Selecting an LC/MS Interface Becky Wittrig, Ph.D.

Similar documents
Chemistry Instrumental Analysis Lecture 37. Chem 4631

M M e M M H M M H. Ion Sources

Molecular Mass Spectrometry

Lecture 15: Introduction to mass spectrometry-i

Chemistry Instrumental Analysis Lecture 34. Chem 4631

Mass Spectrometry (MS)

LC-MS Based Metabolomics

Molecular Mass Spectrometry

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Mass Spectrometry. General Principles

TANDEM MASS SPECTROSCOPY

1) In what pressure range are mass spectrometers normally operated?

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

Mass Spectrometry. Electron Ionization and Chemical Ionization

Mass Spectrometry in MCAL

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n

API Optimization, Source Design, and Other Considerations of LC/MS

Chemistry 311: Topic 3 - Mass Spectrometry

Mass Spectrometry for Chemists and Biochemists

MS Goals and Applications. MS Goals and Applications

Lecture 8: Mass Spectrometry

Ionization Techniques Part IV

Lecture 8: Mass Spectrometry

Ionization Methods in Mass Spectrometry at the SCS Mass Spectrometry Laboratory

Electrospray Ion Trap Mass Spectrometry. Introduction

MS Goals and Applications. MS Goals and Applications

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

MASS SPECTROMETRY. Topics

Mass Spectrometry: Introduction

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Analysis of Polar Metabolites using Mass Spectrometry

IDENTIFICATION OF ORGANOMETALLIC COMPOUNDS USING FIELD DESORPTION IONIZATION ON THE GCT

Introduction to LC-MS

Mass spectrometry.

Technical Note. Introduction

Introduction to GC/MS

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Ch 13. Basics of Mass Spectrometry (I) : Principles & Ionization Sources

Analytical Technologies and Compound Identification. Daniel L. Norwood, MSPH, PhD SCĪO Analytical Consulting, LLC.

Ch.28 HPLC. Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography)

CEE 772: Instrumental Methods in Environmental Analysis

Mass Spectrometry. Fundamental LC-MS. Electrospray Ionisation Instrumentation

Liquid Chromatography in conjunction with Mass Spectrometry (LC-MS)

Agilent G3212 GC-APCI Source

Auxiliary Techniques Soft ionization methods

Mass spectrometry gas phase transfer and instrumentation

Qualitative Organic Analysis CH 351 Mass Spectrometry

Presentation Basic Introduction to Instrumentation Matrix Effects Challenges

Chapter 5 Basic Mass Spectrometry

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Hypericin Analysis by LC/MS

CHROMATOGRAPHY AND MASS SPECTROMETER

Introduction. Chapter 1. Learning Objectives

Harris: Quantitative Chemical Analysis, Eight Edition

1. The range of frequencies that a measurement is sensitive to is called the frequency

Mass spectrometry and elemental analysis

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

HPLC solvents and mobile phase additives

CEE 772: Instrumental Methods in Environmental Analysis

The AccuTOF -DART 4G: The Ambient Ionization Toolbox

MS/MS .LQGVRI0606([SHULPHQWV

Translational Biomarker Core

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York

Quattro Micro - How does it work?

4. How can fragmentation be useful in identifying compounds? Permits identification of branching not observed in soft ionization.

Selecting Detectors for Compounds with No Optical Absorbance

BIOINF 4399B Computational Proteomics and Metabolomics

Mass Spectrometry. A truly interdisciplinary and versatile analytical method

Choosing the metabolomics platform

Mass Spectrometry. Introduction EI-MS and CI-MS Molecular mass & formulas Principles of fragmentation Fragmentation patterns Isotopic effects

Construction and Use of API Single- Quadrupole MS Spectral Libraries.

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

15.04.jpg. Mass spectrometry. Electron impact Mass spectrometry

Propose a structure for an alcohol, C4H10O, that has the following

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Welcome to Organic Chemistry II

WADA Technical Document TD2003IDCR

Mass spectrometry: forming ions, to identifying proteins and their modifications Stephen Barnes, PhD

Mass Spectrometry Course

Part I Introduction to MS in bioanalysis

Finnigan LCQ Advantage MAX

Basics of Mass Spectrometry

MS Interpretation I. Identification of the Molecular Ion

Simultaneous Determination of Paraquat and Diquat in Environmental Water Samples by HPLC-MS/MS

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

LC MS analysis of metabolites. Basis of Chromatography

2. Separate the ions based on their mass to charge (m/e) ratio. 3. Measure the relative abundance of the ions that are produced

Ch. 22 Mass Spectrometry (MS)

Supporting Information

Lecture 1 Introduction- Protein Sequencing Production of Ions for Mass Spectrometry

Webinar: How to be More Successful with your ACQUITY QDa Detector Eric van Beelen

Fast Atom Bombardment (FAB)-MS

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications

LC and LC/MS Column Selection Flow Chart

6.5 Optical-Coating-Deposition Technologies

Hplc Lc Ms And Gc Method Development And Validation Guideline For Academic And Industrial Scientists Involved In Method Development And Validation

IONIZATION. LC/EI MS may extend the simplicity. of GC/MS to many HPLC-amenable. compounds without compromising

The role of LC-MS in Toxicology

Transcription:

Selecting an LC/MS Interface Becky Wittrig, Ph.D. RESTEK CORPORATION

LC/MS Interfaces I. Background of LC/MS I. Historical Perspective II. Reasons for use II. Interfaces I. Transport devices II. Particle Beam III. Thermospray IV. Atmospheric Pressure Interfaces

Historical Perspective Goldstein 1886 Existence of positively charged particles Wein 1898 Positively charged ions can be deflected in electrical and magnetic fields J.J. Thomson 1913 Demonstrated isotopes of Neon Father of mass spectrometry First GC-MS interface - 1960 s

Historical Perspective First LC-MS interface developed - 1969 1uL/min flow into an EI source Transport devices applied to LC/MS - 1970 s Loss of volatile components Thermally-reactive compound losses Thermospray (TSI) gains popularity - 1983 1.0 1.5 ml/min Mobile phases consist primarily of an aqueous buffer

Historical Perspective Atmospheric Pressure Interfaces (API) Early 1990 s (commercialization) Now most common interface Electrospray (ESI) ¾Initial interfaces required lower flows (1-5 ul/min) ¾Able to produce multiply-charged molecules Atmospheric Pressure Chemical Ionization ¾Similar to Thermospray ¾ Solvent-mediated ionization

Why Use Mass Spectrometry? Spectral resolution is possible Chromatographic coelutions Compound identification from spectral data Mass spectrum is very dependant on the ionization Limited availability of LC/MS libraries High degree of specificity

LC/MS Interfaces I. Background of LC/MS I. Historical Perspective II. Reasons for use II. Interfaces I. Transport devices II. Particle Beam III. Thermospray IV. Amospheric Pressure Interfaces

Components of an MS Detector Ions Made Ions Selected Ions Detected

LC/MS Systems Interfaces Analyzers Transport Devices Particle Beam Thermospray Atmospheric Pressure Electrospray APCI Others Quadrupoles Ion Trap MS n High Resolution Time-of-Flight (TOF) Others

Ion Sources Analytes must be charged (ions) Needed to separate and detect Where ionization occurs In the interface In a separate ion source Types of ionization Hard - considerable fragmentation Soft - molecular ion is main product Ions can be positive or negative

Requirements of LC/MS Interfaces Elimination of the mobile phase Most difficult step Can use splitters Volatilized solvent vapor removed under vacuum Often where ionization occurs Vacuum required by mass analyzers

Challenges in Interfacing LC to MS ƒ Flow Rate Concerns ƒ Differential pumping can only handle 2 µl/min of water ƒ For maximum sensitivity, want to use all of the eluent ƒ Use of Buffers and Additives ƒ Non-volatile buffers a concern ƒ Some additives suppress ionization ƒ Wide Range of Analytes ƒ Many are nonvolatile, thermally labile

Transport Devices One of the first commercial interfaces Sample deposited onto a moving belt or wire Sample passes through multiple vacuum zones Solvent elimination Sample is desorbed into source using heat Electron impact ionization Belt/wire cycles back

Moving Belt Interface From: Niessen

Particle Beam Interface Column effluent is nebulized Pneumatic or thermospray nebulization Desolvation chamber is under a moderate vacuum A momentum separator is used for analyte enrichment High MW compounds favored Analytes into the EI or CI source as small particles Evaporative collisions with the walls

Particle Beam Interface From LC Nebulization Desolvation Skimmers Helium added Pumping System Source EI or CI typical Momentum Separation

Analyte Enrichment in PB Interfaces Analyte Ion Solvent Pumping System Analyte enrichment with a molecular beam approach Heavier molecules are in the core of the vapor jet and are sampled through the skimmer

Electron Impact Ionization LC interfaces with the ion source Electrons are boiled from a hot wire (filament), and accelerated (70eV) As electrons pass neutral molecules, they may remove outer shell electrons M + e - Æ M +. + 2e - M +. Æ F + + N.

Electron Impact Ionization Produces positively charged ions Fragmentation is generally significant for most molecules hard ionization Masses of these fragments is the information used in interpretation Efficiency of ionization is 1/10 5 Bulk of molecules are removed by vacuum pump Use Traps

Electron Impact Ionization (EI) filament Sample from LC e - e - e - e - e - + + + + Fragment ion Mass Analyzer

Thermospray Interface Nebulization of the eluent from a heated transfer tube Uses a reagent gas Mobile phase buffer Added buffer solution Similar spectra to GC/MS CI Reagent gas is ionized Volatilization EI with high energy electrons Charge transfer to the analyte(s)

Thermospray Interface Effectively replaced transport systems Ionization in a medium pressure environment Approx. 1000 Pa or 0.01 atm Inlet flows of 1 to 2 ml/min Can use standard LC column flows Positive and negative ions are possible Temperature optimization is critical Maintain gas phase

Thermospray LC/MS System From: McMaster and McMaster

Thermospray Ionization Positive Ion Mode ƒ M + SH + MH + + S PA(M) > PA(S) ƒ M + SH + MHS + PA(M) PA(S) Negative Ion Mode ƒ M + [S-H] - [M-H] - + S gpa(m) > gpa(s) ƒ M + A - MA - polar molecules gpa = gas phase acidity

Typical Proton Affinities Compound Water Methanol Acetonitrile Ethers, esters, ketones PAHs Carboxylic Acids Alcohols Peptides Proton Affinity 723 773 797 630-670 710-800 <800 750-840 880-1000 From: Niessen

Thermospray Modes of Operation Solvent-Mediated CI Typical mobile phase: MeOH or MeCN + 0.1M NH 4 OAc Ionization by ion evaporation (pre-formed ions) or ion-molecule reactions Discharge or Filament On Mode High energy electrons (0.5-1 kev) ionize the reagent gas Ion-molecule reactions

Electrospray Interface High electric potential applied to eluent from transfer capillary Atmospheric pressure Droplet formation Ionization in the solution phase Orthogonal sampling of ions Reduces contamination of the sampling orifice Z-Spray devices

Electrospray Interface Higher flows now possible Pneumatic + thermal nebulizers Applicable to volatile and nonvolatile analytes Need to be ionizable Can create multiply-charged ions Allows for analysis of large molecules

Electrospray LC/MS System From: McMaster and McMaster

Electrospray Interface Nebulizer gas +/- 3 kv Skimmers Quadrupole Transfer capillary Electrospray (Taylor cone) Focusing Lenses First pumping stage Second pumping stage

Electrospray Needle Design From LC Nebulizer gas Sheath Liquid Electrospray Droplets Example of an electrospray needle design (coaxial flow) The nebulizer gas disrupts the liquid surface so that small droplets are formed and then dispersed by the gas

Theory of API Electrospray Charged Droplets + + + - + + - - + - - + + Evaporation + + Analyte Ions Solvent Ion Clusters Salts/Ion pairs Neutrals Rayleigh Limit Reached + + ++ - - - - - + + + + + + + + - - + + Coulomb Explosions + + + + + - - + + + + + Solvent Ion Cluster Analyte Ion Courtesy of Agilent Technologies

Agilent 1100 LC/MSD Electrospray HPLC inlet Nebulizer gas inlet Nebulizer Skimmers Lenses HED detector Capillary Octopole Quadrupole Neutral Molecules Analyte Ions Clusters Salts + + + + + + + + + + + + + + + + + + heated N 2 Waste Fragmentation zone (CID) Courtesy of Agilent Technologies

Electrospray Interface Best vaporization with higher % organic and lower flow rates Cluster ion formation possible Solvent clusters Analyte/salt clusters Salts and sample impurities can affect the response TFA causes signal suppression TFA anion masks the analyte ion

APCI Interface Atmospheric Pressure Chemical Ionization Initially investigated in 1974 Popular in late 1980 s Uses an API (ESI) source Column effluent nebulized into heated vaporizer tube Solvent vapor acts as a reagent gas Charge transfer to the analytes Can be very sensitive

APCI Interface Results in a chemical ionization spectrum [M+H] + or [M-H] - Products depend on equilibrium (concentration) conditions Analytes must have sufficient proton affinities May be simplest interface to operate Liquid flow rates of 0.2-2.0 ml/min [Solv+H] + + M --> Solv + [M+H] +

APCI Interface From: Scott

APCI Probe (Heated) From LC Auxiliary gas Nebulizer Gas Vaporizer The vaporizer tube temperature is optimized for complete transfer to the vapor state

Theory of APCI Analyte containing aerosol Heat + + + + + + + + + + + + + + + + + + + + + + + Charge transfer to analyte Vapor Charged reagent gas formed + + + + + Analyte ions Courtesy of Agilent Technologies

Agilent 1100 LC/MSD - APCI Nebulizer Skimmers Lenses HED detector Capillary Octopole Quadrupole Heater + + + + + + + + + + + + + + + + Corona needle heated 2 N2 Waste Courtesy of Agilent Technologies Fragmentation zone (CID) Vaporize to the gas phase and ionize the gas with a discharge

Agilent 1100 LC/MSD APCI Ion Source Corona Needle Vaporize into the gas phase and ionize the gas with a discharge Courtesy of Agilent Technologies

API-ES vs. APCI for Triamterene 400000 254 M+ H 2E6 254 M+ H 300000 API-ES 1.5E6 APCI 200000 1E6 100000 255 276 M+ Na 500000 255 0 0 100 200 300 100 200 300 m/z Comparison of the mass spectra typical of Electrospray vs. APCI

Optimizing Mobile Phases for API-MS Ion-Pair Agent Alternatives ¾ ¾ ¾ Use highly bases-deactivated silica columns Use low ph (3-4) to reduce tailing Use columns that retain based on polar interactions (e.g. CN, IBD) Using Ion-Pair Reagents ¾ Use low amounts (<0.02%) ¾ Use post-column addition to negate the effect of the ion-pair agent

Sample Considerations for LC/MS The Analyte Must Have Ionizable Groups ¾ ¾ ¾ Amines Carboxylic Acids Ketones, Aldehydes For Best Sensitivity, Work at a ph Where the Analyte is Ionized ¾ ¾ Neutral to basic ph (7-9) for acids Acidic ph (3-4) for bases

Sample Considerations Positive Ion Mode ¾ Analyte = (M+H) + Negative Ion Mode ¾ Analyte = (M-H) -

Basic Compound Sensitive in Positive Ion Mode

Acidic Compound Sensitive in Negative Ion Mode

Selecting an Interface Advances in LC/MS, Waters Corporation, Milford, MA.

Electron Ionization (EI) Analytes Suitable for EI ¾ ¾ ¾ Small molecules with rings and double bonds Compounds that would need derivatization for GC/MS Pesticides, PAHs, natural products Compound Identifications ¾ Fragmentation is possible Poor Detection Limits Will Not Tolerate Non-Volatile Buffers

Comparison of API vs. EI Atmospheric Pressure Ionization (API) MW confirmation Good for fragile compounds Able to fragment in the source Low (ppb) LODs in SIM mode Electron Ionization (EI) Ionization occurs in a vacuum Standard libraries are available Classical EI spectra (similar to GC/MS) Higher LODs (ppm to high ppb)

Atmospheric Pressure Ionization (API) Electrospray Ionization (ESI) Uses Solution Phase Ionization Atmospheric Pressure Ionization (APCI) Uses Gas Phase Ionization Products are [M+H]+ and [M-H]-, adducts Suitable for Analyzing Drugs, Small Molecules, Dyes, Peptides Good Sensitivity Thermal Degradation is Possible (APCI)

Adduct Formation with API H 3 C K + H 3 C H 3 N O O H H 3 C H 3 C Adducts can form between polar molecules and sample or solvent components. For example, adducts with Na+, K+, NH 4 +, MeOH, MeCN, and H 2 O are common.

Electrospray Ionization + Good molecular weight information, including high MW compounds + Can be used for volatiles, nonvolatiles, ionic/polar compounds + Good sensitivity Need relatively low flow rates Need to be able to form ions in solution Limited structural information Problems with high aqueous and buffer solutions

Atmospheric Pressure Chemical Ionization + Gives molecular weight information + Easy to use, rugged + Can use higher LC flow rates (up to 2 ml/min) Thermal degradation can occur Limited structural information Not appropriate for higher MW (e.g., >1000 Da)

MALDI Matrix-assisted laser desorption/ionization Sample is deposited on a target and cocrystallized with a solid matrix (dihydroxybenzoic acid) Desorption/ionization occurs using a laser such as Nd-YAG (266nm, v 4 ) Energy transferred to matrix, then analyte Useful in excess of 200kDa (biomacromolecules)

Fast Atom Bombardment (FAB) ƒ Cf-FAB = Continuous Flow FAB ƒ Column eluent mixes with a matrix (glycerol) ƒ Eluent + matrix deposited on a target ƒ Analyte film hit with fast atoms or ions ƒ Only low flow rates used

Acknowledgements Charlie Schmidt - Thermo Instruments Agilent Technologies Liquid Chromatography-Mass Spectrometery, Marcel Dekker, New York, NY, W.M.A. Niessen ed. 1999 Chromatographic Detectors, Marcel Dekker, New York, NY, Raymond P.W. Scott ed. 1996 GC/MS A Practical Users Guide, Marvin C McMaster and Christopher McMaster, Wiley-VCH, New York, NY 1988