Multi-Channel SFC System for Fast Chiral Method Development and Optimization

Similar documents
Expediting Achiral SFC Method Development Using a Multi-Channel SFC System with MS Detection

Basic Principles for Purification Using Supercritical Fluid Chromatography

Enantiomeric and Diastereomeric Separations of Pyrethroids Using UPC 2

UPC 2 Strategy for Scaling from Analytical to Preparative SFC Separations

Peptide Isolation Using the Prep 150 LC System

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Effects of Aqueous Sample Content and Aqueous Co-Solvent Composition on UPC 2 Separation Performance

Routine MS Detection for USP Chromatographic Methods

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Analysis and Purification of Diastereomeric Steroid Saponins from the Trigonella foenum-graecum Seed Extract Using SFC/ELSD

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

Application Note. Authors. Abstract. Pharmaceuticals

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

ChromegaChiral TM CSP Media and Columns

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

Separation of Chiral Pesticides by SFC and HPLC

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

ChromegaChiral TM CSP Media and Columns

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

[application note] ACQUITY UPLC/SQD ANALYSIS OF POLYMER ADDITIVES. Peter J. Lee, and Alice J. Di Gioia, Waters Corporation, Milford, MA, U.S.A.

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Chiral Separation Using SFC and HPLC

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Method Development for Chiral LC/MS/MS Analysis of Acidic Stereoisomeric Pharmaceutical Compounds with Polysaccharide-based Stationary Phases

viridis Columns Bringing a New Dimension to SFC Combining state-of-the-art media manufacturing with industry leading column technology, Viridis

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

Varian Galaxie Chromatography Data System for Preparative HPLC

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Implementation of Methods Translation between Liquid Chromatography Instrumentation

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Preparative Chromatography with Improved Loadability

ImprovIng ADmE ScrEEnIng productivity In Drug DIScovEry

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

A Protocol for High-Throughput Drug Mixture Quantitation: Fast LC MS or Flow Injection Analysis MS?

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Nexera UC Unified Chromatography

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Analysis of Organic Light Emitting Diode Materials by UltraPerformance Convergence Chromatography Coupled with Mass Spectrometry (UPC 2 /MS)

PosterREPRINT AN INTERACTIVE PHYSICOCHEMICAL PROPERTY PROFILING SOFTWARE FOR EARLY CANDIDATE ANALYSIS IN DRUG DISCOVERY INTRODUCTION

Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate Yu 3

Chiral Method Development in LC/MS Using Macrocyclic Glycopeptide CSPs

An Investigation of the Impact of Common Experimental Parameters on Signal Intensity in SFC ESI MS

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Determination of Enantiomeric Excess of Metolachlor from Chiral Synthesis using the Agilent 1260 Infinity Analytical SFC System

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

A novel approach to the analysis of multivitamin by online supercritical fluid extraction/supercritical fluid chromatography

LC Technical Information

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition

Chiral separations efficient, fast and productive

HPLC Praktikum Skript

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

APPLICATIONS TN HPLC Conditions. Lux Cellulose-1 Cellulose tris(3,5-dimethylphenylcarbamate)

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Automatization for development of HPLC methods

Fast methods for the determination of ibuprofen in drug products

ANALYSIS OF BASIC COMPOUNDS

Quality control analytical methods- Switch from HPLC to UPLC

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

Solvent Flexibility for Size-Based Polymer Analysis Using the Advanced Polymer Chromatography (APC) System

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Supercritical Fluid Chromatography

ChromegaChiral TM Columns. The right choice for chiral purification

High Pressure/Performance Liquid Chromatography (HPLC)

Optimized Flash Chromatography Purification: From TLC to large scale in three steps

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

[ a ppl ic at ion no t e ]

UPC 2 Method Development for Achiral Impurity Analysis

Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software. Application

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

APPLICATIONS TN Peptide Purification Method Development: Application for the Purification of Bivalirudin. Introduction

Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

SCPC Systems for Liquid-Liquid Extraction

Stephen McDonald and Mark D. Wrona Waters Corporation, Milford, MA, USA INT RO DU C T ION. Batch-from-Structure Analysis WAT E R S SO LU T IONS

Evaluation of Ketone Containing Terpene Compounds using Evaporative Light Scattering Detection

Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC

Sample Preparation. Approaches to Automation for SPE

The Nanostream (Pasadena, CA) Veloce system,

MODERN HPLC FOR PRACTICING SCIENTISTS

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Use of the SFC-MS Splitter Kit G

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

Quantitative Analysis of Caffeine in Energy Drinks by High Performance Liquid Chromatography

Transcription:

Multi-Channel SFC System for Fast Chiral Method Development and ptimization Lakshmi Subbarao, Ziqiang Wang, Ph.D., and Rui Chen, Ph.D. Waters Corporation, Milford, USA APPLICATIN BENEFITS The Method Station X5 SFC System offers parallel screening for method development and further optimization in single channel mode, on one platform, for chiral separations by SFC. The proprietary plumbing enables software controlled switching between parallel and single channel mode without the need for any hardware change. WATERS SLUTINS Method Station X5 SFC System 2998 Photodiode Array (PDA) Detector 2489 UV/Visible Detector SuperChrom Software KEYWRDS SFC Chiral chromatography Chiral stationary phase Method development INTRDUCTIN According to the FDA policy statement, developing stereoisomeric drugs, each enantiomer should be evaluated. 1 As a result, the pharmaceutical industry has escalated its emphasis on the generation of enantiomerically pure compounds before undertaking pharmacokinetic, metabolic, physiological, and toxicological evaluations. 2 Chiral chromatography, especially SFC, is the most widely used technique for obtaining milligrams to multi-grams of pure enantiomers in drug discovery. In SFC, supercritical C 2 combined with polar organic solvent(s), most commonly methanol, are used as the mobile phase. Due to the higher diffusivity and lower viscosity of supercritical fluid, SFC often provides a 3- to 8-fold faster separation than normal-phase HPLC. For chiral purification, SFC also offers significant cost savings by reducing organic solvent usage and removal as well as the time and energy required post-purification. As SFC instrumentation continues to improve, it is gradually overtaking HPLC as the first choice for chiral separations and purification. At the drug discovery stage, it is important to minimize the time required to obtain pure enantiomers, from milligrams to multi-grams. With stacked injections, such quantities can be obtained using SFC in a matter of hours. However, stacked injections can only be performed under isocratic conditions. Consequently, users often start method development with a screening of multiple chiral stationary phases (CSPs) using a generic gradient and finish the optimization with an isocratic method for scale-up. Despite the speed advantage SFC can offer, method development still remains one of the major bottlenecks for chiral separations by SFC. Due to the lack of a universal CSP, trial-and-error screening of a set of CSPs has been the predominant approach to chiral method development. Currently, most commercial instruments employ automated column and solvent switching to facilitate this screening process. However, it is time-consuming because each column has to be tested individually in a temporal manner. To circumvent this problem, multi-column, parallel screening approaches in both LC 3 and SFC 4 have been proposed. Typically, an injected sample was carried by the mobile phase and divided into multiple columns with varying detection schemes. The extent of improvement in throughput Parallel screening 1

depends upon the number of channels the main flow splits. These proposed approaches, however, suffer from the lack of independent control of multi-channel and single-channel required by method development and optimization. After the parallel screening, users would have to either re-plumb the system to the single channel configuration or resort to the use of another dedicated single channel system to perform isocratic optimization. In this application note, we introduce the first commercially available parallel SFC system, Method Station X5 SFC System, from its design objectives to operation using 4-benzoyloxy-2-azetidinone as a model compound. EXPERIMENTAL The Method Station X5 SFC System consists of a Fluid Delivery Module (FDM), Alias Autosampler, Analytical-2-Prep Column ven, Automated Back Pressure Regulator (ABPR), 2998 PDA Detector, and four 2489 UV/Visible detectors. SuperChrom Software was used for data acquisition and analysis. In the current study, each injection is split into five individual channels. All columns were purchased from commercial resources. Figure 1 shows 4-benzoyloxy-2-azetidinone, obtained from Sigma Aldrich (St. Louis, M, USA). A stock solution of 2.5 mg/ml was prepared by dissolving the compound in methanol. Key experimental parameters are listed as follows: Column Parallel D-H, AD-H, J-H, AS-H, and CC4, all 4.6 x 150 mm Single-Channel D-H, 4.6 x 150 mm Flow rate 12.5 ml/min 2.5 ml/min Modifier Methanol Methanol Injection volume 10 µl 2 µl Temperature 40 C 40 C System pressure 120 bar 120 bar Gradient 5% for 1 min, 5-40% in 5 min, 40% for 1 min, 40-5% in 1 min, and 5% for 1 min UV wavelength 254 nm 254 nm Isocratic at 5%, 6%, and 7% N H Figure 1. The chemical structure of 4-benzoyloxy-2-azetidinone. 2 Multi-Channel SFC System for Fast Chiral Method Development and ptimization

RESULTS AND DISCUSSIN Despite the lack of one universal column for chiral separations by SFC, it is the experience of many users that more than 75% of the pharmaceutical relevant chiral molecules can be separated by one or multiple of the following CSPs: D-H, AD-H, J-H, and AS-H. 5 The advancements in CSPs, especially the newly commercialized IC (Chiral Technologies Inc., West Chester, PA) and Cellulose-LUX-2 (Phenomenex, Torrance, CA), have further improved the success rate of chiral separations by SFC. The Method Station X5 SFC System is designed to primarily deal with those chiral molecules that can be separated using the aforementioned four to six CSPs, to quickly select the appropriate CSP in a parallel format, and to optimize the isocratic condition in a single channel mode for ensuing purification. The proprietary plumbing of the Method Station X5 SFC System enables a software controlled switching between parallel and single channel mode without any hardware change. Figure 2 shows the SFC chromatograms of 4-benzoyloxy-2-azetidinone obtained in parallel screening mode. Both AD-H and AS-H columns yielded baseline resolution of the enantiomers. D-H generated partial separation. J-H and CC4 offered no resolution. Based on the screening results, the system should then be switched to the single channel mode with either AD-H or AS-H for an optimal isocratic method for scale-up. For illustration purposes, we selected the D-H column instead, for further method optimization to mimic a common scenario where only partial separation is achieved with a gradient in the parallel mode. Based on the chromatograms shown in Figure 2, we ran a sequence consisting of three methods with 7%, 6%, and 5% modifier respectively. AD-H D-H - - 0.0 0.6 J-H AS-H 0.4 - - - CC4 - Figure 2. SFC chromatograms of 4-benzoyloxy-2-azetidinone obtained in parallel mode. Multi-Channel SFC System for Fast Chiral Method Development and ptimization 3

The results are shown in Figure 3. In this particular experiment, the 5% isocratic method appears to yield the best resolution. It is noted that the Method Station X5 SFC System is applicable for achiral method development and optimization as well, as long as a limited number of columns can be selected to encompass a sufficiently wide selectivity range. 7% Methanol - - 6% Methanol - - 5% Methanol - - Figure 3. SFC chromatograms of 4-benzoyloxy-2-azetidinone obtained under isocratic conditions using various volumes of methanol. 4 Multi-Channel SFC System for Fast Chiral Method Development and ptimization

CNCLUSINS The Method Station X5 SFC System offers a streamlined process for SFC method development and optimization. The proprietary plumbing enables a software controlled switching between parallel and single channel mode. The system will expedite the method development by five-fold; hence, shortening the turnaround time for those molecules that can be separated by the commonly used CSPs. References 1. FDA s Policy Statement for the Development of New Stereoisomeric Drugs May 1, 1992; Corrections made on January 3, 1997. http://www.fda.gov/drugs/ GuidanceComplianceRegulatoryInformation/Guidances/ ucm122883.htm 2. Zhang Y, Wu D, Wang-Iverson DB, and Tymiak A. A. Drug Discov. Today, 2004; 10 (8): 571-577. 3. 4. Zhang Y, Watts W, Nogle L, and McDonnell. J. Chromatogr. A., 2004; 1049: 75-84. Zeng L, Xu R, Laskar DB, and Kassel DB. J. Chromatogr. A., 2007; 1169: 193-204. 5. Francotte ER. J. Chromatogr. A., 2001; 906: 379-397. Waters is a registered trademark of Waters Corporation. SuperChrom, Analytical-2-Prep, and The Science of What s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 2010 Waters Corporation. Printed in the U.S.A. ctober 2010 720003540EN LL-LCGC Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com