Partitioning. Separation is based on the analyte s relative solubility between two liquid phases or a liquid and solid.

Similar documents
Chromatography. Gas Chromatography

HPLC Background Chem 250 F 2008 Page 1 of 24

Instrumental Chemical Analysis

Course goals: Course goals: Lecture 1 A brief introduction to chromatography. AM Quality parameters and optimization in Chromatography

Chapter 27: Gas Chromatography

Introduction to Gas Chromatography

Introduction to Chromatography

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

CHAPTER 6 GAS CHROMATOGRAPHY

Chromatographic Analysis

Chapter 31 Gas Chromatography. Carrier Gas System

Introduction to Chromatographic Separations

High Performance Liquid Chromatography

High Performance Liquid Chromatography

Open Column Chromatography, GC, TLC, and HPLC

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

Gas Chromatography. Introduction

GC Instruments. GC Instruments - Sample Introduction

LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction.

CHEM340 Tutorial 4: Chromatography

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

GC Instruments. GC Instruments - Columns

Gas Chromatography (GC)

Instrumental Analysis II Course Code: CH3109. Chromatographic &Thermal Methods of Analysis Part 1: General Introduction. Prof. Tarek A.

Liquid Chromatography

10/27/10. Chapter 27. Injector typically 50 C hotter than oven

Understanding Gas Chromatography

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications

GAS CHROMATOGRAPHY. Mobile phase is a gas! Stationary phase could be anything but a gas

Gas Chromatography. Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid

2401 Gas (liquid) Chromatography

Analytical Chemistry

Luminescence transitions. Fluorescence spectroscopy

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011


Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

Chromatographic Separation

High Performance Liquid Chromatography

Chemistry Instrumental Analysis Lecture 27. Chem 4631

CHROMATOGRAPHY. The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments.

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

Volatile organic compounds (VOCs):

CHEM 429 / 529 Chemical Separation Techniques

High Pressure/Performance Liquid Chromatography (HPLC)

High Performance Liquid Chromatography

HPLC. High Performance Liquid Chromatography (HPLC) Harris Chapter 25

Introduction to Capillary GC

GAS CHROMATOGRAPHY (GC)

2. a) R N and L N so R L or L R 2.

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

Selection of a Capillary

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers.

Chapter 11 Conventional Gas Chromatography

Chromatography. Chromatography is a combination of two words; * Chromo Meaning color * Graphy representation of something on paper (writing)

Chromatography and other Separation Methods

Introduction and Principles of Gas Chromatography

Gas Chromatography. Chromatography Laboratory Course. Dr. Christian Jungnickel Chromatography Course GC September 2005

CEE 772: Instrumental Methods in Environmental Analysis

LIQUID CHROMATOGRAPHY

for Acclaim Mixed-Mode HILIC-1 Column

Gas Chromatography. Rosa Yu, David Reckhow CEE772 Instrumental Methods in Environmental Analysis CEE 772 #16 2

Disadvantage: Destructive Technique once analyzed by GC, the sample is lost

Chromatography- Separation of mixtures CHEM 212. What is solvent extraction and what is it commonly used for?

Ch.28 HPLC. Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography)

Gas chromatography. Advantages of GC. Disadvantages of GC

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

Selection of a Capillary GC Column

Biochemistry. Biochemical Techniques. 12 Gas Liquid Chromatography

How To Select the Correct GC Column. Simon Jones Application Engineer

Introduction to Chromatographic Separations (Chapter 1) Many determinations involve separation followed by analysis chromatography electrophoresis

Principles of Gas- Chromatography (GC)

Course CHEM Chromatography

Chapter 1. Chromatography. Abdul Muttaleb Jaber

M > ACN > > THF

LC III: HPLC. Originally referred to as High-Pressure Liquid Chromatography. Now more commonly called High Performance Liquid Chromatography

Chemistry 311: Instrumental Analysis Topic 4: Basic Chromatography. Chemistry 311: Instrumental Analysis Topic 4: Basic Chromatography

Skoog/Holler/Crouch Chapter 26 Principles of Instrumental Analysis, 6th ed. CHAPTER 26

G a s C h r o m a t o g r a p h y Determination of Ethanol in Wine by Head-Space Gas Chromatography

Chapter content. Reference

Chromatography Outline

What is Chromatography?

Gas Chromatography (Chapter 2 and 3 in The essence of chromatography)

ERT320 BIOSEPARATION ENGINEERING CHROMATOGRAPHY

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Trajan SGE GC Columns

Chemistry Instrumental Analysis Lecture 28. Chem 4631

Biochemistry. Biochemical Techniques HPLC

INSTITUTE OF MEDICAL BIOCHEMISTRY FIRST FACULTY OF MEDICINE, CHARLES UNIVERSITY IN PRAGUE. Chromatography. in biochemistry

Chemistry Gas Chromatography: Separation of Volatile Organics

Speakers. Moderator. John V Hinshaw GC Dept. Dean CHROMacademy. Tony Taylor Technical Director CHROMacademy. Dave Walsh Editor In Chief LCGC Magazine

Chromatography. writing in color

Introduction to Capillary GC


Theory and Instrumentation of GC. Chromatographic Parameters

Separation Methods Based on Distributions in Discrete Stages (02/04/15)

LC Technical Information

Gas Chromatography notes

Understanding the Capillary GC Column: How to Choose the Correct Type and Dimension

Transcription:

Chromatography Various techniques for the separation of complex mixtures that rely on the differential affinities of substances for a gas or liquid mobile medium and for a stationary adsorbing medium through which they pass.

Partitioning Separation is based on the analyte s relative solubility between two liquid phases or a liquid and solid Mobile Phase Stationary Phase Solvent or gas Bonded Phase Solid or liquid

Separation The individual components are retained by the stationary phase differently. The components separate from each other since they are running at different speeds through the column with the eluent [gas or solvent(s)]. At the end of the column they elute one at a time depending on their retention governed by their polarity.

Separation Boiling point- lower bp compounds spend more time in gas phase. Temperature of column does not have to be above boiling pt. Solids have vapor pressure. High vapor pressure liquids used as solvents (water 25 mmhg whereas ether 520 mmhg/25c)

Separation Polarity like absorbs likes. Polar compounds have longer retention time on Polar columns. Non-polar compounds have longer retention on non-polar column. Two types of columns in GC3800, a silicone based column and a PEG column.

Separation Carrier gas flow- high flow rate decreases retention time. High gas flow may cause poor separation since components have little time to react with stationary phase. There is optimum gas flow for various columns. Too high a flow causes excessive back pressure.

Separation Column length- longer column usually improves resolution, but increases back pressure. Doubling length will not double resolution (resolution increases according to square root of length). Generally a 30 meter column gives best resolution, analysis time and head pressure

Separation An effective means to increase resolution is to decrease column ID. The efficiency of a capillary column increases (number of theoretical plates per meter) as the ID of column decrease. Makes sharper peaks,& decreases column bleed Decreasing ID decreases sample capacity. Ideal (most popular) ID is 0.25 mm.

Some Types of Chromatography Lid Paper chromatography (PC), Thin-layer chromatography (TLC), Paper Solvent front solvent Two experiments in MCAL demonstrate: Liquid chromatography (LC, including highperformance liquid chromatography, or HPLC), Gas chromatography (GC).

GC Experiment 1) To compare the separation of alcohols of increasing carbon number using a carbowax column and a FID detector in the Varian 3900 GC. 2) To compare separation of various compounds on a carbowax versus a silicone based column. 3) To measured sensitivity of instrument (LOD).

Gas Chromatography (a review) Chromatographic technique that can be used to separate volatile organic compounds. A gas chromatograph consists of a flowing mobile phase, an injection port, a separation column containing the stationary phase, detector and a controller (integrator) and data collection and storage. Organic compounds are separated due to differences in their partitioning behavior between the mobile gas phase and the stationary phase in the column.

Bruker Improving the Quality of our Beer New 456-GC Off-Flavor Beer Analyzers

Gas tanks He, N 2, H 2, Air Varian (Agilent) 3800 GC Detectors Autosampler Injectors Oven Computer control and data acquisition

Stationary Phases The most common stationary phases in gaschromatography columns are polysiloxanes, which contain various substituent groups to change the polarity of the phase. The nonpolar end of the spectrum is polydimethyl siloxane, which can be made more polar by increasing the percentage of phenyl groups on the polymer. Polyethylene glycol (carbowax) is commonly used as the stationary phase for more polar analytes.

Polysiloxanes

Polyethylene glycol

Low Bleed Phases (Arlenes)

Varian (Agilent) Factor Four Capillary GC Columns. Capillary columns are a thin fused-silica capillary. Typically 10-100 m in length and 250 µm inner diameter. The stationary phase is coated on the inner surface.

Capillary Columns

FSOT Column

Capillary Columns Capillary columns provide much higher separation efficiency than packed columns but are more easily overloaded by too much sample. You need to dilute your sample to get good symmetrical (Gaussian) peaks Solvent peak dominates chromatogram due to normalization of peaks.

GC Column Oven

MCAL GC Columns GC 3800- two columns FID and ECD detector: 1) CP-sil 8 50 m, 0.2mm,.33um 5% phenyl 95% dimethyl polysiloxane 2) CP-sil 5CB 100% dimethyl polysiloxane 15 m 0.25mm 0.25um GC 3900 one column FID detector: 1) Supelco wax-10 15m x 0.32 mm, 0.5 u Polyetyylene glycol

Simple GC Gas control valve Sample injector port Computer FID, ECD, MS Detector Gas is He, N 2 or H 2 Separation Column Temperature controlled oven

Sequence of GC Injection 1. A small amount of liquid (microliters) is injected through a silicon rubber septum into the heated (>200 o C) GC injector that is lined with an inert glass tube. 2. The sample is immediately vaporized. 3. A pressurized, inert, carrier gas-which is continually flowing from a gas regulator through the injector and into the GC column-sweeps the gaseous sample, solvent, and analyte, onto the column. 4. Septum purge: a small ancillary flow of carrier gas bathes the underside of the injector's septum so that hot vaporized sample gases can't interact and possibly stick to the septum. This improves peak shape and reproducibility.

Split / Splitless Injector Helium

Split / Splitless Injector

ECD detector FID Detector Injectors

FID Detector Organic compounds burning in a hydrogenoxygen flame produce ions and electrons. These charged particles created in the combustion process create a current between the detector's electrodes. One electrode is the metallic jet itself, the other is above the jet. The detector housing is heated so that gases produced by the combustion (mainly water) do not condense in the detector before leaving the detector chimney.

Insulator FID Detector Cathode Anode Air Hydrogen Sample plus Helium from Column

FID Detector

The ECD or Electron Capture Detector The ECD or electron capture detector measures electron capturing compounds (usually halogenated) by creating an electrical field in which molecules exiting a GC column can be detected by the drop in current in the field

ECD Detector

ECD Detector Electrode insulator waste + electrode -- - - - ----- ----- Radioactive Beta emitter (electrons) - electrode GC column Electron cloud insulator

Separation of Alcohols C1 to C8 on carbowax column

GC Exp: ECD Detector Dichlorobenzene Hexane Solvent??

GC Exp: Fid Detector Solvent and contamination dichlorobenzene

GC Exp: FID Detector Expanded dichlorobenzene

HPLC in MCAL Solvents rescence ctor Autosampler Manual injector V Vis iode array etector Pumps

HPLC Image System control- Data storage HPLC column (packing) Injector Autosampler Sample Solvent(s) Pump(s) Detector(s) (Diode Array & Fluorescence) Waste

Separation in HPLC Seperation depends on : Polarity of molecule RP verses NP, Hilic, Hydrophilic-Interaction Chrpmatography Hic, Hydrophobic-interaction Chromatography Electrical charge (ion exchange). Anion exchanger attracts charged analyte Cation exchanger attracts + charged analyte Molecular size Size exclusion gel permeation or size exclusion

Columns - Heart of System Solid Support (column packing) - Backbone for bonded phases. Usually 10µ, 5µ, 3µ, 2.5µ siloxane or polymeric particles. Porous or solid core Packing has different degrees of polarity depending on functional groups attached Bonded Phases - Functional groups chemically bound to the solid support.

HPLC - Modes Normal Phase - Polar stationary phase and nonpolar solvent. Reverse Phase (used for caffeine and quinine separation) - Non-polar stationary phase and a polar solvent.

Normal Phase versus Reverse Phase HPLC Packing Polar packing Non-polar packing C4 C8 C18

HPLC Solid Support C18RP Si O CH 3 Si CH 2 CH 2 CH 2 CH 2 (CH 2 ) 14 O O CH 3 Si OH O O Si O CH 3 Si (CH 2 ) 18 Gel Surface CH 3 Bonded phase

HPLC Solid Support C18RP End Capping Si O CH 3 Si CH 2 CH 2 CH 2 CH 2 (CH 2 ) 14 O O CH 3 Si OH + Si (CH 2 ) 2 O O Si O CH 3 Si (CH 2 ) 18 Gel Surface CH 3 Bonded phase

HPLC Solid Support C18RP Polymeric Si O CH 3 Si CH 2 CH 2 CH 2 CH 2 (CH 2 ) 14 O O CH 3 Si O Si (CH 2 ) 18 O O Si O CH 3 Si (CH 2 ) 18 Gel Surface CH 3 Bonded phase

HPLC Solid Support C18RP Polymeric Si O OH Si CH 2 CH 2 CH 2 CH 2 (CH 2 ) 14 O O O Si O Si (CH 2 ) 18 O O Si O CH 3 Si (CH 2 ) 18 Gel Surface CH 3 Bonded phase

Solid Core Particle Solid Core Porous Outer Layer

Core-Shell Particle

Core-Shell Particle

HPLC Columns (MCAL) C18 Microsorb (Varian)5 u 100 A 250 x 4.5 mm C18 Polar Synergi (Phenomenex) 4 u 80 A 250 x 4.5 mm C18 RP Particil column (Varian) 3 u 300 A 100 x 2.0 mm C18 Hilic column (Phenomenex) 5 u 200 A 150 x 4.6 mm XB-C18 RP Kinetex (Phenomenex) 2.6 u 100 A 100 x 4.6 mm C18 Microsorb (Varian) 5 u 100 A 250 x 4.5 C8 Polaris (Varian) 3 u A 50 x 20 mm

nual ctor HPLC columns

Solvents for RP H 2 O most polar with decreasing polarity as solvent B added to mix. Solvent B usually MeOH, acetonitrile, iso-propanol Can be isocratic mixture - 1 or 2 pump system or Gradient need at least two pumps and mixing apparatus. Caffeine-quinine exp. uses buffer-acetonitrile or buffer- methanol gradient mixing.

The Theoretical Plate Model of Chromatography The plate model supposes that the chromatographic column is contains a large number of separate layers, called theoretical plates (N). Separate equilibrations of the sample between the stationary and mobile phase occur in these "plates". The analyte moves down the column by transfer of equilibrated mobile phase from one plate to the next.

Chromatographic Resolution Depends On: 1.Column efficiency (N) peak width Particle diameter Flow rate Sol. Viscosity Temp. Col. Length Col. diameter 2. Solvent efficiency (α) retention time Stationary phase Mobile phase Solvent composition Temperature

Column Efficiency -Theoretical plate numbers (N) Theoretical plate numbers are indirect measure of peak width for a peak at a specific retention time. Columns with high plate numbers are considered to be more efficient, that is, have higher column efficiency, than columns with a lower plate count. A column with a high number of theoretical plates will have a narrower peak at a given retention time than a column with a lower N number. N=5.545(t r /w h ) 2

Column Efficiency -Height equivalent to theoretical plates Measure of the resolving power of the column HETP = L/N L = Length of column N = Theoretical plates At each theoretical plate one equilibrium distribution of solute between mobile phase and stationary phase occurs.

Peak Seperation (α) On stationary phases where the alphas (α) are small, more efficient columns (higher N) are beneficial. Column efficiency is a function of: a) the column dimensions (diameter, length and film thickness), b) the type of carrier solvent and its flow rate or average linear velocity, and c) the compound and its retention.

Band broadening theory (Van Deemter equation) column band broadening originates from three main sources: multiple path of an analyte through the column packing; molecular diffusion; effect of mass transfer between phases.

Multiple path The velocity of mobile phase in the column may vary significantly across the column diameter, depending on the particle shape, porosity, and the whole bed structure.

Measure of the Resolving Power of the Column HETP = A + B / u + C u A = eddy diffusion take different paths (different lengths). B = longitudinal diffusion analyte diffuses from centre to outer edge. C = resistance to mass transfer: analyte strong affinity for stationary and veloscity of mobile phase is high. U = avg veloscity of mobile phase

Resolving Power α N

Column efficiency is reported as plates per meter (N/Meter). Smaller average packing particle size = Larger N Broader particle size distribution = Smaller N Better packing procedures = Larger N

Polarity - Elotropic Series Water is at the polar end of mobile-phasesolvent scale, Hexane, an aliphatic hydrocarbon, is at the non-polar end. In between, single solvents, as well as miscible-solvent mixtures

Stationary Phase Polarity Spectrum Silica has an active, hydrophilic [waterloving] surface containing acidic silanol [silicon-containing analog of alcohol] The polarity of the silica surface may be modified selectively by chemically bonding to it less polar functional group like n-octylsilyl- [C8], and n-octadecylsilyl- [C18, ODS] moieties. The latter is a hydrophobic [water-hating], very non-polar packing.

HPLC Experiment Seperation and identification of caffeine, quinine and added theophylline (internal standard) in commercial drinks Comparison of different quantitation methods (external standard, internal standard) Uv verses fluorescent detection Diode array detection

HPLC Experiment The separation is done on a C18 RP column. The solvent is a 20% methanol in water (low ph with TEA added) increased to 30% methanol. The caffeine comes off column before the quinine.

HPLC Exp: Caffeine and Quinine Quinine Rt = 11.6 min Caffeine wavelength 206 Rt = 9.2 min Caffeine wavelength 273 Rt = 9.2 min

Diode Array Detector Caffeine

Detection with HPLC Ultraviolet light source (deuterium lamp 190 to 950nm) Visible light source (quartz iodide 350 nm and beyond) Flurescent light (Xenon lamp) Excitation and emmission wavelength selection

Concentration Of Analytes Concentration is determined by measuring the area of known concentrations of caffeine and quinine, and making a standard curve. Concentration of caffeine is also determined by adding an internal standard (IS) and calculating caffeine based on the response factor of the IS.