Development of a Portable Preconcentrator/lon Mobility Spectrometer System for the Trace Detection of Narcotics

Similar documents
Development of a High Intensity EBIT for Basic and Applied Science

August 3,1999. Stiffness and Strength Properties for Basic Sandwich Material Core Types UCRL-JC B. Kim, R.M. Christensen.

Magnetic Measurements of the Elliptical Multipole Wiggler Prototype

Modeling Laser and e-beam Generated Plasma-Plume Experiments Using LASNEX

The Radiological Hazard of Plutonium Isotopes and Specific Plutonium Mixtures

Simulation of Double-Null Divertor Plasmas with the UEDGE Code

m w n? r i OF,THISDOCUMENT IS UNLIMITED

BWXT Y-12 Y-12. A BWXT/Bechtel Enterprise SMALL, PORTABLE, LIGHTWEIGHT DT NEUTRON GENERATOR FOR USE WITH NMIS

RESEARCH OPPORTUNITIES AT THE CNS (Y-12 AND PANTEX) NUCLEAR DETECTION AND SENSOR TESTING CENTERS (NDSTC)

Protective Coatings for Concrete

BASAL CAMBRIAN BASELINE GEOLOGICAL CHARACTERIZATION COMPLETED

ABSTRACT INTRODUCTION

Data Comparisons Y-12 West Tower Data

Determine the Inside Wall Temperature of DSTs using an Infrared Temperature Sensor

C o f l F - 9m307-- SPREADSHEET APPLICATION TO CLASSIFY RADIOACTIVE MATERIAL FOR SHIPMENTa. A. N. Brown

OAK RIDGE NATIONAL LABORATORY

Parametric Instabilities in Laser/Matter Interaction: From Noise Levels to Relativistic Regimes

Determine the Inside Wall Temperature of DSTs using an Infrared Temperature Sensor

John Falconer Richard Noble

Plutonium 239 Equivalency Calculations

HEAT TRANSFER AND THERMAL STRESS ANALYSES OF A GLASS BEAM DUMP

Abstract of paper proposed for the American Nuclear Society 1997 Winter Meeting Albuquerque, New Mexico November 16-20, 1997

Inside Wall Temperature Measurements of DSTs Using an Infrared Temperature Sensor

Safety Considerations for Laser Power on Metals in Contact with High Explosives-Experimental and Calculational Results

INTERMOLECULAR POTENTIAL FUNCTIONS AND HIGH RESOLUTION MOLECULAR SPECTROSCOPY OF WEAKLY BOUND COMPLEXES. Final Progress Report

Controlling Backstreaming Ions from X-ray Converter Targets with Time Varying Final Focusing Solenoidal Lens and Beam Energy Variation

PROJECT PROGRESS REPORT (03/lfi?lfibr-~/15/1998):

DETECTING AND QUANTIFYING LEWISITE DEGRADATION PRODUCTS I N ENVIRONMENTAL SAMPLES USING ARSENIC SPECIATION*

Optimization of NSLS-II Blade X-ray Beam Position Monitors: from Photoemission type to Diamond Detector. P. Ilinski

Paper to be submitted to. Thermal Management Technology for Hydrogen Storage: Fullerene Option?

Valley-Fill Sandstones in the Kootenai Formation on the Crow Indian Reservation, South-Central Montana

Destruction of 2,4,6=Trinitrotoluene using Ammonium Peroxydisulfate

A TER. Observation of Xe with the PNNL ARSA System PNNL UC-7 13

Valley-Fill Sandstones in the Kootenai Formation on the Crow Indian Reservation, South-Central Montana

Three-Dimensional Silicon Photonic Crystals

J. C. Batchelder', K. S. Toth2, D. M. Moltz3, E. F. Zganjarl, T. J. Ognibene3, M. W. Rowe3, C. R. Binghan12.~,J. Powell3, and B. E.

AC dipole based optics measurement and correction at RHIC

CQNl_" RESPONSE TO 100% INTERNAL QUANTUM EFFICIENCY SILICON PHOTODIODES TO LOW ENERGY ELECTRONS AND IONS

Multicusp Sources for Ion Beam Lithography Applications

Clifford K. Ho and Michael L. Wilson Sandia National Laboratories. P.O. Box Albuquerque, NM

: Y/LB-16,056 OAK RIDGE Y-12 PLANT

RECEIVED SEQ

(4) How do you develop an optimal signal detection technique from the knowledge of

Solid Phase Microextraction Analysis of B83 SLTs and Core B Compatibility Test Units

Determinktion of Iodine to Compliment Mass Spectrometric Measurements

PRINCETON PLASMA PHYSICS LABORATORY PRINCETON UNIVERSITY, PRINCETON, NEW JERSEY

Synthesis of Methyl Methacrylate from Coal-derived Syngas

Multi-Scale Chemical Process Modeling with Bayesian Nonparametric Regression

Tell uric prof i 1 es across the Darrough Known Geothermal Resource Area, Nevada. Harold Kaufniann. Open-file Report No.

ADSORPTION ON NANOSURFACES: A DETAILED LOOK AT METAL CLUSTERS USING INFRARED SPECTROSCOPY

FUSION TECHNOLOGY INSTITUTE

ASTER CONF-~W OSTI MAY DISTRIBUTION OF THIS

~ _- IDOCLRXKT DMSP SATELLITE DETECTIONS OF GAMMA-RAY BURSTS. J. Terrell, NIS-2 P. Lee, NIS-2 R W.

Use of Gamma Rays from the Decay of 13.8-sec "Be to Calibrate a Germanium Gamma Ray Detector for Measurements up to 8 MeV

Applications of Pulse Shape Analysis to HPGe Gamma-Ray Detectors

Excitations of the transversely polarized spin density. waves in chromium. E3-r 1s

Quarterly Report April 1 - June 30, By: Shirley P. Dutton. Work Performed Under Contract No.: DE-FC22-95BC14936

Application of a Three-Dimensional Prognostic Model During the ETEX Real-Time Modeling Exercise: Evaluatin of Results (u)

GA A22677 THERMAL ANALYSIS AND TESTING FOR DIII D OHMIC HEATING COIL

GA A23736 EFFECTS OF CROSS-SECTION SHAPE ON L MODE AND H MODE ENERGY TRANSPORT

National Accelerator Laboratory

GA A26474 SYNERGY IN TWO-FREQUENCY FAST WAVE CYCLOTRON HARMONIC ABSORPTION IN DIII-D

GA A24166 SUPER-INTENSE QUASI-NEUTRAL PROTON BEAMS INTERACTING WITH PLASMA: A NUMERICAL INVESTIGATION

ITP Filtrate Benzene Removal Alternatives

The Graphite Isotope Ratio Method (GIRM): A Plutonium Production Verification Tool

GA A26686 FAST ION EFFECTS DURING TEST BLANKET MODULE SIMULATION EXPERIMENTS IN DIII-D

PROGRESS REPORT JULY TO SEPTEMBER

DE '! N0V ?

UPGRADED CALIBRATIONS OF THE THOMSON SYSTEM AT DIII D

Experiment. The Pinhole Neutron. Pinex. c. c. sartain UCRL-ID November 21,1958

c*7 U336! On Satellite Constellation Selection LA MS ^«fc ^Xiyy PLEASE RETURN TO:

To be published in the Proceedings of ICEC-22, Seoul Korea, July 2008 MICE Note 232 1

SRL TEE DRY DEPOSITION OF PARTICULATE MATTER ABOVE. Robert Lorenz and Charles E. Murphy, Jr.

&OAE- 9b I 0 IS& - - I Determination of Plutonium in Urine: Evaluation of Electrothermal Vaporization Inductively Coupled Plasma Mass Spectroscopy.

12/16/95-3/15/96 PERIOD MULTI-PARAMETER ON-LINE COAL BULK ANALYSIS. 2, 1. Thermal Neutron Flux in Coal: New Coal Container Geometry

0STI. E. Hammerberg, XNH MD SIMULATIONS OF DUSTY PLASMA CRYSTAL FORMATION: PRELIMINARY RESULTS M. J. S. Murillo, XPA

MAGNETIC RESONANCE IMAGING OF SOLVENT TRANSPORT IN POLYMER NETWORKS

Bulk Modulus Capacitor Load Cells

UNIT OPERATIONS AND CHEMICAL PROCESSES. Alan M. Krichinsky

Plasma Response Control Using Advanced Feedback Techniques

Half-Cell, Steady-State Flow-Battery Experiments. Robert M. Darling and Mike L. Perry

Analysis of Shane Telescope Aberration and After Collimation

AEC Research and Development REPORT

A Geological and Geophysical Assessment of the Royal Center Gas Storage Field in North-Central Indiana, a Joint NIPSCO, DOE & GRI Case Study

ASTER. Dose Rate Visualization of Radioisotope Thermoelectric. Generators. RECElVED BEC OF THIS DOCUMEMT IS UNLlMIi L, Hanford Company

Introduction to Instrumental Detection Technology: IMS and DMS

9 7og$y4- International Conference On Neutron Scattering, Toronto August Spin Dynamics of the reentrant spin glass Fe0.7A10.3.

J. R Creighton and C. M. Truong Org. 1126, MS 0601 P.O. Box 5800 Sandia National Laboratories Albuquerque, NM

Status and Progress of a Fault Current Limiting HTS Cable To Be Installed In The Consolidated Edison Grid

A Few-Group Delayed Neutron Model Based on a Consistent Set of Decay Constants. Joann M. Campbell Gregory D. Spriggs

LLNL DATA COLLECTION DURING NOAAETL COPE EXPERIMENT

PROCEEDINGS THIRD WORKSHOP GEOTHERMAL RESERVOIR ENGINEERING. December 14-15,1977

AlliedSianal. Measurement U ncertai n ty of Silicone Fluid Leakage Testing for Rolamites. W. E. Holland. Federal Manufacturing & Technologies

FUSION WITH Z-PINCHES. Don Cook. Sandia National Laboratories Albuquerque, New Mexico 87185

ORNL/TM-2002/118 Plutonium Production Using Natural Uranium From the Front-End of the Nuclear Fuel Cycle

PLASMA MASS DENSITY, SPECIES MIX AND FLUCTUATION DIAGNOSTICS USING FAST ALFVEN WAVE

PRECISION STUDIES OF NUCLEI

Lawrence Berkeley National Laboratory Lawrence Berkeley National Laboratory

Equivalencing the Collector System of a Large Wind Power Plant

The Concentration of 236 Pu Daughters in Plutonium for Application to MOX Production from Plutonium from Dismantled US Nuclear Weapons

Transcription:

SANDIA REPORT SAND97-2037 Unlimited Release Printed August 1997 UC-900 Development of a Portable Preconcentrator/lon Mobility Spectrometer System for the Trace Detection of Narcotics Prepared by ornia 94550 United States

Issued by Sandia National Laboratories, operated for the United States Department of Energy by Sandia Corporation. NOTICE: This report was prepared as an account of work sponsored by an agency of the United States Government. Neither the United States Government nor any agency thereof, nor any of their employees, nor any of their contractors, subcontractors, or their employees, makes any warranty, express or implied, or assumes any legal liability or responsibility for the accuracy, completeness, or usefulness of any information, apparatus, product, or process disclosed, or represents that its use would not infringe privately owned rights. Reference herein to any specific commercial product, process, or service by trade name, trademark, manufacturer, or otherwise, does not necessarily constitute or imply its endorsement, recommendation, or favoring by the United States Government, any agency thereof, or any of their contractors or subcontractors. The views and opinions expressed herein do not necessarily state or reflect those of the United States Government, any agency thereof, or any of their contractors. Printed in the United States of America. This report has been reproduced directly from the best available copy. Available to DOE and DOE contractors from Office of Scientific and Technical Information P.O. Box 62 Oak Ridge, TN 37831 Prices available from (615) 576-8401, FTS 626-8401 Available to the public from National Technical Information Service U.S. Department of Commerce 5285 Port Royal Rd SpringGeld, VA 22161 NTIS price codes Printed copy: A03 Microfiche copy: A01

SAND97-203 7 Unlimited Release Printed August 1997 Distribution Category UC-900 Development of a Portable Preconcentratorfion Mobility Spectrometer System for the Trace Detection of Narcotics John E. Parmeter and Chad A. Custer Contraband Detection Technologies Sandia National Laboratories P. 0. Box 5800 Albuquerque, NM 871 85-0782 Abstract This project was supported by LDRD funding for the development and preliminary testing of a portable narcotics detection system. The system developed combines a commercial trace detector known as an ion mobility spectrometer (IMS) with a preconcentrator originally designed by Department 5848 for the collection of explosives molecules. The detector and preconcentrator were combined along with all necessary accessories onto a push cart, thus yielding a fully portable detection unit. Preliminary testing with both explosives and narcotics molecules shown that the system is operational, and that it can successfully detect drugs such as marijuana, methamphetamine (speed), and cocaine based on their characteristic IMS signatures. 3

Acknowledgments The authors are indebted to the Albuquerque Public School s police force, and in particular to officers Mark Shea and Dennis Mascarenaz, for providing the drug samples used in this study. We also gratefully acknowledge the contributions to this work of Sandia employees Mary Green, Charles Rhykerd, Jr., Kevin Linker, Nathan Varley, and James Chapek. 4

1

Contents Introduction... Background... 6 Design ofthe Portable Narcotics Detection Unit... Figure 1... 6 7 8 Figure 2... 9 Preliminary Test Results... Mu.ana..... 9 10 Methamphetamine... 10 LSD... 10 Cocaine... 11 Summary and Future Outlook... 11 References... 12 5

Development of a Portable Preconcentrator/Ion Mobility Spectrometer System for the Trace Detection of Narcotics Introduction Trace detection of narcotics is a technological area that is of great importance to the law enforcement community. In the middle of FY 96, department 5848 received a limited amount of LDRD funding in order to pursue the adaptation of our technology for the trace chemical detection of explosivesto the detection of narcotics. The overall goals of this project were to construct a portable drug detection system on a wheeled cart, and to demonstrate proof of principle, i.e. show that the system is capable of detecting key narcotics molecules in a laboratory environment. These goals have been met, though we have not yet been able to advance to the stage of field testing the system. Background The detection technology employed in this study has been developed for a walk-through portal for the trace detection of explosives which department 5848 has developed under an interagency agreement with the Federal Aviation Administration. The portal s detection apparatus combines a unique preconcentrator developed at Sandia with a trace chemical detector known as an ion mobility spectrometer. Air flows over a test subject s body, entraining any explosive material that is present in the form of either vapor or particulate matter, and this airflow is directed into the preconcentrator where the explosive material is collected and concentrated. The explosives material is then delivered into a trace chemical detector known as an ion mobility spectrometer (IMS). The operation of the IMS and the preconcentrator are discussed separately below. Ion mobility spectrometry is one of the most powerfid analytical techniques currently available for the trace detection of chemicals. Past research in department 5848 has focused on using IMS to detect explosives, but the technique has also been applied to the detection of narcotics [l]. When an IMS is operated, air is pulled into an inlet at the front of the spectrometer, typically at a rate of 200 cubic centimeters per minute. This air flow, along with any entrained analyte molecules, first enters the ionization region, where analyte molecules are ionized via the process of atmospheric pressure chemical ionization. The ionization source is most commonly a beta emitter (Le., a radioactive source that emits electrons). A sample of the partially ionized gases in the ionization region is periodically pulsed into the drift region, a section of the instrument with an applied electric field, where analyte molecules are identified by the time it takes them to travel to a collector plate at the end of that region. Depending on the field polarity, the collected ions can be either positive or negative ions. In general, an IMS is run in negative 6

ion mode to detect explosives and in positive ion mode to detect narcotics, based on the stable molecular ions that these types of molecules form. A dopant called the carrier gas is generally added to the IMS flow in order to help initiate the ionization process. An IMS is capable of parts per trillion (ppt) sensitivity for some molecules in air, and has a nearly instantaneous response time. Current detectors provide adequate sensitivity only if a sample can be efficiently collected and delivered to the detector. Sample collection often limits system performance in field applications of trace chemical detectors. This is especially true in applications such as Sandia s explosives detection portal, where any explosive sample that is collected is initially contained in a very large volume (hundreds of liters) of air. This is also likely to be a problem in numerous applications of narcotics detection, e.g. in sampling air inside crates or lockers. Proprietary preconcentration technology recently developed in department 5848 has addressed t h i s issue for our explosives detection portal, and one of the aims of the current project was to apply this same technology to narcotics. In brief, the Sandia preconcentrator functions by allowing a large air flow to be drawn rapidly through a high density screen. The air flows on to an exhaust line, while heavy organic molecules originally in the flow (in the form of either vapor or particles) are collected on the screen. Once this is done, the screen is rapidly heated to desorb the explosives molecules into the gas phase, and these are then pulsed into the IMS using a much smaller air flow. Experiments with various explosives have shown that this preconcentrator is. approximately 50 % efficient in collecting explosives molecules, and - more importantly that this process increases the density of analyte molecules in the air delivered to the IMS by about two orders of magnitude [2]. In the remainder of this document, we describe the portable drug detection system that has been constructed, and discuss initial test results sampling various narcotics molecules. These results demonstrate that the system is functional. Further testing and refinement of the system is highly desirable, if funding becomes available to pursue this work. Design of the Portable Narcotics Detection Unit A schematic representation of the portable narcotics detection system is shown in Figure 1, and a photo is shown in Figure 2. The entire system is mounted on a wheeled cart with dimensions 63 x 30.25 x 44, making it fully portable. These dimensions are actually considerably larger than necessary, and it is envisioned that a second generation model would be substantially smaller. The IMS used in this system is a PCP, Inc. [3] Phemto Chem 100, with approximate dimensions of 11 x 18 x 15. This IMS was chosen simply because it was already in our group s possession, since our funding level did not allow the purchase of a new IMS (typically costing $50-75 K). The IMS uses a 10 millicurie Ni-63 source as a source of electrons in the ionization region. In the experiments discussed below, the IMS was operated with a drift region temperature of 200 C, and the carrier gas used was 7

methylene chloride in the experiments involving explosives (looking at negative ions) and nitrogen in the experiments involving narcotics (looking at positive ions). Flow parameters were inlet flow, 201 cc/min; carrier gas flow, 20 cc/min; exit flow, 700 cc/min; and drift gas flow (nitrogen), 500 cc/min. Figure 1. Schematic representation of the portable drug detection system. The two inlets below the main figure show the direction of air flow in the preconcentrator during the adsorption (left) and desorption (right) cycles. The IMS is connected to the preconcentrator via 1.5 of one quarter inch diameter stainless steel tubing. The preconcentrator has approximate dimensions of 7 x 6 x 3, with an internal volume of 600 cubic centimeters and an air inlet diameter of 4. Air can be sucked either directly into the preconcentrator inlet, or through an attached flexible sampling hose. In the experiments described in this report, no sampling hose was used. A Dayton one third horse power blower is used to pull air through the preconcentrator during the adsorption cycle, as shown in the left hand inset of Figure 1. This air flow goes through the face of the preconcentrating screen. During the desorption cycle, the 8

during the adsorption cycle, as shown in the left hand inset of Figure 1. This air flow goes though the face of the preconcentrating screen. During the desorption cycle, the screen is heated to approximately 200 "Cto desorb the explosive or narcotics material into the gas phase, and the malyte molecules are then pulsed into the IMS with a desorption air flow of 0.016 literslsec. As can be seen in the right hand inset to Figure 1, the direction of this desorb (or purge) air flow is perpendicular to the direction of the intake air flow. The desorption flow is provided by a compressed air cylinder, mounted on a lower level of the cart. Data acquisition is provided by an IBM compatible 486 personal computer using ASP sohare.... Figure 2. Photograph of the portable drug detedion system. 9

Preliminary Test Results Initially, the new system was tested using the explosives RDX and TNT, with which we have considerable previous experience. Operating the IMS in negative ion mode, it proved straightforward to detect both of these molecules when a small amount of vapor was desorbed directly into the preconcentrator inlet. In the case of TNT, a limit of detection (LOD) of approximately 6 nanograms was estimated for the combined preconcentrator and IMS. This is consistent with previous studies using a similar IMS and preconcentrator, where an LOD of approximately 1 nanogram was measured [2]. The higher value in the present study probably results from the fact that very little time or effort was spent trying to optimize the system for explosives detection. Testing the new system with narcotics had to be extremely limited due to the fact that we do not currently have a license to store narcotics in ow laboratory. On the afternoon of November 8, 1996, two officers fiom the Albuquerque Public Schools police force, Mark Shea and Dennis Mascarenaz, visited our laboratory with the following narcotics samples: marijuana, 3 grams; LSD, 1.5 grams; methamphetamines (speed), 10 tablets; and cocaine, 1 gram. The detection of these molecules was tested by holding samples directly up to the preconcentrator inlet while air was pulled through the preconcentrator, then subsequently heating the preconcentrator screen and pulsing the air containing the resulting narcotics vapor into the IMS. Because we were limited to four hours time we could not prepare quantitative solutions or other such samples, so the results are of a purely qualitative nature. Results for the individual drug molecules are as follows: Marijuana (active chemical tetrahydrocannabiol or THC, molecular weight = 3 14.45): Marijuana was easily detectable, producing an IMS spectrum characterized by peaks at drift times of 12.5, 13.8, 16.9, and 18.6 milliseconds (ms). The peak at 16.9 ms was quite small; the others were larger with the relative intensities being time dependent (and also depending on precise experimental conditions, such as preconcentrator screen heating temperature). Initially, the 18.6 ms peak was observed to be the most intense, and this is probably the parent peak (i.e. molecular THC or some adduct thereof), while the peaks at lower retention time probably represent decomposition products. Methamphetamine (speed, molecular weight = 135.20): Speed was also easily detectable, with an IMS spectrum characterized by a single, sharp peak at a drift time of 12.6 ms. Note that this is much lower than the value of 18.6 ms for the presumed parent peak for marijuana, consistent with the much lower molecular weight of speed (the lighter the molecule, the faster it will travel through the drift region). LSD (molecular weight = 268.32): The LSD sample tested gave a single strong peak at 18.6 ms. This looked identical to some marijuana spectra, and since the LSD sample smelled like marijuana we believe that the sample was contaminated and that marijuana is what actually was detected. This test was thus inconclusive, because it is still possible that LSD could be detected in the absence of marijuana. Any time two 10

species are sampled together in an IMS, they compete for ionizing charge, and the presence of one molecule may be completely masked if the other is much more easily ionized. We did not have time to pursue this issue further. Cocaine (molecular weight = 303.35): Cocaine was readily detected, producing a spectrum with three peaks at 10.7,13.2, and 16.8 ms. The first two peaks are large and remain in the spectrum for a relatively large time, while the peak at 16.8 ms decays rapidly. We suspect that this last peak is the parent ion, and that the two peaks at lower drift times are decomposition products. Summary and Future Outlook The results reported here demonstrate that the preconcentratorams technology developed by department 5848 for the trace detection of explosives can also be applied to the trace detection of important narcotics molecules, and that the entire detection system can be constructed so as to be easily portable. Due to a combination of factors (the difficulty of obtaining a permit to have narcotics samples in our laboratory, plus time constraints and the very limited funding available for this project) it was not possible to test the detection system with all narcotics molecules of interest. For example, heroin was never tested and the test results for LSD were inconclusive. Nevertheless, it appears that this system is probably capable of detecting any type of narcotic when operating in a the positive ion detection mode. Future work with this portable drug sniffer should focus on (1) obtaining more quantitative test data such as limits of detection for the various narcotics molecules, (2) reducing the size of the various components and of the cart itself to make the system lighter and simpler to transport, and (3) field testing. Our original intent for this project was to test the system in public schools by sampling air out of vents in lockers. This sort of testing would still be highly desirable. However, we are not working on this project at present since we currently have no funding for this research. 11

References [ 11 G. A. Eiceman and Z. Karpas, in Ion Mobility Spectrometry, CRC Press, Boca Raton, 1994, pp. 151-153. [2] Evaluation of Prototype Explosives Preconcentrators, a report written by Charles L. Rhykerd, Jr., and submitted to the Department of Energy s Office of Safeguards and Securities, February, 1996. [3] PCP, Inc., 2155 Indian Road, West Palm Beach, Florida, 33409-3287. 12

Distribution: 1 MS 9018 5 MS 0899 2 MS 0619 Central Technical Files, 8940-2 Technical Library, 491 6 Review & Approval Desk, 12690 For DOE/OSTI 24 2 1 MS 0782 MS 1436 MS 0161 Contraband Detection Technologies, 5848 Laboratory Directed R & D, 4523, Attn: D. Chavez Patent & Licensing Office, 1 1500 13