Automated LC Method Development: Fundamentals and Columns, Hardware, and Software

Similar documents
LC and LC/MS Column Selection Flow Chart

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

High Resolution Fast LC

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

Fast Screening Methods for Steroids by HPLC with Agilent Poroshell 120 Columns

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

UPLC Method Development and Validation

Agilent LC & LC-MS Cannabis Testing

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

penta-hilic UHPLC COLUMNS

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

LC Technical Information

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

Fusion QbD and the Perfect Storm of Technologies Driving QbDaligned LC Method Development

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Chromatography. Gas Chromatography

Basic Principles for Purification Using Supercritical Fluid Chromatography

Routine MS Detection for USP Chromatographic Methods

Modernizing the USP Monograph for Acetaminophen

FUSION METHOD DEVELOPMENT SOFTWARE

A Look at Agilent HPLC Columns Choices

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

Product Bulletin. ACE LC/MS and Rapid Analysis HPLC Columns. HPLC Columns

Multi-residue analysis of pesticides by GC-HRMS

Reversed Phase Solvents

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

New 5-micron HALO-5 columns based on Fused- Core particle technology boost the performance of HPLC. Compared to other HPLC columns, HALO-5 columns

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Validated HPLC Methods

Sample Preparation. Approaches to Automation for SPE

Pure Chromatography Consumables Pure flexibility. Pure specialization. Pure convenience.

Successfully Scaling and Transferring HPLC and UPLC Methods

Guide to Peptide Quantitation. Agilent clinical research

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Application Note. Abstract. Authors. Pharmaceutical. William J. Long Agilent Technologies, Inc Centerville Road Wilmington, DE USA

Too Polar for Reversed Phase What Do You Do?

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

penta-hilic UHPLC COLUMNS

Automated method development using Agilent 1100 Series HPLC systems, Agilent ChemStation and ChromSword software. Application

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

High Performance Liquid Chromatography

Varian Galaxie Chromatography Data System for Preparative HPLC

Advanced HPLC Method Development Using Quality by Design

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

C18 Column. Care & Use Sheet

LECTURE 2. Advanced Separation Science Techniques Present and Future Separation Tools

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

for Acclaim Mixed-Mode HILIC-1 Column

Nexera UC Unified Chromatography

A Quality-by-Design Methodology for Rapid HPLC Column and Solvent Selection

Figure 1. Structures for Vitamin B2 and Vitamin B1.

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

HPLC Method Development with Eclipse Plus: Standard Practices and New Columns. Agilent Technologies

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

Alternatives to Sub-2 µm UHPLC Columns

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

Luminescence transitions. Fluorescence spectroscopy

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Open Column Chromatography, GC, TLC, and HPLC

High Performance Liquid Chromatography

ACD/Labs Software Impurity Resolution Management. Presented by Peter Russell

Packings for HPLC. Packings for HPLC

Implementation of Methods Translation between Liquid Chromatography Instrumentation

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Fast Screening Methods for Beta Blockers by HPLC with Agilent Poroshell 120 Columns

FUSION QBD. QbD-aligned LC Method Development with Fusion QbD

Fast methods for the determination of ibuprofen in drug products

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

Hypersil BDS Columns TG 01-05

Automatization for development of HPLC methods

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

New Product Update. HPLC Columns

Basic chromatographic parameters and optimization in LC

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs

How proteins separate on reverse-phase HPLC

High Pressure/Performance Liquid Chromatography (HPLC)

ACD/AutoChrom Assisted Method Development for Challenging Separations. Vera Leshchinskaya February 7, 2018

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Transcription:

An Executive Summary Automated LC Method Development: Fundamentals and Columns, Hardware, and Software Dr. William Long Application Scientist in the LC Columns marketing group Agilent Technologies Andreas Borowiak Product Manager Analytical HPLC Agilent Technologies Overview Method development for high performance liquid chromatography (HPLC) is a critical task, but with the wide array of available columns, eluent, equipment, and operating parameters, it can seem like an extremely complex and time-consuming process to complete. Moreover, if individuals involved in HPLC method development do not establish a clear plan for how they will tackle the project from the start, it can quickly become even more complicated. In this manuscript, experts lay out clear objectives for HPLC method development as well as identify key elements to keep in mind during the development process. Several HPLC method development strategies are discussed. In the end, efficient method development requires an appropriate balance between the initial investment and spending on developing the methods. When evaluating new technologies, the authors recommend choosing instruments that are scalable, flexible with upward and downward compatibility, and those that provide automation. First Steps: HPLC Method Development To ensure a smooth HPLC method development process, objectives must take into consideration: The kind of sample that will be analyzed and its complexity Whether high-efficiency separations are important, and which column particle sizes (e.g., sub-2-µm, 5-µm) may work best for the separation at hand The speed at which samples must be run Any potential instrument limitations in the lab The operators skillset, especially if the method requires a highly skilled analyst Figure 1 offers several examples of method performance criteria, many of which will be required during method validation. To avoid potential problems, leave out any system suitability criteria that will lock in the use of specific instrument platforms or columns. For instance, specifying absolute retention times or column efficiency of an exact number of plates could stop labs from taking advantage of future developments in equipment and instrumentation. Strategy 1: Hands-on HPLC Method Development One standard method development practice involves altering various selectivity-changing parameters such as ph, columns, and mobile phase types to establish the best separation conditions. After every step, technicians must analyze the results and determine how to proceed. SPONSORED BY

Individuals who prefer a hands-on HPLC method development technique should begin by selecting a highquality modern C18 or C8 column (e.g., Agilent InfinityLab Poroshell 120). Newer, shorter columns packed with small particles can help provide faster analysis with high efficiency and resolution. Second, choose a simple mobile phase that is reliable and will work with numerous samples. Many compounds are ionizable, and thus their retention and selectivity are greatly affected by ph. Therefore, screening ph can be an important way to identify good separation conditions. Consider starting with a low ph in the aqueous portion (e.g., phosphate buffer ph 3, trifluoroacetic acid, or formic acid) and with acetonitrile or methanol as the organic modifier. Then, optimize the mobile phase to achieve the desired retention and resolution. One can consider optimization achieved when there is adequate peak resolution (Rs 2); the first peak retention factor is k'=1 or better; and the analysis takes 20 minutes or less to complete. Selectivity. How can one control the resolution from a chromatographic separation? As Figure 2 shows, resolution is affected by three important factors: number of plates (i.e., efficiency), selectivity (i.e., separation), and retention (i.e., capacity, k'). One can see that increasing the number of theoretical plates (N) by using a longer column, using a column with smaller particles, or moving to a superficially porous column increases the efficiency of a separation. It is also possible to improve resolution by adjusting k'. This can be achieved by altering the ratio of organic to buffer mobile phase, but it is not the most effective approach. Conversely, changing selectivity (alpha) is clearly the most powerful and easiest way to affect resolution; this can be accomplished by changing either the bonded or mobile phase. Other factors that affect selectivity are shown in Figure 3. Adjusting selectivity is easier if various column choices are Selectivity is Powerful Resolution 7.00 6.00 5.00 4.00 3.00 2.00 1.00 Increasing N Increasing Alpha Increasing k' available such as with the InfinityLab Poroshell line. These superficially porous columns offer high efficiencies, relatively low back pressures, and good lifetimes. The line has a wide variety of stationary phases, which supports fast and easy method development. Recent updates to the InfinityLab Poroshell line include: a new column ID tag to help track usage, a high ph column that helps maximize selectivity and improve peak shape for basic compounds, and the availability of three particle sizes (1.9 µm, 2.7 µm, and 4 µm). A recent study emphasizes the benefits of using new technology during process development. Researchers found a Poroshell 120 EC-C8 column slashed ibuprofen analysis time from 15 minutes to 2.4 minutes with pressure of <400 bar. Figure 1: Examples of common separation goals and method performance criteria. Good System Suitability Parameters Resolution: 2 Peak shape: USP T f close to 1 (<2) Injection Repeatability: areas, T f, etc. (RSD 0.1-0.25%) Absolute retention factors: 1< k<10 Relative Retention: α or k 2 /k 1 Signal-to-Noise Ratio: >10 AVOID THESE for System Suit. Criteria: Column efficiency (theoretical plates) Absolute retention time These may prevent the ability to speed up your method in the future! Figure 2: Selectivity is powerful. Method Performance Criteria Accuracy Precision Ruggedness Robustness Analytical Selectivity/Specificity Linearity Range Quantitation Limit (LOQ, 10x S/ N) Detection Limit (LOD, 3x S/N) R s = N ½ /4 (α-1)/α k /(k +1) 0.00 Plates: 5000 10000 15000 20000 25000 Alpha: 1.10 1.35 1.60 1.85 2.1 k 2.0 4.5 7.0 9.5 12.0 Selectivity Impacts Resolution Most Change bonded phase Typical Method Development Parameters Change mobile phase Plates are easiest to increase

Putting this into perspective with an assay method, one could run 5,000 injections at 3.5 ml/min in about 10.5 days; with an old column, the analysis would take 70 days. injections; a short high-efficiency InfinityLab Poroshell 120 EC-C18 column was at the heart of this process that took less than one hour. Scouting gradient. Af ter selecting a shorter efficient column, another way to jumpstart method development is with the use of a scouting gradient (e.g., 5 95% acetonitrile gradient at low ph, depending on the column length). If the gradient has spaced-out peaks across the gradient width, then method developers may find that a gradient method is ideal. If all peaks are pushed to one side of the chromatogram, an isocratic method might be more appropriate. Figure 4 lists several column volumes and recommended scouting gradient times. In this example, a scouting method was used to produce a basic method in three Figure 3: Factors that influence selectivity. Isocratic Separations Column stationary phase Organic modifier ph (ionized analytes only) % Organic modifier Buffer selection Column temperature Buffer concentration MOST Important LEAST Important Gradient Separations All parameters for isocratic Gradient steepness k* Dwell volume Column dimensions Adapted from Introduction to Modern Liquid Chromatography, 3rd Edition, Snyder, Kirkland, Dolan, 2010, p.29, Wiley & Sons. Figure 4: Starting point: scouting gradient. Good Starting Point for work is a scouting Gradient. Conditions recommended by John Dolan are 5-95 % Acetonitrile low ph and are dependent on the column length. Where 10 cm columns are chosen use a 10 minute gradient. This example shows a 150 mm column.

Changing solvent. During HPLC method development, one should optimize the organic modifier before the bonded phase. There are several reasons for completing these steps in this order. First, most HPLC instruments will easily accommodate a quick change of mobile phase; replacing acetonitrile with methanol (or vice versa) is easy, both are readily available, and they work with any bonded phase. A simple move from acetonitrile to methanol may provide a better peak shape with bases, for instance. Controlling mobile phase ph. One can also control the mobile phase ph to adjust peak shapes. To do so, one must first consider whether the compounds are ionizable, and thereby ph sensitive. Figure 5 shows how various compounds retentions change with ph. Standard columns can have ph limitations; however, the InfinityLab Poroshell HPH column may be helpful in that it can be used in high-ph mobile phases without any loss of retention and with excellent lifetime efficiency. Changing the column. If the previous steps have not fully optimized the HPLC method, one may next consider changing the bonded phase, which can help quickly and easily improve selectivity and resolution. Figure 6 illustrates how a simple change in the column phase, while using the same gradient, altered the compounds elution order. Note that this run on a triple-quadrupole mass spectrometer involved a C18, a Phenyl- Hexyl, and the Bonus RP embedded polar columns. Figure 5: Change in retention with ph for ionizable (compound-dependent). More retention for non-charged analytes (i.e. acids at low ph and bases at high ph) 12 Retention Time 10 8 6 4 2 0 Acetylsalicylic acid (pka 3.5) Pyridine (pka 5.2) Codeine (pka 8) Procainamide (pka 9.2) Amphetamine (pka 9.9) Caffeine (pka 14) ph 2.5 ph 6.5 ph 8 ph 11.5 Mobile Phase: 45% MeOH, 55% 20 mm Phosphate Buffer Figure 6: Selectivity: comparison of sulfa drugs on different phases with 0.1 % formic acid. C18 3 x 100 mm 1.Sulfadiazine 2.Sulfathiazole 3.Sulfapyridine 4.Sulfamerazine 5.Sulfamethazine 6.Sulfamethizole 7.Sulfamethoxypyridazine 8.Sulfachloropyridazine 9.Sulfamethoxazole 10.Sulfamethoxine Phenyl Hexyl 3 x 100 mm Bonus RP 3 x 100 mm G6240 TOF 1200 SL

How else can column selection affect method development? Studies demonstrate that column elution order can be very different on an endcapped versus un-endcapped column. The use of Cyano columns can be helpful when separating mixtures of polar and non-polar columns. HILIC columns may work well for compounds not very well retained on a reversed-phase column, which is helpful when exploring polar compounds (e.g., dopamine, epinephrine, and norepinephrine). As will be explored in the next section, it may also be helpful to use automation software for column screening. One system from Agilent (the 1260 Infinity II Method Development System), for instance, can automatically screen up to 100 different column conditions, 15 mobile phases, or four different columns, while another model (the 1290 system) can automatically screen even more conditions and columns. Strategy #2: Automated HPLC Method Development Another option for HPLC method development is to use dedicated method development software to predict the best possible separation conditions. A workflow solution for automated method development has numerous benefits in terms of improved efficiency in methods development, such as: Reduced instrument idle time Less manual interaction with the LC system Less time spent on manual data analysis, yielding more time for other tasks Better system performance Cost efficiency Automated column switching. As noted, trying different columns is an important aspect of optimizing separation conditions. Screening stationary phases with automated column-switching valves makes this task faster and easier. For instance, the Agilent 1290 Infinity II column compartment and the column selection valve enables a method developer to test a variety of stationary phases and easily assess a wide range of column selectivities. The 1290 Infinity II column compartment can hold as many as eight columns per compartment. Four compartments can be added to a system, allowing a total of 32 columns to be tested without manual system interaction. The Agilent 1290 Infinity II column compartment has two independent heating zones, which is useful for preheating the next column if temperatures are screened, thus resulting in reduced equilibration time. Colum ID tags. If process developers do not know the history of a column s use, it can make it difficult to interpret results (e.g., Could poor results be attributable to a broken or old column?). Insufficient tracking of column usage, therefore, might lead to false negative results. It is critical to know the number of injections performed on columns as well as the ph and temperatures that the column was exposed to in the past. One solution are column ID tags, which are available on Agilent InfinityLab Poroshell columns. These tags store important information about columns such as the number of injections performed, the dates of the first/last injections, ph and temperature stability information, a full product description (including serial and batch numbers), and more. The affiliated software will also warn users if incompatible methods are selected for a column that could harm the column. Automated solvent selection. When developers want to optimize a separation method by screening mobile phase solvents, automation provides great efficiency. In a manual method, re-priming solvent lines and re-equilibrating a system in new mobile phase solvents can take a considerable amount of time. Instead, using a solvent selection valve with Agilent technology lets users choose from as many as 12 different mobile phases (e.g., buffers, organic solvents, or various modifiers); additional solvent selection valves can be added. Method development software. Two major software approaches can be used to achieve automated method development. The first involves method prediction software, which uses algorithms to predict and test separation conditions in iterative cycles. A combination of Agilent hardware (e.g., 1260 Infinity II and 1290 Infinity II) and third-party software can be used to accomplish such method prediction. Examples of third-party software, all aligned with quality by design (QbD) as the foundation of their methodology, include: ChromSwordAuto, an intelligent chromatography data system with four modules, viz. for method scouting (e.g., to quickly find the most promising combination of stationary and mobile phases); optimization of isocratic and gradient methods and fine optimization for late-stage projects; development of robust quality control methods and their transfer through the automatic creation, performance, and analysis of different experiment designs; and data browsing analyzing and report generation (see Figure 7). Fusion QbD, a modular informatics platform for various chromatography types including reversed phase, normal phase, chiral, HILIC, ion exchange, size exclusion, and SFC. Modules include those for LC method development with intelligence that minimizes run time and maximizes consistency; analytical method validation with a suite of experiments required for the validation of analytical methods with full CDS automation; product development for formulation and process optimization as well as non-lc method development (e.g., GC, CD, dissolution, and MS

detection optimization); respiratory drug development; and QC testing (see Figure 8). Equipment setup is simple and conforms to the instrument s configuration, though it can expand beyond the equipment s capabilities if need be. The software automatically generates a complete experiment to support the variables that have been aggregated for the study and then reconstruct that experiment in the chromatography data software. Figure 7: ChromSword: Supports method development strategies for earlier and later stages of product development. The ACD/AutoChrom software was designed to address the challenges that analysts face when developing stabilityindicating methods that require baseline resolution between the drug product, process impurities, degradants, and excipients (see Figure 9). The software uses a five-step process centered on QbD quality, which includes instrument control (e.g., compatible with Agilent ChemStation, Agilent OpenLAB ChemStation Edition, and others); automated data processing and peak tracking using a solvent subtracted UV spectrum and LC/MS deconvolution as well as identification of molecular ions; an advanced processing Full QbD Experiment Automation ChemStation OpenLAB interface to review and adjust processing and peak tracking results, if necessary; a dashboard design that allows simultaneous comparison of various pieces of data (chromatographic data and associated spectral, structural, and text information) to enable faster more informed decisions; chromatographic simulation for method optimization and robust methods design; Figure 8: Fusion QbD: Modular software platform. LC Method Development Analytical Method Validation Product Development Formulation & Process Development Non-LC Method Development Respiratory Drug QC Testing and the ability to generate high-quality results or to upload the data to a searchable and sharable knowledge base. The user can edit the QbD strategy and set critical quality attributes at any stage. Method scouting software. Another approach to achieving automated method development is method scouting. For

instance, one can use Agilent s 1260 Infinity II and 1290 Infinity II instruments combined with the firm s Method Scouting Wizard software to set up even complex screening campaigns in a simple and automated manner. In 10 steps, experienced or novice developers can design workflows for even advanced workflows such as chiral separations. Users define their process; select columns and solvents from a preloaded drop-down menu; define binary, tertiary, or quaternary gradients and temperatures; review and select screening methods; and define injection volumes and repetitions. The system automatically warns users if something is incompatible. Once the scouting runs are completed and the data are collected for the screening conditions, the OpenLAB ChemStation (CDS) data analysis and reporting functionalities that help users process the scouting data quickly and efficiently. Find your most suitable results by visualizing data in bubble plots and filtering results can for different quality criteria (e.g., number of peaks, maximum/minimum resolution, and maximum symmetry). Simplified Methods Transfer One factor in simplified methods transfer is the choice of columns. The Poroshell family of columns was designed with various particle sizes (1.9 µm, 2.7 µm, and 4 µm) for scalability and simpler transferability. For instance, the selectivity developed on a 2.7-µm Poroshell column can be readily transferred to either a 4-µm or a 1.9-µm Poroshell column. To reduce unwanted variations in retention behavior after transfer of your method to other target systems, rely on Figure 9: ACD/AutoChrom: Five steps to improve complex chromatography. Communicate Results Chromatographic Simulation Control of Agilent ChemStation Automated Peak Tracking Data Management Intelligent System Emulation Technology (ISET) as a part of the 1290 Infinity II Method Development System. ISET gives you the opportunity to directly develop methods with on the fly target system emulation. With ISET, your method development system mimics the characteristics of the legacy target instruments and reduces variations to a minimum. Summary HPLC method development can be made more efficient in several ways such as by proper column selection, a wellplanned and systematic off-line approach, or the use of hardware and software for automated method development. In the end, efficient method development requires an appropriate balance between the initial investment and spending on developing methods. When evaluating new technologies, protect your investment by choosing instruments that are scalable, flexible with upward and downward compatibility with existing LC systems, and those that provide the highest degree of automation possible. Agilent publication number: 5991-8285EN Published in the USA, July 17, 2017