Low level Os isotopic measurements using multiple ion counting

Similar documents
Trace Analyses in Metal Matrices Using the ELAN DRC II

Thermo Scientific Neptune XT MC-ICP-MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

of mass spectrometry

Rapid and precise calcium isotope ratio determinations using the Apex-ACM desolvating inlet system with sector-field ICP-MS in low resolution

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Fergus Keenan Shona McSheehy Julian Wills

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Improvement of bulk analysis of environmental samples by using a multiple collector ICP-MS

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Precise Pb isotope analysis of igneous rocks using fully-automated double spike thermal ionization mass spectrometry (FA -DS- TIMS)

Error Analysis. The big questions are: - What value best represents a set of measurements and how reliable is it?

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

ARGUS VI. Static Vacuum Mass Spectrometer. Static Vacuum ARGUS VI. Multicollection Low Volume Precision

SmartNotes. What is meant by the term interference?

Trace Elements - Definitions

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Automated Determination of PPQ Levels of Thorium in High Purity Copper

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Samples compliant with the WATER

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Finnigan TRITON Finnigan NEPTUNE. High Resolution Multicollector Mass Spectrometers

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

HELIX MC Plus. Static Vacuum. ARGUS VI Thermo Scientific HELIX MC Plus Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer

HELIX SFT. Static Vacuum ARGUS VI HELIX SFT. Static Vacuum Mass Spectrometer Static Vacuum Mass Spectrometer

Defining quality standards for the analysis of solid samples

Use of ICP-MS in analysing radioisotopes. Per Roos Risø National Laboratory for Sustainable Energy, Technicial University of Denmark

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Thermo Scientific TRITON Thermal Ionization MS. Thermo Scientific NEPTUNE Multicollector ICP-MS. High Resolution Multicollector Mass Spectrometers

Uncertainty in Measurement of Isotope Ratios by Multi-Collector Mass Spectrometry

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Sample Analysis Design PART II

Sample Analysis Design Internal Standard

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Speciation and δ 34 S analysis of volatile organic compounds in crude oil by GC-MC-ICP-MS using the Thermo Scientific GCI 300 Interface

The 30-Minute Guide to ICP-MS

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Nu Plasma II - Collector Configuration

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Determination of Ni isotopes in nickel material from the Rossi reactor

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Exploring new fields of Applications using Multi Collector Magnetic Sector MS

JRP SIB09 Elements GOOD PRACTICE GUIDE ON DEALING WITH INTERFERENCES IN ICP-MS FOR INVESTIGATING HIGH PURITY MATERIALS. Deliverable 1.2.

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

UNIVERSITY OF CALGARY. Measurement of the absolute isotopic composition and atomic weight

Interface (backside) & Extraction Lens

Overcoming Interferences with the Thermo Scientific icap 7000 Plus Series ICP-OES

Sources of Errors in Trace Element and Speciation Analysis

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Plasma Interface Element2

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Sample Analysis Design PART III

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Analysis of noble metals at low levels in geological reference materials and ores

2101 Atomic Spectroscopy

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Prima PRO Process Mass Spectrometer

Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products

Additional Analytical Methods. Detrital zircon samples were collected from nine fine-, medium-, and coarse-grained

The NKS-NORCMASS guide to beginners in ICP-MS

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

Atomic Spectroscopy AA/ICP/ICPMS:

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

GSA Data Repository Denyszyn, et al., 2018, A bigger tent for CAMP: Geology,

A new variable dispersion double-focusing plasma mass spectrometer with performance illustrated for Pb isotopes

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Partial Energy Level Diagrams

Automating Sample Purification MC-ICPMS. Elemental Scientific ICP ICPMS AA

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Determination of Trace Metals in Human Urine Using the NexION 300 ICP-MS

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

FRONTIER RESEARCH ON EARTH EVOLUTION, VOL. 2. Research Program for Data and Sample Analyses, Institute For Research on Earth Evolution (IFREE) 2

Autenticity control, provenance testing and fraud detection using mass spectrometry

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed Date 2017.

ENVG FALL ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques

Measurement of 234 U/ 238 U and 230 Th/ 232 Th in volcanic rocks using the Neptune MC-ICP-MS

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

ICP-OES DETERMINATION OF IRON. Introduction:

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

Determination of Elements at Sub-ppb Concentrations in Naphtha Mixtures Using the NexION 300 ICP-MS

Analysis Repeatability of Trace and Major Elements in a Water Sample

Emission spectrum of H

Transcription:

APPLICATION NOTE 30355 Low level Os isotopic measurements using multiple ion counting Authors Introduction Jean Louis Birck, Delphine Limmois, Institut de Physique du Globe de Paris, Sorbonne Paris Cité, France Lionnel Mounier, Nicholas Lloyd, Johannes B. Schwieters, Thermo Fisher Scientific, Germany The Re-Os chronometer is an important chronometer in earth sciences. The most striking feature is that both elements preferentially partition into the metallic core of planets rather than into the crust and mantle. In silicate partitioning, Os prefers the earth s mantle rather than the crust, although this is a little less true for Re, resulting in crustal rocks with very high Re/Os ratios, leading to variations of the 187Os/xOs (x=186 or 188 or 189) from radioactive decay of 187Re into 187Os. Relative to the other common radioactive chronometers, the 187Os/188Os variations are large and extend over at least 3 orders of magnitude, e.g. ranging from around 0.12 for mantle rock to ~1 for average terrestrial crustal rock to up to several 100 in molybdenites or Archean basalts. Nevertheless, there are caveats to these attractive properties: the very low level of Re and Os concentration in most terrestrial surface rocks on one hand and the high ionization potential of Os on the other. Hence, until the development of high efficiency n-tims, the Re-Os chronometer has been only marginally used for the study of earth s surface rocks. Although n-tims ionization yields range up to 20% for Re-Os, silicate rocks in general have such low concentrations of Os, that the use of an electron multiplier is required. Until recently, the later was used in a single detector mode using magnetic field peak hopping to measure the required number of isotopes. Keywords Isotope ratio, osmium, multicollector ICP-MS, Neptune Plus, multiple ion counting, MIC

Plus Platform IC IC IC IC IC L4 L3 L2 L1 C/IC H1 H2 H3 H4 202Hg 204Pb 206Pb 207Pb 208Pb 185Re 187Os 189Os 190Os 192Os Figure 1. Multiple Ion Counting configuration for Pb and Os isotope ratio analysis on Neptune Plus MC-ICP-MS. The purpose of this note is to extend the low level Os isotopic measurements to multicollector ICP-MS instrumentation. Peak hopping on a single electron multiplier is excluded for two major reasons: 1) ICP-MS is at least 5 times less sensitive than n-tims and 2) beam stability in plasma sources is usually poorer than for TIMS. The possibility to use static multiple collection on secondary electron multiplier ion counters can compensate for these two drawbacks. Here we describe a procedure for calibrating the multiple ion counting system using Os standards using at least ten times larger intensities than the samples. This strategy allows for the precise correction of detector yield drift. Detector configuration A Thermo Scientific Neptune Plus MC-ICP-MS was configured with a multi-ion-counting array that is designed for the collection of Pb isotopes (Figure 1), with the collector spacings fixed for Pb isotopes. The dynamic zoom lens system of the Neptune Plus MC-ICP-MS was used to adjust the dispersion of the Re-Os ion beam into these collectors (dispersion lens +90 V, focus lens -15 V). Figures 2 and 3 show the peak shape and overlap for the Re-Os isotopes in low resolution and medium resolution modes. The peak shape proves to be adequate at least for low level Re-Os measurements, where the precision is limited by counting statistics only. Figure 2. Mass overlap in the Os mass range in the low resolution mode. 1 ppt Os standard solutions are used for these displays. For measurements in the low resolution mode, the mass is set at 206.60 on the axis of the instrument. The peak shapes drastically improve when using the Medium Resolution mode (Figure 3), implying that both the off-axis location of the multiple ion counter and the extreme values of the zoom lens voltages do not produce visible optical aberrations in the Os mass spectrum. Figure 3. Mass overlap in the Os mass range in the medium resolution mode. A 1 ppt Os standard solution is used to produce these spectra. When measuring a sample, mass is set at 206.66 on the axis of the instrument. 2

Background spectra and ion source settings Usually Os is stored in solution in more than 4N HBr media provided it is stabilized as an Os(Br) 6 2- complex. In our opinion, it is not wise to use such high halogenate solutions in an ICP-MS as it would lead to undesired interferences in the mass spectrum and possible corrosion problems on the metal parts of the source region. Solutions of Os were prepared in nitric acid. Nitric acid is not oxidizing below 2N, but because of the absence of the stabilizing HBr, we added a reducing agent which prevents the escape of Os from solution as OsO 4 and retention in a reduced form in the plastic bottles or vials. All final solutions are in 0.3% HNO 3 + 0.03% hydroxylamine. Our experience has shown that such solutions are stable in concentration over a few years. The introduction system is an Apex desolvation nebulizer (ESI, Omaha, USA) with a self-aspirating nebulizer and an uptake rate of 0.1 ml per minute. The cones are a Jet sample cone and a standard H skimmer cone (for enhanced heavy mass sensitivity with low oxides). Typical beam sizes for a 100 ppq Os solution are about 1600 cps of 192 Os translating into an ion per atom yields around 0.6%. The minimum amount of Os required to carry out a measurement within the precision frame of this technical note is close to 33 fg. Unlike TIMS, which usually displays less than 0.5 cps in the OsO 3 - mass range, ICP-MS always displays residual signals in the 180 190 mass range. The spectrum depends on the history of the instruments in the days before the Os measurement and is a combination of memory effects on Re-Os and interferences resulting from molecular ions which include combinations of iron group elements with Ar. The later are mass shifted downwards relative to elemental Os in medium resolution mode (Figure 4). In this figure these polyatomic interference ions were deliberately enhanced for visibility. Satisfactory background levels (typically less than 50 cps 192 Os+), equivalent to about 1 ppq Os can be obtained by using clean cones and alternating: the above running solution, 5% HNO 3 and dilute hydroxylamine in water. Measurement procedure The instrument is tuned on Faraday cups using a tungsten standard. The zoom voltages are adjusted using a 1 ppt Re-Os standard solution to fit the Os isotopes in the detectors according to the schematics in Figure 1. Figure 4a. Interference on mass 187 (IC5, light brown or CDD 3 in Figure 1) from a low mass cluster on mass 187 shown in the medium mass resolution mode. For a Re-Os measurement, the axis mass is set on 206.66 to discard the cluster from mass 187 Os or 187 Re. Figure 4b. For demonstration in medium mass resolution mode, the interfering clusters on mass 187 have been enhanced by adding V to the running solution. Every three samples, and after a few minutes of rinsing, a blank (on-peak zero) is measured for 3 4 minutes. The average of the background levels on the different masses taken before and after each set of three sample measurements is subtracted from the sample signals. Generally this background is stable over the day or tends to decrease very slowly as far as < 10 ppt Os solutions are introduced into the instrument and bromides are only at trace levels in the solution. A single sample measurement consists of 210 seconds data (50 cycles of 4.2 seconds integrations). Standards solutions from 0.1 to 50 ppt were used in this study. 3

Ratios are all referred to 189 Os, because the collector setup of our instrument does not permit the measurement of 188 Os together with 187 Os, 189 Os, 190 Os (spike), 192 Os and 185 Re. Instead, we use 189 Os as reference isotope. An additional advantage of using 189 Os is that 189 Os is not affected by interferences from clustering ions of iron group elements with argon (resulting from impure solutions or insufficient instrument cleaning). The following data evaluation procedure is used: For each ratio after background correction, the measurements of all solution equal or above 10 ppt are averaged and a bias factor is calculated relative to the reference value obtained on a Faraday cup. This bias factor is then applied to all individual ratios to obtain the corrected values. It combines the mass bias and ion counter yield corrections. No further normalization is applied. This procedure assumes that once the instrument is tuned, the instrumental fractionation drift over the session duration is insignificant to the uncertainty from counting statistics. The data show that this assumption is valid. A rhenium correction is applied for mass 187 using the measured mass 185 with a reference value measured for 187 Re/ 185 Re from a 0.2 ppt Re standard solution measured within the same session. The normal running setup uses the Elemental Scientific Apex desolvating nebulizer system and the Low Resolution mode in order to have the highest possible signal level. A series of standards at various concentrations are measured in a row and also by alternating two solutions different by a factor of 30. The results are displayed in Figures 5 and 6. The results show that for 187 Os/ 189 Os the reproducibility increases from the percent level for 0.1 ppt solutions to the permil level for solutions above 4 ppt. For 190 Os/ 189 Os and 192 Os the 2 permil level of reproducibility is already reached for 0.3 ppt Os solution. The error bars from the different Os concentrations overlap both for the individual measurements and for the average ratios and show that there is no significant signal dependence within the explored range from 1,500 to 500,000 cps 192 Os signal. We mention here that the Os standard has a very low 187 Os/ 189 Os mantle-like isotopic composition. Low Resolution measurement sequence with Apex inlet system 187 Os/189 Os 190 Os/189 Os 192 Os/189 Os 0.094 0.092 0.090 0.088 0.086 0.084 0.082 1.65 1.64 1.63 1.62 1.61 1.60 2.56 2.55 2.54 2.53 2.52 2.51 2.50 2.49 0.1 ppt 0.3 ppt 1 ppt 4 ppt 10 ppt 40 ppt 2.48 0 5 10 15 20 25 30 35 Figure 5. Individual measurements of 210 s over the whole concentration range of this study in Low Resolution mode. Error Bars from Standard Error of the Mean for each measurement (1σ). A B C 4

With seawater-like compositions, which have about ten times more 187 Os relative to e.g. 189, a reproducibility of ca. 0.5% is anticipated with a 0.1 ppt solution. A similar data set was obtained by replacing the Apex desolvation system by a cyclonic spray chamber (wet plasma) reducing signal levels by a factor of 4: range 330 to 170,000 cps for solutions from 0.1 ppt to 50 ppt with an injection rate of 0.1 ml/min. Due to poorer counting statistics, the 187 Os/ 189 Os reproducibility is better than 1% only above 1 ppt Os concentration. 187 Os/189 Os Medium Resolution measurement sequence with cyclonic spray chamber 0.094 0.092 0.090 0.088 0.086 0.084 1 ppt 2 ppt 5 ppt 10 ppt 20 ppt 50 ppt A The Medium Resolution mode (R = 8,000) was also investigated with the cyclonic spray chamber, reducing the signal level by another factor of 4, typically from 90 cps to about 45,000 cps of 192 Os beam for the standards from 0.1 to 50 ppt total Os. 187 Os signal are very low ranging from 3 to 1,500 cps. Measured 185 Re beams are below 0.5 cps. Nevertheless the resulting data fit very well within increased error bars with higher signal data. This shows that there is no residual bias from the data correction process at very low signal. These results are displayed in Figure 6. A similar data set was also obtained in the Medium Resolution mode using the Apex desolvating nebulizer and are not displayed here. The conclusion is the same: the data fit to the expected results within the error bars. 190Os/189 Os 192 Os/189 Os 0.082 1.65 1.64 1.63 1.62 1.61 1.60 2.56 2.55 2.54 2.53 2.52 2.51 B C Conclusion The dynamic zoom lens optics of the instrument allowed precise Os isotopic ratios in the ppt and subppt concentration range. With 30 fg Os, a significant 187 Os measurement can be reached with the Pb MIC configuration on a Neptune Plus MC-ICP-MS. Depending on the 187 Os/ 188 Os ratio, such low amounts allow precisions from 1.5% for mantle type ratios to less than 0.5% for crustal radiogenic ratios. Within the range investigated here (0.05 to 20 pg), the final error is limited by counting statistics. 2.50 2.49 2.48 0 10 20 30 40 50 60 Run No. Figure 6. Average ratios for the different concentrations of solution: 1, 2, 5, 10, 20, and 50 ppt. Find out more at www.thermofisher.com/irms 2016 Thermo Fisher Scientific Inc. All rights reserved. Apex is a trademark of Elemental Scientific. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. AN30355-EN 0117S