Eppley PIR Precision Infrared Radiometer

Similar documents
Program Notes Use 2:1 voltage divider on Differential Input channel 1 to get full output range from C7

3WHB10K 3-WIRE HALF BRIDGE TERMINAL INPUT MODULE INSTRUCTION MANUAL

CR10XTCR THERMOCOUPLE REFERENCE

CR10XTCR THERMOCOUPLE REFERENCE

SP-LITE SILICON PYRANOMETER INSTRUCTION MANUAL

CR10TCR THERMOCOUPLE REFERENCE

MODEL 107 TEMPERATURE PROBE INSTRUCTION MANUAL

CM3 PYRANOMETER INSTRUCTION MANUAL

IRTS-P PRECISION INFRARED TEMPERATURE SENSOR INSTRUCTION MANUAL

INSTRUCTION MANUAL. CS547 Conductivity and Temperature Probe and A547 Interface. Revision: 9/00. Copyright (c) Campbell Scientific, Inc.

PTA-427 BAROMETRIC PRESSURE TRANSDUCER OPERATOR S MANUAL

INSTRUCTION MANUAL. SP Lite2 Pyranometer. August Copyright 2017 Campbell Scientific (Canada) Corp.

SI-111 Precision Infrared Radiometer Revision: 7/13

INSTRUCTION MANUAL. CR10XTCR Thermocouple Reference. Revision: 1/03. Copyright (c) Campbell Scientific, Inc.

3WHB10K 3-Wire Half Bridge Terminal Input Module

MODEL HMP45C TEMPERATURE AND RELATIVE HUMIDITY PROBE INSTRUCTION MANUAL

WARRANTY AND ASSISTANCE

3WHB10K 3-Wire Half Bridge Terminal Input Module

Module 1, Add on math lesson Simultaneous Equations. Teacher. 45 minutes

INSTRUCTION MANUAL. Model HFP01SC Self-Calibrating Soil Heat Flux Plate Revision: 3/14. Copyright Campbell Scientific, Inc.

What to Add Next time you update?

Exercise 1: Thermocouple Characteristics

INSTRUCTION MANUAL. Model 207 Temperature and Relative Humidity Probe. Revision: 2/94. Copyright (c) Campbell Scientific, Inc.

Exercise 1: Thermistor Characteristics

Model 108 Temperature Probe Revision: 4/07

CVD20 20:1 Voltage Divider Revision: 5/12

INSTRUCTION MANUAL. CS616 and CS625 Water Content Reflectometers Revision: 2/14. Copyright Campbell Scientific, Inc.

INSTRUCTION MANUAL. LI200X-LC Pyranometer for MetData1 4/97. Copyright (c) Campbell Scientific, Inc.

Parallel Circuits. Chapter

Chapter 18. Direct Current Circuits

Sensors Conditioning Module

4WFB120, 4WFB350, 4WFB1K 4 Wire Full Bridge Terminal Input Modules Revision: 5/07

227 Delmhorst Cylindrical Soil Moisture Block Revision: 5/13

PHYSICS EXTENDED ESSAY

1. Mark the correct statement(s)

THERMOCOUPLE CHARACTERISTICS TRAINER

Experiment 5: Thermocouples (tbc 1/14/2007, revised 3/16/2007, 3/22,2007, 2/23/2009, 3/13/2011)

Chapter 1. Blackbody Radiation. Theory

INSTRUCTION MANUAL. 023/CO2 Bowen Ratio System with CO2 Flux. Revision: 4/98. Copyright (c) Campbell Scientific, Inc.

US ARMY INTELLIGENCE CENTER CIRCUITS

TP01 Thermal Properties Sensor

EXPERIMENT NO. 4. Thermal Radiation: the Stefan-Boltzmann Law

STEAM Clown Production. Series Circuits. STEAM Clown & Productions Copyright 2017 STEAM Clown. Page 2

Temperature Sensors & Measurement

Resistance Learning Outcomes

Solved Problems. Electric Circuits & Components. 1-1 Write the KVL equation for the circuit shown.

Chapter 3. Chapter 3

MASSACHUSETTS INSTITUTE OF TECHNOLOGY /6.071 Introduction to Electronics, Signals and Measurement Spring 2006

Resistance Learning Outcomes. Resistance Learning Outcomes. Resistance

Experiment 4: Resistances in Circuits

LM34 - Precision Fahrenheit Temperature Sensor

IRTS-P Precision Infrared Temperature Sensor

Using Newton s Method to Solve the Clasic Diode Problem

MET ONE RAIN GAGE MODELS 380, 385, 380M, 385M

ECE2019 Sensors, Circuits, and Systems A2015. Lab #2: Temperature Sensing

Chapter 3. Chapter 3

Lab E3: The Wheatstone Bridge

Switched-Capacitor Circuits David Johns and Ken Martin University of Toronto

STRAIN GAUGE MEASUREMENT

St art. rp m. Km /h 1: : : : : : : : : : : : :5 2.5.

COURSE OF Prepared By: MUHAMMAD MOEEN SULTAN Department of Mechanical Engineering UET Lahore, KSK Campus

EXPERIMENT ET: ENERGY TRANSFORMATION & SPECIFIC HEAT

Figure 1: Capacitor circuit

2.004 Dynamics and Control II Spring 2008

Physics Investigation 10 Teacher Manual

DC Circuits Analysis

INSTRUCTION MANUAL. CS500 Temperature and Relative Humidity Probe. Revision: 7/04. Copyright (c) Campbell Scientific, Inc.

As light level increases, resistance decreases. As temperature increases, resistance decreases. Voltage across capacitor increases with time LDR

Test Review Electricity

Temperature Measurement

PGA309 Programmable Sensor Signal Conditioner. Bringing Linearity to a Non-Linear Pressure Sensor World

Lecture 36: Temperatue Measurements

Model HFP01 Soil Heat Flux Plate Revision: 7/12

EE301 RESISTANCE AND OHM S LAW

University of Connecticut Department of Electrical and Computer Engineering. ECE 4901: Fall 2017 Spring 2018

A.M. WEDNESDAY, 13 May minutes

Notes on Electricity (Circuits)

An Ultra Low Resistance Continuity Checker

PHYSICS 171. Experiment 3. Kirchhoff's Laws. Three resistors (Nominally: 1 Kilohm, 2 Kilohm, 3 Kilohm).

Millivolt Output Pressure Sensors

4. Heat and Thermal Energy

Trade of Electrician. Resistance Network Measurement

Electronics Resistive Sensors and Bridge Circuits

Circuit Analysis and Ohm s Law

coil of the circuit. [8+8]

Infrared Sensor Application Note 2 Signal Processing for Infrared Gas Sensors

EE 321 Analog Electronics, Fall 2013 Homework #3 solution

Keywords: sensor compensation, integrated circuit, ic, compensation algorithm, piezoresistive, prt, pressure sensors, temperature

Last Revision: August,

The Farad is a very big unit so the following subdivisions are used in

DATA PLOT Calibration Report: Sales Order: 65301

2007 The McGraw-Hill Companies, Inc. All rights reserved.

HIGH CURRENT BRIDGE DRIVER and 4-20mA Transmitter

Physics 17 Spring 2003 The Stefan-Boltzmann Law

4) What is the direction of the net force on the charge at the origin due to the other three?

Phys 2025, First Test. September 20, minutes Name:

Switch or amplifies f. Capacitor i. Capacitance is measured in micro/pico farads ii. Filters frequencies iii. Stores electrical energy

Physics 102: Lecture 05 Circuits and Ohm s Law

Simple calibration for ceramic sensing elements using an ME651

Exam 1--PHYS 202--Spring 2013

Transcription:

App. Note Code: 2RA-H Revision: 1 Eppley PIR Precision Infrared Radiometer CAMPBELL SCIENTIFIC, INC. W H E N M E A S U R E M E N T S M A T T E R Copyright 2001-2007 Campbell Scientific, Inc.

Eppley PIR Precision Infrared Radiometer The Eppley Laboratory, Inc. Precision Infrared Radiometer (PIR) consists of a thermopile and two thermistors and provides several measurement options. This application note describes interfacing Campbell Scientific dataloggers with the PIR, explains how to measure the thermopile and thermistors, and apply the temperature compensation to the thermopile output. This note assumes previous experience with Campbell Scientific dataloggers. Wiring The thermopile is connected, using wires A and C, to a differential analog channel. Connect C to the high (H) side and A to the low (L) side of the differential channel. The two thermistors are each wired to a single-ended analog channel and analog ground; thermistors are not polarity sensitive. Wires D and E are the case thermistor. Wires F and G are for the dome thermistor. The dome thermistor allows you to correct for the dome temperature. This is not necessary for many applications and is not discussed in this application note. Information about correcting for dome temperature is provided in the Eppley PIR manual. To complete the circuit, a bridge completion resistor must be wired between an excitation channel and the single-ended measurement channel. The bridge completion resistor should be a 1 kohm, precision- and temperature-stable resistor. The power (watt) rating of the resistor is not important. Other resistors can be used, but the datalogger must be programmed accordingly. Programming the Datalogger The datalogger program performs three basic tasks, typically in the following order: Measures the sensor (voltage and resistance measurements). Converts the measurements to appropriate units. Writes the data to final storage. Copyright 2001-2007 Campbell Scientific, Inc. 1 815 W. 1800 N., Logan, UT 84321-1784 (435) 753-2342 App. Note: 2RA-H

Measuring the Sensors Thermopile Thermistor Program Instruction 2 (P02) is used to make the differential voltage measurement. The PIR sensor has a very small full-scale output range, less than 1.5 mv. Using the smallest voltage range (±2.5 mv) of the CR10X will return the best measurement. Use a differential channel from 1 through 6, ensuring it's the same channel in which the sensor is wired. The output is in millivolts. 1: Volt (Diff) (P2) 2: 1 2.5 mv Slow Range 3: 1 DIFF Channel 4: 1 Loc [ PIR_mV ] 5: 1.0 Mult 6: 0.0 Offset Each thermistor s half-bridge circuit is measured with Program Instruction 5 (P5). Select the ±2500 mv Fast Range and a singleended channel from 1 through 12, ensuring the single-ended channel is not used by another input. Choose an excitation channel and use ±2500 mv for the excitation voltage. The same excitation channel can be used by both thermistors, but don't share bridge completion resistors. 2: AC Half Bridge (P5) 2: 15 2500 mv Fast Range 3: 3 SE Channel 4: 1 Excite all reps w/exchan 1 5: 2500 mv Excitation 6: 2 Loc [ Case_Res ] 7: 1.0 Mult 8: 0.0 Offset Converting to Appropriate Units The thermistor measurements must be converted to temperature in K. The mv output from the thermopiles is converted to W m -2, then corrected for the temperature effects on the PIR s case. 2 Copyright 2001-2007 Campbell Scientific, Inc. App. Note: 2RA-H 815 W. 1800 N., Logan, UT 84321 (435) 753-2342

Converting Thermistor Measurements to Resistance The P5 instruction used to measure the thermistor returns a ratiometric output. To convert to resistance, use Instruction 59 (P59). Note the multiplier of 1000; this is the value of the bridge completion resistor. 3: BR Transform Rf[X/(1-X)] (P59) 2: 2 Loc [ Case_Res ] 3: 1000 Multiplier (Rf) With the resistance of the thermistor in an input location, convert the resistance to temperature. Converting Resistance to Temperature Equation 1 converts the resistance of the thermistor to temperature in K. This conversion requires some basic math functions that can be entered in a subroutine or in the main body of the program table. T = 1/(A + B * Ln(R) + C * (Ln) 3 ) (Eq. 1) Where: T = temperature in degrees Kelvin A = 0.0010295 or 1.0295E -3 B = 0.0002391 or 2.391E -4 C = 0.0000001568 or 1.568E -7 R = the measured resistance of the thermistor Ln(R) = the natural log of the thermistor resistance ; Load constants A, B and C to input locations 4: Z=F (P30) 1: 1.0295 F 2: -3 Exponent of 10 3: 6 Z Loc [ ConstA ] ; *** A term 5: Z=F (P30) 1: 2.391 F 2: -4 Exponent of 10 3: 7 Z Loc [ ConstB ] 6: Z=F (P30) 1: 1.568 F 2: -7 Exponent of 10 3: 8 Z Loc [ ConstC ] Copyright 2001-2007 Campbell Scientific, Inc. 3 815 W. 1800 N., Logan, UT 84321-1784 (435) 753-2342 App. Note: 2RA-H

; Convert resistance of thermistor to natural log of resistance 7: Z=LN(X) (P40) 1: 2 X Loc [ Case_Res ] 2: 9 Z Loc [ Ln_ResS1 ] ;*** Natural log of resistance ; Multiply constant B with Natural log of resistance ; Store result in "B term", bln_res 8: Z=X*Y (P36) 1: 7 X Loc [ ConstB ] 2: 9 Y Loc [ Ln_Res ] 3: 10 Z Loc [ bln_res ] ; *** B term ; Square natural log of resistance 9: Z=X*Y (P36) 1: 9 X Loc [ Ln_Res ] 2: 9 Y Loc [ Ln_Res ] 3: 11 Z Loc [ Ln_res2 ] ; Natural log of resistance squared ; Cube natural log of resistance 10: Z=X*Y (P36) 1: 9 X Loc [ Ln_Res ] 2: 11 Y Loc [ Ln_res2 ] 3: 12 Z Loc [ Ln_res3 ] ; Natural log of resistance cubed ; Multiply constant C with natural log of resistance raised to the third power 11: Z=X*Y (P36) 1: 8 X Loc [ ConstC ] 2: 12 Y Loc [ Ln_res3 ] 3: 12 Z Loc [ Ln_res3 ] ; *** C term ; Add A term to B term, store in location case_temp 12: Z=X+Y (P33) 1: 6 X Loc [ ConstA ] 2: 10 Y Loc [ bln_ress1 ] 3: 2 Z Loc [ case_temp ] ; Add C term to A and B term, store in location Case_temp 13: Z=X+Y (P33) 1: 2 X Loc [ case_temp ] 2: 12 Y Loc [ Ln_res3] 3: 2 Z Loc [ case_temp ] 4 Copyright 2001-2007 Campbell Scientific, Inc. App. Note: 2RA-H 815 W. 1800 N., Logan, UT 84321 (435) 753-2342

; Take the reciprocal of the "Case_temp", result is temperature in degrees Kelvin 14: Z=1/X (P42) 1: 2 X Loc [ case_temp ] 2: 2 Z Loc [ case_temp ] ; *** Temp in degrees K Converting Thermopile Measurements to W m -2 The thermopile measurement output is in mv. To convert mv to Wm -2, the PIR sensitivity is divided into the mv output. Sensitivity is given in units of µvw -1 m 2. The proper multiplier is: 1/PIR sensitivity. Example: Thermopile sensitivity = 3.41 µvw -1 m 2 or.00341 mv / W/m 2 The thermopile sensitivity is obtained from the PIR calibration sheet. Multiplier = (1)/(.00341 mv / W/m 2 ) = 293.255 Now that the multiplier has been calculated, use Instruction 37 to convert the thermopile mv output to W m -2. 3: Z=X*F (P37) 1: 1 X Loc [ PIR_mV ] 2: 293.255 F ; *** Multiplier 3: 14 Z Loc [ PIR_Aterm ] Correcting for PIR Case Temperature You must account for the effects of the PIR case temperature. The thermopile output is adjusted using Equation 2: Corrected Thermopile Output = A + (C * T 4 ) (Eq. 2) Where: A = Thermopile output in W m -2 C = Stefan-Boltzmann constant = 5.6697E -8 W m -2 K -4 T 4 = the case temperature in degrees K raised to the fourth power Datalogger program instructions can be placed in a subroutine or the main body of the program. Copyright 2001-2007 Campbell Scientific, Inc. 5 815 W. 1800 N., Logan, UT 84321-1784 (435) 753-2342 App. Note: 2RA-H

Use the following instructions: ; load number 4 into input location Power4 16: Z=F (P30) 1: 4 F 2: 0 Exponent of 10 3: 11 Z Loc [ Power4 ] ; Raise temperature to the fourth power 17: Z=X^Y (P47) 1: 10 X Loc [ case_temp ] 2: 11 Y Loc [ Power4 ] 3: 10 Z Loc [ case_temp ] ; Load 5.6697E -8 into PIR_Bterm 18: Z=F (P30) 1: 5.669 F 2: -8 Exponent of 10 3: 13 Z Loc [ PIR_Bterm ] ; Multiply 5.6697E -8 by Temp K raised to the fourth ; This is the PIR B term 19: Z=X*Y (P36) 1: 13 X Loc [ PIR_Bterm ] 2: 10 Y Loc [ case_temp ] 3: 13 Z Loc [ PIR_Bterm ] ; Add PIR A term to PIR B term. Output is in Watts per meter squared. 20: Z=X+Y (P33) 1: 14 X Loc [ PIR_Aterm ] 2: 13 Y Loc [ PIR_Bterm ] 3: 15 Z Loc [ PIR_Watt ] Writing Data to Final Storage First determine when to write data to final storage then what data to write. The program should set the output flag before any output processing instructions are executed. Below is the portion of a datalogger program that outputs the data to final storage. ; Every 60 minutes at the top of the hour, set the Output Flag 6: If time is (P92) 1: 0 Minutes (Seconds --) into a 2: 60 Interval (same units as above) 3: 10 Set Output Flag High (Flag 0) 6 Copyright 2001-2007 Campbell Scientific, Inc. App. Note: 2RA-H 815 W. 1800 N., Logan, UT 84321 (435) 753-2342

;Output the day, hour, and minute 7: Real Time (P77) 1: 0220 Day, Hour/Minute (midnight = 2400) ; Store the PIR hourly average 8: Average (P71) 2: 15 Loc [ PIR_Watt ] Sample CR10X Program ; Store the PIR minimum value for the past hour 9: Minimum (P74) 2: 00 Time Option 3: 15 Loc [ PIR_Watt ] ; Store the PIR maximum value for the past hour 10: Maximum (P73) 2: 00 Time Option 3: 15 Loc [ PIR_Watt ] ;{CR10X} ; *Table 1 Program 01: 10 Execution Interval (seconds) ; Measure thermopile 1: Volt (Diff) (P2) 2: 1 2.5 mv Slow Range 3: 1 DIFF Channel 4: 1 Loc [ PIR_mV ] 5: 1.0 Mult 6: 0.0 Offset Copyright 2001-2007 Campbell Scientific, Inc. 7 815 W. 1800 N., Logan, UT 84321-1784 (435) 753-2342 App. Note: 2RA-H

; Measure case thermistor 2: AC Half Bridge (P5) 2: 15 2500 mv Fast Range 3: 3 SE Channel 4: 1 Excite all reps w/exchan 1 5: 2500 mv Excitation 6: 2 Loc [ Case_Res ] 7: 1.0 Mult 8: 0.0 Offset ; Apply eppley calibration to thermopile output 3: Z=X*F (P37) 1: 1 X Loc [ PIR_mV ] 2: 256.891 F 3: 14 Z Loc [ PIR_Aterm ] ; Call Subroutine 1, used to calculate case temperature 4: Do (P86) 1: 1 Call Subroutine 1 ; Call Subroutine 2, Used to correct thermopile output 5: Do (P86) 1: 2 Call Subroutine 2 ; Store the data hourly 6: If time is (P92) 1: 0 Minutes (Seconds --) into a 2: 60 Interval (same units as above) 3: 10 Set Output Flag High (Flag 0) 7: Real Time (P77) 1: 0220 Day, Hour/Minute (midnight = 2400) 8: Average (P71) 2: 15 Loc [ PIR_Watt ] 9: Minimum (P74) 2: 00 Time Option 3: 15 Loc [ PIR_Watt ] 8 Copyright 2001-2007 Campbell Scientific, Inc. App. Note: 2RA-H 815 W. 1800 N., Logan, UT 84321 (435) 753-2342

10: Maximum (P73) 2: 00 Time Option 3: 15 Loc [ PIR_Watt ] *Table 2 Program 02: 0.0000 Execution Interval (seconds) *Table 3 Subroutines 1: Beginning of Subroutine (P85) 1: 1 Subroutine 1 ; Equation to convert YSI 10 K thermistor resistance to temperature ; in degrees K ; 1/T = a + b(lnr) + c(lnr)^3 ; Coefficients used in equation above ; a = 0.0010295 = ConstA ; b = 0.0002391 = ConstB ; c = 1.568E-7 = ConstC ; Calculate resistance from P5 instruction (Table 1, line 2) 2: BR Transform Rf[X/(1-X)] (P59) 2: 2 Loc [ Case_Res ] 3: 1000 Multiplier (Rf) ; Load constants a, b and c to input locations 3: Z=F (P30) 1: 1.0295 F 2: -3 Exponent of 10 3: 3 Z Loc [ ConstA ] ; *** A term 4: Z=F (P30) 1: 2.391 F 2: -4 Exponent of 10 3: 4 Z Loc [ ConstB ] 5: Z=F (P30) 1: 1.568 F 2: -7 Exponent of 10 3: 5 Z Loc [ ConstC ] Copyright 2001-2007 Campbell Scientific, Inc. 9 815 W. 1800 N., Logan, UT 84321-1784 (435) 753-2342 App. Note: 2RA-H

; Convert resistance of thermistor to natural log of resistance 6: Z=LN(X) (P40) 1: 2 X Loc [ Case_Res ] 2: 6 Z Loc [ Ln_Res ] ; *** Natural log of resistance ; Multiply constant B with Natural log of resistance ; Store result in "B term, bln_res 7: Z=X*Y (P36) 1: 4 X Loc [ ConstB ] 2: 6 Y Loc [ Ln_Res ] 3: 7 Z Loc [ bln_res ] ; *** B term ; Square natural log of resistance 8: Z=X*Y (P36) 1: 6 X Loc [ Ln_Res ] 2: 6 Y Loc [ Ln_Res ] 3: 8 Z Loc [ Ln_res2 ] ; Natural log of resistance squared ; Cube natural log of resistance 9: Z=X*Y (P36) 1: 6 X Loc [ Ln_Res ] 2: 8 Y Loc [ Ln_res2 ] 3: 9 Z Loc [ Ln_res3 ] ; Natural log of resistance cubed ; Multiply constant C with natural log of resistance raised to the third power 10: Z=X*Y (P36) 1: 5 X Loc [ ConstC ] 2: 9 Y Loc [ Ln_res3 ] 3: 9 Z Loc [ Ln_res3 ] ; *** C term ; Add A term to B term natural log of resistance, store in location ; case_temp 11: Z=X+Y (P33) 1: 3 X Loc [ ConstA ] 2: 7 Y Loc [ bln_res ] 3: 10 Z Loc [ case_temp ] ; Add C term to A and B term, store in location case_temp 12: Z=X+Y (P33) 1: 10 X Loc [ case_temp ] 2: 9 Y Loc [ Ln_res3 ] 3: 10 Z Loc [ case_temp ] 10 Copyright 2001-2007 Campbell Scientific, Inc. App. Note: 2RA-H 815 W. 1800 N., Logan, UT 84321 (435) 753-2342

; Invert A plus B plus C term. Result is temperature in degrees ; Kelvin 13: Z=1/X (P42) 1: 10 X Loc [ case_temp ] 2: 10 Z Loc [ case_temp ] ; *** Temp in degrees K 14: End (P95) 15: Beginning of Subroutine (P85) 1: 2 Subroutine 2 ; Apply case thermistor correction to PIR thermopile output ; Correction is PIR output in watts per meter squared plus ; 5.669E -8 multiplied by the case temperature raised to the fourth power. ; Load number 4 into input location Power4 16: Z=F (P30) 1: 4 F 2: 0 Exponent of 10 3: 11 Z Loc [ Power4 ] ; Raise temperature to the fourth power 17: Z=X^Y (P47) 1: 10 X Loc [ case_temp ] 2: 11 Y Loc [ Power4 ] 3: 10 Z Loc [ case_temp ] ; Load 5.669E -8 into PIR_Bterm 18: Z=F (P30) 1: 5.669 F 2: -8 Exponent of 10 3: 13 Z Loc [ PIR_Bterm ] ; Multiply 5.669E -8 by Temp K raised to the fourth ; This is the PIR B term 19: Z=X*Y (P36) 1: 13 X Loc [ PIR_Bterm ] 2: 10 Y Loc [ case_temp ] 3: 13 Z Loc [ PIR_Bterm ] Copyright 2001-2007 Campbell Scientific, Inc. 11 815 W. 1800 N., Logan, UT 84321-1784 (435) 753-2342 App. Note: 2RA-H

; Add PIR A term to PIR B term. Output is in Watts per meter squared. 20: Z=X+Y (P33) 1: 14 X Loc [ PIR_Aterm ] 2: 13 Y Loc [ PIR_Bterm ] 3: 15 Z Loc [ PIR_Watt ] 21: End (P95) End Program 12 Copyright 2001-2007 Campbell Scientific, Inc. App. Note: 2RA-H 815 W. 1800 N., Logan, UT 84321 (435) 753-2342