Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX)

Similar documents
Determination of Alkaloids in Goldenseal Using Agilent Bond Elut Plexa Solid Phase Extraction Sorbent for Cleanup and HPLC-DAD Analysis

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

The Effect of Contact Angle and Wetting on Performance in Solid Phase Extraction

Yuhui Yang, An Trinh, Michael Ye, and Tom Henderson 595 North Harrison Road, Bellefonte, PA T HKB

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

A New Polymer-based SPE Sorbent to Reduce Matrix Effects in Bioanalytical LC-MS/MS

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

DEVELOPMENT AND VALIDATION OF A HPLC METHOD FOR IN-VIVO STUDY OF DICLOFENAC POTASSIUM

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

with diode array detection

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

EVOLUTE WCX Columns for Solid Phase Extraction of Quaternary Amines and Strongly Basic Compounds from Aqueous Samples

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Application Note. Agilent Application Solution Analysis of acaricides in honey. Authors. Abstract. Food

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Application. Authors. Abstract. Introduction. Hydrocarbon Processing

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

HPLC Determination of Antidepressants and their Metabolites in Plasma

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Detection of 9-tetrahydrocannabinol ( 9-THC) in human urine by Solid Phase Extraction and HPLC.

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Department of Quality Assurance, Luqman College of Pharmacy, GULBARGA (K.S.) INDIA ABSTRACT

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

7. Stability indicating analytical method development and validation of Ramipril and Amlodipine in capsule dosage form by HPLC.

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Determination of 24 PAHs in Drinking Water

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

[ Care and Use Manual ]

Solid Phase Extraction Method Development Tips and Tricks

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

Impact factor: 3.958/ICV: 4.10 ISSN:

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC

Chemical Analysis Problem

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Available online Journal of Chemical and Pharmaceutical Research, 2012, 4(6): Research Article

Determination of trace anions in concentrated hydrofluoric acid

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

International Journal of Pharmacy and Pharmaceutical Sciences Vol 2, Issue 1, 2010

Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Electronic Supplementary Information. The selective response of templated polymer for the cationic drug

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Supporting Information

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Quantitative analysis of small molecules in biological samples. Jeevan Prasain, Ph.D. Department of Pharmacology & Toxicology, UAB.

CHAPTER V ANALYTICAL METHODS ESTIMATION OF DICLOFENAC. Diclofenac (gift sample from M/s Micro Labs Ltd., Pondicherry)

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Analysis of Saponins in Spine Date Seed using Agilent ZORBAX Solvent Saver HT (1.8 µm) Columns

Rapid quantification of Chinese medicine Zuo Jin Pill using rapid resolution liquid chromatography

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Development of analytical methods for Catecholamines, Metanephrines and related metabolites in urine and plasma matrices

Pharmacophore 2011, Vol. 2 (4), ISSN Pharmacophore. (An International Research Journal)

Abstract. Experimental Sample Preparation

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

Characterization of Carbon SPE for the Extraction of Polar Analytes

Phenobarbital analysis in biological matrix (blood) by high performance liquid chromatography (HPLC)

Modernizing the USP Monograph for Acetaminophen

Novel LC/MS-Compatible Stationary Phase with Polar Selectivity

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Shodex TM ODP2 HP series columns

Journal of Pharmaceutical and Biomedical Analysis Letters. Analysis Letters

Analysis - HPLC A.136. Primesep 5 µm columns B.136

Transcription:

Fractionation of Acidic, Basic, and Neutral Drugs from Plasma with an SPE Mixed Mode Strong Cation Exchange Polymeric Resin (Agilent SampliQ SCX) Application Note Forensic Toxicology Authors Bellah. Pule, Lesego C. Mmualefe, and Nelson Torto Department of Chemistry, Rhodes South Africa Abstract A method for the simultaneous extraction of drugs (amphetamine, acetaminophen, p-toluamide, m-toluidine, and phenobarbital) from spiked human plasma sample was developed. This procedure employed solid phase extraction with a mixed mode strong cation exchange resin, Agilent SampliQ SCX. The chromatographic separation and analysis of solid phase extraction extracts were achieved using 3% methanol and 7% potassium dihydrogen phosphate as a mobile phase under isocratic conditions on an Agilent ZRBAX Eclipse Plus C8 4.6 mm 75 mm, 3.5 µm column at ml/min flow rate and a diode array detector (DAD) set at 2 nm. High and reproducible recoveries (> 8%) for all the analytes were obtained. The limits of detection (LD) and quantification (LQ) were.39 and.7 µg/ml for acetaminophen,.84 and.87 µg/ml for amphetamine,.36 and.6 µg/ml for m-toluidine,.66 and.7 µg/ml for p-toluamide, as well as.8 and.89 µg/ml for phenobarbital, respectively.

Introduction Sample preparation prior to chromatographic analysis presents a major challenge for the determination of drugs and their metabolites in complex matrices, such as biological fluids (for example, blood, plasma, urine, serum, saliva). Drugs normally exhibit a diverse polarity with acidic, basic, or neutral functionalities depending on the ph of the matrix. Liquid-liquid extraction (LLE) and solid phase extraction (SPE) have traditionally been employed for the extraction of drugs, their metabolites, and endogenous compounds from plasma [-2]. Their quantification at a low concentration has proven to be a difficulty in forensic toxicology analyses. SPE is, however, the preferred method as it is not only employed for class fractionation, but it is also for trace enrichment and purification. Available commercial sorbents, such as chemically modified silica gel, polymer, and graphitized or porous carbon, are used [3]. These offer separations based on normal phase, reversed phase, ion exchange, and mixed mode ion exchange (combination of reversed phase and ion exchange) sorbents. The mixed mode sorbents have proven to give cleaner extracts and better separations than standard reversed phase or ion exchange sorbents as they take advantage of both the ion exchange and hydrophobic interactions [4,5]. For the extraction of compounds with a wide variety of polarity, polymeric sorbents have proven to be superior to other sorbents (for example, alkylated silica) and are, therefore, the choice of sorbent for this study [6 8]. In this application note, a method employing solid phase extraction was developed for the fractionation of acidic, basic, and neutral drugs from plasma with Agilent SampliQ SCX, a mixed mode strong cation exchange polymer. The resin is a sulfonic acid modified divinyl benzene polymer that exhibits both cation exchange and reversed phase behavior. In addition, it provides excellent reproducibility and enables a simple extraction protocol. Specific drugs (Figure ) have been used as representatives of the three classes of drugs, for example, p-toluamide (acidic), amphetamine (basic), and acetaminophen (neutral). Experimental Chemicals Acetaminophen, phenobarbital, p-toluamide, amphetamine, m- toluidine, and ranitidine (IS) were purchased from Sigma- Aldrich Chemicals (St. Louis, M, USA). Phosphoric acid, formic acid, and potassium hydroxide were purchased from Merck Chemicals (Gauteng, South Africa) while the HPLC grade methanol was from Merck KGaA (Darmstadt, Germany), H NH 2 Amphetamine _ N H p-toluamide N + NH NH m-toluidine Ranitidine (IS) H 2 N dipotassium hydrogen phosphate and potassium dihydrogen phosphate were purchased from Saarchem Analytical (Krugersdorp, South Africa). The mobile phase was prepared with ultrapure water (8.2 M Wcm) from a MilliQ system by Millipore (Milford, Mass, USA) and filtered through a Whatman membrane filter (47 mm diameter and.2 µm pore size). The stock solutions of the drugs and the internal standard were prepared in methanol (, µg/ml) and kept at 4 C while the working solutions were prepared daily by diluting the stock solutions, to appropriate concentrations, also in methanol. The plasma (ECZ HQ Donation 549778, + ) was from SANBS (Port Elizabeth, South Africa). Instrumental The analysis was performed on an Agilent 2 Series LC composed of a binary pump and a DAD set at l = 2 nm. Separation of the compounds was achieved on an Agilent ZRBAX Eclipse Plus C8 column (4.6 mm 75 mm, 3.5 µm, p/n 959933-92, Agilent Technologies, Santa Clara, CA). The data was processed by Agilent LC 2D ChemStation software. The SPE cartridges were Agilent SampliQ SCX, ml/3 mg, p/n 5982-323, a polymeric strong cation exchanger with 25 35 µm average particle sizes. A Jenway 35 ph meter (London, UK) was employed for ph adjustments. NH 2 Acetaminophen Figure. Structures of drugs used in present study. S H N NH Phenobarbital N 2

Sample pretreatment: SPE procedure The plasma sample ( ml) was hydrolyzed with % formic acid (3 ml) for 3 minutes. The sample, spiked with drugs, was then loaded onto the SampliQ SCX cartridges as described in Figure 2. An internal standard, 5 µl of the ranitidine stock solution, was added to each SPE fraction. A blank plasma sample ( ml) was carried through the procedure also. Separation and analysis Table shows the reversed-phase chromatographic conditions. All drugs were separated within 5 minutes. Table. HPLC conditions. Column: Flow rate: Column temperature: 35 C Injection volume: 5 µl Mobile phase: Agilent ZRBAX Eclipse Plus C8 4.6 mm 75 mm, 3.5 µm.5 ml/min Isocratic elution: 3/7 A/B A: CH 3 H B: 25 mm K 2 HP 4, ph 7 Condition: ml CH 3 H Equilibrate: ml H 2 Load: ml spiked plasma Wash: ml 2% HCH Elute : ml CH 3 H (acidic and neutral) Elute 2: ml 5% NH 4 H/CH 3 H (basic analytes) Elution Add 5 µl of IS Elution 2 Add 5 µl of IS Analyze Fraction using LC-UV Figure 2. SPE procedure. 3

Results and Discussion Separation The standard mixture of the analytes was separated with the set chromatographic conditions (Table ), and the chromatogram is reported in Figure 3. Figure 3. Chromatogram of a standard mixture (7 µg/ml): () acetaminophen, (2) amphetamine, (3) p-toluamide, (4) m-toluidine, and (5) phenobarbital ranitidine (IS). Analysis of standard solutions Calibration curves were processed at µg/ml concentration ranges for all the analytes. They were linear with coefficient of regression (R 2 ) greater than.999 as shown in Figure 4. 4

Peak area ratio.8.7.6.5.4.3 Acetaminophen calibration curve y =.692x R 2 =.9993 Peak area ratio.8.7.6.5.4.3 Amphetamine calibration curve y =.699x R 2 =.9994.2.2...4.2 2 4 6 8 2 Conc (µg/ml) p-toluamide calibration curve y =.45x R 2 =.9992 2.5 2 2 4 6 8 2 Conc (µg/ml) m-toluidine calibration curve y =.842x R 2 =.999 Peak area ratio.8.6 Peak area ratio.5.4.2.5 Peak area ratio 2 4 6 8 2.2.8.6.4.2 y =.72x R 2 =.9995 Conc (µg/ml) Phenobarbital calibration curve 2 4 6 8 2 Conc (µg/ml) 2 4 6 8 2 Conc (µg/ml) Figure 4. Calibration curves of the standards. SPE procedure A mixed mode strong cation exchange sorbent was used to extract basic drugs from the acidic and neutral drugs in plasma. The acidic and the neutral drugs, which exhibit a similar retention mechanism to that of the undissociated acidic compounds, were adsorbed in the hydrophobic portion of the sorbent and eluted in the neutral fraction while the basic drugs were retained by the cation exchange interactions with the sorbent and eluted in the ammoniated fraction. First, blank urine containing no drugs was treated using the SPE method. Figure 5A, for the acidic and neutral elution conditions (Elution ), showed nothing eluting in the region of the acidic and neutral drugs in the standards. Figure 5B, which depicts a blank urine using the basic elution conditions (Elution 2), also showed nothing eluting in the region of the acidic drugs. For the spiked urine samples, the neutral and acidic drugs were eluted in the neutral fraction (Figure 6A, Elution ) because they were retained through hydrophobic interactions while the basic drugs, retained by the strong anion exchange functionalities of the sorbent, eluted separately in the basic fraction as shown in Figure 6B, Elution 2. A small amount (< %) of the neutral/acidic drugs were also found in the basic fraction. A larger volume of methanol in the prior step could have been used to improve extraction efficiency. 5

6 5 A Plasma blank Elution 4 mau 3 2 2 3 4 5 6 Time (min) Figure 5A. Chromatograms of blank plasma sample from Elution. 9.5 8.5 7.5 6.5 5.5 B Plasma blank Elution 2 mau 4.5 3.5 2.5.5.5 _.5 2 3 4 5 6 Time (min) Figure 5B. Chromatograms of blank plasma sample from Elution 2. 6

9 8 A Elution : acidic and neutral drugs 7 Acetaminophen 6 m AU 5 4 Ranitidine (IS) p-toluamide 3 2 Phenobarbital 2 3 4 5 6 Time (min) Figure 6A. Chromatograms of drugs in plasma, elution :. acetaminophen, 2. ranitidine (IS), 3. p-toluamide, and 4. phenobarbital. Elution 2: basic drugs 4 3.5 3 B Amphetamine Ranitidine (IS) m-toluidine 2.5 mau 2.5.5 2 3 4 5 6 Time (min) Figure 6B. Chromatograms of drugs in plasma, elution 2:. amphetamine, 2. ranitidine (IS), and 3. m-toluidine. Linearity, limits of detection, and limits of quantification The blank plasma was spiked with the analytes at five different concentrations and subjected to the SPE procedure in Figure 2. The internal standard, 5 µl, was added. Each spiked plasma sample was prepared in triplicate. The analyte/is peak area ratios were plotted against the corresponding concentrations. All the analytes were linear in the chosen concentration range ( 8 µg/ml) with R 2 >.999 as shown in Table 2. Table 2. Linearity of the method employing SPE. Concentration range ( 8 µg/ml) Linear equation Correlation Drugs equation (y) coefficient (R 2 ) Acetaminophen.99x.9999 Amphetamine.853x.999 p-toluidine.38x.999 m-toluamide.84x.999 Phenobarbital.526x.999 7

Equations and 2 were used to calculate LD and LQ, where Syx is the standard error of the regression line and b is the gradient. Equation Equation 2 LD = 3.3 Syx () b LQ =. Syx (2) b Table 3 shows the LD and LQ determined for each analyte. Table 3. LD and LQ for the analytes. Drug LD (µg/ml) LQ (µg/ml) Acetaminophen.39.85 Amphetamine.7.87 p-toluamide.66.7 m-toluidine.35.6 Phenobarbital.82.89 Recovery and reproducibility studies Spiked plasma samples at three concentration levels:.5, 2.5, and 5 µg/ml, corresponding to the lower, middle, and upper limit of the linearity curve, were subjected to SPE cleanup. The analyte chromatographic peak areas obtained were compared to those obtained from standard solutions at the same concentration, and the percentage extraction yield was calculated. To demonstrate reproducibility, the samples were analyzed at the three mentioned concentration levels (n = 6). Good recoveries were obtained as indicated in Table 4. Conclusions The SPE procedure was successfully carried out on Agilent SampliQ SCX sorbents for the simultaneous extraction of acidic, basic, and neutral drugs from plasma with high and reproducible recoveries (> 8%). The LD ranged from.39 to.84 µg/ml for the drugs studied. The LQ ranged from.7 to.89 µg/ml. This method can be applied to compounds that exhibit a diverse polarity and acidic, basic, or neutral functionalities. Table 4. Recoveries for the drugs in the study. Spike level (µg/ml).5 2.5 5 Drugs Class Recovery %RSD Recovery %RSD Recovery %RSD Acetaminophen Neutral 8.45.4 95.8.48 92.62.22 Amphetamine Basic 83.6.62 88.83.9 86.64.28 p-toluamide Acidic 85.88.65 96.86.8 94.4.7 m-toluidine Basic 8.97.5 89.6.33 94.86. Phenobarbital Acidic 85.49.39 9.3.25 8.48.2 8

References. A. Zwir-Ference, M. Bizziuk, J. Env 5 (26) 677-69. 2. K. B. Borges, Talanta 78 (29) 233-24. 3. K. A. Shaikh, S. D. Patil, A. B. Devkhile, J. Pharm. Biomed. Anal 48 (28) 48-484. 4. M. S. Landis J. Pharm. Biomed. Anal. 44 (27) 29-39. 5. L. Mercolini, G. Gerra, M Consorti, L. Somaini, M. A. Raggi, Talanta 78 (29) 5-55. 6. M. A. Allabdalla, J. Clinical Forensic Medicine. 2 (25) 3-35. 7. N. H Yu, E. N. M Ho, F. P. W Tang, T. S. M Wan, A. S. Y. Wong, J. Chromatogr. A 89 (28) 426-434. 8. S. Weigel, R Kallenborn, H. Huhnerfuss J. Chromatogr. A 23 (24) 83-95. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. 9