Triwave. More Complete Characterization of Mixtures and Molecules

Similar documents
Multi-stage Mass Spectrometry up to MS 4 on a QTof System

Improved Ion Mobility Separation of Protein Conformations in the Gas Phase with SYNAPT G2 HDMS

[ instrument specifications ]

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS

APPLICATION SOLUTIONS FOR STRUCTURAL PROTEOMICS

Electron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Stephen McDonald and Mark D. Wrona Waters Corporation, Milford, MA, USA INT RO DU C T ION. Batch-from-Structure Analysis WAT E R S SO LU T IONS

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Stephen McDonald, Mark D. Wrona, Jeff Goshawk Waters Corporation, Milford, MA, USA INTRODUCTION

MALDI-HDMS E : A Novel Data Independent Acquisition Method for the Enhanced Analysis of 2D-Gel Tryptic Peptide Digests

Analysis of Labeled and Non-Labeled Proteomic Data Using Progenesis QI for Proteomics

INTRODUCTION OVERVIEW

MASS ANALYSER. Mass analysers - separate the ions according to their mass-to-charge ratio. sample. Vacuum pumps

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

Translational Biomarker Core

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Estimating Collision Cross Sections from Travelling Wave IM-MS Data: How it is done and what are the Problems?

- Supporting Information -

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

FORENSIC TOXICOLOGY SCREENING APPLICATION SOLUTION

Introduction to the Q Trap LC/MS/MS System

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

TANDEM MASS SPECTROSCOPY

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

Mass Analyzers. Principles of the three most common types magnetic sector, quadrupole and time of flight - will be discussed herein.

Fereshteh Zandkarimi, Samanthi Wickramasekara, Jeff Morre, Jan F. Stevens and Claudia S. Maier

CEE 772 Lecture #27 12/10/2014. CEE 772: Instrumental Methods in Environmental Analysis

CEE 772: Instrumental Methods in Environmental Analysis

Conformational Characterization of Calmodulin by Hydrogen Deuterium Exchange Mass Spectrometry

Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate Yu 3

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

High-Field Orbitrap Creating new possibilities

Analysis of Polar Metabolites using Mass Spectrometry

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365

LECTURE-11. Hybrid MS Configurations HANDOUT. As discussed in our previous lecture, mass spectrometry is by far the most versatile

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

[ xevo G2 tof ] Your Quantitative and Qualitative Questions Answered. Exactly.

Types of Analyzers: Quadrupole: mass filter -part1

MS/MS .LQGVRI0606([SHULPHQWV

Quantitative and Qualitative Determination of Organics in Water Using Electrospray Ionization Coupled with Ion Mobility Spectrometry

Mass Spectrometry and Proteomics - Lecture 2 - Matthias Trost Newcastle University

Confirmation of In Vitro Nefazodone Metabolites using the Superior Fragmentation of the QTRAP 5500 LC/MS/MS System

Instrumentation development for coupling ion/ion reactions and ion mobility in biological mass spectrometry

Harris: Quantitative Chemical Analysis, Eight Edition

Powerful Scan Modes of QTRAP System Technology

PosterREPRINT COMPARISON OF PEAK PARKING VERSUS AUTOMATED FRACTION ANALYSIS OF A COMPLEX PROTEIN MIXTURE. Introduction

Thermo Finnigan LTQ. Specifications

Lecture 8: Mass Spectrometry

UPLC Intact MASS Analysis Application Kit

Metabolic Phenotyping Using Atmospheric Pressure Gas Chromatography-MS

Ion Mobility Mass Spectrometry of Complex Carbohydrates Collision Cross Sections of Sodiated N- linked Glycans. - Supporting Information -

Lecture 8: Mass Spectrometry

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

IN QUALITATIVE ANALYSIS,

Peptide Isolation Using the Prep 150 LC System

Ion trap. 3D Quadrupole Trap. -Mass spectrometer -Ion trap (Quadrupol Ion Storage, Quistor) 18. April

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition

Other Methods for Generating Ions 1. MALDI matrix assisted laser desorption ionization MS 2. Spray ionization techniques 3. Fast atom bombardment 4.

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

EXPRESSION MORE PEPTIDES. MORE PROTEINS. MORE REPRODUCIBILITY. MORE QUANTIFICATION.

THE IMPLEMENTATION OF A SCREENING WORKFLOW FOR ION MOBILITY QUADRUOPOLE TIME-OF-FLIGHT MASS SPECTROMETRIC ANALYSIS OF PFOS ISOMERS

ABI 3200 Q TRAP LC/MS/MS System

Mass Analyzers. Ion Trap, FTICR, Orbitrap. CU- Boulder CHEM 5181: Mass Spectrometry & Chromatography. Prof. Jose-Luis Jimenez

Secondary Ion Mass Spectroscopy (SIMS)

Olumide Adebolu. Chromatographic Fidelity and Matrix /Analyte Solubility in Complex Polymer Systems using HPLC-MALD/I TOF MS

Quality control analytical methods- Switch from HPLC to UPLC

Everyday NMR. Innovation with Integrity. Why infer when you can be sure? NMR

Mass Spectrometry. What is Mass Spectrometry?

Tendenze nell innovazione della strumentazione in spettrometria di massa:

C HA R AC T E RIZ AT IO N O F INK J E T P RINT E R C A RT RIDG E INK S USING A CHEMOMETRIC APPROACH

Incorporation of post-ionization, gas-phase separations

Ion Mobility-Mass Spectrometry: A Tool for Characterizing the Petroleome

Guide to Peptide Quantitation. Agilent clinical research

Designed for Accuracy. Innovation with Integrity. High resolution quantitative proteomics LC-MS

This is the total charge on an ion divided by the elementary charge (e).

LC-MS Based Metabolomics

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

GRADUATE COURSE IN MASS SPECTROMETRY: LECTURE 2

Chapter 1. Introduction

Finnigan LCQ Advantage MAX

Mass Analyzers. mass measurement accuracy/reproducibility. % of ions allowed through the analyzer. Highest m/z that can be analyzed

Atomic masses. Atomic masses of elements. Atomic masses of isotopes. Nominal and exact atomic masses. Example: CO, N 2 ja C 2 H 4

LECTURE-13. Peptide Mass Fingerprinting HANDOUT. Mass spectrometry is an indispensable tool for qualitative and quantitative analysis of

MetWorks Metabolite Identification Software

Chemistry Instrumental Analysis Lecture 35. Chem 4631

1. The range of frequencies that a measurement is sensitive to is called the frequency

Choosing the metabolomics platform

Introduction to LC-MS

CHROMATOGRAPHY AND MASS SPECTROMETER

ICP-MS. High Resolution ICP-MS.

Transcription:

Triwave More Complete Characterization of Mixtures and Molecules

INTRODUCTION This white paper describes: Benefits of T-Wave Ion Mobility Separations (IMS) Triwave design and implementation of T-Wave IMS The continued need for more comprehensive characterization of complex mixtures or complex molecules in today s analytical applications is a primary driver for the development of powerful new mass spectrometry technologies. With higher levels of separation power (selectivity) and new experimental capabilities (versatility), scientists from a wide range of disciplines can carry out more complete profiling of mixtures and structural characterization. T-Wave IMS benefits There are a number of benefits to using T-Wave ion mobility that have been reported in over 200 peer-reviewed papers. 5 Orthogonal separation power The second generation Triwave device has an IMS resolution sufficient to separate: Isobaric ions, ions of very similar m/z but different compounds. Conformers or isomers that have multiple species with subtle changes in size and shape, as shown in Figure 1. Compounds of different classes. of different charge state. 3 Based on Travelling wave 1,2 (T-Wave is a type of stacked ring ion guide, also referred to as T-Wave ion guide, or TWIG) technology, Triwave is a unique device designed to maximize sample separation in an analytical workflow using ion mobility separations (IMS). 3,4 Conventional Mass Spectrometry enables separation of ions based on m/z, the addition of IMS further enables the differentiation of ions by size, shape and charge, as well as m/z. Triwave therefore provides an additional, orthogonal dimension of separation in a short timescale enabling scientists to: Increase analytical peak capacity with an extra orthogonal dimension of separation power. Figure 1. Separating small molecule drug isomers 3 with T-Wave ion mobility. Ability to identify and quantify with more coverage and confidence. Derive information on conformation and structure with mass spectrometry. Deliver the most comprehensive structural characterization with novel fragmentation routines. It is shown that IMMS reveals 2 to 3 gas-phase conformer populations for IgG2s. In contrast, a single gas phase conformer is revealed using IMS for both an IgG1 antibody and a Cys-232 Ser mutant IgG2, both of which are homogeneous with respect to disulfide bonding. This provides strong evidence that the observed IgG2 gasphase conformers are related to disulfide bond heterogeneity. Additionally, IMS MS analysis of redox enriched disulfide isoforms allows unambiguous assignment of the mobility peaks to known disulfide structures. These data clearly illustrate how IMMS can be used to quickly provide information on the higher order structure of antibody therapeutics. Bagal D., et al. Rapid Commun. Mass Spectrom. 2008; 22: 2898-2904. 2

Figure 2. Visualization of the distributions of two ions (shown in red in the m/z versus drift time plot), which differ significantly in drift time but have a mass difference so small that it would require a m/z resolving power in excess of 1.8 million to achieve differentiation of the ions without T-Wave IMS separation. The spatial distributions of the two ions are clearly different, when plotted using m/z and drift time to generate the images, as shown in the two panels for m/z 545.9499 and m/z 545.9502. Improved coverage and confidence Due to the compatibility of analytical timescales, shown in Figure 3, T-Wave IMS can be employed in hyphenated approaches, without detriment to duty cycle or data quality. This can provide significant gains in selectivity (analytical peak capacity, shown in Figure 4) and specificity (confidence in results, shown in Figure 5), when compared to the use of m/z separation alone: TOF MS IMS LC -9-8 -7-6 -5-4 -3-2 -1 0 1 2 3 4 10 n seconds Figure 3. Compatible timescales IMS separations integrate perfectly between those of UPLC and Tof. Up to 100% increase in identifications compared to non-ims based approaches. 7 Improved LOD and LOQ where new detections are typically at the lowest concentrations. 6,7 Improved confidence in identification or confirmation with cleaner, interference-free fragment ion spectra. 7 Figure 4. A plot of retention time versus drift time, illustrating how peak capacity can be increased by embedding T-Wave IMS with UPLC/MS. 3

Figure 5. Generating cleaner spectra for more confident detection and identification from complex mixtures (above spectrum shows identified peptide from a protein digest mixture). Analysis with IMS eliminates significantly reduces interference (grey peaks in upper spectrum) from ions of similar m/z or rt, to those of interest. The Waters SYNAPT G2 with ion mobility has greatly enhanced selectivity for proteomic analysis. An orthogonal measure, ion mobility dramatically improves the label-free MS E methodology used in our research. Andrew K. Ottens, Ph.D. Assistant Professor of Anatomy & Neurobiology and Biochemistry, Virginia Commonwealth University, VA, USA 4

Providing information on conformation T-Wave IMS enables the characterization of molecular structure through the determination of collision cross section values (CCS), which is a measure of the average rotational collision cross section of a molecule. T-Wave IMS can be used as a powerful complementary approach to traditional structural methods 8 (X-ray crystallography, Nuclear Magnetic Resonance spectroscopy, or Electron Microscopy) through its ability to: Separate isomers, isoforms, and isobaric molecules from mixtures. Provide conformational measurements at very low (physiological) concentrations. Enable analysis of compounds over a very wide m/z range (20 to 100,000 m/z). Elucidate structural changes, e.g. biotransformation. 9,10 Provide novel research insights, with numerous publications 5 reporting the use of CCS values from protein complexes, to small molecules. More detailed structural characterization The geometry of Triwave Technology provides a range of options to increase information from fragmentation experiments, shown in Table 1. These techniques have been applied successfully to a wide range of applications including, but not limited to: Structural elucidation of small molecules 10,11 and lipids 12,13,14 (TAP fragmentation). Top-down sequencing of proteins. Simultaneous confirmation of site of modification and peptide sequence with ETD or dual-stage CID. TRAP T-Wave IMS T-Wave TRANFER T-Wave - IMS - CID or ETD IMS - - IMS CID CID or ETD IMS CID Table 1. Modes of operation available with the Triwave geometry. One particularly powerful mode of operation is Time Aligned Parallel (TAP) fragmentation, where molecules of interest are analyzed by CID/IMS/CID. This process generates first and second generation product ions with accurate mass measurement in a very rapid and sensitive manner, enabling more detailed and confidence in structural characterization. [On the use of CCS values to elucidate the position of biotransformations on drug candidates] If we had to synthesise just these three metabolites, it would probably have taken several months, whereas the ion mobility LC-MS experiment and modelling calculations is probably around a week s worth of work. Gordon Dear, GSK Chemistry World UK 5

Analyzer Geometry and O p eration Triwave design Principles of T-Wave separation Triwave is the technology that provides ion mobility separations in the SYNAPT family of MS products, shown in Figure 6. It is comprised of three travelling wave devices that are used to guide (trap, accumulate, release, separate, and fragment) ions in a very precise and efficient manner.3 The IMS T-Wave is the region where ion mobility separations are performed and the TRAP and TRANSFER T-Wave regions play an important role in facilitating separation by ion mobility. They can also operate as collision cells which broadens the experimental possibilities using T-Wave IMS. The Triwave design of SYNAPT G2 and G2-S MS instruments provides ion mobility resolution (of 40Ω/ΔΩ) capable of separating small or large molecular species,4 and also ensures the high duty cycle (sensitivity) of the MS instrument is maintained. By combining Triwave with Quantitative Tof15 (QuanTof ) Technology, the SYNAPT G2 and G2-S platforms also provide exact mass data over a wide dynamic range for ion mobility-separated ion species. Ion mobility separation is achieved by driving ions through a neutral buffer gas with an electric field. As ions are driven forward, they interact/collide with the neutral buffer gas, which slows them down in a mobility-dependent manner, causing ions of different size, shape, charge, and mass to transit through a mobility device at different rates. To put it simply, ions with more compact structure transit more quickly through the buffer gas than ions with more open structure. A In Transfer IMS 0.05 mbar 0.5 mbar Ar Figure 6. Schematic showing (a) the Triwave geometry, typical operating pressures and pumping configuration; (b) the SYNAPT G2-S high resolution Mass Spectrometry system, containing Triwave. N2 Gate Trap The mobility separation of ions in a drift tube (the classic form of IMS employed in research MS instrumentation6) and in a T-Wave ion guide both use electric fields to drive the ions through a region of gas. The classical drift tube16 uses a uniform, static electric field/voltage gradient to separate ions, as shown in Figure 7. However, the IMS T-Wave uses non-uniform, moving electric fields/voltage pulses to separate ions, as shown in Figure 7. 0.05 mbar Out Turbomolecular Pump B ANALYTE SPRAY LOCKMASS SPRAY STEPWAV E ION GUIDE QUADRUPOLE T RAP ION MOBILITY SEPARATION HIGH FIELD PUSHER T RANSFER ION MIRROR ION DETECTION SYSTEM HELIUM CELL RY PUMP ROTARY AIR-COOLED TURBOMOLECULAR PUMPS 6 DUAL STAGE REFLECT RON 6

Ring Electrode Pair 1.5 mm (center-to-center) Ring Electrodes Drift Tube/TWIG 5 mm diameter Time 0.5 mm thick Drift Tube Voltage TWIG DC Pulse Figure 7. Electrical fields used for IM separation. Figure 8. Operation of the T-Wave. A T-Wave ion guide, along with typical dimensions, is shown schematically in Figure 8. Opposite phases of RF voltage are applied to adjacent electrodes to provide radial ion confinement and high transmission. The travelling wave DC voltages are applied to electrode pairs in repeating sequence along the device and step to the next adjacent electrode pair at fixed intervals. In this manner, ions are propelled along the device. Due to relaxation of the applied DC voltage in the IMS T-Wave, the voltage profile driving the ions through is not sharp-edged, as in Figure 8, but has a smoothly varying profile, as shown in Figure 9. Figure 9. Profile of electrical field used in T-Wave IM separation. 7

Due to the high gas pressure (typically 2-5 to 3 mbar N 2 ) in the IMS T-Wave, the ion species experience multiple collisions (with the buffer gas molecules) as they are being propelled through by the field. The ability of an ion species to keep up with the wave depends on its mobility. Those species with high mobility surf more on the wave front and are overtaken by the wave less often than those species of low mobility, hence for a packet of ion species entering the cell, mobility separation occurs. Accordingly, ions of high mobility have shorter transit times through the device than those of low mobility. N 2 is used as the IMS gas in most cases because it enables effective separation of molecules of all classes, while the T-Wave speed is typically the main parameter the user needs to optimize the separation. However, the extent of separation can also be improved with the use of different IM gases (e.g. CO 2 ) 17, and a range of other parameters can be utilized for more advanced analysis. T-Wave IMS in practice The IMS T-Wave is present in the Triwave region of all SYNAPT MS instruments. A simplified schematic of the Triwave region is shown in Figure 10. Gate In Trap TWIM Transfer Out Ion species continually enter the TRAP T-Wave region and are prevented from exiting by a potential barrier applied at the gate; they accumulate during the previous mobility separation. There is no Travelling wave applied in the TRAP T-Wave region during ion accumulation. Once the previous mobility separation is complete, the gate is opened for a period of time (by removal of potential barrier) to admit a packet of ions into the IMS T-Wave, in which the ions then separate according to their mobility by the action of the T-Wave in the presence of the buffer gas. Once the mobilityseparated ions exit the IMS T-Wave they are transported through the TRANSFER T-Wave cell with a travelling wave, which maintains the mobility separation, to the ToF analyzer. Once all of the ions have been detected in the ToF analyzer, the gating in of the next packet of ions occurs and the cycle is repeated. This synchronized process ensures that no ions are lost during IMS in order to maintain full duty cycle of the analysis. This is in contrast to many conventional drift tube IMS instruments where there is often no trapping capability and limited ion confinement (resulting in an overall duty cycle which can be as low as 1%). The Triwave design ensures that the IMS separations are carried out with near maximal transmission, therefore providing the benefits of IMS at the limits of MS sensitivity. The helium cell at the entrance of the IMS T-Wave plays an important role by minimizing scatter or fragmentation of ions as the transit from a high vacuum to high pressure environment. 4 Figure 10. Conducting IM separations in the Triwave. 8

The T-Waves that drive the ions through the IMS cell are implemented with a repeating pattern, shown in Figure 11, in order to optimize the speed and extent of the separation. Ion mobility separations are typically performed in the tens of milliseconds timeframe. A B In Helium Out Helium Helium Cell ~90% of Applied Voltage Stacked Ring Ion Guide T-Wave IMS Cell Nitrogen Max. Field 25V/cm (10V applied) Out Nitrogen Out SUMMARY The development of Triwave has transformed the analytical scope and capability of tandem mass spectrometry-based analysis provides a unique opportunity to extend scientific understanding and makes new discoveries possible that were not feasible with other analytical approaches. The orthogonal nature of ion mobility separations provides a significant improvement in analytical selectivity, specificity, and versatility that has been described in over 200 peer reviewed papers from small molecules to peptides and large protein complexes. The rapid and continued development of software tools to automate processing of acquired data will continue to allow more scientists from all backgrounds to access the unique power of Triwave on a routine basis. The experimental capabilities described provide organizations with a new, efficient discovery tool that can help them better compete in a wide range of industries and research disciplines. C 4 Repeat Pattern Figure 11. (a) Schematic of the IMS T-Wave. (b) The applied T-Wave DC voltage in the IMS cell of the SYNAPT G2-S steps to the next adjacent electrode pair in the ion guide and so on along the device, propelling in the direction of motion. (c) The repeat pattern of the T-Wave in the IMS cell of the SYNAPT G2-S. 4 9

Further Reading 1. Waters T-Wave white paper. (www.waters.com) Literature No. 720004177en. 2. Giles K, Pringle SD, Worthington KR, Little D, Wildgoose JL, Bateman RH. Applications of a Travelling Wave-Based Radio-Frequency Only Stacked Ring Ion Guide. Rapid Commun Mass Spectrom. 2004;18(20):2401-14. DOI: 10.1002/rcm.1641 3. Pringle SD, Giles K, Wildgoose JL, Williams JP, Slade SE, Thalassinos K, Bateman RH, Bowers MT, and Scrivens JH. An Investigation of the Mobility Separation of Some Peptide and Protein Using a New Hybrid Quadrupole/ Traveling Wave IMS/oa-ToF Instrument. International Journal of Mass Spectrometry. 2007 Mar 1;261(1):1-12. DOI: 10.1016/j.ijms.2006.07.021 4. Giles K, Williams JP, Campuzano I. Enhancements in Travelling Wave Ion Mobility Resolution. Rapid Commun Mass Spectrom. 2011 Jun 15;25(11):1559-66. DOI: 10.1002/rcm.5013 5. Waters HDMS peer-reviewed reference list. (www.waters.com) Literature No. 720002633en. 6. Unparalleled Specificity Using HDMS Technology in a MALDI Imaging Experiment. Waters Technical Brief (www.waters.com) Literature No. 720004259en. 7. Rodríguez-Suárez E, Hughes C, Gethings L, Giles, K, Wildgoose J, Stapels M, Fadgen KE, Geromanos SJ, Vissers JP, Elortza F, Langridge JI: An Ion Mobility Assisted Data Independent LC-MS Strategy for the Analysis of Complex Biological Samples, Current. Anal. Chem. special issue: Ion Mobility Spectrometry: Using Size and Shape to Understand Real-World Systems at the Molecular Level, HT-SBJ-CAC-0005. 8. Scarff CA, Thalassinos K, Hilton GR, Scrivens JH. Travelling Wave Ion Mobility Mass Spectrometry Studies of Protein Structure: Biological Significance and Comparison with X-Ray Crystallography and Nuclear Magnetic Resonance Spectroscopy Measurements. Rapid Commun Mass Spectrom. 2008 Oct;22(20):3297-304. DOI: 10.1002/rcm.3737 9. Cuyckens F, Wassvik C, Mortishire-Smith RJ, Tresadern G, Campuzano I, Claereboudt J. Product Ion Mobility as a Promising Tool for Assignment of Positional Isomers of Drug Metabolites. Rapid Commun Mass Spectrom. 2011 Dec 15;25(23):3497-503. DOI: 10.1002/rcm.5258 10. Dear GJ, Munoz-Muriedas J, Beaumont C, Roberts A, Kirk J, Williams JP, Campuzano I. Sites of Metabolic Substitution: Investigating Metabolite Structures Utilising Ion Mobility and Molecular Modelling. Rapid Commun Mass Spectrom. 2010 Nov 15;24(21):3157-62. DOI: 10.1002/rcm.4742 11. Chan EC, New LS, Yap CW, Goh LT. Pharmaceutical Metabolite Profiling Using Quadrupole/Ion Mobility Spectrometry/Time-of-Flight Mass Spectrometry. Rapid Commun Mass Spectrom. 2009 Feb;23(3):384-94. DOI: 10.1002/ rcm.3887 12. Castro-Perez J, Roddy TP, Nibbering NM, Shah V, McLaren DG, Previs S, Attygalle AB, Herath K, Chen Z, Wang SP, Mitnaul L, Hubbard BK, Vreeken RJ, Johns DG, Hankemeier T. Localization of Fatty Acyl and Double Bond Positions in Phosphatidylcholines Using a Dual Stage CID Fragmentation Coupled with Ion Mobility Mass Spectrometry. J Am Soc Mass Spectrom. 2011 Sep;22(9):1552-67. Epub 2011 Jun 24. DOI: 10.1007/s13361-011-0172-2 13. Kim H. Structural Characterization of Unsaturated Phospholipids Using Traveling Wave Ion Mobility Spectrometry. DOI: 10.1007/978-1-4419-7601-7_6 14. Olivova P, Chen W, Chakraborty AB, Gebler JC. Determination of N-Glycosylation Sites and Site Heterogeneity in a Monoclonal Antibody by Electrospray Quadrupole Ion-Mobility Ttime-of-Flight Mass Spectrometry. Rapid Commun Mass Spectrom. 2008;22(1):29-40. DOI: 10.1002/rcm.3330 15. Waters Application Note. SYNAPT G2: Breakthrough Quantitative and Qualitative Performance for UPLC/MS and MS/MS (MS E ) Applications. Literature No. 720003057en. 16. Kanu AB, Dwivedi P, Tam M, Matz L, Hill HH Jr. Ion Mobility-Mass Spectrometry. J Mass Spectrom. 2008 Jan;43(1):1-22. DOI: 10.1002/jms.1383 17. Waters Application Note. SYNAPT HDMS: Improving Ion Mobility Separation by Increasing Drift-Gas Polarizability. Literature No. 720003201en. Find further reading and animations online: www.waters.com/wavedevices Waters, SYNAPT, UPLC, UltraPerformance LC, and Triwave are registered trademarks of Waters Corporation. The Science of What s Possible, T-Wave, and DriftScope are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 2012 Waters Corporation. Produced in the U.S.A. April 2012 720004176EN LB-PDF Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com