Research on Air density Measurement for measuring weights

Similar documents
Traceable Mass Determination and Uncertainty Calculation

MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT

Estimation of Uncertainty of Air Buoyancy Correction for Establishment of Primary Mass Standards


Problem 1 in the section "self-thinking" Team «MG 12»

Chapter 1 Solutions Engineering and Chemical Thermodynamics 2e Wyatt Tenhaeff Milo Koretsky

Introduction to Thermodynamic Cycles Part 1 1 st Law of Thermodynamics and Gas Power Cycles

Operating Instructions MagCal - Susceptometer Magnets Calibration System

Modeling Hydraulic Accumulators for use in Wind Turbines

Calculation methods for the physical properties of air used in the calibration of microphones

Lect-19. In this lecture...

PC TOOL FOR DATA ANALYSIS IN CALIBRATION OF SPECIAL WEIGHTS

SELECTION, SIZING, AND OPERATION OF CONTROL VALVES FOR GASES AND LIQUIDS Class # 6110

Outline. Example. Solution. Property evaluation examples Specific heat Internal energy, enthalpy, and specific heats of solids and liquids Examples

Experiment: measure of the elastic deformation in three point bending

Purpose of the experiment

Kinetic Theory. Reading: Chapter 19. Ideal Gases. Ideal gas law:

Water Vapour Effects in Mass Measurement

Design of Intelligent Meteorological System Based on MEMS Wang Hao 1,a, Nie Meng 2,b, Huang Qing an 3,c

EVALUATION OF ADSORPTION EFFECT BY USING SILICON SPHERE AND SILICON SURFACE ARTEFACTS

Mass determination of 1 kg silicon spheres for Avogadro project

Web Resource: Ideal Gas Simulation. Kinetic Theory of Gases. Ideal Gas. Ideal Gas Assumptions

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure

OIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air

Pure Component Equations

IPC-TM-650 TEST METHODS MANUAL

Performance Measurements for Mechanical Air Handling Systems

Assembly and Maintenance manual according to regulation DIN 82079

Technical Note Development of a New, High Sensitivity 2000 kg Mechanical Balance

NUMERICAL EVALUATION OF THE THERMAL PERFORMANCES OF ROOF-MOUNTED RADIANT BARRIERS

Getting started with BatchReactor Example : Simulation of the Chlorotoluene chlorination

Air Density Determination Using 1 kg Buoyancy Mass Comparison(III)

O P E R A T I N G M A N U A L

Exact expression for the lifting condensation level. David M. Romps

1D-HAM. Coupled Heat, Air and Moisture Transport in Multi-layered Wall Structures. Manual with brief theory and an example. Version 2.

ISOCON-6. 24V AC or DC POWERED ISOLATING SIGNAL CONVERTER

Chapter 7: The Second Law of Thermodynamics

Heat stress index. Dewpoint. pressure. altitude

Total, active and effective alkali

Development of standard calibration equipment for the rain gauges

Greenhouse Steady State Energy Balance Model

Fu Yuhua 1. Beijing, China

6XMag-1 Magnifier User manual

Supplementary comparison SIM.M.FF-S12. Final Report for Volume of Liquids at 20 L

RheolabQC. Rotational Rheometer for Quality Control. ::: Intelligence in Rheometry

PRODUCT SPECIFICATIONS

SOLUTION MANUAL ENGLISH UNIT PROBLEMS CHAPTER 3 SONNTAG BORGNAKKE VAN WYLEN. FUNDAMENTALS of. Thermodynamics. Sixth Edition

The International Association for the Properties of Water and Steam

New Functionalities of Thermo-Calc and DICTRA

DESIGN, DEVELOPMENT, AND COMMISSIONING OF A 120 kn DEADWEIGHT FORCE STANDARD MACHINE

The uncertainty evaluation for the hydrometer calibration at NMISA. Presenter: Bongani Ndlovu Date: 09/10/2018

R I T. Title: GCA Stepper Operations. Semiconductor & Microsystems Fabrication Laboratory Revision: F Rev Date: 08/09/ SCOPE

LabVIEW and PCI DAQ Card Based HTS Test and Control Platforms

ln P s T and P s T where R 22,105, D A 27, E B 97.

Design of Integrated Error Compensating System for the Portable Flexible CMMs

ISOCON-3 MAINS POWERED ISOLATING SIGNAL CONVERTER

IPC-TM-650 TEST METHODS MANUAL

Calibration Routine. Store in HDD. Switch "Program Control" Ref 1/ Ref 2 Manual Automatic

473-SHX Dew Point Mirror

d dt T R m n p 1. (A) 4. (4) Carnot engine T Refrigerating effect W COPref. = 1 4 kw 5. (A)

ON SITE SYSTEMS Chemical Safety Assistant

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

Edgetech Instruments Inc.

Minimum (-31) every 10 s -35 (-31)

Section 6: PRISMATIC BEAMS. Beam Theory

SOFTWARE USER MANUAL. Weather Capture Advance WS1640 WM9280

DUST SETTLING USING A NEGATIVE IONIZATION SYSTEM

Karl-Fischer Titration the method for determining water

CityGML XFM Application Template Documentation. Bentley Map V8i (SELECTseries 2)

ph electrode Instruction Manual For use with the RAH-210 & RPH-250

Air velocity & IAQ measuring instruments

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

MS BA Micro Altimeter Module

PiCUS Calliper Version 3

Research on the Laser Doppler Torque Sensor

OIML R 50-3 RECOMMENDATION. Edition 2014 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

Thermodynamics Introduction and Basic Concepts

The General Conference on Weights and Measures (CGPM), at its 24th meeting,

Lecture 3: DESIGN CONSIDERATION OF DRIERS

Innovative Sustainable Technology

Stress and Strain State in the Segmental Linings during the Excavation by Tunnel Boring Machines

Determining the Concentration of a Solution: Beer s Law

EZ PUSH-TO MANUAL EZ PUSH TO MANUAL. romer-optics.com.

Determination of the Equivalent Weight and the K a or K b for a Weak Acid or Base

EVEN THE FINEST CANNOT ESCAPE. Wöhler SM 500 Suspended Particulate Analyser

Exploitation of ground based GPS for Climate and Numerical Weather Prediction applications COST action 716

CALYS 80P. Field pressure calibrator with high accuracy 0.01%

Chapter 4: Properties of Pure Substances. Pure Substance. Phases of a Pure Substance. Phase-Change Processes of Pure Substances

AW/AHW Professional Weighing Scale. User Manual

The Ordering Code for various standard model Recorders with an AC supply and without any additional options are as follows:

Automated volume measurement for weihts using acoustic volumeter

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications

An ultra-sensitive, versatile, and affordable single tube luminometer for Life Science Research.

Solution to 1-D consolidation of non-homogeneous soft clay *

BERMAD Irrigation. BIC2000 Weather Station SYSTEM GUIDE VERSIONS : BIC ; PC 4.90

THE METEOROLOGICAL DATA QUALITY MANAGEMENT OF THE ROMANIAN NATIONAL SURFACE OBSERVATION NETWORK

The International Association for the Properties of Water and Steam

GW Professional Weighing Scale. User Manual SNOWREX INTERNATIONAL CO., LTD. SRGW

Titrator 3.0 Tutorial: Calcite precipitation

Transcription:

POSTER SESSIO Research on Air density Measurement for measuring weights WAG Xiao-lei, WAG Jian, ZHOG Rui-lin, ZHAG ue, CAI Chang-qing, AO Hong, DIG Jing-an (ational Institute of Metrology, Beijing 100013, China) Abstract: According to the CIPM-007 equation, an equipment of air density measurement that for high accuracy measurement of measuring weights is deeloped. Measurement of temperature, humidity, pressure and CO are integrated on it to measure air density automatically. The experimental results proe that additional measurement of CO can improe accuracy of air density measurement. Key words: Air density; Weight; Air buoyancy correction; CIPM-007 equation 1. Introduction Measurement of measuring weights will be influenced by air buoyancy when measuring in air. This kind of influence is ery important for high accuracy measurement of measuring weights and air buoyancy correction need be processed to reduce it. The major component of air buoyancy correction is the determination of the density of moist air. The air density is commonly determined from the recommended equation of state. CIPM-007 equation is the newest equation that was endorsed by the International Committee for Weights and Measures (CIPM). The method that most of measurement inspection institutions and laboratories used is base on approximation formula for air density presented in [1-].In this method, both measurements of temperature, relatie humidity, pressure and calculation of air density are manual, thus measurement work is complex and deiation of measurement data may be high. Further more, approximation formula for air density can t meet the need of high accuracy measurement of measuring weights because of its low precision. To change this situation, we deelop equipment that integrated measurement of temperature, humidity, pressure and CO to determine air density auto automatically according CIPM-007 equation. In this paper, the designing of equipment and support software is introduced.. Air density measurement principle According CIPM-007 equation, the density of moist air is ealuated using an equation of state a p M ZRT a [1 x M (1 M a )], (1) where the quantities and units are P /Pa: pressure, t/ : air temperature, T /K: thermodynamic temperature = 73.15 + t/, x : mole fraction of water apour, Ma/(g mol 1 ): molar mass of dry air, M /(g mol 1 ): molar mass of water, Z : compressibility factor, R /(J mol Page 77

TC3 : FORCE, MASS, TORQUE & DESIT 1 K 1 ): molar gas constant. If a measurement of CO is aailable, then () should be used to improe the estimate of the molar mass of dry air[3]: M a xco [ 8.96546 1.011 ( where x co is mole fraction of carbon dioxide. x is ealuated using (3): 0.0004)] 10 3 kg mol 1 () x ps( t) hf ( p, t) p (3) where h is the relatie humidity and has the following range: 0 h 1. For example, an instrumental relatie humidity reading of 53% is expressed as h = 0.53. And f is enhance factor, p s is apour pressure at saturation. (4) and (5) can be used to ealuate f and p s : p s exp( AT BT C D / T) (4) f p t (5) where A, B, C, D,,, are constant, t is air temperature, T is thermodynamic temperature. where The equation for Z is (6): p p Z 1 [ a0 a1t at ( b0 b1t ) x ( c0 c1t) x ] ( d ex ) (6) T T a 0, a1, a, b0, b1, c0, c1, d, e are constant. Through equations listed aboe, we can know that to improe the determination of the air density, four enironment parameters, temperature, humidity, pressure and CO, need be measured. Thus we design equipment of air density base on measurement of these four enironment parameters. 3. Designing of air density measurement equipment 3.1 Design Structure Figure 1 shows the design structure of the air density measurement equipment. Measurement of temperature, humidity, pressure and CO are assembled in equipment, all measurement data is sent to module of data processing, and then air density is calculated with these data according to CIPM-007 equation. Page 78

POSTER SESSIO Humidity measurement deice Temperature measurement deice Pressure measurement deice CO measurement deice humidity temperature pressure CO Air density Figure 1. Design structure of equipment 3. Hardware and Software designing temperature pressure humidity CO Communication module MCU Communication module computer Figure. Equipment module structure Figure shows the structure of the equipment hardware module. Four corresponding sensors take on measurement of enironment parameters (temperature, humidity, pressure and CO ). Communication module is used to control communication between sensors and MCU or MCU and computer. MCU handles collection of measurement signals and calculation of air density. All of these deices are installed in a box which has small size for carrying and using coneniently. More functions can be used through connect with computer. Software designing includes two parts: software in MCU and application software installed in Page 79

TC3 : FORCE, MASS, TORQUE & DESIT computer. Figure 3 shows the MCU program flow chart. Software codes design and optimize base on this flow chart. To make up for lack of SCM functionality, application software installed in computer is deeloped. Start Communication initialization Send measure command Measurement data receied Show error message Get Air density Measurement finished End Figure 3. MCU program flow chart start Measure data file export Import saed data file Function select Air density measurement Send measure command Measure data chart Import saed data file Error message Data correct Export file Error message Measurement data receied Get air density Sae measurement data Error messag e Data correct Show data chart Measurement finished Exit end Figure 4. Application software program flow chart Page 80

POSTER SESSIO Application software extends the capabilities of equipment. It proides higher accuracy of floating-point calculation and measurement data storage serice in long time. User can export data stored to XLS format for using easily. Data stored also can be loaded to create analysis chart to support user s study. Figure 5 shows the main interface of application software, eery function can be found on here. Interface of data analysis chart is shown in Figure 6, analysis chart can be Figure 5. Application software main interface Figure 6. Application software program flow chart Page 81

TC3 : FORCE, MASS, TORQUE & DESIT exported to JPEG format for printing. 4. Experimental Analysis An experiment using this equipment connected with computer has been made in an air-conditioned room at 010/3/4 1:00 to 14:00. The results of measurement of CO are shown in Figure 7. The content of CO generally remained at around 400ppm from 1:00 to 1:40. From 1:40 it increased rapidly and reached its peak of 937ppm at 13:3, then began to decline slowly. The reason of CO changed is that one worker went into the room at 1:40 and left at 13:3, the breath of worker increase the content of CO in the room. Figure 8 shows the change of air density in same period. Two cures represent the air density which calculated by CIPM-007 equation and approximation formula for air density. From charts we can know that, the influence of CO for air density is neglected by approximation formula. And 10 group measurement data and comparisons are gien in Table 1. These experiment results proe that we can obtain a more accurate air density by using CIPM-007 equation than approximation formula. Figure 7. Measurement data of CO Page 8

POSTER SESSIO Figure 8. Measurement data of air density Table 1. Data comparison t ( ) RH(%) Pa (hpa) CO Air density (kg/m 3 ) approximation CIPM-007 formula equation Relatie error (%) 1.3 4.48 1006.800 858.5 1.187084 1.18730 0.018 1.3 4.49 1006.796 860.6 1.187078 1.18797 0.018 1.3 4.46 1006.84 865.0 1.187115 1.187336 0.019 1.5 4.43 1006.801 870. 1.187004 1.1878 0.019 1.6 4.39 1006.801 877.9 1.186966 1.187193 0.019 1.3 4.37 1006.80 886.8 1.187099 1.187331 0.00 1.5 4.38 1006.797 893.7 1.187005 1.18740 0.00 1.5 4.38 1006.797 901.0 1.187005 1.18744 0.00 1.1 4.37 1006.794 906.8 1.187176 1.187418 0.00 1.5 4.45 1006.713 937.4 1.186897 1.187154 0.0 Page 83

TC3 : FORCE, MASS, TORQUE & DESIT 5. Conclusions The equipment of air density measurement introduced in this paper has good features: small size, portable and high accuracy. Furthermore, it can support user s research work by using application software. For high accuracy measurement of measuring weights, the method that using CIPM-007 equation to determining air density is a better choice. References [1] aohong et. Weights. JJG 99-006. China Metrology Publishing House, 006. [] Weights of classes E 1, E, F 1, F, M 1, M 1, M, M 3, and M 3 Metrological and technical requirements (OIML R 111-1:004) [3] A Picard et. Reised formula for the density of moist air (CIPM-007).al 008 Metrologia 45 149-155 Page 84