Evaluation of Indoor Exposition to Benzene, Toluene, Ethylbenzene, Xylene, and Styrene by Passive Sampling with a Solid-Phase Microextraction Device

Similar documents
HYDROCARBONS, AROMATIC 1501

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Validation of. Tetrahydrofuran Using. SKC Passive Sampler Cat. No

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Analysis of BTEX in Natural Water with SPME

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Magnitudes of Back Diffusion During Long-Term Diffusive Sampling of Volatile Organic Compounds Using Carbotrap and Chromosorb 106

Method for the determination of 1,3-butadiene

OPTIMISATION OF SOLID PHASE MICROEXTRACTION (SPME) CONDITIONS FOR HEADSPACE ANALYSIS OF ORGANOPHOSPHATE PESTICIDES IN WHOLE BLOOD

AYOUB ET AL.: JOURNAL OF AOAC INTERNATIONAL VOL. 85, NO. 6,

Calibration for On-Site Analysis of Hydrocarbons in Aqueous and Gaseous Samples Using Solid- Phase Microextraction

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES

SOLID PHASE MICROEXTRACTION FOR TRACE ANALYSIS OF BENZENE IN ENVIRONMENTAL MONITORING

Methods for the determination of vinyl chloride

Air Sampling with Solid Phase Microextraction

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

The Optimal Analytical Method of NDMA(N-Nitrodimethylamine) by Solid-Phase Microextraction Analysis (SPME)

Air Sampling and Analysis of Volatile Organic Compounds with Solid Phase Microextraction

Adsorption Characteristics of Granular Activated Carbon and SPME Indication of VOCs Breakthrough

1. A gas chromatograph equipped with flame ionization detection and a computing integrator.

Determination of Volatile Substances Proof of Food Adulteration

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

VINYL CHLORIDE. Recommended by NIOSH, partially evaluated by OSHA Laboratory.

A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC

Vapor Intrusion Sampling Options: Performance Data for Canisters, Badges, and Sorbent Tubes for VOCs

Application. Gas Chromatography February Introduction

Laser induced fluorescence of polycyclic aromatic hydrocarbons: An approach to gas standards

STANDARD OPERATING PROCEDURES

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

TOTALLY INNOVATIVE MULTIMODE AUTOSAMPLER NEW KONIK ROBOKROM Laboratory Gas Generators An Overview +8 OPERATIONAL MODES MAIN FEATURES

PAL SPME Arrow The Better SPME

PAL SPME Arrow The Better SPME

Determination of Benzene, Toluene, Ethylbenzene and Xylene in River Water by Solid-Phase Extraction and Gas Chromatography

Method for the determination of dimethyl sulfate

ChromTech options for PAL systems

Solid-phase Microextraction in Analysis of Pollutants in the Field

Gas Chromatography (GC)

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009)

anthracene Figure 1: Structures of Selected Polyaromatic Hydrocarbons (PAHs)

FORMALDEHYDE by GC 2541

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Chemistry Gas Chromatography: Separation of Volatile Organics

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

DEVELOPMENT OF A SENSITIVE THERMAL DESORPTION GC-MS METHOD USING SELECTIVE ION MONITORING FOR A WIDE RANGE OF VOCS

Thermal Conductivity Detector Analysis of Hydrogen Using Helium Carrier Gas and HayeSep D Columns

Volatile organic compounds (VOCs):

Chapter 27: Gas Chromatography

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Determination of Limonene Oxidation Products Using SPME and GC MS

DETERMINATION OF UPTAKE RATES FOR VOCs IN AMBIENT AIR BY USING AXIAL TYPE THERMAL DESORPTION PASSIVE TUBES

Application Note 032

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

GC/MS Application Note

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Tar & (SPA) tar analysis

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Rapid Catalyst Screening Reactors

Headspace Solid Phase Microextraction/Gas Chromatography for Determination of Aldehydes

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

GAS CHROMATOGRAPHY (GC)

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Formation of artefacts during air analysis of volatile amines by solid-phase micro extraction

Real-Time Detection: From Gisclard et al.: A Simple Device for Air Analysis. AIHA Quarterly, 14(1):23-25 (1953)

METHYLTIN CHLORIDES 5526

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

GB Translated English of Chinese Standard: GB NATIONAL STANDARD

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

An Innovative Field-Based Analytical Method For Low-Level Detection of Chlorinated Solvents in Groundwater and Soil Samples

Analysis of Pressure Sensitive Adhesives by GC/MS and GC/AED with Temperature Programmable Pyrolyzer

Real Time On-Site Odor and VOC Emission Measurements Using a znose

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Introduction to IH Analytical Chemistry

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Selection of a Capillary

Optimization of a Polydimethylsiloxane Based Passive Sampler of Common Household Volatile Organic Compounds

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Determination of Particulate-Bound Formaldehyde from Burning Incense by Solid Phase Microextraction

Sulfotepp impurities in Chlorpyrifos EC formulations

Equilibrated Vapor Concentrations for Bicomponent Organic Solvents

Experimental Determination of Pyrolysis Products from Carbon/Resin Ablative Materials

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Selection of a Capillary GC Column

Tar measurement by the Solid Phase Adsorption (SPA) method

Pei-Hua Juan and Shih-Wei Tsai

The Effects of Carrier Gas Viscosity and Diffusion on Column Efficiency in Capillary Gas Chromatography

Transcription:

Evaluation of Indoor Exposition to Benzene, Toluene, Ethylbenzene, Xylene, and Styrene by Passive Sampling with a Solid-Phase Microextraction Device Fabrício V. Parreira, Ciomara R. de Carvalho*, and Zenilda de L. Cardeal Chemistry Department ICEx, Federal University of Minas Gerais, Av. Antônio Carlos, 6627, 31270-901 Belo Horizonte, M.G., Brazil Abstract A solid-phase microextraction (SPME) sampling method is developed to evaluate indoor exposure to benzene, toluene, ethylbenzene, xylene, and styrene with gas chromatography and flame ionization detection for quantitative analysis. An SPME holder with a 100-µm polydimethylsiloxane (PDMS) and 65-µm PDMS divinylbenzene fiber coating is tested in different air relative humidity conditions. The method gives good resolution, shows a linear response, is repeatable, and presents high sensitivity. This method is compared with National Institute of Occupational Safety and Health (NIOSH) active sampling. Introduction The monitoring of internal atmospheres is very important because a large number of potentially hazardous volatile organic compounds are produced or used indoors. Benzene, toluene, ethylbenzene, m-xylene, o-xylene, p-xylene, and styrene (BTEXS) are residues present in most solvents. The method most commonly used to determine low-concentration organic compounds in internal atmospheres is calculating the average concentration over a period of time (time weighted average concentration) with passive or active sampling to adsorb the gas on a solid and a later analysis of the components by gas chromatography (GC). There are three types of passive samplers: tube, badge, and radial. Active-sampling methods draw air through a sampling tube by means of a pump. Passive sampling is advantageous to active sampling because * Permanent address: Setor de Medições Ambientais, Fundação Centro Tecnológico de Minas Gerais CETEC, Belo Horizonte, M.G., Brazil. Author to whom correspondence should be addressed: email zenilda@dedalus.lcc.ufmg.br. it does not require a careful calibration routine and expensive air-sampling pumps. However, potentially faulty batteries tend to present constant hurdles to the air-sampling professional, and samplers are affected by ambient pressure and temperature changes during sampling (1 3). Also, passive-sampling devices are currently expensive and are not reusable. High humidity can alter the sorption capacity of a sorbent or the sorption behavior of the exposed inner wall of tube-type samples if condensation occurs (4). A single solid-phase microextraction (SPME) device is an attractive alternative to conventional passive-sampling methods (5 9) because this technique is fast, economical, and the fiber is reusable. This study presents an efficient method to determine indoor exposition to BTEXS by GC analysis with a flame ionization detector (FID) and an SPME sampling device. Experiments with polydimethylsiloxane (PDMS) and PDMS divinylbenzene (DVB) fibers were carried out at different air relative humidity conditions. The fiber that presented the best result in this study was subjected to experiments to determine the equilibrium time, linearity range, repeatability, and detection limit (LOD). In addition, analyses of real samples were carried out in parallel with the National Institute of Occupational Safety and Health (NIOSH) 1501 method. Experimental Chemicals and materials BTEXS were obtained from Merck (Darmstadt, Germany). Carbon disulfide (CS 2 ) was obtained from Riedel-de Haën (Seelze, Germany). Activated charcoal (20 35 mesh), magnesium chloride hexahydrate, sodium chloride, potassium chloride, and potassium nitrate were also obtained from Merck. The 122 Reproduction (photocopying) of editorial content of this journal is prohibited without publisher s permission.

stock standard solution of BTEXS was prepared in a concentration of 1000.0 mg/l using CS 2 as a solvent. Intermediate concentration solutions were prepared according to experimental conditions. The nitrogen, hydrogen, and synthetic air for FID and the synthetic air for the standard gas generation device were obtained from White Martins (Belo Horizonte, Brazil). The synthetic air had a water concentration below 1 ppm. The SPME holder with 100-µm PDMS and PDMS DVB 65-µm-fiber assemblies were purchased from Supelco (Bellefonte, PA). Equipment The chromatographic analyses were performed using an HP 5890 Series II GC (Hewlett Packard, Avondale, PA) equipped with an FID. The analyses were carried out in an HP Ultra 1 capillary column (cross-linked phenyl methylsiloxane film) with a 50- m length and a 0.20-mm i.d. with a 33-µm phase thickness. The oven temperature program for GC began at 35 C and was then raised to 125 C at 10 C/min and then to 250 C at 20 C/min. The carrier gas was hydrogen at a flow rate of 0.90 ml/min at 100 C. A 0.75-mm-i.d. SPME insert was used in the split/splitless injector selected in the splitless mode at 220 C. The detector was set at 280 C in the analysis with the GC FID system. Humidity generation The system used for the dynamic generation of humidity with stopped flow was based on Nelson s system (10) and is illustrated in Figure 1. The dry air is a gas stream prepared using compressed air and scrubbers. The humid air was generated by passing a dry-air flow through water flask 6. The different humidity values inside of flask 10 were obtained by varying the proportion of dry air and humid air. The baths for the generation of humid air (flask 6) and the mixture/measurement chamber (flask 8) were maintained at 30 C. In order to obtain the desired humidity value, the required proportion of dry air and humid air was flushed through chamber number 8 for 10 min. Shortly thereafter, the vacuum pump was turned down and the humidity reached equilibrium in approximately 10 min. After measuring the humidity value with a TESTO (Lenzkirch, Germany) Model 635 hygrometer (flask 9), the air with known humidity was carried through a stainless steel tube with an outer diameter of 0.56 mm to the sampling chamber (flask 14). Standard gas generation The static generation of the standard mixture was carried out in a stirrer/heater with an aluminum block containing a 22-mL Pyrex vial (Supelco) (as illustrated in Figure 1). The vials were calibrated and sealed with aluminum caps containing Teflon-rubber septa. The standard gaseous mixture of BTEXS was obtained through the injection of the standard liquid inside the vial with a 5.00-mL Hamilton (Reno, NV) syringe. Then, the vial was left in the system (described in Figure 1) for 30 min at 30 C ± 0.5 C to generate the standard gas with a known relative humidity. A stirring bar of 14 5 mm at 1000 rpm was used to homogenize the standard gas generated. After 30 min of homogenization, stirring was stopped and the fiber was introduced into the vial for the sorption of the analytes. The BTEXS compounds were desorbed in the injector for 3 min at 220 C. The theoretical concentration was calculated in parts per million for volume (ppm v ) at 25 C and 1 atm using the equation: ppm v = (24.5 10 6 W) / (V M) Eq. 1 where W is the weight of the compound (g), V the volume of the sample in the vial (L), and M the molecular weight of the compound (g/mol). Results and Discussion Evaluation of the PDMS and PDMS DVB fibers in different humidity conditions The concentration of standard gas mixtures used to study the effect of humidity on analyte sorption by SPME fibers were Figure 1. System for dynamic generation of humidity: air dryer, 1; activated charcoal tube, 2; regulator, 3; dry air, 4; humid air, 5; ultrapure water flask, 6; bath at 30 C, 7; mixture and humidity measurement chamber, 8; TESTO hygrometer, 9; block of aluminum, 10; heater and stirrer, 11; SPME device, 12; chromatographic syringe, 13; sampling chamber, 14; and vacuum pump, 15. 123

1.43, 1.20, 1.00, 1.05, 1.07, and 1.05 ppm v for benzene, toluene, ethylbenzene, m- and p-xylene, styrene, and o-xylene, respectively, at 25 C and 1 atm. The effect of relative humidity on the sorption of BTEXS using PDMS and PDMS DVB is plotted in Figures 2 and 3, respectively. The results obtained with both fibers indicated that the sorption of each BTEXS compound remained nearly the same at different humidity values. Therefore, humidity appeared not to interfere with fiber sorption. The PDMS DVB fiber had a high BTEXS sorption capacity in relation to the PDMS fiber. This sensitivity can be observed through the area values of Figures 2 and 3. Therefore, the results obtained supported the choice of the PDMS DVB fiber for BTEXS sampling in ambient air. the adsorption extraction process, the isotherm is highly nonlinear for higher concentrations when the surface coverage is substantial (11). The repeatability experiment of the PDMS DVB fiber was carried out through the sampling and analysis of two ranges of different concentrations of the standard gas samples. The first range of concentration (data 1) was 0.07 ppm v for benzene; 0.06 ppm v for toluene; and 0.02 ppm v for ethylbenzene, m- and p- xylene, styrene, and o-xylene. In the second range (data 2), the values of concentration were 22.81, 19.34, 16.02, 16.78, 17.11, and 16.78 ppm v for benzene, toluene, ethylbenzene, m- and p- xylene, styrene, and o-xylene, respectively. These analyses were PDMS DVB equilibrium time The development of an SPME method requires the determination of the minimum time necessary for the analytes to reach equilibrium between a gaseous mixture and the PDMS DVB fiber. For this purpose, the PDMS DVB fiber was exposed to the gas standard mixtures in concentrations of 1.43, 1.20, 1.00, 1.05, 1.07, and 1.05 ppm v for benzene, toluene, ethylbenzene, m- and p-xylene, styrene, and o-xylene, respectively, at 25 C and 1 atm. The fiber was exposed for the following times: 0.25, 1, 5, 15, 30, and 60 min. These experiments were carried out with atmospheric air in a relative humidity of 46% ± 2%. The results are plotted in Figure 4. The equilibrium time was relatively fast (approximately 1 min for benzene and toluene and 5 min for ethylbenzene, m- and p-xylene, styrene, and o-xylene). Accordingly, the time of 5 min was used in the partial validation experiments. SPME methods The linearity study is summarized in Table I. Although the correlation coefficients were in a range of 0.992 and 0.999, the analysis of variance indicated that there was a linear relationship between concentration and area. The relationship between the regression average square (QM reg ) and the residue average square (QM r ) was larger than the tabulated F value for 1.4 df (F = 7.71). Concentrations of benzene, toluene, ethylbenzene, m- and p-xylene, styrene, and o-xylene over 45.62, 38.68, 32.04, 33.57, 33.57, and 34.22 ppm v, respectively, at 25 C and 1 atm did not present a linear relationship between the concentration and area. In Figure 2. Effect of relative humidity on the sorption of analytes in the PDMS fiber. Figure 3. Effect of relative humidity on the sorption of analytes in the PDMS DVB fiber. 124

carried out in the same laboratory by the same operator using the same apparatus over a relatively short time span (same day). The results (summarized in Table II) indicate that this standard gas generation device, combined with sampling by PDMS DVB fibers, is precise for equilibrium sampling conditions. The results of the method detection limit (MDL) are summarized in Table II. The MDL was determined through the sampling and analysis of seven gas standards (12). The statistical results obtained for this series of analyses were substituted into the equation: Figure 4. Determination of the gas/fiber equilibrium time with PDMS DVB. Table I. Linear Regression Analysis Parameters* of BTEXS in the SPME GC Method Range Analyte (ppm v ) Equation R QM reg /QM r Benzene 0.07 45.62 y = 1045x + 2037 0.997 809.20 Toluene 0.06 38.6 y = 4287x + 5949 0.999 2061.61 Ethylbenzene 0.02 32.04 y = 10432x + 14049 0.997 640.54 m- and p-xylene 0.02 33.57 y = 24610x + 39735 0.995 449.70 o-xylene 0.02 33.57 y = 23778x + 32339 0.997 611.34 Styrene 0.02 34.22 y = 19039x + 38924 0.992 241.71 * These experiments were carried out with atmospheric air in a relative humidity of 46% ± 2% at 30 C and a PDMS DVB fiber. R, correlation coefficient. Table II. Analysis Parameters* of Repeatability and MDL of BTEXS in the SPME GC Method MDL = t (N 1, 1 α = 0.99) S c Eq. 2 where t is the student value for N 1 degrees of freedom to 99% confidence and S c is the standard deviation of the seven analyses. Analysis of real samples The application of the SPME method was carried out in indoor air passive sampling at two fuel analysis laboratories (A and B) and in an office with wet latex paint. The SPME sampling was made in parallel with NIOSH method 1501 (13). Integrated samples were collected for a period of 180 min. The sampling time was long in order to verify the emission of new analytes in the indoor air and because the BTEXS concentration in air was very low for the NIOSH method. After the sampling period, the fiber and cartridge were refrigerated to 3 C for 1 h until analysis time (14). Then, the cartridges were desorbed using the parameters described in NIOSH method 1501 and the PDMS DVB fibers were desorbed in the injector for 3 min at 220 C. The results were determined from calibration curves. The analysis by the NIOSH method did not detect any compound of the BTEXS group in laboratories A and B. In the office, the NIOSH method and the SPME method had similar determinations for the compounds detected. The LOD of NIOSH method 1501 is 0.001 for 0.01 mg. The results (shown in Table III) indicate that SPME with a PDMS DVB fiber is more sensitive than NIOSH at trace quantities of analytes in the sampling conditions used. Repeatability (%RSD) MDL Analyte Data 1 Data 2 (ppb v ) Benzene 5.7 8.9 16.70 Toluene 7.0 4.7 20.07 Ethylbenzene 6.3 5.3 3.33 m- and p-xylene 5.9 4.1 3.45 o-xylene 5.9 3.8 3.03 Styrene 5.7 4.4 3.51 * These experiments were carried out with atmospheric air in a relative humidity of 46% ± 2% at 30 C and a PDMS DVB fiber. Conclusion This study proposes a method for the passive sampling of BTEXS in indoor air with an SPME device. The sorption study of the PDMS and PDMS DVB fibers in different relative air humidity conditions shows that the fibers do not suffer from interference in the large range of humidity of BTEXS sampling. The PDMS DVB fiber was selected because it presented higher sorption capacity. Analysis of real samples was made to compare the classic active-sampling and the SPME passive-sam- 125

Table III. Analysis of Real Samples* with the NIOSH Active-Sampling Method and SPME Passive-Sampling Method Using a PDMS DVB Fiber pling method. This comparison shows that the SPME method is economical and capable of achieving a very low LOD. Because of the advantages of the SPME passive-sampling method, this technique can be used for the quantitative analysis of BTEXS indoor exposition. Acknowledgments This study was supported by FAPEMIG (Fundação de Amparo a Pesquisa do Estado de Minas Gerais) and CNPq (Conselho Nacional de Pesquisa e Desenvolvimento) (Brasilia, Brazil). References Toluene Ethylbenzene m- and p-xylene o-xylene (ppm v ) (ppm v ) (ppm v ) (ppm v ) Sampling Place Day SPME NIOSH SPME NIOSH SPME NIOSH SPME NIOSH Laboratory A 1 2.74 n.d. 0.61 n.d. 0.21 n.d. 0.09 n.d. 2 1.77 n.d. 0.80 n.d. 0.28 n.d. 0.13 n.d. 3 3.06 n.d. 0.27 n.d. 0.43 n.d. 0.20 n.d. 4 1.41 n.d. 0.11 n.d. 0.17 n.d. 0.07 n.d. Laboratory B 1.19 n.d. 0.24 n.d. 0.35 n.d. 0.15 n.d. Office n.d. n.d. n.d. n.d. 13.93 10.39 15.42 16.91 * Temperature and relative humidity were 27 C and 57% ± 3%, respectively. n.d., not detected (below LOD). Sampling time was 8 h. Average flow of 130 ml/min and 180 min. 1. C. Chung, M.T. Morandi, T.H. Stock, and M. Afshar. Evaluation of a passive sampler for volatile organic compounds at ppb concentrations, varying temperatures, and humidities with 24-h exposures. 1. Description and characterization of exposure chamber system. Environ. Sci. Technol. 33: 3661 65 (1999). 2. C. Chung, M.T. Morandi, T.H. Stock, and M. Afshar. Evaluation of a passive sampler for volatile organic compounds at ppb concentrations, varying temperatures, and humidities with 24-h exposures. 2. Sampler performance. Environ. Sci. Technol. 33: 3666 71 (1999). 3. J. Begerow, E. Jermann, T. Keles, U. Ranft, and L. Dunemann. Passive sampling for volatile organic compounds (VOCs) in air at environmentally relevant concentration levels. Fresenius J. Anal. Chem. 351: 549 54 (1995). 4. R.H. Brown. Monitoring the ambient environment with diffusive samplers: theory and pratical considerations. J. Environ. Monit. 2: 1 9 (2000). 5. J. Pawlisyn. Solid-Phase Microextraction, 1st ed. Wiley-VCH, New York, NY, 1997. 6. P.A. Martos and J. Pawlisyn. Timeweighted average sampling whith solidphase microextraction device: implications for enhanced personal exposure monitoring to airbone pollutants. Anal. Chem. 71: 1513 20 (1999). 7. J. Koziel, M. Jia, A. Khaled, J. Noah, and J. Pawlisyn. Field air analysis with SPME device. Anal. Chim Acta. 400: 153 62 (1999). 8. P.A. Martos and J. Pawlisyn. Calibration of solid-phase microextraction for air analyses based on physical chemical properties of coating. Anal. Chem. 69: 206 15 (1999). 9. L. Müller, T. Górecki, and J. Pawlisyn. Optimization of the SPME device: design for field applications. Anal. Chem. 71: 1513 20 (1999). 10. G.O. Nelson. Controlled Test Atmospheres Principles and Techniques, 2nd ed. Ann Arbor Publisher, Inc., Ann Arbor, MI, 1972. 11. J. Pawliszyn. Theory of solid-phase microextraction. J. Chromatogr. Sci. 38: 270 78 (2000). 12. J.A. Glaser, D.L. Foerst, G.D. Mckee, S.A. Quave, and L.W. Budde. Trace analyses for wastewaters. Environ. Sci. Technol. 15(12): 1426 35. (1981). 13. Method 1501 Hydrocarbons Aromatics. NIOSH. In Manual of Analytical Methods NMAM, 4th ed. United States Department of Health Education and Welfare (NIOSH), Cincinnati, OH, 1994. 14. L. Müller, T. Górecki, and J. Pawliszyn. Optimization of the SPME device design for field applications. Fresenius J. Anal. Chem. 364: 610 16 (1999). Manuscript accepted January 8, 2002. 126