Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Similar documents
Analysis of domestic sludge using the Agilent 4200 MP-AES

The Measurement of Lead in Food Products by Graphite Furnace AA and SpectraAA-40

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

NOTICE: This method: The laboratories have to study the instrumental conditions appropriate for their own instrumentation.

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Direct Determination of Aluminium in Milk by Graphite Furnace Atomic Absorption Spectrometry

Direct Determination of Cadmium and Lead in honey by Electrothermal Atomization Atomic Absorption Spectrometry with Zeeman Effect Correction

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

ICP-OES Application Note Number 35

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Quality Control Procedures for Graphite Furnace AAS using Avanta Software

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

UNITED NATIONS. ENVIRONMENT PROGRAMME MEDITERRANEAN ACTION PLAN 25 May 2015 Original: English. UNEP(DEPI)/ MED WG.417/Inf.14

Test Method: CPSC-CH-E

Title Revision No date DETERMINATION OF HEAVY METALS (ARSENIC, CADMIUM, LEAD AND MERCURY) IN COSMETIC PRODUCTS

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Sodium Chloride - Analytical Standard

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Prof. Dr. Biljana Škrbić, Jelena Živančev

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Selenium Determination in Blood using Zeeman Background Correction and Palladium/Ascorbic Acid Chemical Modification

Cadmium and lead content

Atomic Absorption Spectrometer ZEEnit P series

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

DISCLAIMER: This method:

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Example method validation and MU approach for Cadmium in seafood test method

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

OPTIMIZING METALS SAMPLE PREP

Not Just for Paint Chips Anymore. High Throughput, Open Vessel Microwave Digestion of Matrices Requiring Moderate Conditions

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Mercury Determination in Lichen, BCR-482, Using the QuickTrace M-7600 CVAAS Mercury Analyzer

Determination of urea in ultrapure water by IC-MS/MS

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

INFLUENCE OF THE MEASUREMENT MEDIUM AND MATRIX MODIFIERS ON THE DETERMINATION OF SILICON IN WATERS BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Basic Digestion Principles

Determinations by Atomic Absorption Spectroscopy and Inductively Coupled Plasma-Atomic Emission

GB Translated English of Chinese Standard: GB

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

Sulfotepp impurities in Chlorpyrifos EC formulations

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Hydra-C Application Note: 1076

novaa 800 D Atomic Absorption Spectrometer

POLYDEXTROSES. SYNONYMS Modified polydextroses; INS No DEFINITION DESCRIPTION. FUNCTIONAL USES Bulking agent, humectant, stabilizer, thickener

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

By Donna L. Damrau U.S. GEOLOGICAL SURVEY. Open-File Report Denver, Colorado

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS

Trace elemental analysis solutions for your application. June 6, 2018

Determination Of Pb (II), Cr (VI) And Cd (II) Contents In Some Ethiopia Spices After Microwave Digestion Using Atomic Absorption Spectroscopy

Single particle analysis using the Agilent 7700x ICP-MS

Full-text Available Online at and

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

P. Perämäki 1*, M. Pesonen 1 and J. Piispanen 2. Introduction

DRAFT EAST AFRICAN STANDARD

U.S. EPA SW-846 Method 6010C Using the Prodigy7 High- Dispersion ICP Introduction. Application Note - AN1305. Experimental

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

Maintenance with OQ. ZEEnit 700. Operational Qualification OQ within the scope of a maintenance for the. Atomic Absorption Spectrometer

Brooklyn College Department of Chemistry

Graphite Block Acid Digestion System ODLAB

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Automated determination of the uniformity of dosage in quinine sulfate tablets using a fiber optics autosampler

Mercury, total-in-sediment, atomic absorption spectrophotometry, nameless, direct

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4

The PinAAcle Family of AAS Features and Benefits

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS

HEAVY METALS IN LAKE NAKURU ATOMIC SPECTROSCOPY

Transcription:

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS Application Note Food Testing Author John Cauduro Agilent Technologies, Australia Introduction Trace elemental analysis of foods is essential to ensure that products are suitable for consumption. Accurate low-level determination of metals is especially important for foods that are consumed on a regular basis, such as grain products like wheat and rice. Analysis of elements such as Pb, Cd, Cr and Ni is important for ensuring product quality. When trace element analysis is required, a high sensitivity technique such as graphite furnace atomic absorption spectrometry (GFAAS) with Zeeman background correction. Features such as high sensitivity, low running costs, and accurate correction for complex backgrounds make GFAAS with Zeeman background correction suitable for the determination of trace levels of Pb, Cd, Cr and Ni in grain. The procedures set out in four Chinese National Methods relate to the determination of Pb, Cd, Cr and Ni in many food products using GFAAS.

In this study, Pb, Cd, Cr and Ni were determined in the Wheat Flour Standard Reference Materials (SRM) using the Agilent 280Z Zeeman GFAAS. The study was conducted in accordance with the following Chinese methods: GB Method 5009.12 2010 for Pb, GB 5009.15 2014 for Cd, GB 5009.123 2014 for Cr and GB 5009. 138 2003 for Ni. Experimental An Agilent 280Z AAS with transverse Zeeman background correction was used for all measurements to determine Pb, Cd, Cr and Ni in wheat flour. The instrument features the highly sensitive and accurate Agilent GTA 120 Graphite Tube Atomizer (GTA) and an Agilent PSD 120 Programmable Sample Dispenser (PSD) autosampler, as shown in Figure 1. The PSD automatically delivers measured volumes of the sample to the furnace. Figure 1. Agilent PSD 120 programmable sample dispenser The Agilent PSD 120 provides the capacity for up to 135 solutions and performs automatic standard preparation, modifier addition and overrange dilution. The 280Z AA comprises patented longitudinal graphite tube heating and a Constant Temperature Zone (CTZ) design. In addition to prolonging the graphite tube lifetime, these features provide consistent, uniform heating, which is essential for the accurate low-level determination of complex samples. The Tube-CAM furnace viewing camera, which is standard on the 280Z AA, provides real time viewing inside the graphite tube via the Agilent SpectrAA Software that controls the instrument. Still images and videos can be recorded to set the probe dispensing height and confirm optimum drying conditions, simplifying method development. For reduced baseline noise and increased sensitivity due to higher emission intensity, a single-element UltrAA hollow cathode lamp light source was used for Pb and Ni. A singleelement hollow cathode lamp light source was used for Cd and Cr. For all elements, a partition tube (coated) was used. Instrument operating conditions are listed in Table 1. Materials and reagents The PSD 120 autosampler rinse solution comprised a 0.01 % v/v non-ionic surfactant and 2% v/v of high purity nitric acid. Chemical modifiers were used to reduce matrix interferences and to stabilize analyte signals. Table 1. Agilent 280Z GFAAS instrument operating conditions. Parameter Setting Element Pb Cd Cr Ni Instrument mode Absorbance Absorbance Absorbance Absorbance Calibration mode Concentration Concentration Concentration Concentration Measurement mode Peak height Peak height Peak Height Peak Height Wavelength (nm) 283.3 228.8 357.9 232.0 Lamp type UltrAA lamp Hollow cathode lamp Hollow cathode lamp UltrAA Lamp Lamp current (ma) 10 4 7 7 Slit width (nm) 0.5 0.5 0.2 0.2 Replicates 2 2 2 2 Calibration standards (μg/l) 10, 30, 50 1, 2, 3 2, 4, 8 10, 25, 50 Sample volume (μl) 10 10 10 10 Sample introduction Auto-mix Auto-mix Auto-mix Auto-mix Background correction On On On On Modifier Premixed 2000 μg/ml NH 4 H 2 PO 4 + 120 µg/ml Mg(NO 3 ) 2 2000 mg/l NH 4 H 2 PO 4 300 mg/l Pd 150 mg/l Pd Modifier injection Type Co injection Co injection Co injection Co injection Modifier volume (μl) 5 5 5 5 2

The following reagents were used for the preparation of calibration standard solutions and samples: Single element standard solutions for Pb, Cd, Cr and Ni, 1000 mg/l (Agilent); Non-ionic surfactant, Trition X-100 surfactant (Agilent); Mixed GFAA matrix modifiers: 10 mg/ml ammonium phosphate, 600 μg/ml magnesium nitrate, 10% ammonium phosphate, 1% palladium nitrate (Agilent); High-purity nitric acid (Ultrapur, Merck); NIST 1567b Wheat Flour SRM, NRC-CNRC Durum wheat flour DUWF-1 (NRC-CNRC), 18 MΩ de-ionized water prepared using a Milli-Q purification system (Merck Millipore). Sample preparation To validate the method for Pb and Cd, NIST 1567b Wheat Flour SRM (Gaithersburg, USA) was prepared for analysis by microwave digestion. To validate the method for Cr and Ni, NRC-CNRC Durum wheat flour DUWF-1 (NRC-CNRC) was prepared for analysis by wet digestion. Microwave digestion for Pb and Cd Digests were prepared in triplicate using a Milestone UltraWAVE Single Reaction Chamber (SRC) microwave digestion system. The system, which serves both as a microwave cavity and reaction vessel, can reach high temperatures. Sealing of the vials was not required since the SRC was pressurized using nitrogen gas at 40 bar. Approximately 1.0 g of the sample was weighed into a quartz reaction vessel. 1 ml of hydrogen peroxide 30% Emsure (Merck) and 5 ml nitric acid 60% Ultrapur (Merck) were added. Each sample was thoroughly mixed using a vortex mixer. The samples were then digested and made up to a volume of 20 ml using 18 MΩ de-ionized water. The temperature program used for the microwave digestion process is shown in Table 2. Table 2. Parameters used for microwave digestion (where t is the time in minutes, and T1 and T2 are the programmed initial and final vessel temperatures). Wet digestion for Cr and Ni For preparation of digests for the determination of Cr and Ni, approximately 0.5 g of the sample was weighed into a Pyrex test tube. 1 ml of hydrogen peroxide 30% Emsure (Merck) and 5 ml nitric acid 60% Ultrapur (Merck) were added and thoroughly mixed using a vortex mixer. The test tubes were placed in a hot block at 80 C. They were removed when the solution level began to rise. The temperature was increased to 120 C for 1 hour and the test tubes were covered with a watch glass to allow reflux. Once the temperature had risen to 150 C, the watch glasses were removed to allow the water to boil off for 1 hour. The digests were made up to a volume of 20 ml using 18 MΩ de-ionized water. Samples were prepared in triplicate. Calibration standards Bulk standards of Pb, Cd, Cr and Ni were prepared from 1000 μg/l single element standards. All calibration standards were automatically prepared by the PSD 120 autosampler from 100 μg/l Pb, 5 μg/l Cd, 20 μg/l Cr and 50 µg/l Ni. The bulk standards were matrix matched to 20% HNO 3 to ensure consistency with the acid content of the digested samples. Results and Discussion The furnace programs used for the analysis of Pb, Cd, Cr and Ni in wheat flour are given in Table 3. Inert high-purity gas was used throughout all furnace programs. Gas flow rates were the same for all elements and were the default values in the SpectrAA Software. The Constant Temperature Zone (CTZ) of the GTA 120 graphite tube atomizer allows rapid heating, with a maximum ramp rate of 2000 C/s. The pre-emptive sampling capabilities of the PSD 120 ensure optimum analysis times as the autosampler proceeds to the next sample while the previous sample is still reading. The combination of these features means that analysis times were significantly decreased without compromising tube lifetime. To verify the four methods, Pb, Cd, Cr and Ni were analyzed in three replicate digests of a wheat flour SRMs. The results are given in Table 4. The agreement between the certified values and the measured results (average of three runs) confirm the accuracy of the methods. Step t (min) T1 ( C) T2 ( C) Power (W) 1 5.5 Ambient 200 1500 2 4.5 200 200 1500 3

Table 3. Furnace program for the analysis of Pb, Cd, Cr, and Ni in wheat flour. Pb Cd Cr Ni Step Temp ( C) Time (s) Temp ( C) Time (s) Temp ( C) Time (s) Temp ( C) Time (s) Read 1 Drying 85 10.0 85 10.0 85 10.0 85 10.0-2 Drying 95 25.0 95 25.0 95 25.0 95 30-3 Drying 120 5.0 120 5.0 120 5.0 120 5.0-4 Ashing 500 5.0 400 5.0 1300 5.0 1200 5.0-5 Ashing 500 10.0 400 5.0 1300 5.0 1200 5.0-6 Ashing 500 2.0 400 2.0 1300 2.0 1200 2.0-7 Atomization 2100 0.8 1800 0.7 2650 0.6 2800 0.8 Yes 8 Atomization 2100 1.0 1800 1.0 2650 1.0 2800 1.0 Yes 9 Clean out 2700 2.0 2300 2.0 2650 2.0 2800 2.0 - Table 4. Results for analysis of NIST 1567b Wheat Flour (Pb and Cd) and DUWF-1 (Cr and Ni) Durum Wheat Flour by GFAAS. Element Certified value (ppb) Measured value (mean, n=3) (ppb) Standard deviation (ppb) Recovery (%) Pb 10.4 ± 2.4 10.9 1.4 105 Cd 25.4 ± 0.9 23.4 0.22 92 Cr 23.0 ±- 9.0 21.5 2.1 93 Ni 170 ± 80 172 7.5 101 Method detection limits Method detection limits (MDLs) for each method required only 10 μl of sample and were calculated from 7 replicate analyses of a 0.3 ppb standard solution for Pb, 0.05 ppb standard solution for Cd, 0.1 ppb standard solution for Cr and 0.5 ppb solution for Ni. The absorbance peaks for Pb and Cd are shown in Figures 2 and 3 respectively. The calculated MDLs for Pb, Cd, Cr and Ni and the required MDLs specified in 5009.12 2010, GB 5009.15 2014, GB 5009.123 2014 and GB 5009. 138 2003 are given in Table 5. The MDLs for both elements obtained using the Agilent 280Z Zeeman GFAAS are lower than the MDLs specified in the GB methods. Figure 3. Absorbance peak from the measurement of a 0.05 ppb Cd standard solution Table 5. MDLs acquired per GB 5009.12 2010, GB 5009.15 2014, GB 5009.123 2014 and GB 5009. 138 2003 guidelines. A Student t factor of 3.143 was applied to give a 99% confidence. Element MDL 1 (µg/l) MDL 2 (µg/kg) MDL 2 (mg/kg) GB specified MDL (mg/kg) Pb 0.13 2.6 0.0026 0.005 Cd 0.019 0.37 0.00037 0.001 Cr 0.038 1.5 0.0015 0.01 Ni 0.16 6.0 0.006 0.06 1. Calculated MDL in solution 2.Calculated MDL in sample Calibration range The calibration curves for Pb, Cd, Cr and Ni are shown in Figures 4, 5, 6 and 7, respectively. All elements display excellent calibrations with correlation coefficients greater than 0.999. All standards were prepared automatically via the PSD 120. Figure 2. Absorbance peak from the measurement a 0.3 ppb Pb standard solution 4

Table 6. %RSD of a 5 ppb Pb standard measured periodically over 7 hours. Sample Measured value, (mean, n=26) (ppb) RSD (%) Wheat Flour 5.22 1.6 Figure 4. Calibration curve for Pb from 0 to 100 µg/l. Figure 5. Calibration curve for Cd from 0 to 5 µg/l. Figure 8. Long-term stability plot for a 5 ppb Pb standard measured periodically during the analysis of a wheat flour digest run over a 7-hour period. Conclusions Figure 6. Calibration curve for Cr from 0 to 20 µg/l. Figure 7. Calibration curve for Ni from 0 to 50 µg/l. Instrument stability The stability of the 280Z Zeeman GFAAS was investigated by analyzing a digest of a commercial wheat flour sample multiple times, interspersed with a 5 ppb Pb standard that was measured after every five samples (without recalibration or reslope). The measured results for the 5 ppb Pb standard were used to determine the long-term stability of the instrument. The %RSD was 1.6 over a seven-hour period. The results are given in Table 6 and with the stability plot displayed in Figure 8. There was only a 4% deviation in concentration from the initial reading. The Agilent 280Z GFAAS was used to determine Pb, Cd, Cr and Ni in digested wheat flour SRMs according to Chinese National Methods GB 5009.12 2010, GB 5009.15 2014, GB 5009.123 2014 and GB 5009. 138 2003. Calibrations were achieved over a large calibration range for all three elements. Good agreement with the certified values was obtained, demonstrating the accuracy of the method. The study showed that the Agilent 280Z GFAAS exceeds the Chinese method requirements for the determination of Pb, Cd, Cr and Ni in food products. With its excellent precision, faster analysis times when used in conjunction with the PSD 120 autosampler, long graphite tube lifetime and low argon gas usage, the 280Z GFAAS is ideally suited to meet the needs of food testing labs. Especially for labs that require accurate, sensitive, and cost effective instrumentation. 5

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2017 Published July 31, 2017 Publication number: 5991-8144EN