GC/MS: A Valid Tool for Soil Gas Hydrocarbons Speciation

Similar documents
Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Analysis of BTEX in Natural Water with SPME

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

Author. Abstract. Introduction

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.

Analytical Trap Comparison for USEPA Method 8260C

Atlantic RBCA. Guidelines for Laboratories. Tier I and Tier II Petroleum Hydrocarbon Methods

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

U.S. EPA Method 8270 for multicomponent analyte determination

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Methanol Extraction of high level soil samples by USEPA Method 8260C

Analysis of Bakken Oil Samples

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Enhanced Simultaneous Total ion and Selective Multi-Ion GC/MS Using a Combination of New Software and Low-Bleed Columns

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

DEVELOPMENT OF A SENSITIVE THERMAL DESORPTION GC-MS METHOD USING SELECTIVE ION MONITORING FOR A WIDE RANGE OF VOCS

Selection of a Capillary

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

CALIBRATION OF GC/MS METHOD AND VALIDATION OF THE MODIFIED SAMPLE PREPARATION FOR DETERMINATION OF POLYCYCLIC AROMATIC HYDROCARBONS

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Maximizing Sample Throughput In Purge And Trap Analysis

HYDROCARBONS, AROMATIC 1501

Spring 2010 updated March 24 Determination of Aromatics in Gasoline by Gas Chromatography- Mass Spectrometry: Comparison of Grades and Brands

Automated Characterization of Compounds in Fire Debris Samples

TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated

Pittcon P

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

The Application of Method TO-15 to Naphthalene Measurements in Indoor Air

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Selection of a Capillary GC Column

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

WADA Technical Document TD2003IDCR

Sulfotepp impurities in Chlorpyrifos EC formulations

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

is given for the isotopic fingerprinting methodology.

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Study of Residual Solvents in Various Matrices by Static Headspace

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.)

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

TDTS 16 Round-the-clock, on-line and cryogen-free monitoring of hydrocarbons from acetylene to trimethylbenzene in ambient air

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

Substance identification and naming convention for hydrocarbon solvents

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

TPH Methods and Measurements and Petroleum Vapor Intrusion (PVI) Risks

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Volatile Organic Compounds in Every Day Food

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Solid Phase Micro Extraction of Flavor Compounds in Beer

GC/MS Application Note

Chem 230, Fall, 2014 Homework Set # 4 LONG ANSWER SOLUTIONS

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

METHOD 3600B CLEANUP

VUV ANALYTICS VGA-100 GC DETECTOR Gas Chromatograph Agilent 6890 equipped with a 7683 model autosampler Restek 30m x 0.25mm x 0.

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Highly Sensitive and Rugged GC/MS/MS Tool

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

EPA TO-17 Volatile Organic Compounds

Introduction to GC/MS

Thermal Desorption Technical Support

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Real Time On-Site Odor and VOC Emission Measurements Using a znose

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

Understanding Gas Chromatography

Ultra-Inert chemistry for Trace Level Analysis

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Transcription:

International Journal of Analytical Mass Spectrometry and Chromatography, 2015, 3, 54-62 Published Online September 2015 in SciRes. http://www.scirp.org/journal/ijamsc http://dx.doi.org/10.4236/ijamsc.2015.33007 GC/MS: A Valid Tool for Soil Gas Hydrocarbons Speciation Laura Clerici 1, Emanuela Mongini 1, Pasquale Robles 1, Pierluisa Dellavedova 2, Maria Antonietta De Gregorio 1* 1 U.O. Laboratorio di Milano, Sede Laboratoristica di via Juvara Settore Laboratori ARPA LOMBARDIA, Milano, Italia 2 Settore Laboratori ARPA LOMBARDIA, Milano, Italia Email: * a.degregorio@arpalombardia.it Received 3 July 2015; accepted 25 September 2015; published 28 September 2015 Copyright 2015 by authors and Scientific Research Publishing Inc. This work is licensed under the Creative Commons Attribution International License (CC BY). http://creativecommons.org/licenses/by/4.0/ Abstract This work describes an alternative method based on GC/MS technique with SCAN-ion approach for speciation of hydrocarbons contained in soil gas matrices and sampled on solid sorbent tubes (coconut shell charcoal). Keywords Hydrocarbons, Soil Gas Matrices, Solid Sorbent Tubes, Soil Gas, GC/MS 1. Introduction Soil gas is defined as the atmosphere present in pore spaces of soil matrices [1]. These pores can undergo contamination phenomena, due to exposure to volatile pollutants. The latter can derive from different sources: contaminated soils and ground waters. Chemical-physical, geological and hydrogeological characteristics of each soil make its contamination highly compound- and site-specific. Therefore, one of the most recent aims of the analytical and environmental chemistry consists in hydrocarbons determination in soil gas matrices and their speciation. Hydrocarbons represent a class of numerous and heterogeneous organic compounds ranging from linear, cyclic and branched saturated and unsaturated alkanes to aromatic and polycyclic ones. Each of these species is characterized by a peculiar value of toxicity and, as a consequence, their effect on the human health and environment is strongly different. More in detail, the speciation consists in the determination and quantification of hydrocarbons fractions in soil gas matrices. * Corresponding author. How to cite this paper: Clerici, L., Mongini, E., Robles, P., Dellavedova, P. and De Gregorio, M.A. (2015) GC/MS: A Valid Tool for Soil Gas Hydrocarbons Speciation. International Journal of Analytical Mass Spectrometry and Chromatography, 3, 54-62. http://dx.doi.org/10.4236/ijamsc.2015.33007

In order to evaluate hydrocarbons carcinogenic effects, some specific target molecules have been identified: polycyclic aromatic hydrocarbons (PAHs), benzene, etc. On the contrary, for toxic non-carcinogenic effects, specific hydrocarbons fractions are taken into account (e.g., in Italy for different environmental matrices C5-C8 aliphatic, C9-C12 aliphatic, C9-C10 aromatic, C11- C12 aromatic, C13-C18 aliphatic, C13-C22 aromatic and C19-C36 aliphatic have been identified) [2]. If hydrocarbons speciation in contaminated soil and water matrices can be easily analyzed by the use of official analytical methods, the discussion about the more appropriate method for the analysis of soil gases collected on different supports is still open. Among the official methods [3]-[5], one is based on hydrocarbon aromatic fraction quantification by extraction of characteristic ions and hydrocarbon aliphatic fraction quantification by total ion integration and subsequent subtraction of internal standard and target compounds peaks. This method is valid up to C12 aliphatic compounds and up to C10 aromatic compounds and is related to the use of canisters in the absence of any extraction solvent [5]. Another official method is useful for soil and water matrices [3]. It is based on the use of traditional detectors (FID and PID) that discriminate the different hydrocarbons fractions only on the base of the retention time and do not provide any information on the chemical species present in the samples. Our needs, however, require not only the determination of C11-C12 aromatic hydrocarbons and C13-C18 aliphatic fractions (higher than C12), but also the use of CS 2 as an extraction solvent to elute hydrocarbons fraction from soil gas matrices sampled on solid sorbent tubes (coconut shell charcoal). Therefore, in order to develop a unique, even empirical method for this purpose, in this work we demonstrate that gas-chromatography coupled with mass spectrometry is an effective technique for the speciation of hydrocarbons collected on chemical desorption solid sorbent tubes. 2. Experimental 2.1. Materials and Instruments A standard solution containing all analytes (1000 µg/ml for each analyte: n-pentane, 2-methylpentane, 2,2,4- trimethylpentane, n-nonane, n-butylcyclohexane, n-decane, 1,2,4-trimethylbenzene, naphthalene, butylbenzene, pentylbenzene, biphenyl, n-tetradecane, n-hexadecane, n-octadecane; Solution A) and an internal standard solution (ISTD, 2,5-dibrometoluene, 5000 µg/ml) were purchased by CPAChem and used as received. They were bought as solutions in CS2. CS2 (99.9% purity low benzene) and solid sorbent tubes (coconut shell charcoal) for chemical desorption process were purchased by Sigma Aldrich. Real samples (SAMPLE 1-8), contaminated with aliphatic hydrocarbons, were sampled on solid sorbent tubes reported above. All samples were analyzed by a GC/MS technique using an Agilent 7890A GC equipped with a 5975C inert MSD (mass spectrometer detector) with triple axis detector. The Agilent 7890A GC was interfaced with an Agilent 7683B Series Injector Automatic Liquid Sampler. A Thermo TR5-MS (60 m 0.25 mm 1 µm) column was employed. 2.2. Certified Reference Materials (CRM) Preparation for Calibration Method Calibration standards were prepared by dilution of Solution A with a diluted internal standard solution (d-istd, with a concentration of 10 μg/ml in CS 2 ), as described in Table 1. At each STD in the Table 1 corresponds a calibration level (CAL LEV). All the STD solutions reported in Table 1 were injected in a GC/MS instrument interfaced with an automatic liquid sampler (Section 2.5). 2.3. Fortified Materials Preparation for Method Validation (Precision and Recovery) The availability of certified reference materials (CRM) is limited [6]. Moreover, because CRM are not commercially available for this kind of analyses (sample of solid sorbent tube coconut shell charcoal), fortified materials were prepared in-house. For this purpose, starting from Solution A, STD 8 and STD 5 were prepared as reported in Table 1 but using pure CS 2 instead of diluted internal standard solution. Therefore, 100 μl of Solution A, STD 8 and STD 5 were spiked on 200 mg of coconut shell charcoal (solid sorbent tube). This procedure was repeated ten times for each solution in order to obtain a series of ten samples for each concentration. 55

2.4. Fortified Materials Analysis from Chemical Desorption of Sorbent Tubes All materials reported in the Section 2.3 were extracted with 2 ml of d-istd (1 ml of d-istd for each 100 mg of carbon) [7]. 2.5. GC/MS Analytical Procedure All samples were processed using the following GC/MS conditions. Oven Program: 40 C for 8 min; then 4 C/min to 100 C for 0 min; then 15 C/min to 320 C for 9 min; then 40 C/min to 330 C for 1 min. Injection Volume: 1 µl. SS Inlet: Mode splitless. Heater: 250 C. Column: TR5- MS 60 m 250 µm 1 µm. Flow: 1.2 ml/min. MSD Parameters: MS Source 230 C, MS Quadrupole 150 C. SIM Parameter: characteristic ion Target and Qualifier for each VOC compounds. SCAN Parameter: Low Mass: 35.0 amu; High Mass: 350 amu. SIM and SCAN modes were simultaneously acquired. 3. Results and Discussion Starting from the data reported in the official methods [3] [5] and on the base of our experience and technical requirements, an accurate mass spectra investigation of numerous chemical species representative of each hydrocarbons fraction (C5-C8 aliphatic, C9-C12 aliphatic, C9-C10 aromatic, C11-C12 aromatic, C13-C18 aliphatic) was carried out. Table 2 reports standard markers and corresponding ions (m/z) used to quantify each hydrocarbons fraction, as well as the hydrocarbons range in terms of retention time. All chromatograms were contemporary acquired by SIM (selected ion monitoring) and SCAN-mode. SIMmode analysis allows the quali-quantitative determination of target compounds, such as BTEX (benzene, toluene, Table 1. Calibration standards (STD) preparation and hydrocarbons concentration in each fraction. STD and calibration level STD prepation (μg/ml) STD concentration (μg/ml) for each analyte STD 8 and CAL LEV8 100 µl SolA/1mL diluted ISTD 100 STD 7 and CAL LEV7 500 µl STD 8/1mL diluted ISTD 50 STD 6 and CAL LEV 6 200 µl STD 8/1mL diluted ISTD 20 STD 5 and CAL LEV5 100 µl STD 8/1mL diluted ISTD 10 STD 4 and CAL LEV4 500 µl STD 5/1mL diluted ISTD 5 STD 3 and CAL LEV3 200 µl STD 5/1mL diluted ISTD 2 STD 2 and CAL LEV2 100 µl STD 5/1mL diluted ISTD 1 STD 1 and CAL LEV1 500 µl STD 2/1mL diluted ISTD 0.5 Table 2. Target ions (m/z), standard markers and retention time window for each hydrocarbons fraction. * From Ref. [5]. Hydrocarbons Fraction m/z Standard Marker Beginning Marker Ending Marker C5-C8 Aliphatic saturated, unsaturated, cyclic hydrocarbons 57-71 - 41 n-pentane 2-methylpentane 2,2,4-trimethylpentane n-pentane n-nonane C9-C12 Aliphatic saturated, unsaturated, cyclic hydrocarbons 57-71 - 41 n-nonane n-butylcyclohexane n-decane n-dodecane n-nonane naphtalene C9-C10 Aromatic 120-134 * 1,2,4-trimethylbenzene butylbenzene 0.1 min after o-xylene naphtalene C11-C12 Aromatic 148-154 - 162-115 Pentylbenzene biphenyl n-tetradecane n-hexadecane n-octadecane pentylbenzene 0.1 min after biphenyl C13-C18 Aliphatic saturated, unsaturated, cyclic hydrocarbons 57-71 - 41 0.1 min after n-dodecane 0.1 min after n-octadecane 56

ethylbenzene, xylenes), MTBE (methyl-ter-butyl-ether), halogenated VOC (volatile organic compounds), etc. SCAN-mode analysis permits the identification and determination of hydrocarbons fractions by characteristic ions extraction (SCAN-ion quantification) within the typical retention time of each fraction. SCAN-ion approach allows achieving many benefits. Among them, the most important one consists in a correct and easy integration of more critical fractions: C5-C8 and C13-C18 aliphatic. In fact, C5-C8 aliphatic fraction could be affect by the presence of solvent front (CS 2 ), whereas column bleed, strictly related to the oven temperature increase, could have a negative effect on the C13-C18 aliphatic fraction determination, in particular at low concentration, causing an analytical sensitivity drop. SCAN-ion approach, based on the characteristic ions extraction, improves analytical performances, as described in Figure 1. 3.1. Initial Calibration The relationship between detector response of each analyte and collective range concentration was calculated by the RRFs approach (Equation (1)): Ax Cis RRF = (1) A C where RRF = relative response factor, A x = sum of ions areas of each fraction to be measured, C is = concentration of internal standard spiking mixture, µg/ml, A is = area of the primary ion for the internal standard, C x = concentration of hydrocarbons fraction in the calibration standard, µg/ml. The calculated %RDS (Equation (2)) the RRF for each hydrocarbons fraction in the calibration method resulted to be less than 30%, as required by the technical acceptance criteria for initial calibration [5]. is x Figure 1. (A) (B) SCAN-mode acquired chromatograms and hydrocarbon fractions quantification using total ion integration; (C) (D) SCAN-mode acquired chromatograms and hydrocarbon fractions quantification using SCAN-ion approach by characteristic ions extraction (m/z 57, 71, 41). 57

( ) ( ) SD RRF %RSD = 100 (2) RRF mean where SD (RRF) = standard deviation of initial response factors for each fraction, RRF(mean) = mean of initial relative response factors for each fraction. However, when these technical criteria were not satisfied, linear and nonlinear regressions were used, as for the C5-C8 aliphatic fraction (linear regression R 2 0.98). Moreover, following the recommendations reported in [8], other technical acceptance criteria, were satisfied, as: The RRT (Relative Retention Time) for each hydrocarbons fraction at each calibration level was within 0.06 RRT units of the mean RRT for the fraction; The Retention Time (RT) shift of ISTD at each calibration level was within 20 s of the mean RT over the initial calibration range for ISTD; The area response of the internal standard at each calibration level was within 40% of the mean area response over initial calibration range for ISTD. 3.2. Precision, Recovery and Uncertainty As described in Section 2.3, ten fortified materials at three different concentrations were prepared, in order to evaluate quality data of the method. Tables 3-5 summarize mean of precision data, CV% and recovery% obtained for all fractions at three concentrations (µg/ml) reported in the Table 1. It is worth to notice that in all data reported in Tables 3-5 CV% and recovery% are in line with the technical acceptance criteria reported in the literature [5]. All data were analyzed by means of statistical tests: Shapiro-Wilk normality test and Dixon test for identify and reject outliers. Table 3. Mean value, CV% and recovery% for each hydrocarbons fraction at concentration corresponding to STD 1 in Table 1. Fractions Mean CV% Recovery% C9-C12 Aliphatic 2.36 7.79 118 C9-C10 Aromatic 1.20 5.41 120 C11-C12 Aromatic 0.91 5.71 91 C13-C18 Aliphatic 1.81 4.80 120 Table 4. Mean value, CV% and recovery% for each hydrocarbons fraction at concentration corresponding to STD 4 in Table 1. Fraction Mean CV% Recovery% C5-C8 Aliphatic 19.42 13.03 129 C9-C12 Aliphatic 16.82 4.08 84 C9-C10 Aromatic 7.86 5.43 79 C11-C12 Aromatic 5.76 5.88 58 C13-C18 Aliphatic 12.77 5.23 85 Table 5. Mean value, CV% and recovery% for each hydrocarbons fraction at concentration corresponding to STD 7 in Table 1. Fraction Mean CV% Recovery% C5-C8 Aliphatic 201.15 4.09 134 C9-C12 Aliphatic 158.79 4.64 79 C9-C10 Aromatic 81.79 3.43 82 C11-C12 Aromatic 67.73 1.71 68 C13-C18 Aliphatic 131.29 6.80 88 58

As an example, in Figure 2 the distribution obtained for C13-C18 fraction at concentration values corresponding to STD 1 in Table 1 is reported. The LOQ of the present method for each hydrocarbons fraction was mathematically estimated multiplying the standard deviation values, obtained from fortified materials at concentration values corresponding to STD 1 (see Table 1), by 5 (Table 6). Uncertainty was estimated by metrological approach calculating each component to the final uncertainty [9]. For all fractions the uncertainty value was estimated about ±60%. 4. Real Samples Analysis Eight real samples collected on solid sorbent tubes (SAMPLE 1-8), contaminated with C5-C8 and C9-C12 aliphatic hydrocarbons fraction, were quantified by both the traditional TIC-SCAN method [5] and the present one (SCAN-ion), in order to prove the performances of our innovative analytical approach. It is worth to notice that the results (Table 7, Table 8 and Figures 3-6) obtained by the two methods are rather comparable and correlated with a mean difference that lower than 30%, which is included in the uncertainty. Figure 2. Normal distribution of C13-C18 fraction data. Table 6. Standard deviation (SD) and limit of quantification (LOQ) for all hydrocarbons fractions. Fraction SD LOQ (µg/ml = 100 mg Carbon) C5-C8 Aliphatic 2.53 12.66 C9-C12 Aliphatic 0.18 0.92 C9-C10 Aromatic 0.07 0.33 C11-C12 Aromatic 0.05 0.26 C13-C18 Aliphatic 0.09 0.43 Table 7. A comparasion between the C5-C8 aliphatic hydrocarbons fraction concentration of real samples quantified by TIC-scan and SCAN-ion approaches. SAMPLES C5-C8 ALIPHATIC by TIC-scan concentration (µg/ml = 100 mg Carbon) C5-C8 ALIPHATIC by SCAN-ion concentration (µg/ml = 100 mg Carbon) SAMPLE 1 51 65 SAMPLE 2 237 190 SAMPLE 3 79 61 SAMPLE 4 113 80 SAMPLE 5 133 88 SAMPLE 6 257 185 SAMPLE 7 108 82 SAMPLE 8 180 131 59

Table 8. A comparasion between the C9-C12 aliphatic hydrocarbons fraction concentration of real samples quantified by TIC-scan and SCAN-ion approaches. SAMPLES C9-C12 ALIPHATIC by TIC-scan concentration (µg/ml = 100 mg Carbon) C9-C12 ALIPHATIC by SCAN-ion concentration (µg/ml = 100 mg Carbon) SAMPLE 1 125 92 SAMPLE 2 90 71 SAMPLE 3 123 84 SAMPLE 4 138 118 SAMPLE 5 25 22 SAMPLE 6 157 107 SAMPLE 7 38 37 Figure 3. Trend of SCAN-ion vs TIC-scan for C5-C8 aliphatic hydrocarbons fraction. Figure 4. Correlation between data (Table 7) derived by TIC-scan and SCAN-ion approaches. 60

Figure 5. Trend of SCAN-ion vs TIC-scan for C9-C12 aliphatic hydrocarbons fraction. 5. Conclusions Figure 6. Correlation between data (Table 8) derived by TIC-scan and SCAN-ion approaches. As an alternative to official methods, in this work an innovative empirical GC/MS method for hydrocarbons speciation in soil gas matrices was proposed. It is based on the use of a mass spectrometer as the detector and on SCAN-ion acquisition approach that allows separating the different fractions and identifying the chemical species present in all fractions, permits the use of an extraction solvent (CS 2 ) to elute hydrocarbons fraction from solid sorbent tubes (coconut shell charcoal) and extends the determination to hydrocarbons fraction up to C18. Good performances in terms of precision, recovery and uncertainty were obtained. The analyses of real samples provided promising results compared to those from traditional approaches. The simplicity and velocity of the method make it a good candidate for wide applications in routine work for analyses of soil gas matrices sampled on solid sorbent tubes. References [1] Mayer, C.L. (In Press) Guidance Document for Soil-Gas Surveying, Prepared under EPA EMSL-LV Contract n. 68033245. Lockheed Engineering and Sciences Company, Las Vegas. 61

[2] Cambiaghi, M., Cogliati, N. and Maspero, M. (2009) Idrocarburi e analisi di rischio: problemi e interpretazioni. Acqua & Aria, 5, 18-23. [3] Department of Environmental Protection (2004) Method for the Determination of Volatile Petroleum Hydrocarbons (VPH), Revision 1.1. Department of Environmental Protection, Massachusetts [4] Department of Environmental Protection (2004) Method for the Determination of Extractable Petroleum Hydrocarbons (EPH), Revision 1.1. Department of Environmental Protection, Massachusetts. [5] Department of Environmental Protection (2009) Method for Determination of Air-Phase Petroleum Hydrocarbon (APH), Final Revision 1. Department of Environmental Protection, Massachusetts. [6] EURACHEM (1998) The Fitness for Purpose of Analytical Methods A Laboratory Guide to Method Validation and Related Topics. EURACHEM Guide, 22. [7] UNI EN 13649-2002. Emissioni da sorgente fissa Determinazione della concentrazione in massa di singoli composti organici in forma gassosa. Metodo mediante carboni attivi e desorbimento con solvente; 7 Procedimento analitico, 7.1. Desorbimento del campione. [8] EPA (1999) Compendium Method TO-15. [9] EURACHEM-CITAC Guide (2011) Quantifying Uncertainty in Analytical Measurement, Third Edition. 62