Development and Validation of Rp-hplc Method for Analysis of Raloxifene in Pharmaceutical Dosage Forms

Similar documents
International Journal of Pharmacy and Pharmaceutical Sciences Vol 2, Issue 1, 2010

RP-HPLC Method Development and Validation of Dapagliflozin in Bulk and Tablet formulation

Development of Validated Analytical Method of Mefenamic Acid in an Emulgel (Topical Formulation)

Development and validation a RP-HPLC method: Application for the quantitative determination of quetiapine fumarate from marketed bulk tablets

Impact factor: 3.958/ICV: 4.10 ISSN:

CHAPTER INTRODUCTION OF DOSAGE FORM AND LITERATURE REVIEW

Int. J. Pharm. Sci. Rev. Res., 30(2), January February 2015; Article No. 09, Pages: 63-68

Development And Validation Of Rp-Hplc Method For Determination Of Velpatasvir In Bulk

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

Validated RP-HPLC Method for Estimation of Cefprozil in Tablet Dosage Form

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD TO DETERMINE CINITAPRIDE HYDROGEN TARTARATE IN BULK AND PHARMACEUTICAL FORMULATION

International Journal of Research in Pharmaceutical and Nano Sciences Journal homepage:

Department of Quality Assurance, Luqman College of Pharmacy, GULBARGA (K.S.) INDIA ABSTRACT

Research Article METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION OF ELBASVIR AND GRAZOPREVIR BY RP-HPLC

Journal of Pharmaceutical and Biomedical Analysis Letters. Analysis Letters

STABILITY INDICATING RP HPLC METHOD FOR ANALYSIS OF DORZOLAMIDE HCl IN THE BULK DRUG AND IT S PHARMACEUTICAL DOSAGE FORM

RP-HPLC Estimation of Trospium Chloride in Tablet Dosage Forms

A Simple, Novel Validated Stability Indicating RP-HPLC method for estimation of Duloxetine HCl in Capsule Pharmaceutical Formulation

Development and Validation of Stability Indicating RP-HPLC Method for the Determination of Anagrelide HCl in Pharmaceutical Formulation

DEVELOPMENT AND VALIDATION OF RP-HPLC METHOD FOR QUANTITATIVE ANALYSIS OF GABAPENTIN IN PURE AND PHARMACEUTICAL FORMULATIONS

Received: ; Accepted:

Dissolution study and method validation of alprazolam by high performance liquid chromatography method in pharmaceutical dosage form

INTERNATIONAL JOURNAL OF RESEARCH IN PHARMACY AND LIFE SCIENCES

Research Article. Figure 1. Chemical structure of doxofylline. Indonesian J. Pharm. Vol. 24 No. 1 : ISSN-p :

Novus International Journal of Analytical Innovations 2012, Vol. 1, No. 3

Validated First Order Derivative Spectroscopic Method for the determination of Stavudine in Bulk and Pharmaceutical Dosage Forms

STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS

Stability indicating RP-HPLC method for determination of azilsartan medoxomil in bulk and its dosage form

Chapter-4 EXPERIMENTAL WORK BY RP-HPLC

SIMULTANEOUS ESTIMATION OF CILOSTAZOL AND ASPIRIN IN SYNTHETIC MIXTURE USING HPTLC METHOD

International Journal of Pharma and Bio Sciences V1(1)2010. HPLC method for analysis of Lercanidipine Hydrochloride in Tablets

Development and Validation of Stability Indicating Assay Method of Etodolac by using UV-Visible Spectrophotometer

Research Article Available online at

Isocraticc Reverse Phase HPLC Method-Determination and Validation of Cilostazol

Pelagia Research Library

Simultaneous estimation and validation of bromhexine and cephalexin in bulk and pharmaceutical dosage form by RP-HPLC method

Validation of Stability-Indicating RP-HPLC Method for the Assay of Ibrutinib in Pharmaceutical Dosage form

International Research Journal of Pharmaceutical and Applied Sciences Available online at Int. Res J Pharm. App Sci.

Journal of Advanced Scientific Research DEVELOPMENT AND VALIDATION OF STABILITY-INDICATING RP-HPLC METHOD FOR ESTIMATION OF DABIGATRAN ETEXILATE

A RP-HPLC METHOD DEVELOPMENT AND VALIDATION OF PARA- PHENYLENEDIAMINE IN PURE FORM AND IN MARKETED PRODUCTS

Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article. Estimation of zaleplon by a new RP-HPLC method

International Journal of Current Trends in Pharmaceutical Research. International Journal of Current Trends in Pharmaceutical Research

Method Development and Validation for the Estimation of Darunavir in Rat Plasma by RP-HPLC

Development And Validation Of Stability Indicating RP-HPLC Method For Estimation Of Ledipasvir And Sofosbuvir

A Reverse Phase Liquid Chromatography Analysis of Citicoline Sodium in Pharmaceutical Dosage Form using Internal Standard Method

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Method Development and Validation Of Prasugrel Tablets By RP- HPLC

Journal of Drug Delivery and Therapeutics

QUANTITATIVE DETERMINATION OF TARTARIC ACID IN TOLTERODINE TARTRATE BY ION CHROMATOGRAPHY USING CONDUCTIVITY DETECTION

Volume 6, Issue 2, January February 2011; Article-015

SIMPLE AND SENSITIVE VALIDATED REVERSE PHASE HPLC-UV METHOD FOR THE DETERMINATION OF LYMECYCLINE IN PHARMACEUTICAL DOSAGE FORMS

Analytical method development and validation of carvedilol in bulk and tablet dosage form by using uv spectroscopic method as per ich guidelines

Pelagia Research Library

Analytical Method Development and Validation of Lafutidine in Tablet dosage form by RP-HPLC

ISSN: ; CODEN ECJHAO E-Journal of Chemistry , 9(1), 35-42

Asian Journal of Pharmaceutical Analysis and Medicinal Chemistry Journal home page:

Zero And First Order Derivative Spectrophotometric Methods For Determination Of Dronedarone In Pharmaceutical Formulation

Development and Validation of Stability-Indicating RP-HPLC Method for Estimation of Atovaquone

Pharmacophore 2014, Vol. 5 (2), USA CODEN: PHARM7 ISSN Pharmacophore. (An International Research Journal)

HPLC Method Development and Validation for the estimation of Esomeprazole in Bulk and Pharmaceutical Dosage Form

RP-HPLC METHOD FOR THE ESTIMATION OF ZIPRASIDONE

Stability Indicating RP-HPLC Method for Determination of Valsartan in Pure and Pharmaceutical Formulation

DEVELOPMENT AND VALIDATION OF NEW RP-HPLC METHOD FOR THE DETERMINATION OF AFLOQUALONE IN HUMAN PLASMA AND FORMULATION

Stability-indicating HPLC determination of tolterodine tartrate in pharmaceutical dosage form

Reverse Phase High Performance Liquid Chromatography method for determination of Lercanidipine hydrochloride in bulk and tablet dosage form

Development and Validation of UV Spectrophotometric Estimation of Diclofenac Sodium Bulk and Tablet Dosage form using Area under Curve Method

Journal of Chemical and Pharmaceutical Research

Available online Research Article

Chapter 4: Verification of compendial methods

Research Article. *Corresponding author S.Ushasri

Development and validation of UV- spectrophotometric method for the estimation of dabigatran etexilate mesylate (dem)

J Pharm Sci Bioscientific Res (4): ISSN NO

Development and Validation of RP-HPLC Method for the Estimation of Repaglinide in Bulk Drug and Pharmaceutical Formulation

Development and validation of RP- HPLC method for determination of ticagrelor in pharmaceutical dosage formulation

Development and Validation of a Stability-Indicating RP-HPLC Method for Analysis of Racecadotril in Pharmaceutical Dosage Forms

Simultaneous estimation of amitryptyline and chlordiazepoxide by RP-HPLC method

International Journal of Pharma and Bio Sciences

Appendix II- Bioanalytical Method Development and Validation

Development and Validation of a HPLC Method for Determination of Anastrozole in Tablet Dosage Form

Available online at Scholars Research Library

Available Online through

Validated spectrophotometric determination of Fenofibrate in formulation

International Journal of Pharmaceutical Research & Analysis

SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS.

2.1 2,3 Dichloro Benzoyl Cyanide (2,3 DCBC) and survey of. manufactured commonly for the bulk drug industry, few references were

Saudi Journal of Medical and Pharmaceutical Sciences

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Pharmacophore 2011, Vol. 2 (4), ISSN Pharmacophore. (An International Research Journal)

Development and validation of RP-HPLC method for determination of marker in polyherbal marketed Kankasava formulations

New Simple UV Spectrophotometric Method for Determination of Mirtazapine in Bulk and pharmaceutical dosage forms

Department of Chemistry, JNTUACE, Kalikiri

B. Anupama et al. IRJP 2 (1)

Research Article. Identification and characterization of unknown impurity in zolmitriptan tablets by a sensitive HPLC method

DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR IN-VIVO STUDY OF DICLOFENAC POTASSIUM

International Journal of PharmTech Research CODEN (USA): IJPRIF, ISSN: , ISSN(Online): Vol.9, No.7, pp , 2016

KEYWORDS: Acetaminophen, Doxylamine succinate, Dextromethorphan hydrobromide.

Analytical method development and validation of gabapentin in bulk and tablet dosage form by using UV spectroscopic method

Liquid Chromatography and Spectroscopic Method for Estimation of Sulphadoxine in Marketed Formulation

Research Article. UV Spectrophotometric Estimation of Alprazolam by second and third order derivative Methods in Bulk and Pharmaceutical Dosage Form

Development and Validation of Sensitive RP-HPLC Method for the Estimation of Glibenclamide in Pure and Tablet Dosage Forms

Transcription:

Research Article JPRR (2018) 2:7 Journal of Pharmaceutical Research and Reviews (ISSN:2576-8417) Development and Validation of Rp-hplc Method for Analysis of Raloxifene in Pharmaceutical Dosage Forms Venu Manukonda 1, Shaheen Sultana 2*, Amit Singh 3, Maria Khan 2, Mohammad Yusuf 2, Ibrahim Abdullah Maghrabi 2 1 RV Northland Institute; 2 Taif University, Saudi Arabia; 3 RV Northland Institute ABSTRACT The aim of present work was to develop and validate a simple, fast and reliable isocratic HPLC method for determination of raloxifene hydrochloride in pharmaceutical dosage form. Chromatographic Separation was achieved by using Agilent technologies model G4288A. Gradient HPLC-1120, Compact low pressure binary Gradient Pump Consisting Of vacuum in built degasser unit non-pda UV-vis detector equipped with a manual injection system with 50 ul loop and the eluted analytes for drug was traced by UV detection at 287nm. The method was found to be linear over a wide concentration range (0.1-20 µg/ml) with a correlation coefficient (r) of 0.9995. The LOD and LOQ were 21 and 63 ng/ml for RLX. These results indicate that method is sensitive enough for therapeutic assay. The % RSD values for three selected concentrations were found to be 0.14, 0.51 and 0.75% revealed the high system accuracy. The relative standard deviations (RSD) obtained at three concentrations (n = 6) in the range of 8-10 µg/ml was 0.016-0.018 % for the intraday assay and 0.016-0.024% for interday assay revealed good precision. The specificity was demonstrated that drugs were free of interference from potential impurities. % RSD, theoretical plates and Rmin were not significantly affected when the experimental parameters (such as flow rate, gradient program) were altered and thus concluded that the method was robust. Keywords: Raloxifene; HPLC, linearity, robustness *Correspondence to Author: Dr. Shaheen Sultana Taif University, Saudi Arabia Phone no: +966544122485 How to cite this article: Venu Manukonda, Shaheen Sultana, Amit Singh, Maria Khan, Mohammad Yusuf, Ibrahim Abdullah Maghrabi. Development and Validation of Rp-hplc Method for Analysis of Raloxifene in Pharmaceutical Dosage Forms. Journal of Pharmaceutical Research and Reviews, 2018; 2:7. escipub LLC, Houston, TX USA. Website: http://escipub.com/ JPRR: http://escipub.com/journal-of-pharmaceutical-research-and-reviews/ 0001

1. INTRODUCTION Raloxifene Hydrochloride belongs to the Benzo Thiophene class Of compound used for the treatment and prevention of postmenopausal disorder inwomen (1). Raloxifene hydrochloride is Chemically Termed as Methanone [6 Hydroxy -2 (4 hydroxyl Phenyl0benzo(B) Thien 3 yl)] [4 [ 2 1 Piperidinyl ) ethoxy] Phenyl ]Hydrochloride Figure 1). Raloxifene Hydrochloride was approved by Food and Drug Administration (FDA) In 1997 (2). A High-Performance liquid chromatographic (HPLC) is a Recent Technique in the liqid Chromatography which enables significant reduction in separation time and solvent consumption. Analytical methods described in literature for determination of raloxifene hydrochloride in biological and other matrices mostly involve sepectrophotometery analysis (3,4). UV spectrophotometric methods are less sensitive, accurate and time consuming. Some Sensitive and accurate methods have already have been reported for estimation of rlx such as LC MS MS (5,6) and Capillary electrophoresis (7) but they are expensive and time consuming as lot of sample preparation and processing steps. HPLC methods also reported for the determination of RLX either for pharmaceutical dosage forms or biological fluids. All these HPLC methods require use of buffer solutions as mobile phase in isocractic or gradient way (8). The aim of present work was to develop and validate a simple, fast and reliable isocratic HPLC method for determination of raloxifene hydrochloride in pharmaceutical dosage form. 2. Chemicals and Reagents Pharmaceutical Grade Raloxifene Hydrochloride certified to be 99.8 % pure according to manufactured recommend spectrophotometry method was generous gifted from Ranbaxy research labs, gurgaon, India. HPLC Grade acetonitrile and Water (HPLC) were purchased from fisher scientific Ltd Mumbai and Rankem RFCL India respectably. The formic acid Ar Grade was provided by Ranbaxy research labs Gurgaon India. Di hydrogen sodium phosphate and sodium hydroxide was purchased from Merck, India. Nylon membrane filter (47 mm, 0.22um porosity) were purchased from Pali corporation, Mumbai, India. The other Chemicals used were of analytical Grade. 3. Instrumenation and analytical Conditions Chromatographic Separation was achieved by using a Agilent technologies model G4288A. Gradient HPLC-1120, Compact low pressure binary Gradient Pump Consisting Of vacuum in built degasser unit non-pda UV-vis detector equipped with a manual injection system with 50 ul loop and the eluted analytes for drug was traced by UV detection at 287nm. The mobile phase consisting of HPLC water (ph 3.2 Adjusted with Formic Acid AR Grade ) and Acetonitrile (HPLC) Grade Degassed for 5min in a bath sonicator. The mobile phase was pumped from the solvent reservoir in the ratio of (40 : 60) was pumped into column at a flow Rate 2.0 ml/min and the run time was 6 min. Prior to injection of the drug solution the column was equilibrated for at least 30 min with the mobile phase following through the system. 4. Preparation of working standard solution of Raloxifene a. Calibration curve and quality control standards A definite amount of RLX (10 mg) was transferred to a 10 ml volumetric flask. Small quantity of methanol was added and the volume was maintained with methanol to give a final concentration of 1 mg/ml (stock solution A). Different sets of working standards at various concentrations were prepared by appropriate dilution of the stock solution A (0.1 to 20 µg/ml) with mobile phase containing Acetonitrile and water at a ratio of 60:40.. The JPRR: http://escipub.com/journal-of-pharmaceutical-research-and-reviews/ 0002

Table 1 Solubility data of the drug in different solvents/ buffers Solvents Mean Concentration (mg/ml) ± SD at room temperature DMSO Water Methanol Acetone Ethanol Chloroform 3.2 ± 0.4 mg/ml 3.6 ± 0.5 mg/ml 3.1 ± 0.3 mg/ml 3.0 ± 0.3 mg/ml 3.3 ± 0.6 mg/ml 3.4 ± 0.5mg/mL Solubility Freely Soluble Insoluble soluble Insoluble Soluble Insoluble Table 2 Characterization of the linear regression analysis of standard RLX Parameters Results Linearity Regression equation Mean slope ± SD** Intercept ± SD Standard error of line Standard error of slope Standard error of estimate Correlation coefficient 0.1 20 µg/ml y= 44246 x + 915.5 44246 ± 123.3 915.5 ± 84.3 0.0064 1.0003 1949.9858 0.9990 * Average from three calibration plots (i.e. n=3) ** Standard deviation Table 3 Accuracy for determination of RLX Sample (µg/ml) % RSD (n=3) Results Mean recovery (%) 2 (80 %) 4 (100 %) 6 (120 %) 0.75 0.51 0.14 96.75 % 97.00 % 98.00 % *Average from three determinations (n=3) RSD, relative standard deviation Table 4 Summarizes within-day and between-day precision data. Sample Added Final Intra-day measured Inter-day measured conc. drug drug assay assay (µg/ml) conc. conc. (µg/ml) (µg/ml) 5 3 8 Mean Conc ± µg/ml 8.98 ± 0.04 % Conc. % RSD Mean conc ± µg/ml. 99.28 % 0.018 8.97 ± 0.05 % Conc. % RSD 99.00 % 0.024 5 4 9 9.98 ± 0.01 99.75 % 0.016 9.97 ± 0.01 99.62 % 0.016 5 5 10 9.99 ± 0.01 99.88 % 0.016 9.97 ± 0.01 99.66 % 0.018 *Average from three determinations (n=3) SD, Standard deviation; RSD, relative standard deviation JPRR: http://escipub.com/journal-of-pharmaceutical-research-and-reviews/ 0003

calibration curves were constructed by plotting peak area ratios of corresponding analytes to the internal standard against their theoretical concentrations which were fitted by a least squares linear regression to the equation: response ratio (y) = slope (m) concentration (x) + intercept (c). Unknown concentrations of analytes were determined with reference to the calibration equation (9). b. Method validation Method was validated according to ICH Harmonized Tripartite, Validation of Analytical procedures and various parameters such as precision, accuracy, specificity, limit of detection, limit of quantitation, linearity and range, suitability and robustness were calculated (10). Linearity Linearity was assessed by making different concentrations of sample solutions (0.1 to 20 µg/ml) and assayed. 50 µl of each solution was injected into the HPLC system and the peak area of the chromatogram obtained was noted. Before the injection, the column was equilibrating for at least 20 min with mobile phase. Each concentration was analyzed in triplicate and curves were constructed using peak area ratio of drug to the internal standard versus nominal concentrations of the analytes. Least square linear regression analysis of the data gave slope, intercept, and correlation coefficient or coefficient of determination. Sensitivity Sensitivity of the method has been tested by examining limit of detection (LOD), the lowest concentration of an analyte in a sample that can be detected but not necessarily quantified, and the limit of quantitation (LOQ), the lowest concentration of analyte in a sample that can be determined with acceptable precision and accuracy under the stated operational conditions of the method. The calculation method is based on the standard deviation of the response (Sb) and the slope of the calibration curve (a). The limit of detection was calculated from calibration graph by the formula: LOD = 3 Sxy/a, and the limit of quantification: LOQ = 10 Sb/a. Accuracy Aliquots of 10 µl of the selected assay solutions at three concentration levels (80, 100, 120) were injected into the HPLC system, and triplicate measurements were recorded for each concentration. The nominal contents of the drug in each solution were calculated from the linear regression equations. The results were expressed as percent recoveries of the particular components in the samples as [mean found concentration/theoretical concentration] 100. Precision The precision of the assay was determined by repeatability (intra-day) and intermediate precision (inter-day). The intra-day precision was calculated as the relative standard deviation (RSD) of results from ten standard samples, during the same day, and the interday precision was studied by comparing the assays on two different days. Six sample solutions of different concentration (8-10 µg/ml) were prepared and assayed, and the standard deviation (SD) and RSD were calculated. Specificity The specificity is the ability to measure accurately and specifically the analyte of interest in the presence of other components that may be expected to be present in the sample matrices. It was demonstrated that analytes were free of interference from degradation products or excipients in pharmaceutical formulations ensuring that peak response in the same retention times is due to examined components only. Robustness The Robustness of an analytical Procedure Refers to its ability to remain unaffected by small deliberated variations in method parameters and provides an indication of its reliability for routine Analysis.. It has been JPRR: http://escipub.com/journal-of-pharmaceutical-research-and-reviews/ 0004

Parameters Table 5 System suitability parameters Results System precision (% RSD) 0.54 Retention time (min.) 4.37±0.61 Peak width (min) 1.41±0.24 Peak width (USP) 0.22 ± 0.04 Asymmetry /Tailing factor 0.44 ± 0.02 Capacity factor 0.20 ± 0.002 Plates/meter (USP) 3503.96 ± 124.21 *Average from three determinations (n=3) SD, Standard deviation; RSD, relative standard deviation Table 6 Chromatographic conditions and range investigated during robustness testing Variables Range % RSD Theoretical Retention time investigated (n=3) plates ± SD (min ± SD) ( n=3) Flow Rate 2.0 Temperature 30 Condition I Wave length 343 0.34 2040±241 4.37± 0.07 Mobile phase ratio 75 : 25 ph 3.0 Flow Rate 1.8 Temperature 25 Condition II Wave length 341 1.83 3671±151 6.512±0.12 Mobile phase ratio 75.5 : 24.5 ph 2.8 Flow Rate 2.2 Temperature 35 Condition III Wave length 345 1.47 11927±471 7.138±0.23 Mobile phase ratio 74.5 : 25.5 ph 3.2 *Average from three determinations (n=3) SD, Standard deviation; RSD, relative standard deviation JPRR: http://escipub.com/journal-of-pharmaceutical-research-and-reviews/ 0005

evaluated by varying method parameters set for optimum conditions such as flow rate, gradient program, concentration of buffer, and determining the effect (if any) on the results of the method. Result and discussion Solubility studies Solubility studies were conducted in different solvents and the results are shown in Table 1. The drug is found to be freely soluble in DMSO and methanol which matches the existing reference. Linearity The analytical curve for RLX was constructed by plotting the area under the curve (AUC) of the main peak versus drug concentration. It was found to be linear over a wide concentration range (0.1-20 µg/ml) with a correlation coefficient (r) of 0.9995. The straight line equation obtained from the experimental results was found to be (Equation 1): y = 44246x + 915.5 (1) There was no significant difference between the slopes of calibration plots prepared on different days (ANOVA, P > 0.05). Sensitivity The LOD and LOQ were 21 and 63 ng/ml for RLX. These results indicate that method is sensitive enough for therapeutic assay. Accuracy Accuracy is one of the most important parameters of an analytical methodology and it can be expressed as the percent recovery of known amounts of drug added to a sample. The accuracy of the analytical method was evaluated in terms of repeatability. Repeatability (intra-day precision) was assessed by injecting the six replicate injections of 2-6 µg/ml drug concentration (Table 3). The % RSD values for three concentrations were found to be 0.14, 0.51 and 0.75%. The value of Relative standard deviation (% RSD) revealed the high system accuracy. The mean (n=3) percentage recovery of standard drug at 80%, 100%, and 120% of specific concentration were found to be satisfactorily high. Precision The precision refers to the variability of the results in repeated analyses of the sample under identical experimental conditions. The method was validated by evaluating the intraand inter-day precision. The precision was calculated from an average of ten determinations of a homogeneous sample. The relative standard deviations (RSD) obtained at three concentrations (n = 6) in the range of 8-10 µg/ml was 0.016-0.018 % for the intraday assay and 0.016-0.024% for interday assay. High recovery percentage (> 99 %) indicates that the method presents a good precision (Table 4). Specificity The specificity was demonstrated showing that drugs were free of interference from potential impurities and degradation products as absence of any peak in the same retention times (Figure 2). Results indicate the high specificity of the method and can be used in a stability assay and routine analysis of the both investigated drugs. The average retention time for RLX was found to be 4.37 min (Table 5). Robustness It was found that the % RSD, theoretical plates and Rmin were excellent undermost conditions, and remained unaffected by small deliberate changes of experimental parameters (Table 6). Variation in the experimental parameters (flow rate, gradient program) provided an indication of its reliability during normal use and concluded that the method was robust. The results obtained were within the acceptable limit and it was observed there were no marked changes in the areas (%RSD is 2), retention time, tailing factor, and theoretical plates of the JPRR: http://escipub.com/journal-of-pharmaceutical-research-and-reviews/ 0006

Figure 1: Structure of Raloxifene Hydrochloride (Molecular Formula C28H27NO4S.HCL) Figure 2. Chromatogram of raloxifene showing sharp peak at 4.353 min. Figure 3. Robustness chromatogram of standard RLX at different condition I, II, and III JPRR: http://escipub.com/journal-of-pharmaceutical-research-and-reviews/ 0007

chromatograms which demonstrated that the RP-HPLC method was robust (Figure 3). Conclusion The method described enables the simple and fast assessment of RLX. The validation data demonstrate good precision and accuracy, which prove the reliability of the proposed References method. Hence, this RP-HPLC method can be used routinely for estimation of RLX. This study demonstrates the applicability of the proposed method for the potency determination of RLX in bulk drug and pharmaceutical formulations. The method can be successfully used for routine quality control and stability assays and offers advantages in speed, simplicity, and reliability 1. Pavan kumar Bathini, Venkata Ramana Kandula, Prashanth Reddy Gaddameedhi (2009). An improved synthesis of raloxifene hydrochoride: a selective estrogen receptor modulator. Coden HLEEAI, 2014; 4(4): 515-518. 2. Enrico M Messalli, Cono Scaffa. Long-term safety and efficacy of raloxifene in the prevention and treatment of postmenopausal osteoporosis: an update. International Journal of Womens Health, 2009; 1: 11 20. 3. Buridi Kalyanaramu, K. Raghubabu. Determination of Raloxifene Hydrochloride By Oxidative Coupling Reaction In Pharmaceutical Formulations. International Journal of Applied Pharmaceutics, 2011; 3(2): 69. 4. Nagaraju PT, K Venu Gopal, Murali Krishna NV, Bhargavi V, Srinivasulu N. UV - Spectrophotometric Method Development and Validation for Determination of Raloxifene in Pharmaceutical Dosage Form. Research & Reviews: Journal of Pharmaceutical Analysis, 2014; 3(4): 27-31. 5. Jurji Trontelj, Marija Bogataj, Janja Marc, Ales Mrhr. Development and validation of a LC MS MS assay for determination of ralo xifene and its metabolites in human plasma. Journal of Chromatography. B, Analytical Technologies in the Biomedical and Life Sciences, 2007; 855(2): 220 227. 6. Basavaiah Kanakapura, Anil Kumar UR, Tharpa K, Vinay KB. Determination of raloxifene in human urine by LC MS MS. Chemical Industry & Chemical Engineering Quarterly, 2009; 15(3): 119 123. 7. Pérez-Ruiz T, Martínez-Lozano C, Sanz A, Bravo E. Development and validation of a quantitative assay for raloxifene by capillary electrophoresis. J Pharm Biomed Anal. 2004 Mar 10;34(5):891-7. 7. D Suneetha, A. Lakshmana Rao. A New Validated Rp- Hplc Method For The Estimation Of Raloxifene In Pure And Tablet Dosage Form. Rasayan Journal of chemistry, 2010; 3(1): 117-121. 8. Mohammad Nasir Uddin, Suman Das, Shakhawat Hasan Khan, Swapon Kumar Shill, Habibur Rahman Bhuiyan, Rezaul Karim. Simultaneous determination of amoxicillin and chloramphenicol and their drug interaction study by the validated UPLC method. Journal Taibah University of Science, 2016; 10: 755 765. 9. ICH Harmonized Tripartite, Validation of Analytical Procedures: Text and Methodology Q2 (R1), Current Step 4 Version, Parent Guidelines on Methodology Dated November 6, 1996, Incorporated in November 2005, in: International Conference on Harmonisation, Geneva, Switzerland, 1996. JPRR: http://escipub.com/journal-of-pharmaceutical-research-and-reviews/ 0008