Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Similar documents
Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Fast methods for the determination of ibuprofen in drug products

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Enhancement of Linearity and Response in Charged Aerosol Detection

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Determination of trace anions in concentrated hydrofluoric acid

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

Advances in Sample Preparation for Accelerated Solvent Extraction

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

APPLICATION NOTE Goal. Introduction. Keywords

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection

Determination of urea in ultrapure water by IC-MS/MS

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

ProPac WCX-10 Columns

Expanded capabilities in ammonium and amine detection

Dionex IonPac AS28-Fast-4µm column

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Discover what you re missing

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

Monitoring Protein PEGylation with Ion Exchange Chromatography

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

Acclaim AmG C18 Column

columns IonPac SCS 1 Silica Cation Separator

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

Tips & Tricks GPC/SEC: Quantify and Get More Than Molar Mass Averages

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Determination of Carbonyl Compounds in Workplace Air

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

[ Care and Use Manual ]

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

HILIC Method Development in a Few Simple Steps

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

PDG.pdf G-20 CHROMATOGRAPHY 3 4 INTRODUCTION

( )( ) Selectivity Choices in Reversed-Phase Fast LC. Introduction. R s = 1 a 1 k 4 a 1 + k

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

High Performance Liquid Chromatography

High Performance Liquid Chromatography

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

Determination of Aniline and Nitroanilines in Environmental and Drinking Waters by On-Line SPE

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Determination of Trace Cations in Power Plant Waters Containing Morpholine

for Acclaim Mixed-Mode HILIC-1 Column

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Thermo Scientific. Acclaim Trinity P2. Column Product Manual. P/N: October Part of Thermo Fisher Scientific

Easy Method Transfer from HPLC to RSLC with the Dionex Method Speed-Up Calculator

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Simultaneous Determination of Aromatic Amines and Pyridines in Soil

Automated and accurate component detection using reference mass spectra

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

of mass spectrometry

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Transcription:

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection Mark Tracy, Xiaodong Liu, and Ian Acworth, Thermo Fisher Scientific, Sunnyvale, CA, USA Application Note 20984 Key Words Acclaim SEC columns, size-exclusion chromatography, Corona Veo charged aerosol detector, polyacrylic acid, boiler water, power generation Abstract Polyacrylic acid (PAA) is widely used as an anti-scaling additive for boiler water in conventional steam generators and, more recently, in nuclear steam generators. This work demonstrates the application of Thermo Scientific Acclaim SEC columns and the Thermo Scientific Dionex Corona Veo charged aerosol detector to the measurement of PAA in boiler water. No sample extraction or pre-concentration is needed to attain detection limits below 10 µg/l. The linear range exceeds 200 µg/l, with a useful dynamic range exceeding 400 µg/l, and recoveries are 60 120% over the range of 3 400 µg/l in a boiler water matrix. Introduction Boiler scale is formed when impurities in water precipitate onto heat transfer surfaces or when suspended matter in the water settles out forming hard, adherent coatings. If unchecked, scaling causes progressive lowering of the boiler efficiency due to heat retardation as it acts as an insulator. Boiler deposits can also cause plugging leading to decreased operating efficiency, boiler damage, unscheduled boiler outages, and the need for costly cleaning procedures. Such scale-related issues are major problems for the generation of electrical power whether using conventional steam generators or nuclear power. Polyacrylic acid, a widely used anti-scaling additive, has both scale inhibition and dispersion properties. It is typically used at concentrations of 1 1000 µg/l, depending on the load of metals in the boiler water. The amount of PAA remaining needs to be measured for both process control and for wastewater discharge. Since PAA is the only macromolecule normally present in boiler water, it can easily be resolved from other components using size-exclusion chromatography (SEC). Size-exclusion chromatography is the dominant mode of separation for polymers. It is based on the discrimination of individual sample components in the pore volume of the packing material. Large sample molecules cannot or can only partially penetrate the pores, whereas smaller molecules can access most or all pores. Thus, large molecules elute first and smaller molecules elute later. Acclaim SEC columns use hydrophilic polymer particles with a well controlled pore-size distribution that gives a quantitative relation between molecular size and retention time. These columns are designed for the analysis of water-soluble polymers, including PAA. Unfortunately, PAA only has a weak UV chromophore so UV absorbance detection is limited to more concentrated samples. Charged aerosol detection (CAD), a masssensitive approach capable of measuring any non-volatile and many semi-volatile species, shows excellent sensitivity and a wide dynamic range and does not require a chromophore for analyte detection. The Corona Veo charged aerosol detector is an excellent choice of detector for polymers, especially ones with weak UV chromophores. Detection begins with the nebulization of the column effluent, followed by the generation of a dry aerosol and the subsequent adsorption of ionized nitrogen onto the particle surface. In a final step, this charge is then

2 measured by an electrometer. With full coverage of the aerosol particles, this charge is proportional to the surface area and is thus as a first approximation to its mass. Since this process does not depend on the chemical or spectroscopic properties of the particles, it is nearly universal for nonvolatile analytes. Compared to light-scattering detection, electrical detection provides superior detection limits. Here we demonstrate direct analysis for polyacrylic acid using SEC separation and charged aerosol detection with detection limits below 10 µg/l, recoveries of 60 110%, and a useful dynamic range greater than 400 µg/l. Experimental Details Consumables Thermo Scientific Chromacol Instrument Select Kit, clear standard opening screw thread vial (8-425), black closure with white silicone/red PTFE septum Acetonitrile, Fisher Chemical Optima LC-MS grade Deionized water Polyacrylic acid ethanolamine salt 10% PAA solution in purified water from a reputable supplier A specimen of boiler water matrix without PAA was kindly provided by a customer. The specimen was preserved by the addition of acetonitrile (10% v/v). Part Number TSL A955 Sample Preparation PAA stock standard, 1000 µg/l, was prepared by serial dilution in water of the 10% PAA solution. Working standards at 200, 100, 50, and 25 µg/l were prepared by diluting the stock standard with water directly into the autosampler vials. The matrix recovery solutions were prepared by diluting the stock standard in preserved boiler water matrix to 400 µg/l, then serially diluting to 80, 16, and 3.2 µg/l, again directly into the autosampler vials. Separation Conditions Instrumentation: Thermo Scientific Dionex UltiMate 3000 Liquid Chromatography System consisting of Part Number DGP-3600RS Pump 50466 TCC-3000RS Column Oven 57300 WPS-3000RS Sampler 58420 DAD-3000RS Diode-array Detector 5082.0020 Corona Veo RS Charged Aerosol Detector 5081.9920 Columns: Acclaim SEC-300 5 µm, 4.6 300 mm 079723 Mobile phase A: Mobile phase B: Isocratic 90% A, Flow rate: Acclaim SEC-1000 7 µm, 4.6 300 mm 079724 Water Acetonitrile 10% B Column temperature: 30 C 0.35 ml/min Injection details: Variable, 2 100 µl Diode array detection: Charged aerosol detection: UV 200 nm, data rate 1 Hz, filter 2 s Evaporator temperature 55 C, gas pressure 60 psi, data rate 2 Hz, filter 5 s, power function 0.80 Data Processing Software: Thermo Scientific Dionex Chromeleon 6.8 SR13 Chromatography System. The CAD data was smoothed by a two-pass 25-point Savitsky-Golay filter.

Results The chromatographic method used a 4.6 300 mm Acclaim SEC-1000 column with water / acetonitrile (90:10 v/v) for the mobile phase. This permitted a direct injection of a relatively large volume (35 µl) of aqueous sample without any pre-concentration. Detection with the Corona Veo charged aerosol detector gave good linearity and satisfactory detection limits. 3 1.4 1.2 1.0 0.8 0.6 PAA 0.10 Polyacrylic acid 0.4 0.2 200 μg/l 100 μg/l 50 μg/l Area (*min) 25 μg/l 0.0 0 100 200 250 Concentration (μg/l) Figure 1: Calibration data for PAA using Acclaim SEC-1000 column and 35 µl injection of standards in water 0.35 0.30 PAA 0.25 0.20 100 μg/l 0.06 Polyacrylic acid 0.15 0.10 0.05 60 μg/l 30 μg/l 20 μg/l 10 μg/l Area (*min) 5 μg/l 0 1 2 3 ng on column Figure 2: Detection limit data for PAA using Acclaim SEC-1000 column and injection volumes as indicated of a 25 µg/l standard in water

4 1.40 PAA 1.20 1.00 0.80 0.60 400 μg/l 80 μg/l 0.40 0.20 16 μg/l 3.2 μg/l matrix Figure 3: Recovery data for PAA using an Acclaim SEC-1000 column and injections of 35 µl of matrix fortified as indicated Figures 1, 2, and 3 show typical calibration, detection limit, and recovery chromatograms using the Acclaim SEC-1000 column. Typical calibration data are presented in Table 1. Calibration was calculated using a quadratic fit forced through zero, with 1/X weighting. Calibration standards in water were 25 400 µg/l. The fit quality was good at 4%RSD for a 35 µl injection. Carryover was not observed in a blank injection following an injection of 400 µg/l PAA. Detection limit data are presented in Figure 2. The limit of detection (LOD) was found to be 125 pg on column (5 µl 25 µg/l), with a signal-to-noise ratio of 3. Data showed an adequate fit with 6% RSD. This data also showed that large injections up to 30 µl caused no peak broadening, and up to 100 µl caused only minor broadening. Data for recoveries are presented in Figure 3 and Table 2, and were found to be 86 109%. The Acclaim SEC-1000 column showed better resolution between the PAA and the interfering peaks in the matrix and provided satisfactory recoveries. Compound Calibration Range (µg/l) Coefficient of Determination Mean Relative Error (%) PAA on Acclaim SEC-300 column PAA on Acclaim SEC-1000 column 25 400 0.996 3.3 25 400 0.999 3.9 Table 1: Accuracy (Acceptance criteria: the back calculated concentrations should fall within ±15% of their nominal values at all levels except for the LLOQ where ±20% is acceptable. No two adjacent levels should be outside their criteria.) Acclaim SEC-300 Column Acclaim SEC-1000 Column Spiking Level µg/l Found µg/l % Recovery Found µg/l % Recovery 400 330 83 358 90 80 74 93 67 84 16 12 75 14 89 3.2 1.9 59 3.5 109 Blank 0.1 n/a 0.0 n/a Table 2: Recovery (Calculate the recovery acceptance criteria: method dependant, target >80% and at least as good as the competition)

Standard SEC methods for measuring molecular weight employ a buffered mobile phase, which is used to suppress electrostatic effects between an ionic polymer and the stationary phase. In this case, the buffer was omitted, which shifted the retention time near to the exclusion volume and sharpened the peak, thereby improving resolution between PAA and interferences. This also reduced baseline noise in the detector. The Corona Veo charged aerosol detector gave better signal-to-noise performance with 10% acetonitrile added to the mobile phase, which had no significant effect on the chromatographic performance. The Corona Veo charged aerosol detector measures the electrical current carried by charge-bearing aerosol particles. The charge on a particle is limited by its surface area, so to relate the charge on a particle to its mass, a linearizing function is applied. After calibration, the detector signal is proportional to the mass of analyte delivered to the detector. In this instance the optimal linearity was obtained with a power function value of 0.80. Application Note 20984 Due to the universal nature of charged aerosol detection and the sensitivity of the Corona Veo, care must be taken to avoid column bleed and contamination of the mobile phase or LC system. The Acclaim SEC columns have extremely low-bleed performance when properly equilibrated and conditioned. For best performance, acetonitrile should be LC/MS grade or pesticide-residue grade and water must have 18.2 MΩ-cm conductivity and 5 ppb total organic carbon. To minimize down-time, an HPLC dedicated for use with volatile mobile phases should be used. Washed glass sample vials are preferable to polypropylene vials as the latter were found to reduce recovery of PAA at low ppb levels. Conclusion The Acclaim SEC columns provide the resolution and low background needed for successful trace analysis of polyacrylic acid. The Corona Veo charged aerosol detector excellent sensitivity for directly measuring polyacrylic acid to < 10 µg/l in boiler water matrix. thermofisher.com/columns 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. USA and Canada +1 800 332 3331 Australia 1300 735 292 (free call domestic) China 800 810 5118 (free call domestic) 400 650 5118 France +33 (0)1 60 92 48 34 Germany +49 (0) 2423 9431 20 or 21 India +91 22 6742 9494 +91 27 1766 2352 Japan 0120 753 670 (free call domestic) 0120 753 671 fax United Kingdom +44 (0) 1928 534 110 New Zealand 0800 933 966 (free call domestic) Singapore +65 6289 1190 All Other Enquiries +44 (0) 1928 534 050 Technical Support For advice and support, please visit our website: www.thermoscientific.com/chromexpert AN20984-EN 0616S