EFFICIENCY AND RESOLUTION OF GERMANIUM DETECTORS AS A FUNCTION OF ENERGY AND INCIDENT GEOMETRY

Similar documents
The Shape of the Full-Energy Gamma Ray Peak in Germanium Detectors for Different Energies and Incident Geometries

PROFILE S and C Series

PROFILE S and C Series P-type Semi-planar and Coaxial HPGe Detectors

Background Reduction Using Collimators on a Portable HPGe Nuclide Identifier

Minimum Detectable Activity Estimates for a Germanium-Detector Based Spectroscopic Portal Monitor

A Digital Method for Dead Time Compensation in Nuclear Spectroscopy

Compton suppression spectrometry

ORTEC. Overview of Semiconductor Photon Detectors. GEM (P-type) and GAMMA-X (N-type) Coaxial Detectors. Efficiency as a Function of Energy

EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD

Performance of the True Coincidence Correction Method in GammaVision. Keyser, Ronald M., Haywood, Susan E., and Upp, Daniel L.

Calibration of A Small Anode Germanium Well Detector

ENERGY PEAK STABILITY WITH COUNT RATE FOR DIGITAL AND ANALOG SPECTROMETERS

COUNT RATE EFFECTS ON PEAK SHAPE AND THROUGHPUT IN DIGITAL AND ANALOG GAMMA RAY SPECTROMETRY

The first model was based only on the CANBERRA information which we. drawing. Canberra refused any other information.

Characterization and Monte Carlo simulations for a CLYC detector

BEGe Detector studies update

Determination of the activity of radionuclides

DESIGN, DEVELOPMENT AND TESTING OF AN AUTOMATED SEGMENTED GAMMA SCANNER FOR MEASURING NUCLEAR POWER STATION RADIOACTIVE WASTE

Collimated LaBr 3 detector response function in radioactivity analysis of nuclear waste drums

Effect of Co-60 Single Escape Peak on Detection of Cs-137 in Analysis of Radionuclide from Research Reactor. Abstract

WM2014 Conference, March 2 6, 2014, Phoenix, Arizona, USA

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data

τ lepton mass measurement at BESIII

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

Digital simulation of neutron and gamma measurement devices

Gamma-ray spectroscopy with the scintillator/photomultiplierand with the high purity Ge detector: Compton scattering, photoeffect, and pair production

arxiv:nucl-ex/ v2 21 Jul 2005

arxiv: v1 [physics.ins-det] 14 Nov 2017

Determination of the shielding power of different materials against gamma radiation

ORTEC. GEM Series Coaxial HPGe Detector Product Configuration Guide. Configuration Guidelines

Non-Destructive Assay Applications Using Temperature-Stabilized Large Volume CeBr Detectors 14277

ISOCS / LabSOCS. Calibration software for Gamma Spectroscopy

AGATA preamplifier performance on large signals from a 241 Am+Be source. F. Zocca, A. Pullia, D. Bazzacco, G. Pascovici

Quartz-Crystal Spectrometer for the Analysis of Plutonium K X-Rays

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

Applications of Pulse Shape Analysis to HPGe Gamma-Ray Detectors

CHAPTER 5 EFFECTIVE ATOMIC NUMBER OF SELECTED POLYMERS BY GAMMA BACKSCATTERING TECHNIQUE

The Best Gamma-Ray Detector

Identification of Naturally Occurring Radioactive Material in Sand

FRAM V5.2. Plutonium and Uranium Isotopic Analysis Software

Gamma-Spectrum Generator

NGN PhD Studentship Proposal

ORTEC. GLP Series Planar HPGe Low-Energy Detector Product Configuration Guide. Configuration Guidelines

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Gamma Analyst Performance Characteristics (MDAs)

Advanced and Special Detector Programs from ORTEC Historical and Next Step Perspective

1 of :32

Introduction to gamma spectroscopy on a budget. Daniel Emes SA Radiation

Gamma ray coincidence and angular correlation

Surface effects in Segmented Germanium Detectors

ISO INTERNATIONAL STANDARD

Semiconductor Detector

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

SOME ASPECTS OF MONTE CARLO SIMULATION FOR EFFICIENCY CALIBRATION OF GERMANIUM DETECTORS

ORTEC was founded in 1960 by researchers from Oak Ridge National Labs to commercialize charged particle detectors

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

NUCL 3000/5030 Laboratory 2 Fall 2013

Interactive Web Accessible Gamma-Spectrum Generator & EasyMonteCarlo Tools

European Project Metrology for Radioactive Waste Management

Renewed whole-body counting chamber in STUK

Design of a virtual model of a hand-held Germanium detector and a voxelized ICRP whole body phantom: A Monte Carlo study

VIRTUAL GAMMA-RAY SPECTROMETRY FOR TEMPLATE-MATCHING NUCLEAR WARHEAD VERIFICATION. Janet Schirm and Alexander Glaser

The Advanced Gamma Ray Tracking Array AGATA

1 Introduction. 2 Method. Robert Metzger 1,*, Kenneth Van Riper 2, and George Lasche 3

Gamma Ray Spectrometry

High Purity Germanium Detector Calibration at ISOLDE

SYSTEM FOR PIXE TRACE ELEMENT ANALYSIS AT MTF. M. Beňo, J. Dobrovodský, D. Vaňa, S. Minárik

in Cross-Section Data

1 cm. Cu electrode. Ge crystal. end cap. Dead layer. crystal cup

A MONTE CARLO SIMULATION OF COMPTON SUPPRESSION FOR NEUTRON ACTIVATION ANALYSIS. Joshua Frye Adviser Chris Grant 8/24/2012 ABSTRACT

Uncertainty Quantification for Safeguards Measurements

CALIBRATION OF SCINTILLATION DETECTORS USING A DT GENERATOR Jarrod D. Edwards, Sara A. Pozzi, and John T. Mihalczo

VALIDATION OF IQ3 MEASUREMENTS FOR HIGH-DENSITY LOW-ENRICHED-URANIUM WASTE DRUMS AT PELINDABA

Characterization of Large Structures & Components

THE USE OF WAVEFORM DIGITIZERS WITH SOLID STATE DETECTORS FOR SPECTROSCOPIC APPLICATIONS. Carly Jean Smith

W. Udo Schröder Departments of Chemistry & of Physics and Astronomy

Energy resolution and absolute detection efficiency for LSO crystals: a comparison between Monte Carlo simulation and experimental data

ORTEC. GEM Series Coaxial HPGe Detector Product Configuration Guide. Configuration Guidelines

Positron Annihilation Spectroscopy

Status of Pulsed Neutron Facility in Korea

The Evaluation of Program ISOTOPIC to Nondestructively Quantify Uranium and Plutonium Samples at the Joint Research Center in Ispra, Italy

High-Performance Germanium Coaxial Detectors for Safeguards and Non-Destructive Assay Product Configuration Guide

HEROICA: a test facility for the characterization of BEGe detectors for the GERDA experiment

Activities of the neutron standardization. at the Korea Research Institute of Standards and Science (KRISS)

Characterizing New Multi-Channel Peak Sensing ADC-Mesytec MADC-32

Experimental Studies on the Self-Shielding Effect in Fissile Fuel Breeding Measurement in Thorium Oxide Pellets Irradiated with 14 MeV Neutrons

Chem 481 Lecture Material 3/20/09

1. Introduction. Etim I.P. 1, William E. S. 2, Ekwe S.O. 3. Licensed Under Creative Commons Attribution CC BY

Preparation of Standard Source as a Petri Dish for Plant by Using

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

Monte Carlo modeling of an electronic brachytherapy source using MCNP5 and EGSnrc

Monte Carlo Characterization of a Pulsed Laser-Wakefield Driven Monochromatic X-Ray Source

Recent improvements in on-site detection and identification of radioactive and nuclear material

Evaluation of the Nonlinear Response Function and Efficiency of a Scintillation Detector Using Monte Carlo and Analytical Methods

Distance from the face of the detector to the surface of the item being measured.

Neutron and Gamma Ray Imaging for Nuclear Materials Identification

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN

REACTION RATES OF RESIDUAL NUCLEI PRODUCED OF 59 Co AT THE TARGET QUINTA

JRPR. The Development of Gamma Energy Identifying Algorithm for Compact Radiation Sensors Using Stepwise Refinement Technique.

Transcription:

EFFICIENCY AND RESOLUTION OF GERMANIUM DETECTORS AS A FUNCTION OF ENERGY AND INCIDENT GEOMETRY R.M. Keyser New Product Development, ORTEC, 801 South Illinois Avenue, Oak Ridge, TN 37831 1 INTRODUCTION The use of germanium detectors for the identification and quantification of radionuclides in unknown and non-standard counting geometries has increased in the recent past with the need to clean up and verify sites used for radionuclide processing, smuggling detection and other purposes. The calculation of the amount of radionuclide present requires a knowledge of the efficiency of the detector in the counting geometry. Several methods of determining the efficiency in these unusual geometries have been developed over the years. These methods are used in waste measurement, field (in situ) measurements and material verification. One method currently being used is the modeling of the detector response using Monte Carlo simulation programs, especially Monte Carlo N-Particle Code (MCNP). The input to MCNP is a detailed knowledge of the detector construction details, detector crystal details and the source being measured. This paper addresses the difficulty in obtaining the required detector crystal details, even when the common specifications are well known. In addition, the Institute of Electrical and Electronic Engineers (IEEE) Germanium Test Standard (325-1996) is widely used for the specification of High Purity Germanium (HPGe) detectors. The two important specifications are the efficiency and the resolution of the detector. However, the IEEE 325-1996 standard only specifies the Full Width at Half Maximum (FWHM) measurement at one geometry and two energies. Thus, this standard does not apply in most counting geometries used today, especially for environmental samples. Modeling programs, such as MCNP, use the physical dimensions of the detector crystal to predict the response of HPGe detectors on the assumption that the detector response can be related to the physical dimensions and that the detector crystal response is independent of the position of the interaction in the crystal. Other investigators 1, have shown that the peak resolution (FWHM, FW.1M and FW.02M) change with position of the incident gamma ray on the front of the detector. Such variability has possible implications for the accuracy of peak shape and area determination, since the calibration is potentially a function of angle of incidence. To demonstrate that the efficiency and resolution vary as a function of energy and gamma-ray point of incidence, measurements have been made on several coaxial detectors of various crystal types and sizes for different energies. The full-energy peaks from 60 kev to 2.6 MeV were used. 2 EXPERIMENTAL SETUP Several detectors were chosen as being representative of the majority of detectors in use today. Large detectors were selected as it was expected that these would show the most variation. However, it is shown that the variations are most significant in older detectors. The detectors were placed in a low-background shield to reduce any contribution from external sources. None of the detectors tested was a low-background type. The detector dimensions are shown in Table 1. The sources used were 241 Am and 60 Co point sources. The 241 Am source was collimated to a 1 mm diameter beam by a 3 cm long by 17 mm diameter lead collimator. The 60 Co source was collimated to a 2 mm diameter beam by a 8 cm long by 8 cm diameter tungsten collimator. In addition, 60 Co data were Detector Number Table 1. Detectors Studied Diameter (mm) Length (mm) N21240A 60.5 60.4 N31626B 53.0 47.8 P41075A 84.5 100.5 P41182A 86.1 112.4 P41181A 81.5 53.7

collected without a collimator. Complete details of the collimators were presented earlier 2. The measurements were made on the top or front of the detector in East-West and North-South modes and on the side of the detector in Up-Down mode as shown in Fig. 1. The actual positions were selected at random with respect to the crystal orientation. The spectra were collected with a Digital Signal Processing MCA (DSPEC Plus), with 12 µs rise time (corresponding to an analog shaping time of about 6 µs), 1 µs flattop and a cusp of 1. The deadtime was always less than 10%. The number of channels was 16k, giving at least 6 channels in the FWHM of the narrowest 59 Figure 1. Definition of Scan Directions kev peak. The peak areas and widths were calculated using the methods described in IEEE 325-1996 (equivalent to International Electrotechnical Commission standard IEC 61976-2000). The position numbers are arbitrary scale readings and are related to the detector crystal position by the count rate data. 3 RESULTS 3.1 Efficiency at 59 kev The first detector studied was an ORTEC GMX type (ntype) with dimensions of 60 mm diameter and 60 mm length. The 241 Am scan on the front is shown in Fig. 2. The crystal diameter is shown by the black bar. Note the uniform sensitivity across the front of the detector, indicating that the crystal has uniform dead layer on the front. The scan down the side of a similar detector is shown in Fig. 3. This shows the 59 kev peak area for the Up-Down scans for 0 degrees and 90 degrees. The intensity of the peak shows the three thick regions in the mounting cup of the detector. More importantly, the two scans are nearly the same. The slight variation between the front of the crystal (left side) and the bottom of the crystal in the reduction in count with position is attributed to the increase in material at the bottom of the mounting cup. The peak area in the regions where the cup is thin, show that the dead layer is uniform along the length of the crystal. Other GMX detectors showed the similar results. The relative dead layer was also measured for larger ORTEC GEM type detectors. Fig. 4 shows the relative intensity of the 59 kev peak for a scan along the length (UD) of the crystal. For comparison, the results of Ref. 3 for a similar, but not identical, detector are shown. Note that the first thick band is visible, but the second band is obscured by the thickening dead layer. In both of these detectors, the dead layer appears to be constant for a portion of the detector (starting at the closed end or front) and then increasing in thickness from some point Figure 2. Peak Area vs Position on Front for 59 kev in NS Direction Figure 3. Peak Count Rate at 59 kev vs Position on Side of Detector Figure 4. Peak Count Rate at 59 kev vs Position on Side of Detector P41075A

to the bottom of the crystal. In contrast to these two detectors, Fig. 5 shows a 59 kev scan of another GEM detector, again of similar length and diameter. In this scan, both the thick bands of the cup can be seen and the dead layer is otherwise nearly uniform from front to rear of the crystal. 3.2 Resolution at 59 kev The peak shape of 241 Am on the front of a GMX (N31626B) is shown in Fig. 6. The crystal diameter is shown by the black bar. Another detector is shown in Fig. Figure 5. Peak Count Rate at 59 kev vs Position on Side of Detector for P41182A 7. Note the uniformity of the peak shape over the front, with the exception of the increase for a region in Fig. 7 and the general broadening near the edges of the crystal. Figure 6. Resolution at 59 kev Across Front of GMX (N31626B). Figure 7. Resolution at 59 kev Across Front of GMX (N21240A). The peak FWHM for the scans down the length of the crystal are shown Figs. 8, 9, and 10. These are again uniform, except for small variations at the ends of the crystals and at some positions down the length. With the exception of Fig. 7, it is expected that the detectors would exhibit this behavior because the low-energy gamma rays interact near the surface. Figure 8. Resolution at 59 kev the Side of GMX (N20140A). Figure 9. Resolution at 59 kev on the Side of GEM (P41182A). Figure 10. Resolution at 59 kev on the Side of GMX (N31626B).

3.3 Resolution at 1.1 and 1.3 MeV The resolution for the 60 Co peaks for scans across the front of a GMX detector is shown in Fig. 12. The GMX detectors do not show the increase in width at the center of the crystal, as seen by previous workers. The broadening of the peak near the center of the crystal is due to the center hole of the detector. This hole extends to about 9 mm of the front surface, giving the possibility of loss of energy into the hole. Figure 11 Resolution at 1332 kev Across Front of GMX (N31626B). The 60 Co scans across the front for the GEM detectors are shown in Figs. 12 and 13. Note that these scans show the same dependance of the FW.1M and FW.04M with front position as seen before. The 1173 and 1332 kev peaks show similar results. This is a general increase in peak width, but is more pronounced at the base of the peak. Figure 12 Resolution at 1332 kev Across Front of GEM (P41075A). Figure 13 Resolution at 1332 kev Across Front of GEM (P41182A). The scan on the side for GEM detectors is shown in Figs. 14 and 15. The shape is very nearly constant over the length except at the bottom of the crystal for Fig. 14, but begins to significantly increase about 75% of the distance from the top in Fig. 15. Figure 14. Resolution at 1332 on Side of GEM (P41075A). Figure 15. Resolution at 1332 on Side of GEM (P41182A). 3.4 Efficiency at 1.1 and 1.3 MeV The sensitivity scan UD of a GEM detector for both 1173 and 1332 kev is shown in Fig. 16. This is the same detector and position as the detector in Fig. 14. The 1173 and 1332 kev UD scan has been plotted UD and DU to show the similarity between the front and rear of the crystal. This plot shows the sensitivity is uniform away from the ends of the crystal.

Fig. 17 shows the same plot for the GEM detector in Fig. 15. The 1173 and 1332 kev UD scan has been plotted UD and DU. Note that the detector is very symmetric top-to-bottom, which is not seen in the 59 kev scan (Fig. 4). Figure 16. Peak Count Rate at 1173 and 1332 kev vs Position on Side of Detector P41182A Figure 17. Peak Count Rate at 1173 and 1332 kev vs Position on Side of Detector P41075A 4 DISCUSSION and CONCLUSION Previous workers 1,3 measured several detectors in preparation for use in MCNP and other programs. These scans showed significant variation in the sensitivity (efficiency) and peak shape with the incident position of the gamma-ray beam. Detectors from several manufacturers were used. In one work, the actual thickness of the dead layer as a function of the position on the detector crystal surface was very important to the measurement geometry. These variations could have a significant impact on the efficacy of the calculations. Thus it is important to measure the variation for more detectors. The peak shape and sensitivity for the detectors studied did not show the wide variations with incident beam position as seen by others. However, similar detectors, in terms of easily measured physical properties (length and diameter), do show this variation. Thus, any model of a detector performance will, of necessity, require a very precise characterization of the exact detector. Future work on other detectors will extend these measurements and compare the MCNP predictions for different assumptions with experimental results. REFERENCES 1. R. J. Gehrke, R. P. Keegan, and P. J. Taylor, Specifications for Today s Coaxial HPGe Detectors, 2001 ANS Annual Meeting, Milwaukee, WI 2. R. M. Keyser, Resolution and Sensitivity as a Function of Energy and Incident Geometry for Germanium Detectors, 2002 IRRMA Meeting, June 2002, Bologna, Italy 3. R. L. Metzger, private communication, see also: R. L. Metzger, K. A. Van Riper, and K. J. Kearfott, Radionuclide Depth Distribution by Collimated Spectroscopy, 2002 ANS Topical Meeting, Santa Fe, NM