Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Similar documents
Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Ed George and Anaïs Viven Varian, Inc.

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Solid Phase Micro Extraction of Flavor Compounds in Beer

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Methanol Extraction of high level soil samples by USEPA Method 8260C

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Optimizing. Abstract: is standardd. procedures. altered to

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Maximizing Sample Throughput In Purge And Trap Analysis

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

EPA TO-17 Volatile Organic Compounds

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Maximizing Production While Minimizing Costs

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

2017 EPA Method Update Rule and EPA Method 624.1

Optimizing Standard Preparation

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Solid Phase Micro Extraction of Flavor Compounds in Beer

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Target Compound Results Summary

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

Determination of Volatile Organic Compounds in Air

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

ORGANIC COMPOUNDS IN WATER BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY USING NITROGEN PURGE GAS

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

Analytical Trap Comparison for USEPA Method 8260C

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

STANDARD OPERATING PROCEDURES

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

Meeting NJ Low Level TO-15 Air Testing Method Requirements

Volatile Organic Compounds in Water PBM

CALA Directory of Laboratories

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

ANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

Building 280A and Building 450 Soil Sampling Results Richmond Field Station Site Berkeley Global Campus at Richmond Bay, Richmond, California

Associates. Project No.: October 25, 2017

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

METHOD MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN WATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY. Revision 4.

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Standard Operating Procedure for the Analysis of Purgeable Organic Compounds (VOC s) in Water by Capillary Column Gas Chromatography/Mass Spectrometry

Method 1624 Revision C Volatile Organic Compounds by Isotope Dilution GCMS

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

OREGON Environmental Laboratory Accreditation Program ORELAP Fields of Accreditation ALS Environmental - Simi Valley

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Thermal Desorption Technical Support

In-Tube Extraction Sample Preparation for Gas Chromatography

McCAMPBELL ANALYTICAL INC Willow Pass Road Pittsburg CA

Copies: Dave Favero, RACER File

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

Study of Residual Solvents in Various Matrices by Static Headspace

METHOD 5030C PURGE-AND-TRAP FOR AQUEOUS SAMPLES

ANALYTICAL REPORT. Job Number: SDG Number: Job Description: Joint Base Cape Cod

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.

Enhanced Preconcentrator for the Analysis of Vapor Phase Volatile Organic Compounds

Headspace Technology for GC and GC/MS: Features, benefits & applications

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

Readiness thru Health

STANDARD OPERATING PROCEDURES

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

/J~ GOLDER. Golder Associates Inc. golder.com. Project No April 25, 2018

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Transcription:

HS-20 & GCMS-QP2010Ultra Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS Introduction Volatile Organic Compounds (VOCs) refer to a group of easily vapourised organic compounds. Under atmospheric pressure, these compounds boil below 250 o C. Studies have shown that prolonged exposure to the VOCs could increase the risk of various health problems, such as cancer. VOCs are commonly present in gasoline, dry cleaning solvents and degreasing agents. Due to improper storage, disposal or spillage of chemicals, these hazardous chemicals could contaminate the drinking water. In order to protect human health and the environment, the U.S. Environmental Protection Agency (EPA) has developed a standard method, namely EPA 524.3 to identify and quantify the purgeable organic compounds in finished drinking water by using purgeand-trap with GC-MS. Headspace with trap mode could serve as an alternative option in analysing the purgeable VOCs. During incubation, VOCs from water samples are effectively partitioned into the headspace. Shimadzu HS-20 headspace sampler is ideal for extraction and concentration of VOCs from water samples. This application news reports a robust analysis method of VOCs complying with the EPA 524.3 criteria based on dynamic headspace coupled with gas chromatography-mass spectrometry. The combination of HS-20 headspace sampler and GCMS-QP2010 Ultra provides an alternative method of choice to extract purgeable organic compounds from drinking water followed by analysis of GC-MS. Experimental Instrument and Analytical Conditions The HS-20 Trap sampler utilizes the headspace technology. A water sample containing VOCs is sealed tightly inside a headspace vial. Under specified conditions of temperature, agitation and time, the VOCs in water sample achieve equilibrium between water phase and the gaseous phase. HS-20 Trap was used as an alternative sample preparation technique for the purge-and-trap. The trap mode HS-20 extracts analytes from the water sample and concentrates on a sorbent trap prior to desorbing to GC-MS. The analytical parameters of both headspace and GC-MS are presented in Table 1. Preparation of Tuning Standard Based on the EPA 524.3, the mass and abundance scales of the mass spectrometer was calibrated in order to meet the ion ratio specification for 4-bromofluorobenzene (BFB). Prior to the analysis, the mass spectrometer was tuned according to Table 1: GC-MS and Headspace sampler conditions HS-20 Mode Oven Temp. Trap Cooling Temp. Trap Equilib Temp. Trap Desorb Temp. Equilibrating Time GC Column Carrier Gas Flow Control Mode Linear Velocity Purge Flow Trap 60 o C 5 o C 25 o C 250 o C 30 min Rtx-VMS 60mx0.25mmx1.4µm He Split Ratio 10 Column Oven Temp MS Ion Source Temp. Interface Temp. Ionization Mode Ionization Voltage Solvent cut Time Acquisition Mode Linear Velocity 31.3 ml/min 3 ml/min 45 o C(4.5min), 12 o C/min 100 o C(0min), 25 o C/min 240 o C (5.32min) 200 o C 200 o C Electron impact (EI) 70 ev 1.5 min SCAN m/z 35-300 the instrument default condition. Approximately 10mL of 0.4 µg/ml BFB was prepared and filled into headspace vials. The BFB solution was introduced into the system by dynamic HS and analysed by GC-MS using the same sample analysis condition (Table 1). Preparation of Primary Dilution Standard Primary dilution standard was the most important as it will be used to prepare the calibration standards and standard for fortification. A series of primary dilution standards, concentration ranged from 2µg/mL to 160µg/mL were

prepared in 2mL microreaction vials with Mininert caps. The EPA 524.3 analytes, internal standards and surrogate standards were commercially available in 2000µg/mL ampules. 20µL of internal standard/surrogate standard mixture was spiked into each primary dilution standard. VOCs standard stock solution was added accordingly as 2, 4, 8, 20, 40, 80 and 160. Finally, methanol was added into the microreaction vials to obtain 2mL of final volume. Preparation of Final Calibration Standard Ascorbic acid and maleic acid with a final concentration of 0.625g/L and 5g/L respectively in reagent water was first prepared. Seven 100mL volumetric flasks were prepared, labelled with respective concentration level and filled with 95mL of reagent water. Subsequently, 25µL of each primary dilution standard was spiked into the respective volumetric flask. Reagent water was filled to reach the gauge line of volumetric flasks. After the analytes and reagent water were mixed homogeneously, 10mL of the sample was taken into headspace vials. Immediately, the headspace vials were crimped tightly to minimize loss of VOCs. Preparation of Field Samples Tap water was used as the field sample. In preparing field samples, water was collected directly from the tap and dechlorinated with ascorbic acid at ph 2.00. 0.625g of ascorbic acid and 5g of maleic acid were added into a 1L volumetric flask. 400mL of tap water was added to volumetric flask to dissolve the solid preservatives. After that, tap water was then used to fill the volumetric flask to the gauge line. Results and Discussion Initial Mass Spectrometry Tune The MS was tuned based on the default parameter. BFB was analysed under the same analytical condition. The system performance was evaluated based on a single spectrum at the apex of the BFB peak with subtraction of the background. Shimadzu GCMSsolution QA/QC function is able to check the BFB tuning results with respect to the EPA 524 method. Figure 1 shows the result of the BFB evaluation using the QA/QC function. All BFB mass spectrum criteria must be achieved before carrying out further analysis. If the MS tune results does not meet the required mass intensity criteria, the MS should be retuned by changing the tuning conditions and repeated with BFB analysis. Chromatogram This revised EPA 524.3 has included some new emerging contaminants, gasoline additives and potential breakdown of MtBE which gives a total of 76 target analytes, 3 internal standards and 3 surrogate standards. Most of the compounds have distinctive mass fragmentation. By matching the retention time, target ions and reference ions, all compounds could be identified. All the target compounds were well separated based on the respective mass chromatogram (MC), except m-xylene and p-xylene. Due to the same elution time and mass fragmentation, these two compounds were integrated together. Figure 2 showed the standard Total Ion Chromatogram (TIC) of all the compounds, including targets, internal standards and surrogate standards. Figure 1: Evaluation results of BFB analysis for the Spectrum Check

Figure 2: TIC chromatogram of VOCs based on EPA method 524.3 and analysis at 20 µg/l Calibration All the target analytes were calibrated by using internal standard technique. Seven calibration standards, ranging from 0.5µg/L to 40µg/L were prepared and analysed. Each calibration point is treated as an unknown. Concentration of the calibration point is calculated based on the calibration curve. According to the EPA 524.3, either linear or quadratic regression is permitted as long as the initial calibration point is within ±50% of its true value and all other points are within ±30% of their true values. An example of Benzene quadratic calibration curve is illustrated in Figure 3. Correlation coefficient for all the compounds were more than 0.995. Some of the compounds gave R 2 -value as high as 1.000. Recovery of each calibration point and correlation regression are tabulated in Table 2. Detection Limit Method detection limit is a statistical method used to determine reported value that is greater than zero with 99% confidence level. The MDL of each analyte was calculated based on the equation below. Seven replicated control samples were prepared at 0.1µg/L. The Shimadzu GCMSsolution QA/QC function will automatically generate the calculation of MDL. As shown in Table 2, target analytes MDL is lower than the first calibration point, except t-butyl alcohol which commonly gives poor MS response. Precision and Accuracy Area Ratio 3.0 2.5 2.0 1.5 1.0 Benzene R 2 = 0.9996 In the demonstration of system stability, seven replicate samples in reagent water were prepared at 5µg/L which is the midrange of the initial calibration curve. Percentage of relative standard deviation (RSD) was calculated based on the equation below to determine the system precision. System accuracy is calculated based on the percentage of recovery using the same seven replicate samples. 0.5 0.0 0.0 2.5 5.0 Conc. Ratio Figure 3: Calibration curve of Benzene from 0.5µg/L to 40µg/L These seven replicate samples concentration must not deviate more than 20% from the true concentration with %RSD less than 20%. All analytes %RSD and %Recovery are found to be within the allowable limit. Chromatogram of chlorodifluoromethane from seven replicate samples were overlaid in Figure 4. 90% of this highly volatile analyte were recovered with deviation of 2.9%.

Table 2: Summary of the response of all the target analytes which showed the correlation coefficient, detection limit and recovery at each calibration level ranged from 0.5 to 40 µg/l. Internal Standards and Surrogate Standards were omitted for the ease of data comparison. ID Compound Correlative Regression (R 2 ) Recovery at each calibration level (%) 0.5 µg/l 1 µg/l 2 µg/l 5 µg/l 10 µg/l 20 µg/l 40 µg/l Method Detection Limit (MDL, µg/l) 1 Dichlorodifluoromethane 0.9988 148 131 119 73 101 102 100 0.100 2 Chlorodifluoromethane 0.9999 118 104 107 92 101 100 100 0.163 3 Chloromethane 0.9996 132 114 104 95 94 103 100 0.113 4 Vinyl Chloride 0.9995 144 121 105 87 97 103 100 0.092 5 1,3-Butadiene 0.9991 150 123 111 77 102 102 100 0.181 6 Bromomethane 0.9999 134 97 103 96 99 101 100 0.179 7 Trichlorofluoromethane 0.9993 134 127 108 80 100 102 100 0.100 8 Diethyl ether 0.9998 114 118 95 91 104 100 100 0.039 9 1,1-Dichloroethene 0.9995 118 123 106 82 106 100 100 0.024 10 Carbon Disulfide 0.9996 112 115 107 85 105 100 100 0.065 11 Methyl Iodide 0.9998 124 114 102 88 103 100 100 0.047 12 Allyl Chloride 0.9996 134 113 98 86 106 99 100 0.028 13 Methylene Chloride 0.9998 110 110 95 93 105 99 100 0.241 14 trans-1,2-dichloroethene 0.9996 108 108 102 88 107 99 100 0.055 15 Methyl Acetate 0.9993 142 124 110 81 101 102 100 0.285 16 Methyl-t-Butyl Ether (MtBE) 0.9999 92 101 102 97 103 99 100 0.033 17 t-butyl Alcohol (TBA) 0.9998 114 75 87 110 102 99 100 0.881 18 Diisopropyl Ether (DIPE) 0.9999 114 99 104 95 102 100 100 0.026 19 1,1- dichloroethane 0.9998 118 105 103 89 105 99 100 0.071 20 t-butyl Ethyl Ether (ETBE) 0.9998 100 96 101 94 106 98 100 0.026 21 cis-1,2-dichloroethene 0.9997 88 97 99 95 107 98 100 0.060 22 Bromochloromethane 1.0000 94 92 102 99 103 99 100 0.037 23 Chloroform 0.9997 86 91 97 98 107 98 100 0.387 24 Carbon Tetrachloride 0.9997 132 114 112 85 99 102 100 0.035 25 Tetrahydrofuran 0.9978 108 97 66 95 119 94 100 0.121 26 1,1,1-Trichloroethene 0.9998 130 110 108 89 100 101 100 0.019 27 1,1-Dichloropropene 0.9996 118 117 102 85 106 99 100 0.039 28 1-Chlorobutane 0.9997 110 108 107 87 105 99 100 0.033 29 Benzene 0.9998 130 117 98 89 103 100 100 0.023 30 t-amyl Methyl Ether (TAME) 0.9999 122 96 106 94 101 100 100 0.038 31 1,2-Dichloroethane 0.9998 86 90 98 100 105 98 100 0.041 32 Trichloroethene 0.9999 98 105 104 92 104 99 100 0.048 33 t-amyl Ethyl Ether (TAEE) 0.9998 104 98 102 92 106 99 100 0.021 34 Dibromomethane 0.9999 98 100 103 94 104 99 100 0.021 35 1,2-Dichloropropane 0.9998 114 123 103 89 100 101 100 0.035 36 Bromodichloromethane 0.9998 102 102 100 93 105 99 100 0.060 37 cis-1,3-dichloropropene 0.9997 92 90 101 96 107 98 100 0.035 38 Toluene 0.9996 104 106 106 88 106 99 100 0.063 39 trans-1,3-dichloropropene 0.9995 80 104 105 90 108 98 100 0.048 40 Tetrachloroethene 0.9995 118 110 110 84 105 100 100 0.047 41 Ethyl Methacrylate 0.9995 94 107 100 90 108 98 100 0.043 42 1,1,2-Trichloropropane 0.9997 100 95 98 96 107 98 100 0.270 43 Dibromochloromethane 0.9998 64 104 108 95 105 99 100 0.054 44 1,3-Dichloropropane 0.9995 100 100 98 92 109 98 100 0.070 45 1,2-Dibromoethane 0.9992 136 94 107 84 108 98 100 0.054 46 Ethylbenzene 0.9998 120 103 105 90 103 100 100 0.173 47 Chlorobenzene 0.9996 108 102 106 88 106 99 100 0.062 48 1,1,1,2-Tetrachloroethane 0.9999 128 97 106 94 100 101 100 0.077 49 m-xylene, p-xylene 0.9992 150 115 101 87 99 103 86 0.249 50 o-xylene 0.9999 94 96 104 96 104 99 100 0.110 51 Styrene 0.9997 112 103 102 91 105 99 100 0.084 52 Bromoform 0.9993 66 71 108 99 109 96 100 0.101 53 Isopropylbenzene 0.9995 138 114 110 85 100 102 100 0.043 54 n-propylbenzene 0.9989 150 133 105 82 97 104 99 0.039 55 1,1,2,2-Tetrachloroethane 0.9999 106 96 102 96 103 99 100 0.196 56 Bromobenzene 0.9995 96 110 98 88 109 98 100 0.182 57 1,3,5-Trimethylbenzene 0.9997 142 119 102 85 102 101 100 0.033 58 2-Chlorotoluene 0.9998 124 104 100 90 105 99 100 0.024 59 1,2,3-Trichloropropane 0.9995 128 76 93 114 93 101 100 0.303 60 4-Chlorotoluene 0.9998 118 111 100 89 104 99 100 0.062 61 t-butylbenzene 0.9998 110 112 99 89 106 99 100 0.030 62 1,2,4-Trimethylbenzene 0.9998 138 120 105 86 100 101 100 0.058 63 Pentachloroethane 0.9998 114 103 113 88 103 100 100 0.067 64 sec-butylbenzene 0.9988 148 133 109 80 98 104 99 0.032 65 4-Isopropyltoluene 0.9995 150 127 104 83 100 102 100 0.061

Peak Area Application ID Compound Correlative Regression (R 2 ) Recovery at each calibration level (%) 0.5 µg/l 1 µg/l 2 µg/l 5 µg/l 10 µg/l 20 µg/l 40 µg/l Method Detection Limit (MDL, µg/l) 66 1,3-Dichlorobenzene 0.9999 108 109 101 93 103 100 100 0.147 67 1,4-Dichlorobenzene 0.9998 112 104 94 94 105 99 100 0.059 68 n-butylbenzene 0.9994 150 125 103 84 99 102 100 0.034 69 Hexachloroethane 0.9997 106 123 95 87 107 99 100 0.107 70 1,2-Dichlorobenzene 1.0000 84 102 103 97 103 99 100 0.065 71 1,2-Dibromo-3-chlropropane 0.9982 112 95 100 78 118 96 100 0.379 72 Hexachlorobutadiene 0.9998 124 116 102 86 105 100 100 0.081 73 1,2,4-Trichlorobenzene 0.9997 138 106 93 89 107 99 100 0.169 74 Naphthalene 0.9998 108 105 97 92 106 99 100 0.122 75 1,2,3-Trichlorobenzene 0.9992 86 106 89 93 112 97 100 0.136 250000 225000 200000 175000 150000 125000 100000 75000 50000 Internal Standards EPA Method 524.3 is quantitated based on the Internal Standards. Internal Standards were spiked into all calibration standards, CCCs and field samples. Stability of internal standards is crucial to obtain precise response factor of the analytes which will be used to plot the calibration curve. Internal standard peak areas of all analyses should not deviate more than 50% for the lowest calibration level and 30% of other calibration levels from the most recent CCC. Figure 5 shows the stability of internal standards for the analysis of 10 field samples. 25000 3.9 4.0 Figure 4: Recovery of Chlorodifluoromethane which is the most volatile compounds of the analysis gave 90% recovery with RSD of only 2.9%. Quality Control The system stability and accuracy of existing calibration were monitored through the Continuing Calibration Checks (CCC). CCC is a calibration standard which contains the method analytes, internal standards and surrogate analytes. CCC was analysed at the beginning of each Analysis Batch, after every tenth field sample and at the end of each Analysis Batch. Each CCC contained 5µg/L of method analytes, internal standards and surrogate analytes. Each analyte was calculated to be within 3.5µg/L and 6.5µg/L, which is equivalent to ±30% of true concentration. Throughout the analysis, all the target analytes in CCC detected were within the CCCs quality control criterion. Table 3 shows the recovery of the most volatile target, Chlorodifluoromethane in the CCCs analyses. Table 3: Recovery of Chlorodifluoromethane which is the most volatile compound of the analysis in the Continuing Calibration Checks was within the QC check criteria. Injection Recovery Requirement (%) Actual Recovery (%) 1 70-130 74 2 70-130 93 3 70-130 98 4 70-130 89 5 70-130 87 500,000 Internal Standard 1 1,4-difluorobenzene Internal Standard 2 Chlorobenene-d 5 Internal Standard 3 1,4-dichlorobenzene-d 4 0 Figure 5: 0.5 Peak areas of Internal Standards in all the field samples are within ±30% of peak areas (highlighted region) of internal standard in CCCs. Surrogate Recovery In order to evaluate the matrix interference on sample, surrogates are spiked into the field samples prior to analyses. Surrogates are compounds which are chemically similar to the target analytes and thus behave similarly in the matrix. In the EPA Method 524.3, methyl-t-butyl ether-d 3, 4- bromofluorobenzene, and 1,2-dichlorobenzne-d 4 were fortified as Surrogate Standards for all analyses including calibration standards, CCCs and field samples. The surrogate standards were closely monitored to confirm that the recovery percentage of samples were within the range of 70% to 130%. Figure 6 shows the recovery of all surrogates in 10 field samples analyses.

Allowable Surrogate Recovery Application 130% 120% 110% 100% 90% 80% methyl-t-butyl ether-d3 4-bromofluorobenzene 1,2-dichlorobenzene 70% 1 2 3 4 5 6 7 8 9 10 No of Field Samples Figure 6: Recovery of all surrogates, namely methyl-t-butyl ether-d 3 (Surrogate 1), 4-bromofluorobenzene(Surrogate 2), and 1,2-dichlorobenzene-d 4 (Surrogate 3) of all 10 CCCs are within the acceptable QC criteria (70 130%). Conclusions The results shows that the HS-20 Trap mode is a viable alternative method in analysis of VOCs following EPA 524.3. The correlation coefficient, recovery and RSD obtained in this study complies with the EPA Method 524.3. Matrix effect of the real samples is insignificant as recoveries of the internal standards and surrogate standards are within the limits of the standard method. In brief, the system of Shimadzu HS-20 Trap sampler coupled with GCMS-QP2010 Ultra provide another choice for analysis of VOCs in drinking and underground water samples. Reference B.Prakash, A.D. Zaffiro, M. Zimmerman, D. J. Munch and B.V. Pepich, Method 524.3 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography/Mass Spectrometry, EPA Document # EPA 815-B-009, Version 1.0, June 2009 SHIMADZU (Asia Pacific) Pte. Ltd 79 Science Park Drive, #02-01/08, Cintech IV, Singapore 118264 www.shimadzu.com.sg Tel: +65-6778 6280 Fax: +65-6778 2050 Copyright 2014 SHIMADZU (Asia Pacific) Pte. Ltd. All rights reserved. No part of this document may be reproduced in any form or by any means without permission in writing from SHIMADZU (Asia Pacific) Pte. Ltd.