Prompt γ radiation measured with a Nal scintillation detector: a beam monitor for neutron scattering instruments which needs no space in the beam

Similar documents
Precision neutron flux measurement with a neutron beam monitor

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

PHYS 3650L - Modern Physics Laboratory

Small Angle Neutron Scattering in Different Fields of Research. Henrich Frielinghaus

EEE4106Z Radiation Interactions & Detection

Detection and measurement of gamma-radiation by gammaspectroscopy

Determination of the boron content in polyethylene samples using the reactor Orphée

PoS(FNDA2006)033. Radiography Using Fission Neutrons. Thomas Bücherl. Christoph Lierse von Gostomski

Applied Nuclear Physics (Fall 2006) Lecture 21 (11/29/06) Detection of Nuclear Radiation: Pulse Height Spectra

Application of positrons in materials research

This paper should be understood as an extended version of a talk given at the

Compton suppression spectrometry

Measurement of nuclear recoil responses of NaI(Tl) crystal for dark matter search

Measurements of Neutron Total and Capture Cross Sections at the TOF spectrometers of the Moscow Meson Factory

Conclusion. 109m Ag isomer showed that there is no such broadening. Because one can hardly

School on Pulsed Neutrons: Characterization of Materials October Neurton Sources & Scattering Techniques (1-2)

Total probability for reaction Yield

The intense, pulsed positron source EPOS at the Research Centre Dresden-Rossendorf

Copyright 2008, University of Chicago, Department of Physics. Experiment VI. Gamma Ray Spectroscopy

Sensitivity of sodium iodide cryogenic scintillation-phonon detectors to WIMP signals

Absorption and Backscattering of β-rays

The 46g BGO bolometer

DETECTORS. I. Charged Particle Detectors

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

An autonomous underwater telescope for measuring the scattering of light in the deep sea

1.4 The Tools of the Trade!

Absorption and Backscattering ofβrays

Good Vibrations Studying phonons with momentum resolved spectroscopy. D.J. Voneshen 20/6/2018

Interactive Web Accessible Gamma-Spectrum Generator & EasyMonteCarlo Tools

V. 3. Development of an Accelerator Beam Loss Monitor Using an Optical Fiber

An idea for the future proton detection of (p,2p) reactions with the R 3 B set-up at FAIR

1.5. The Tools of the Trade!

Week 7: Ch. 10 Spec. w/ Scintillation Ctrs. Photomultiplier Devices

GAMMA RAY SPECTROSCOPY

Radioactivity. Lecture 6 Detectors and Instrumentation

Neutron and x-ray spectroscopy

PoS(TIPP2014)033. Upgrade of MEG Liquid Xenon Calorimeter. Ryu SAWADA. ICEPP, the University of Tokyo

New Concept of EPOS Progress of the Mono-energetic Positron Beam (MePS) Gamma-induced Positron Spectroscopy (GiPS)

SECTION 8 Part I Typical Questions

Study well-shaped germanium detectors for lowbackground

The Ones That Got Away: Neutron Leakage Spectra

An industrial radiography exposure device based on measurement of transmitted gamma-ray intensity

A Measurement of Monoenergetic Neutrons from 9 Be(p,n) 9 B

PoS(ICHEP2016)474. SoLid: Search for Oscillations with a Lithium-6 Detector at the SCK CEN BR2 reactor

Lecture 3 - Compton Scattering

AGATA preamplifier performance on large signals from a 241 Am+Be source. F. Zocca, A. Pullia, D. Bazzacco, G. Pascovici

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Nuclear Reactions A Z. Radioactivity, Spontaneous Decay: Nuclear Reaction, Induced Process: x + X Y + y + Q Q > 0. Exothermic Endothermic

RDCH 702 Lecture 8: Accelerators and Isotope Production

Gamma-Spectrum Generator

Scintillation Detector

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

Gamma ray coincidence and angular correlation

Preparatory experiments for cold-neutron induced fission studies at IKI

Hands on LUNA: Detector Simulations with Geant4

FEASIBILITY STUDY ON A LARGE CHOPPER DISC FOR A TOF SPECTROMETER

TIIVISTELMÄRAPORTTI (SUMMARY REPORT)

MEASUREMENT OF T-ODD EFFECTS IN THE NEUTRON INDUCED FISSION OF

MEASUREMENT AND DETECTION OF RADIATION

Project Memorandum. N N o. = e (ρx)(µ/ρ) (1)

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Radiation Detection and Measurement

X-ray absorption. 4. Prove that / = f(z 3.12 ) applies.

Simulation and optimization of a new focusing polarizing bender for the diffuse neutrons scattering spectrometer DNS at MLZ

Quality Assurance. Purity control. Polycrystalline Ingots

This is a repository copy of TACTIC : The TRIUMF Annular Chamber for Tracking and Identification of Charged particles.

Neutron Instruments I & II. Ken Andersen ESS Instruments Division

Detekce a spektrometrie neutronů. neutron detection and spectroscopy

Research Center Dresden Rossendorf

solenoid and time projection chamber for neutron lifetime measurement LINA

Nuclear Physics and Astrophysics

Simulating Gamma-Ray Telescopes in Space Radiation Environments with Geant4: Detector Activation

Neutron and gamma ray measurements. for fusion experiments and spallation sources

Measurements of liquid xenon s response to low-energy particle interactions

Correction for PMT temperature dependence of the LHCf calorimeters

THE COMPTON EFFECT Last Revised: January 5, 2007

Nuclear Physics. (PHY-231) Dr C. M. Cormack. Nuclear Physics This Lecture

Gamma Spectroscopy. References: Objectives:

The EDM Polarimeter Development at COSY-Jülich

Chemistry Instrumental Analysis Lecture 19 Chapter 12. Chem 4631

DETERMINATION OF CORRECTION FACTORS RELATED TO THE MANGANESE SULPHATE BATH TECHNIQUE

Design, Construction, Operation, and Simulation of a Radioactivity Assay Chamber

Figure 1. Decay Scheme for 60Co

Gamma Ray Spectroscopy

Electromagnetic modulation of monochromatic neutrino beams

Neutronics Experiments for ITER at JAERI/FNS

Prompt γ-rays from Neutron Inelastic

Comparison of the Photo-peak Efficiencies between the Experimental Data of 137 Cs Radioactive Source with Monte Carlo (MC) Simulation Data

Development of a Dosimetric System using Spectrometric Technique suitable for Operational Radiation Dose Measurements and Evaluation

GAMMA RAY OPTICS. Michael Jentschel Institut Laue-Langevin, Grenoble, France

Plasma Accelerator for Detection of Hidden Objects by Using Nanosecond Impulse Neutron Inspection System (NINIS)

Rb, which had been compressed to a density of 1013

The interaction of radiation with matter

The Mössbauer Effect

Shielded Scintillator for Neutron Characterization

6 Neutrons and Neutron Interactions

Science of Nuclear Energy and Radiation a Comprehensive Course for Science Teachers June 22-25, 1998 McMaster University

Slides by: Prof. Abeer Alharbi

THE INSTALLATION FOR EXPERIMENTAL NEUTRON SPECTRA RESEARCH IN REACTOR MATERIALS COMPOSITIONS Hliustin D.V. Institute for Nuclear Research, Moscow,

Physics 23 Fall 1989 Lab 5 - The Interaction of Gamma Rays with Matter

Transcription:

Journal of Physics: Conference Series OPEN ACCESS Prompt γ radiation measured with a Nal scintillation detector: a beam monitor for neutron scattering instruments which needs no space in the beam To cite this article: O Holderer et al 2014 J. Phys.: Conf. Ser. 528 012038 Related content - Grazing incidence neutron spin echo spectroscopy: instrumentation aspects and scientific opportunities O Holderer, H Frielinghaus, S Wellert et al. - Extension of the VITESS polarized neutron suite towards the use of imported magnetic field distributions S Manoshin, A Rubtsov, V Bodnarchuk et al. - The first prototype diamond monochromator at the Institut Laue- Langevin P Courtois, M T Fernandez-Diaz, G Nenert et al. View the article online for updates and enhancements. This content was downloaded from IP address 37.44.198.239 on 05/02/2018 at 23:51

Prompt γ radiation measured with a NaI scintillation detector: a beam monitor for neutron scattering instruments which needs no space in the beam O. Holderer 1, M. Zamponi 1, M. Monkenbusch 2, R. Engels 3 1 Jülich Centre for Neutron Science (JCNS) at Heinz Maier-Leibnitz Zentrum (MLZ), Forschungszentrum Jülich, Lichtenbergstr. 1, 85747 Garching, Germany 2 Jülich Centre for Neutron Science (JCNS) and Institute for Complex Systems, Forschungszentrum Jülich, D-52425 Jülich, Germany, 3 Zentralinstitut für Engineering, Elektronik und Analytik, Systeme der Elektronik (ZEA-2), Forschungszentrum Jülich, D-52425 Jülich, Germany E-mail: o.holderer@fz-juelich.de Abstract. We investigate the possibility of using the prompt γ rays emitted by aluminum windows in order to monitor the neutron flux of the beam. A NaI scintillation detector is used to detect the prompt γ rays. No additional material apart from the unavoidable Al windows along the flight path is placed in the beam. The performance of the monitor is compared to that of a standard BF3-monitor placed in the beam. Influences of a magnetic field on the photomultiplier of the NaI monitor is discussed, as well as the influence of activation gammas. At an instrument using a beam chopper the time behaviour is discussed. 1. Introduction Usually, beam monitors are placed directly in the beam, with a low sensitivity of the order of 10 4 10 5, for example He3- or BF3 monitors. Two Al-windows are unavoidably in the beam, producing some absorption, small angle scattering and prompt γ radiation. Prompt γ rays from absorption of neutrons in aluminum, which is the main material used for neutron windows, have a significant contribution of high energetic gammas, e.g. for aluminum at 7.7 MeV [1] [2], compared with the activation γ s of Al which have the main contribution at 1.7 MeV [3]. With a scintillation detector sensitive also to high energy gammas, where the lower energies are filtered and discriminated away, a beam monitor without additional material besides the unavoidable Al windows can be realized. Fig. 1 shows the absorption coefficient of lead as a function of energy. At the J-NSE neutron spin echo spectrometer [4, 5, 7] at the FRM II research reactor of the TU München this type of monitor has been tested and showed that it is a relieble and easy to handle possibility as a beam monitor. Time behaviour has been investigated at the SPHERES backscattering spectrometer [8], where the phase space transform chopper imposes a time structure on the incoming neutron beam. 2. Experiment Neutron capture releases the total binding energy of a neutron in the order of 8-10 MeV mostly in the form of one to several γ-photons. A fraction of them has energies close to the total binding Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. Published under licence by Ltd 1

Figure 1. Absorption coefficient of lead as a function of γ energy and the relative intensities of the prompt γ lines of Al (with more than 10% rel. intensity). Lead acts as a filter for the low energy γ s from the produced radioactive Al-28 isotope and others. A very strong prompt γ line is above 7 MeV. Data from [9] and [2]. Figure 2. Pulse height spectra of the NaI monitor. The peak at closed beam is attributed to the 511 kev line of positron-electron annihilation. The low energy part contains contributions from activation decay, the high energy part is used for monitoring the beam intensity. energy. To detect them with a scintillation detector, a rather large scintillation crystal is needed. In the present experiment, we used a SCIONIX scintillation detector with a NaI crystal with 76 mm diameter and 102 mm height to detect the prompt γ s emitted from the aluminum windows in front of the first main coil of the J-NSE spectrometer. A first data acquisition of the analog output has been taken with the monitor placed 2

Figure 3. Monitor at a position about 1.4m away from the beam, without lead shielding. Count rate at the analog output with open beam (blue) and closed beam (green) as a function of time. Counting time was 1 sec. The activation gammas are counted with these settings. Fits are made with the Al decay time of T 1/2 =2.246 min. underneath the lead shielding of the polarizing unit. The built up and decay of the activation γ s from Al is clearly visible in the analog pulses of the monitor (see Fig. 3). Without the lead acting as a high energy filter, the NaI scintillator alone is therefore not suitable as a beam monitor. The situation can be improved by pulse height discrimination, by this means the obvious Al-28 decay signal with a half life time of 2.2414 minutes (BNL data) (β-decay with associated 1.78 MeV γ [6] ) can be suppressed. However, the high count rate of low energy γ s leads to saturation and dead-time effects. 2.1. Low γ discrimination The use of the NaI-counter as a beam monitor requires that the low energy γ s are cut away. As a filter, the lead hutch of the second polarizer of the J-NSE spectrometer has been used with its 50 mm thickness to absorb preferentially the low energy γ rays. Fig. 2 shows pulse height spectra of the analog signal of the NaI-monitor with the shutter of the neutron beam open ( beam open ) and closed ( beam closed ) respectively. With the beam closed, there is still a significant number of pulses at lower energies. They can be filtered to a large amount by some (typically 5 cm) Pb. With an adequate threshold for the digital signal it is possible to be only sensitive to the high energy part of the spectrum, where most of the prompt γ-lines from Al are found. Measurements with a Pb filter realized by placing 50 mm lead between the NaI detector and the beam (i.e. placing the detector outside the lead hutch of the long wavelength polarizer) lead to a significantly reduced count rate of the analog output by one order of magnitude. The ratio of the count rate with beam open and close is about 15 and can be largely improved for the digital output in combination with the appropriate trigger level. A good sensitivity to the prompt γ rays could be achieved in this way. In this monitor position, the count rate with the reactor not running is at about 2-3 cps. The remaining dark count rate is due to the unshielded position towards other instruments in the neutron guide hall. 3

Figure 4. Detector counts versus number of measurement. Top: BF3-monitor, Bottom: NaImonitor. Spectra are time normalized (counts per second), resulting in the different levels of noise. 2.2. Reliability The time behaviour of the NaI monitor has been compared to the J-NSE standard BF3 monitor which covers the whole beam cross section. A series of 220 measurements is shown in Fig. 4. The standard deviation has been compared to the square root of the count rate as the expected deviation. The standard deviation of the BF3-Monitor is 2369, the square root of the count rate 829, while for the NaI monitor (with a much lower overall absolute count rate) the two values are 73 and 69 respectively. It has to be pointed out that above a magnetic field of 2 Gauss at the position of the monitor the photomultiplier tube of the NaI counter has been disturbed despite of it s µ-metal shielding. The count rate drops for a field of 4 Gauss by 1/3 compared to the undisturbed count rate. A suitable magnetically stable position is therefore needed. 2.3. Time behaviour At the backscattering spectrometer SPHERES the time response of the monitor has been tested. At the end of the neutron guide, inside the housing of the spectrometer, is the phase space transform chopper with added Pb shielding on the front side. The monitor was placed on top of the neutron guide shielding towards the chopper. Shown are counts versus chopper phase for monitor (shifted by 20000) and a large angle detector tube (separated into elastic and inelastic counts). The count rate of the NaI monitor followed exactly the opening and closing of the chopper. The Al windows in front of the chopper produce a constant high count rate which is then modulated with the frequency of the chopper. The response is as fast as the response of the detectors just behind the chopper (Fig. 5). The steps in the first 24 degree of the chopper open and reflecting phase corresponds to a flight distance of 1.6 m. We attribute it to prompt γ radiation from neutrons passing the Al window in front of the Doppler monochromator. The wiggling structure during the reflection phase of the chopper is in our opinion due to slightly varying absorption of the chopper wheel and reflection of the different crytals. The average count rate is nontheless a measure of the incoming neutron flux. A really time resolved operation needs a good shielding towards all unwanted sources of γ radiation. 4

Figure 5. NaI monitor at a chopper instrument, the backscattering spectrometer SPHERES. The time structure of the monitor follows the opening and closing of the chopper (see text). 3. Practical setup as a beam monitor Fig. 6 shows a sketch of how the prompt γ radiation can be used. A thin lead shielding (depending of the γ intensity) of about 5 cm filters all low energy γ rays, a thicker lead shielding (about 10 cm) blocks the main part of the remaining γ rays from other sources. The general hall background can be substracted with a second background monitor, which has to be placed in a equivalent throughout thick shielding (without the thin lead filter towards the beam). 4. Conclusion The NaI monitor is an easy to use plug-and-play device which performed similarly to the currently used traditional BF3-beam monitor for the 60x60 mm 2 beam. The stability and reproducibility at the same settings of the instrument seems to fulfil the requirements of the NSE spectrometer. It has the advantage that no additional material is needed in the beam since Al-windows producing the required prompt γ s are available anyway and thus no losses of intensity due to the monitor are encountered. The monitor has a slight dependence on the magnetic field of the main coils, it is visible at the present position as a drop of count rate as the B-field is at about 4 Gauss. The signal-to-background ratio at the current position is 20. The monitor was not shielded against γ s coming from other experiments or neutron guides. A better placement and a lead shielding around the monitor will certainly improve the background situation. The magnetic field sensitivity might be reduced with a µ-metal shielding. One only needs to ensure that the influence of the shielding on the magnetic field homogeneity of the instrument is negligible. Between neutron guide and second polarizer would be such a position at the J-NSE spectrometer, which is further away from the main coils with two Alwindows present. The temperature dependence has not been studied, here the monitor has been operated in an air conditioned stable environment. Variations in temperature will have a certain influence on the efficiency of the monitor. The sensitivity to external background could be further reduced by a differential measurement with a second NaI-monitor, placed in an equivalent position concerning the γ sources from other experiments, but shielded against the prompt γ s of the own instrument. 5

Figure 6. Measurement principle for the NaI beam monitor. A thin lead shielding of about 5 cm filters all low energy γ rays, a thicker lead shielding (about 10 cm) blocks the main part of the γ rays from other sources. The general hall background can be substracted with a second background monitor. 5. Acknowledgement We thank Karl Zeitelhack (TU München) for fruitful discussion. [1] http://www-nds.iaea.org/pgaa/tecdoc.pdf. [2] http://www.nndc.bnl.gov/capgam/byn/. [3] G. Pfennig, H. Klewe-Nebenius, W. Seelmann-Eggebert, Karlsruher Nuklidkarte, 6th edition (1998). [4] F. Mezei, ed., Neutron Spin Echo, no. 128 in Lecture Notes in Physics Vol. 128 (Springer, Berlin, Heidelberg, New York, 1980). [5] F. Mezei, P. C., and G. T., eds., Neutron Spin Echo Spectroscopy, no. 601 in Lecture Notes in Physics (Springer, Berlin, Heidelberg, New York, 2003). [6] H.T. Motz and D.E. Alburger, Phys. Rev. 86 (1952) 165. [7] M. Monkenbusch, R. Schätzler, and D. Richter, Nuclear Instruments & Methods In Physics Research Section A-accelerators Spectrometers Detectors And Associated Equipment 399 (1997) 301. [8] J. Wuttke, A. Budwig, M. Drochner, et. al, Rev. Sci. Instrum 83 (2012) 075109. [9] http://physics.nist.gov/physrefdata/xraymasscoef/elemtab/z82.html. [10] D. Richter, M. Monkenbusch, A. Arbe, and J. Colmenero, Adv. Polym. Sci. 174 (2005) 1-221. [11] M. Kerscher, P. Busch, S. Mattauch, H. Frielinghaus, D. Richter, M. Belushkin, and G. Gompper, Phys. Rev. E 83 (2011) 030401(R). [12] H. Frielinghaus, M. Kerscher, O. Holderer, M. Monkenbusch, and D. Richter, Phys. Rev. E 85 (2012) 041408. [13] A.G. Zilman, R. Granek, Phys. Rev. Lett. 77 (1996) 4788-4791. [14] H. Frielinghaus, V. Pipich, A. Radulescu, M. Heiderich, R. Hanslik, K. Dahlhoff, H. Iwase, S. Koizumi, and D. Schwahn, J. App. Cryst. 42 (2009) 681. 6