Comparison of Several Detector Technologies for Measurement of Special Nuclear Materials i

Similar documents
2008 Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies SPECTRAL ANALYSIS OF RADIOXENON

Quantitation and Ratio Determination of Uranium Isotopes in Water and Soil Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Report Documentation Page

REGENERATION OF SPENT ADSORBENTS USING ADVANCED OXIDATION (PREPRINT)

CRS Report for Congress

Analysis Comparison between CFD and FEA of an Idealized Concept V- Hull Floor Configuration in Two Dimensions. Dr. Bijan Khatib-Shahidi & Rob E.

System Reliability Simulation and Optimization by Component Reliability Allocation

Report Documentation Page

Use of Wijsman's Theorem for the Ratio of Maximal Invariant Densities in Signal Detection Applications

Erik L. Swanberg 1 and Steven G. Hoffert 2. Veridian Systems 1, Autometric 2. Sponsored by Defense Threat Reduction Agency

Attribution Concepts for Sub-meter Resolution Ground Physics Models

Diagonal Representation of Certain Matrices

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies AUTOMATED QA/QC CHECK

Closed-form and Numerical Reverberation and Propagation: Inclusion of Convergence Effects

7. PERFORMING ORGANIZATION NAME(S) AND ADDRESS(ES) 8. PERFORMING ORGANIZATION REPORT NUMBER

Metrology Experiment for Engineering Students: Platinum Resistance Temperature Detector

REPORT DOCUMENTATION PAGE

G.H. Share, W.N. Johnson, J.D. Kurfess, R.J. Murphy. A. Connors. B.L. Dingus, B.E. Schaefer. W. Collmar, V. Schonfelder

Optimizing Robotic Team Performance with Probabilistic Model Checking

Thermo-Kinetic Model of Burning for Polymeric Materials

P. Kestener and A. Arneodo. Laboratoire de Physique Ecole Normale Supérieure de Lyon 46, allée d Italie Lyon cedex 07, FRANCE

Report Documentation Page

AUTOMATIC AND INTERACTIVE ANALYSIS SOFTWARE FOR BETA- GAMMA COINCIDENCE SYSTEMS USED IN CTBT MONITORING

PIPS 3.0. Pamela G. Posey NRL Code 7322 Stennis Space Center, MS Phone: Fax:

2010 Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Ocean Acoustics Turbulence Study

INFRARED SPECTRAL MEASUREMENTS OF SHUTTLE ENGINE FIRINGS

Real-Time Environmental Information Network and Analysis System (REINAS)

Effects of Constrictions on Blast-Hazard Areas

DIRECTIONAL WAVE SPECTRA USING NORMAL SPREADING FUNCTION

SW06 Shallow Water Acoustics Experiment Data Analysis

Broadband matched-field source localization in the East China Sea*

Predictive Model for Archaeological Resources. Marine Corps Base Quantico, Virginia John Haynes Jesse Bellavance

MODELING SOLAR UV/EUV IRRADIANCE AND IONOSPHERE VARIATIONS WITH MT. WILSON MAGNETIC INDICIES

Crowd Behavior Modeling in COMBAT XXI

REPORT DOCUMENTATION PAGE. Theoretical Study on Nano-Catalyst Burn Rate. Yoshiyuki Kawazoe (Tohoku Univ) N/A AOARD UNIT APO AP

Exact Solution of a Constrained. Optimization Problem in Thermoelectric Cooling

Advanced Numerical Methods for NWP Models

Super-Parameterization of Boundary Layer Roll Vortices in Tropical Cyclone Models

Coastal Engineering Technical Note

High-Fidelity Computational Simulation of Nonlinear Fluid- Structure Interaction Problems

Scattering of Internal Gravity Waves at Finite Topography

Comparative Analysis of Flood Routing Methods

Nonlinear Internal Waves: Test of the Inverted Echo Sounder

STUDY OF DETECTION LIMITS AND QUANTITATION ACCURACY USING 300 MHZ NMR

Estimation of Vertical Distributions of Water Vapor and Aerosols from Spaceborne Observations of Scattered Sunlight

Predicting Tropical Cyclone Formation and Structure Change

Design for Lifecycle Cost using Time-Dependent Reliability

A report (dated September 20, 2011) on. scientific research carried out under Grant: FA

Estimation of Vertical Distributions of Water Vapor from Spaceborne Observations of Scattered Sunlight

On Applying Point-Interval Logic to Criminal Forensics

AN OPTICAL-IR COUNTERPART FOR SGR B1900+l4? and. 1. Introduction

VLBA IMAGING OF SOURCES AT 24 AND 43 GHZ

HIGH-POWER SOLID-STATE LASER: LETHALITY TESTING AND MODELING

Dynamics of Droplet-Droplet and Droplet-Film Collision. C. K. Law Princeton University

Catalytic Oxidation of CW Agents Using H O in 2 2 Ionic Liquids

Improvements in Modeling Radiant Emission from the Interaction Between Spacecraft Emanations and the Residual Atmosphere in LEO

Development and Application of Acoustic Metamaterials with Locally Resonant Microstructures

Measurement of Accelerated Particles at the Sun

USMC Enlisted Endstrength Model

USER S GUIDE. ESTCP Project ER

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

Experimental and Theoretical Studies of Ice-Albedo Feedback Processes in the Arctic Basin

Super-Parameterization of Boundary Layer Roll Vortices in Tropical Cyclone Models

FRACTAL CONCEPTS AND THE ANALYSIS OF ATMOSPHERIC PROCESSES

Parametric Models of NIR Transmission and Reflectivity Spectra for Dyed Fabrics

SENSORS FOR MEASURING THE VOLUME SCATTERING FUNCTION OF OCEANIC WATERS

Guide to the Development of a Deterioration Rate Curve Using Condition State Inspection Data

Flocculation, Optics and Turbulence in the Community Sediment Transport Model System: Application of OASIS Results

High Resolution Surface Characterization from Marine Radar Measurements

Extension of the BLT Equation to Incorporate Electromagnetic Field Propagation

PREDICTION OF HIGH TEMPERATURE GREASE LIFE USING A DECOMPOSITION KINETIC MODEL

A Bound on Mean-Square Estimation Error Accounting for System Model Mismatch

The Navy Precision Optical Interferometer for SSA applications: an update

Mine Burial Studies with a Large Oscillating Water-Sediment Tunnel (LOWST)

Experimental and Theoretical Studies of Ice-Albedo Feedback Processes in the Arctic Basin

IMPROVED SYNTHESIS AND REACTION CHEMISTRY OF FN 3

Internal Tide Generation in the Indonesian Seas

Computer Simulation of Sand Ripple Growth and Migration.

Reverse Ion Acceleration by Laser-Matter Interaction

Assimilation of Synthetic-Aperture Radar Data into Navy Wave Prediction Models

Molecular Characterization and Proposed Taxonomic Placement of the Biosimulant 'BG'

Coupled Ocean-Atmosphere Modeling of the Coastal Zone

Regional Stratification and Shear of the Various Streams Feeding the Philippine Straits ESR Component

Direct Numerical Simulation of Aeolian Tones

Hurricane Wave Topography and Directional Wave Spectra in Near Real-Time

Sensitivity of West Florida Shelf Simulations to Initial and Boundary Conditions Provided by HYCOM Data-Assimilative Ocean Hindcasts

Modulation Instability of Spatially-Incoherent Light Beams and Pattern Formation in Incoherent Wave Systems

HIGH PERFORMANCE CONTROLLERS BASED ON REAL PARAMETERS TO ACCOUNT FOR PARAMETER VARIATIONS DUE TO IRON SATURATION

Grant Number: N IP To compare obtained theoretical results with NPAL experimental data.

Investigation of the Air-Wave-Sea Interaction Modes Using an Airborne Doppler Wind Lidar: Analyses of the HRDL data taken during DYNAMO

Range-Dependent Acoustic Propagation in Shallow Water with Elastic Bottom Effects

NAVGEM Platform Support

Analysis of Infrared Measurements of Microbreaking and Whitecaps

Swash Zone Dynamics: Modeling and Data Analysis

uniform distribution theory

A 1/10th Degree Global Ocean Simulation Using the Parallel Ocean Program

An Examination of 3D Environmental Variability on Broadband Acoustic Propagation Near the Mid-Atlantic Bight

Determining the Stratification of Exchange Flows in Sea Straits

Convection and Shear Flow in TC Development and Intensification

Transcription:

Comparison of Several Detector Technologies for Measurement of Special Nuclear Materials i A. E. Proctor, K. R. Pohl Constellation Technology Corporation, 7887 Bryan Dairy Road, Largo Fl 33777,U.S.A. Abstract Gamma spectral analysis of measurements made on Special Nuclear Materials (SNM) provides identification and sometimes quantification of the SNM present. Different radiation detection technologies have been used for measurement of SNM. Selection of one of these for a particular application depends on the information that can be obtained using that detector and on the support needed to use the detection technology, especially if it is to be applied at sites where limited support is available. In some cases, detector selection may be determined by a need to limit the information available from measurements due to international treaty requirements. There are many studies of system capabilities in SNM measurement. Most of these used either scintillator or high-resolution (HPGe) detectors embedded in a unique electronics and software combination. Their respective developers claimed good results for their instrument. Unfortunately, the packaging often obscured the question of what can be obtained from various detector technologies. While instrumentation and analysis is important, a study comparing only detector characteristics would serve as a starting point for development of systems for SMN measurement. Our detector study consisted of measurements on shielded and unshielded high-enriched uranium and Plutonium sources using detectors: NaI(Tl), Cadmium Zinc Telluride (CZT), Mercuric Iodide (HgI 2 ), and two HPGe detectors, one with liquid nitrogen cooling and one with electromechanical cooling. Measurements were made simultaneously using all detectors. As expected, all detectors are suitable for SNM measurements in cases where one radionuclide, such as 235 U, is the dominant constituent, but only the high-resolution detectors are suitable for isotopic composition determinations using standard spectroscopic analysis. Room temperature semiconductor detectors (CZT, HgI 2 ) can provide limited isotopic information. Development of improved analysis algorithms for room-temperature semiconductor detectors would be desirable because of their attractive characteristics: small sizes and minimal support requirements. Introduction We undertook this task to provide an equal comparison of radiation detector technologies for the measurement of SNM. There are (and have been) many studies of system capabilities in SNM measurement. Most of these used either scintillator or high-resolution (HPGe) detectors embedded in a unique electronics and software combination. Their respective developers claimed good results for their instrument. The packaging often obscured the question of what can be obtained from various detector technologies. While instrumentation and analysis is important for this work, and user-friendly features are also important, a study comparing detector characteristics would serve as a starting point for the development of systems for arms control. Moreover, knowing the capabilities of various detectors would be valuable when we are the inspected party rather than the inspector, required by various bilateral agreements. The detector study consisted of measurements on shielded and unshielded SNM using detectors: NaI (Tl), Cadmium Zinc Telluride (CZT), Mercuric Iodide (HgI 2 ), and two HPGe detectors, one with liquid nitrogen cooling ( LN 2 ) and one with electromechanical cooling ( EMC ). A high pressure xenon ionization chamber was to be tested but did not arrive in time to be included. Gamma Spectroscopy of Special Nuclear Materials The individual gamma ray energies used to establish the presence and quantify SNM are well established. ii These include the 143.7 and 185.6 kev gamma rays of 235 U and the 1001 kev emission of 234m Pa, a daughter of 238 U. Plutonium gamma spectra are more complex: significant emission occurs at 59.5 kev from 241 Am

Report Documentation Page Form Approved OMB No. 0704-0188 Public reporting burden for the collection of information is estimated to average 1 hour per response, including the time for reviewing instructions, searching existing data sources, gathering and maintaining the data needed, and completing and reviewing the collection of information. Send comments regarding this burden estimate or any other aspect of this collection of information, including suggestions for reducing this burden, to Washington Headquarters Services, Directorate for Information Operations and Reports, 1215 Jefferson Davis Highway, Suite 1204, Arlington VA 22202-4302. Respondents should be aware that notwithstanding any other provision of law, no person shall be subject to a penalty for failing to comply with a collection of information if it does not display a currently valid OMB control number. 1. REPORT DATE 2003 2. REPORT TYPE 3. DATES COVERED 00-00-2003 to 00-00-2003 4. TITLE AND SUBTITLE Comparison of Several Detector Technologies for Measurement of Special Nuclear Materials 5a. CONTRACT NUMBER 5b. GRANT NUMBER 5c. PROGRAM ELEMENT NUMBER 6. AUTHOR(S) 5d. PROJECT NUMBER 5e. TASK NUMBER 5f. WORK UNIT NUMBER 7. PERFORMING ORGANIZATION NAME(S) AND ADDRESS(ES) Constellation Technology Corporation,7887 Bryan Dairy Road,Largo,FL,33777 8. PERFORMING ORGANIZATION REPORT NUMBER 9. SPONSORING/MONITORING AGENCY NAME(S) AND ADDRESS(ES) 10. SPONSOR/MONITOR S ACRONYM(S) 12. DISTRIBUTION/AVAILABILITY STATEMENT Approved for public release; distribution unlimited 13. SUPPLEMENTARY NOTES 14. ABSTRACT see report 15. SUBJECT TERMS 11. SPONSOR/MONITOR S REPORT NUMBER(S) 16. SECURITY CLASSIFICATION OF: 17. LIMITATION OF ABSTRACT a. REPORT unclassified b. ABSTRACT unclassified c. THIS PAGE unclassified 18. NUMBER OF PAGES 8 19a. NAME OF RESPONSIBLE PERSON Standard Form 298 (Rev. 8-98) Prescribed by ANSI Std Z39-18

(daughter of 241Pu), 129.3 and 413.7 kev for 239 Pu, 45.2 kev, and 160.3 kev for 240 Pu. Both uranium and Plutonium emit many less intense gamma rays. Spectra of Plutonium mixtures are complex. Measurement of 235 U is more straightforward than measurement of the Plutonium isotopes. The gamma line at 185.7 kev is used most often. Unfortunately, in areas of high natural background radiation, the 186.2 and 187.1 kev gamma lines of 226 Ra, a naturally occurring radionuclide, interfere with the 185.7 kev line of 235 U. The three gamma lines can only be separated using high-resolution detectors under ideal measurement conditions. Use of the 143 kev gamma line of 235 U is an alternative (the 143 kev line is used extensively in gamma spectroscopy in support of cleanup operations). Practical measurements of Plutonium are more complex. There are five isotopes of interest: 238 Pu, 240 Pu, 239 Pu, 241 Pu, and 242 Pu that are produced from irradiation of 238 U. 239 Pu is of most interest to weapons inspections. Due to spontaneous neutron emission, 238 Pu and 240 Pu are undesirable in weapons but are tolerable in Plutonium used in nuclear reactors. The ratio of 240 Pu to 239 Pu in weapons grade materials is ~5%, while this ratio is ~15% for reactor grade Plutonium. From an arms control viewpoint, the presence of any form of Plutonium may be significant, but only weapons grade material is typically used in nuclear weapons. Fortunately, the individual isotopes of interest emit gamma rays having unique energies; unfortunately the spectral are complex and require high-resolution measurements to separate the individual gamma lines. Codes have been developed for the determination of SNM content based on spectra. For example, the MGA code iii, available in several different forms, is often used for determining the isotopic composition of Plutonium samples. Additionally, some gamma rays from the daughter isotopes of SNM are useful in identification of SNM by association. For example, 232 U has been suggested as a means to identify the presence of 235 U, using the 2614 kev gamma line of 208 Tl, a daughter of 232 U. iv. 232 U is a small constituent of high enriched uranium (HEU). 208 Tl produces a 2614 kev gamma ray, however, 208 Tl is also produced from the natural 232 Th decay chain. Using the 2614 kev (or 583 kev) gamma line as an indication of 235 U requires a correction for the 208 Tl produced by natural sources. Measurement of the 59.5 kev gamma ray emitted by 241 Am (a daughter of 241 Pu) is often used as an indication of Plutonium; the 59.5 kev gamma has been used as a measure of Plutonium contamination in environmental surveys. v 241 Am concentration for Plutonium samples may be very different. Plutonium samples used for this study both emit gamma rays at 59.5 kev, with the weapons grade sample having much higher emission. This difference is not a characteristic of reactor and weapons grade Plutonium, but an indication of the time that has elapsed since the Plutonium sample was separated from neutron-irradiated Uranium (i.e., U-238). Lastly, the gamma ray background in the facility in which the inspector performs his measurements must to be considered. Background contributions to the measured spectrum determine the minimum detectable activity in spectral measurements. For identification purposes, extra gamma ray lines complicate the analysis. The facility background in which these measurements were done contains contributions due to 137 Cs, 40 K, 228 Ac, 214 Bi, 226 Ra, and 212 Pb. Intercomparison Measurements Detectors based on several different technologies were selected for comparison. These included high resolution HPGe: a 55% efficiency detector with EMC, a 40% efficient HPGe with LN 2 cooling, a 10 mm x 10 mm x 7.5 mm coplanar grid CZT detector, a 10 mm x 10mm x 2.8 mm HgI 2 detector, and a 2 diameter x 2 long NaI (Tl) detector with a photomultiplier. These detectors represent a cross-section of available technology. To aid in comparisons of detectors of varying sizes, we experimentally determined the efficiency vs. energy for each detector and calculated the area-normalized efficiencies : ratios of photopeak count rate / gamma rays striking the detector measured at multiple specific energies. Figure 1 shows the

area-normalized efficiencies vs. energy for our detector group. Absolute efficiencies can be estimated by multiplying the area-normalized efficiency by the source solid angle fraction viewed by the detector. Figure 1 Area-normalized Detector Efficiencies Information shown in Figure 1 includes the efficiency vs. energy trends for the CZT and HgI 2 detectors a more rapid decrease in efficiency than the others due to smaller detector active thicknesses. The CZT detector efficiency never reaches unity; counts are always lost even at low energies. The HPGe efficiency is representative of both the liquid nitrogen and electromechanically cooled P-type detectors. HPGe and NaI(Tl) detector efficiency curves have shapes similar to those reported previously. The detectors were arranged in a circle surrounding a test source in the TA-18 facility at Los Alamos National Laboratory (LANL). Detectors and sources rested on a wooden floor built over a concrete basement. Metal walls and roof covered the wooden floor. Locations of the various detectors were selected to provide a good count rate without significant dead time. Figure 2 shows the arrangement. This geometry was maintained during SNM measurements. Counting (live) times were 20 minutes for the LN 2 and EMC detectors and 60 minutes for the others. All detectors were aimed at the source to achieve the highest possible count rates. Figure 2 Measurement Arrangement

The LANL NIS-6 group provided small cylindrical SNM samples of four isotopic types: 1) HEU (93.15 % 235 U, remainder 238 U), 2) depleted uranium (0.22 % 235 U by weight), 3) weapons grade Plutonium (6% 240 Pu by weight, assumed 94% 239 Pu), and 4) high burnup Plutonium (21.94% 240 Pu by weight, assumed 78% 239 Pu). Sources from the four isotopic types in both shielded and unshielded configurations were measured individually to generate the spectra shown in this paper. SMN Measurement Results All detectors were tested using several HEU samples, both shielded and unshielded. Comparisons of the detectors were made based on good resolution of the 143, 160, and 185 kev gamma lines needed to positively identify HEU. The 185 kev gamma line is often used alone for enrichment measurements, but use of the other lines is needed when little is known about the sample in advance. Detector comparisons are shown below. Comparisons of shielded and unshielded sources are made in the next section. The first spectrum (Figure 3) shows data from a 295 g unshielded HEU sample with a spectrum of facility background, both measured with the EMC detector. Counting times were 20 minutes for the sample and overnight for the background. Data using the LN 2 detector were identical to those shown for the EMC. The majority of gamma lines visible are due to natural background radiation, including 137 Cs (661 kev), 40 K (1460 kev) and 214 Bi (1764 kev). This spectrum illustrates the complexity of spectra measured in a legacy nuclear facility with considerable industrial use. Figure 3 Facility Background and HEU Sample High-resolution spectra Note the abundance of gamma lines from the background. Only a few low energy lines due to 235 U are visible. Ideally, the sample-detector geometry can be adapted to minimize the effects of natural background. Figure 4 shows HEU spectra from the four detectors. As expected, the HPGe detectors resolved the 143, 160, 185, and 205 kev line grouping and the uranium x-rays around 95 kev. The room-temperature semiconductor detectors (CZT and HgI 2 ) resolved the four gamma lines but not the x-rays and the NaI(Tl) detector only resolved the 185 kev gamma line. Count rates in this and many figures have been scaled to show several spectra on a single plot. Peak heights may not represent actual count rates measured during this work.

Figure 4 HEU Spectra Obtained Using Several Detectors. The 185 kev Peak is Marked. We obtained spectra of measurements of weapons grade Plutonium. Figures 5-6 show these results. 10 7 10 6 10 5 HPGe ("EMC") HgI 2 x0.1 CZT x 0.01 NaI(Tl) 10 4 10 3 10 2 10 1 10 0 10-1 0 200 400 Energy (kev) Figure 5 Weapons Grade Plutonium Detector Comparison, Low-energy Region 10 4 10 3 HPGe ("EMC") HgI 2 x0.1 CZT x 0.01 NaI(Tl) 10 2 10 1 10 0 10-1 10-2 400 500 600 700 800 Energy (kev) Figure 6 Weapons Grade Plutonium detector Comparison, High-energy Region

Figures 5 and 6 show that all the detectors are suitable for semi- quantitative measurements based on the 414 kev region, but the higher resolution spectra are required for isotopic analysis of Plutonium samples. CZT and mercuric iodide spectra can be used for isotopic analysis if state-of-the-art computer codes are available. Spectra acquired with a best of the best CZT detector were used as input to the MGA code for determining Plutonium isotopics. vi In the high energy region, however, only the spectra acquired using the HPGe detector shows significant detail. Differentiation between weapons grade and high-burnup Plutonium is important because high burnup Plutonium is typically not used in weapons. We compare detector performance in measuring both weapons grade and high burnup Plutonium in Figures 7, 8, and 9. Figure 7 High Resolution Spectra of Weapons Grade and High Burnup Plutonium, Low-energy Region Numerous examples of Figure 7 can be found in published data. For comparison, Figure 8 shows spectra from measurement of the same samples using the CZT detector, and Figure 9 shows spectra of the high energy region using the CZT detector. Spectra acquired using the mercuric iodide detector closely resemble their CZT counterparts. Low energy (50-450 kev) data from the CZT or mercuric iodide detectors could be useful in differentiating between weapons grade and high burnup Plutonium, perhaps by using spectral peak ratios, with some difficulty. Data above 450 kev is degraded due to the decreasing detector efficiency at higher energies. Data acquired using the NaI(Tl) detector did not show significant differences. The 59.5 kev peak from 241 Am, typically present in weapons grade Plutonium, may not be present in more recently processed material or material that did not originally have much 241 Pu. Figure 8 Weapons Grade and High Burnup Plutonium, 0-450 kev

10 3 10 2 CZT detector Weapons grade Pu High burnup Pu 10 1 10 0 400 500 600 700 800 900 Energy (kev) Figure 9 Weapons Grade and High Burnup Plutonium, 400-900 kev We also investigated the effects of shielding on spectra taken with the different detectors. Shielding in our tests consisted of 1 of steel added between the source and detector. While actual shielding of SNM covers an unpredictably wide range, we sought to determine the effects of shielding on recognizable peak shapes that would be used as input to radionuclide identification algorithms. We did not attempt to determine quantitative results. Figure 10 shows the effects of shielding on HEU spectra acquired using HPGe and HgI 2 detectors. The 185 kev peak is attenuated but still present and the higher energy peaks are less attenuated. The uranium x-rays are highly attenuated, thereby complicating the process of differentiating uranium from natural radium, for example. Similar effects can be observed in spectra of shielded and unshielded Plutonium samples. Figure 10 Spectra from Shielded and Unshielded HEU Samples (scaled to separate spectra)

Examination of the spectra show that a typical peak search algorithm would find the 235 U peaks in all four spectra, although peaks are clearly better defined in the HPGe data. Again spectra acquired using mercuric iodide and CZT detectors are almost identical in terms of peak resolutions and relative peak heights. HPGe and CZT detectors have been successfully implemented in radionuclide identification devices. HPGe Detector Comparison High-resolution spectra shown in previous figures were obtained using Constellation Technology s EMC HPGe detector. The EMC HPGe detector was developed for arms control inspectors and others who may not have access to liquid nitrogen at the measurement locations. In all measurements, the EMC HPGe detector and the conventional liquid nitrogen-cooled detectors performed identically. This is shown in Figure 11, below, showing a portion of the weapons grade Plutonium spectrum. Figure 11 HPGe Detector Comparison i This work was supported by the Defense Threat Reduction Agency ii For example, see: D. Reilly, et. al Passive Nondestructive Assay of Nuclear Materials. NUREG/CR-5550 (1991), p 18. iii R. Gunnink, W. Ruhter. :MGA: A Gamma-Ray Spectrum Analysis Code for Determining Plutonium Isotopic Abundances, vols 1-3, UCRL-LR-103220 (1990). iv A. Perrung, Predicting U-232 Content in Uranium, PNNL-12075 (1998) v See for example, A.Proctor, Aerial Radiological Surveys, DOE/NV/11718-127 (1997). vi R. Gunnink, private communication to A. Proctor ( 2002)