OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD

Similar documents
CALIBRATION CERTIFICATE FAILURE EXPLANATION FOR AEROTRAK PARTICLE COUNTERS

NANOSCAN SMPS SPECTROMETER COMPARED TO THE TSI SMPS SPECTROMETER

SIZE-SELECTED AEROSOL FILTER ANALYSIS USING ELECTROSTATIC CLASSIFICATION AND LIBS

MEASURING NANOPARTICLE EXPOSURE

MACROIMS MACROION MOBILITY SPECTROMETER MOLECULAR WEIGHT AND PARTICLE SIZE DETERMINATION OF POLYVINYLPYRROLIDONE (PVP)

CARBON NANOTUBE ANALYSIS USING THE TSI LIBS DESKTOP ANALYZER

TSI AEROTRAK REMOTE 4-20 ma SENSORS LIFE SCIENCE

ELECTROSTATIC CLASSIFIER MODEL 3082

Fundamentals of Particle Counting

SPECIFICATIONS PARTICLE SENSOR KS-28B Higashimotomachi, Kokubunji, Tokyo , Japan

Product data sheet Palas U-SMPS 1050 / 1100 / 1200

Estimation of Mass with the Model 3321 APS Spectrometer

Experiment C-10 Titration of a Strong Acid and a Strong Base

Heat Flow Calculations Made Using the Models 8386 and 8386A VELOCICALC Plus Air Velocity Meters Application Note ITI-124

Pre-Lab Exercise Full Name:

Spectrometer User s Guide

VIBRATING ORIFICE AEROSOL GENERATOR MODEL 3450

Fast, Effective XPS Point Analysis of Metal Components

Particle number concentration standards at NMIJ, including a new calibration service with an inkjet aerosol generator.

Photoluminescence Spectrometer (FLS980)

ncounter PlexSet Data Analysis Guidelines

Calibration checks of particle counter using primary and other techniques at the laboratory level

Determination of the Rydberg constant, Moseley s law, and screening constant (Item No.: P )

ENVR 116 Aerosol Technology Laboratory Session Fall 2005

Model 2300XP PSL & Process-Particle Wafer Deposition System

Waters GPC User Guide and Tutorial for Using the GPC in the Reynolds Research Group 2 nd Edition: April 2012

Connect the Vernier spectrometer to your lap top computer and power the spectrometer if necessary. Start LoggerPro on your computer.

and Environmental Science Centre

General Physics I Lab (PHYS-2011) Experiment MECH-2: Newton's Second Law

Work and Energy. This sum can be determined graphically as the area under the plot of force vs. distance. 1

A new versatile condensation particle counter for research and environmental monitoring

Newton's 2 nd Law. . Your end results should only be interms of m

Technical Paper 3 Verifying Particle Counter Calibration: Size Verification

Multi-Instrument Manager Tool for Data Acquisition and Merging of Optical and Electrical Mobility Size Distributions

L03 The Coefficient of Static Friction 1. Pre-Lab Exercises

ISO INTERNATIONAL STANDARD

COPLEY S C I E N T I F I C. A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for:

Kinetics of Crystal Violet Bleaching

SIEMENS DIMENSION VISTA 1 The NGAL Test Reagent Kit

The RX-6000 Sulfur in Oil Analyzer For the analysis of Sulfur in Oil and other elements from S to U

NATURAL ROCK SAMPLE. Performance Monitoring Accessory for the RAD7 User Manual

Jasco V-670 absorption spectrometer

Measuring the time constant for an RC-Circuit

Force vs time. IMPULSE AND MOMENTUM Pre Lab Exercise: Turn in with your lab report

Technical notes: using the optical/ellipsometric film thickness measurement setup.

Technical Note. How to make Concentration Measurements using NanoSight LM Series Instruments

Lab 1 Uniform Motion - Graphing and Analyzing Motion

Product data sheet Palas Nanoparticle measurement system DEMC 2000

SIEMENS DIMENSION VISTA 1 The NGAL Test Reagent Kit

ELEC 1908 The Electric Potential (V) March 28, 2013

Measuring ph with Smart Cuvettes

Traceability research activities in the field of airborne particle number/ mass concentration measurement Liu Junjie, Zhang Wenge, Song Xiaoping

Workshop: Build a Basic HEC-HMS Model from Scratch

Lab 1a Wind Tunnel Testing Principles & Drag Coefficients of Golf balls

Application of Micro-Flow Imaging (MFI TM ) to The Analysis of Particles in Parenteral Fluids. October 2006 Ottawa, Canada

Conductometric Titration & Gravimetric Determination of a Precipitate

Exercise 4-4. Titration of a Buffer Solution EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Buffer solutions

The solution for all of your

Lab: Newton s Second Law

Natural Rock Sample Manual

Size Determination of Gold Nanoparticles using Mie Theory and Extinction Spectra

What is the velocity profile downstream of the sudden expansion? What is the relationship between the velocity profile and the flow rate?

ab83360 Ammonia Assay Kit

ICP-OES DETERMINATION OF IRON. Introduction:

15 Temperature volume relationship in a gas (Charles law, Gay-Lussac s law)

Lab 3 Momentum Change and Impulse

Gamma-Spectrum Generator

Ocean Optics Red Tide UV-VIS Spectrometer (Order Code: SPRT-UV-VIS)

Energy loss of alpha particles - Prelab questions

FireFamilyPlus Version 5.0

ISO INTERNATIONAL STANDARD

Sample Alignment Part

Eurostat Business Cycle Clock (BCC): A user's guide

Information Dependent Acquisition (IDA) 1

Supplementary information

The NGAL Test Reagent Kit

Manual X-ray fluorescence Manual remote experiment Project e-xperimenteren+

The New Legend. Eppendorf Reference 2 Pipette

Collisions Impulse and Momentum

NMRPredict Functional Block Diagram

The ultimate in desktop particle characterization

Experiment P-5 Motion of a Cart on an Inclined Plane

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

protein interaction analysis bulletin 6300

Predictor Assay Setup Guide on the Tecan Safire 2 Microplate Reader

Sample Alignment (2D detector) Part

SOLIDWORKS Simulation Time Based Thermal Stress

Static and Kinetic Friction

The ultimate in desktop particle characterization

Intercomparison of Mobility Particle Size Spectrometers

Relativistic Electrons

Analyst Software. Automatic Optimization Tutorial

Intercomparison of Mobility Particle Size Spectrometers

Finite Element Modules for Enhancing Undergraduate Transport Courses: Application to Fuel Cell Fundamentals

This is the third of 3 parts detailing my experience of auto guiding for astrophotography.

Physics 1020 Experiment 6. Equilibrium of a Rigid Body

Hooke s Law. Equipment. Introduction and Theory

Manual Laser Doppler Anemometry Manual remote experiment Project e-xperimenteren+

Calibrating Thermo Exactive with the Direct Analysis in Real Time (DART) Ambient Ionization Source (Protocol Adapted from US FDA FCC)

Chem 1 Kinetics. Objectives. Concepts

Transcription:

OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD APPLICATION NOTE OPS-001 The Optical Particle Sizer (OPS) is calibrated for size with Polystyrene Latex (PSL) spheres (per ISO 12501-1/4) at TSI. This method of calibration is common for all optical particle sizing instruments. In order to calculate mass from this size information there are assumptions that must be made. Standard assumptions are that the particles are spheres and by supplying a density the mass of the particles can be determined. In reality, the particles may not be spheres and the light scatter of the particles can be significantly different than that of the calibration particles (white PSL spheres). Corrections for these differences may be incorporated into an effective density which is the material density multiplied by a correction factor for the aerosol being measured. The calibration consists of three main steps: 1. Create a mass reference 2. Calculate the effective density 3. Apply the effective density to the OPS Note that the OPS Pump Always On and Dead Time Correction options should be enabled. The former is to avoid contamination of the optical chamber by the test aerosols, and the latter is to minimize coincidence error. During the test, the total particle concentrations also need to be lower than the OPS upper concentration limit which is 3000 particles/cm 3. The following steps illustrate how to calibrate the OPS for mass measurement. Once calibrated, the OPS will measure mass with reasonable accuracy as long as the aerosols of interest are the same as the aerosol used for calibration, and the size distributions remain the same. If the aerosol types or size distributions are different, a new calibration may be required. Note: Mass is a post-processing calculation. Density information can be entered either before or after data has been taken and it will be reflected in the graphs, tables and statistics. Create a Mass Reference The reference mass can be created using the OPS internal filter or an external filter. The external filter method is preferred since the mass collected by the OPS internal 37 mm internal filter may be biased by transport losses inside the OPS. Aerosol Instrument Manager, TSI, and TSI logo are trademarks of TSI Incorporated.

External Filter (Preferred Method) The calibration steps are outlined below: 1. Setup an external filter measurement system. To avoid sampling bias, the flow rate and sample time of the external filter measurement system should be the same or very close to the OPS flow rate and sample time. 2. Weight the new filter with a microbalance. You may need to minimize the static charge effect by exposing the filter to a bipolar ion source before weighting it. 3. Clean the filter cartridge before placing the new filter in it. 4. Co-locate the two instruments as shown in figure 1. Aerosol Souce Optical Particle Sizer OPS To Flow Control and Vacuum Figure 1. Experimental setup for gravimetric calibration 5. To avoid test aerosol filling up the optical chamber, DO NOT hook up the test aerosol to the OPS before starting the OPS pump. 6. Start the OPS. If the Pump Always On is enabled, the pump should be running after warm up. 7. Introduce the aerosol of interest into the OPS and the external filter measurement system, and start the measurement. Since the amount of mass collected by the filter depending on the aerosol density, size distribution and concentration, you need to adjust the sample time appropriately to get reliable mass measurement. 8. Stop the measurement. Take the filter out and measure it with the microbalance. 9. Using the mass collected on the external filter, calculate the mass concentration by using the equation shown below: Mass Collected Mass concentration = Flow Rate Sample Time Internal Filter The OPS has a 37-mm filter downstream of the optical chamber that can be used for mass calibration. A new filter is required for every calibration. Note that the mass collected by this filter may be biased by the transport losses inside the OPS, so it is recommended to use external filter measurement system if possible. The calibration steps are outlined below: -2-

1. Weight the new filter with a microbalance. You may need to minimize the static charge effect by exposing the filter to a bipolar ion source before weighting it. 2. Clean the filter cartridge before placing the new filter in it. Put the assembly back to the OPS. Make sure the assembly is secured to avoid leaking. 3. To avoid test aerosol filling up the optical chamber, DO NOT hook up the test aerosol to the OPS before the pump starts. 4. Start the OPS. If the Pump Always On is enabled, the pump should be running after warm up. 5. Introduce the aerosol of interest into the OPS, and start the measurement. Since the amount of mass collected by the filter depending on the aerosol density, size distribution and concentration, you need to adjust the sample time appropriately to get reliable mass measurement. 6. Stop the measurement. Take the filter out and measure it with the microbalance. 7. Using the mass collected on the 37 mm filter, calculate the mass concentration by using the equation shown below: Mass Collected Mass concentration = Flow Rate Sample Time The flow rate and sample time are the OPS flow rate and the OPS sample time. Calculate the effective density 1. Set the OPS particle density to 1 g/cm3, and record the integrated mass concentration. The effective density can be calculated by: Effective density = mass concentration from the filter 3 1 g/cm OPS integrated mass concentration Apply the effective density to the OPS (on instrument) 1. Set the OPS particle density to the effective density calculated above. a. On the OPS select the Setup tab at the bottom of the screen. b. Select the Sampling icon. c. Select the User Calibration icon. d. Enter your effective density in the Density box as shown in Figure 2. -3-

Figure 2. User Calibration screen 2. The OPS will now use this effective density in the displayed mass results both in graphs and tables. Apply the effective density to the OPS (in software) 1. If data is being collected using Aerosol Instrument Manager software, the density information is entered using the OPS Properties Dialog Box. 2. When you start the Aerosol Instrument Manager software and select the OPS 3330 Data type, the OPS 3330 Properties window opens. Select the Data Setup tab and select the Set Up Channels button. 3. If you select the Particle Density check box, you should then enter your effective density in the Density box. 4. If density varies by particle size, it is possible to do a more advanced correction and supply different densities for each size channel. If you uncheck the box in Particle Density portion of the Channel Setup window, you can enter different densities for each of the Channels. Figure 3 shows an example of Channel Setup using different densities. -4-

Figure 3. Channel Setup Screen with different densities 5. The Aerosol Instrument Manager software will now reflect your effective density value(s) in the mass results both in graphs, tables and statistics. -5-

TSI Incorporated Visit our website www.tsi.com for more information. USA Tel: +1 800 874 2811 UK Tel: +44 149 4 459200 France Tel: +33 4 91 11 87 64 Germany Tel: +49 241 523030 India Tel: +91 80 67877200 China Tel: +86 10 8251 6588 Singapore Tel: +65 6595 6388 OPS-001 Rev. B (A4) 2012 TSI Incorporated Printed in U.S.A.