during setting. The second situation compares black rubber samples similar to those used in gasket manufacturing. The third study involves two, blown polymer foam materials. This range of experiments illustrates both the power and the flexibility of TGA-IR. Experimental An in-sample compartment TGA accessory equipped with a double pass gas cell heated to 220 C was coupled to the TGA with a heated transfer line. The sampling accessory was inserted into a Thermo Scientific Nicolet iz 10 sampling module coupled to the Nicolet is 10 FT-IR spectrometer; shown in Figure 1. The FT-IR was driven using OMNIC spectroscopy software and OMNIC Series time-base software. In all cases, data were collected at 4 cm -1 resolution with background and Gram-Schmidt (GS) basis vector collection occurring just before data collection. The GS display is, in short, a simple way to visualize the total change in the spectrum from the initial time. Multi-component searching was performed by sending the OMNIC data to OMNIC Specta. Figure 1 The thermal analysis was performed with a TA Instruments Q5000 TGA with autosampler. Pre-tared platinum pans were used to contain the samples, and the samples were between 10 and 50 mg in size. The same computer was used to control both instruments, and a trigger cable allowed simultaneous initiation of data collection. Various temperature ramps were applied to the materials, as noted below. The TGA data was imported to OMNIC via the Universal Analysis (UA) software (TA Instruments). Similar export of experimental profiles, including the Gram-Schmidt profile, to UA is available.

Results Epoxy A single drop of a freshly mixed two-part epoxy (normal 12 hour cure time) was placed in a TGA sample pan. The temperature was ramped from ambient to 500 C at the rate of 15 C/min. Figure 2 shows the FT-IR data in OMNIC Series. The upper trace reveals the evolution of the overall IR signal in time (the GS plot), while the lower trace is the IR spectrum present at the time point indicated by the scroll bar in the upper window. The small signals in the GS plot around 11-12 minutes (190 210 C) result from CO 2 and water bubbles evolving from the sample. The larger peak rising after 15 minutes (about 250 C) is associated with many features in the infrared. The single spectrum shown in Figure 2 (bottom) demonstrates the multiple gas evolution discussed above. A simple search on this spectrum indeed extracts one of the components, but leaves the user in need of multiple subtraction steps. Unfortunately, gas phase spectra are very sensitive to environmental conditions (temperature, pressure, mixtures), which affect both peak shapes and line intensities, making spectral subtraction based on library spectra a difficult proposition. Figure 3 shows a 3-dimensional view (time-wavenumber-intensity) of the entire file in the OMNIC Series software the overlapping emissions and complex time behavior are quite apparent. Figure 3 Figure 2 OMNIC Specta permits these complexities to be unraveled. Choosing the HR (High Resolution) Nicolet TGA Vapor Phase library and executing the multicomponent search in OMNIC Specta provides the results shown in Figure 4 for multiple time points in the OMNIC Series file. The visual agreement between the original spectrum and the composite is outstanding in all cases, giving the user high confidence in the results. Further, each of these multi-component searches required less than one minute. The slight differences seen in the composite and the original are due to the same factors discussed above regarding spectral subtraction spectral changes due to environmental variations between the experimental and library collection conditions. OMNIC Specta s multi-component search algorithm identifies three unique compounds in an epoxy in less than a minute.

a b The first gases to emerge are clearly from the unreacted base materials of the epoxy and hardener, including the common epoxy ingredient bisphenol-a. As the temperature rises, ester peaks appear, indicative of the accelerated curing. Late in time, methane evolution along with low levels of bisphenol-a are evident as the epoxy breaks down. Broadening the idea of using a TGA to drive more general reactions, the ability to ramp and hold, or to change the purge gas during a run, gives great flexibility. Indeed, elevated temperature often accelerates reactions, and the purge gas could even be a reactant. Black Rubber Samples Pieces of raw rubbers (one light, one dark) used in producing high performance black rubbers were analyzed using a two-step TGA ramp. Figure 5 shows GS profiles for the two rubber samples, along with the derivative weight loss and temperature ramp information imported from the TGA. The GS curves clearly point to differences in the formulations. Below 200 C, neither material shows major gas evolution, but above 350 C, strong signals appear. The derivative weight loss and the GS trace are also highly correlated in time, which allows the temperature profile to be coordinated with the IR data confidently. c Figure 5 Figure 4

Comparing spectra from various time intervals in the two samples shows both similarities and differences. The high carbon-filled (dark colored) rubber releases many of the same gases as the light rubber, but at a much lower temperature. Figure 6 shows spectra from the two samples at different times, and the same basic materials are present in both but separated in temperature of release. Figure 7, in contrast, shows the 40 minute time points (same temperature point) for both samples. The dark rubber shows CO 2, CO and water while the light rubber is showing the strong release of many components. The multiple component gas phase spectra were exported to OMNIC Specta. Preliminary searches over the entire spectrum showed the expected hits CO, CO 2 and water plus one rubber component. However, we focused the multi-component search algorithm on a narrow range of the spectrum where there were small features. The narrow range, multi-component search result, shown in Figure 8, reveals four materials, included butenes, methanol and methyl-ethyl ketone. Figure 8 Figure 6 In normal infrared spectroscopy, water and CO 2 often originate from the atmosphere, and produce undesired signals in the spectrum, which is why atmospheric corrections and instrument purging are often used. However, in the TGA experiment, both water and CO 2 provide information, as they are generated by the sample during heating. Thus, software corrections should be avoided, while purge, if applied, must be maintained at a constant level. Normally atmospheric water and carbon dioxide are to be avoided in FT-IR. With TGA they both provide useful information as the sample is heated. Figure 7 The two rubbers are apparently similar in underlying chemistry. However, the large difference in the temperature of release shows the formulations or treatments applied to the samples cause differences. In a high performance application, it may well be that the lower temperature of evolution causes a failure; at the least, it would indicate a potential target for further analytical investigation.

Blown Polymeric Foam Decorative or functional covers for many manufactured items are often produced in a blowing process. A two (or more) component mixture of a base polymer and a blowing agent are inserted into a mold. The mold is then subjected to heating, which melts the base polymer. At a higher temperature, the blowing agent releases gas, usually with a sharp onset, which quickly expands the material to fill the mold. In the present study, the mechanical properties of two plastics made with the same base polymer, but different blowing agents, were similar. Either would have been useable mechanically, but one of the materials emitted unacceptable levels of noxious vapors. Pieces of the blown polymers and pellets of the two blowing agents and the base polymer were analyzed by TGA-IR; only a critical subset of this data is shown here. The temperature ramp ended at 200 C, as the thermal decomposition of the base polymer was not of interest only the behavior of the residual blowing agents or trapped gases. The entire piece melted around 190 C, releasing gases caught in the matrix, but there was little or no decomposition of the base polymer at this temperature. GS profiles for the blowing agents and a cover blown with Agent 2 are shown in Figure 9. Agent 1 shows a single, rapid-onset, gas emission (which is almost explosive in nature). Agent 2 shows two transitions, and blown cover material shows the same transitions. The 3-D plot in Figure 10 for Agent 2 shows only CO 2 in either of the two emissions. In Figure 11, the same region is shown for Agent 1. The onset is rapid and only occurs at one temperature, but the signals show more than one gas is coming off. The OMNIC Specta multi-component search result, shown in Figure 12, reveals that the second peak is isocyanate. A very small shoulder on the low wavenumber side of the isocyanate is due to CO. The OMNIC Series results for the base polymer alone (not shown here) reveal none of these signals, so the problem lies with the choice of blowing agent. Figure 9 Figure 10 The additional information provided by FT-IR enables the analyst to identify which Agent is proper to use in blown polymeric foam. Figure 11

Both blowing agents work very well so far as the mechanical properties of the final material are concerned. However, the additional gases evolved permitted the manufacturer to prefer Agent 2. Note that the weight loss profiles alone would be insufficient to discriminate between the two materials. The additional information provided by the infrared analysis provides the needed differentiation. Figure 12 Conclusions TGA-IR is a general purpose, powerful technique for the deformulation of materials in many industries, including plastics, resins, rubbers, coatings and pharmaceuticals. The melding of the two techniques more than doubles the analytical power for these applications. With the addition of OMNIC Specta multi-component searching, the identification of the overlapped vapor emissions no longer requires spectral interpretation skills, greatly reducing the time per analysis for experienced and inexperienced users alike. Consistency in the analysis is also assured. OMNIC Specta is shown to greatly improve an analyst s to interpret overlapped gas phase spectra. The key conclusion from the examples shown here is that TGA-IR coupled with OMNIC Specta provides actionable results. In essence, the operations personnel can make process modification decisions based on a solid foundation. In addition to these offices, Thermo Fisher Scientific maintains a network of representative organizations throughout the world. Africa +43 1 333 5034 127 Australia +61 2 8844 9500 Austria +43 1 333 50340 Belgium +32 2 482 30 30 Canada +1 800 530 8447 China +86 10 8419 3588 Denmark +45 70 23 62 60 Europe-Other +43 1 333 5034 127 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9434 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 608 276 5659 Middle East +43 1 333 5034 127 Netherlands +31 76 579 55 55 South Africa +27 11 570 1840 Spain +34 914 845 965 Sweden/Norway/ Finland +46 8 556 468 00 Switzerland +41 61 48784 00 UK +44 1442 233555 USA +1 800 532 4752 www.thermo.com 2008 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Electron Scientific Instruments LLC, Madison, WI USA is ISO Certified. AN51694_E 08/08M Part of Thermo Fisher Scientific