Ground Calibration of X-ray CCD Detectors with Charge Injection for the X-ray Imaging Spectrometer on Astro-E2

Similar documents
Front- and back-illuminated X-ray CCD performance in lowand high-earth orbit: Performance trends of Suzaku XIS and Chandra ACIS detectors

Progress in X-ray CCD Sensor Performance for the Astro-E2 X-ray Imaging Spectrometer

The onboard calibration for the spaced-row charge injection of Suzaku XIS

Quantum e$ciency of the CCD camera (XIS) for the ASTRO-E mission

A charge transfer inefficiency correction model for the Chandra Advanced CCD Imaging Spectrometer

X-ray CCD response functions, front to back

An Experimental Study of Charge Diffusion in the Undepleted Silicon of X-ray CCDs

X-ray CCD calibration system using #uorescent lines

A Technique to Measure Trap Characteristics in CCDs Using X-rays. Center For Space Research, Massachusetts Institute of Technology,

Characterization of low energy ionization signals from Compton scattering in a CCD Dark Matter detector

Focal plane instrumentation for the Wide-Field X-ray Telescope

X-ray absorption near edge structure in the quantum efficiency of x-ray charge-coupled devices

The High Definition X-ray Imager (HDXI) Instrument on the Lynx X-Ray Surveyor

Semiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH

X- & γ-ray Instrumentation

X-Ray and Gamma-Ray Telescopes and Instruments for Astronomy, Joachim E. Trümper, Harvey D. Tananbaum, Editors, Proceedings of SPIE Vol.

Status and Calibration of the erosita X ray Telescope

Soft X-ray transmission of optical blocking filters for the X-ray CCD cameras onboard Astro-E 2

The X-Ray Universe. Potsdam University. Dr. Lidia Oskinova Wintersemester 2008/09

Centre for Electronic Imaging

Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors

TIMEPIX3 First measurements and characterization of a hybrid pixel detector working in event driven mode

Detector R&D at KIPAC

arxiv: v1 [physics.ins-det] 19 Apr 2014

CCD OPERATION. The BBD was an analog delay line, made up of capacitors such that an analog signal was moving along one step at each clock cycle.

Detection of X-Rays. Solid state detectors Proportional counters Microcalorimeters Detector characteristics

Development of CCDs for REXIS on OSIRIS-REx

Detector R&D at KIPAC. Hiro Tajima Kavli InStitute of Particle Astrophysics and Cosmology

1. Give short answers to the following questions. a. What limits the size of a corrected field of view in AO?

Detecting high energy photons. Interactions of photons with matter Properties of detectors (with examples)

Distribution and Identification of ACIS-S4 Streak Events. March 16, 2004

arxiv:astro-ph/ v1 31 Oct 2001

Detector R&D in KAPAC

Improving the Relative Accuracy of the HETGS Effective Area

This work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory under contract

Point-Source CCD Photometry with STIS: Correcting for CTE loss

The Open University s repository of research publications and other research outputs

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

X-ray Fluorescence Imaging Following Synchrotron Beam Excitation

Status / Progress of the erosita X ray Observatory:

Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers

Characterization of the radiation damage in the Chandra X-ray

Chandra Source Catalog Energy Bands

Metallic magnetic calorimeters. Andreas Fleischmann Heidelberg University

ACIS Gain Studies Temporal, Spatial, and Temperature Dependencies

Special SLS Symposium on Detectors

Tsukuba Space Center, Sengen Tsukuba Ibaragi , Japan 3 Solid State Division, Hamamatsu Photonics K.K., (Received )

RADIATION DAMAGE IN HST DETECTORS

M. De Napoli, F. Giacoppo, G. Raciti, E. Rapisarda, C. Sfienti. Laboratori Nazionali del Sud (LNS) INFN University of Catania. IPRD Oct.

AN EVALUATION OF A BAKE-OUT OF THE ACIS INSTRUMENT ON THE CHANDRA X-RAY OBSERVATORY

Development of a Radiation Hard CMOS Monolithic Pixel Sensor

Lessons learned from Bright Pixels and the Internal Background of the EPIC pn-ccd Camera

Calibration of the IXPE Instrument

Portable type TXRF analyzer: Ourstex 200TX

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

Characterization of the silicon drift detector for NICER instrument

Phonon-Mediated Distributed Transition- Edge-Sensor X-ray Detectors

Neutron Capture Experiments with DANCE

Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy

Multilayer coating facility for the HEFT hard X-ray telescope

MEASUREMENT OF TEMPORAL RESOLUTION AND DETECTION EFFICIENCY OF X-RAY STREAK CAMERA BY SINGLE PHOTON IMAGES

WFC3 TV2 Testing: UVIS-2 Dark Frames and Rates

Charge sharing in silicon pixel detectors

Open Research Online The Open University s repository of research publications and other research outputs

CRaTER Pre-Environmental Review (I-PER) Science Requirements Update

Development of New MicroStrip Gas Chambers for X-ray Applications

Studies on 3D Fusion Reactions in TiDx under Ion Beam Implantation

An Interdigitated Pixel PIN Detector for Energetic Particle Spectroscopy in Space

The Gaia CCD radiation damage

High-Resolution Gamma-Ray and Neutron Detectors For Nuclear Spectroscopy

On ground and in-flight calibration. Fabio Muleri INAF-IAPS

Semiconductor Detectors

Chapter 4 Scintillation Detectors

Results of a Si/CdTe Compton Telescope

BENEFITS OF IMPROVED RESOLUTION FOR EDXRF

CRaTER Pre-Ship Review (PSR) Instrument Calibration Science Requirements Compliance

MoBiKID Kinetic Inductance Detectors for up-coming B-mode satellite experiments

Quantum efficiency measurements of erosita pnccds

Cadmium Zinc Telluride (CZT) Detectors

Precise Low-Energy Electron Tracking Using a Gaseous Time Projection Chamber for the Balloon-Borne Gamma Ray Compton Telescope

Study of Edgeless TimePix Pixel Devices. Dylan Hsu Syracuse University 4/30/2014

arxiv:physics/ v1 3 Aug 2006

normalized occurence

Direct search for low mass dark matter particles with CCDs. Juan Estrada - Fermilab 10/13/2009

The High Resolution X-ray Spectrometer, SXS, on the Astro-H mission

Characterization of a high-performance Ti/Au TES microcalorimeter with a central Cu absorber

THE spectroscopic performance of large volume CdZnTe

Photon Instrumentation. First Mexican Particle Accelerator School Guanajuato Oct 6, 2011

WFC3 TV2 Testing: IR Channel Read Noise

Silicon Drift Detectors for gamma-ray detection: 15 years of research (and collaboration between Politecnico and INAF-Milano)

arxiv: v1 [astro-ph.im] 31 Jul 2014

Requirements for the Star Tracker Parallel Science Programme

EDS Mapping. Ian Harvey Fall Practical Electron Microscopy

The HXDS Solid State Detectors

Radiation Tolerance of Fully-Depleted P-Channel CCDs Designed for the SNAP Satellite

X-ray fluorescence analysis - calibration of the X-ray energy detector

Silicon Drift Detectors: Understanding the Advantages for EDS Microanalysis. Patrick Camus, PhD Applications Scientist March 18, 2010

Queen s University PHYS 352

Understanding X-rays: The electromagnetic spectrum

THE ENIGMATIC X-RAY JET OF 3C120

Transcription:

Ground Calibration of X-ray CCD Detectors with Charge Injection for the X-ray Imaging Spectrometer on Astro-E2 Beverly LaMarr a, Mark Bautz a, Steve Kissel a, Gregory Prigozhin a Kiyoshi Hayashida b Takeshi Tsuru c and Hironori Matsumoto c a Massachusetts Institute of Technology, Cambridge MA, USA; b Osaka University, Osaka, Japan; c Kyoto University, Kyoto, Japan ABSTRACT The X-ray Imaging Spectrometer (XIS) on the Astro-E2 mission contains X-ray CCDs fabricated at MIT Lincoln Laboratory. We present results from CCD subassembly calibration of these detectors. Absolute detection efficiency and spectral resolution in the 0.3-10 kev band are presented for four front-illuminated and two backilluminated flight candidate devices. In addition, we characterize the charge injection capabilities of these devices. The quantum efficiency measurements have been used to model the various dead layers in the gate structure and to constrain the depletion depth of the detectors. Noise levels of 2 electrons, RMS, and spectral resolution better than 50 ev (FWHM) at 525 ev have been achieved. The small but measurable charge transfer inefficiency ( less than 1e-5 per transfer at 5.9 kev) of these devices has been characterized as a function of X-ray energy. Keywords: XIS, X-ray performance, CTI, charge injection 1. INTRODUCTION As part of our work developing the X-ray Imaging Spectrometer (XIS) 1 for the joint Japanese/US mission Astro- E2 2 we have characterized a number of X-ray CCDs. These CCDs, designed and fabricated at MIT Lincoln Laboratory, include a novel charge injection structure and are described in greater detail in two companion papers in these proceedings. 3 4 The work we describe here is one part of the XIS calibration program. It was carried out at the MIT Center for Space Research. Additional calibration of these sensors, performed at Osaka University and Kyoto University, will be described elsewhere. 2. SETUP AND MEASUREMENTS The calibration facilities used in the measurements have been described elsewhere 5 6 and we present only a brief summary here. Two kinds of X-ray sources are used: the High Energy X-ray Source (HEXS) is an X-ray fluorescence source excited by a tungsten anode electron-bombardment source. A number of targets are available and lines at energies between 1.49 and 9.9 kev have been used here. At lower energies, a tritium source is used to excite fluorescence from targets of Teflon and boric oxide. Quantum efficiency measurements are made by comparing the XIS devices to a reference CCD that has been calibrated absolutely using undispersed synchrotron radiation at a PTB beam line of a BESSY synchrotron storage ring. The reference detector and the test detector are both mounted on a translation stage so that they can be exposed to the X-ray source with the same illumination geometry. Each of the flight candidate devices was calibrated at MIT relative to an ACIS flight spare front-illuminated device, w168c2r. In order to minimize the possible effects of pileup to the relative quantum efficiency measurements, the reference chip was clocked with the same frame time, 8 seconds, as the flight candidate devices. Despite this, pileup effects are likely relevant at the lowest energies for the back-illuminated devices because of their much higher quantum efficiency. Send correspondence to B.L. B.L.: E-mail: fergason@space.mit.edu, Telephone: 617 258 8153 SPIE USE, V. 1 5501-51 (p.1 of 7) / Color: No / Format: Letter/ AF: Letter / Date: 2004-05-28 17:01:34

System Energy or Band Total Counts No. of (kev) (/energy/ccd quad.) Frames HEXS 2.622 (Cl-K) 2 10 5 per energy 500 6.404 (Fe-K) 450 9.886 (Ge-K) 900 4.952 (V-K) 2 10 5 per energy 400 6.930 (Co-K) 600 7.478 (Ni-K) 550 8.639 (Zn-K) 900 Table 1. Spectral Resolution Measurements Spectral resolution measurements were made at the energies listed in Table 1. We characterize the FWHM by fitting a single Gaussian function to the recorded pulse height spectrum. For each of the four detector output nodes typically at least 2.5E5 counts were detected under the peak of the primary line at each energy. The quantum efficiency measurements were made at the energies listed in Table 2. Typically at least 1E6 counts were collected at each energy over the array. While we have enough data to map the quantum efficiency on scales smaller than the telescope PSF (about 2.5 mm half power diameter) we present only the median quantum efficiency for each detector. System Energy Total Counts No. of (kev) (/energy/) Frames ccd quad.) Tritium 0.277 (C-K) 2.5 10 4 2000 0.525 (O-K) 2.5 10 5 800 0.677 (F-K) 1200 HEXS 1.489 (Al-K) 2.5 10 5 500 1.740 (Si-K) 450 2.014 (P-K) 500 4.511 (Ti-K) 500 5.899 (Mn-K) 2000 8.048 (Cu-K) 750 Table 2. Quantum Efficiency Measurements in timed-exposure mode (at 8s per frame) with CCD at -90C. 3. RESULTS 3.1. Spectral Resolution The spectral resolution of the front-illuminated flight candidate devices ranges from around 50eV to 180eV over the energy range of 0.3 kev to 9.9 kev as shown in Figure 1. The measured resolution at the lowest energy, 277 ev, suffers both from a very low count rate and spectral contamination in the X-ray source. At the highest energies the resolution suffers slightly from the parallel charge transfer inefficiency (CTI) described in section 3.3. SPIE USE, V. 1 5501-51 (p.2 of 7) / Color: No / Format: Letter/ AF: Letter / Date: 2004-05-28 17:01:34

The spectral resolution of the back-illuminated flight candidate devices ranges from around 60eV to 170eV over the energy range of 0.3 kev to 9.9 kev as shown in Figure 2. The resolution at the lowest energies, 277 ev and 525 ev, suffer from spectral contamination in the X-ray source and, possibly, pileup of events. The best fit to the FWHM as a function of energy as described by FWHM(E) = (FWHM 0 ) 2 +2.35 2 G E (1) for each device is overplotted as solid lines in both Figures 1 and 2. Here FWHM 0 is the (readnoise limited) minimum FWHM, and G is expected to equal the product of the Fano factor and w, the ionization yield in silicon. Values of FWHM 0 and G fit for each devices are listed in Table 3. We define the effective readout noise, σ eff = FWHM 0 /2.35/w (2) Taking w =3.65eV/e, we obtain values for σ eff listed in Table 3. We note that σ eff exceeds the measured readout noise of 1.5-2.5 e- RMS, probably because many X-ray events occupy more than one pixel. This effect is larger, at low X-ray energies, for the back-illuminated devices. Chip FWHM 0 G σ eff w1.7c5 31.34 0.529 3.65 w1.7c6 28.08 0.510 3.27 w1.3c6 29.27 0.510 3.41 w1.14c7 31.51 0.508 3.67 w1.8c5 49.89 0.439 5.82 w1.8c2 55.17 0.437 6.43 Table 3. FWHM Fit Values for Flight Candidate Devices The spectral resolution for the front-illuminated devices was calculated from the 1-, 2-, and 3-pixel events with a split threshold of 13 adu (one adu is approximately one electron). The same grade selection was used for the back-illuminated devices, but a split threshold of 7 adu was applied in order to improve the resolution. The values plotted are the average of all four readout nodes for each device. 3.2. Quantum Efficiency The median absolute quantum efficiency of each front-illuminated flight candidate is plotted in Figure 3. The solid lines are the best fit to a simple dead layer model. The dashed line in Figure 3 is the best fit to the model for a back-illuminated device, w1.11c6. This device is an engineering rather than a flight device. The quantum efficiency analysis for the flight back-illuminated detectors is in process at this writing. 3.3. Charge Transfer Inefficiency At the nominal operating temperature of -90C, the flight candidate devices all have a small but measurable charge transfer inefficiency (CTI) which is plotted, as a function of energy, in Figure 4. The CTI at each energy is determined by binning the center pixel pulse heights of X-ray events according to their row location and then calculating their centroid. A straight line fit to the centroids yields the amount of charge lost to each transfer. SPIE USE, V. 1 5501-51 (p.3 of 7) / Color: No / Format: Letter/ AF: Letter / Date: 2004-05-28 17:01:34

Figure 1. Resolution of Flight-candidate Front-illuminated Devices 3.4. Charge Injection The charge injection study on the devices is described in more detail elsewhere in the proceedings. 3 4 Charge injection was characterized for each of the flight candidate devices. The charge was injected in a checkerboard pattern such that each column got a single injected packet separated from other injected charge packets to allow the standard event processing analysis to detect the charge as if it were an X-ray event. The injected charge is then treated precisely like X-rays would be, including event grade selection and pulse height summation. The injected charge as a function of commanded input gate volts for each flight device appears in Figure 5. Note that we can inject charge with amplitudes ranging from a few hundred to a few thousand ev. The charge injection capabilities and characteristics are described in much greater detail in Reference 3. ACKNOWLEDGMENTS We thank John Doty, Ed Boughan and Rick Foster at MIT CSR for design and construction of the XIS analog electronics, Barry Burke and Jim Gregory at MIT Lincoln Laboratory for design and fabrication of the XIS CCDs, respectively, and our colleagues at ISAS/JAXA, Osaka University and Kyoto University for the opportunity to participate in the XIS program. This work was supported by ISAS/JAXA via NASA contract NAS5-02094. REFERENCES 1. T. Dotani and et al., The x-ray imaging spectrometer (xis) on board astro-e2, in X-ray and Gamma-Ray Telescopes and Instruments for Astronomy, H. D. T. J. E. Trümper, ed., Proc. SPIE 4851, pp. 289 292, 2002. 2. H. Inoue, The astro-e2 mission, in X-ray and Gamma-Ray Telescopes and Instruments for Astronomy, H. D. T. J. E. Trümper, ed., Proc. SPIE 4851, pp. 289 292, 2002. SPIE USE, V. 1 5501-51 (p.4 of 7) / Color: No / Format: Letter/ AF: Letter / Date: 2004-05-28 17:01:34

Figure 2. Resolution of Back Illuminated Devices 3. G. Y. Prigozhin, B. Burke, M. Bautz, S. Kissel, B. LaMarr, and M. Freytsis, X-ray ccd with charge injection structure, in High-Energy Detectors in Astronomy, A. D. H.& and L. Strüder, eds., Proc. SPIE 5501 (these proceedings), 2004. 4. M. Bautz, G. Prigozhin, S. Kissel, B. LaMarr, G. Ricker, B. Burke, and J. Gregory, Progress in x-ray ccd sensor performance for the astro-e2 x-ray imaging spectrometer, in High-Energy Detectors in Astronomy, A. D. H. & and L. Strüder, eds., Proc. SPIE 5501 (these proceedings), 2004. 5. S. Jones, M. Bautz, S. Kissel, and M. Pivovaroff, Using tritium and x-ray tubes as x-ray calibration sources for the axaf ccd imaging spectrometer, Proc. SPIE 2080, p. 158, 1996. 6. M. Pivovaroff, S. Kissel, M. Bautz, G. Prigozhin, T. Isobe, and J. A. Gregory, Flight x-ray ccd selection for the axaf ccd imaging spectromeeter, Proc. SPIE 2080, p. 182, 1996. SPIE USE, V. 1 5501-51 (p.5 of 7) / Color: No / Format: Letter/ AF: Letter / Date: 2004-05-28 17:01:34

Figure 3. Absolute Quantum Efficiency Figure 4. Parallel Charge Transfer Inefficiency SPIE USE, V. 1 5501-51 (p.6 of 7) / Color: No / Format: Letter/ AF: Letter / Date: 2004-05-28 17:01:34

Figure 5. Injected Charge (in units of Equivalent X-ray Energy) for Flight Candidate Devices SPIE USE, V. 1 5501-51 (p.7 of 7) / Color: No / Format: Letter/ AF: Letter / Date: 2004-05-28 17:01:34