The Agilent 6495 Triple Quadrupole LC/MS: Peptide Quantitation Performance

Similar documents
Agilent 6495 Triple Quadrupole LC/MS System EXPERIENCE A NEW LEVEL OF CONFIDENCE

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Guide to Peptide Quantitation. Agilent clinical research

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

László Tölgyesi LC/MS scientist EFSC Demo Group. International Symposium on Food Safety and Quality: Applications of Nuclear and Related Techniques

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Chemical Labeling Strategy for Generation of Internal Standards for Targeted Quantitative Proteomics

Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 *

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS

Agilent ESI and APCI sources: for polar to non-polar compounds

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Agilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Translational Biomarker Core

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Protein Quantitation II: Multiple Reaction Monitoring. Kelly Ruggles New York University

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Protein Quantitation II: Multiple Reaction Monitoring. Kelly Ruggles New York University

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

6 x 5 Ways to Ensure Your LC-MS/MS is Healthy

How proteins separate on reverse-phase HPLC

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Peptide Targeted Quantification By High Resolution Mass Spectrometry A Paradigm Shift? Zhiqi Hao Thermo Fisher Scientific San Jose, CA

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

SRM assay generation and data analysis in Skyline

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Tutorial 1: Setting up your Skyline document

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

Quantification Strategies Using the High Sensitivity Protein 250 Assay for the Agilent 2100 Bioanalyzer. Technical Note. Abstract

UPLC Intact MASS Analysis Application Kit

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Calibration in Proteomics. Proteomics 202 :: Practical Proteomics Using the Skyline Software Ecosystem Lindsay K. Pino Monday, Jan 22

Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Protocol. Product Use & Liability. Contact us: InfoLine: Order per fax: www:

A High Sensitivity Dual Solid Phase Extraction LC-MS/MS Assay for the Determination of the Therapeutic Peptide Desmopressin in Human Plasma

A Sensitive and Reproducible Signature Peptide MRM Based Quantitation Method for Pegylated Interferon α-2b in Serum

The Power of LC MALDI: Identification of Proteins by LC MALDI MS/MS Using the Applied Biosystems 4700 Proteomics Analyzer with TOF/TOF Optics

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Agilent 6400 Series Triple Quadrupole LC/MS/MS Users Session

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Reagents. Affinity Tag (Biotin) Acid Cleavage Site. Figure 1. Cleavable ICAT Reagent Structure.

Analysis of Labeled and Non-Labeled Proteomic Data Using Progenesis QI for Proteomics

The new Water Screening PCDL

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

WADA Technical Document TD2003IDCR

Separation of Large and Small Peptides by Supercritical Fluid Chromatography and Detection by Mass Spectrometry

Quantitation of Cystine and Identification of Related Metabolites in White Blood Cells Using a High Resolution Accurate Mass LC/MS Approach

Ultrafast Analysis of Methadone and Metabolite EDDP in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Proven robustness for large-scale metabolomics studies using the Phenomics Workhorse

Application Note. Abstract. Authors. Environmental

Analyst Software. Peptide and Protein Quantitation Tutorial

Agilent All Ions MS/MS

LC and LC/MS Column Selection Flow Chart

Transcription:

The Agilent 495 Triple Quadrupole LC/MS: Peptide Quantitation Performance Technical Overview Introduction Sample complexity and the low concentration of certain biomarkers are the major challenges encountered when performing biomarker verification using multiple reaction monitoring (MRM)-based methods. For this reason, the development of MRM-based analytical methods using stable isotope-labeled standards for the quantitation of peptides in biological fluids has focused on improving method sensitivity and increasing dynamic range by lowering detection limits. Robustness and reproducibility are essential when deploying these methods in high-throughput translational research environments. This Technical Overview presents the performance of the Agilent 495 Triple Quadrupole LC/MS for the quantitation of peptides. Sensitivity performance for a synthetic peptide standard was evaluated in both standard flow and nanoflow configurations. The Agilent Jet Stream ionization source was used for standard-flow chromatography, and the Agilent 120 Infinity HPLC-Chip/MS system was used for nanospray LC/MS. Instrument robustness under challenging conditions was evaluated in the standard flow configuration with direct injection of plasma digest. The instrument was not cleaned or tuned during the three and one half week robustness test. To assess overall quantitative performance, data from the 495 Triple Quadrupole LC/MS system were submitted to the MRM Protein Quantitation contest held in conjunction with the Mikromethoden in der Proteinchemie Meeting of July 2014. Contest samples provided by the University of Victoria - Genome BC Proteomics Centre were comprised of lyophilized mouse plasma digests spiked with the light/heavy peptide pairs of six proteins. The results from the participating labs were judged based on quantitative accuracy.

Overview: Agilent 495 Triple Quadrupole LC/MS Enhancements Built on Agilent dual-stage ifunnel technology, the 495 Triple Quadrupole LC/MS delivers attogram to zeptomole limits of detection (LOD) and limits of quantitation (LOQ) over a wide linear dynamic range. The instrument s remarkably low LODs are primarily due to enhanced ion utilization, starting with the ifunnel ion sampling through the high-transmission ion optics, and ending with the high efficiency detector. The key design features that boost performance for peptide-specific applications include: Quantitation of Peptide Standards in Human Plasma Sensitivity performance for quantitation of the synthetic peptide standard LVNEVTEFAK from human serum albumin (p/n G2455-85001) was evaluated in both standard flow and nanoflow LC/MS instrument configurations. To minimize loss of the target peptide due to adsorption, the standard curves for 5 amol to 5 pmol on-column were prepared in a simple background matrix of 10 fmol/µl of trypsinized enolase (Sigma, trypsinized using a 2,2,2-trifluroethanol-based protocol developed in-house). All data analysis, including standard curve generation, was performed using Agilent MassHunter WorkStation software. Standard Flow LC/MS System Evaluation To evaluate quantitative performance at standard flow rates, an Agilent 1290 Infinity Binary LC system was interfaced to a 495 Triple Quadrupole LC/MS using the Agilent Jet Stream ionization source. Separations were performed on an Agilent ZORBAX RRHD 300SB-C18 RRHD, 2.1 50 mm, 1.8 µm column using a five-minute LC/MS analysis performed in MRM mode. Following a blank injection to establish system cleanliness, replicate (n = 10) injections were made at 10 levels from 5 amol to 5 pmol (Table 1) to evaluate reproducibility. Retention time (RT) reproducibility was determined across all samples (n = 100), and area reproducibility was determined for each level. Enhanced Q1 ion optics that improve ion transmission and robustness, and reduce contamination Curved, tapered hexapole collision cell that improves MS/MS spectral fidelity New detector with high energy conversion dynode (HED) that improves ion detection efficiency and lowers noise at higher HED voltages (±20 kv) Table 1. Precision and accuracy for standard flow LC/MS analysis of the peptide standard LVNEVTEFAK in 10 fmol/µl trypsinized enolase, for all levels tested. Amount on-column %RSD (n = 10) % Accuracy 5 amol 14.0 109.8 7.5 amol 1.0 108.7 15 amol 9.4 105.0 30 amol 9.0 87.1 300 amol 1. 85.2 3 fmol 1.2 81.4 30 fmol 0. 8.4 300 fmol 0.7 87.4 3 pmol 2.1 105. 5 pmol 1.0 97.5 Extended mass range to 2,250 m/z for the analysis of higher molecular weight peptides such as those with modifications such as glycosylation To learn more about the Agilent 495 Triple Quadrupole LC/MS system download the brochure at http://www.agilent. com/cs/library/brochures/5991-4541en_495_qqq_brochure_ MedRes.pdf 2

Figure 1 and Table 1 show that the standard flow LC/MS system demonstrated outstanding quantitative performance with: Low attomole-level sensitivity with a lower limit of quantitation (LLOQ) of 5 amol on-column (Figure 1A) Six orders of linear dynamic range (5 amol 5 pmol on-column) with R 2 = 0.998 for LVNEVTEFAK (Figure 1B) Excellent precision and accuracy observed at all levels, including the LLOQ level (Table 1) Excellent RT reproducibility (RSD = 0.12 % for n = 100) 10 7 3.0 2.8 2. 2.4 2.2 2.0 1.8 1. 1.4 1.2 1.0 0.8 0. 0.4 0.2 0-0.2 Responses Counts Counts Counts Counts 3.0 amol 2.314 LOD 20 A Blank 5.0 amol LLOQ 7.5 amol 2.309 29 2.311 42 1. 1.8 2.0 2.2 2.4 2. 2.8 3.0 Acquisition time (min) B y = 5.88490*x 2.45328 R 2 = 0.997 0 5 10 15 20 25 30 35 40 45 50 Concentration (amol on-column 10 5 ) Responses 10 2 Zoom: 5 500 amol 2.8 y = 5.88490*x 2.45328 2.4 R 2 = 0.997 2.0 1. 1.2 0.8 0.4 0 5 9 13 17 21 25 29 Concentration (amol on-column) Figure 1. Standard flow LC/MS results for the peptide standard LVNEVTEFAK in 10 fmol/µl trypsinized enolase. A) Stacked extracted ion chromatograms showing the LOD and LLOQ. B) Standard curve with inset showing detail for the range of the curve from 5 to 500 amol. 3

Nanoflow LC/MS System Evaluation To evaluate quantitative performance at nanoflow rates, the 495 Triple Quadrupole LC/MS was interfaced to a 120 Infinity HPLC-Chip/MS system. Chromatography was performed using a Polaris-HR-Chip-3C18 HPLC-Chip with 20-minute LC/MS runs in MRM mode. Following a blank injection to establish a system background, replicate (n = 5) injections were made at seven levels from 500 zmol to 100 fmol (Table 2) to evaluate %RSDs. Retention time reproducibility was evaluated across all samples, and area reproducibility was calculated for each concentration level. Figure 2 and Table 2 show that the nanoflow LC/MS configuration provided the ultimate in sensitivity performance including: Zeptomole-level sensitivity, with an LLOQ of 500 zmol and an LOD of 250 zmol (Figure 1A) Greater than five orders of linear dynamic range (500 zmol 100 fmol on-column) with R 2 = 0.9999 for LVNEVTEFAK (Figure 1B) Excellent precision and accuracy observed at all levels, including the LLOQ (Table 2) Very good RT reproducibility (RSD = 1.4 % for n = 35) Table 2. Precision and accuracy for nanoflow LC/MS analysis of the peptide standard LVNEVTEFAK in 10 fmol/µl trypsinized enolase at all levels tested. Amount on-column %RSD (n = 5) % Accuracy 500 zmol 103.2 1 amol 8.5 80.2 10 amol 5.8 84. 100 amol 4.7 88.9 1 fmol 1.2 85. 10 fmol 1.1 97. 100 fmol 1.7 100.4 Counts Counts Counts Counts Responses.0 5.5 5.0.0 5.5 5.0.0 5.5 5.0.0 5.5 5.0 A Blank 250 zmol LOD 500 zmol LLOQ 1 amol 9.54 10 9.743 14 9.3 27 8.5 9.0 9.5 10.0 10.5 11.0 Acquisition time (min) 10 B 5.5 y = 54.025590*x 13.708172 5.0 R 2 = 0.9997 4.0 3.5 3.0 2.5 2.0 1.5 1.0 0.5 0 Responses 7 5 4 3 2 1 0 Zoom: 0.5 10 amol y = 54.025590*x 13.708172 R 2 = 0.9997 1 2 3 4 5 7 8 9 10 Concentration (amol on-column) 0 10,000 20,000 30,000 40,000 50,000 0,000 70,000 80,000 90,000 100,000 Concentration (amol on-column) Figure 2. Nanoflow LC/MS results for the peptide standard LVNEVTEFAK in 10 fmol/µl trypsinized enolase. A) Stacked extracted ion chromatograms showing the LOD and LLOQ. B) Standard curve with inset showing detail for the range of the curve from 0.5 to 10 amol. 4

Robustness: Complex Sample Analysis Robustness was evaluated using the standard flow instrument configuration through the direct injection of plasma digest (no SPE cleanup nor immunodepletion) onto an Agilent AdvanceBio Peptide Mapping 2.1 100 mm, 2.7 µm column without column eluent diversion at the start or end of analyses. Protein assay kits (MRM Proteomics from Cambridge Isotope Laboratories) were purchased to provide standard plasma digest and stable-labeled synthetic peptides for 42 proteins at different concentrations in plasma. Both a daily QC kit and a monthly calibration curve kit were used. To assess system robustness, the QC samples were interspersed between sets of hundreds of injections of a high load of plasma digest (40 µg per injection). The recommended kit gradient was used for the QC samples. To accelerate the testing to three and one half weeks, a rapid 15-minute LC/MS method was used to analyze the plasma samples. Over the course of the test, the instrument was not cleaned or tuned, and because of the highly reproducible standard flow chromatography, there was no adjustment of scheduled retention time windows in the dynamic MRM (DMRM) analytical method. The Agilent DMRM algorithm automatically constructs DMRM timetables for multiple analytes throughout the LC/MS analysis based on the retention time window for each analyte, an advantage for high-throughput protein and peptide quantitation. Figure 3 shows the results for selected peptides in the QC sample, normalized to the response obtained on the first day of the experiment. The system demonstrated excellent robustness with: Extremely stable retention times (average of 0.4 %RSD for all peptides across all runs) No significant signal degradation observed after 853 injections of 40 µg of plasma digest per injection over three and one half weeks of operation Stable response (%RSD = 15 for all peptides over all runs) Percent of first day response 140.00% 120.00% 100.00% 80.00% 0.00% 40.00% 20.00% Apolipoprotein E L-selectin Plasminogen Albumin serum Kininogen Transthyretin Hemopexin 0.00% 01/0/14 01/11/14 01/1/14 01/21/14 01/2/14 01/31/14 02/05/14 Figure 3. Results for selected peptides in the QC sample, normalized to Day 1 response. 5

Overall Performance: Analysis of Real-world Samples To assess the accuracy of quantitative performance for the analysis of a typical biological sample, data from the 495 Triple Quadrupole LC/MS system were submitted to the MRM Protein Quantitation contest held in association with the Mikromethoden in der Proteinchemie Meeting of July 2014. Contest samples provided by the University of Victoria Genome BC Proteomics Centre were comprised of lyophilized mouse plasma digests (n = 5) spiked with light/heavy peptide pairs of six proteins at different concentrations. The contest rules specified sample rehydration, transitions, and analysis using LC-MRM/MS. The data analysis strategy was chosen at the contestants discretion, and participants were asked to report both the light natural (NAT) and heavy stable isotopic labeled standard (SIS) peptide peak areas as well as the NAT concentration. The results from the seven sites/systems that participated in the contest were judged based on their quantitative accuracy. Sample preparation Rehydrated raw mouse plasma was diluted with 25 mm ammonium bicarbonate. Proteins in the diluted plasma were denatured (deoxycholate), reduced (TCEP), alkylated (iodoacetamide), and tryptically digested. The digests were then spiked with a constant amount of each NAT peptide and variable amounts of SIS peptide for the six proteins (Table 3) for the creation of standard curves. The samples were extracted using solid phase extraction (SPE), concentrated, and desalted. LC-MRM/MS was performed in the 495 Triple Quadrupole LC/MS standard flow and nanoflow configurations described above. The NAT and SIS transitions were found and quantified in the same plasma mix. The calculated natural/endogenous concentrations were correlated to the reference peptide to assess quantitative accuracy. Table 3. Standard curve concentration levels for the contest samples. Contest results Table 4 presents the actual NAT concentration and the Agilent contest results for the target peptides. The 495 Triple Quadrupole LC/MS systems yielded superior quantitative accuracy for four of the six peptides. With 95.30 % accuracy and 3.3 % CV, the Agilent standard flow system produced the best contest results. The Agilent nanoflow system placed second overall. UniProt Acc. Concentration (fmol/µg of total protein) Protein No. Peptide Std A Std B Std C Std D Std E Heparin Cofactor 2 P0554 TLEAQLTPR 20 40 100 200 1,000 Adiponectin Q15848 GDIGETGVPGAEGPR 0.8 1. 4 8 40 Apolipoprotein E P0249 LGPLVEQGR 0.8 1. 4 8 40 Leptin P41159 INDISHTQSVSSK 0.032 0.04 0.1 0.32 1. Plasminogen P00747 LFLEPTR 0.032 0.04 0.1 0.32 1. Clusterin P10909 ELDESLQVAER 0.032 0.04 0.1 0.32 1. Table 4. Agilent contest results for the target peptides. Standard flow Nanoflow Peptide Actual concentration (fmol/µg) Concentration determined (fmol/µg) Quantitative accuracy (%) Concentration determined (fmol/µg) Quantitative accuracy (%) GDIGETGVPGAEGPR 2.49 2.37 95.20 2.30 92.23 LGPLVEQGR 5.49 5.38 97.90 5.42 98.70 ELDESLQVAER 0.45 0.42 92.05 0.37 82.34 TLEAQLTPR 32.79 92.21 91.42 5.93 95.02 INDISHTQSVSSK 0.09 0.09 99.49 0.09 99.49 LFLEPTR 0.03 0.03 95.71 0.03 95.71 Average 95.30 93.92

Conclusion The sensitivity, dynamic range, robustness, and reproducibility of MRM-based LC/MS methods for the quantitation of peptides in biological fluids are important considerations for high-throughput biomarker verification applications. This Technical Overview describes the performance of the Agilent 495 Triple Quadrupole LC/MS for the quantitation of peptides in plasma for each of these considerations. Sensitivity performance for the quantitation of a tryptic peptide standard in plasma was evaluated in both standard flow and nanoflow system configurations. The 495 Triple Quadrupole LC/MS with Agilent Jet Stream Technology provided outstanding sensitivity at standard flow rates with an LLOQ of 5 amol on-column. The highest sensitivity, 500 zmol LLOQ, was achieved at nanoflow rates using the Agilent Infinity HPLC-Chip/MS technology. For the analysis of a complex plasma digest matrix over an extended period of three and one half weeks, the 495 Triple Quadrupole LC/MS maintained excellent response and RT reproducibility, without cleaning or tuning. Comparative quantitative performance was assessed through participation in the MRM Protein Quantitation contest held in association with the Mikromethoden in der Proteinchemie Meeting. Results from LC-MRM/MS analyses of contest samples of lyophilized mouse plasma digests spiked with light/heavy peptide pairs of six proteins were judged based on quantitative accuracy. The 495 Triple Quadrupole LC/MS produced superior results for four of the six peptides, and produced the highest average quantitative accuracy for the six proteins measured. The unequalled performance of the 495 Triple Quadrupole LC/MS in peptide quantitation applications is the result of design enhancements including increased precursor ion transmission, improved MS/MS spectral quality, enhanced detection efficiency (HED voltage up to 20 kv), and extended m/z range. 7

www.agilent.com/chem For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc., 201 Published in the USA, June 1, 201 5991-898EN