Data report for elemental analysis of IMPROVE samples collected during OCTOBER, NOVEMBER, DECEMBER 2005 UC Davis SUMMARY

Similar documents
Data report for elemental analysis of IMPROVE samples collected during April, May, June 2006 UC Davis - Submitted May 14, 2008 SUMMARY

Data report for elemental analysis of IMPROVE samples collected during April, May, June 2009 UC Davis Submitted June 18, 2010 SUMMARY

Quantitative XRF Analysis. algorithms and their practical use

Overview of X-Ray Fluorescence Analysis

DOCUMENT HISTORY. Initials Section/s Modified Brief Description of Modifications

ARTICLE IN PRESS. Journal of Aerosol Science

FINDING A NEEDLE IN A HAYSTACK: PERFORMANCE EVALUATION OF PORTABLE XRF INSTRUMENTS FROM THREE MANUFACTURERS

REGIONAL TRAINING COURSE ON METHODS AND TOOLS TO IDENTIFY SOURCES OF AIR POLLUTION

LAB REPORT ON XRF OF POTTERY SAMPLES By BIJOY KRISHNA HALDER Mohammad Arif Ishtiaque Shuvo Jie Hong

X-Ray Fluorescence and Natural History

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF

WHEN IS IT EVER GOING TO RAIN? Table of Average Annual Rainfall and Rainfall For Selected Arizona Cities

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

ENGINE SERIAL NUMBERS

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

GAMINGRE 8/1/ of 7

Chiang Rai Province CC Threat overview AAS1109 Mekong ARCC

AMMS-100 Atmospheric Metal Monitoring System

Worldwide Open Proficiency Test for X Ray Fluorescence Laboratories PTXRFIAEA13. Determination of Major, Minor and Trace Elements in a Clay Sample

Altitude influence of elemental distribution in grass from Rila mountain. Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory

Mountain View Community Shuttle Monthly Operations Report

Inspection Documents for Type 2.1 Coverage:

Inspection Documents for Type 2.1 Coverage:

GTR # VLTs GTR/VLT/Day %Δ:

In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer*

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

A. Identification of Temperature as the Cause of Non-Attainment

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS

Location. Datum. Survey. information. Etrometa. Step Gauge. Description. relative to Herne Bay is -2.72m. The site new level.

Jackson County 2013 Weather Data

Location. Datum. Survey. information. Etrometa. Step Gauge. Description. relative to Herne Bay is -2.72m. The site new level.

MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question.

Scarborough Tide Gauge

Technical note on seasonal adjustment for M0

Inspection Documents for Type 2.1 Coverage:

P7.7 A CLIMATOLOGICAL STUDY OF CLOUD TO GROUND LIGHTNING STRIKES IN THE VICINITY OF KENNEDY SPACE CENTER, FLORIDA

Inspection Documents for Type 2.1 Coverage:

XUV 773: X-Ray Fluorescence Analysis of Gemstones

Reporting Category 1: Matter and Energy

Chemistry 31A Autumn 2004 Professors Chidsey & Zare Exam 2 Name:

Inspection Documents for Type 2.1 Coverage:

Unit 1 Part 2 Atomic Structure and The Periodic Table Introduction to the Periodic Table UNIT 1 ATOMIC STRUCTURE AND THE PERIODIC TABLE

Enhancements to the CTBTO IDC radionuclide processing pipeline for particulate samples achieving significant improvement of automatic products

Monthly Magnetic Bulletin

Reporting Category 1: Matter and Energy

ATLAS Tile Calorimeter Calibration and Monitoring Systems

Microsoft Excel Directions

Inspection Documents for Type 2.1 Coverage:

Wind Resource Data Summary Cotal Area, Guam Data Summary and Transmittal for December 2011

Introduction to X-ray Photoelectron Spectroscopy (XPS) XPS which makes use of the photoelectric effect, was developed in the mid-1960

The European Activation File: EAF-2005 decay data library

Proposed Procedures for Determining the Method Detection Limit and Minimum Level

Operation of the Coupled Cyclotron Facility at Michigan State University

Three hour lab. Chem : Sept Experiment 2 Session 2. Preparation Pre-lab prep and reading for E2, Parts 3-5

Creating Empirical Calibrations

Name: SCH3U Worksheet-Trends

Long-term Water Quality Monitoring in Estero Bay

Pre-Calc Chapter 1 Sample Test. D) slope: 3 4

Three hour lab. Chem : Feb Experiment 2 Session 2. Experiment 2 Session 2 Electrons and Solution Color

Page 1 of 9. Website: Mobile:

Give the number of protons, neutrons and electrons in this atom of aluminium. Why is aluminium positioned in Group 3 of the periodic table? ...

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

Course theme. Three hours of lab Complete E1 (Parts 1, 2, 3, 4, and 5B) Prepare discussion presentation Prepare team report.

Three hours of lab Complete E1 (Parts 1, 2, 3, 4, and 5B) Prepare discussion presentation Prepare team report. Course theme

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017

Latest advances in identifying mineral composition variation by the M4 TORNADO AMICS

Life Cycle of Convective Systems over Western Colombia

CHARACTERIZATION OF Pu-CONTAINING PARTICLES BY X-RAY MICROFLUORESCENCE

X-ray Absorption Spectroscopy

Horizontal rows are called periods. There are seven periods on the periodic chart.

Chapter 3: Stoichiometry

Champaign-Urbana 1999 Annual Weather Summary

FUNDAMENTAL PARAMETERS ANALYSIS OF ROHS ELEMENTS IN PLASTICS

Chapter 1. I- Fill the following table. Element symbol and the mass no. n p n n n e. number. II - Choose the correct answer for the following: Ca-40

Honors Chemistry - Unit 4 Bonding Part I

Published by ASX Settlement Pty Limited A.B.N Settlement Calendar for ASX Cash Market Products

DAILY QUESTIONS 28 TH JUNE 18 REASONING - CALENDAR

Introduction to LIBS COMMUNITY USER WORKSHOP ON PLANETARY LIBS (CHEMCAM) DATA. Sam Clegg and the ChemCam team

Analysis of Soil and Sewage Sludge in the Field with a Portable ED-XRF Spectrometer

SYSTEM BRIEF DAILY SUMMARY

ORBITAL DIAGRAM - A graphical representation of the quantum number "map" of electrons around an atom.

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Sampling, Storage and Pre-Treatment Techniques

Principles of Mathematics 12: Explained!

Chemistry 1 First Lecture Exam Fall Abbasi Khajo Levine Mathias Mathias/Ortiz Metlitsky Rahi Sanchez-Delgado Vasserman

Monthly Magnetic Bulletin

WIND DATA REPORT. Vinalhaven

INCREASING TREND IN DUST CLOUD INTRUSIONS FROM THE SAHARA OVER ISRAEL. Department of Geophysics and Planetary Sciences, Tel Aviv University

SYSTEM BRIEF DAILY SUMMARY

Spectroscopy on Mars!

KING EDWARD POINT OBSERVATORY MAGNETIC DATA

Electron probe microanalysis - Electron microprobe analysis EPMA (EMPA) What s EPMA all about? What can you learn?

Astrological Calendar. for Central Time

4.1 Atomic structure and the periodic table. GCSE Chemistry

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

Halogens HALOGENS. Parts 2A and 2B. Chem : Feb. 19, 20 and March 3. Compare the properties and reactivity of the halogens and halides

Mendeleev arranged the elements in order of their atomic mass (atomic weight).

Calculations Equation of Time. EQUATION OF TIME = apparent solar time - mean solar time

High-resolution study of Gamow- Teller transitions in pf-shell nuclei. Tatsuya ADACHI

Transcription:

Data report for elemental analysis of IMPROVE samples collected during OCTOBER, NOVEMBER, DECEMBER 2005 UC Davis SUMMARY This report summarizes the quality assurance performed during elemental analysis of the IMPROVE samples collected in October, November and December of 2005. The elemental analyses include the determination of most elements with atomic numbers from to 26 (Na-Fe) by energy dispersive X-ray fluorescence (XRF) with a Cu-anode system, most elements from 27 to 40 (Ni-Zr) and 82 (Pb) by XRF with a Mo-anode system, and hydrogen by Proton Elastic Scattering Analysis (PESA) with the Crocker cyclotron. The following data assessments and quality controls are obtained for all analyses: Concentration calibration and verification (calibration check) Energy calibration Laboratory replicates (reanalysis) Systems comparison Field blanks The procedures and their results are presented below. Two significant changes were introduced in the XRF analysis of October - December 2005 samples. One was the placing into operation of an additional Cu-anode system, as described in a data advisory. The added system has already increased analytical throughput. The second change was to the calibration procedures for the Mo-anode system ; the current used for single-element foils was raised from 2.6 ma to 0 ma to better match the livetimes encountered with actual samples. Section. Overview of Elemental Analysis Systems The elements Na and Mg (considered qualitative only) and Al to Fe are reported from an XRF system with a Cu-anode grounded X-ray tube. This system operates under vacuum (< 350 microns Hg). There are now two Cu systems, which will be distinguished hereafter as Cu-Vac and Cu-Vac2. Data acceptance is based solely on each system s performance for single-element calibration foils; as the data advisory noted, the systems exhibit some detectable differences for actual samples. All July 2005 samples were analyzed on both systems, with results that are shown in Attachment. Cu-Vac was used to analyze the samples from November 2005, and Cu-Vac2 was used to analyze the samples from October 2005 and December 2005. All October-December 2005 samples were analyzed for 000 seconds at 0 ma and 20 kv (default settings for sample analysis). The elements Ni to Zr and Pb are reported from a similar system with a Mo-anode grounded X- ray tube operating in ambient air. Samples were analyzed for 000 seconds at 23 ma and 35 kv (default settings for ). The PESA system operates under vacuum (< 0 microns Hg) and uses a proton beam (4.5 MeV H+) from the Crocker cyclotron to quantify the concentration of hydrogen (H). Samples were analyzed for 5 seconds, with an average typical current value of approximately 50 na collected on a Faraday cup.

Section 2. General Statistics of October, November and December 2005 data XRF and PESA analyses were carried out on 85 samples collected in October 2005, 637 samples collected in November 2005 and 688 samples collected in December 2005. All samples collected in the forth quarter of 2005 were analyzed between 6 April 2006 and 29 June 2006 on the Mo-anode XRF system, between 4 December 2006 and 2 January 2007 on the Cuanode XRF system (Cu-Vac), between 2 November 2006 and 7 January 2007 on Cu-Vac2, and on December 20 th, 2006 and from January 6 th to January 8 th, 2007 on the PESA system. Table summarizes the forth quarter 2005 detection rates on the three systems, with rates for September 2005 included for comparison. PESA Z element 0-2005 -2005 2-2005 9-2005 H 99% 99% 99% 00% Cu-anode XRF Z element 0-2005 -2005 2-2005 9-2005 Na 56% 3% 62% 40% 2 Mg 37% 7% 3% 20% 3 Al 76% 6% 65% 78% 4 Si 96% 93% 92% 94% 5 P 3% % 3% % 6 S 99% 00% 00% 00% 7 Cl 4% 2% 5% 9% 9 K 99% 00% 99% 00% 20 Ca 98% 00% 99% 00% 22 Ti 96% 95% 92% 98% 23 V 88% 8% 78% 94% 24 Cr 4% 37% 50% 42% 25 Mn 95% 94% 93% 98% 26 Fe 99% 00% 00% 00% Mo-anode XRF Z element 0-2005 -2005 2-2005 9-2005 28 Ni 44% 4% 46% 62% 29 Cu 8% 78% 76% 94% 30 Zn 99% 00% 99% 00% 33 As 65% 63% 46% 27% 34 Se 76% 64% 75% 90% 35 Br 99% 99% 99% 00% 37 Rb 54% 52% 68% 63% 38 Sr 9% 92% 9% 94% 40 Zr 4% 2% 6% 33% 82 Pb 99% 99% 94% 99% Table. Percentage of cases in which the element was detected on each system. September 2005 data included for reference.

Section 3. Quality Control 3. Concentration calibration and verification (calibration checks) Both XRF systems are calibrated with thin film foil standards produced by Micromatter. The standards used for samples from the forth quarter of 2005 are listed below in Table 2. Because their concentrations are relatively high, standards are analyzed at reduced X-ray tube current (2.6 ma on XRF-Cu systems and 0mA on XRF-Mo system) to maintain counting live times comparable with those of actual IMPROVE samples. Standard Certified Elemental Concentrations Serial # +/- 5% (µg/cm 2 ) NaCl Na: 9., Cl: 29.4 658 MgF 2 Mg: 20.6 659 Al Al: 40.7 6520 SiO Si: 23.9 652 GaP* P: 4.5 6500 CuS x S: 2.9 Cu: 37.6 6523 KCl Cl: 22.5 K: 24.9 6296 CaF 2 Ca: 24.9 6525 Ti Ti: 3.7 6504 V V: 2.2 6505 Cr Cr: 5.8 6507 Mn Mn: 4.6 6506 Fe Fe: 4.7 6508 Ni Ni: 0.5 6509 Cu Cu: 2.4 650 ZnTe* Zn: 5.2 65 GaAs* Ga: 8 As: 8.7 652 Se Se: 2.9 653 CsBr Br: 5. 654 RbI Rb: 5.7 655 SrF 2 Sr: 0.9 656 Pb Pb La: 6 Pb Lb: 6 657 Table 2. Micromatter standard foils used for all analyses..some standards (*) have variable stoichiometry; they are not use directly in calibration of the systems but serve only as indicators. Spectra from the foil standards are processed and analyzed by the same software used for samples. Calibration factors relating spectral counts to elemental concentrations are determined from the ratio of an element s observed peak area to the concentration quoted by Micromatter. For elements such as Cl, which are found in more than one standard, calibration factors are based on the average of the several ratios. During the time of Oct-Dec 2005 samples analyses, the performance of the Mo XRF system was monitored by weekly calibration checks, and the performance of both Cu XRF systems was monitored once or twice a month. Fifteen standards are analyzed on the Cu system and 9 on the Mo system, and the ratios of reported to quoted values are calculated. If the ratios lie within the acceptance limits 0.9. for all elements, then the system is considered stable and the existing calibration factors continue to be used. Deviations beyond 0% for the quantitative reported elements trigger an investigation of the problem and possible system recalibration. After a recalibration, all samples analyzed since the last successful calibration verification are reanalyzed with the new calibration factors.

Figures to 3 below show the calibration checks and system recalibrations performed during the period in which samples of interest were analyzed on each system. The analysis dates for each sample month are listed in the legends. September 2005 samples analysis dates are included for reference. The y-axes indicate the ratio of the values reported for each standard to the value quoted by Micromatter. Ratio Found/Quoted Standard..05 0.95 0.9 Ga Cu S K Cl Ca Ti V Cr Mn Fe Fe Ni Cu Zn Ga As Se Br Rb Sr Pb La Pb Lb Sep 2005 3/2/06-4/5/06 Oct 2005 4/6/06-5/4/06 Nov 2005 5/4/06-5/26/06 Dec 2005 6//06-6/29/06 0.85 0.8 3//06 4//06 5//06 6//06 7//06 Figure. Mo XRF system performance chart based on standards. Na Cl Mg Al Si P S Cl K Ca Ti V Cr Mn Fe Ni new cal new cal new cal new cal new cal Ratio Found/Quoted Standard.05 0.95 0.9 0.85 Nov 2005 2/4/06 - /2/07 0.8 //2006 2//2006 //2007 Figure 2. XRF Cu-Vac system performance chart based on standards. 2//2007

. Na Cl Mg Al Si P S Cl K Ca Ti V Cr Mn Fe Ni new calib Ratio Found/Quoted Standard.05 0.95 0.9 Oct 2005 /2/06 -/28/06 Dec 2005 2/7/06 - /7/07 0.85 0.8 9//06 0//06 //06 2//06 //07 2//07 Figure 3. XRF Cu-Vac2 system performance chart based on standards. The XRF-Mo system was recalibrated after change was made on April 3 th 2006 to the operational parameters for standards. The current was changed from 2.6 ma to 0 ma to better reflect the livetimes for the IMPROVE samples. All calibration checks on the Mo system during the analysis of Oct-Dec 2005 samples were within our criteria. November 2005 samples were analyzed on the XRF Cu-Vac system and processed with new concentration calibration factors obtained in December 2006 to reflect maintenance done on the system (including a change of the vacuum pump) in September 2006 October and December 2005 samples were analyzed on XRF Cu-Vac2 system. This system was set up and calibrated in late September 2006. All calibration checks during the analysis of the October-December 2005 samples for the elements considered quantitative were successful. However, the following calibration check on January 8 th 2007, showed the Si standard slightly falling below acceptable range (0.89). None of the samples were re-analyzed. The PESA system is calibrated with six /8 mil thick Mylar blanks whose areal densities are determined from their weights and the chemical composition of Mylar. These foils have served as the PESA calibration standards for many years. The average hydrogen concentration for these PESA standards is calculated to be 20 μg/cm 2. As with XRF, the calibration factor is based on the ratio of observed counts for the six PESA standards to their calculated H concentration. The PESA system is recalibrated at the beginning of every analytical session, because of variations in the ion source production, amplitude harmonics, and optics. The six Mylar blanks used as calibration standards are reanalyzed approximately every 200 samples to verify the calibration throughout the session. If the ratio of reported to calculated concentrations for these standards drifts outside the 0.95-.05 range during an analysis run., the cyclotron is re-tuned, system is recalibrated, and samples reanalyzed. Figure 4 shows calibration verifications and recalibrations during the analysis of Oct-Dec 2005 samples. The cyclotron was re-tuned twice

during analysis of these samples. One of the calibration checks on Jan 8 th (run number 27839) resulted in the average ratio falling outside the criteria. The cyclotron was not retuned. The following checks were all successful..2 /8 mil Mylar standards October, November, December 2005 samples.5 Ratio Found/Standard..05 0.95 0.9 cyclotron re-tuned Oct 05 Nov 05 Dec 05 cyclotron re-tuned 0.85 0.8 22000 23000 24000 25000 26000 27000 28000 29000 30000 Run Number Figure 4. PESA standards for Oct-Dec 2005 samples 3.2 X-ray energy calibration In addition to the peak counts associated with a known concentration (concentration calibration), the energy channel associated with a known fluorescence line must also be determined; this is the energy calibration. Energy calibrations were performed for the analyses of each sample month on each system to establish relationships of the form energy = intercept + slope * channel The following energy calibration equations (in energy units of KeV) were used for the analysis: XRF-Cu change XRF-Cu2 change XRF-Mo change intercept slope full scale from Sep intercept slope full scale from Sep intercept slope full scale from Sep Oct-05-0.0358 0.076 8.750 N/A -0.087 0.03607 8.385 0.09% Nov-05-0.044 0.0667 8.494-0.02% -0.0797 0.03604 8.374 0.04% Dec-05-0.0358 0.076 8.750 N/A -0.0862 0.03606 8.378 0.06% 3.3 Reanalysis The reproducibility of XRF and PESA data is tracked over time by reanalyzing selected sample filters. Different reanalysis protocols are used for the XRF and PESA reanalyses, reflecting the different impacts of their exciting beams on the Teflon filter substrate, as explained in previous reports.

Filters to be reanalyzed by PESA are selected from the previous quarter s X-module (collocated A-module) samples. During the analysis of Oct 2005 samples, 56 PMRFX and OLYMX filters from Jan-Feb and July 2005 were reanalyzed. Similarly, Nov 2005 and Dec 2005 reanalyses were performed, respectively, on 52 SAFOX, SAMAX AND TRCRX filters from August 2005 and 56 MEVEX, OLYMX and PMRFX filters from August 2005. Figure 5 compares the original and repeat analyses. 0000 0000 reanalyzed with Oct 2005 samples 000 reanalyzed with Nov 2005 samples 000 00 00 000 0000 original OLYMX, PMRFX, Jan-Feb, Jul 2005 (ng/cm²) 00 00 000 0000 original SAFOX, SAMAX, TRCRX, Aug 2005 (ng/cm²) reanalyzed with Dec 2005 samples 0000 Figure 5. PESA reanalysis of selected Jan, Feb, Jul and Aug 2005 samples during analyses of Oct-Dec 2005 network samples. Reported uncertainties are indicated by error bars; agreement is indicated by sloping lines. 000 000 0000 original MEVEX, OLYMX, PMRFX, Aug 05 (ng/cm²) XRF reanalyses are conducted repeatedly on a fixed collection of sample filters referred to as REANAL and REANAL2 and described in previous reports. The trays were reanalyzed with the XRF-Mo and both vacuum XRF-Cu systems (Cu-Vac and Cu-Vac2) approximately monthly during 2006/2007 analyses of Oct-Dec 2005 samples. The results are summarized in the figures below. Figures 6- show the consistency of reanalysis results from both systems during the analysis of the Oct-Dec 2005 samples. The horizontal axis of each figure indicates the dates of successive reanalyses. The vertical axis shows for each element the average ratio, over all sample days, of observed deviations (from the mean of all reanalyses) to reported measurement uncertainties. This format highlights any systematic trend in the measurements and provides a test, at actual sample loadings, of the stability of calibrations based on the heavily-loaded foil standards.

At present, sample reanalysis is used as a qualitative check on system performance. 5 Reanalysis R on Mo Av erage Relative Deviation (Observed/Expecte d) 4 3 2 0 - -2-3 -4 S K CA TI V CR MN FE NI CU ZN AS PB SE BR RB SR Sep 2005 3/2/06-4/5/06 Oct 2005 4/6/06-5/4/06 Nov 2005 5/4/06-5/26/06 Dec 2005 6//06-6/29/06-5 mar6a may06 may6a jun06 jul06 Reanalysis Date Figure 6. Reanalyses of REANAL samples on XRF-Mo system. Horizontal arrow indicates when Oct-Dec 2005 network samples were analyzed. 5 Reanalysis R2 on Mo Average Relative Deviation (Observed/Expected) 4 3 2 0 - -2-3 -4 S K CA TI V CR MN FE NI CU ZN AS PB SE BR RB SR Sep 2005 3/2/06-4/5/06 Oct 2005 4/6/06-5/4/06 Nov 2005 5/4/06-5/26/06 Dec 2005 6//06-6/29/06-5 apr6a may06 may6a jun06 jul06 Reanalysis Date

Figure 7. Reanalyses of REANAL2 samples on XRF-Mo system. Horizontal arrow indicates when Oct-Dec 2005 network samples were analyzed. 5 4 REANAL on XRF-CuVac YNA YMG YAL YSI YS YCL YK YCA YTI YV YCR YMN YFE Average Relative Deviation (Observed/Expected) 3 2 0 - -2-3 -4 Nov 2005 2/4/06 - /2/07-5 Sep06 Oct6a Oct6b Nov06 Dec06 Jan7a Feb7a Reanalysis Date Figure 8. Reanalyses of REANAL samples on XRF-Cu system. Horizontal arrow indicates when Nov 2005 network samples were analyzed. Average Relative Deviation (Observed/Expected) 5 4 3 2 0 - -2-3 -4 REANAL2 on XRF-CuVac YNA YMG YAL YSI YS YCL YK YCA YTI YV YCR YMN YFE Nov 2005 2/4/06 - /2/07-5 Sep06 Oct6a Oct6b Nov06 Dec06 Jan7a Feb7a Reanalysis Date Figure 9. Reanalyses of REANAL2 samples on XRF-Cu system. Horizontal arrow indicates when Nov 2005 network samples were analyzed.

Average Relative Deviation (Observed/Expected) 5 4 3 2 0 - -2-3 -4-5 REANAL on XRF-CuVac2 YNA YMG YAL YSI YS YCL YK YCA YTI YV YCR YMN YFE Nov6a Dec6a Jan7a Jan7b Mar7a Mar7b Reanalysis Date Oct 2005 /2/06 /28/06 Dec 2005 2/7/06 /7/07 Figure 0. Reanalyses of REANAL samples on XRF-Cu2 system. Horizontal arrow indicates when Oct, Dec 2005 network samples were analyzed. Average Relative Deviation (Observed/Expected) 5 4 3 2 0 - -2-3 -4 REANAL2 on XRF-CuVac2 YNA YMG YAL YSI YS YCL YK YCA YTI YV YCR YMN YFE Oct 2005 /2/06 /28/06 Dec 2005 2/7/06 /7/07-5 Nov6a Dec6a Jan7a Jan7b Mar7a Mar7b Reanalysis Date Figure. Reanalyses of REANAL2 samples on XRF-Cu2 system. Horizontal arrow indicates when Oct, Dec 2005 network samples were analyzed. 3.4 Systems comparison

Additional comparison between elements measured independently by the Cu and Mo systems allows identification of problems that may not be evident in repeated measurements by the same system. The elements Calcium and Iron are reported from the Cu system but are also quantified by the Mo system. Figures 2 and 3 compare the two measurements of these two elements for the samples from Oct-Dec 2005. Since the October 2005 samples are the first to be reported from Cu-Vac2 system, selected sites (ISLE - MEAD) were reanalyzed on Cu-Vac, and these comparisons are also shown. Additional comparisons between Cu-Vac and Cu-Vac2 are presented in Attachment. Reported uncertainties are shown as bars for each sample, and reported MDL s are indicated by green and pink points for both systems. The increase in analytical uncertainty closer to the MDL s can be observed for all cases. 00000 October 2005 000 October 2005 0000 FE from Mo (ng/cm²) 000 00 0 Fe from CuVac 2, ng/cm 2 00 0 0. 0.0 0.0 0. 0 00 000 0000 00000 FE from Cu2 (ng/cm²) 0. 0. 0 00 000 Fe from CuVac, ng/cm 2 November 2005 December 2005 00000 00000 0000 0000 000 000 FE from Mo (ng/cm²) 00 0 FE from Mo (ng/cm²) 00 0 0. 0. 0.0 0.0 0. 0 00 000 0000 00000 FE from Cu (ng/cm²) 0.0 0.0 0. 0 00 000 0000 00000 FE from Cu2 (ng/cm²) Figure 2. Comparison of Iron data obtained independently from Cu (x-axis) and Mo (y-axis) systems. Cu refers to Cu-Vac and Cu2 to Cu-Vac2. October 2005

October 2005 00000 0000 0000 000 CA from Mo (ng/cm²) 000 00 0 Ca from CuVac 2, ng/cm 2 00 0 0. 0. 0 00 000 0000 00000 CA from Cu2 (ng/cm²) 0. 0. 0 00 000 0000 Ca from CuVac, ng/cm 2 November 2005 December 2005 00000 00000 0000 0000 000 CA from Mo (ng/cm²) 00 0 CA from Mo (ng/cm²) 000 00 0 0. 0.0 0.0 0. 0 00 000 0000 00000 CA from Cu (ng/cm²) 0. 0. 0 00 000 0000 00000 CA from Cu (ng/cm²) Figure 3. Comparison of Calcium data obtained independently from Cu (x-axis) and Mo (y-axis) systems. Calcium and iron determinations by the Mo system contain more uncertainty than those from the Cu system, and are accordingly not used to report concentrations. Their value in these system comparisons is the additional qualitative check they provide on both systems performance. 3.5 Field blanks Twenty four field blanks for November 2005 samples were exposed at selected sites on selected sampling events. The filed blanks were analyzed on XRF-Mo system on March 6, 2006, but were not analyzed yet on the Cu system. Therefore, the results are from XRF-Mo analysis only. Ninty-fifth, 90 th and 75 th percentile field blank loadings are shown in the table below. They are given as percentiles of well measured network sample loadings during November 2005. Loadings are considered well measured when their uncertainties are less than 0%.

Thus, the 95 th percentile field blank loading for Fe was as large as 7% of all well measured sample loadings. Field Blanks Nov 2005 Fe Cu Zn Pb Se Br 95 %ile 7.% 0.0% 36.5% 3.% 0.0% 0.0% 90 %ile 0.7% 0.0% 23.7%.5% 0.0% 0.0% 75 %ile 0.0% 0.0% 4.6% 0.2% 0.0% 0.0% In addition to field blanks, fifteen laboratory blanks were analyzed on the Mo system. Table below shows the percentages of values in the network below laboratory blank concentrations. Laboratory Blanks Nov 2005 Fe Cu Zn Pb Se Br 95 %ile 0.3% 0.0%.53% 0.0% 0.0% 0.0% 90 %ile 0.2% 0.0% 0.8% 0.0% 0.0% 0.0% 75 %ile 0.% 0.0% 0.00% 0.0% 0.0% 0.0% The results show that for all the elements, with the exception of Zn, the field and lab blanks loadings are negligible compared to samples. The field blank results for Zn suggest contamination that has been noted before.

Attachment All July 2005 IMPROVE samples were analyzed on both Cu-anode XRF systems, the original (Cu-Vac) and the duplicate (Cu-Vac2). The following is an album of comparisons for each CuXRF element. Each point is (Y,Y2), where Yi is val(i) if detected and mdl(i) if not detected. Samples with a non-detect in one analysis are plotted in red. Samples with non-detects in both analyses are not plotted. 00 Na 0 Mg 0 CuVac 2, ug/cm 2 0. 0. 0.0 0.0 0.0 0. 0 00 0.0 0. 0 CuVac, ug/cm 2 CuVac, ug/cm 2 00 0 Al 00 0 Si CuVac 2, ug/cm 2 0. 0. 0.0 0.0 0.00 0.00 0.00 0.0 0. 0 00 0.00 0.0 0. 0 00 CuVac, ug/cm 2 CuVac, ug/cm 2

0 P 00 S CuVac 2, ug/cm 2 0. 0 0.0 0.0 0. 0 00 0 Cl 0. 0. 0 00 00 0 K CuVac 2, ug/cm 2 0. 0.0 0. 0.0 0.00 0.00 0.000 0.000 0.00 0.0 0. 0 00 0 Ca 0.000 0.000 0.00 0.0 0. 0 00 0 Ti CuVac 2, ug/cm 2 0. 0.0 0. 0.0 0.00 0.00 0.000 0.000 0.000 0.00 0.0 0. 0 CuVac, ug/cm 2 0.0000 0.0000 0.000 0.00 0.0 0. 0 CuVac, ug/cm 2

V Cr 0. 0. CuVac 2, ug/cm 2 0.0 0.00 0.0 0.00 0.000 0.000 0.00 0.0 0. CuVac, ug/cm 2 0.000 0.000 0.00 0.0 0. CuVac, ug/cm 2 0. Mn 00 0 Fe CuVac 2, ug/cm 2 0.0 0.00 0. 0.0 0.00 0.000 0.000 0.00 0.0 0. CuVac, ug/cm 2 0.000 0.000 0.00 0.0 0. 0 00 CuVac, ug/cm 2