Ion sources. Ionization and desorption methods

Similar documents
Ionization Techniques Part IV

Fundamentals of Mass Spectrometry. Fundamentals of Mass Spectrometry. Learning Objective. Proteomics

Lecture 15: Introduction to mass spectrometry-i

Chemistry 311: Topic 3 - Mass Spectrometry

Mass Spectrometry. Electron Ionization and Chemical Ionization

M M e M M H M M H. Ion Sources

1) In what pressure range are mass spectrometers normally operated?

Chemistry Instrumental Analysis Lecture 34. Chem 4631

Mass Spectrometry (MS)

Harris: Quantitative Chemical Analysis, Eight Edition

4. How can fragmentation be useful in identifying compounds? Permits identification of branching not observed in soft ionization.

Mass Spectrometry. General Principles

TANDEM MASS SPECTROSCOPY

Lecture 8: Mass Spectrometry

LC-MS Based Metabolomics

high temp ( K) Chapter 20: Atomic Spectroscopy

Lecture 8: Mass Spectrometry

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n

Emission spectrum of H

Mass spectrometry and elemental analysis

Chemistry Instrumental Analysis Lecture 37. Chem 4631

MS Goals and Applications. MS Goals and Applications

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K. Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

MASS SPECTROMETRY. Topics

Introduction to Pharmaceutical Chemical Analysis

Qualitative Organic Analysis CH 351 Mass Spectrometry

Secondary Ion Mass Spectrometry (SIMS)

3 - Atomic Absorption Spectroscopy

A process whereby an electron is either removed from or added to the atom or molecule producing an ion in its ground state.

Matti Laan Gas Discharge Laboratory University of Tartu ESTONIA

Mass Spectrometry in MCAL

Other Methods for Generating Ions 1. MALDI matrix assisted laser desorption ionization MS 2. Spray ionization techniques 3. Fast atom bombardment 4.

Electron temperature is the temperature that describes, through Maxwell's law, the kinetic energy distribution of the free electrons.

INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms

Laser Dissociation of Protonated PAHs

Molecular Mass Spectrometry

Questions on Instrumental Methods of Analysis

1. The range of frequencies that a measurement is sensitive to is called the frequency

(Refer Slide Time 00:09) (Refer Slide Time 00:13)

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

Mass Spectroscopy. Base peak. Molecular Ion peak. The positively charged fragments produced are separated, based on their mass/charge (m/z) ratio. M+.

very high temperature for excitation not necessary generally no plasma/arc/spark AAS

Mass spectrometry gas phase transfer and instrumentation

Welcome to Organic Chemistry II

Chemistry Instrumental Analysis Lecture 17. Chem 4631

( 1+ A) 2 cos2 θ Incident Ion Techniques for Surface Composition Analysis Ion Scattering Spectroscopy (ISS)

Theory of Gas Discharge

Instrumental Analysis

2101 Atomic Spectroscopy

TMT4320 Nanomaterials November 10 th, Thin films by physical/chemical methods (From chapter 24 and 25)

Chemistry Instrumental Analysis Lecture 18. Chem 4631

Ionization Methods in Mass Spectrometry at the SCS Mass Spectrometry Laboratory

for the Novice Mass Spectrometry (^>, John Greaves and John Roboz yc**' CRC Press J Taylor & Francis Group Boca Raton London New York

Ch 13. Basics of Mass Spectrometry (I) : Principles & Ionization Sources

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

MS731M: Chemical Analysis of Materials

Auxiliary Techniques Soft ionization methods

Chapter 4 Scintillation Detectors

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP

PHYSICAL VAPOR DEPOSITION OF THIN FILMS

MS/MS .LQGVRI0606([SHULPHQWV

Chapter 14. Molar Mass Distribution.

Ch. 8 Introduction to Optical Atomic Spectroscopy

Combinatorial RF Magnetron Sputtering for Rapid Materials Discovery: Methodology and Applications

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

MS Goals and Applications. MS Goals and Applications

Chem 550, Spring, 2012 Part I: OVERVIEW OF MASS SPECTROMETRY:

CHROMATOGRAPHY AND MASS SPECTROMETER

15.04.jpg. Mass spectrometry. Electron impact Mass spectrometry

Optical Atomic Spectroscopy

12. Structure Determination: Mass Spectrometry and Infrared Spectroscopy

Huashun Zhang. Ion Sources. With 187 Figures and 26 Tables Э SCIENCE PRESS. Springer

Extrel Application Note

Mass spectrometry.

Mass Spectrometry. A truly interdisciplinary and versatile analytical method

1.4 The Tools of the Trade!

Introduction to Spectroscopic methods

Mass Spectrometry (MS)

CHAPTER A2 LASER DESORPTION IONIZATION AND MALDI

Atomic Emission Spectroscopy

Gas 1: Molecular clouds

Secondaryionmassspectrometry

Vacuum-Ultraviolet-Excited and CH 2 Cl 2 /H 2 O-Amplified Ionization- Coupled Mass Spectrometry for Oxygenated Organics Analysis

ICPMS Doherty Lecture 1

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

Methods of surface analysis

Characteristics and classification of plasmas

Selecting an LC/MS Interface Becky Wittrig, Ph.D.

SCALING OF PLASMA SOURCES FOR O 2 ( 1 ) GENERATION FOR CHEMICAL OXYGEN-IODINE LASERS

Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT.

[2] (b) An electron is accelerated from rest through a potential difference of 300 V.

Introduction to the Q Trap LC/MS/MS System

Plasma Radiation. Ø Free electrons Blackbody emission Bremsstrahlung

a. An emission line as close as possible to the analyte resonance line

Chapter III: III: Sputtering and secondary electron emission

Total probability for reaction Yield

Ch. 9 Atomic Absorption & Atomic Fluorescence Spectrometry

Choosing the metabolomics platform

Current Research and Future Plans

Transcription:

Ion sources Ionization and desorption methods 1

2

Processes in ion sources 3

Ionization/ desorption Ionization Desorption methods Electron impact ionization Chemical ionization Electro-spray ionisation Ion-spray ionisation Inductively coupled plasma MS Surface ionization Field ionisation Secondary ion mass spectrometrysims Liquid-SIMS, LSIMS Fast atom bombardment, FAB Laser desorption, LD Laser ablation/ laser ionization, LA/LI, Matrix assisted laser desorption, MALDI Plasma desorption, PD 4

Introduction: different types of ion sources 5

Ionization 6

Concept underlying the electron impact ionization The thermoionic electron emission is described by the Richardson equation: (W-tungsten) For T=2 10 3 K, the power of thermoionic emission is described: 7

Electron impact ionization The gas phase sample: -N 2, C 6 H 6 etc -Sample diluted in carrier gas Molecular beams, effusion beams Ionization by 70-100 ev electrons The solid phase sample can be evaporized in an oven T.D. Märk & G.H. Dunn, Electron Impact Ionisation, Springer 1985 8

Characteristic of some suitable materials ev 9

Single charged species 10

For multiple charged species 11

Ionization potentials for selected molecules 12

Ionization by collision with electrons 13

Excitation scenario diagram 14

Appearance potential ionization energy The ionization efficiency curve can be produced by the normal electron impact I ion (Uel). The ionization efficiency curve can be modeled and appearance potential can be determined. One should include into model that electron energy Is not mono-chromatic but it is described by Maxwell-Boltzman thermal distribution. 15

Electron Impact In the electron impact ionization method one can generate pool of the excited molecules, which can fragment during the flight to the detector leading to so-called ghosts mass speaks (Geister Linien) : 16

Electron impact generation of the negatively charged ions Dissociative ion pair formation Dissociative electron attachment Electron attachment 17

Ion formation formulae With the gas density n 0, the cross section σ, the electron pathway s, electron current I e, and probability β that ion will be formed in ionization volume: the ion current can be determined to be: n 0... Density gradient can occur Ie... Ions can be lost on electrodes, Repeller, Wände,... β... depends on formation location and can vary significantly depends on the gas temperature, initial energy of ions and fragments generated in dissociative ionization, The average values are taken for the calculation 18

Example 19

The ionization cross section, σ 20

Partial cross sections for multi-ionization Total cross-section: Ionization is stepwise: 21

examples 22

The ionization cross section obtained for electrons scattering with 90 ev The absolute cross sections can only be determined for the limmited number of molecules T.D. Märk & G.H. Dunn, Electron Impact Ionisation, Springer 1985 23

24

25

26

27

28

29

Chemical ionization In the chemical ionization, the sample is ionized in ion-molecule reaction in the environment saturated by ions. One uses Methan, Iso-Buthan, or Ammoniak atmospheres, to generate their cations. 30

Chemical ionization Hybrid abstraction Charge exchange The ionization potential of the ionized gas must be larger than reactant. If the IP>5 ev the spectra will be similar to that obtained by electron ionization Ion molecule reactions J. Watson, Introduction to Mass Spectrometry, 2nd Ed., Raven Press, New York, 1985 A.G. Harrison, Chemical Ionisation Mass Spectrometry, CRC Press, Boca Raton, 1992 31

Chemical Ionization 32

Chemical Ionization 33

34

Proton transfer mass spectrometry 35

Proton transfer mass spectrometry Chemical ionization for the diagnostic of organic molecules: Hollow cathode discharge generates H 3 O + Drift Tube very small fragmentation Quantitative no GC neccesary 1 part of 2 10 11 can be detected Application in the field of environmental research -Medicine, food industry 36

Atmospheric pressure ion source Ionization under atmospheric pressure in corona discharge or 63 Ni source API sources ionized in sensitive way trace elements in the air It can be coupled to liquid chromatography (LC): Ion spray ion spray ionization inductively coupled plasma (ICP) High energy electrons interact with Nitrogen contained in the air or through the interface One obtains thermalized electrons: 37

Electrospray ionization (ESI) The sample is transported in capillary A high voltage is applied to the end of capillary 38

ESI 39

Electrospray ionization (ESI) 40

Electroionization: determination of the mass from the spectra 41

42

Inductively coupled plasma mass spectrometry * Highly precise analysis and detection of trace elements: geochemistry, biology * Sample is provided through cappillare and laminar Ar beam is used * High frequency (e.g., 30 MHz) discharge * Atomization and generation of single ionized cations mass spectra are simple to interpret * High sensitivity typicaly 50 ng/l 50 ppt 43

Glow discharge ion source/planar cathode 44

Gas discharge ion source /pin-shaped cathode 45

Grimm-type dc glow discharge ion source 46

Surface ionization Ions emited from the hot metal surfaces are partially ionized. The ionization yield: N + is number of evaporated ions per time unit (rate of ion generation), N 0 is number of neutral particles evaporated from the surface, g + statistical weigthing of ions (number of states), g 0 statistical weighting of neutrals Φ the surface exit energy I Ionization potential of evaporated ions Valid for molecular ionization molecules Valid at thermostatic equilibrium between the surface and the sample 47

48

Example of the thermoionic efficiencies (k B T 1 ev for T 11 000 K). In general, the dependance is more complex, thickness of the surfaces, contamination, lack of thermal equilibrium! 49

Elements analyzed by thermal surface ionization 50

Field ionization the potential for a external electron in the molecule in the field-free space and in the external field Elektrons originated from A level are in a strong electric field and are not bound; electrons from B level can tunnel through the potential barrier Field ionization conditions requires the electric field of a few V pro atom or molecule radi size, e.g., 1 V/0.1 nm, gives electric field strength of 10 10 V/m 51

Field ionization The stark field can be obtained using a sharpe edges. Approximating the edge by rounded surface one can estimated the field parameters and radius of the edge: For E(R 0 ) = 10 10 V/m and to get U = 10 kv One need the radius of the edge of: The molecule adsorbed on the edges are desorbed by the field and ionized: field desorption and ionization 52

Field ionization Ions generated by field ionization possess relatively narrow energy distribution The fragmentation is minimal Practically no excited ions Ion currents of 10 12 A - 10 10 A Generation of M +, [M+H] +, [M+Na] +, [M+K] +, H.D. Beckey, Field Ionisation Mass Spectrometry, Pergamon Press, Oxford, 1970 53

Desorption methods 54

Particle desorption by energetic ion or neutral beams Atoms and molecules can be sputtered from the surface by fast ion or neutral beams (kev). A part of sputtered material is ionized. Application of neutral beams to sputter the material is known as Fast Atom Bombardment FAB. The efficiency of sputtering depends on kind of primary particles used and sample material, energy and the incident angle charge state, crystal structure and material reorganization. Benninghoven, Rüdenauer, Werner, Secondary Ion Mass Spectrometry, Chemical Analysis, Vol 86 (1987). 55

Sputtering 56

Sputtering 57

α... Dimensionless parameter dependant on M2 / M1 C 0... Constant from the Sputtering theory Sn... Nuclear scattering losses for primary energy Ep US... Binding energie of surface 58

Surface oxidation 59

Fast atom bombardement FAB 60

How to get fast atoms 61

FAB Since 1980, FAB is used as a soft ionization technique in organic mass spectrometry. It is convinient method because of a low fragmentation efficiency. The mother molecule (protonated or di-protonated) is the main mass. Sometimes Na is used. Peptides are up to 10 4 amu/q0 observed. A large mass range and a high resolution spectrometers are required. 62

Photoionization The ionization energies are larger than 5 ev, in most cases larger than 9 ev. For the one-photon ionization radiation energy of the photons is in VUV region (the sample and the radiation source have to be used in the vacuum environment). High photon energy can be generated using hydrogen, deuterium lamps, synchrotron radiation or in by the laser wave mixing methods. The photoionization cross sections are 10-100 times smaller than that of electron impact ionization. (hν > IP). For NO (IP = 9.5 ev), Lyman-α radiation with λ =121.6 nm (hν = 10.2 ev) can be used for the ionization. 63

Photoionization vs electron impact The photoionization is convinient to measure ionization energy (e.g., synchrotron radiation, laser VUV source). Photoionization spectra are simple with a small fragmentation efficiency. 64

Laser photoionization resonant nonresonant IP Resonance ionization mass spectrometry (RIMS) -low laser power resonant steps -selective -high efficiency, low fragmentation Multiphoton ionization (MPI) -high laser power, non-resonant steps -non-selective, but power dependance on laser power -low efficiency, high fragmentation 65

Selectivity and high ionization rate are high for atoms 66

67

Resonance Ionization Mass Spectrometry: example 187 Os / 187 Re measurement 68

Resonance Ionisation Mass Spectrometry: example 187 Os / 187 Re measurement 69

Photoionisation: non-resonant Laser Ionization 70

Non-resonant photoionization:sims-sali SIMS secondary ion mass spectrometry SALI surface analysis mass spectrometry 71

SIMS 72

Non-resonant photionization 73

Non-resonant ionization 74

Non-resonant ionization of molecules 75

Laser desorption/laser ablation 76

Laser desorption 77

Laser Ablation / Laser Ionisation 78

Laser ablation 79

Laser Ablation / Laser Ionization 80

Matrix assisted laser desorption and ionization MALDI is soft ionization method for detection of large molecules for which other methods or not sensitive. MALDI can be well applied for the detection of biopolymers (Proteine, Glycoproteine, Oligosaccharide, Oligonucleotide). The samples are cristilized in a matrix. The trix acts as absorber of laser radiation and subsequently enables its evaporation. The sample is mixed with the ratio 1: 1000-1:5000 with the matrix molecules. Matrix UV Laser (broadband UV absorption): cinnamic acid sinapinic acid... IR Laser: nicotinic acid glycerin Succinsäure, Harnstoff Kaffeesäure,... 81

Diagram of MALDI source 82

Characteristics: MALDI > Mass range: up to 500 kdalton > Positive Ions Protonated, [M+H] + Protonated, [M+2H] 2+ Negative Ions [M H] > positive und negative cluster Ionen [2M+H] +, [2M H] > No fragmentation > Sensitivity in fmol region > Sample preparation is alchemie > Matrix and sample should have non-overlap absorption bands > Laser pulses should be short 83

MALDI matrix 84

85

MALDI: example 86

87

Summary of ionization methods 88

Summary of desorption methods 89