Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin

Similar documents
A Comparison of Arsenic Speciation

Elemental Scientific. Stability. Elemental Scientific

Elemental Scientific. Spike Recovery. Elemental Scientific

Automated Determination of PPQ Levels of Thorium in High Purity Copper

Automating Sample Purification MC-ICPMS. Elemental Scientific ICP ICPMS AA

A Comparison of Three Methods for Arsenic Speciation in Biological Tissues. May Nguyen Brooks Rand Labs Seattle, WA

Elemental Scientific. Spike Recovery. Elemental Scientific

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS

DISCLAIMER: This method:

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Characterization of Arsenic Species in Apple Juice using a NexSAR HPLC-ICP-MS Speciation Analysis Ready Solution

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

PROCEDURES. Pharmacopeial Forum 2 Vol. 36(1) [Jan. Feb. 2009]

Arsenic Speciation in Chicken. Mary Carson, Ph.D. FDA/CVM/OR/ Division of Residue Chemistry

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Pharmacopeial Forum Vol. 36(1) [Jan. Feb. 2010] 1

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Determination of trace anions in concentrated hydrofluoric acid

J. Serb. Chem. Soc. 78 (10) S117 S126 (2013) Supplementary material. SUPPLEMENTARY MATERIAL TO Analytical methods for arsenic speciation analysis

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

Intercomparison of Extraction and Analytical Techniques for Inorganic Arsenic in Seaweed

Determination of urea in ultrapure water by IC-MS/MS

Development of a Sampling and Analysis Method for the Measurement of Hexavalent Chromium in Ambient Air

Ion Chromatography (IC)

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

GAS CHROMATOGRAPHY (GC)

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Samples compliant with the WATER

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

BECKMAN COULTER QbD1200 Total Organic Carbon Analyzer Critical Measurements Made Simple

ION CHROMATOGRAPHY SYSTEM S 150

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

110 Lowell Rd Hudson, NH USA Tel:

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

Automation of the radiochemical procedures for the sequential separation of radionuclides

Hydra-C Application Note: 1076

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Available online Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article

Rapid screening of arsenic species in urine from exposed human by inductively coupled plasma mass spectrometry with germanium as internal standard

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

á233ñ ELEMENTAL IMPURITIES PROCEDURES

MM800 (a) Ion Exchange and ICP/MS of Uranium in Water. 1.0 Scope and Application

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Product Descriptions and Offerings

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Volume 6, Issue 2, January February 2011; Article-015

Analytical Approaches to Arsenic

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Multi-residue analysis of pesticides by GC-HRMS

Dartmouth Trace Element Analysis Core: Lab methods

Innovative. Intuitive. Concentration (ppb) R = Versatile. Hydra II C Mercury Analyzer

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Appendix: Laboratory Testing Methods

DETECTING AND QUANTIFYING LEWISITE DEGRADATION PRODUCTS I N ENVIRONMENTAL SAMPLES USING ARSENIC SPECIATION*

881 Compact IC pro 882 Compact IC plus. Compact ion chromatography for intelligent, reliable and precise analysis

Schedule. Draft Section of Lab Report Monday 6pm (Jan 27) Summary of Paper 2 Monday 2pm (Feb 3)

HPLC Winter Webinars Part 2: Sample Preparation for HPLC

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Advances in Interference Removal for Accurate Arsenic Analysis in Food and Beverages

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

JOURNAL OF. Flow Injection. Analysis FIA 研究懇談会会誌

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Agilent Technologies at TIAFT 2013

The Theory of HPLC. Quantitative and Qualitative HPLC

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Determination of the Halogen Elements in the Deep UV Region of the Spectrum by ICP - OES

Pelagia Research Library

Examples of Method Validation Studies Conducted in Different Economies

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Introduction to Chromatographic Separations (Chapter 1) Many determinations involve separation followed by analysis chromatography electrophoresis

On the use of Deming and quadratic calibrations in analytical chemistry

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Automated and accurate component detection using reference mass spectra

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Application note. Determination of exchangeable cations in soil extracts using the Agilent 4100 Microwave Plasma-Atomic Emission Spectrometer

Transcription:

Elemental Scientific prepfast IC: Inline Autodilution Method to Eliminate Species Interconversion for LC-ICPMS Based Applications Authors: C. Derrick Quarles, Jr. *, Patrick Sullivan, M. Paul Field, Hwan Kim, and Daniel R. Wiederin Derrick.Quarles@icpms.com Use of an Inline Autodilution Method to Eliminate Species Interconversion for LC-ICPMS with Elemental Scientific's prepfast IC Introduction Trace element analysis is vital for industrial, government, and academic communities. Knowing the total concentration of an element in a given sample is important, but it does not tell the whole story. The chemical species (e.g., As (III), As (V), or AsCH 5 O 3 ) plays an important part in understanding how it can affect the environment, its bioavailability, or its toxicity. For example, inorganic arsenic (As (III)) is more toxic than organic arsenic (monomethylarsonic acid), whereas organic mercury (methylmercury) is more toxic than its inorganic form (Hg (II)). 1 The most common technique for trace metals analysis is inductively coupled plasma (ICP) coupled with either an optical emission spectrometer (-OES) or mass spectrometer (-MS). However, to determine the species of an element the chemical forms need to be selectively separated and the most common technique for this has become liquid chromatography (LC). Combining the two techniques, LC- ICPMS has become a very powerful and important tool for speciation of clinical, biological, food, pharmaceutical, and environmental based samples. 2 Typically, trace elemental analysis measurements are done by acidifying the sample (e.g., 2% HNO 3 ) and introducing it into an ICPMS (or ICP-OES). When determining the chemical species in a given sample, the sample integrity is paramount, and thus the sample must be kept in its native form. Species interconversion (the process of a chemical species changing form) becomes more likely during sample preparation as it is handled and exposed to external conditions. Yet most analytical columns used for chromatography have a low mass loading capacity which require diluted samples and/or small sample volumes (µl s). In order to address this issue we at Elemental Scientific (ESI) have developed an all-inclusive autosampler instrument, prepfast IC, which can be used for both total concentration analysis and speciation analysis. One of the key features of this instrument is that the samples are autodiluted inline, eliminating the need for sample preparation thus keeping the sample integrity intact. The other added benefit is that the analyst has the option to use the instrument for totals or speciation depending on the need at that given time, without having to attach multiple pieces of equipment to the frontend of the ICPMS. The prepfast IC is completely automated giving the user two instruments in one. The following study will be focused on arsenic species in urine in order to demonstrate the capabilities of the prepfast IC. This study will look at the precision and robustness for a syringe driven LC method, the advantages of autocalibration and autodilution of samples versus manual preparation, and figures of merit. Elemental Scientific 7277 World Communications Drive Omaha, NE 68122 Tel: 1-402-991-7800 sales@icpms.com www.icpms.com 1

Instrumentation In many laboratories adding an additional LC-ICPMS system may be too costly, thus ESI has developed and introduced the prepfast IC that is capable of providing total elemental analysis and elemental speciation within the same system. This system allows the user to seamlessly switch between total metal analysis and speciation without having to change any hardware, solutions, or samples. In addition no sample preparation is needed as standards and samples can be autocalibrated/autodiluted inline. Load column with sample Eluent 1 separates AsB, DMA, and As (III) Eluent 2 removes MMA immediately Auto switching between speciation and total metals Autocalibrate - Total metals - Speciation Autodilute - Total metals - Speciation Ion chromatography Gradient elution syringe pump Automated tuning Syringe loading - Viscous samples - Allows for all sample types Completely metal free system Eluent 2 separates MMA and As (V) Cl is retained onto column and eluted during wash step Figure 2. Illustration of the chromatographic separation of AsB, DMA, As (III), MMA, and As (V) as they elute off of the anion exchange column. Figure 1. Chemical structure of each arsenic species analyzed in this method. 2

Column Chemistry An anion exchange column is used for separating 5 species of arsenic (Fig. 1) in urine, aresenobetaine (AsB), dimethylarsinic acid (DMA), arsenite (As (III)), monomethylarsonic acid (MMA), and arsenate (As (V)). The separation involves a two-step elution gradient that varies in ph and ionic strength (Fig. 2). AsB, DMA, and As (III) elute in step 1 and MMA and As (V) elute in step 2. The Cl matrix, which can act as an interference and potentially cause high backgrounds, is eluted during the wash step using the FAST valve, ensuring that it never reaches the nebulizer. An ideal method allows for the user to have a range of sample dilutions, therefore dilution factors from 0X 100X were evaluated. All sample dilutions were performed using the inline autodilution feature. After the sample is diluted, it is injected directly onto the column; this minimizes the time that the sample can undergo any species interconversions. Figure 3 shows the chromatogram for 30X, 50X, and 100X urine samples that were spiked with 10 ppb arsenic species. Dilution factors from 0X 20X displayed varying retention times, suggesting that amount of sample (~ 23 µl of sample injected) was beyond the loading capacity of this anion exchange column. Inline autodilution factors of 30X up to 100X urine samples displayed the exact same retention times. Table 2 displays the retention times for each species using different dilution factors and different columns. A very small variation was seen for As (III) retention times (± 2.7 s) when changing columns, while all other species were ~ 1 s or less in peak variation. To test the reproducibility of the column and the syringe driven chromatography, 20 urine samples spiked with 5 ppb arsenic species were analyzed. In between each spiked urine sample a urine blank sample was analyzed to check for carryover from sample to sample. Table 1 displays the precision for each arsenic species tested during this experiment. The blank had no noticeable carryover and the background changed very little from sample to sample, 1.2 %RSD. The 5 ppb spiked samples showed good precision with the highest variation of 2.6 %RSD for AsB and all other species 1.4 %RSD. Autodilution of 10 ppb Spiked Urine Samples Table 1. Reproducibility of the column with a syringe driven chromatography method. % RSD (n=20) AsB DMA As (III) MMA As (V) Blank 1.2 1.2 1.0 0.8 0.7 5 ppb Spiked Urine 2.6 1.3 1.4 1.4 1.4 Alternating urine blank and spiked urine samples. 40 samples in total (20 blank and 20 samples). Figure 3. 10 ppb spiked urine samples analyzed at a 30X, 50X, and 100X dilution factor using the inline autodilution feature. Table 2. Effects of autodilution and changing of columns on the retention time for each arsenic species. Retention Times (s) Dilution Factor AsB DMA As (III) MMA As (V) Column A 30X 103.3 ± 0.8 125.4 ± 0.8 160.8 ± 2.1 228.9 ± 0.6 270.7 ± 0.8 50X 103.1 ± 0.5 125.6 ± 0.8 161.9 ± 1.5 229.2 ± 0.7 271.4 ± 0.6 100X 103.2 ± 0.6 126.0 ± 0.9 162.9 ± 0.9 229.5 ± 0.6 271.6 ± 0.6 Average All 103.2 ± 0.1 125.7 ± 0.3 161.9 ± 1.1 229.2 ± 0.3 271.2 ± 0.5 Column B 30X 103.0 ± 0.9 123.8 ± 0.8 156.6 ± 1.4 228.1 ± 0.5 268.9 ± 0.7 50X 102.5 ± 0.4 124.3 ± 0.3 157.1 ± 0.9 228.5 ± 0.7 269.3 ± 0.6 100X 102.7 ± 0.6 123.6 ± 0.4 157.5 ± 0.8 228.4 ± 0.7 269.5 ± 0.9 Average All 102.7 ± 0.3 123.9 ± 0.4 157.1 ± 0.5 228.3 ± 0.2 269.2 ± 0.3 Column A & B All 103.0 ± 0.3 124.8 ± 1.0 159.5 ± 2.7 228.8 ± 0.5 270.2 ± 1.2 3

Manual Dilution vs Inline Autodilution Autocalibration of Arsenic Species A comparison of manual dilution calibrations to inline autocalibrations showed similar chromatograms (Fig. 4) and resulted in almost identical sensitivity. The calibration curves ranging from 0 25 ppb for these two methods were used to evaluate the effects of time until analysis for a manual sample preparation method compared to the inline autodilution method. Preliminary tests with spikes of 1 ppb arsenic suggested up to a 43 % conversion of As (III) to As (V) after 24 h. Urine samples (from same urine pool) were spiked with 10 ppb of each arsenic species, individually. The urine samples were analyzed directly after the spiking process (t = 0 h) and at 1 h, 3 h, 6 h, 12 h, and 24 h after spiking. The overlay of the time study is displayed for the manual dilution method (Fig. 5) and the inline autodilution method (Fig. 6). The As (III) showed no conversion to As (V) at t = 0 h for manual and inline dilution methods. However, at t = 6 h the manual dilution showed a detectable peak at both the As (III) and As (V) elution times. The measured concentration for the complete 24 h study can be found in Table 3. The average recovery for the manual dilutions was found to be 4 11 % less than that of the autodilution method for AsB, DMA, MMA, and As (V). The poor recovery for As (III) in the manual dilution mode is due to the species interconversion to As (V). Figure 4. Chromatograms for the manual dilution calibration (top) and inline autocalibration (bottom). Table 3. Manual dilution vs inline autodilution. Measured Concentration (ppb) % Recovery Manual Dilution Time (h) AsB DMA As (III) MMA As (V) AsB DMA As(III) MMA As (V) 0 9.5 10.3 9.9 8.8 9.6 95 103 99 88 96 1 9.5 7.9 9.5 9.5 9.5 95 79 95 95 95 3 9.0 9.1 9.2 9.4 9.4 90 91 92 94 94 6 8.7 8.7 9.1 9.3 9.3 87 87 91 93 93 12 8.7 9.0 7.8 9.1 9.4 87 90 78 91 94 24 9.2 8.9 6.1 9.4 9.4 92 89 61 94 94 Measured Concentration (ppb) % Recovery Inline Autodilution Time (h) AsB DMA As (III) MMA As (V) AsB DMA As(III) MMA As (V) 0 10.9 10.2 9.7 11.0 11.0 109 102 97 110 110 1 10.2 10.0 10.0 10.6 10.1 102 100 100 106 101 3 9.0 9.1 10.6 9.7 11.3 90 91 106 97 113 6 9.5 8.4 10.8 10.6 10.2 95 84 108 106 102 12 9.1 8.8 9.2 9.4 10.0 91 88 92 94 100 24 9.9 9.8 10.6 9.5 10.4 99 98 106 95 104 4

Manual Dilution Inline Autodilution Figure 5. Manual dilutions for each arsenic species at time = 0, 1, 3, 6, 12 and 24 h. Individual arsenic species spiked into urine at a concentration of 10 ppb. Figure 6. Inline autodilutions for each arsenic species at time = 0, 1, 3, 6, 12 and 24 h. Individual arsenic species spiked into urine at a concentration of 10 ppb. 5

Table 4. The amount of As (III) that converted to As (V) over a 24 h time period for manual dilution compared to inline autodilution sample preparation. Manual Dilution Inline Autodilution 10 ppb As (III) Spiked Urine Samples Time (h) As (III) % As (V) % As (III) % = (x h As (III) / 0 h As (III)) * 100 As (V) % = (x h As (V) / 0 h As (III)) * 100 As (III) + As (V) = As (III) % + As (V) % As (III) + As (V) 0 100.0 0.0 100.0 1 95.7 0.0 95.7 3 92.8 0.2 93.0 6 91.3 1.6 92.9 12 78.5 12.0 90.6 24 61.9 21.4 83.4 10 ppb As (III) Spiked Urine Samples Time (h) As (III) % As (V) % As (III) + As (V) 0 100.0 0.0 100.0 1 102.8 0.0 102.8 3 105.2 0.0 105.2 6 106.4 0.0 106.4 12 95.1 0.5 95.6 24 102.9 1.3 104.2 Table 5. Limits of detection (LOD) and limits of quantification (LOQ) determined over 9 different analytical runs for each arsenic species. Stats are based on LODs and LOQs from 9 different studies during the method development/evaluation. AsB DMA As (III) MMA As (V) LOD (ppt) 1.7 0.7 0.6 1.0 0.3 LOQ (ppt) 5.8 2.5 1.9 3.2 1.0 LOD = (3 x σblank)/m; LOQ = (10 x σblank)/m The % recovery decreased for As (III) during the manual dilution experiments. This was expected since there was a noticeable species interconversion from As (III) to As (V). The interconversion was noticeable in these urine samples after 6 h from sample preparation. Assuming that the measured sample at t = 0 h is 100%, Table 4 shows the % conversion of As (III) to As (V) over the 24 h study as compared to t = 0 h. After 24 h, there was 21.4 % As (V) converted from As (III) in the manual dilution mode as compared to 1.3 % converted in the inline autodilution method. Species Interconversion Figure 7. Effect of time on manual sample preparation vs inline autodilution. The limits of detection (LOD) and limits of quantification (LOQ) were determined for 9 different analytical runs. The values listed in Table 5 are on par or better than what is found in the published literature. 3,4 The 9 analytical runs were collected by different users over a 1 month time period on a NexION 350d ICPMS (in KED mode) that is shared for multiple users within the facility for multiple types of applications. The aforementioned spiked samples used in the species interconversion study represented good accuracy/recovery, but were prepared in house. To test the method accuracy for samples not prepared in house, NIST SRM 2669 arsenic species in frozen urine was used. Table 6 displays the accuracy results for NIST SRM 2669. The measured values are compared to the NIST certified values and the minimum and maximum reported values from the SRM certification process. 5 6

Table 6. Accuracy to NIST SRM 2669 (arsenic species in frozen urine) levels 1 and 2. Measured values from 3 separate days (n = 3), AC = Arsenocholine, TMAO = trimethylarsine oxide. ^ AsB + TMAO = 3.37 ppb (2.44 ppb (min) 4.86 ppb Conclusions All Methods Combined (ppb) Measured Values (ppb) NIST Level 1 Certified Value (ppb) Min Max Avg ± 1 SD AsB 12.4 ± 1.9 10.48 14.68 12.3 ± 1.0 DMA 3.47 ± 0.41 2.98 4.09 3.39 ± 0.35 As (III) 1.47 ± 0.10 1.14 1.86 1.34 ± 0.37 MMA 1.87 ± 0.39 1.5 2.55 1.54 ± 0.17 As (V) 2.41 ± 0.30 1.9 2.88 2.57 ± 0.53 Total 22.2 ± 4.8 18 26.06 20.85 All Methods Combined (ppb) Measured Values (ppb) NIST Level 2 Certified Value (ppb) Min Max Avg ± 1 SD AsB 1.43 ± 0.08 1.26 1.84 2.45 ± 0.54^ DMA 25.3 ± 0.7 24.1 27.1 26.0 ± 3.6 As (III) 5.03 ± 0.31 4 6.2 5.58 ± 0.24 MMA 7.18 ± 0.56 6.51 8.24 6.96 ± 0.79 As (V) 6.16 ± 0.95 4.99 7.48 6.68 ± 0.85 AC 3.74 ± 0.35 3.14 4.27 n/a TMAO 1.94 ± 0.27 1.18 3.02 n/a Total 50.7 ± 6.3 45.18 58.15 47.62 The data reported here suggest that a syringe driven chromatography system is fully capable of reliable and reproducible results. There were no matrix effects seen for urine that was diluted from 30-100X, giving the user a range of sample dilutions from which to choose. The manual dilutions revealed that the longer a sample is diluted and sits out before analysis, the more likely it is to undergo species interconversion. This study resulted in up to 21% conversion of As (III) to As (V). The inline autodilution option allows for longer time to analysis for samples to sit on the deck of an autosampler with little to no species interconversion. References 1. Wang, T., Liquid chromatography-inductively coupled plasma-mass spectrometry (LC-ICP-MS), J. Liq. Chrom. & Related Tech., 2007, 30, 807-831. 2. Montes-Bayon, M., DeNicola, K., Caruso, J.A., Liquid chromatography-inductively coupled plasma-mass spectrometry, J. Chrom. A, 2003, 1000, 457-476. 3. Carioni, V.M.O., McElroy, J.A., Guthrie, J.M., Ngwenyama, R.A., Brockman, J.D., Fast and reliable method for As speciation in urine samples containing low levels of As by LC-ICP-MS: Focus on epidemiological studies, Talanta, 2017, 165, 76-83. 4. Xie, R., Johnson, W., Spayd, S., Hall, G.S., Buckley, B., Arsenic speciation analysis of human urine using ion exchange chromatography coupled to inductively coupled plasma mass spectrometry, Anal. Chim. Acta, 2006, 578, 186-194. 5. Yu, L.L., Verdon, C.P., Davis, W.C., Turk, G.C., Caldwell, K.L., Jones, R.L., Buckley, B., Xie, R., A human urine standard reference material for accurate assessment of arsenic exposure, Anal. Methods, 2011, 3, 1107-1115. 7

Elemental Scientific 7277 World Communications Drive Omaha, NE 68122 Tel: 1-402-991-7800 sales@icpms.com www.icpms.com 8 17202