Simultaneous quantification of peptides and phosphopeptides by caplc-icp- MS using the Agilent 8800/8900 Triple Quadrupole ICP-MS

Similar documents
Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Application note. Accurate determination of sulfur in biodiesel using Isotope Dilution-Triple Quadrupole ICP-MS (ID-ICP-QQQ) Petrochemical

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Agilent 8900 Triple Quadrupole ICP-MS

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Agilent s Solutions for Nanoparticles Analysis ICP-MS, ICP-MS/MS, software. Alain Desprez Agilent Technologies

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

IMPROVING DATA QUALITY WITH ADVANCED INSTRUMENTATION UTILIZING MULTIPLE INTERFERENCE REDUCTION TECHNOLOGIES

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Quantitation of a target protein in crude samples using targeted peptide quantification by Mass Spectrometry

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Supporting Information

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline

JRP SIB09 Elements GOOD PRACTICE GUIDE ON DEALING WITH INTERFERENCES IN ICP-MS FOR INVESTIGATING HIGH PURITY MATERIALS. Deliverable 1.2.

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Analysis of selenoproteins in rat serum by Triple Quadrupole ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS

materials analysis Solutions for Your Analytical Business Markets and Applications Programs

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Accurate Determination of TiO 2 Nanoparticles in Complex Matrices using the Agilent 8900 ICP-QQQ

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

A rapid and highly selective colorimetric method for direct detection of tryptophan in proteins via DMSO acceleration

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Agilent ESI and APCI sources: for polar to non-polar compounds

An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining

Københavns Universitet

Application Note. Edgar Naegele. Abstract

Highly sensitive quantitative estimation of genotoxic impurities from API and drug formulation using LC/MS/MS

Simultaneous Arsenic and Chromium Speciation by HPLC/ICP-MS in Environmental Waters

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

A Q-TOF Generated, Metabolomics- Specifi c LC/MS/MS Library Facilitates Identifi cation of Metabolites in Malaria Infected Erythrocytes

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Fast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns

Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application

Mass Spectrometry. Quantitative Mass Spectrometry Chiral Mass Spectrometry

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

The Agilent 6495 Triple Quadrupole LC/MS: Peptide Quantitation Performance

Translational Biomarker Core

TANDEM MASS SPECTROSCOPY

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Transcription:

Simultaneous quantification of peptides and phosphopeptides by -ICP- MS using the Agilent 8800/8900 Triple Quadrupole ICP-MS Application note Proteomics Authors Silvia Diez Fernández 1, Naoki Sugiyama 2, Jorge Ruiz Encinar 1, and Alfredo SanzMedel 1 1 Department of Physical and Analytical Chemistry, University of Oviedo, Julián Clavería 8, 33006 Oviedo, Spain 2 Agilent Technologies, Tokyo, Japan Abstract In combination with capillary LC () separation, the Agilent 8800 Triple Quadrupole ICP-MS, operating in MS/MS mass shift mode, was used for the trace measurement of phosphorus (P) in phosphopeptides and sulfur (S) in S-containing peptides. The lowest absolute detection limits ever reported for the analysis of S- and P-containing species (11 fmol and 6.6 fmol, respectively) using LC-ICP-MS were achieved. Effective interference removal was confirmed by good agreement with the theoretical isotope ratio for S. Excellent peak profiles and signal to noise (S/N) ratios were observed for both P and S. The great potential of triple quadrupole -ICP-MS for sensitive and simultaneous absolute quantitation of P- and S-containing peptides using non-specific standards is demonstrated.

Introduction LC-MS/MS is used for the quantification of target proteins in pharma/biopharma and clinical research. The approach generally relies on the use of synthetic isotopically labeled proteins and peptides, which are used as internal standards for the quantification of the corresponding target compound. In contrast, LC ICP-MS enables non-species specific quantitation of proteins and peptides by measuring the heteroatoms S and P contained in the target compounds. In this way, different compounds containing the heteroatom may be used and thus can be quantified using a single nonproteinaceous heteroatom-containing compound as a generic standard. Unfortunately, detection limits (DLs) for P and S are compromised by their high ionization potential and by multiple polyatomic interferences with conventional quadrupole ICP-MS (ICP-QMS) even when a collision/reaction cell (CRC) is employed. Sector field high resolution ICP-MS (HR-ICP-MS) has also been applied, but so far no ICP-MS instrument has achieved DLs (especially for sulfur) equivalent to compound DLs reported using ESI-MS/MS with isotopically labeled synthetic peptides. Table 1 lists published data for P and S DLs using LC-ICP-MS, showing the ICP-MS technology and separation technique used. In this work, the Agilent 8800 Triple Quadrupole ICP-MS (also referred to as ICP-QQQ) was applied to the determination of proteins and peptides by the absolute quantitation of P and S heteroatoms. Compared to conventional or ICP-QMS, the Agilent 8800/8900 features an additional quadrupole mass filter (Q1), situated in front of the Octopole Reaction System (ORS) cell and quadrupole mass filter (now called Q2). In this tandem MS configuration, Q1 operates as a mass filter, allowing only the target analyte mass to enter the cell, and rejecting all other masses. Because plasma and matrix ions are eliminated by Q1, this ensures that the reaction processes in the ORS are precisely controlled and so accurate measurements can be performed even in complex, high matrix samples, while sensitivity is significantly increased. Chromatographic separation was performed by an Agilent cap-lc system coupled to the 8800 ICP-QQQ. Experimental Reagents and materials Ultra pure water (Millipore, Tokyo, Japan), HPLC and analytical reagent grade acetonitrile and formic acid (Wako Pure Chemical, Ltd., Osaka, Japan). ICP-MS standards (SPEX CertiPrep, Inc., Metuchen, NJ, USA). Bis(4-nitrophenyl) phosphate (BNPP, 99% purity) and methionine ( 99% 82 purity) (Sigma Aldrich, Steinheim, Germany). Amino acid sequences of the phosphopeptide standards were LRRApSLG and KRSpYEEHIP, and the S- containing peptide standards were ACTPERMAE and VPMLK. All peptides were purchased from AnaSpec (Fremont, CA, USA) with purity 95%. Table 1. Critical comparison between detection limits obtained for the analysis of S- and P-containing species using different ICP-MS instruments Phosphorus Instrument Separation technique DL (compound) Reference Separation technique DL (compound) Reference ICP-QMS Collision cell (He) ICP-QMS Collision cell (Xe) LC CE 110 fmol 0.97 pmol 63 fmol 19 fmol 1 2 3 2 nanolc CE ICP-QMS DRC (O 2 ) LC 2.2 pmol 4 microlc microlc LC HR-ICP-MS 100 fmol 50 fmol 5 10 microlc Sulfur 1.5 pmol 2.6 pmol 2.7 pmol 3.1 pmol 2.3 pmol 9 pmol 11 pmol 10 pmol ICP-MS Triple Quad 6.6 fmol This paper 11 fmol This paper 6 7 8 9 4 11 12 13 2

Preparation of P- and S-containing standards Calibration standards containing P and S at 0, 25, 50, 100 and 200 ng/ml (as the element) using BNPP and methionine were prepared in a matrix of 1% mobile phase B (which was the starting point of the LC gradient). Preparation of the phosphopeptide solutions To confirm the retention time and purity of each peptide, three separate solutions (100 ng/ml, as the element) were prepared. Mixture 1 contained LRRApSLG and ACTPERMAE and mixture 2 contained KRSpYEEHIP and VPMLK. Both mixtures contained approximately 100 ng/ml of BNPP and methionine. Mixture 3 was prepared after 1:1 mixing of mixtures 1 and 2 by weight. Instrumentation system An Agilent 1200 Series with Agilent Zorbax SB C18, 5 µm, 150 x 0.3 mm reversed phase capillary column was used. Mobile phases A and B consisted of water and acetonitrile respectively. Both phases contained 0.1% formic acid and 10 ng/ml Ge (used as ICP-MS ISTD). Injection volume was 1-2 µl. LC flow rate was 5 µl/min, and the LC gradient was 0-3 min: 1 % B, isocratic. 3-35 min: 1-60% B linear. ICP-MS An Agilent 8800 Triple Quadrupole ICP-MS was used. Oxygen (Ar:O 2 mixture, 8:2) was added to the nebulizer gas at 80 ml/min. Cell gas: oxygen at 0.35 ml/min. Sample flow rate 5 µl/min. Integration time was 150 ms (one point per peak) for P, S and 34 S. The Agilent 8800 was optimized for lowest P and S DLs using a standard containing 100 ng/ml P and S in 30% acetonitrile with a flow rate of 5 µl/min, introduced with a syringe pump. Operation of the LC and integration of the chromatographic peaks was performed using the Agilent ICP-MS MassHunter software. Removal of interferences on S and P using the Agilent 8800/8900 in MS/MS mode The ICP-QQQ was operated in MS/MS mass shift mode to remove interferences on P and S. In Figure 2, the upper diagram (a) shows Q1 set to m/z which allows S + to pass through to the cell, rejecting interferences at all other masses (including those at m/z 48). S + is converted to SO + in the cell by reaction with O 2. Q2 is set to m/z 48, which allows SO + to pass to the detector, rejecting all the original interferences on S at m/z. Interferences at m/z 48 were previously rejected by Q1. This is the key to the interference removal power of the QQQ configuration it allows controlled, accurate, and interference-free measurement of reaction product ions. -ICP-MS interface The column was connected to the ICP-MS using the Agilent interface (Agilent G3680A capillary LC interface kit, Figure 1), which consists of a total consumption nebulizer inside a small quartz spray chamber. Figure 1. Agilent capillary LC interface kit (G3680A) 3

The lower diagram (b) in Figure 2 shows the settings for P determination: Q1 is set to m/z, which allows P + to pass through, rejecting interferences at all other masses (including m/z 47). P + is converted to PO + in the cell by reaction with O 2. Q2 is set to m/z 47 to allow the reaction product PO + to pass to the detector. Interferences at m/z 47 were previously rejected by Q1. The interferences removed by Q1: ( 15 N 16 O +, 14 N 16 OH +, 14 N 17 O +, 12 C 18 OH + on P + and 16 O 2+, 15 N 17 O +, 15 N 16 OH +, 14 N 18 O + on S + ) and Q2 ( 47 Ti +, S 14 NH +, S 15 N +, 15 N 16 O 2+, 12 C 35 Cl + on P 16 O + at m/z 47, and 48 Ti +, 36 Ar 12 C +, P 16 OH +, P 17 O +, 48 Ca + on SO + at m/z 48), are shown in each diagram. Note also that without the presence of Q1, reaction product ions derived from P and S would interfere with each other, which could affect P/S ratios, measurement of P- and S-containing species (leading to errors in the calculation of the degree of phosphorylation), and S isotope ratios (e.g. leading to errors in isotope dilution calculations). Using the 8800/8900 in MS/MS mass shift mode, reaction product ions retain the correct isotopic pattern of the parent ions this is not the case in reaction cell ICP QMS. In ICP-QMS there is no Q1, so all isotopes of S enter the cell and react to produce SO reaction product ions which overlap each other (e.g. S 18 O + overlaps the 34 S 16 O + product ion at m/z 50). Initially, DLs were measured in a high organic matrix (30% acetonitrile, delivered continuously via the syringe pump) to assess the interference removal efficiency of the Agilent 8800. DLs achieved were 0.6 ng/ml for P and 1.2 ng/ml for S with BECs of 9 ng/ml for P and 5 ng/ml for S. These were significantly lower in aqueous solution, which suggests the presence of P and S in the acetonitrile. To confirm interference removal on S, the 34/ ratio was measured. Since interferences are much higher on S, a 34/ ratio matching the theoretical value for 34/ S indicates effective removal of interferences. The 34/ ratio in the blank (0.0484 ±0.0017) agreed well with the standard (0.0487 ±0.0012). The IUPAC value for 34/ S is 0.0447 ±0.0025. The difference between observed and theoretical is due to mass bias (the 8800 ratios were not mass bias corrected). The -4.0% to -4.3% mass discrimination factor matches that reported in the literature [14-16], demonstrating that all interferences on S are effectively removed by the 8800, and that S is present in the acetonitrile. (a) Oxygen 16 + O 2 S + 48 Ca +, 36 Ar 12 C +, other ions (b) SO + Q1= cell Q2=48 detector Oxygen 15 N 16 O +, 14 N 16 OH + P + 47 Ti +, 12 C 35 Cl +, other ions PO + Q1= cell Q2=47 detector Figure 2. 8800/8900 ICP-QQQ MS/MS operation in mass shift mode to remove interferences on S (a) and P (b) 4

Results and discussion Switching to -ICP-MS operation, calibration standards containing 0, 25, 50, 100 and 200 ng/ml of both P and S (BNPP and methionine, respectively) were injected and measured: excellent linearity and RDSs of <4% were obtained (Figure 3). The chromatogram obtained for the 50 ng/ml standard is shown in Figure 4. Signal to noise (S/N) ratio observed was excellent for both elements. In LC ICP MS operation, the background due to P and S in the acetonitrile mobile phase can be effectively subtracted, allowing the detection power of the Agilent 8800 to be accurately measured. DLs obtained were 0.10 ng/ml P and 0.18 ng/ml S (6.6 fmol P, 11 fmol S absolute). We believe these to be the lowest absolute DLs ever reported for S and P by LC-ICP-MS, as summarized in Table 1. Measurement of phosphopeptides and S-containing peptides The LC-ICP-MS system was then used to measure phosphopeptides and S-containing peptides. The retention time for each peptide was calculated by two different solutions mixtures 1 and 2 as described in the experimental section. Mixture 3, which contained approximately 45 ng/ml of each of the peptides and 105 ng/ml (as P and S) of the standard was measured under the same instrumental conditions. The chromatogram obtained from mixture 3 is shown in Figure 5. Both peptides and the standard for both P and S are clearly shown. Only the ptyr-containing peptide exhibited a small shoulder, likely corresponding to a phosphopeptide impurity, accounting for 9% of the signal. S/N ratios observed were excellent (no smoothing performed). The average DL for the S-containing peptides (0.26 ng/ml S), was slightly higher than that obtained for the methionine standard (0.18 ng/ml S), possibly due to slightly wider base peak widths (24 s) than those observed for the methionine standard (9 s). In the case of P, the average detection limit obtained (0.33 ng/ml P) was significantly higher than that obtained for the BNPP standard (0.10 ng/ml P), because phosphorus suffers a sensitivity enhancement in the presence of carbon (gradient elution). Figure 3. Calibration plots for S and P (methionine and BNPP, respectively) at 25, 50,100 and 200 ng/ml (as element). Injection volume = 2 μl, flow rate = 5 µl/min, gradient given in text. 5

250000 200000 P 47 PO + 150000 S 48 SO + Cps Cps 100000 50000 0 0 500 1000 1500 2000 (sec.) Figure 4. Chromatogram obtained for the 50 ng/ml standard 500000 450000 400000 350000 300000 250000 200000 150000 100000 50000 P 47 PO + S 48 SO + 0 0 500 1000 1500 2000 (sec.) Figure 5. Chromatogram of mixture 3 containing LRRApSLG, ACTPERMAE, KRSpYEEHIP and VPMLK plus BNPP and methionine 6

Conclusions Highly sensitive and interference free quantitation of ICP-MS-detectable elements naturally present in proteins will expand the use of LC-ICP-MS in proteomics. In combination with cap-lc separation, the Agilent 8800 Triple Quadrupole ICP-MS generated the lowest absolute detection limits ever reported for the analysis of S- and P-containing species (11 fmol and 6.6 fmol, respectively) using LC-ICP-MS. A significantly lower detection limit for sulfur should be achievable using the Agilent 8900 ICP-QQQ #100 (Advanced Applications Configuration), as this newer instrument has a re-engineered argon gas flow path that utilizes low sulfur (and silicon) materials. This has the effect of minimizing the elemental background for these two elements. Operating in MS/MS mass shift mode, the Agilent 8800/8900 eliminates interferences on both P and S by reacting the analytes with O 2 and measuring the oxide ions produced. Effective interference removal on S was confirmed by good agreement with the theoretical isotope ratio for S, and because correct isotope patterns are retained with QQQ MS, isotope dilution calibration can be used to correct for S sensitivity changes due to gradient dilution can be performed. Excellent peak profiles and S/N ratios were observed for both S and P species and peptides. This is the first time that generic standards have been applied to the simultaneous absolute quantification of peptides and phosphopeptides. It is envisaged that this new powerful approach for P and especially for S detection will lead to the use of LC ICP-MS in fields such as pharmaceutical research (drugs and metabolites), environment analysis (pesticides) and nanotechnology (characterization of engineered nanoparticles). More Information For a full account of this application see publication: Triple Quad ICPMS (ICPQQQ) as a New Tool for Absolute Quantitative Proteomics and Phosphoproteomics, Silvia Diez Fernández, Naoki Sugiyama, Jorge Ruiz Encinar, Alfredo Sanz-Medel, Anal. Chem., 2012, 84 (14), pp 5851 5857, Publication Date (Web): June 18, 2012 References 1. Pereira Navaza, A.; Ruiz Encinar, J.; Sanz-Medel, A. Angew. Chem. Int. Ed. 2007, 46, 569-671. 2. Pröfrock, D.; Leonhard, P.; Prange, A. J. Anal. At. Spectrom. 2003, 18, 708-713. 3. Pröfrock, D.; Leonard, P.; Ruck, W.; Prange, A. Anal. Bioanal. Chem. 2005, 381, 194-204. 4. Smith, C. J.; Wilson, I. D.; Weidolf, L.; Abou-Shakra, F.; Thomsen, M. Chromatographia. 2004, 59, S165-S170. 5. Wind, M.; Edler, M.; Jakubowski, N.; Linscheid, M.; Wesch, H.; Lehmann, W. D. Anal. Chem. 2001, 73, 29-35. 6. Schaumlöffel, D.; Giusti, P.; Preud Homme, H.; Szpunar, J.; Lobinski, R. Anal. Chem. 2007, 79, 2859 2868. 7. Pröfrock, D.; Leonhard, P.; Prange A. Anal. Bioanal. Chem. 2003, 377, 1-139. 8. Hann, S.; Koellensperger, G.; Obinger, C.; Furtmüller, P. G.; Stingeder, G. J. Anal. At. Spectrom. 2004, 19, 74-79. 9. Stürup, S.; Bendahl, L.; Gammelgaard, B. J. Anal. At. Spectrom. 2006, 21, 201 203. 10. Zinn, N.; Hahn, B.; Pipkorn, R.; Schwarzer, D.; Lehmann, W. D. J. Proteome Res. 2009, 8, 4870 4875. 11. Hann, S.; Koellensperger, G.; Obinger, C.; Furtmüller, P. G.; Stingeder, G. J. Anal. At. Spectrom. 2004, 19, 74 79. 12. Zinn, N.; Krüger, R.; Leonhard, P.; Bettmer, J. Anal. Bioanal. Chem. 2008, 391, 537-543. 7

13. Wind, M.; Wegener, A.; Eisenmenger, A.; Kelner, R.; Lehmann, W. D. Angew. Chem. Int. Ed. 2003, 42, 3425 3427 14. Clough, R.; Evans, P.; Catterick, T.; Evans, E. H. Anal. Chem. 2006, 78, 6126-61. 15. Becker, J. S. J. Anal. At. Spectrom. 2002, 17, 1172 1185. 16. Mason, P. R. D.; Košler, J.; de Hoog, J. C. M.; Sylvester, P. J.; Meffan-Main, S. Anal. At. Spectrom. 2006, 21, 177 186. www.agilent.com/chem For Research Use Only. Not for use in diagnostic procedures. This information is subject to change without notice. Agilent Technologies, Inc. 2016 Published June 1, 2016 Publication number: 5991-1461EN