Sep-Pak XPoSure Aldehyde Sampler

Similar documents
ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

Determination of Carbonyl Compounds In Water by Dinitrophenylhydrazine Derivatization and HPLC/UV*

[ a ppl ic at ion no t e ]

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

METHOD 0100 SAMPLING FOR FORMALDEHYDE AND OTHER CARBONYL COMPOUNDS IN INDOOR AIR

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Determination of Polymer Modifier in Asphalt

CORESTA Recommended Method No. 86

Sulfotepp impurities in Chlorpyrifos EC formulations

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

INTERNATIONAL CHEMISTRY TESTING. 258 Main Street- Suite 311- Milford, MA Tel: Fax:

Determination of Carbonyl Compounds in Workplace Air

Determination of trace anions in concentrated hydrofluoric acid

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

12 Nicarbazin Nicarbazin (4,4 -dinitro carbanilid (DNC) and 2-hydroxy-4,6-dimethyl pyrimidine (HDP))

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

PYRIPROXYFEN TECHNICAL

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

EPA IP-6 METHOD UPDATE. Year DETERMINATION OF FORMALDEHYDE AND OTHER ALDEHYDES IN INDOOR AIR

Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

TECHNICAL BRODIFACOUM

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION

FORMALDEHYDE by GC 2541

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

Basic Principles for Purification Using Supercritical Fluid Chromatography

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

HIPPURIC ACID and o-m-p- METHYLHIPPURIC ACIDS IN URINE BY UV CODE Z06010

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin

METHYLTIN CHLORIDES 5526

Spectrophotometric Determination of Ferrocyanide in Effluents

Introduction to IH Analytical Chemistry

for Acclaim Mixed-Mode HILIC-1 Column

VINYL CHLORIDE. Recommended by NIOSH, partially evaluated by OSHA Laboratory.

STANDARD OPERATING PROCEDURES

Micro Volume QuEChERS kit

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

METHYL ETHYL KETONE PEROXIDE 3508

Zearalenone Plate Kit PN 53018B

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

SOIL ORGANIC CONTENT USING UV-VIS METHOD

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

LC Technical Information

Certified Reference Material - Certificate of Analysis

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

High Performance Liquid Chromatography

GB/T Translated English of Chinese Standard: GB/T

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

Chemistry Gas Chromatography: Separation of Volatile Organics

Peptide Isolation Using the Prep 150 LC System

E.Z.N.A. MicroElute Clean-up Kits Table of Contents

Cadion Method Method to 0.30 mg/l Cd TNTplus 852

Certified Reference Material - Certificate of Analysis

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Functional Genomics Research Stream. Lecture: February 17, 2009 Masses, Volumes, Solutions & Dilutions

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

IonPac Trace Cation Concentrator

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

ACETONE IN URINE BY UV CODE Z42010

637. Thiamethoxam. HPLC method

RayBio Rat TNF-alpha ELISA Kit (For Lysates)

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

HYDROCARBONS, AROMATIC 1501

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit

PAR Method Method to 2.0 mg/l Pb TNTplus 850

Method 630. The Determination of Dithiocarbamate Pesticides in Municipal and Industrial Wastewater

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Test Method: CPSC-CH-E

METHOD 7199 DETERMINATION OF HEXAVALENT CHROMIUM IN DRINKING WATER, GROUNDWATER AND INDUSTRIAL WASTEWATER EFFLUENTS BY ION CHROMATOGRAPHY

Determination of Orthophosphate Ion

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

RayBio Human IDUA ELISA Kit

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Chromazurol S Method Method to 0.50 mg/l Al TNTplus 848

1. Preliminary qualitative analysis of unknown substances (liquid or solid).

The analysis of organic acid content of additives, premix, feed, and water.

Oxygen Demand, Chemical

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

H 3 CO H 3 CO S CH 3

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Transcription:

[ CARE AND AND USE USE MANUAL MANUAL ] ] CONTENTS I. INTRODUCTION a. Sep-Pak DNPH-Silica Cartridge Description II. USING THE SEP-PAK XPOSURE ALDEHYDE SAMPLER a. Theory of Operation b. Preventing Contamination c. Collecting the Sample d. Eluting the Derivatives from the Sampler I. INTRODUCTION Waters Sep-Pak XPoSure Aldehyde Samplers are convenient, reproducible sampling devices for quantifying formaldehyde concentrations in the workplace and indoor air within a range of 0.001 to 5 parts per million (ppmv). III. IV. ANALYZING THE DNPH DERIVATIVES a. Theory of Operation b. Preventing Contamination c. Collecting the Sample d. Calculating Results APPLICATION EXAMPLES a. Formaldehyde in Laboratory Air - STEL b. Glutaraldehyde in Laboratory Air STEL V. STORAGE AND DISPOSAL OF THE SAMPLERS a. Storing Unused Samplers b. Storing Exposed Samplers c. Disposing of used cartridges VI. TROUBLESHOOTING VII. REFERENCES AND BIBLIOGRAPHY VIII. ORDERING INFORMATION IX. APPENDICES a. Appendix A: Measuring Acetontrile Purity b. Appendix B: Synthesizing the DNPH-Derivative Standards c. Appendix C: Measuring Breakthrough d. Appendix D: Useful Conversion Factors d.1. Carbonyl to Hydrazone Conversion Factors d.2. Equation for converting µg/l to ppmv d.3. Conversion Factors: µg/l to ppmv

a. Description and Features s contain acidified 2,4-Dinitrophenylhydrazine(DNPH)-coated silica, packed in Waters Sep-Pak Plus cartridges. The samplers are constructed from high-purity and highdensity polyethylene components, triaxially-compressed packed beds and Luer fittings equipped with end caps and plugs. The samplers are designed for flow rates of up to 1.5 L/min with typical personal pumps. The gold-colored aluminum compression ring on the allows for easy identification. II. USING THE SEP-PAK XPOSURE ALDEHYDE SAMPLER a. Theory of Operation s trap aldehydes in air by reacting them with acidified 2,4-dinitrophenylhydrazine(DNPH), forming stable hydrazone derivatives. The derivitization reaction, as shown in Figure 2, takes place during sample collection. The derivatives are later eluted and analyzed using HPLC. Luer Connector Aluminum Compression Ring Polyethylene Filter DNPH-Silica Polyethylene Filter Figure 2. Derivitization Reaction. b. Preventing Contamination Contamination is most likely to occur during sample collection. Before eluting the derivatives, clean all glassware by rinsing with acetonitrile and heating to 60 C in a vacuum oven for at least 30 minutes. Eluting the samples in a nitrogen-purged glove box further reduces the risk of contamination. Luer Connector Figure 1. Cutaway View. Table 1: Physical and Chemical Properties Hold-up Volume 0.7 ml Collection Efficiency >95% for sampling rates up to 1.5 L/min The acetonitrile you use to elute the DNPH derivatives can also be a source of contamination. Even HPLC-grade acetonitrile may have unacceptable levels of carbonyl contaminants and should be stored in a carbonyl-free environment. A concentration of 10 µg/l of any aldehyde or ketone contaminant will add 0.1 µg to the blank values determined for the DNPH-derivatives per cartridge. Follow the procedure in Appendix A, Measuring Acetonitrile Purity, to pre-qualify your acetonitrile. Capacity a Quantity of DNPH Silica Approximately 70 µg formaldehyde 0.35 g/sampler (~1 mg DNPH) Sampling Temperature b 10 C to 100 C Dimensions 4.3 cm total length 2.0 cm o.d. at widest point 1.0 cm i.d. 0.9 cm bed length a. Based on 50% consumption of DNPH. b Evaluate sampler performance for individual high-temperature method. 2

c. Collecting the Sample Caution: Beware of unintentional exposure of the samplers and eluted samples to aldehyde and ketone sources. Laboratory air often holds high concentrations of acetone. Labeling inks and adhesives as well as packaging containers (including vials with plastic caps) are all possible sources of contamination. Sampling methods have been developed and validated for both 15 minute short-term exposure limit (STEL), and in 8 hour personal exposure limit (PEL) measurements following the National Institute for Occupational Safety and Health (NIOSH) guidelines 1,2. XPoSure Aldehyde Samplers have been tested under controlled laboratory conditions. Table 2 lists the equipment needed to collect air samples using XPoSure Aldehyde Samplers. The recommended measurement range for formaldehyde is: STEL: 0.022 to 2 ppmv PEL: 0.01 to 1 ppmv Table 2: Sample Collection Equipment Suggested Personal Pump Specifications Operating Range 100 to 1,500 ml/min Compensation Range 1,500 ml/min 0 to 20 inches water Flow Control ±5% set point constant flow Flow Indicatora Built-in flow indicatora a. A flow calibrator may also be required. The background levels of aldehydes and ketones in the sampler determine the sensitivity of the method. The volume of air passed through the sampler must be large enough for the quantity of DNPH derivatives formed to be several times greater than the background level. The United States Environmental Protection Agency (US EPA) recommends that this level be at least 10 times the background level. 3 Table 3 lists the sampler background specifications. Sample Collection STEL (0.02 to 2.0 ppmv) To collect the STEL air sample: 1. 1Calibrate the sampling pump with a representative sampler in line. Set the flow to 1.5 L/min. Figure 3 shows the flow rate through a sampler versus applied vacuum. Once calibrated, remove and store this representative sampler for future calibrations. 2. Take a fresh sample from its pouch. Remove and save the end cap and plugs. 3. Connect the sampler to a pump with flexible plastic tubing. The sampler is bidirectional (flow can be in either direction). 4. Draw air for 15 minutes, yielding a sample volume of 22.5 liters. 5. Reseal the sampler with its end cap and plug. 6. Store the sampler in the pouch provided with appropriate identification. Seal the pouch by folding the edge over twice and stapling it shut. Avoid exposing the samplers to heat. A 22.5 L air sample is sufficient for quantifying formaldehyde in the range of 0.02 to 2 ppmv. Formaldehyde concentrations lower than 0.02 ppmv in air will require longer sampling times and a larger air sample. Conversely, formaldehyde concentrations that exceed 2.0 ppmv will require shorter sampling times or reduced sampling flow rates in order to avoid overloading the sampler and obtaining nonlinear results. Note: The maximum recommended sampler capacity is 2.3 µmoles total carbonyl species. This calculates to 50% of the DNPH consumed. Contaminated air may contain significant concentrations of other aldehydes and ketones and the total may exceed the capacity of the sampling device. Follow the procedure in Appendix C, Measuring Breakthrough for more information. Table 3: Sampler Background Specifications Compound µg DNPH Derivatives per Sampler µg Carbonyl Compounds per Sampler Formaldehyde <0.45 <0.06 Adetaldehyde <0.75 <0.15 Acetone <1.5 <0.38 Othersa <0.75 - a. Individually, as acetone-dnph. 3

2.0 1.8 1.6 1.4 1.2 Note: The maximum recommended sampler capacity is 2.3 µmoles total carbonyl species. This calculates to 50% of the DNPH consumed. Contaminated air may contain significant concentrations of other aldehydes and ketones and the total may exceed the capacity of the sampling device. Follow the procedure in Appendix C, Measuring Breakthrough for more information. 1.0 0.8 0.6 0.4 0.2 5 15 Vacuum (inches water) Figure 3: Typical Sampler Backpressure Profile. 25 30 d. Eluting the Derivatives from the Sampler To elute the derivatives from the sampler: 1. Remove the sampler from the stapled pouch. 2. Elute the DNPH derivatives from the sampler with pre-qualified acetonitrile directly into a 10 ml volumetric flask. Use a flow rate of <3 ml/min. Higher flow rates (>3 ml/min) can result in reduced recovery. 3. Cap the volumetric flask and mix by inverting it several times. Sample Collection PEL (0.01 to 1.0 ppmv) To collect the PEL air sample: 1. Calibrate the sampling pump with a representative sampler in line. Set the flow to 100 ml/min. Figure 3 shows the flow rate through a sampler versus applied vacuum. Once calibrated, remove and store this representative sampler for future calibrations. 2. Take a fresh sample from its pouch. Remove and save the end cap and plugs. 3. Connect the sampler to a pump with flexible plastic tubing. The sampler is bidirectional (flow can be in either direction). 4. Draw air for 8 hours, yielding a sample volume of 48 liters. 5. Reseal the sampler with its end cap and plug. 6. Store the sampler in the pouch provided with appropriate identification. Seal the pouch by folding the edge over twice and stapling it shut. Avoid exposing the samplers to heat. A 48 L air sample is sufficient for quantifying formaldehyde in the range of 0.01 to 1 ppmv. Formaldehyde concentrations lower than 0.01 ppmv in air will require longer sampling times and a larger air sample. Conversely, formaldehyde concentrations that exceed 1.0 ppmv will require shorter sampling times or reduced sampling flow rates in order to avoid overloading the sampler and obtaining nonlinear results. 4. Analyze the eluate using HPLC. Note: Since background levels may change during storage, always compare samples to a blank sampler from the same lot, stored under the same conditions. III. ANALYZING THE DNPH DERIVATIVES a. Operating Guidelines To ensure success in your HPLC analysis: Use a pre-column filter between the injector and column. Use HPLC-grade water and HPLC-grade acetonitrile. Degas the mobile phases by simultaneously applying vacuum and ultrasound to the mobile phases for 30 seconds. If you are using a low-pressure mixing gradient system, sparging with helium may be necessary. Waters Symmetry C 18 columns are shipped containing water/acetonitrile. Before the first analysis, equilibrate the column with mobile phase at 1.3 ml/min for 10 minutes in mobile phase or until the baseline is stable. See Table 4 for separation conditions. 4

b. Performing the HPLC Analysis To analyze the sample: 1. Prepare the standard solution of the DNPH derivatives that you need to quantify. The concentrations of the standards should be in the same range as the expected concentrations in the sample. To synthesize DNPH derivatives, see Appendix B. 2. Prepare a cartridge blank from the same sample lot as the cartridge used to collect the sample, using the sample procedure and same bottled solvent. 3. Analyze the standard solution to determine the response factor for each derivative. Due to the high linearity of the detector response, a single point calibration is sufficient for Waters detectors. Note: Use an injection volume appropriate for your column. Inject 20 µl for a 3.9 x 150 mm Waters Symmetry C 18 column, and 10 µl for a 3.0 x 75 mm Waters Symmetry C 18 column. 4. Analyze the cartridge blank to determine background levels. Ensure that the blank values are in the normal range (see Table 3). Figures 4 and 5 show a representative separation and blank sample, respectively. 5. Analyze the samples. 6. Subtract the blank values from the sample values. Run standards at regular intervals between samples. Table 4: HPLC Separation Conditions Column Waters Symmetry C 18, 3.9 x 150 mm Mobile Phase 45:55 Acetonitrile/Water Flow Rate 1.3 ml/min Injection Volume 20 µl Detection Wavelength Absorbance at 360 nm (<50 mv of baseline noise) Limit of Detection for Formaldehyde Limit of Quantification for Formaldehyde Based on a signal to noise ration of 3 (55 picograms in 20 µl injection) Based on a signal to noise ratio of 10 (183 picograms in 20 µl injection) Figure 4. Isocratic Separation of C 1 -C 3 Hydrazone Derivatives..0030.0025.0020.0015.0010.0005 0 Figure 5: Typical Sampler Blank. c. Analyzing a Cartridge Blank Analyze a blank to determine background levels. Since background levels may change during storage, always compare samples to blank cartridges from the same lot stored under the same conditions. Note: When preparing a blank sample, ensure that you use the exact bottled reagents that were used for the preparation of the sample. To prepare a cartridge blank: 1 2 1. DNPH 2. Formaldehyde-DNPH 3. Acetaldehyde-DNPH 4. Acetone-DNPH 5. Acrolein-DNPH 6. Propionaldehyde-DNPH Time (minutes) 3 1. DNPH 2. DNPH-Formaldehyde 3. DNPH-Acetaldehyde 4. DNPH-Acetone 0 2 4 6 8 10 12 14 4 1. Elute a fresh DNPH-Silica Sep-Pak cartridge from the same lot as the cartridges used to collect your sample 2. Analyze the solution by HPLC using the same conditions as those used for the sample. 3. Multiply the concentration of each DNPH derivative by the volume of the eluate to determine the amount of background DNPH derivative. 5

d. Calculating Results To calculate the aldehyde concentration in air: 1. Prepare a calibration line of peak area to standard concentration on µg/ml of analyte as carbonyl compound. 2. Use the calibration line to determine the analyte concentration in the blank and sample eluates. 3. Determine the mass of each analyte in the blank (W b ) in µg. This is done by multiplying each analyte concentration by the eluate volume in ml. Compare these vales to the specifications listed in Table 3. If these values are higher than the specifications, contamination occurred during sample preparation or storage. 4. Determine the mass of each analyte in the sample (W s ) in µg. This is done by multiplying each analyte concentration by the eluate volume in ml. 5. Calculate the analyte concentration (C) in µg/l*. This is done by dividing the weight of the analyte in the air (Ws-Wb) by the volume of the air sampled (V) in liters. C = W s W b V *The units, µg/l, is equivalent to mg/m 3. For converting µg/l to ppmv at 1 atm and 20 C, see Appendix D. IV. APPLICATION EXAMPLES a. Formaldehyde in Laboratory Air - STEL The sample in Figure 6 was collected in a chemical research laboratory using a portable sampling pump. The 22.5 liter air sample was collected at 1.5 L/min. The chromatogram shows 0.02 ppmv formaldehyde, with trace levels of both acetonitrile and acetone detected..004.002 1 2 1. DNPH 2. DNPH-Formaldehyde 3. DNPH-Acetaldehyde 4. DNPH-Acetone 3 4 0 0 2 4 6 8 10 12 14 b. Glutaraldehyde in Laboratory Air STEL The sample in Figure 7 was collected in chemical research laboratory using a portable sampling pump. The 22.5 liter air sample was collected at 1.5 L/min. The chromatogram shows 0.02 ppmv of glutaraldehyde. Under these chromatographic conditions the DNPH-glutaraldehyde diastereomers are resolved..004.002 1 0 0 2 4 6 8 Figure 7. Analysis of Glutaraldehyde in Lab Air. 1. DNPH 2a & 2b. DNPH-Glutaraldehyde Diastereomers 2b V. STORAGE AND DISPOSAL OF THE SAMPLERS a. Storing unused samplers Always store any unused Waters Sep-Pak XPoSure Aldehyde Samplers in their protective pouches to prevent contamination. Store the sealed pouches in a refrigerator at (4 C ± 2 C) for up to six months. Cartridges may be stored in their unopened pouches at room temperature (20 to 25 C) for up to two weeks. Background levels of hydrazone derivatives increase slightly with time and temperature. Before using cartridges exposed to high temperatures or stored longer than the recommended periods, run a blank. b. Storing exposed samplers Once a sampler has been used for collection, be careful to cap and seal it until it is time to elute it. Inadvertent exposure of an exposed cartridge can ruin a carefully collected sample. Elute the derivatives from the cartridge within two weeks. The sample eluates are stable at 4 C ± 2 C for up to one month. c. Disposing of used cartridges Dispose of used cartridges according to applicable government regulations. 2a Figure 8. Analysis of Formaldehyde in Lab Air. 6

VI. TROUBLESHOOTING Table 5 describes solutions to problems that may arise while using the samplers. Most errors occur as a result of contamination during sample preparation. To resolve chromatographic problems not listed, refer to your HPLC system manual. 4 Table 5: Troubleshooting Symptom Possible Cause Solution High carbonyl values in sampler blank High formaldehyde levels in sampler blank Broad peaks Contaminated acetonitrile Contaminated glassware Air contamination during elution Sampler age and storage conditions Coelution of formaldehyde with an impurity Injection volume too high System or column failure Certify acetonitrile quality prior to use, see Appendix A. Use only pre-cleaned glassware. Prepare sample in a glove box. Replace samplers. Refrigerate fresh samplers. Prepare a fresh mobile phase or decrease the acetonitrile content. Reduce the injection volume. Consult the HPLC System manual. VII. REFERENCES AND BIBLIOGRAPHY 1. Committee on Aldehydes, Board of Toxicology and Environmental Hazards, National Research Council, Formaldehyde and Other Aldehydes; National Academy Press, Washington, DC, 1981. 2. Tejada, S. B. Evaluation of Silica Gel Cartridges Coated in Situ With Acidified 2,4-Dinitrophenylhydrazine for Sampling Aldehydes and Ketones in Air, Intern. J. Environ. Chem. 1986, 26, 167-185. 3. Riggins, R.M. Compendium of Methods for the Determination of Toxic Organic Compounds in Ambient Air, U.S. Environmental Protection Agency Report EPA-600/4-84-041, U.S. Environmental Protection Agency; Research Triangle Park, NC, 1984. 4. Guide to Successful Operation of Your LC System; Waters Corporation, Milford, MA 1991. 5. ASTM Method E411; Standard Test Method for Trace Quantities of Carbonyl Compounds with 2,4-Dinitrophenylhydrazine. VIII. ORDERING INFORMATION Waters s are shipped in boxes of 20 individually-packaged cartridges. Pouches are supplied for storage after sampling. To order, contact your Waters representative, call Waters at (800) 252-4752 or order online at www.waters.com. Table 6: Ordering Information Product Part Number s, 20/box WAT047205 In-Line Pre-Column Filter WAT084560 Symmetry C 18 Column, 3.9 x 150 mm WAT046980 Solvent Filter, PVDF, 47 mm, 0.45 µm, Protein, 100/pkg WAT200530 Beginners Guide to Liquid Chromatography 715001531 Guide to Successful Operation of Your LC System WAT022378TP IX. APPENDICES a. Appendix A: Measuring Acetontrile Purity HPLC-grade acetonitrile may contain traces of aldehydes and ketones, and especially acetone. A concentration of 10 µg/l of an aldehyde or ketone in the acetonitrile adds 0.1 µg to the blank values determined for the DNPH-derivatives per cartridge. If you acetonitrile is unacceptable for your application, contact your solvent supplier, to purify the acetonitrile. To purify acetonitrile, distill it from an acidified DNPH solution, using a procedure analogous to the one described in ASTM Method E411 for the purification of methanol. 5 To measure acetonitrile purity: 1. Clean all glassware by rinsing with acetonitrile and heating in a 60 C vacuum oven for at least 30 minutes. 2. Elute a fresh sampler with 3 ml acetonitrile. 3. Within 3 minutes, inject the eluate into the HPLC system to measure the concentration of DNPH derivatives. 4. Add 1 drop of concentrated HCl to the eluate, and allow to react at room temperature for 30 minutes. 5. Remeasure the concentration of DNPH derivatives by HPLC. 6. Calculate the difference in the concentration of each DNPH derivative measured in steps 3 and 5 to yield the contribution form the acetonitrile. 7. Calculate the percent hydrazone contributed by the acetonitrile relative to the background level. The value for any hydrazone should not exceed 25% of its value in the blank. 7

Example: Measuring Acetonitrile Purity 1. HPLC analysis of a fresh sampler shows the sample contains: Derivative Formaldehyde- DNPH Acetaldehyde- DNPH Contribution from Acetonitrile Divided by Background Value Times 100 Equals Percent Relative to Background 0.01 0.08 x 100 = 12% 0.02 0.12 x 100 = 17% Acetone-DNPH 1.60 0.40 x 100 = 400% 2. Analysis of the concentrations of hydrazones after reacting with acid yields: Derivative Concentration (µg/ml) Formaldehyde-DNPH 0.08 Acetaldehyde-DNPH 0.12 Acetone-DNPH 0.40 All other hydrazones <0.05 3. The difference between the concentrations of hydrazone from steps 3 and 5 represents the amount of hydrazone contributed by the acetonitrile: Derivative Concentration (µg/ml) Formaldehyde-DNPH 0.09 Acetaldehyde-DNPH 0.14 Acetone-DNPH 2.00 All other hydrazones <0.05 4. The percent of the hydrazones contributed by the acetonitrile is: Derivative Formaldehyde- DNPH Acetaldehyde- DNPH Concentration after Reaction with Acid Minus Concentration in Blank Equals Contribution form Acetonitrile 0.09 µg/ml - 0.08 µg/ml = 0.01 µg/ml 0.14 µg/ml - 0.12 µg/ml = 0.02 µg/ml Acetone-DNPH 2.00 µg/ml - 0.40 µg/ml = 1.60 µg/ml Since the percent for formaldehyde and acetaldehyde arising from the acetonitrile is less than 25% of the background level in the sampler, the acetonitrile is considered clean for these compounds. If the analysis considers only these compounds, the acetonitrile is acceptable. However, the amount of acetone arising for the acetonitrile is 4 times the amount found in the the background level. Therefore, it is suggested that this lot of acetonitrile may be unacceptable for use in the analysis of acetone. b. Appendix B: Synthesizing the DNPH-Derivative Standards High purity (99%) DNPH derivatives are commercially available or can be synthesized from DNPH supplied by Aldrich Chemical Company (70% DNPH and 30% water). To synthesize 98-99% pure hydrozones: 1. Prepare one liter of 2 M HCl solution: Add 172 ml concentrated reagent-grade hydrochloric acid (HCl) to a 1 L volumetric flask. Fill the flask to the mark with distilled deionized water. 2. Saturate the 2 M HCl solution with DNPH: Add 8 g DNPH and stir for one hour at 20 to 25 C. Filter through a 0.45 µm hydrophilic membrane (HVLP) filter (Waters Part number: WAT200530). 3. Form the hydrazone derivative by adding a two-fold molar excess of reagent-grade aldehyde or ketone to the filtered 2 M HCl DNPH solution. Stir for 30 minutes to one hour at 20 to 25 C. 4. Filter the hydrazone slurry. Wash the hydrazone with 50 ml 2 M HCl 3 times. Wash with 50 ml water 3 times. Dry the filter cake in an oven at 50 to 60 C overnight. 5. Prepare a standard stock solution of the DNPH-derivatives by dissolving an accurately weighed amount in acetonitrile. Prepare a set of calibration standards using the stock solution. The concentration of the standards should be in the same range as the expected concentration of the samples. The solutions are stable for at least one month when stored in tightly-capped glass vials at 4 C ± 2 C. 8

c. Appendix C: Measuring Breakthrough Note: If several aldehydes and ketones are present in significant concentration, estimate the maximum sample size from the total concentration of all species. Collection efficiency determinations are best made during times expected to yield peak formaldehyde concentrations. This will enable appropriate sampling rates and intervals to be selected to avoid breakthrough. Figure 8 shows the predicted total carbonyl concentration versus the range of sample volumes. To measure Waters for collection efficiency: 1. Connect two unused cartridges together by the Luer fittings and mark each cartridge for identification. 2. Connect the cartridges to a calibrated pump with a short length of flexible tubing. 3. Collect the sample. 4. Elute both cartridges and a third blank cartridge. 5. Analyze all three cartridges by HPLC. 6. Subtract the blank value from the values determined from the other two cartridges. 7. Calculate and sum of the total captured DNPH-derivative from both cartridges. 8. Divide the amount of DNPH-derivative determined from the first cartridge by the total amount determined form cartridges 1 and 2. Multiply by 100. This is the percentage of DNPH-derivatives captured on the first cartridge. This value should exceed 95%; otherwise, some of the sample broke through to the second cartridge. Example: Measuring Sample Breakthrough The expected concentration of formaldehyde is 0.66 ppmv (µl/l). Flow rate is 1.25 L/min for 80 minutes. A sample volume of 100 liters is collected. The theoretical quantity of carbonyl is: Analyte ppmv Carbonyl Collected concentration x molecular x air volume = µg Carbonyl weight Figure 8. Total Carbonyl Concentration vs. Range of Sample Volumes. Analyte molar volume at 1 atm/25 C Collection efficiency for Waters Sep-Pak XPoSure Aldehyde Sampler is greater than 95% for air sampling rates of up to 1.5 L/min. The sampler may exhibit breakthrough if: The sampling flow rate is greater than 1.5 L/min The amount of sample collected is enough to react with more than 50% of the DNPH (~2.3 µmoles) This calculates to: 0.66 µl x 30.03 g/mole x 100 L = 81 µg formaldehyde 24.46 L/mole The actual results are shown in Table 7. To calculate the percent captured on the first sampler, divide the quantity captured on sampler 1 by the total quantity captured, then multiply by 100. Since this value is less than 95%, and the total carbonyl amount exceeded 2.3 µmoles, breakthrough occurred. 9

In the above example, only a single carbonyl source was present. Under many test conditions more than one carbonyl source may be present in significant concentrations. These other compounds will consume DNPH, effectively reducing the capacity of the sampler for the compound of interest. To assure that the capacity of the sampler has not been exceeded, compare the DNPH peak areas of the sample to a similarly eluted blank. The DNPH peak area in all samples must be no less than 50% of the DNPH peak area of the blank. This ensures the sampler capacity has not been exceeded. Table 7: Breakthrough Example HPLC Results Sampler Amount (µg) Quantity Captured Sampler blank (µg) Percent Captured on Sampler Sampler 1 75.06 75.00 91.8 Sampler 2 6.72 6.66 8.2 Blank 0.06 - - Table 9: Conversion Factors, Carbonyl Weights Derivative Weights Carbonyl Compounds Carbonyl Derivative, (MW D /MW C ) Derivative Carbonyl, (MW C /MW D ) Formaldehyde 6.998 0.1429 Adetaldehyde 5.089 0.1965 Adetone 4.101 0.2438 Glutaraldehyde 4.589 0.2175 d.2. Equation for converting µg/l to ppmv Carbonyl concentrations can be converted from µg/l to ppmv (µl/l) by using the following expression: (Result in ppmv) = (Result in µg/l) x 22.41 x T2 x P1 MWC T1 P2 d. Appendix D: Useful Conversion Factors This appendix contains: Carbonyl to hydrazone conversion factors Equation for converting µg/l to ppmv Conversion factors: µg/l ppmv Obtaining carbonyl concentrations in eluates and air samples required the use of several constants and conversion factors. The factors described in this appendix can be used when converting carbonyl weights to: Equivalent derivative weights for preparing standard solutions Volumes for reporting air samples in ppmv d.1. Carbonyl to Hydrazone Conversion Factors Table 8 lists the molecular weights (MW) for some carbonyl compounds. These values were used to derive the conversion factors listed in Table 9. Multiply the carbonyl or derivative weights by the appropriate factor for the desired conversion. Table 8: Carbonyl and Hydrazone Molecular Weights Carbonyl Compounds Carbonyl Compounds Molecular Weight, (MW C ) Hydrazone Derivative Molecular Weight (MW D ) Formaldehyde 30.03 210.15 Adetaldehyde 44.05 224.17 Adetone 58.08 238.20 Glutaraldehyde 100.12 460.36 Where values are: 22.41 = Molar volume of an ideal gas at STP (273.15 K and 1 atm), L/mole MWC = Molecular weight of carbonyl, g/mole T1 = Standard temperature, 273.15 K T2 = Air sample temperature, ºK P1 P2 = Standard pressure, 1 atm = Air sample pressure, atm d.3. Conversion Factors: µg/l to ppmv Table 10 lists the factors for converting between µg/l and ppmv at 25 C and 1 atm. Results are converted between µg/l (or mg/m 3 ) and ppmv, by multiplying by the appropriate factor. Table 10: Factors for Converting Between µg/l and ppmv at 25 C and 1 atm. Carbonyl ppmv µg/l µg/l p ppmv Compounds Formaldehyde 1.23 0.813 Adetaldehyde 1.80 0.555 Adetone 2.38 0.420 Glutaraldehyde 4.09 0.244 10

Waters, The Science of What s Possible, Symmetry, and Sep-Pak are registered trademarks of Waters Corporation. XPoSure is a trademark of Waters Corporation. All other trademarks are the property of their respective owners. 2014 Waters Corporation. Produced in the U.S.A. March 2014 Rev C WAT047204EN KP-PDF Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com