The Determination of Tertiary Dodecyl Mercaptan by Low Thermal Mass Gas Chromatography Dual Plasma Sulfur Chemiluminescence Detection

Similar documents
Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Analysis of Sulfur-Containing Flavor Compounds By GC/MS With A PFPD

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Application Note. Abstract. Authors. Introduction

Abstract. Introduction

Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Capillary Flow Technology with Multi-Dimensional Gas Chromatography for Trace Analysis of Oxygenated Compounds in Complex Hydrocarbon Matrices

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Continuous Improvement in Petroleum/Petrochemical Analysis HP s Family of Innovative PLOT Columns

Introduction to Capillary GC

Author. Abstract. Introduction

GAS CHROMATOGRAPHY (GC)

Understanding Gas Chromatography

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

CHAPTER 6 GAS CHROMATOGRAPHY

New ZB-5HT Inferno The World s Highest Temperature Non-Metal GC Column

Deposition and Detection System. Trace Level Analysis of Chemical Warfare Agents (CWAs) Related Chemicals by GC-IR Solid Phase System

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Application Note S/SL

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Contents. Efficient synthesis of 5-(chloromethyl)furfural (CMF) from high fructose corn syrup (HFCS) using continuous flow processing

Introduction to Capillary GC

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Selective Formation of Benzo[c]cinnoline by Photocatalytic Reduction of 2,2 Dinitrobiphenyl with TiO 2 and UV light irradiation

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Study of Residual Solvents in Various Matrices by Static Headspace

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

STANDARD OPERATING PROCEDURES

Fast and Accurate Analysis of PBDEs in a Single Run, Including

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Residual solvents analysis using an Agilent Intuvo 9000 GC system

AppNote 6/2004. Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons SUMMARY

Analysis of Pressure Sensitive Adhesives by GC/MS and GC/AED with Temperature Programmable Pyrolyzer

U.S. EPA Method 8270 for multicomponent analyte determination

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Application Note 032

Virtually Particle-Free Rt -Silica BOND Columns

Industrial Applications with a New Polyethylene Glycol-Based GC Column

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

Hydrogen as a Carrier Gas in the Analysis of Polar/Non-Polar Compounds Using the Polyarc System. Application Note. Author. Introduction.

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

Sulfotepp impurities in Chlorpyrifos EC formulations

Pittcon P

Gas Chromatography. Introduction

DETERMINATION OF THE TOTAL SULPHUR CONTENT IN GAS OILS WITH A FLAME PHOTOMETRIC DETECTOR

Activity in the FID Detection Port: A Big Problem if Underestimated

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Standard Test Method for Determination of Trace Thiophene in Refined Benzene by Gas Chromatography 1

Fast and Accurate Analysis of PBDEs in a Single Run, Including BDE-209

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Principles of Gas- Chromatography (GC)

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

Monomer Analysis. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Chromatography. Gas Chromatography

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

2401 Gas (liquid) Chromatography

Muliple Gas#3 plus Sulfur GC Configuration

Determination of the Boiling-Point Distribution by Simulated Distillation from n-pentane through n-tetratetracontane in 70 to 80 Seconds

Spring 2010 updated March 24 Determination of Aromatics in Gasoline by Gas Chromatography- Mass Spectrometry: Comparison of Grades and Brands

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Gas Chromatography. Chromatography Laboratory Course. Dr. Christian Jungnickel Chromatography Course GC September 2005

Application. Gas Chromatography February Introduction

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

APPLICATION NOTE. A Capillary Approach to ASTM D3606: Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline

The Need for Multidimensionality in Analytical Thinking. Chromatographic Developments as an Example

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Maximizing Sample Throughput In Purge And Trap Analysis

Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD)

JWH-018 and JWH-022 as Combustion Products of AM2201

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Method for the determination of 1,3-butadiene

Detection of Chemical Warfare Agents by Transportable GC/MS

Volatile organic compounds (VOCs):

Transcription:

The Determination of Tertiary Dodecyl Mercaptan by Low Thermal Mass Gas Chromatography Dual Plasma Sulfur Chemiluminescence Detection Ronda Gras 1, Jim Luong 1, *, Patric Eckerle 2, and Randy Shearer 3 1 Dow Chemical Canada, PO BAG 16, Highway 15, Fort Saskatchewan, Alberta, Canada T8L 2P4; 2 Dow Chemical Germany, Werk Rheinmuenster, Industriestrasse 1, Dow Deutschland GMBH & Co., Germany; 3 Rentech Inc, 12401 E. 37th Avenue, Denver, CO 80239 Abstract Tertiary dodecyl mercaptan (TDM) is commonly used as a chain transfer agent in the manufacturing process of styrene/butadiene latex for use in carpet and paper industries. A gas chromatographic technique has been successfully developed for the measurement of TDM based on its sulfur content for material identification, trend analysis, or for the monitoring of un-reacted residual material in final products. The method employs low thermal mass gas chromatography (LTM-GC) and a dual-plasma sulfur chemiluminescence detector (DP-SCD) to attain a high degree of sensitivity and selectivity. Using the technique described, a detection limit in the range of 0.5 ppm (v/v) TDM and less than 1 min analysis time can be achieved. Response is linear over four orders of magnitude with high degree of repeatability of less than 5% RSD. tetramer structure, and hence the location of the C=C bond, the thiol group can be located in many different positions, resulting in a product mixture of isomers with an average boiling point range around 230 C. Recently, there has also been some increased concerns regarding to the accumulation of TDM in the environment (3,4). The open literature contains little to no information on the analysis of TDM. This is partly due to the fact that the matrix can be quite complex. An example would be water soluble emulsion polymer, comprizing hundreds of components which can cause chromatographic interference. Also, alkyl mercaptans such as TDM are difficult to analyze due to reasons such as the alkyl Introduction In the process of manufacturing latex such as styrene-butadiene, a chain transfer agent is required. The chain transfer agent assists in the polymerization to make products of the desired molecular distribution. Previously, chlorinated compounds such as carbon tetrachloride and chloroform have been used for this application, but because of their toxicity and negative environmental effects, it is no longer a practice to employ said compounds for the manufacturing of latex used for the carpet and paper industries. Instead, use of tertiary dodecyl mercaptan (TDM) is preferred for the applications described. As a result of the world demand for latex and the magnitude of the associated industries, TDM has become a chemical of industrial significance. From a manufacturing standpoint, TDM is a mixture of isomeric thiols produced from oligomers of propylene tetramer or sometimes, isobutylene trimer. Propylene tetramer is produced by oligomerization of propylene in the presence of a Friedel- Crafts type catalyst such as sulfuric acid. TDM is produced by passing hydrogen sulfide and either propylene tetramer or isobutylene trimer over a catalyst such as boron trifluoride (1,2). Because of the fact that there are many permutations of the * Author to whom correspondence should be addressed. Figure 1. An injection of 1 µl of iso-octane (solvent) on the analytical system. Note the high degree of selectivity (lack of response to hydrocarbons) towards sulfur compounds. Figure 2. A chromatogram of 10 ppm (v/v) TDM in iso-octane. Chromatographic conditions: 50 C held for 1 min, increased 30 C/min to 280 C and held for 2 min. Reproduction (photocopying) of editorial content of this journal is prohibited without publisher s permission. 665

chains are C 8 to C 15 in size and cover a wide range of boiling points, the polarity of the individual components in TDM varies with the degree of thiolation, the location of the R-SH moiety. In addition to the differences in polarity and boiling points of the TDM components, the product can also contain a fraction of relatively non-polar, non-thiolated tetramer. For the measurement of TDM, an internal method involves the use of headspace gas chromatography in combination with flame photometric detection (FPD) had been developed (5). The method, however, has its constraints, including competing vapour liquid equilibrium of solutes in the sample and the lack of linear dynamic range of the FPD. As a result, a new chromatographic method is required for raw material identification of TDM, for trend analysis, and for the monitoring of residual material in the final products. The new chromatographic method was developed with three enablers: (i) Liquid liquid extraction to remove TDM isomers from their respective matrices; (ii) Low thermal mass gas chromatography to deliver the flexibility of either speciation of individual sulfur compounds, or peak compression to combine individual sulfur compounds into one discreet peak with high temperature programming capability and to improves overall sample to sample throughput; (iii) Dual plasma sulfur chemiluminescence detector (DP-SCD) to offer the highest degree of selectivity for TDM isomers, equi-molar response and a respectable linear dynamic range. This report summarizes the method development and analytical results obtained. Figure 6. A chromatogram of TDM isomers at 240 C in the total sulfur mode: Note the improved peak shape due to effect of temperature. Figure 3. Latex sample type 1 extracted with iso-octane for TDM. Figure 7. A chromatogram of TDM isomers at 280 C in the total sulfur mode: Note the improved peak shape due to effect of temperature. Figure 4. Latex sample type 2 extracted with iso-octane for TDM. Figure 5. A chromatogram of TDM isomers at 200 C in the total sulfur mode: Note the broad peak obtained. Figure 8. A chromatogram of TDM isomers under temperature programming condition: 150 C held for 0.35 min, increased 250 C/min to 280 C and held for 1 min. 666

Experimental An Agilent HP-6890N gas chromatograph (Wilmington, DE) equipped with two split/splitless injectors, a flame ionization detector, a DP-SCD (Agilent Technologies, Boulder, CO) and an Agilent 7673C autosampler were used for the application development. The gas chromatographic conditions used were as follows: split/splitless injector in split mode, at 250 C, Focus liner, with split ratio of 10:1 for trace analysis or 100:1 where the level of TDM exceeds the maximum calibration range of 1000 ppm. The oven was held at 275 C isothermally. The DP-SCD was operated at 802 C, with hydrogen at 47 ml/min, air at 62 ml/min, and pressure at 357 torr. 1 µl was injected for each analysis. A RVM LTM-68A LTMGC module was also used. The column module was connected directly to the injector and the detector of the Agilent HP-6890N GC using standard column ferrules and nuts. A 18 metre 0.25 mm i.d. 0.25 micron Varian VF-1ms column was used in the module for separation. Helium was used as carrier gas, with average linear velocity of 50 cm/s at 20 psig inlet pressure. For speciation, the following temperature programming was employed: 50 C, held for 1 min; increased at 30 C/min to 280 C and held for 2 min. For total sulfur measurement, the following temperature programming was selected: 200 C, held for 0.15 min, increased at 500 C/min to 250 C and held for 2 min. In 2001, Mustacich et al. was granted two patents for the invention of low thermal mass gas chromatography (LTM-GC) (6,7,8). This led to the recent commercialization of a radically new disruptive technology to achieve ultra fast temperature programming and unprecedented cool down time, with a power consumption of only about 1% of conventional GC. Based on resistive heating, the low thermal mass design, coupled with the high surface area offers a theoretical rapid temperature programming rate of up to 30 C per second, or 1800 C/min, in addition to an unprecedented cooling down time. Power consumption is approximately 1% that of conventional GC, primarily due to the effective heating of interstitial air with the random positioning of the heating wire in the toroid, and the low A B Figure 9. A chromatogram of TDM isomers under temperature programming condition: 200 C held for 0.15 min, increased 500 C/min to 250 C and held for 1 min. Figure 11. Linear curves of TDM in iso-octane with various analytical approaches. Speciated: Individual isomer, area summed between 3.5 and 6.0 minutes (A) and Total sulfur: single peak (B). Figure 10. A chromatogram of TDM isomers under temperature programming condition: 220 C held for 0.15 min, increased 500 C/min to 250 C and held for 1 min. Figure 12. A chromatogram of 1 ppm (v/v) of TDM in iso-octane in the speciated mode. Concentration approaching actual detection limit of the system. 667

thermal mass of the assembly (8). The described features, such as rapid heating which lead to rapid remobilization of trapped solutes, rapid cooling, low power consumption in addition to independent oven control of up to four modules per GC make LTMGC ideal for use with the application described herein. TDM standard was obtained from Aldrich Chemicals (Oakville, Ontario, Canada), while iso-octane was used to extract residual TDM from various commercially available latex products. The extraction procedure involved adding 5 ml of iso-octane to 2.5 gram of sample and mechanically agitating for 120 min. The resulting extract analyzed by the procedures described herein. Discussions The GC technique involved offers dual-analytical capability of either speciation or rapid measurement for total sulfur compounds using the same hardware (9). Because TDM is a mixture of isomeric thiols with a range of boiling points, both the classical separation and the total sulfur approach were considered and performance data summarized below: The classical separation approach With the employment of the highly selective and sensitive DP- Table I. A Comparison of Precision Between Two Approaches Speciated Individual isomers Single peak Total sulfur Run Counts Run Counts 1 2,734,206 1 188,565 2 2,831,196 2 198,393 3 2,877,138 3 198,552 4 3,021,790 4 201,533 5 2,876,081 5 203,116 6 2,779,599 6 192,999 7 3,026,342 7 191,584 8 3,062,429 8 198,297 9 2,857,473 9 187,647 10 2,809,881 10 194,463 Average 2,887,614 11 191,737 SD* 112,243 12 188,257 %RSD 3.9 13 201,053 14 195,142 15 196,791 16 191,908 17 186,948 18 193,095 19 192,697 20 200,286 Average 194,653 SD 4920 %RSD 2.5 *SD = standard deviation. SCD, under optimized conditions, the possibility of false positives from interfering hydrocarbons such as residual propylene tetramer or isobutylene trimer can be mimized, if not eliminated (9). If sulfur-containing bactericide is used and its boiling point is outside the range of TDM such as in the case of KATHON, it can be chromatographically separated from the distribution of TDM isomers. For quantitative analysis, the areas of the distribution of TDM isomers were summed together. Method optimization includes mainly sample size so that the selectivity of the detector towards sulfur containing compounds is not compromised and oven temperature profile. Figure 1 shows a chromatogram of a 1 µl injection of isooctane. Note the lack of response of the detector against the matrix used in the retention time of interest, specifically in the region between 3.5 and 6.0 min. Figure 2 shows a chromatogram of 10 ppm (v/v) of TDM. Note the respectable sensitivity obtained. Figures 3 and 4 show chromatograms of typical iso-octane extracts of residual TDM in two different type of styrene-butadiene latex. The GC conditions used were as follows: Split/splitless injection system, 1.0 µl injection; SR: 100:1 or 10:1. The LTMGC and temperature profile were: VF-1ms (18 m 0.25 mm i.d. 0.25- µm); 50 cm/s helium (20 psig inlet pressure); Temperature: 50 C held for 1 min, increased at 30 C/min to 280 C and held for 2 min; Host oven, 275 C; DP-SCD; Burner, 802 C; Vacuum, 357 Torr; H 2 : 47mL/min; Air 62 ml/min. The total sulfur approach In the total sulfur approach, the separation power of the column is compressed by operating the column at an elevated temperature. The rationale of this approach is that because all the isomers of TDM are compressed into one discreet, Gaussian peak, in theory, the sensitivity of the method can be improved and since chromatographic separation is not required, shorter analytical time can be attained. The downside, however, is if there is any sulfur containing compounds in the sample retainable by the chromatographic column, it will also be measured as TDM. Method optimization involved selecting the appropriate operating temperature for the analytical column to obtain a symmetric peak for reliable quantitative work. Peak symmetry quality was compared between maintaining the column temperature isothermally versus a slight temperature program. Figure Figure 13. A comparison of TDM results obtained by two different approaches. 668

5, 6, and 7 show chromatogram of TDM isomers at 200 C, 240 C, and 280 C while Figures 8, 9, and 10 show chromatogram of TDM isomer under various temperature programming conditions with initial temperature at 150 C, 200 C, and 220 C. The gas chromatographic conditions used were as follows: 1.0 µl injection; SR, 100:1 or 10:1. LTMGC and temperature profile were as follows: VF-1ms (18 m 0.25 mm i.d. 0.25 µm); 50 cm/sec Helium (20 psig inlet pressure); temperature, 200 C held for 0.15 min, increased at 500 C/min to 250 C; Host oven, 275 C. DP-SCD: Burner, 802 C; Vacuum, 357 Torr; H 2, 47 ml/min; Air, 62mL/min Comparison of performance between the two approaches In terms of precision, as stated earlier for the speciated method, the distribution of individual isomers of TDM between 3.5 and 6.0 min was integrated, whereas for the total method, the discreet peak representing TDM was integrated. Standards containing 1000 ppm (v/v) of TDM in iso-octane were used for the evaluation. A relative standard deviation of 2.5% (n = 20) was obtained for the speciated method while a relative standard deviation of 3.9% (n = 10) was obtained for the total sulfur method. The results obtained were tabulated in Table I. In terms of linearity, over the range from 1 ppm to 1000 ppm (v/v) TDM, correlation coefficients R 2 of 0.9994 and of 0.9995 were obtained for the speciated and the total sulfur method, respectively. The detection limit for TDM by the total sulfur method was found to be 0.5 ppm (v/v) TDM whereas 1.0 ppm (v/v) for the speciated method as shown in Figures 11 and 12. Table II shows a comparison of five iso-octane extract samples containing TDM. It was found the results obtained were comparable amongst the two methods; despite there is a trend that the TDM results obtained by the total sulfur method is consistently Table II. A Comparison of Five Iso-Octane Extract of TDM From Latex Samples Under Two Operating Conditions: Speciate and Total Speciate: Individual isomers Sample # Area counts ppm TDDM 1 2,943,432 4638 2 3,387,471 5338 3 2,991,420 4714 4 2,742,534 4321 5 2,734,206 4308 5 2,831,196 4461 (control duplicate) Single peak: Total sulfur Sample Area Counts ppm TDDM 1 1,275,009 5190 1 1,313,176 5345 (control duplicate) 2 1,573,582 6405 3 1,323,256 5386 4 1,381,066 5621 5 1,256,298 5113 elevated as shown in Figure 13. Some plausible explanations for this bias include the samples might have a different distribution of isomers than TDM used for calibration, or more of the TDM is detected in the samples as isomers are thermally band compressed into a much shorter peak width than classical method. Nevertheless, the results obtained show that the concept of tracking for the presence of TDM by measuring its sulfur content and associated retention time range in the speciated method or by measuring its sulfur content alone can be employed for the material identification, trend monitoring, or the measurement of residual TDM in various matrices. If a high degree of accuracy is required, the results obtained by using said techniques must be compared to other assaying techniques. Conclusions A gas chromatographic technique has been successfully developed for the measurement of TDM based on its sulfur content for raw material identification, trend analysis, or for the measurement of un-reacted material in the final products. The method employs LTM-GC offering the flexibility either for speciation of individual sulfur compounds or delivering peak compression to combine individual sulfur compounds into one discreet peak without changing of hardware, and a DP-SCD to attain a high degree of sensitivity and selectivity. Using the technique described, a detection limit in the range of 0.5 ppm (v/v) TDM with less than 1 min analysis can be achieved. Response is linear over four orders of magnitude with a high degree of repeatability of less than 5%. Acknowledgements Special thanks to the Global Separations Leadership Team and the Global GC Steering Team for their support of the project. Dr. Mary Fairhurst, Lyndon Sieben, Myron Hawryluk, Vicki Carter, Michelle Baker of Analytical Sciences are acknowledged for their support. Dr. Robert Mustacich of RVM Scientific is acknowledged for the fruitful discussion and support of this work. The authors would also like to thank Dr. Terry McCabe of Dow, Analytical Sciences and the editors for their help in preparing the manuscript. This project is partially funded by Analytical Sciences, Core Technologies, 2007 Competency Improvement Funds (CIF). Siltite is a trademark of SGE Analytical Science, Australia KATHON LXE is a trademark of Rohm and Haas, Philadelphia, USA. References 1. W. Schulze, J. Lyon, and G. Short. Synthesis of tertiary alkyl mercaptans. Ind. Eng. Chem. 40(12): 2308 13 (1948). 2. J. Frantz and K. Glass. A new process for dodecyl mercaptan. Chem. Eng. Prog. 59(7): 68 73 (1963). 669

3. DEFRA, UK Co-coordinated Chemical Risk Management Program Environmental Chemicals, Department of Environment, Food and Rural Affair, 2008. http://www.defra.gov.uk/environment/ chemicals/ukrisk.htm, accessed date: January 13, 2008 4. A. Girling, L. McLauchlin, and R. Wildey. Environmental Risk Assessment Report: Tert-dodecanethiol (CAS 25103-58-6). Environmental Agency, Rio House, Waterside Drive, Aztec West, Bristol BS324UD, 2004. 5. P. Eckerle and B. Fouchet. Residual TDM in emulsion polymers by headspace capillary chromatography, using flame photometric detection. Dow Internal Report DOC 7443, Rheimunster, Germany, February 28, 2005. 6. R. Mustacich and J. Everson. Reduced power consumption gas chromatograph system. US Patent 6,217,829, April 17, 2001. 7. R. Mustacich and J. Richards. Electrically insulated gas chromatograph assembly and method of fabricating same. US Patent 6,209,386, April 3, 2001. 8. J. Luong, R. Gras, R. Mustacich, and H. Cortes. Low thermal mass gas chromatography: principles and applications. J. Chromatogr. Sci. 44(4): 253 61 (2006). 9. R. Gras, J. Luong, and R. Shearer. A unified approach for the measurement of individual or total volatile organic sulfur compounds in hydrocarbon matrices by dual-plasma chemiluminescence detector and low thermal mass gas chromatography. J. Chromatogr. Sci. 45(10): 671 76 (2007). 10. R. Benner and D. Stedman. Universal Sulfur Detection by Chemiluminescence, Analytical Chemistry, 61 (1989), 1268 71. 11. R. Shearer. Development of Flameless Sulfur Chemiluminescence Detetion: A pplications to Gas Chromatography. Journal of High Resolution Chromatography 421 26 (1990). 12. C. Poole. The Essence of Chromatography. Elsevier Science B.V., Amsterdam, the Netherlands, 2003. 13. W. Jennings. Analytical Gas Chromatography, Second Edition, Academic Press, New York, 1997. 14. R. Nadkami and A. Kishore. Determination of sulfur in petroleum products and lubricants: a critical review of test performance. American Laboratory 32: 16 29 (2000). Manuscript received January 24, 2008; Revision received March 24, 2008. 670