Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Similar documents
Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Fergus Keenan Shona McSheehy Julian Wills

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Trace Analyses in Metal Matrices Using the ELAN DRC II

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Accurate measurement of elemental impurities in metals and metal alloys using the Thermo Scientific icap TQ ICP-MS

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

ICP/MS Multi-Element Standards

Samples compliant with the WATER

IMPROVING DATA QUALITY WITH ADVANCED INSTRUMENTATION UTILIZING MULTIPLE INTERFERENCE REDUCTION TECHNOLOGIES

Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Agilent s Solutions for Nanoparticles Analysis ICP-MS, ICP-MS/MS, software. Alain Desprez Agilent Technologies

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Secondary Ion Mass Spectrometry (SIMS) Thomas Sky

Handling Interferences in the Modern Laboratory with Advanced Triple Quadrupole ICP-MS Technology

Secondary ion mass spectrometry (SIMS)

Ar Ar + e - INDUCTIVELY+COUPLED+ PLASMA+SPECTROMETRY+ What+is+Plasma?+ FuncDon+of+Plasma+

Using cold vapor generation atomic absorption to determine mercury impurities in pharmaceutical products

Sample Analysis Design PART III

Sample Analysis Design PART II

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

ICP-MS. plasma 17OO. High Resolution Multi-Collector ICP-MS.

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Validating performance of an Agilent 7800 ICP MS for USP <232>/<233> & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Analytical Methods. Electronic Supplementary Information File

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Partial Energy Level Diagrams

á233ñ ELEMENTAL IMPURITIES PROCEDURES

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

Elemental Scientific. Spike Recovery. Elemental Scientific

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Chapter 3: Stoichiometry

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Chem Exam 1. September 26, Dr. Susan E. Bates. Name 9:00 OR 10:00

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

The New Agilent 7700 Series ICP-MS

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Sample Analysis Design Internal Standard

Lecture 7: Atomic Spectroscopy

ELEMENTAL ANALYSIS OF GLASS EXAMINATIONS (PART 1) Module 4

Analytical Approaches to Arsenic

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Analysis of noble metals at low levels in geological reference materials and ores

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Opening the analytical black box: Insights into interferences, corrections, and data quality

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:

Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Debjani Banerjee Department of Chemical Engineering IIT Kanpur

Nucleus. Electron Cloud

OES - Optical Emission Spectrometer 2000

Determination of Trace Metals in Human Urine Using the NexION 300 ICP-MS

ICP-Mass Spectrometer

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Simplified workflows for the Thermo Scientific icap TQ ICP-MS

CHE 105 Exam 1 Spring 2016

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

Department of Chemistry, The University of Hong Kong, Pokfulam Road, Hong Kong.

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

7) Applications of Nuclear Radiation in Science and Technique (1) Analytical applications (Radiometric titration)

M11/4/CHEMI/SPM/ENG/TZ2/XX CHEMISTRY STANDARD LEVEL PAPER 1. Monday 9 May 2011 (afternoon) 45 minutes INSTRUCTIONS TO CANDIDATES

CHEM 171 EXAMINATION 1. October 9, Dr. Kimberly M. Broekemeier. NAME: Key

Agilent 8800 Triple Quadrupole ICP-MS: Understanding oxygen reaction mode in ICP-MS/MS

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

All chemical bonding is based on the following relationships of electrostatics: 2. Each period on the periodic table

Using FIMS to Determine Mercury Content in Sewage Sludge, Sediment and Soil Samples

(please print) (1) (18) H IIA IIIA IVA VA VIA VIIA He (2) (13) (14) (15) (16) (17)

ENVG FALL ICP-MS (Inductively Coupled Plasma Mass Spectrometry) Analytical Techniques

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

Sources of Errors in Trace Element and Speciation Analysis

CHEM 172 EXAMINATION 1. January 15, 2009

Reporting Category 1: Matter and Energy

Agilent Technologies at TIAFT 2013

Reporting Category 1: Matter and Energy

Stability Nuclear & Electronic (then ion formation/covalent bonding)

Transcription:

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Why ICP Mass Spectrometry? Ultra-trace multi-element analytical technique Ultra-Trace atomic absorption techniques are single element Applying Option 1 (10-g) dosing at the control threshold (30% of PDE) too low for most ICP optical emission methods for Class 1 and 2A EIs Eli Lilly and Company 2

Fundamental Instrument Configurations Commercial ICP-MS Instrumentation configurations Quadrupole based (unit resolution) Time of Flight (unit resolution) Double focusing magnetic sector based (high resolution) The majority of pharmaceutical analyses will be conducted using quadrupole based ICP mass spectrometers..so we will focus on that configuration To illustrate application of ICP mass spectrometry, experimental results generated on two test materials: parenteral sucrose and sodium chloride Eli Lilly and Company 3

Experimental details for analysis of sucrose and NaCl Dose, g 10 Specimen weight, g 0.2 Final Volume, ml 100 Class Element Parenteral PDE, µg/day Permitted Conc., µg/g Control Threshold, µg/g Control Threshold Spike Conc., Experimental Spike Conc., 1 As 15 1.5 0.45 0.9 0.12 1 Cd 2 0.2 0.06 0.12 0.12 2A Co 5 0.5 0.15 0.3 0.12 3 Cu 300 30 9 18 2.4 1 Hg 3 0.3 0.09 0.18 0.12 3 Li 250 25 7.5 15 2.4 2A Ni 20 2 0.6 1.2 0.6 1 Pb 5 0.5 0.15 0.3 0.12 3 Sb 90 9 2.7 5.4 2.4 2A V 10 1 0.3 0.6 0.6 Eli Lilly and Company 4

Standard curve specifics-used for both test materials Final Volume, ml 0.0 0.05 0.10 0.25 0.50 1.00 0.06 Element Stock Standard Conc., µg/ml Blank, Std 1, Std 2, Std 3, Std 4, Std 5, Spiking Conc., Final Volume, 2% HNO 3 As 0.1 0 0.1 0.2 0.5 1 2 0.12 50 Cd 0.1 0 0.1 0.2 0.5 1 2 0.12 50 Co 0.1 0 0.1 0.2 0.5 1 2 0.12 50 Cu 2 0 2 4 10 20 40 2.4 50 Hg 0.1 0 0.1 0.2 0.5 1 2 0.12 50 Li 2 0 2 4 10 20 40 2.4 50 Ni 0.5 0 0.5 1 2.5 5 10 0.6 50 Pb 0.1 0 0.1 0.2 0.5 1 2 0.12 50 Sb 2 0 2 4 10 20 40 2.4 50 V 0.5 0 0.5 1 2.5 5 10 0.6 50 All test solutions contain 10ppb internal standard and 100ppb Au. Eli Lilly and Company 5

Spike Recovery (Accuracy) Results for Sucrose and NaCl (both prepared at 0.2g-% in 2% Nitric Acid) 7Li, 45Sc IS 51V, 59Co, 60Ni, 65Cu, 75As, 89Y IS 114Cd, 115In IS 123Sb, 159Tb IS 175Lu IS Description Sucrose Unspiked Average (n=3) 0.044 98 0.146 0.005 0.018 0.053 0.003 98 0.001 99 0.107 99 97 0.007 0.004 97 Sucrose Spiked Average (n=3) 2.381 99 0.692 0.126 0.629 2.507 0.135 97 0.124 99 2.521 98 97 0.131 0.127 98 Spiking Concentration 2.40 0.12 0.12 0.60 2.40 0.12 0.12 2.40 0.12 0.12 Sucrose Recovery, % 97 91 101 102 102 110 103 101 103 102 Description NaCl Unspiked Average (n=3) NaCl Spiked Average (n=3) 7Li, 45Sc IS 51V, 59Co, 60Ni, 65Cu, 75As, 89Y IS 114Cd, 115In IS 123Sb, 159Tb IS 175Lu IS 202Hg, 202Hg, 208Pb, 208Pb, 0.105 89 2.262 0.049 0.062 0.051 1.832 80 0.001 70 0.103 69 66 0.012 0.003 47 2.309 94 3.149 0.203 0.770 2.601 2.125 81 0.104 69 2.251 66 63 0.107 0.096 45 Spiking Concentration 2.40 0.12 0.12 0.60 2.40 0.12 0.12 2.40 0.120 0.120 NaCl %Recovery 92 148 128 118 106 244 86 90 79 77 209Bi IS 209Bi IS Only 115In used for internal standard. Note even IS response across the mass range for sucrose. Note attenuation of internal standards as mass increases in NaCl test solutions. Note concentrations of V (1.13 µg/g) and As (0.92 µg/g) in NaCl. Why the dramatic differences between these two materials? Eli Lilly and Company 6

Quadrupole ICP Mass Spectrometer Instrumentation: Plasma/spectrometer interface, mass interferences collision-reaction cell http://www.jsg.utexas.edu/icp-ms/files/schematic.jpg Eli Lilly and Company 7

Polyatomic interferences at unit resolution specificity 51V (99% abundant) Interfering species is 35Cl16O 75As (100% abundant) Interfering species is 40Ar35Cl By what means is this corrected? Inter-element correction Monitor count-rate at 37Cl16O (53Cr) Monitor count-rate at 40Ar37Cl (77Se) Collision cell/reaction cell Kinetic Energy Discrimination (non-reactive gases-collision cell) Dynamic Reaction Cell (reactive gases e.g. ammonium, oxygen) Eli Lilly and Company 8

Repeat analysis of sucrose and NaCl in KED mode (He).. Dramatically improve specificity of isotopes with polyatomic isobaric interferences Attenuate count-rate for analytical isotopes as well For some systems (especially older collision cell equipped systems) cannot attenuate both 35ClO16 and 40ArCl35 equally well with a single set of parameters.so in some cases, they have to be analyzed separately Eli Lilly and Company 9

Slopes and background equivalent concentrations for standard curves in standard and KED modes. Element STD Mode Slope (CPS/) KED Mode Slope (CPS/) STD Mode BEC, KED Mode BEC, Cd 104991 45105 0.00049 0.00052 Pb 338524 290879 0.00424 0.00440 As 28870 3368 0.00970 0.00330 Hg 47252 42875 0.00143 0.00215 Co 243487 63625 0.00139 0.00028 V 233273 20815 0.01435 0.00059 Ni 52487 17862 0.00781 0.00690 Li 141966 219 0.61960 0.59361 Sb 68504 19923 0.00307 0.00335 Cu 46749 20362 0.01940 0.02276 Eli Lilly and Company 10

Spike Recovery (Accuracy) Results for Sucrose and NaCl acquired in KED mode.. Description 7Li, 45Sc IS 51V, 59Co, 60Ni, 65Cu, 75As, 89Y IS 114Cd, 115In IS 123SbIS 159Tb IS 175Lu IS Average Sucrose Unspiked (n=3) 0.198 101 0.011 0.003 0.015 0.113-0.004 100 0.003 100 0.048 100 100 0.005 0.006 101 Average Sucrose Spiked (n=3) 2.409 102 0.601 0.121 0.609 2.502 0.120 100 0.122 100 2.447 99 100 0.123 0.126 101 Spike Concentration 2.40 0.60 0.12 0.60 2.40 0.12 0.12 2.40 0.12 0.12 %Recovery 92 98 98 99 100 104 99 100 98 100 202Hg, 208Pb, 209Bi IS Description Average NaCl Unspiked (n=3) Average NaCl Spiked (n=3) 7Li, 45Sc IS 51V, 59Co, 60Ni, 65Cu, 75As, 89Y IS 114Cd, 115In IS 123SbIS 159Tb IS 175Lu IS 202Hg, 208Pb, 0.291 142 0.033 0.004 0.039 0.065-0.001 111 0.002 93 0.029 88 85 0.006 0.005 59 2.906 145 0.933 0.155 0.737 2.633 0.143 111 0.104 93 2.326 87 85 0.096 0.094 58 Spike Concentration 2.40 0.60 0.12 0.60 2.40 0.12 0.12 2.40 0.12 0.12 %Recovery 109 150 126 116 107 119 85 96 75 74 Why does the internal standard response trend downward with increased analyte mass in NaCl? 209Bi IS Eli Lilly and Company 11

Impact of Internal Standard Selection on Spike Recovery (Accuracy) Only 115In IS Corrected: Description 7Li, 45Sc IS 51V, 59Co, 60Ni, 65Cu, 75As, 89Y IS 114Cd, 115In IS 123SbIS 159Tb IS 175Lu IS Average NaCl Unspiked (n=3) 0.291 142 0.033 0.004 0.039 0.065-0.001 111 0.002 93 0.029 88 85 0.006 0.005 59 Average NaCl Spiked (n=3) Spike Concentration 2.40 0.60 0.12 0.60 2.40 0.12 0.12 2.40 0.12 0.12 %Recovery 109 150 126 116 107 119 85 96 75 74 202Hg, 208Pb, 2.906 145 0.933 0.155 0.737 2.633 0.143 111 0.104 93 2.326 87 85 0.096 0.094 58 209Bi IS Corrected by Interpolation between ISs by mass: Description 7Li, 45Sc IS 51V, 59Co, 60Ni, 65Cu, 75As, 89Y IS 114Cd, 115In IS 123Sb, 159Tb IS 175Lu IS Average NaCl Unspiked (n=3) -0.017 142 0.017 0.002 0.023 0.034-0.002 111 0.002 93 0.031 88 85 0.010 0.010 59 Average NaCl Spiked (n=3) 1.683 145 0.624 0.109 0.517 1.903 0.108 111 0.103 93 2.351 87 85 0.143 0.151 58 Spike Concentration 2.40 0.60 0.12 0.60 2.40 0.12 0.12 2.40 0.12 0.12 % Recovery 70 101 89 82 79 90 84 98 111 118 202Hg, 208Pb, 209Bi IS Eli Lilly and Company 12

Element Characteristics Element Ionization Potential (ev) Isotopes (amu) Abundance 6 7.52 Li 5.39 7 92.48 Lu 5.43 In 5.79 Tb 5.85 159 100 204 1.37 Pb 6.11 208 52.38 Y 6.38 89 100 Sc 6.54 45 100 V 6.74 Bi 7.29 209 100 175 113 206 50 97.4 4.16 25.15 0.24 176 115 207 51 2.6 95.84 21.11 99.76 Eli Lilly and Company 13

Element Characteristics, Cont d Element Ionization Potential (ev) Isotopes (amu) Abundance 58 67.76 60 26.16 Ni 7.63 61 1.25 62 3.66 64 1.16 Cu 7.73 63 69.09 65 30.91 Sb 7.85 121 57.25 123 42.75 Co 7.86 59 100 106 1.22 108 0.89 110 12.43 Cd 8.99 111 12.86 112 23.79 113 12.34 114 28.81 116 7.66 As 9.81 75 100 196 0.146 198 10.02 199 16.84 Hg 10.44 200 23.13 201 13.22 202 29.8 204 6.85 Eli Lilly and Company 14

Impact of Internal Standard Selection on Spike Recovery (Accuracy), cont d No IS Correction: Description 7Li, 45Sc IS 51V, 59Co, 60Ni, 65Cu, 75As, 89Y IS 114Cd, 115In IS 123Sb, 159Tb IS 175Lu IS Average NaCl Unspiked (n=3) 0.108 142-0.001-0.002 0.005-0.065-0.007 111-0.004 93-0.098 88 85-0.001-0.002 59 Average NaCl Spiked (n=3) 2.585 145 0.851 0.141 0.665 2.365 0.128 111 0.092 93 2.075 87 85 0.084 0.082 58 Spike Concentration 2.4 0.6 0.12 0.6 2.4 0.12 0.12 2.4 0.12 0.12 %Recovery 103 142 117 110 99 107 77 86 70 68 202Hg, 208Pb, 209Bi IS Selected IS Correction: Description 7Li, 45Sc IS 51V, 59Co, 60Ni, 65Cu, 75As, 89Y IS 114Cd, 115In IS 123Sb, 159Tb IS 175Lu IS Average NaCl Unspiked (n=3) 0.291 142 0.015 0.004 0.033 0.065-0.001 111 0.003 93 0.038 88 85 0.010 0.010 59 Average NaCl Spiked (n=3) 2.906 145 0.604 0.132 0.623 2.633 0.120 111 0.110 93 2.467 87 85 0.143 0.151 58 Spike Concentration 2.4 0.6 0.12 0.6 2.4 0.12 0.12 2.4 0.12 0.12 %Recovery 109 98 107 98 107 100 90 101 111 118 202Hg, 208Pb, 209Bi IS Eli Lilly and Company 15

Summary Unit resolution quadrupole based ICP mass spectrometers have interfering species generated within the mass spectrometer from constituent and atmospheric elements. Charging a collision/reaction cell with either inert (KED) or reactive (DRC) gases will attenuate the isobaric polyatomic species. It also will attenuate analyte response, severely for light elements such as lithium. Because the physical and chemical attributes of the interfering species for 51V and 75As are different, finding compromise analytical parameters to include both in a single analysis can be challenging using some collision cell equipped ICP mass spectrometers. The introduction of constituent ions (Na, Ca, K, Si, Ti, Mg, Zn) will cause mass biased results (attenuation as masses increase) due to changes in the electrostatic characteristics of the ion beam. Selection of multiple internal standards across the analytical mass range is highly recommended. Matrix matching of working standards and/or further dilution of test solution along with the use of a collision cell and multiple internal standards may be required to develop a suitable analytical method applicable to all EI s within a single analytical run. Eli Lilly and Company 16