PERFORMANCE IMPROVEMENT OF CZT DETECTORS BY LINE ELECTRODE GEOMETRY

Similar documents
THE spectroscopic performance of large volume CdZnTe

Development and characterization of 3D semiconductor X-rays detectors for medical imaging

Final report on DOE project number DE-FG07-99ID High Pressure Xenon Gamma-Ray Spectrometers for Field Use

New trends in CdTe detectors for X and γ-ray applications

Measurement of material uniformity using 3-D position sensitive CdZnTe gamma-ray spectrometers

Nuclear Instruments and Methods in Physics Research A

THE mobility-lifetime product (μτ) is used to characterize

CADMIUM telluride (CdTe) and cadmium zinc telluride

Stability and Characteristics of Large CZT Coplanar Electrode Detectors

Development of Gamma-ray Monitor using CdZnTe Semiconductor Detector

A CdTe detector with a Gd converter for thermal neutron detection

Investigation of the Asymmetric Characteristics and Temperature Effects of CdZnTe Detectors

Estimate of Large CZT Detector Absolute Efficiency

Application Note ANCZT-2 Rev. 3 Charge Trapping in XR-100T-CdTe and -CZT Detectors

Spatial resolved efficiency determination of CdZnTe semiconductor detectors with a collimated gamma-ray source for the C0BRA experiment

arxiv:astro-ph/ v1 6 Mar 2001

Efficiency and Attenuation in CdTe Detectors

MINIATURE CdZnTe DETECTORS FOR APPLICATION IN HIGH INTENSITY RADIATION FIELDS

Hard X- and g-ray measurements with a large volume coplanar grid CdZnTe detector

Development of semiconductor imaging detectors for a Si/CdTe Compton camera

Review of the Shockley Ramo theorem and its application in semiconductor gamma-ray detectors

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data

SENSITIVITY ANALYSIS TO EVALUATE THE TRANSPORT PROPERTIES OF CdZnTe DETECTORS USING ALPHA PARTICLES AND LOW-ENERGY GAMMA-RAYS

Semiconductor Detectors

Advances in Compound Semiconductor Radiation Detectors. a review of recent progress

Nuclear Instruments and Methods in Physics Research A

In order to get the energy response of the CZT detector, the measurements were done using the mono-energetic X-ray beam sources from 10 to 40 kev and

Development of New MicroStrip Gas Chambers for X-ray Applications

A Prototype of LaBr3:Ce in situ Gamma-Ray Spectrometer for Marine Environmental Monitoring

Simulation of charge transport in pixelated CdTe

Pixels GaAs Detectors for Digital Radiography. M.E. Fantacci. and. Abstract

Status and Perspectives of the COBRA-Experiment

EEE4106Z Radiation Interactions & Detection

Chem 481 Lecture Material 3/20/09

Development of a Spectral Model Based on Charge Transport for the Swift/BAT 32K CdZnTe Detector Array

GEM: A new concept for electron amplification in gas detectors

Semiconductor Detectors

Study of PC-HPGe detector for dark matter search

Maximum-Likelihood Deconvolution in the Spatial and Spatial-Energy Domain for Events With Any Number of Interactions

The Best Gamma-Ray Detector

Research Article Performance Test and Evaluation for Pixel CdZnTe Detector of Different Thickness

1. Introduction. Study of Thick CZT Detectors for X-ray and Gamma-Ray Astronomy

Cadmium Zinc Telluride (CZT) Detectors

Position Sensitive Germanium Detectors for the Advanced Compton Telescope

A Study On Radioactive Source Imaging By Using A Pixelated CdTe Radiation Detector

Silicon Drift Detectors for gamma-ray detection: 15 years of research (and collaboration between Politecnico and INAF-Milano)

Gamma-ray Spectroscopy with LaBr 3 :Ce Scintillator Readout by a Silicon Drift Detector

Gamma and X-Ray Detection

Exploring the limiting timing resolution for large volume CZT detectors with waveform analysis

Polaris 3-D CdZnTe (CZT) Gamma-Ray Imaging Spectrometers

Detectors for High Resolution Gamma-ray Imaging Based on a Single CsI(Tl) Scintillator Coupled to an Array of Silicon Drift Detectors

SUPPLEMENTARY INFORMATION

Development of a Radioxenon Detection System Using CdZnTe, Array of SiPMs, and a Plastic Scintillator

Performance study of the ceramic THGEM *

DETECTORS. I. Charged Particle Detectors

Effects of Etching and Chemo-Mechanical Polishing on the Electrical Properties of CdZnTe Nuclear Detectors

Development of High-Z Semiconductor Detectors and Their Applications to X-ray/gamma-ray Astronomy

X-ray spectrometry with Peltier-cooled large area avalanche photodiodes

X-ray induced radiation damage in segmented p + n silicon sensors

New perspectives in X-ray detection of concealed illicit materials brought by CdTe/CdZnTe spectrometric detectors

LAAPD Performance Measurements in Liquid Xenon

NUCL 3000/5030 Laboratory 2 Fall 2013

Determination of the shielding power of different materials against gamma radiation

& SUN ZhiJia 1;2. Abstract

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber

arxiv: v1 [physics.ins-det] 29 Jun 2011

Boron-based semiconductor solids as thermal neutron detectors

On the development of compound semi-conductor thallium bromide detectors for astrophysics

Development of Radiation Detectors Based on Semi-Insulating Silicon Carbide

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona, USA

arxiv:physics/ v1 3 Aug 2006

Results of a Si/CdTe Compton Telescope

A POSITION SENSITIVE ALPHA PARTICLE DETECTOR BASED ON A LYSO CRYSTAL AND A MICRO-PIXEL AVALANCHE PHOTODIODE

X-ray ionization yields and energy spectra in liquid argon

arxiv:physics/ v2 [physics.ins-det] 22 Nov 2005

Detection of X-Rays. Solid state detectors Proportional counters Microcalorimeters Detector characteristics

Alpha-Gamma discrimination by Pulse Shape in LaBr 3 :Ce and LaCl 3 :Ce

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

GAMMA-RAY IMAGING AND POLARIZATION MEASUREMENT USING 3-D POSITION-SENSITIVE CdZnTe DETECTORS

Chapter 4 Scintillation Detectors

Advances in the Micro-Hole & Strip Plate gaseous detector

List of Nuclear Medicine Radionuclides. Nuclear Medicine Imaging Systems: The Scintillation Camera. Crystal and light guide

SCI-O11. Design of a Compton Spectrometer Experiment for Studying Electron Response of a Scintillator

PoS(PD07)020. Timing Properties of MCP-PMT. Kenji Inami. Nagoya university, Nagoya, Japan

Modeling of charge collection efficiency degradation in semiconductor devices induced by MeV ion beam irradiation

Neutron Induced Nuclear Counter Effect in Hamamatsu Silicon APDs and PIN Diodes

Study well-shaped germanium detectors for lowbackground

An industrial radiography exposure device based on measurement of transmitted gamma-ray intensity

RESPONSE CALCULATIONS OF THE CdZnTe DETECTOR USING EGS4. James C. Liu, W. R. Nelson and R. Seefred

FISSILE MATERIALS DETECTION VIA NEUTRON DIFFERENTIAL DIE-AWAY TECHNIQUE

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Chapter Seven (Nuclear Detectors)

Measurement of nuclear recoil responses of NaI(Tl) crystal for dark matter search

ONE of the most noticeable facts in the field of room

Researchers at the University of Missouri-Columbia have designed a triple crystal

Energy and position resolution of a CdZnTe gamma-ray detector with orthogonal coplanar anodes

Energetic particles and their detection in situ (particle detectors) Part II. George Gloeckler

Radioactivity. Lecture 6 Detectors and Instrumentation

Solid State Detectors

MEASUREMENT OF DIAGNOSTIC X-RAY SPECTRA USING CdZnTe DETECTOR

Transcription:

Applications of Nuclear Techniques (CRETE3) International Journal of Modern Physics: Conference Series Vol. 27 (24) 4644 (8 pages) The Authors DOI:.42/S294546446 PERFORMANCE IMPROVEMENT OF CZT DETECTORS BY LINE ELECTRODE GEOMETRY Int. J. Mod. Phys. Conf. Ser. 24.27. Downloaded from www.worldscientific.com by 46.3.23.86 on 2/7/8. For personal use only. LAN ZHANG, WEI ZHANG, JUN LI, YANQING LIU Nuctech Company Limited Beijing, 84, China zhanglan@nuctech.com YULAN LI, YUANJING LI, YINONG LIU, JIANQIANG FU, HAO JIANG, ZHI DENG, TAO XUE, JIE LUO Tsinghua University Beijing, 84, China yulanli@mail.tsinghua.edu.cn Published 25 February 24 Due to their high efficiency and high energy resolution, applications of CdZnTe (CZT) detectors have spread into many areas such as room temperature detectors. To improve the performance and decrease the effects of hole trapping and crystal non-uniformity, special contact geometries are applied to the detectors, such as Coplanar, 3D pixel, Frisch grid, Capture and so on. In this paper, we introduce a new line electrode geometry to greatly improve CZT detector performance. A line anode collects electrons, while a planar cathode collects holes. Due to the low electric field close to the cathode and the low hole µτ value, this geometry is also a single charge sensitive structure. The calculated energy resolution of a 2 mm x 2 mm x 5 mm detector could be improved up to.5% for 662 kev gamma rays. Both simulations and experimental results are presented here. Keywords: CdZnTe; CdTe; semiconductor detectors.. Introduction CdZnTe (CZT) detectors are widely applied in different fields. Spectroscopy and photon counting imaging are two main applications. Here we discuss the CZT detectors used for gamma and X-ray detection. As opposed to thin CdZnTe pixel detectors used for X-ray imaging at energies lower than 3 kev, a much thicker CZT crystal should be used for high energy gamma ray detection. However, crystal defects and charge trapping greatly decrease the detector performance. In order to obtain high energy resolution together with high detection efficiency, several contact geometries have been applied, such as Coplanar, 3D pixel, Frisch grid, Capture and so on. 4 These geometries improve the CZT performance mainly due to a higher single polarity charge sensitivity. This is an Open Access article published by World Scientific Publishing Company. It is distributed under the terms of the Creative Commons Attribution 3. (CC-BY) License. Further distribution of this work is permitted, provided the original work is properly cited. 4644-

L. Zhang et al. We propose a new contact structure a line electrode geometry. This structure effectively improves the CZT detector performance and has a simple readout system. The preliminary results are provided in the paper. 2. The Line Electrode Geometry Int. J. Mod. Phys. Conf. Ser. 24.27. Downloaded from www.worldscientific.com by 46.3.23.86 on 2/7/8. For personal use only. This new line electrode geometry, shown in Fig., is an efficient method to improve the performance of thick CZT detectors. A line electrode is deposited on one surface of the detector. The planar cathode is on the opposite surface. The two side surfaces that are parallel to the line electrode are also planar cathodes. The three cathodes are connected together with metal contacts. The signal will be read out from the line anode. The electrons migrate toward the line anode and the holes migrate to the cathode. Due to the low electric field and low hole mobility, holes will be strongly trapped, and the signal is mainly induced by the electron movement. This structure is also a single charge sensitive detector. Fig.. Line electrode geometry scheme. Compared to coplanar and pixel CZT detectors, this structure is easy to fabricate, and the read out system is very simple. This contact geometry is to some extent similar to high-purity, coaxial germanium detectors, but the CZT crystal is softer and therefore more prone to cleavage. Drilling a hole in CZT can cause severe damage, so it is not an easy operation. The germanium crystal volume is usually large enough for a mm diameter hole, but for CZT crystals, it is quite difficult to obtain a large volume, up to 5 mm in diameter, with good uniformity and high quality. With a simple line anode, the operation is simplified and the width of the anode can be adjusted freely to obtain better performance, even for a width of only mm. 3. Simulation In order to analyze the performance of CZT detectors with the line electrode geometry, we simulated the weighting potential, charge collection and spectrum. First, the radiation response was simulated using GEANT4, and then we calculated the electric field and weighting potential using Maxwell s equations. The charges created from incident 4644-2

Performance Improvement of CZT Detectors by Line Electrode Geometry photons will move along the electric field lines: the electrons move toward the line anode and the holes move to the cathode. The induced signal at the electrodes was calculated using the ICC code developed by our research group. The µτ values of holes and electrons assumed for the calculation were 5x -5 cm 2 /V and 7x -3 cm 2 /V respectively. The spectra shown in Fig. 2 for three sizes of line anode CZT detectors were simulated. Usually the 662 kev photon peak is not obvious in the spectra from 3-mm thick CZT detectors with planar contacts. From our simulations, the best 662 kev energy resolution would be.7%, 2.87% and.5% for the 5 x 5 x 5 mm, x x mm and 2 x 2 x 5 mm line anode CZT detectors, respectively. The resolution varies with the Int. J. Mod. Phys. Conf. Ser. 24.27. Downloaded from www.worldscientific.com by 46.3.23.86 on 2/7/8. For personal use only. 662keV Spectra of 5-5-5mmLine-anode detector.8.5 2 2 3 4 5 6 7 8 2.5 6 2 8 4.8 anode width Line anode width (mm).8.5 2 662 kev resolution (%).27.2.735.26 Fig. 2(a). The simulated 662 kev spectra for 5x5x5 mm CZT detectors with different line anode widths. 662keV Spectra of --mmline-anode detector 2 3 4 3 2 2 3 4 5 6 7 8 2 3 anode width 4 Line anode width (mm) 2 3 4 662 kev resolution (%) 4. 3.4 2.87 2.4 Fig. 2(b). Simulated 662 kev spectra for xx mm CZT detectors with different line anode widths. 4644-3

L. Zhang et al. 662keV Spectra of 2-2-5mm Line-anode detector 3 5 3 2 Int. J. Mod. Phys. Conf. Ser. 24.27. Downloaded from www.worldscientific.com by 46.3.23.86 on 2/7/8. For personal use only. 2 3 4 5 6 7 8 Line anode width (mm) 3 5 662 kev resolution (%) 3.85.5 2. Fig. 2(c). Simulated 662 kev spectra for 2x2x5 mm CZT detectors with different line anode widths. line anode width. We found that the best resolution is obtained through a balance between many detector parameters, such as the area, thickness, line anode width, etc. The narrowest line anode structure did not give the best results, because the electric field is weak and the charge trapping will be strong. The weighting potential of a wider line anode would be spread out and the single charge sensitivity will be affected. We observed that a high energy tail appeared in all the simulations. To analyze the cause, we calculated the signal amplitude variation with the photon interaction position inside the CZT detector (see Fig. 3). The weighting potential distribution was also calculated. Figure 3 shows that the high energy tail comes from the interaction position close to the area between the side cathode surface and line anode surface. The electric field, the weighting potential and the distance from the readout electrode are suitable for the optimum charge collection at this position.. 5*5*5 detector.9 Energy Collected.8.7.6 X=-2.5 X=-2 X=-.5 X=- X=-.5 x=.5..5..5 2. 2.5 3. 3.5 4. 4.5 5. Z/mm Fig. 3. The relationship between charge collection efficiency and photon interaction position. 4644-4

Performance Improvement of CZT Detectors by Line Electrode Geometry Int. J. Mod. Phys. Conf. Ser. 24.27. Downloaded from www.worldscientific.com by 46.3.23.86 on 2/7/8. For personal use only. To adjust the signal amplitude to be coincident for good resolution, we decreased the side cathode height. The calculated results are shown in Fig. 4. We can see that the high energy tail has significantly decreased, but it has not disappeared completely. When blocking electrodes were placed between the line anode and the side cathode, the high energy tail has notably diminished, as shown in Fig. 5. The 2 x 2 x 5 mm detector was used for this calculation. The widths of the line anode and the blocking electrodes were mm. The blocking electrodes were placed at both sides of the line anode and powered with a lower voltage than the line anode. The 662 kev energy resolution was improved to.99%, but the detection efficiency decreased due to the charge collection by the blocking electrode. To obtain the best energy resolution, the electrode width and location, as well as the voltage difference need to be considered. 4mm anode bias voltage V mm side cathode height 8mm side cathode height 7mm side cathode height 6mm side cathode height 2 3 4 5 6 7 8 4 3 2 6 45 3 5 4mm anode bias voltage=v 6 625 65 675 7 mm side cathode height 8mm side cathode height 7mm side cathode height 6mm side cathode height Side cathode height (mm) 5 6 7 8 9 662 kev resolution (%) 2.27.6.94.46 -- 2.4 Fig. 4. (Left) The 662 kev spectrum for decreased side cathode height. (Right) Enlarged peak area for comparison. 662 kev spectra of line anode CZT with blocking electrodes 2 blocking electrode 2V 2 blocking electrode 5V 4 blocking electrode V 4 blocking electrode 2V 8 6 4 2 2 3 4 5 6 7 8 Number of blocking electrodes 2 4 Voltage difference (V) 2 5 2 662 kev resolution (%).38.82.4.99 Fig. 5. 662 kev spectra of line anode CZT with two and four blocking electrodes. 4644-5

L. Zhang et al. Through our research, we found a big difference for lower energy photons incident on the detector from different directions as seen in Fig. 6. The best spectrum is for 6 kev photons incident on the bottom cathode. This is easy to explain: all 6 kev photons were deposited within a mm thick layer close to the bottom cathode surface. Holes are collected very quickly, and electrons (which all move a similar distance) contribute to the main part of signal. The spectrum for photons incident on the side cathode surface is seriously affected. Although electrons contribute almost to the whole signal, the electron moving distance varies largely, and both the electron trapping and moving distance affect the spectrum. Int. J. Mod. Phys. Conf. Ser. 24.27. Downloaded from www.worldscientific.com by 46.3.23.86 on 2/7/8. For personal use only. 5 2 9 6 3 ---3-7 irradiate from different sides @6keV incident on side cathode incident on side surface without cathode incident on line anoe surface incident on plane cathode 2 3 4 5 6 7 8 induced charge. incident on.9 plane cathode.8.7.6.5.4.3 Induced Charge Map @ Y= incident on side surface without cathode.2 2 3 4 5 6 7 8 9 Z/mm Fig. 6. Spectra of 6 kev photons incident from different directions. incident on side cathode incident on line anode surface The spectrum for photons incident on the side surface without a cathode shows less peak area. Charges from the interactions spread widely from cathode to anode. It is similar to 662 kev photons penetrating the whole crystal, and the interaction position spreads between the cathode and anode uniformly. The pulse amplitude varies with the interaction position, and the fluctuation decreases the spectrum peak. The spectrum with the peak moving toward the low energy is for photons entering on the line anode surface, and the hole trapping is strong. Use of blocking electrodes could solve the problem (see Fig. 7). These blocking electrodes collect most of the charge induced by photons incident on both the line anode surface and the side cathode. 4. Experiments In order to verify the simulations and the performance improvement, we fabricated a line electrode geometry CZT detector (see Fig. 8). The size of the detector was x x mm. The top gold line anode was 4 mm wide. The height of the side cathode was 7 mm. 4644-6

Performance Improvement of CZT Detectors by Line Electrode Geometry 2-2-5mm irradiate from different sides @6keV incident on plane cathode incident on side surface withou cathode incident on side cathode incident on line anode 8 6 4 2 Int. J. Mod. Phys. Conf. Ser. 24.27. Downloaded from www.worldscientific.com by 46.3.23.86 on 2/7/8. For personal use only. 2 3 4 5 6 7 8 Fig. 7. Spectra of 6 kev photons incident from different directions for blocking electrode structure. Fig. 8. Line electrode CZT detector of size x x mm. Cs37-8V-.5µs-Gain6 25 2 5 5 2 4 6 8 channel Fig. 9. Measured 662 kev spectrum of the line anode CZT detector. We irradiated the bottom cathode of the detector with a 662 kev 37 Cs source and obtained the spectrum shown in Fig. 9. The energy resolution was found to be 2.7%. Using the test pulse to measure the electronic system noise, the pulse FWHM was calculated to be.58%. So the line anode CZT resolution was approximated as.49%. It 4644-7

L. Zhang et al. was shown that the shaping time and the bias are important parameters for good resolution. When 6 kev 24 Am photons entered the CZT detector from four directions, the spectra shown in Fig. were obtained. We can see that the simulations (Fig. 6) and experiments (Fig. ) are generally in agreement. Photons incident on the side cathode need to pass 5 mm thick Teflon, while photons incident on the line anode surface have to go through two PCB board and the distance they have to travel is greater than others. Also, a plane source was used in the simulations while a point source was used for the experiments. This leads to the difference between simulations and experiments. Further experiments are in progress, and the results will be reported in the near future. Int. J. Mod. Phys. Conf. Ser. 24.27. Downloaded from www.worldscientific.com by 46.3.23.86 on 2/7/8. For personal use only. 2 5 5 incident on planar cathode incident on side cathode incident on side surface without cathode incident on line anode surface Am24-8V-µs-Gain6 25 5 75 channel Fig.. Measured 6 kev spectra incident from four directions on the line anode CZT detector. 5. Conclusions This paper proposes a new line electrode geometry CZT detector. A line electrode is deposited on one surface of the detector, and the planar cathode is on the opposite surface. The two side surfaces parallel to the line electrode are also planar cathodes. Through simulations and experiments, we determined that the CZT detector performance was obviously improved. Modifying the side cathode height and inserting blocking electrodes further enhanced the energy resolution, especially for low energy photons incident from different directions. Results of the ongoing experiments will be reported in the future. References. D. S. McGregor, Z. He, H. A. Seifert et al., IEEE Trans. Nucl. Sci. 45(3), 443 (998). 2. P. N. Luke, Applied Physics Letters 65, 2884 (994). 3. Csaba Szeles, Derek Bale, Joseph Grosholz et al., Proceedings of SPIE 639, 6399 (SPIE, Bellingham, WA, 26), doi:.7/2683552. 4. A. Shor, Y. Eisen and I. Mardor, Nucl. Instr. Meth. Phys. Res. A 458, 47 (2). 4644-8