Basic Principles for Purification Using Supercritical Fluid Chromatography

Similar documents
UPC 2 Strategy for Scaling from Analytical to Preparative SFC Separations

viridis Columns Bringing a New Dimension to SFC Combining state-of-the-art media manufacturing with industry leading column technology, Viridis

Peptide Isolation Using the Prep 150 LC System

Multi-Channel SFC System for Fast Chiral Method Development and Optimization

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Dilution(*) Chromatography

Expediting Achiral SFC Method Development Using a Multi-Channel SFC System with MS Detection

Implementation of Methods Translation between Liquid Chromatography Instrumentation

Effects of Aqueous Sample Content and Aqueous Co-Solvent Composition on UPC 2 Separation Performance

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

LC Technical Information

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

Routine MS Detection for USP Chromatographic Methods

[ a ppl ic at ion no t e ]

ANALYSIS OF BASIC COMPOUNDS

Enantiomeric and Diastereomeric Separations of Pyrethroids Using UPC 2

Solvent Flexibility for Size-Based Polymer Analysis Using the Advanced Polymer Chromatography (APC) System

KromasilSFC. Designed for green technology. Introducing Kromasil SFC XT

Analysis of Organic Light Emitting Diode Materials by UltraPerformance Convergence Chromatography Coupled with Mass Spectrometry (UPC 2 /MS)

UPLC Method Development and Validation

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

Minimizing Solvent Impact on Purification of Nitrogencontaining

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

Supercritical Fluid Chromatography

Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

UPC 2 Method Development for Achiral Impurity Analysis

What is Chromatography?

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Application Note. Authors. Abstract. Pharmaceuticals

ChromegaChiral TM CSP Media and Columns

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

Optical Isomer Separation Columns and Packing Materials

Analytical Chemistry

Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate Yu 3

Chiral separations efficient, fast and productive

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Nexera UC Unified Chromatography

C HA R AC T E RIZ AT IO N O F INK J E T P RINT E R C A RT RIDG E INK S USING A CHEMOMETRIC APPROACH

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Biochemistry. Biochemical Techniques HPLC

Preparative Chromatography with Improved Loadability

[application note] ACQUITY UPLC/SQD ANALYSIS OF POLYMER ADDITIVES. Peter J. Lee, and Alice J. Di Gioia, Waters Corporation, Milford, MA, U.S.A.

Automating Gradient Method Development in Flash Chromatography For Greater Productivity and Minimizing Solvent Use

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

3µm Fortis C18. Fortis 2.1µm. Increased Efficiency. Scalability and Sensitivity

ChromegaChiral TM CSP Media and Columns

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

penta-hilic UHPLC COLUMNS

Pure Chromatography Consumables Pure flexibility. Pure specialization. Pure convenience.

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

OBSERVATIONS ON THE WETTING OF REVERSED-PHASE HPLC PACKINGS

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

ImprovIng ADmE ScrEEnIng productivity In Drug DIScovEry

CHEM 429 / 529 Chemical Separation Techniques

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

Application. Introduction. Authors. Abstract. Hydrocarbon Processing

Definitive EtG/EtS LC-MS/MS Analysis:

Remember - Ions are more soluble in water than in organic solvents. - Neutrals are more soluble in organic solvents than in water.

Strategies for the Flash Purification of Highly Polar Compounds

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

SEPARATION OF FLURBIPROFEN AND IBUPROFEN ENANTIOMERS ON A CHIRAL STATIONARY PHASE USING SUPERCRITICAL FLUIDS

New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

Fast methods for the determination of ibuprofen in drug products

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Successfully Scaling and Transferring HPLC and UPLC Methods

1.7 m Fortis UHPLC Columns

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

Introduction to Chromatographic Separations

IN QUALITATIVE ANALYSIS,

SUPERCRITICAL FLUID CHROMATOGRAPHY AS SUCCESSFUL SEPARATION TOOL IN CHEMICAL AND PHARMACEUTICAL INDUSTRY

Liquid Chromatography

NEVIRAPINE ORAL SUSPENSION Final text for addition to The International Pharmacopoeia (February 2009)

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Chromatographic Separation

The Effect of Contact Angle and Wetting on Performance in Solid Phase Extraction

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns

TCI Chiral HPLC Column

Principles of Gas- Chromatography (GC)

LECTURE 2. Advanced Separation Science Techniques Present and Future Separation Tools

A novel approach to the analysis of multivitamin by online supercritical fluid extraction/supercritical fluid chromatography

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Use of High Speed/High Resolution Size-Based Chromatographic Separation of Polymeric Mixtures with Offline Infrared Detection

Quality control analytical methods- Switch from HPLC to UPLC

Transcription:

Basic Principles for Purification Using Supercritical Fluid Chromatography Jo-Ann M. Jablonski, Christopher J. Hudalla, Kenneth J. Fountain, Steven M. Collier, and Damian Morrison Waters Corporation, Milford, USA APPLICATION BENEFITS Basic principles for the separation of compounds from mixtures using supercritical fluid chromatography (SFC) are similar to the fundamentals of preparative liquid chromatography. Fast, large-scale separations increase throughput DMSO can be used for samples with limited solubility INTRODUCTION Although supercritical fluid chromatography (SFC) is not a new technique, preparative SFC is becoming increasingly more popular with advances in instrumentation, software and chemistry. Basic principles for isolating compounds with SFC are similar to the fundamental rules for large-scale preparative liquid chromatography. In this study, we illustrate the effect of flow rate, sample dissolution solvent, and column selectivity on the isolation of compounds from a mixture. Columns packed with Optimum Bed Density (OBD) Technology and with matching chemistry ensure easy and predictable scalability from the small scale to larger preparative separations. Different column selectivities improve compound isolation Column scalability ensures reproducible chromatograms at the analytical and preparative scales RESULTS AND DISCUSSION Flow Rate The flow rate used in preparative LC separations is limited by a number of factors including the length and diameter of the column, the particle size, and the system backpressure. Low viscosity carbon dioxide, the main component of the mobile phase in SFC, enables the use of higher flow rates than those generally used in liquid chromatography. The backpressure associated with column length is also lower, enabling usage of longer columns to increase resolution. Higher flow rates Waters solutions Waters Prep 00 SFC UV Directed System Waters Method Station SFC System Viridis SFC Columns Optimum Bed Density (OBD ) Technology Flow rate: 80 g/min Gradient time:.0 min 0.5.0.5.0 min Flow rate: 90 g/min Gradient time:.6 min Viridis SFC -Ethylpyridine OBD 9 x 00 mm, 5 µm Part number: 86004944 Mobile phase: 5-5% B Methanol with no additive Gradient slope: %/column volume Temperature: 40 C Wavelength: 0 nm System backpressure: 0 bar. Coumarin. Ibuprofen. Carbamazepine key words 0.5.0.5.0 min Supercritical fluid chromatography Viridis SFC columns Flow rate: 00 g/min Gradient time:.4 min Preparative SFC OBD technology 0.5.0.5.0 min Figure. Compound mixture run with constant gradient slope at 80, 90, and 00 g/min. The gradient separation time decreases with increased flow rate while maintaining resolution.

EXPERIMENTAL Sample preparation Compound mixture : Compound mixture : Compound mixture : Compound mixture 4: Coumarin, ibuprofen, and carbamazepine; 0 mg/ml each in methanol Ibuprofen and carbamazepine; 0 mg/ml each in methanol Ibuprofen and carbamazepine; 0 mg/ml each in DMSO (dimethyl sulfoxide) 0 mg/ml Fenoprofen, 5 mg/ml Ketoprofen, 0 mg/ml Flavone, 0 mg/ml Ibuprofen, and 0 mg/ml Coumarin in methanol Experimental conditions Prep SFC system: Waters Prep 00 SFC UV Detection: Columns: Column temperature: 40 C Injection volume: Flow rate: Co-solvent: Gradient: Waters 998 Photodiode Array Viridis SFC -Ethylpyridine, 4.6 x 50 mm, 5 µm, Part Number 8600497 Viridis SFC -Ethylpyridine OBD, 9 x 50 mm, 5 µm, Part Number 86004945 Viridis SFC Silica OBD, 9 x 50 mm, 5 µm, Part Number 8600498 Viridis SFC -Ethylpyridine OBD, 9 x 00 mm, 5 µm, Part Number 86004944 Methanol Analytical SFC system: Waters Method Station SFC System increase throughput by reducing the amount of time to complete a separation. As shown in Figure, the chromatographic profile is the same at different flow rates provided that the slope of the gradient expressed in percent change per column volume remains constant. This is accomplished by decreasing the run time in the same proportion as the flow rate is increased. Sample Dissolution and Loading Solvent Isolating compounds from mixtures requires that the sample be completely soluble before injection. Methanol, the most common co-solvent in SFC, does not solubilize all samples at the high concentrations usually required for typical preparative separations. Large volumes of solvents like methanol can limit the mass capacity of the column due to strong solvent effects that can distort the chromatography. Modifier stream injection, the standard patented configuration of all Waters Prep SFC systems, was designed to reduce this solvent effect. Dimethyl sulfoxide (DMSO) is used regularly in purification labs to solubilize many different types of.0 0.4 0.8..6.0.4.8..6 4.0 min.0 Viridis SFC -Ethylpyridine OBD, 9 x 00 mm, 5 µm Part number: 86004944 Mobile phase: 5-5% B Methanol in min with no additive Flow rate: 00 g/min @ 40 C Wavelength: 0 nm System backpressure: 0 bar Injection volume: 400 µl in methanol. DMSO. Ibuprofen. Carbamazepine Injection volume: 400 µl in DMSO 0.4 0.8..6.0.4.8..6 4.0 min Figure. The resolution between the compounds dissolved in DMSO is greater than for the same size injection with the compounds dissolved in methanol..0 0.4 0.8..6.0.4.8..6 4.0 min.0 Viridis SFC -Ethylpyridine OBD, 9 x 00 mm, 5 µm Part number: 86004944 Mobile phase: 5-5% B Methanol in min with no additive Flow rate: 00 g/min @ 40 C Wavelength: 0 nm System backpressure: 0 bar Injection volume: ml in methanol. DMSO. Ibuprofen. Carbamazepine Injection volume: ml in DMSO 0.4 0.8..6.0.4.8..6 4.0 min Figure. The difference in resolution between the sample mixture dissolved in methanol and in DMSO is more pronounced at a ml injection volume. Better resolution improves the mass loading leading to higher purity and increased purification process efficiency. Basic Principles for Purification Using Supercritical Fluid Chromatography

compounds. Figures and illustrate how DMSO injections in SFC can result in higher loading and better resolution for achiral applications as compared to samples injected in methanol. Note: For chiral applications, caution should be considered as DMSO can severely damage a traditional coated chiral stationary phase. If DMSO should be required for sample solubility, an immobilized chiral column should be selected. Selectivity In SFC, multiple columns are routinely tested to determine which provides the best resolution and peak shape for components in a mixture. Good resolution between compounds leads to isolations of target compounds with high purity. In the example below, the -ethylpyridine column shows excellent separation of all five compounds in the sample mixture. The same sample mixture analyzed on a silica column shows only three component peaks because it has a different selectivity. Coumarin and ibuprofen coelute at about 0.8 minutes and flavone and fenoprofen coelute slightly later. Flavone and ibuprofen also change elution order on the silica column. Ketoprofen is well-separated from all of the other peaks on both columns. 8.0e- 6.0e- 4.0e-.0e- Viridis SFC -Ethylpyridine OBD Column 9 x 50 mm, 5 µm Part Number: 86004945 4 5 0.4 0.8..6.0.4.8..6 4.0 min,,4 Mobile phase: 5-0% B Methanol in min with no additive Flow rate: 00 g/min @ 40 C Injection volume: 00 µl Wavelength: 0 nm System backpressure: 0 bar Concentration: 85 mg/ml 8.0e- 6.0e- 4.0e-.0e- 5 Viridis SFC Silica OBD Column 9 x 50 mm, 5 µm Part Number: 8600498. Coumarin. Flavone. Ibuprofen 4. Fenoprofen 5. Ketoprofen 0.4 0.8..6.0.4.8..6 4.0 min Figure 4. The selectivity difference between the -ethylpyridine and silica columns illustrates the effect that column chemistry can have on the separation. Basic Principles for Purification Using Supercritical Fluid Chromatography

Scaling Crude sample mixtures are usually analyzed with a screening gradient that starts from a low percentage of organic solvent and ends at a higher percentage of organic solvent in a relatively short time. If the compounds of interest are well resolved, the gradient may be scaled directly for purification. When all compounds elute before the completion of the gradient and are fully resolved, a simple reduction in the length of the gradient is acceptable. The fast flow rates used in SFC combined with modified gradients considerably shorten the total time required for purification of target compounds. Scaling separations requires matching column chemistry and particle size as well as appropriately scaled injection volumes and gradients. Waters preparative SFC columns are packed with OBD Technology, ensuring excellent bed stability and comparable performance to Waters analytical SFC columns. OBD preparative columns are packed to bed densities which closely match the equivalent analytical column. This innovative procedure produces preparative columns with excellent stability, reproducibility, and efficiency. A loading study is performed at the small scale to determine how much sample mass can be loaded on the column. The goal is to maximize the load without compromising the resolution. Improved peak resolution leads to higher loads and better purity for isolated compounds. Higher column loading reduces the number of runs required for obtaining enough material for subsequent experiments. The chromatography from the loading study is depicted in Figure 5. The conservative 5 µl injection volume shows good resolution between all of the compounds in the mixture and is subsequently scaled for the preparative run on the large column. A total volume of 600 µl was injected on the 9 x 50 mm preparative SFC column. Figure 6 compares the chromatography using the small scale modified gradient at maximum load with the large-scale chromatography used for the isolation. The selectivity is identical, a critical factor when using analytical chromatography as a scouting tool for isolation and purification. The differences in retention time and resolution between the analytical and preparative chromatograms can be attributed to factors including system volume and injection mode. In particular, the patented modifier stream injection mode* is employed for all Waters preparative SFC systems, minimizing the solvent effect and therefore improving throughput and efficiency. * US Patent # 6,576,5 Viridis -Ethylpyridine, 4.6 x 50 mm Mobile Phase: 5-0% MeOH in min with no additive Flow Rate: 5.9 ml/min at 40 C System Backpressure: 50 bar Injection Volume: 5, 0, 5, 40, and 50 µl Analyte mg/ml. Coumarin 0. Flavone 0. Ibuprofen 0 4. Fenoprofen 0 5. Ketoprofen 5 Viridis -Ethylpyridine Mobile Phase: 5-0% MeOH in min with no additive Injection Volume: 5, 5 and 600 µl Analyte mg/ml. Coumarin 0. Flavone 0. Ibuprofen 0 4. Fenoprofen 0 5. Ketoprofen 5 Total 86 mg/ml (in MeOH) Total 86 mg/ml (in MeOH) Inj. Vol. 4 5 50 µl 40 µl 5 µl 0 µl 5 µl Inj. Vol. 600 µl 5 µl 5 µl 4 5 Prep 9 x 50 mm OBD 00 g/min Analytical 4.6 x 50 mm 5.9 ml/min 0. 0.4 0.6 0.8..4.6.8.0..4.6.8.0 min 0. 0.4 0.6 0.8..4.6.8.0..4.6.8.0 min Figure 5. A loading study on the analytical column determines how much mass can ultimately be injected on the preparative column. Shown here, the increasing peak width is a result of the solvent effect associated with larger injection volumes. Figure 6. Preparative chromatography scaled from the analysis completed on the small scale column. Slight differences in retention time are related to the differences in the plumbing configuration and dwell volumes between the analytical and preparative systems. 4 Basic Principles for Purification Using Supercritical Fluid Chromatography

CONCLUSIONS The basic principles of liquid chromatography can be applied to supercritical fluid chromatography (SFC). Supercritical fluid chromatography can be run at increased flow rates making isolations faster and increasing throughput. References. Optimum Bed Density (OBD) Columns: Enabling Technology for Laboratory-Scale Isolation and Purification. Waters White Paper 700099EN 006 Nov. Samples with limited solubility can be dissolved in dimethyl sulfoxide and purified using SFC. Using DMSO for sample dissolution makes the purification process easier by reducing the uncertainty associated with sample solubility. Resolution and loading of samples dissolved in DMSO can be improved when compared to the same samples dissolved in methanol. Better sample resolution leads to products with higher purity. Higher loading improves the efficiency of the purification process. Column selectivity impacts the separation of mixture components. Screening different columns with fast gradients leads to a decision about which preparative column to use for isolating the compounds in the sample. The ability to scale between analytical and preparative OBD columns leads to identical separations at both the small and large scales, provided that the injection volume, slope, and gradient time are scaled correctly and the chemistry and particle size remain constant. Column scalability removes uncertainty in peak identification, thus facilitating higher purification process efficiency. Waters and ACQUITY UPLC are registered trademarks of Waters Corporation. OBD, Viridis, and The Science of What s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 00 Waters Corporation. Printed in the U.S.A. October 00 70007EN LB-LCGC Waters Corporation 4 Maple Street Milford, MA 0757 U.S.A. T: 508 478 000 F: 508 87 990 www.waters.com