EA-IRMS: Fast and Precise Isotope Analysis of Liquids on a Delta V Isotope Ratio MS with a High Temperature Conversion Elemental Analyzer

Similar documents
Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Finnigan TC/EA. High Temperature Conversion Elemental Analyzer

Thermo Scientific GasBench II

Thermo Scientific FLASH HT Plus Elemental Analyzer for IRMS

High-temperature elemental analysis and pyrolysis techniques for stable isotope analysis

Basic concepts and the Use of Reference Materials in Stable Isotope Metrology

Determination of Volatile Substances Proof of Food Adulteration

δ 13 C of DIC and carbonate samples: comparison of traditional mass spectrometry methods with an infrared spectrometry method

Food authenticity and adulteration testing using trace elemental and inorganic mass spectrometric techniques

Forensic Applications of Isotope Ratio Mass Spectrometry

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Isotope Ratio Mass Spectrometric Method for the On-line Determination of Oxygen-18 in Organic Matter

GC-MS-IRMS: Undisputable results by coupling of GC-IRMS with high-resolution mass spectrometry for final confirmation in sports drug testing

METHOD VALIDATION REPORT

Determination of Carbon d13c isotopic composition from DIC samples

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Chemistry Gas Chromatography: Separation of Volatile Organics

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

ConFlo III An Interface for High Precision d 13 C and d 15 N Analysis with an Extended Dynamic Range

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

costech instruments ECS 4010 Elemental Combustion System CHNS-O

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Stable isotope analytical methods

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Speciation and δ 34 S analysis of volatile organic compounds in crude oil by GC-MC-ICP-MS using the Thermo Scientific GCI 300 Interface

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Gas Chromatography (GC)

Prima PRO Process Mass Spectrometer

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Agilent G3212 GC-APCI Source

RESOLUTION OIV-OENO

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

of mass spectrometry

Structural characterization begins with a purity check!

Addressing adulteration with stable isotopes.

Multiphase Oscillatory Flow Strategy for in Situ Measurement and Screening of Partition Coefficients

Thermo Scientific LC IsoLink

*Correspondence to:

ION CHROMATOGRAPHY SYSTEM S 150

METHOD VALIDATION REPORT

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Thermo Scientific 253 Plus 10 kv Isotope Ratio MS. Utilize the Gold Standard. in isotope ratio MS. Robustness Sensitivity Precision

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Accurate and Precise Sulfur Analysis using the Thermo Scientific FlashSmart Elemental Analyzer

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Dr. Jonathan Gutow Fall Looking for PCBs in Water or Can PCBs Wash out of Landfills and Contaminate Ground Water?

Thermo Scientific DELTA V Advantage DELTA V Plus

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

TR ACE ORGANIC. lgcstandards.com/foodandenvironment

Routine MS Detection for USP Chromatographic Methods

Nitrogen/Protein determination of infant food by the Thermo Scientific FlashSmart Elemental Analyzer using helium or argon as carrier gases

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Headspace Technology for GC and GC/MS: Features, benefits & applications

ECS8020 ORGANIC ELEMENTAL ANALYZER CHNS-O Analyzer

Analysis of trace elements in polymer samples to ensure compliance with WEEE-RoHS standards using the Thermo Scientific icap 7200 ICP-OES

Defining quality standards for the analysis of solid samples

S2 PICOFOX. Innovation with Integrity. Spectrometry Solutions TXRF

Rapid Catalyst Screening Reactors

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

ABB Los Gatos Research Triple Isotope Water Analyzer (TIWA)

Supporting Information

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

Expanded capabilities in ammonium and amine detection

Supporting Information. Electrochemical Reduction of Carbon Dioxide on Nitrogen-Doped Carbons: Insights from Isotopic Labeling Studies

IRMS. perspective. Isotope Ratio Mass Spectrometry.

Benzoylecgonine in Urine by SAMHSA GC/MS

Application Note PB 401. Improve HPLC Sample Preparation in an Analytical Laboratory A New Automated Sample Preparation Process

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Elemental Analysis: CHNS/O characterization of polymers and plastics

Quantification of Rice Aroma, 2-Acetyl-1-Pyrroline (2-AP), Using TurboMatrix Headspace Trap Coupled with GC/NPD and GC/MS

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

ChromTech options for PAL systems

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Application Note LCMS-112 A Fully Automated Two-Step Procedure for Quality Control of Synthetic Peptides

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

PERSPECTIVE STABLE ISOTOPE RATIO MS

GC/MS Analysis Laboratory Report

(CATION EXCHANGE AND LN RESIN, WITH VACUUM BOX SYSTEM)

APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEY WORDS. Simultaneous high-throughput screening of melamine (MEL) and cyanuric acid (CYA).

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

Uranium from water sample

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

Chapter 27: Gas Chromatography

CHAPTER 6 GAS CHROMATOGRAPHY

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Influence of the Temperature on the Liquid-Liquid- Solid Equilibria of the Water + Ethanol + 1-

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Transcription:

APPLICATION NOTE EA-IRMS: Fast and Precise Isotope Analysis of Liquids on a Delta V Isotope Ratio MS with a High Temperature Conversion Elemental Analyzer AN30180 Oliver Kracht, Andreas Hilkert, Thermo Fisher Scientific, Bremen, Germany Key Words EA-IRMS, Isotope Ratio MS, TC/EA, Water Analysis Introduction 18 O and 2 H isotope ratio analysis of aqueous solutions such as water, urine, blood plasma, wine, can be performed by a variety of different techniques: equilibration, chromium reduction and carbon reduction (also referred to as high temperature conversion or pyrolysis). The high temperature conversion (TC) method allows a direct isotope ratio analysis of both oxygen and hydrogen in continuous flow mode. It offers the analysis of two isotopes in five minutes from only one injection of water samples as small as 0.1 μl. Analytical Method High temperature conversion in continuous flow mode became available with the introduction of the glassy carbon reactors. The conversion is based on the reaction: H 2 O + C H 2 + CO 1 Water samples are entrained in a He carrier gas stream and passed through a glassy carbon reactor at 1400 C. The tube-in-tube principle (glassy carbon tube in a ceramic tube, Figure 1) prevents oxidation of the glassy carbon by keeping the tube in an oxygenfree atmosphere, constantly flushed by a flow of He. Contamination or exchange of the O in the sample CO with O in the surrounding ceramic (Al 2 O 3 ) tube is prevented because the reaction takes place inside the glassy carbon tube. Isotope Ratio Analysis of Aqueous Solutions, e.g. Water Waste water Beer Wine Fruit juice Urine Blood plasma Ice Atmospheric water

Autosampler for Liquids He High Temperature Conversion Glassy Carbon Reactor Ceramic Tube Table 2. AS 3000 Autosampler settings. Analytical Conditions Injection Volume 0.1 μl Plunger Strokes 3 Rinses 0 Pre-Injection Dwell Time 0 s Post-Injection Dwell Time 16 s Solvent Wash Cycle 0 GC Column He+H 2 +CO He Dilution Open Split Figure 1. Schematic of the Thermo Scientific high temperature conversion elemental analyzer. The products of the reduction reaction, H 2 and CO, are separated on a gas chromatographic column and analyzed directly in a Thermo Scientific Delta V Isotope Ratio Mass Spectrometer. After collecting the ion beams of m/z 2 and m/z 3 in the H 2 peak a fast switch of the magnetic field allows the subsequent collection of m/z 28 and m/z 30 in the CO peak (dual measurement mode). A narrow stainless steel insert is used at the top of the reactor to ensure both rapid and low memory transfer of the water into the hot zone. The system settings are given in Table 1 and 2. It is essential that the injection of the water sample takes place immediately after the needle penetrates the septum to avoid isotopic fractionation resulting from evaporation in the needle tip. After the sample injection, the needle stays in the hot reactor for a certain period of time in order to remove any sample residue prior to the next injection. Table 1. High temperature conversion elemental analyzer (TC/EA) settings for H and O isotope ratio analysis. Analytical Conditions Reactor Temperature 1400 C GC Oven Temperature 100 C Carrier Flow 100 ml/min Autosampler Type Thermo Scientific AS 3000 Syringe Size 0.5 μl Results Accuracy Table 3 shows data from an installation of a TC/EA at a customer site. International water standards (IAEA, Vienna) plus a lab standard (Sample A) were injected in the given sequence. The water standards cover the full range of and values of natural water. The specifications of the TC/EA are < 2 for and < 0.2 for. The typical precision achieved during routine analysis can be < 1 for and < 0.1 for. All samples were analyzed in dual measurement mode. The uncorrected raw data in Table 3 show that there is no need for any post-acquisition evaluation or correction. The direct conversion of the sample in the reactor, together with a highly precise, small and stable H 3 + -factor of 6.2 ppm/na provides very accurate delta values. This is a clear advantage over equilibration techniques. Common approaches for post-evaluation of raw delta values like the SMOW/SLAP scaling and drift 2, 3, 4 corrections have been discussed by several authors. Memory In Table 3 only the first data point of each set of five injections shows the influence of preceding samples. Large differences between delta values of adjacent samples can corrupt the first analysis due to memory effects mainly caused by the syringe. The implementation of post-injection dwell time of the syringe in the reactor removes most of this memory. Washing cycles for the syringe do not help here because any solvent would contaminate the syringe with H- and O-isotopes. The data presented in Figure 2, Figure 3, Table 3, Table 4 and Table 5 are not warranted because they exceed product specifications. The warranted product specifications for is ±2 (1 sd) measured on 0.5 μl of water and for is ±0.2 (1 sd) measured on 0.5 μl of water.

Table 3. and Standard Deviation () of and values of five consecutive injections of international water standards GISP, SLAP, SMOW and a lab standard. The data is given in sequence of injection to demonstrate memory effects. The table shows raw data and data after scaling with SMOW and SLAP. The data was acquired during field installation. Strike-through data is not considered. GISP SLAP SMOW Sample A [ ] Raw [ ] After Scaling IAEA Accepted Value [ ] Raw [ ] After Scaling IAEA Accepted Value 185.03 24.23-186.95-24.67-186.37-24.75-187.09-24.75-186.62-186.76 0.32-188.99-189.5-24.84-24.75 0.07-24.73-24.5-420.53-55.16-422.13-55.36-422.35-55.42-424.45-55.89-422.67-422.90 1.06-428.00-428.0-55.52-55.54 0.24-55.50-55.5-1.76-0.32-0.29-0.04 0.31-0.06-0.02-0.02-0.11-0.03 0.25 0.00 0.0 0.13 0.01 0.09 0.00 0.0-47.10-7.44-42.47-6.99-41.40-6.82-41.32-6.66-40.11-41.32 0.96-41.79-6.76-6.81 0.14-6.81 Analysis Time Table 4 shows a sequence of injections with samples of Antarctic precipitation with similar isotope ratio values and Standard Deviation () were calculated from five consecutive injections per sample. The mean precision for is << 0.6 which is a requirement for ice core analysis by paleoclimatologists. High resolution data of long ice cores demand high sample throughput. Figure 2 and Figure 3 compare the mean value with the first injection of each set to see if the number of repetitions can be reduced without affecting the accuracy. All isotope values of the first injection fall within the except for Sample H showing that already the first injection gives correct results. About half of the values of the first injection fall outside the range requiring repetition. Table 4. and Standard Deviation () of and values of five consecutive sample injections (N = 5) with similar isotope ratio values. No data was excluded as memory effects are negligible. Standard: N = 3. Sample Standard -413.82 0.40-39.48 0.07 Sample A -429.78 0.43-41.46 0.03 Sample B -423.93 0.79-40.68 0.09 Sample C -422.30 0.55-40.52 0.08 Sample D -418.52 0.55-40.02 0.06 Standard -414.56 0.28-39.58 0.11 Sample E -402.63 0.48-37.94 0.04 Sample F -398.87 0.25-37.54 0.08 Sample G -392.22 0.88-36.56 0.25 Sample H -394.69 0.48-36.83 0.13 Standard -412.87 0.76-39.15 0.13

δ²h V-SMOW [ ] -390-395 -400-405 -410-415 -420-425 and 1st Injection of Standard: -413.75 ± 0.86-430 Std. Sa. Sa. Sa. Sa. Std. Sa. Sa. Sa. Sa. Std. A B C D E F G H Figure 2. mean (sample N = 5, Standard N = 3) with Standard Deviation. Red circle data points represent the first injection of each set, all first injections fall within the range except for Sample H. Intensity [mv] 12000 10000 8000 6000 19.10 58.68 4000 2000 0 99.62 188.69 278.49 m/z 2 m/z 3 0 50 100 150 200 250 300 350 400 Time [s] Figure 4. Single measurement with multiple injections of the same sample to account for two repetitions. The analysis time is six minutes. V-SMOW [ ] -36-37 -38-39 -40-41 and 1st Injection of Standard: -39.40 ± 0.22 Intensity [mv] 20000 16000 12000 8000 4000 30.35 83.06 163.04 296.08 m/z 2 m/z 3 m/z 28 m/z 29 m/z 30-42 Std. Sa. Sa. Sa. Sa. Std. Sa. Sa. Sa. Sa. Std. A B C D E F G H 0 0 50 100 150 Time [s] 200 250 300 350 Figure 3. mean (sample N = 5, standard N = 3) with Standard Deviation. Red circle data points represent the first injection of each set. Most first injections fall within the range. In Short Single Measurement 6 minutes analysis time, no repetitions for screening required (N = 1), 10 injection per hour, 240 samples per day Single Measurement 6-7 minutes analysis time, 2 repetitions (multiple injections, N = 3), 9 samples per hour, 216 samples per day See Figure 4 Figure 5. Dual measurement of and from a single injection with a fast magnet switch at 120 s. The analysis time can be reduced to less than five minutes. Table 5. and Standard Deviation () of and values of two standards and ethanol demonstrating the possibility to run organic liquids. N = 10 for standards, N = 5 for ethanol. Sample Lab Standard -53.69 0.20-7.84 0.04 V-SMOW 0.00 0.34 0.00 0.08 GISP -189.50 0.40-24.80 0.05 Ethanol -234.80 0.12-24.18 0.08 Lab Standard -53.76 0.39-7.85 0.03 Dual Measurement and 5-6 minutes analysis time, 2 repetitions (N = 3), 9 injections per hour, 72 samples per day (9 injections x 24 h/3(triplicate) = 72) See Figure 5

Other Liquid Samples While isotope ratio analysis of water is by far the predominant application, other substrates have been tested, including purified blood plasma, urine, and ethanol (see Table 5), see also Calderone et al. for ethanol applications. 5 The application lab of Thermo Fisher Scientific, Bremen, Germany, took part in an interlaboratory study, the object of which was to evaluate the use of tetramethylurea (TMU) as a reference compound for TC/EA-IRMS analysis of wine and alcohols as defined in European Commission Regulation (EEC) N 2676/90. Since 2003, the oven design and reactor technology of the TC/EA has been integrated in the Thermo Scientific EA IsoLink IRMS System, so that Dumas combustion for the analysis of N, C and S isotopes is combined with high temperature conversion for analysis of O and H isotopes in a single system. The EA IsoLink IRMS System enables the user to switch from N-C isotope ratio analysis to H-O isotope ratio analysis within just a few minutes. The performance of water analyses in both elemental analyzers the TC/EA and the EA IsoLink System are identical. The final report of the TC/EA-IRMS interlaboratory study states the applicability and emerging potential of EA-IRMS for food authenticity testing. 6 Even the analysis of organic substances like dodecane, methyl dodecanoate and methyl N-methylanthranilate produced good results. Conclusion Since its introduction in 1997, the Thermo Scientific high temperature conversion elemental analyzer TC/EA has created a new field of application for the O- and H-isotope analysis world-wide. TC/EA-IRMS is since then used for the analysis of liquids. References and Related Literature 1. Koziet J.J., Isotope Ratio Mass Spectrometric Method for the Online Determination of Oxygen-18 in Organic Matter, J. Mass Spectrom 1997, 32, 103-108. 2. Coplen T., Normalization of oxygen and hydrogen isotope data, Chem. Geol. 1988, 72, 293-297. 3. Nelson S.T., A simple, practical methodology for routine VSMOW/SLAP normalization of water samples analyzed by continuous flow methods, RCM 14, 2000, 1044-1046. 4. 4 Werner R. & Brand W.A., Referencing strategies and techniques in stable isotope ratio analysis, RCM 15, 2001, 501-519. 5. Calderone G., Reniero F. & Guillou C., 18O/16O measurements on ethanol, RCM 20, 2006, 937-940. 6. Bréas O., Thomas F., Zeleny R., Calderone G., Jamin E. & Guillou C., Performance evaluation of elemental analysis/isotope ratio mass spectrometry methods for the determination of the D/H ratio in tetramethylurea and other compounds results of a laboratory intercomparison, RCM 21, 2007, 1555-1560. The advantage of fast sub-μl analysis of liquid samples (only five minutes for two isotope ratios) by far outweighs the eventually lower precision. Low investment with high sample throughput leads to low cost per analysis. In addition, an autosampler for solids with minor modifications on the reactor packing makes the TC/EA also a valuable system for the isotope ratio analysis of solid samples. Find out more at thermofisher.com/eaisolink For Research Use Only. Not for use in diagnostic procedures. 2016 Thermo Fisher Scientific Inc. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. AN30180-EN 0916