Burning position correction for measurements with the turning luminaire LGS 1000

Similar documents
Impact of New SSL Standards on Goniospectroradiometric Measurements

Measurement method for the proficiency testing program

CIE The Measurement of Luminous Flux. XIN Hongzheng

Morris Products 53 Carey Road, Queensbury, NY 12804

AN AUTOMATED GONIOPHOTOMETER FOR LUMINAIRE CHARACTERIZATION

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD

NVLAP LAB CODE LM Test Report. For DONGGUAN THAILIGHT SEMICONDCTOR LIGHTING CO.,LTD

NVLAP LAB CODE LM Test Report. For. Shenzhen Guanke Technologies Co. Ltd. (Brand Name: )

Optical Measurement Guidelines for High-Power LEDs and Solid State Lighting Products

PRACTICAL UNCERTAINTY BUDGETS FOR SPECTRAL MEASUREMENTS OF LEDS

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK)

RoHS. Dimming: PWM/VR/0-10V QPL ID # PL21D5Q1B95E. Cat# W LED FlatPanel Slipfitter Mount. EPA Rating 2.

LM Test Report. For. Beyond LED Technology. (Brand Name: Beyond LED) 2725 Mountain Industrial Blvd. Suite A-1 Tucker, GA 30084

NVLAP LAB CODE LM Test Report. For REMPHOS TECHNOLOGIES LLC. (Brand Name: ) 90 Holten St, Danvers, MA 01923,USA.

NVLAP LAB CODE LM Test Report. For AOK LED LIGHT COMPANY LIMITED. (Brand Name: AOK)

LightLab I N T E R N A T I O N A L

NIST Role in Supporting Solid State Lighting Initiative

NVLAP LAB CODE LM Test Report. For LED PANEL LIGHTING CO.,LTD. (Brand Name: N/A)

LightLab I N T E R N A T I O N A L

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006

NVLAP LAB CODE LM Test Report. For MAXLITE INC. (Brand Name: MaxLed) 12 York Ave West Caldwell NJ 07006

NVLAP LAB CODE LM Test Report. For CE INNOVATIONS LTD. (Brand Name: N/A) 911 Denison St Markham, ON L3R 3K4 Canada

Cat# LED High Bay

Wet Locations Use 71400A 120 W PF V 2.9A 93% 144 pcs. Philips 130 LM/W 5000K Hrs. 5 Years. 24-1/4 Lbs 665*530*120 1 PCS

LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl350u1

NVLAP LAB CODE LM Test Report. For. LED Wav es,inc. (Brand Name: LED Wav es)

PRODUCT FAMILY DATASHEET LEDVANCE FLOODLIGHT LED

application note LED measurement radiant power measurement

NVLAP LAB CODE LM Test Report. For IGT LIGHTING INC. (Brand Name: IG T ) 3755 Lincoln St. Suite B, Riverside, CA 92503

MODEL NAME : LLDMWW0-15K*0*A

NVLAP LAB CODE LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl5s50u1

New Method for Realizing a Luminous Flux Scale using an Integrating Sphere with an External Source

NVLAP LAB CODE LM Test Report. For. Simkar Corporation. (Brand Name: ) LED Luminaire. Model name(s):salho2xl4s50u1

solid state LED LED LED LED LED LED Light Emitting Diode: LED LED module Self - ballasted LED module Integral LED module

LED Entryway Light RoHS

LightLab I N T E R N A T I O N A L

Improvement Uncertainty of Total luminous Flux Measurements by Determining Some Correction Factors

integrating-sphere-construction-kits

Report No.: HZ p. (Lumens) Stabilization Time CRI (K)

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING)

Page 2 of 14 Report No.: Test sample... : Type of test object... : LED Downlight LED Downlight

Solid State Lighting Annex: Interlaboratory Comparison Test Method

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

PrevaLED COIN 50 DC COB G1

TEST REPORT IES LM For Signlite Opto-electronics Co.,Ltd.

Photometric Report. 42 Partners Limited 21 Woden Road Wolverhampton WV10 0AU UK

Photometric Analysis for Triple Tennis Court

LightLab I N T E R N A T I O N A L

Project description: This is a sample showing an Oil Tank Storage (ATEX) application using Spartan Flood White-Light.

Flux-O-Meter. Making luminous flux measurements more accessible. Andrew Bierman, Martin Overington

LightLab I N T E R N A T I O N A L

NVLAP LAB CODE LM Test Report. For. Hocan Group Co.,Ltd (Brand Name: SEPICN LED LIGHTING)

LM Test Report. for. Forest Lighting Corporate. LED Tube. Model: FL-T8U BN

TeraLED. Thermal and Radiometric Characterization of LEDs. Technical Specification.

PrevaLED COIN 50 COB G1

LEDs in ERCO luminaires

LightLab I N T E R N A T I O N A L

LightLab I N T E R N A T I O N A L

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Best color perception in spotlighting and downlighting

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Cree Racine Engineering Services Testing Laboratory (RESTL) Photometric Testing and Evaluation Report

Test Report. Report No.: EED35I Page 1 of 12. Customer : Revolution Lighting Technologies : 4139 Guardian Street, Simi Valley, CA93063

TeraLED. Thermal and Radiometric Characterization of LEDs. Technical Specification.

A NEW DIGITAL ILLUMINANCE-METER WITH A SPECIAL LUMINANCE TUBEADAPTER FOR UNIVERSAL APPLICATIONS IN LIGHTING ENGINEERING

Aurora Ltd 6 Apex Park Welwyn Garden City AL7 4SW T: +44 (0)

Checked X.CAO

Checked X.CAO

NVLAP LAB CODE LM Test Report. For SIMKAR CORPORATION. (Brand Name: ) LED Lamp. Model name(s): SALXS350U1. Model differences: N/A

NVLAP LAB CODE LM Test Report. For SIMKAR CORPORATION. (Brand Name: ) LED Lamp. Model name(s): SALXS450U1. Model differences: N/A

Photometric Test Report

HONOURTEK Delux Array Series HRA0303 Preliminary

Calibration Due Date Chroma Programmable AC Source 61604

NEAR-FIELD GONIOPHOTOMETRY: A METROLOGICAL CHALLENGE

Uncertainty determination of correlated color temperature for high intensity discharge lamps.

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

UNIVERSITY OF HONG KONG Faculty of Engineering M.Sc.(Eng) in Building Services Engineering MEBS6002 LIGHTING ENGINEERING

LEDs in ERCO luminaires

SOCCER FIELDS LEVELS FOR UEFA COMPETITIONS ELITE LEVEL A LEVELS : A-B-C-D

Using a Quantitative Measure to Achieve a Qualitative Lighting Design Effect! Introduction What is Luminance?

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

Collective Road Lighting class ME 5

LED light engine / OLED LED compact. Uready2apply DLA G1 SNC China domestic

Book 14 Edition 1.1 March Socketable Flat LED Module and LED Light Engine

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT8N20-A

Number Eight Lighting Company MPT2-HI-R-6-BK/40/DIM1-M-1000

Number Eight Lighting Company MPT2-HI-R-6-BK/25/DIM1-M-1400

Uncertainty determination of correlated color temperature for high intensity discharge lamps

Tunable White LED compact

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT5Z02-6

LM Test Report. for MLS CO., LTD. Linear Ambient Luminaires Model: MT5Z02-9

Cat# LED Linear High Bay 100W

Dune MC DW. Date: Operator: A. Fantoni

Test Report. Report No.: EED35I R1. Address : 701 Millennium Boulevard Greenville, SC : P , P , P , P

Test Report. Report No.: EED35I Address : 701 Millennium Boulevard Greenville, SC : P , P , P , P

Transcription:

APPLICATION NOTE Burning position correction for measurements with the turning luminaire LGS 1000 Dr. Đenan Konjhodžić SUMMARY This application note describes the implementation of the auxiliary photometer method according to the new Standard CIE S025 on the turning luminaire LGS 1000 and validates the corrected data for various different types of light source. The results of the sample measured in its recommended burning position are used for comparison. On this basis, recommendations are made for the appropriate use of the burning position correction to perform measurements of various light sources in conformance with the Standard. 1. Introduction Goniophotometers are normally used for determining luminous intensity distribution and measuring the total luminous flux by suitable numerical integration of these distributions. A turning luminaire, such as the LGS 1000 from Instrument Systems, is the most practical, space-saving and inexpensive version of a far-field goniophotometer. However, as the name implies, during measurement the luminaires are rotated around axes C and γ, thus changing their burning position. This could be a problem in the measurement of position sensitive light sources. Figure 1: Turning luminaire LGS 1000 (left) and stray light tube with the detector (photometer head or EOP) on a stand (right). 1

As semiconductor light sources, the LEDs used in solid state lighting (SSL) sources are in fact insensitive to positional changes. However, the passive heat sinks of SSL sources are usually optimized for the operating position of the proposed use. The convective air flow through the cooling fins guarantees the required cooling capacity and thus the desired operating temperature of the LEDs. Any hindrance to air convection, e.g. by operating the light source in a burning position different from that prescribed by the manufacturer, can influence the optical characteristics of the luminaire. In this case a correction of the goniophotometric measurement is necessary. The new and first international Standard CIE S025 [1] or its European counterpart EN 13032-4:2015 [2] in principle allows goniophotometric measurements to be carried out in a burning position other than specified. In order to conform to the requirements set forth in the standard, a suitable burning position correction must be performed, e.g. by the auxiliary photometer method. Based on this method, Instrument Systems offers an overall solution for the LGS 1000, which is described in more detail in the following. 2. Performing a correction of the burning position The fundamental concept of the auxiliary photometer method is to monitor the relative photocurrent of the light source at a fixed point while the latter changes its orientation in space. The measured value in each position, weighted by the reference in the specified operating position, serves as a correction factor for the actual measurement with the turning luminaire. In practice, initially an additional sample plate is mounted on the LGS 1000. The sample is adjusted and burned in, and a normal sequence is carried out with the turning luminaire. The advantage of the additional sample plate is that following the sequence measurement a positioning correction can be carried out without the need for major reconfiguration and without repeat burn-in of the sample. Only the photometer arm is mounted and connected together with the auxiliary photometer. A software wizard guides the user through the entire process to create a positioning correction file. First of all, the fixture is brought into the home position for a sequence measurement. In the first correction step, the same sequence is performed as for the previous measurement (Figure 2). Even if the previous measurement was performed with a spectroradiometer, the correction file is continuously recorded with the auxiliary photometer. It is only important that the same angle positions are logged. The additional measurement time required for positioning correction is thus minimized. Sample C Sample plate Swivel arm LGS1000 Auxiliary photometer Photometer arm γ Figure 2: Mounting an auxiliary photometer for performing the first correction step. 2

In the second correction step the sample is placed in the burning position prescribed by the manufacturer, in order to acquire the reference value with the auxiliary photometer (Figure 3). For this purpose a cantilever arm is mounted on the sample plate. The sample can slide along this arm in the switched-on state. The carriage is clamped at an adequate distance and the cantilever arm rotated by 90 to the C-axis and secured. Now the sample can be turned into the prescribed burning position with the aid of the motorized C axis. The user can specify the desired burning position by entering an angle. After a suitable stabilization time, the reference value can be recorded. This step completes the acquisition of the required correction factors and the positioning correction file thus created can be saved. The positioning correction file can then be used on the previously recorded sequence measurement with the turning luminaire, by which the values for the photometric integral (or luminous intensity) are multiplied by the respective correction factor. The luminous flux also changes accordingly as an integral sum. The positioning correction file can also be applied on other samples of the same type, saving valuable measuring time. 1. 2. 3. Figure 3: The sample is placed in the prescribed burning position with the aid of a cantilever arm and the reference value taken for the positioning correction. 3

3. Measurement comparison To determine when a burning position correction is expedient, luminous flux measurements of various light sources using the so-called luminous flux integrator, by which the source did not change its burning position, were compared to the readings for the turning luminaire, by which the source changed position (Table 1). The readings of the luminous flux integrator with the photometer as a detector serve as reference values for the calculation of percentage deviations with the turning luminaire. Since the measurement deviations for all tested SSL downlights and an LED module with inbuilt cooler with 1-2% are very small and clearly lie within the range of typical measurement uncertainties for the respective systems, a dependence on burning position cannot be concluded. Most users, however, have no opportunity to perform such comprehensive comparative measurements. Fast measurement of a positioning correction file as proof of no or negligible position dependence is thus a convenient solution. With this proof and accounting for the positional influence in the measurement uncertainty budget a standard CIE S025 compliant measurement can also be carried out with a turning luminaire. Application of the burning position correction file is however worthwhile for the two other light sources 30W LED floodlight and 300W halogen sun lamp, in order to reduce measurement uncertainties due to position dependence. Table 1: Comparison of luminous fluxes measured with the luminous flux integrator and the turning luminaire for various light sources [3]. Light source Emission angle Ф v [lm] Luminous flux integrator Ф v [lm] Turning luminaire SSL downlight 1 97 798.0 806.9 1.1% SSL downlight 2 80 665.3 671.9 1.0% SSL downlight 3 52 1834 1878 2.4% SSL downlight 4 29 403,0 409.4 1.6% LED module with cooler 104 1167 1178 0.9% LED floodlight 103 1697 1743 2.7% Sun lamp (halogen) 32 4150 4021-3.1% 4. Use of burning position correction A burning position correction file was recorded for all named light sources, as described in Chapter 2. Figure 4 shows an example. Following the software routine, first of all the individual C-planes are measured with exactly the same settings as for measurement with the turning luminaire. Subsequently, the luminaire is brought into the operating position and the reference value measured after the stabilization phase. 4

5,7 Kartesisch 5,68 Photometer-Wert / klx 5,66 5,64 5,62 5,6 5,58-80 -60-40 -20 0 20 40 60 80 Gamma / C=0 C=10 C=20 C=30 C=40 C=50 C=60 C=70 C=80 C=90 C=100 C=110 C=120 C=130 C=140 C=150 C=160 C=170 Figure 4: Example of a burning position correction file shown in Cartesian coordinates. During one scan (C-plane) of the positioning correction with the auxiliary photometer practically negligible oscillations are measured. The individual planes among one another, however, demonstrated deviations of up to 2% for the LED floodlight and 3% for the sun lamp, caused by the lamp (Figure 5). As expected, the SSL downlights show a negligible deviation of less than 1%. It is not surprising that the rotation of the downlights around the C axis during measurement with the turning luminaire causes no change in the luminous flux, as the fins on the heat sink are symmetrically arranged around the luminaire. The LED module with cooler is likewise not positiondependent, due to active cooling. For these sources one could dispense with burning position correction of the C-planes for measurement purposes. This is however required by Standard CIE S025, and the user must be able to correct all measurement steps. Table 2: Application of burning position correction to measurements with the turning luminaire. Ф v [lm] Turning luminaire Ф v [lm] Turning luminaire with position correction LED floodlight 1743 2.7% 1702 0.3% Sun lamp 4021-3.1% 4111-0.9% SSL downlight 1 at 222 V 765.9-4.0% 805.6 1.0% SSL downlight 1 at 215 V 721.8-9.5% 805.8 1.0% SSL downlight 1 at 200 V 650.9-18.4% 806.9 1.1% SSL downlight 1 at 185 V 588.7-26.2% 799.8 0.2% Upon application of the burning position correction on the measurements of the 30W LED floodlight and 300W halogen lamp with a gas mixture to simulate the sun, the corrected luminous fluxes showed an improvement compared to the luminous flux integrator and amount to only 0.3% deviation for the LED floodlight and -0.9% for the sun lamp (see Table 2). One reason for the 5

position dependence of LED-floodlight could be the arrangement of the cooling fins on the heat sink, which run parallel in one direction, and upon rotation on the turning luminaire can result in slightly differing cooling conditions. In the case of the sun lamp, rotation can lead to movement of the gas mixture within the lamp and consequently a measuring error. 100,5% 100,0% 99,5% 99,0% 98,5% 98,0% 97,5% 97,0% LMR4 Downlight 3000K1 LR6 Downlight neutral 2 Philips Downlight smart3 Ph Downlight 4 Module mit with Kühler cooler LED Floodlight floodlight Heimsonne Sun lamp 300W 96,5% 0 30 60 90 120 150 180 C [ ] Figure 5: Relative deviation of individual C-planes over angle C when γ=0. 5. Induced position dependence The observed effects are in part still within the typical measurement uncertainties, and the position dependence of the sample should be higher in order to show that the positioning correction can also be successfully applied in this case. Because an SSL downlight with a correspondingly high position dependence was not available, this was created artificially. SSL downlight 1 was measured at lower operating voltages (222, 215, 200 and 185 V) with the turning luminaire, which of course results in lower luminous fluxes (Table 2). For the subsequent burning position correction the reference value in the standard burning position at a nominal 230 V was recorded after a sufficient burn-in time. After each use of the position correction the high deviations of between -4% and -26% to the measurement with the luminous flux integrator were corrected to about 1%. It is thus also possible to measure light sources which may demonstrate a high position dependence with the turning luminaire and subsequently perform a position correction according to CIE S025. 6. Conclusion The correction of the burning position by the auxiliary photometer method permits goniophotometric measurements conforming to CIE S025 also for light sources with a high position dependence. Should a burning position correction not appear expedient, a simple proof of position independence can be provided with the measurement and measurement uncertainties in the budget thus reduced. References [1] CIE S 025/E:2015, Test Method for LED Lamps, LED Luminaires, and LED Modules, CIE Vienna 2015. [2] EN 13032-4:2015, Light and Lighting. Measurement and presentation of photometric data of lamps and luminaires. LED lamps, modules and luminaires. [3] Konjhodzic D., Goniophotometry and spectral radiometry for SSL light sources, LICHT 03/2015, 73 77. 6