Determination of formic and acetic acids in petroleum products by ion chromatography

Similar documents
Determination of trace anions in concentrated hydrofluoric acid

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Dionex IonPac AS28-Fast-4µm column

Determination of urea in ultrapure water by IC-MS/MS

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Expanded capabilities in ammonium and amine detection

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Inline sample preparation for the determination of anions in sodium hydroxide

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Determination of silicate in high-purity water using ion chromatography and online sample preparation

Thermo Scientific. Anion-Exchange Column

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Determination of Adsorbable Organic Halogen in Wastewater

Accurate and Reproducible Determination of Organic Halogens Using Combustion Ion Chromatography

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

columns IonPac AS17-C Anion Exchange Column

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

The determination of trace anions in concentrated phosphoric acid

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Automated and accurate component detection using reference mass spectra

Determination of Trace Anions in Organic Solvents

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Monitoring Protein PEGylation with Ion Exchange Chromatography

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Fast methods for the determination of ibuprofen in drug products

The Determination of Trace Anions in Concentrated Phosphoric Acid

Determination of trace anions in concentrated glycolic acid

Determination of trace anions in organic solvents using matrix elimination and preconcentration

Determination of Methylamine in Drug Products

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Determination of trace anions in high-nitrate matrices by ion chromatography

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

Thermo Scientific Dionex IonPac AS26 Anion-Exchange Column

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

Utility of the Charge Detector in Ion Chromatography Applications

Determination of Silicate in High Purity Water Using Ion Chromatography and AutoPrep

columns IonPac AS18 Anion-Exchange Columns

Determination of chlorine, bromine, and sulfur in polyethylene materials using combustion IC with a carbonate/bicarbonate eluent

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

Application of Eluent Generation for Trace Anion Analysis of Borated Waters

columns IonPac SCS 1 Silica Cation Separator

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Determination of adsorbable organic halogen in wastewater using a combustion ion chromatography system

Thermo Scientific Dionex IonPac AS11-HC-4µm Anion-Exchange Column

anion-exchange column

ProPac WCX-10 Columns

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Determination of Organic Acids and Inorganic Anions in Lithium-Containing Boric Acid-Treated Nuclear Power Plant Waters

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

columns IonPac AS22 Anion-Exchange Column

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Thermo Scientific Dionex IonPac AS24A Anion-Exchange Column

Determination of trace anions in basic solutions by single pass AutoNeutralization and ion chromatography

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Determination of Trace Anions in Concentrated Glycolic Acid

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

Determination of Trace Anions in Concentrated Hydrofluoric Acid

Chemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Quantitative determination of ethephon in soluble concentration (SL) by Ion chromatography

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Determination of Monochloroacetic Acid in Carbocisteine

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Transcription:

CUSTOMER APPLICATION NOTE 72204 Determination of formic and acetic acids in petroleum products by ion chromatography Authors Joachim Trick 1, Cornelia Wanner 1, Cees Bruggink 2, Detlef Jensen 3 ; 1 Daimler AG, Stuttgart, Germany; 2 Thermo Fisher Scientific, The Netherlands; 3 Thermo Fisher Scientific GmbH, Dreieich, Germany Keywords Fuel, RFIC System, In-Line Sample Preparation, Dionex IonPac AS11-HC column, Corrosion, Organic s Goal To develop and evaluate a novel automated analytical IC method for the direct determination of acetic acid and formic acid in diesel and oil-containing diesel mixtures. Introduction Organic acids in crude oil, refined products, fuel, etc., are one of the biggest concerns of the oil, oil refining, and car manufacturing industries. As a result of their corrosive properties and the associated costs of corrosion, organic acids have become a topic of great industrial interest 1. Crude oil may contain more than a thousand different organic acids with a mass range up to 1500 Da and a carbon number of about C 802. To assess the acidity of petroleum products the total acid number (TAN), defined by the ASTM as milligrams of KOH required to neutralize one gram of oil or petroleum product 3, is commonly used. Recently, it was determined that TAN is no longer a reliable indicator as a surrogate parameter. The corrosivity of organic acids is related to their molecular mass, size, and structure, 1 so more information about the chemical identity and concentration of organic acids is needed. In particular, low molecular weight organic acids are reported as the most aggressively corrosive in the refining process 4,5.

Different chromatographic approaches are used to identify and quantify organic acids in petroleum products with gas chromatography (GC) being the traditional choice. The organic acids typically with a chain length above C 6 are extracted by solid phase extraction (SPE), then methylated and analyzed by GC-MS 6,7. Yang et al. followed a different approach by injecting the aqueous extracts from oil cracking fractions directly onto a polyethylene glycol coated capillary GC column, detecting a small number of low molecular weight acids, except for formic acid 8. Wang and Kasperski used HPLC-MS/MS for the determination of organic acids (C 6 and above) in aqueous oil sand extracts 9. None of these methods allowed an easy to setup and reliable determination of smaller organic acids in petroleum products. Ion chromatography (IC) can be used for the selective and sensitive determination of low molecular weight organic acids in aqueous or water-miscible matrices 10,11. At first glance, water-immiscible samples, like diesel and diesel mixed with oil, appear to be challenging for the application of IC, as they cannot be directly injected. In modern IC applications, however, in-line matrix elimination techniques are available to remove the hydrophobic matrix and specifically retain ionic analytes prior to the injection. A comparable approach was applied to determination of sodium in diesel by IC analysis 12, while the application of IC to the determination of short-chained organic acids in petroleum products has not been described. This application note summarizes the use and evaluation of an automated analytical IC method for the direct determination of acetic acid and formic acid in diesel, oil, and diesel/oil mixtures, to support research on corrosion processes of industrial production installations and car engines. Equipment Thermo Scientific Dionex ICS-2100* with Degasser (P/N 069576) with additional auxiliary-valve (P/N 069472) Thermo Scientific Dionex UltiMate LPG-3400 SD (P/N 5040.0031) Thermo Scientific Dionex UltiMate WPS-3000 RS (P/N 5840.0010) Reagents and standards Deionized (DI) water, Type I reagent grade, Barnstead GenPure UV, Thermo Fisher Scientific, (P/N 50132156) 2-Propanol, Optima LC/MS, Thermo Fisher Scientific, (P/N A-461-1) 1-Butanol p.a., Chem-Lab NV, Zedelgem, Belgium, (P/N CL00.0220) acid, 99.6%, Across Organics, Geel, Belgium, (P/N 29532) Formic acid, 99%, Across Organics, Geel, Belgium, (P/N 27048) Samples Diesel and synthetic motor oil (5W-40) were bought at a local gas (petrol) station. Conditions Columns Thermo Scientific Dionex IonPac AG11-HC, 2 50 mm (P/N 052963) Dionex IonPac AS11-HC, 2 250 mm (P/N 052961) Dionex IonPac UTAC-LP1 Trace Anion Concentrator Column (P/N 063079) Eluent KOH-Gradient (Table 1) Eluent Source Thermo Scientific Dionex EGC KOH Eluent Generator Cartridge (P/N 074532) with Dionex IonPac CR ATC II (P/N 060477) Flow Rate 0.38 ml/min Inj. Volume 10 µl Temperature 30 C Detection Suppressed Conductivity, Thermo Scientific Dionex AERS 500 Anion Self-Regenerating Suppressor (2 mm), (P/N 082541) Regenerant 0.4 ml/min (external water mode) Flow Rate System 1700 psi Backpressure Background <0.8 µs/cm Conductance Typical Noise <2 ns/cm Thermo Scientific Chromeleon Chromatography Data System Software, Version 7.2 (or higher) * Thermo Scientific Note: Methods performed using a Dionex ICS-2100 system are now run on the Thermo Scientific Dionex Integrion HPIC system. 2

Table 1. Experimental conditions for matrix elimination and chromatography. Time (min) Sample transfer flow rate (ml/min) -5.0 0.1-4.5-3.0-2.5 0.0 5.0 9.0 1.0 Sample transfer and matrix elimination solvent* H 2 O 2-Propanol KOH (mm) 100 1 Suppressor current (ma) 20.0 H 2 O 100 22.5 0.1 95 95 2 Comment Begin conditioning of concentrator and analytical column Begin matrix elimination Matrix elimination finished Injection and begin analysis 37.0 20 34 End sample run * To remove possible contaminants, the solvents used for sample transfer and matrix elimination were pumped over a Dionex ATC-500 polymeric anion exchange column (P/N 075976). Preparation of solutions and reagents Calibration solutions An aqueous stock solution with a final concentration of 1000 mg/l was prepared for acetic acid and formic acid. External standard calibration solutions were obtained by diluting the stock solution with water. Solutions used to evaluate the potential influence of organic diluents on the analytical results were made by mixing the stock solution with 2-propanol. Sample preparation Samples were diluted with the same volume of 1-butanol and stored in glass HPLC autosampler vials until injection. Mixtures of diesel and the synthetic motor oil were prepared at different volume ratios to investigate the potential influence of oil on the method. For standard addition experiments a small volume of the analyte stock solution was added to the selected diesel/oil mixture, e.g., 1 ml of the analyte stock solution to 100 ml of the diesel/oil mixture. The acids were quantitatively extracted into the organic layer by vortexing for 10 min. The non-aqueous layer was then diluted with the same volume of 1-butanol. Aliquot volumes were stored in glass HPLC autosampler vials until injection. In-line matrix elimination The sample was delivered to the Dionex IonPac UTAC-LP1 concentrator column using 2-propanol. The analytes were extracted and the organic sample matrix was removed with an excess of 2-propanol. The Dionex IonPac UTAC-LP1 column was then rinsed with water, removing the 2-propanol. The sample was injected by switching the valve holding the concentrator column into the eluent. Analytes were then eluted and separated on the analytical column (for analytical conditions see Table 1). Figure 1 shows a schematic of the setup. Auto- Sampler W Waste L S P L C Transfer Pump Valve (I) with Injection Loop Dionex ATC-500 3 Waste 4 2 Conductivity Detector 5 1 6 RFIC Eluent Dionex AERS Dionex IonPac AS11-HC Dionex IonPac AG11-HC Valve (II) with Dionex IonPac UTAC-LP1 Figure 1. Schematic configuration of the analytical system. 3

12.0 10.0 8.0 6.0 4.0 2.0 0.0 Figure 2. Chromatogram of a standard solution containing acetic acid and formic acid at 5 mg/l each. 7.00 6.00 5.00 4.00 3.00 2.00 1.00 0.00 Figure 3. acid and formic acid in diesel. 4

Results and discussion External calibrations were carried out using five concentration levels in the range of 1 mg/l to 20 mg/l, with two separate preparations and duplicate injections of each level. A typical chromatogram of a standard solution is shown in Figure 2. Best correlations were obtained using a linear calibration for formic acid and a quadratic calibration for acetic acid. The latter finding results from the concentration-dependent dissociation of a weakly dissociated acid impacting peak area in conductivity detection 13. The coefficients of determination, reflecting the deviation of the measured data points from the calibration curve, resulted to r 2 (acetic acid)=0.9992 and r 2 (formic acid)=0.9996. The calibrations were checked every working day and they remained stable for several weeks. The analytical figures of merit were calculated as described in DIN 32645 14. The detection limit for acetic acid was found to be 3.8 mg/l, and for formic acid 1.5 mg/l. The limit of determination for acetic acid was 6.3 mg/l and for formic acid 2.7 mg/l. The potential impact of using organic diluents on the analytical recovery was tested comparing the response factors for standard solutions prepared in water and 2-propanol. The differences of response factors for both organic acids in both matrices were in the range of one percent, indicating the equivalency of the two approaches. Consequently, all external calibration experiments were performed using aqueous standards, simplifying the daily routine. Figure 3 presents the analysis of acetic acid and formic acid in pure diesel, showing the elution of other components close to the acetic acid peak. This indicates the presence of additional, short-chained organic acids, though at a lower concentration than acetic acid. Intra-day and inter-day repeatability and recovery were determined using diesel samples with 2% motor oil, being spiked with 10 mg/l of each acid. Intra-day repeatability was in the range of 2% for both analytes, with increasing variation towards lower analyte concentration. Intra-day recovery was determined to be 107% for acetic acid and 105% for formic acid. Corresponding chromatograms, with and without the addition of acetic and formic acid, are shown in Figure 4. Inter-day recovery for acetic acid over four days was 101% ± 5% and for formic acid it was 90% ± 9%. Inter-day repeatability for the analytes on the spiked levels was 4% for acetic acid and 9% for formic acid. Recovery for acetic acid remained stable over the time investigated, while the concentration of formic acid trended down, possibly due to evaporative losses as a consequence of higher vapor pressure compared to acetic acid. The determination of formic acid should therefore be done as fast as possible after the sample s arrival in the laboratory. 7.00 6.00 5.00 4.00 3.00 2.00 1.00 0.00 (B) (A) Figure 4. Determination of acetic acid and formic acid in diesel containing 2% motor oil. (A) original sample (B) sample spiked with 10 mg/l of each organic acid (an offset of 20% was applied). 5

The possible impact of varying oil contents was evaluated by injecting diesel samples containing 2-, 5- and 10% motor oil, as well as being spiked with 10 mg/l of both organic acids. As an example, the chromatogram of the diesel sample containing 10% oil is presented in Figure 5. While the oil used to spike the diesel sample contained a negligible amount of formic acid, acetic acid was present at a significant level (278 mg/l). The values for acetic acid were blank-corrected, resulting in an average recovery rate of 101% ± 4% for both organic acids and for samples with varying oil content. Applying the standard addition approach of adding oil to diesel allows the determination of both organic acids even in pure oil samples basically using diesel as additional diluent for the oil (Figure 6). 7.00 6.00 5.00 4.00 3.00 2.00 1.00 0.00 Figure 5. acid and formic acid in diesel plus 10% motor oil. Standard addition of 10 mg/l of each acid. 40 35 30 c (mg/l) 25 20 15 y = 2.78 x + 10.08 r = 0.9999 5 0 0 2 4 6 8 10 12 v Oil (ml/100 ml) Figure 6. Addition of oil to diesel spiked with 10 mg/lof each acid. 6

In addition to clean diesel and lab-made mixtures of diesel with oil, the method was applied to samples taken from stressed car engines. An example is shown in Figure 7. It contained 8.4 mg/l acetic acid and 1.8 mg/l formic acid. In addition to the two focus analytes of this study, additional peaks were detected, suggesting the presence of other short-chained organic acids. Identification of these unknowns could be done by applying MS-detection in conjunction with IC. The use of continuously regenerated suppressors permits the use of high-ph eluents, as well as gradients due to the neutralization of the column s effluent prior to the MS-interface. Results of initial experiments already identified unexpected organic acids, suggesting additional sources contributing to the potpourri of organic acids found in real world samples. Conclusions A new IC method was developed allowing the determination of acetic acid and formic acid in diesel and oil-containing diesel samples. Prior to the chromatographic analysis, the analytes were extracted and the matrix was eliminated based on a non-aqueous anion exchange step. With detection limits and quantification limits in the single-digit mg/l range and high repeatability and recovery both intra-day and inter-day the organic acids can be determined at low concentrations in diesel, diesel oil mixtures, and motor oil. Mixtures with up to 10% oil content were analyzed, showing no impact on the analytical performance, or method stability. 7.00 6.00 Diesel: TAN 0.06 5.00 4.00 3.00 2.00 1.00 0.00 Figure 7. Diesel sample taken from a stressed diesel engine (TAN = 0.06). 7

References 1. Barrow, M. P.; McDonnel, A.; Feng, X.; Walker, J.; Derrick, P. J. Determination of the Nature of Naphthenic s Present in Crude Oils Using Nanospray Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. Analytical Chemistry, 2003, 860 866. 2. Terra, L. A.; Filgueiras, P. R.; Tose, L. V.; Romão, W.; de Souza, D. D.; de Castro, E. V. R.; de Oliveira, M. S. L.; Dias, J. C. M.; Ronei, P. J. Petroleomics by electrospray ionization FT-ICR mass spectrometry coupled to partial least squares with variable selection methods: prediction of the total acid number of crude oils. Analyst, 2014, 4908 4916. 3. ASTM International. ASTM D664-11ae1, 2016. http://www.astm.org (accessed September 7, 2016). 4. Kane, R. Corrosion in Petroleum Refining and Petrochemical Operations. In ASM Metals Handbook; ASM International: Materials Park (OH), 2006; Vol. 48, pp 967 1014. 5. Kane, R. D. ATI Corrosion Conference, 2011. ATI Corrosion Conference. 6. Jones, D. M.; Watson, J. S.; Meredith, W.; Chen, M.; Bennett, B. Determination of Naphthenic s in Crude Oils Using Nonaqueous Ion Exchange Solid- Phase Extraction. Analytical Chemistry, 2001, 703 707. 7. Damasceno, F. C.; Gruber, L. D. A.; Geller, A. M.; Vaz de Campos, M. C.; Gomes, A. O.; Guimarães, R. C. L.; Péres, V. F.; Jacques, R. A.; Caramão, E. B. Characterization of naphthenic acids using mass spectroscopy and chromatographic techniques: study of technical mixtures. Analytical Methods, 2014, 807 816. 8. Yang, B.; Xu, C.; Zhao, S.; Hsu, C. S.; Chung, K. H.; Shi, Q. Thermal transformation of acid compounds in high TAN crude oil. Science China, 2013, 848 855. 9. Wang, X.; Kasperski, K. L. Analysis of naphtenic acids in aqueous solution using HPLC-MS/MS. Analytical Methods, 2010, 1715 1722. 10. Weiss, J. Handbook of Ion Chromatography, 3rd ed.; WILEY-VCH Verlag GmbH & Co. KGaA.: Weinheim, 2004. 11. Fischer, K. Environmental analysis of aliphatic carboxylic acids by ion-exclusion chromatography. Analytica Chimica Acta, 2002, 157 173. 12. Thermo Scientific Customer Application Note 118. 2014. https://tools.thermofisher.com/content/sfs/ brochures/can-118-ic-trace-sodium-diesel- Biodiesel-Fuel-AN71197-EN.pdf (accessed November 4, 2016). 13. Brinkmann, T.; Specht, C. H.; Frimmel, F. H. Non-linear calibration functions in ion chromatography with suppressed conductivity detection using hydroxide eluents. Journal of Chromatography A, 2002, 99 109. 14. DIN 32645:2008-11. Chemical analysis Decision limit, detection limit and determination limit under repeatability conditions Terms, methods, evaluations; Beuth Verlag GmbH: Berlin, 2008. Find out more at thermofisher.com/chromatography 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. CAN72204-EN 1216M