Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Similar documents
Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k

Developing Large Volume Injection (LVI) in Split / Splitless Inlets. Philip J. Koerner Phenomenex Inc. NEMC 2014

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS

So Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Selection of a Capillary GC Column

Selection of a Capillary

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Choosing the Correct GC Column Dimensions and Stationary Phase

Rely on Rxi -PAH Columns

Faster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit

Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms

The Suite for Environmental GC Analysis

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

A New Web-Based Application for Modeling Gas Chromatographic Separations

A New Web-Based Application for Modeling Gas Chromatographic Separations. Dr. Hansjoerg Majer Restek Corporation Market Development Manager Europe

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

HILIC Method Development in a Few Simple Steps

Volatile organic compounds (VOCs):

The Role of the Carbon Cartridge in the One-Pass System

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Determination of 24 PAHs in Drinking Water

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Advances in Sample Preparation for Accelerated Solvent Extraction

Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

The Design of High Temperature Capillary Gas Chromatography Columns Based on Polydimethylsiloxane

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Gas Chromatography. 1. Experiment Category: 2. Experiment Name: 3. Date and Issue number: 4. Instructor Name: 5. Institution: Ain Shams University

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Theory and Instrumentation of GC. Chromatographic Parameters

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Introduction to Capillary GC

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Secrets of GC Column Dimensions

Trajan SGE GC Columns

Get Selective. By Jaap de Zeeuw

Capillary GC Column Selection and Method Development A Primer on Column Parameters and Instrument Conditions

LC Technical Information

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Introduction to Capillary GC

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Amec Foster Wheeler Environment & Infrastructure, Inc BIG SHANTY ROAD, NW, SUITE 100 KENNESAW, GEORGIA (770)

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Chromatography. Gas Chromatography

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Analysis. Priority LC GC. Zebron PAH Accurately quantitate EU and EPA PAHs in less than 28 minutes

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

Determination of Polycyclic Aromatic Hydrocarbons (PAHs) in Seafood Using GC/MS

Ultra-Inert chemistry for Trace Level Analysis

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Column Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010

Non-Potable Water A2LA Certificate

Fast and Accurate Analysis of PBDEs in a Single Run, Including

Katherine K. Stenerson, Michael Ye, Michael Halpenny, Olga Shimelis, and Leonard M. Sidisky. Supelco, Div. of Sigma-Aldrich Bellefonte, PA USA

This project has been funded with support from the European Commission. This publication reflects the views only of the authors, and the Commission

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

GAS CHROMATOGRAPHY (GC)

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

ENVIRONMENTAL analysis

Activity in the FID Detection Port: A Big Problem if Underestimated

Monitoring Protein PEGylation with Ion Exchange Chromatography

CHEM Experiment Five Gas Chromatographic Analysis Of A Phenolic Extract. Objectives:

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Introduction and Principles of Gas Chromatography

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Your Reversed Phase Method

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

CHAPTER 6 GAS CHROMATOGRAPHY

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Evaluation of Capillary Columns for General Performance Parameters

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Transcription:

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation Anila I. Khan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 243 Key Words TraceGOLD fast GC analysis Abstract Today s laboratories are increasingly looking for ways to speed up analysis time in order to increase sample throughput and reduce analysis costs, without compromising results. This article discusses the use of Thermo Scientific TraceGOLD Fast GC columns and demonstrates that methods can be transferred from conventional GC columns onto shorter, narrower I.D. columns with minimal method development and no compromise in the quality of results. The result was also up to 5% faster analysis time. Introduction The primary goal of developing a chromatographic separation is to resolve a mixture of analytes. From the general resolution equation (Equation 1), and the graph shown in Figure 1, it is evident there are three parameters that control resolution, namely efficiency (N), selectivity (α), and the retention (k') factor. Selectivity and the retention factor can be improved by changing the column chemistry, carrier gas linear velocity, and temperature ramp rate. The selectivity relates to the separation of two components. Although this is the most sensitive variable for resolution, for complex mixtures, efficiency is used to provide a wider chromatographic applicability. Equation 1 Chromatographic efficiency, and therefore resolution, can be increased by the use of a longer column, but this will result in a longer analysis time. The favored methods of increasing chromatographic efficiency are to reduce the column internal diameter (d c ) or reduce the film thickness (d f ). An advantage of using narrower columns and thinner films is that users of conventional GC methods can reduce the analysis time with minimal method development. Significant improvements in the assay performance can be achieved without the need to make changes to the system setup using a fast GC column (2 m.15 mm.15 μm) compared to conventional column dimensions (3 m.25 mm.25 μm). Improved peak efficiencies can be obtained using a fast GC column without compromise in peak resolution, provided: The ratio of column length to I.D. is the same The column stationary phase remains the same The column phase ratio (β) is kept the same where possible Further improvement in the productivity can be obtained by combining higher optimal linear velocity with an increase in the temperature ramp rate. The effects of different system parameter changes on chromatographic efficiency and speed of analysis are discussed in this note.

2 3. 2.5 α 2. Resolution (R) 1.5 N 1..5 k. 1. 1.5 1.1 1.15 1.2 1.25 5 1 15 2 25 5 1 15 2 25 α k N Figure 1: Graphical representation of the general resolution equation Column Length The separation of sixteen polyaromatic hydrocarbons (PAHs) on a Thermo Scientific TraceGOLD TG-5SilMS GC column under a gradient temperature program is demonstrated in Figure 2. By halving the column length and using the same chromatographic conditions, the speed of analysis was increased by 34%. Resolution of the critical pairs is lost as a result of decreased efficiency. 1 14 12 1 1 2 3 4 R s = 2.1 5, 8 R s = 1.2 9, 1 TG-5SilMS 3 m.25 mm.25 µm 1.21 min R s = 1.21 13 R s = 1.2 1, 12 14, 15 8 TG-5SilMS 15 m.25 mm.25 µm 4 2 R s = 2. 5, R s = 1.2 9, 1 R s =.92, 12 12.1 min R s = 1. 1 14, 15 2 4 8 1 12 14 1 18 2 Figure 2: Column length effect on analysis time and resolution for the separation of 1 PAH standard mix Experimental conditions: Inlet SSL at 25 C; Carrier gas 1.2 ml/min helium, constant flow; Split injection 3:1; Injection volume 1. μl; Oven 9 C (1 min), 25 C/min, 28 C, 4 C/min, 32 C (5 min); Detector FID at 28 C Analytes: 1. Naphthalene, 2. Acenaphthylene, 3. Acenapthene, 4. Flourene, 5. Phenanthrene,. Anthracene,. Fluoranthene, 8. Pyrene, 9. Benzo[a]anthracene, 1. Chrysene,. Benzo[b]fluoranthene, 12. Benzo[k]fluoranthene, 13. Benzo[a]pyrene, 14. Indeno[1,2,3-cd]pyrene, 15. Dibenzo[a,h]anthracene, 1. Benzo[g,h,i]perylene.

Column Internal Diameter (d c ) Figure 3 illustrates that by reducing the column diameter, efficiency increases and as a consequence so does the resolution. The efficiency is always greater in the narrower bore column. Therefore, shorter columns can be used to reduce analysis time while offsetting the reduced efficiency coming from a shorter column length. Table 1 shows normalized efficiency for column length and diameter. Another advantage of using a narrow bore column is that optimal linear velocity of the carrier gas also increases, which allows shorter analysis time. From the Golay plots for a fast GC column, the optimal linear velocity that provides the highest efficiency is 32. cm/s compared to 28.5 cm/s for the conventional GC column. There are some practical considerations with the use of narrow bore columns, including lower sample loading capacity, which means that higher split ratios or reduced sample injection may be required to prevent column overload. 3.9.8.. 3 m.53 mm.25 µm 3 m.32 mm.25 µm 3 m.25 mm.25 µm 2 m.15 mm.15 µm U opt = 1. cm/s HETP.5.4 U opt = 2. cm/s.3.2 U opt = 28.5 cm/s U opt = 32. cm/s.1 1 2 3 4 5 Linear velocity cm/s Figure 3: Influence of column diameter on optimal gas velocity and HETP Column I.D. Column length (m) 3 2 15 1.15 3.3 1. 1.1.8..18 2.8 1.4.9..5.25 2. 1...5.3.32 1..8.5.4.3.53.9.5.3.2.2 Table 1: Normalized efficiency relative to column length and diameter. Normalized efficiency is calculated using 3 m.25 mm column dimension. Phase Film Thickness (d f ) and Phase Ratio (β) The separation of eleven compounds on a Thermo Scientific TraceGOLD TG-5MS GC column under a gradient temperature program is demonstrated in Figure 4. By reducing the film thickness from.5 μm to.25 μm and using the same chromatographic conditions, the speed of analysis increases by 13%. The resolution of the critical pairs increases considerably for these semi-volatile components as a result of increased efficiency.

4 12 TG-5MS 3 m.25 mm.5 µm 1 R s = 3.1 8 R s =. 1 2 5 1 1.9 min TG-5MS 3 m.25 mm.25 µm 4 2 R s = 3.4 R s =.3 15.3 min 4 8 1 12 14 1 18 Figure 4: Film thickness effect on analysis time and resolution for the separation of an eleven phenol standard mix Experimental conditions: Inlet SSL at 25 C; Carrier gas 1.2 ml/min helium, constant flow; Split injection 8:1; Injection volume 1. μl; Oven C (1 min), 1 C/min, 24 C; Detector FID at 28 C Analytes: 1. Phenol, 2. 2-Chlorophenol, 3. 2-Nitrophenol, 4. 2,4-Dimethylphenol, 5. 2,4-Dichlorophenol,. 4-Chloro-3-methylphenol,. 2,4,-Trichlorophenol, 8. 2,4-Dinitrophenol, 9. 4-Nitrophenol, 1. 2-Methyl-4,-dinitrophenol,. Pentachlorophenol. If the phase ratio (Equation 2) is kept consistent, then the elution order of compounds is anticipated to be the same. As Table 2 shows, replacing a.25 mm.25 μm GC column with a.15 mm.15 μm column and thus maintaining the same phase ratio, the same elution order of compounds will be seen provided the column stationary phase is kept the same. However, the efficiency on the.15 mm column diameter will be greater, allowing for a similar separation to be performed with a shorter column length. Equation 2. β Phase ratio of the column d c Column diameter (μm) d f Film thickness (μm) Column diameter, d c (mm) Film thickness, d f (µm).15.18.25.5 1 1.4 1.5 1.8 2.5 3 5.15 25 28 15 5 38 2 25 21 14 13 8.18 3 25 18 9 45 32 3 25 1 15 9.25 41 34 25 125 3 45 42 35 24 21 13.32 533 444 32 1 8 5 53 44 3 2 1.53 883 3 53 25 133 95 88 4 5 44 2 Table 2: Phase ratio values to ensure correct dimensions are selected for optimizing methods

Temperature Ramp Rate For the example shown in Figure 5, the separation of a phenol mix, the speed of analysis increases with increasing GC oven temperature ramp rate. Each 1 C/min increase in temperature ramp rate reduces the retention factor by approximately 5%, but at the expense of resolution. However, if resolution is sufficient then high temperature ramp rate can be used. implication is that the temperature of late-eluting analytes is also increased. The operating temperature of the column may not be high enough to elute compounds. In addition, at a rate above 4 C/min, there will be a greater variation in retention time of the late eluting compounds. This is due to inaccurate reading of the GC oven temperature >4 C/min. 5 The performance of the column also reduces with increasing ramp rate. In this case, efficiency cannot be used as a measure of column performance, instead peak width or peak capacity are generally used. The peak capacity (P c ) decreases by 22% as temperature ramp rate increases from 1 C/min to 2 C/min. The practical 24 22 2 18 ºC (1min), 1 ºC/min 1 2 R s = 3.4 5 R s =.3 1 Elution temp: 28 ºC 15.3 min P c = 3 1 14 12 1 8 4 2 ºC (1min), 15 ºC/min ºC (1min), 2 ºC/min 4 R s = 2.3 R s = 1.12 R s = 3.23 R s = 4.5 8 1 12 14 1 4. min Elution temp: 238 ºC 9.9 min Elution temp: 222 ºC P c = 38 P c = 318 Figure 5: The effect of oven temperature ramp rate on analysis time, resolution and elution temperature Experimental conditions: Column TG-5MS 3 m.25 mm.25 μm; Inlet SSL at 25 C; Carrier gas 1.2 ml/min helium, constant flow; Split injection 8:1; Injection volume 1. μl; Detector - FID at 28 C Analytes: 1. Phenol, 2. 2-Chlorophenol, 3. 2-Nitrophenol, 4. 2,4-Dimethylphenol, 5. 2,4-Dichlorophenol,. 4-Chloro-3-methylphenol,. 2,4,-Trichlorophenol, 8. 2,4-Dinitrophenol, 9. 4-Nitrophenol, 1. 2-Methyl-4, -dinitrophenol,. Pentachlorophenol. Peak capacities are calculated over peaks and an average is taken. Linear Velocity of Carrier Gas Using the same phenol mixture, the effect of varying the linear velocity (u, cm/s) of the carrier gas can be shown in Figure. It can be seen that the speed of analysis and the resolution increases with increasing linear velocity. Under the isothermal conditions, if the linear velocity deviates from the optimum linear velocity (U opt ), relative peak broadening and loss of resolution is observed. At linear velocities above the optimum flow, chromatographic efficiency decreases due to a nonequilibrium of the solute between the stationary and mobile phases. Although, the efficiency drops, the absolute peak width reduces because of the increase of the linear velocity driving the peak off the column in a reduced time. The effect of the increased carrier gas flow on the temperature gradient is to decrease the temperature gradient relative to the time the compounds stay on the column. This is seen in Figure, where peak capacity increases. The increased flow rate results in an increase in the peak capacity, due to the peak widths getting narrower and the temperature gradient effectively being decreased.

28 2 24 22 2 18 1 14 12 1 8 4 2 Elution temp: 22 ºC Linear velocity 3 cm/s (1 ml/min) ºC (1min), 8 ºC/min, 24 ºC R s =.2 R s = 3.1 18.99 min 5 1 1 2 Linear velocity 4 cm/s (2 ml/min) ºC (1min), 8 ºC/min, 24 ºC Linear velocity 5 cm/s (3 ml/min) ºC (1min), 8 ºC/min, 24 ºC 3 R s =.1 R s =.52 4 5 8 9 1 12 13 14 15 1 1 18 19 2 R s = 9. R s = 9.88 Elution temp: 189 ºC 1.8 min Elution temp: 181 ºC 1.19 min P c = 39 P c = 442 P c = 431 Figure : The effect of linear velocity on analysis time, resolution and elution temperature Experimental conditions: Column TG-5MS 3 m.25 mm.25 μm; Inlet SSL at 25 C; Carrier gas helium, constant flow; Split injection- 8:1; Injection volume 1. μl; Detector FID at 28 C. Analytes: 1. Phenol, 2. 2-Chlorophenol, 3. 2-Nitrophenol, 4. 2,4-Dimethylphenol, 5. 2,4-Dichlorophenol,. 4-Chloro-3-methylphenol,. 2,4,-Trichlorophenol, 8. 2,4-Dinitrophenol, 9. 4-Nitrophenol, 1. 2-Methyl-4,-dinitrophenol,. Pentachlorophenol. Peak capacities are calculated over peaks and an average is taken. Temperature Ramp Rate vs. Linear Velocity of Carrier Gas Figure illustrates the separation of a phenol mix on a TG-5MS GC column, where the resolution of peaks is decreased but the baseline is still resolved, as the linear velocity is increased. The ramp rate and isothermal times were adjusted using the simple equations displayed in Equation 3 and this reduced the analysis time by up to 5%. 2 24 22 2 18 1 14 12 1 8 4 2 Linear velocity 3 cm/s (1 ml/min) ºC (1min), 1 ºC/min, 24 ºC Linear velocity 4 cm/s (2 ml/min) ºC (.5min), 15 ºC/min, 24 ºC Linear velocity 5 cm/s (3 ml/min) ºC (.5min), 2 ºC/min, 24 ºC Elution temp: 28 ºC R s = 4.4 R s = 2.19.3 min 3 R s = 3.22 R s =.5 1 1 2 5 R s =.13 P c = 4 4 5 8 9 1 12 12 14 15 1 1 Elution temp: 2 ºC 1.9 min R s = 5.9 P c = 488 Elution temp: 28 ºC 1.8 min P c = 351 Figure : Effect of adjusting the linear velocity combined with varying the oven temperature ramp rate Experimental conditions: Column TG-5MS 3 m.25 mm.25 μm; Inlet SSL at 25 C; Carrier gas helium, constant flow; Split injection 8:1; Injection volume 1. μl; Detector FID at 28 C. Analytes: 1. Phenol, 2. 2-Chlorophenol, 3. 2-Nitrophenol, 4. 2,4-Dimethylphenol, 5. 2,4-Dichlorophenol,. 4-Chloro-3-methylphenol,. 2,4,-Trichlorophenol, 8. 2,4-Dinitrophenol, 9. 4-Nitrophenol, 1. 2-Methyl-4,-dinitrophenol,. Pentachlorophenol.

Method Transfer to a Fast GC Column To demonstrate the applicability of this approach, a fatty acid methyl ester (FAME) separation was transferred from a conventional column to a fast GC column. Table 3 displays the equivalent conventional and fast GC column configurations, which allow method analysis times to be readily reduced with only minor modifications to the method. The ratio of column length to I.D. and phase ratio are kept approximately the same, with the carrier gas flow rate and oven ramp rate variable to ensure that a similar chromatographic performance is obtained, while reducing the overall analysis time. Original column Replaced by 15 m.25 mm.25 µm 1 m.15 mm.15 µm 3 m.25 mm.25 µm 2 m.15 mm.15 µm m.25 mm.25 µm 4 m.15 mm.15 µm 15 m.32 mm.25 µm 1 m.15 mm.15 µm 3 m.32 mm.25 µm 15 m.15 mm.15 µm m.32 mm.25 µm 3 m.15 mm.15 µm Table 3: Faster analysis with same separation achieved using shorter, narrower bore columns The following equations (Equation 3) were used to determine the system parameters required to optimize performance using a fast GC column. Separation of FAMEs standard mix C8-C24 is used as an application illustrated in Figure 8. Equation 3. t g2 = t g1 v 2 v 1 β 2 β 1 l 1 l 2 T 2 = T 1 v 1 v 2 β 1 β 2 l 2 l 1 Where t g1, t g2 - temperature gradient for original and new conditions v 1, v 2 - linear velocity of gas for original and new conditions T 1, T 2 - hold time for isothermal part of separation for original and new conditions β 1, β 2 - phase ratio for original and new conditions l 1, l 2 - length of column for original and new conditions Standard method (I): TG-WaxMS 3 m.25 mm.25 μm, β = 25 Carrier gas: 1.2 ml/min helium flow rate, Linear velocity 3 cm/s, constant flow Split injection: 5:1, 1. μl Oven: 1 C (.5 min), 15 C/min, 22 C, 5 C/min, 25 C (5 min), total run time 19.5 min Fast method (II): TG-WaxMS 2 m.15 mm.15 μm, β = 25 Carrier gas:. ml/min helium flow rate, Linear velocity 3 cm/s, constant flow Split injection: 5:1,.5 μl Oven: 1 C (.3 min), 22.5 C/min, 22 C,.5 C/min, 25 C (3.5 min), total run time 13.13 min Faster method (III): TG-WaxMS 2 m.15 mm.15 μm, β = 25 Carrier gas: 1. ml/min Helium flow rate, linear velocity 43 cm/s, constant flow Split injection: 5:1,.5 μl Oven: 1 C (.25 min), 3 C/min, 22 C, 1 C/min, 25 C (3 min), total run time 1.25 min

35 325 3 25 25 225 2 15 15 125 1 5 5 25 Original method TG-WaxMS 3 m.25 mm.25 µm 1 1 2 3 4 Fast method TG-WaxMS 2 m.15 mm.15 µm 5, R s = 5.92 R s = 5.52 5, Faster method TG-WaxMS 2 m.15 mm.15 µm 5, R s = 5.52 9.13 min 14 2 3 4 5 8 9 1 12 13 14 15 1 1 18 19 8 9 1 12.21 min 14 Linear velocity 3 cm/s 12 13 1.8 min 14 Linear velocity 3 cm/s Linear velocity 43 cm/s Technical Note 243 Figure 8: Chromatogram for FAME standard mix C8 C24 illustrating method transfer for faster analysis Experimental conditions: Inlet SSL at 22 C; Carrier gas helium, constant flow; Detector FID at 24 C. Analytes: 1. C8:, 2. C1:, 3. C12:, 4. C14:, 5. C1:,. C1:1 [cis-9],. C18:, 8. C18:1 [cis-9], 9. C18:2 [cis-9,12], 1. C18:3 [cis-,9,12],. C2:, 12. C22: FAME, 13. C22:1 [cis-13], 14. C24:. Figure 8 shows the analysis time was decreased by 3% by using a modified method incorporating a fast GC column compared to a standard geometry column. The reduction in analysis time does not compromise the resolution. The speed of separation was further increased by increasing the linear velocity by approximately 4% 5%. Overall, the analysis time was reduced by approximately 5% of the original method, with no loss of resolution. Conclusion GC analysis time can be significantly reduced by transferring a method to fast GC columns. Performance not need be compromised by careful consideration of: Column length Column I.D. Column film thickness Carrier gas linear velocity Temperature ramp rate This approach has been used to transfer a FAMEs C8 C24 mix from a standard 3 m.25 mm.25 μm GC column to a fast GC column, giving up to 5% faster analysis time, with no compromise in resolution and with no changes to system configuration. References 1. Modern Practice of Gas Chromatography; Grob, R.L. and Barry, E.F., Eds.; Wiley: Hoboken, 24. thermoscientific.com/columnsforgc 213 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. USA and Canada +1 8 332 3331 France +33 ()1 92 48 34 Germany +49 () 2423 9431 2 or 21 United Kingdom +44 ()1928 534 Japan +81 3 582 115 TN243_E 2/13S China +8 21 854588 +8 1 84193588 +8 2 83145199 8 81 58 India +91 22 42 9494 +91 2 1 2352 Australia 1 3 35 292 (free call domestic) New Zealand 8 933 9 (free call domestic) All Other Enquiries +44 () 1928 534 5 Technical Support North America +1 8 332 3331 Outside North America +44 () 1928 534 44