Towards SI traceable humidity measurements with radiosondes

Similar documents
A calibration facility for automatic weather stations

The Vaisala Reference Radiosonde Program: First Results and Future Plans

GRUAN Station Report for Ny-Ålesund

e-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard

From MeteoMet to MeteoMet2. and more. Andrea Merlone

Specifications for a Reference Radiosonde for the GCOS Reference. Upper-Air Network (GRUAN)

CERTIFICATE OF CALIBRATION

2 Calibration and test facilities

Overview. TC Thermometry Graham Machin. Introduction to TC-T Annual meeting Selected comparisons TC-T training for future thermal metrologists

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

New Peltier-based chilled-mirror hygrometer SKYDEW

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

The Meisei sonde data product

The Fundamentals of Moisture Calibration

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C

Uncertainty Evaluation of the New Setup for Measurement of Water-Vapor Permeation Rate by a Dew-Point Sensor

REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER

VAISALA RADIOSONDE RS41 COVER IMPROVEMENT

Calibration of temperature sensors within Length Standard Section of NMIJ

The calibration of Hygrometer (LectureandTraining )

BIPM, EURAMET AND METEOMET

Field experiments at Sodankylä

Development of the High-Temperature Dew-Point Generator Over the Past 15 Years

Introduction of RIC Tsukuba (Japan, RAII)

Change management for the RS92 phase out

Correlation Effects in the Uncertainty Estimation of Two-Pressure Humidity Generators

Report of CoreTemp2017: Intercomparison of dual thermistor radiosonde (DTR) with RS41, RS92 and DFM09 radiosondes

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements

DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR

GRUAN Data Flow and RS92 Data Product v2

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

LAPBIAT ATMOSPHERIC SOUNDING CAMPAIGN IN 2010: UPPER AIR AND REMOTE SENSING OBSERVATIONS OF WATER VAPOR

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

Calibration capabilities at PTB for radiation thermometry, quantitative thermography and emissivity

IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT

UPDATE on: Consistency for water vapour of GRUAN, LBLRTM and IASI

13B.2 COMPARISON OF SELECTED IN-SITU AND REMOTE SENSING TECHNOLOGIES FOR ATMOSPHERIC HUMIDITY MEASUREMENT

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

Uncertainty Propagation for Force Calibration Systems

CALIBRATION REPORT FOR THERMOHYDROMETER

Quality assurance for sensors at the Deutscher Wetterdienst (DWD)

The GRUAN Observing Station Payerne - Switzerland

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

Components of uncertainty in measurements by means of AWS

CERTIFICATE OF CALIBRATION

GRUAN-wide strategy for the transition away from the RS92 radiosonde

Infrared Thermometer Calibration A Complete Guide

Intercomparison of Mobility Particle Size Spectrometers

DEVELOPMENT OF MEASUREMENT STANDARD FOR DYNAMIC PRESSURE AT MIKES

Chilled Mirror Hygrometry

How do you really know what the temperature is? Michael de Podesta

CONTENT 2. ORGANIZATION 3. SERVICES. instruments. (3)Activities of RIC Tsukuba

Science of Stability 2017

Establishing traceability and estimating measurement uncertainty in physical, chemical and biological measurements

Interaction with the metrology community

473 Dew Point Hygrometer

Water Vapour Effects in Mass Measurement

Ny-Ålesund: Site Update & Atmosphere Flagship Program. Marion Maturilli, Sandro Dahlke, Roland Neuber, Andrea Merlone

Interlaboratory Comparison & Proficiency Testing Program

Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge

Version 4.0 (09/2017) Version 3.0 (02/2015) Version 2.0 (03/2011) Version 1.0 (07/2007) EURAMET e.v. Bundesallee 100 D Braunschweig Germany

WG 1 workshop on the LAPBIAT Atmospheric Sounding campaign at the Finnish Meteorological Institute in Helsinki, September 2-3, 2010.

Instruments and Methods of Observation Programme, the Report of the President of CIMO. Report to Cg-XV May 2007 Dr. J. Nash President of CIMO

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

TC Thermometry Graham Machin

473 Dew Point Hygrometer

OverviewofRIC Tsukuba (Japan,RAII )

Withdrawn at December 31, Guideline DKD-R 5-7. Calibration of Climatic Chambers DEUTSCHER KALIBRIERDIENST

Reference quality upper-air measurements: GRUAN data processing for the Vaisala RS92 radiosonde

Edgetech Instruments Inc.

Metrology Principles for Earth Observation: the NMI view. Emma Woolliams 17 th October 2017

I. Yang, C. H. Song, Y.-G. Kim & K. S. Gam

Traceable Mass Determination and Uncertainty Calculation

Comparison of GRUAN profiles with radio occultation bending angles propagated into temperature space

Temperature and Length. Pieter Greeff October 2013

AirCore flights at Sodankylä Rigel Kivi, Pauli Heikkinen, Juha Hatakka, Tuomas Laurila, Leif Backman, Jouni Pulliainen (1), Huilin Chen (2, 3)

Site Report: Lauder, New Zealand

Results from WMO High Quality Radiosonde Comparison, Mauritius As related to planning Reference Upper Air Observations for GCOS

AN EXPERIMENTAL STUDY OF THE FROST FORMATION ON A COLD SURFACE IN FREE CONVECTIVE FLOW

Certificate of Accreditation to ISO/IEC 17025:2005

Humidity Calibration Solutions

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

The GRUAN Observing Station Payerne - Switzerland

Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C

Study on a thermal diffusivity standard for the laser flash method measurements 1

* * * Table (1) Table (2)

Humidity Calibration Solutions

Final report to the CCT on key comparison CCT-K6 Comparison of local realisations of dew-point temperature scales in the range -50 C to +20 C

BUREAU INTERNATIONAL DES POIDS ET MESURES

EVALUATION OF THE ACCURACY OF A THERMOMETRIC RESISTANCE BRIDGE TO DIFFERENT VALUES OF ATMOSPHERIC PRESSURE

Commissioning of the Beta Secondary Standard (BSS2)

Climate & Earth System Science. Introduction to Meteorology & Climate. Chapter 05 SOME OBSERVING INSTRUMENTS. Instrument Enclosure.

473-SHX Dew Point Mirror

Proper Platinum Resistance Thermometer Calibration Uncertainty Analysis

Vocabulary of Metrology

A COMPARISON OF TWO WEATHER DATA ACQUISITION METHODS FOR THE CALIBRATION OF THE PTB GEODETIC BASELINE

The Payerne Meteolabor Radiosonde

Transcription:

VTT TECHNICAL RESEARCH CENTRE OF FINLAND LTD Centre for Metrology MIKES MIKES METROLOGY Towards SI traceable humidity measurements with radiosondes GRUAN ICM-9, MeteoMet Training Course FMI, Helsinki, 15 th June 2017 Martti Heinonen

Outline Introduction SI traceability in humidity measurements Challenges of establishing traceability in radiosonde-based humidity measurements Development of a reference humidity calibration system for radiosondes at VTT MIKES Future Conclusion 23/06/2017 2

Introduction SI system 23/06/2017 3

SI traceability in humidity measurements Measurement For each measurement/calibration of the chain: Uncertainty estimation Documented procedures and results Competence Validity for the application (interval of calibrations, conditions etc.) Calibration Calibration standard Calibration Calibration standard Calibration Reference hygrometer or calibrator Calibration standard Realisation of metre Calibration Realisation of pressure Calibration of barometer Realisation of dew-point temperature Calibration of thermometer SPRT Calibration of SPRT Realisation of kg SI system Realisation of kelvin (ITS-90) 23/06/2017 4

Challenges of establishing traceability in radiosonde-based humidity measurements Uncertainties due to measurement conditions Calibration of a radiosonde It s not feasible to calibrate each device by user Disposable => re-calibration is not possible Representativity of the calibration: Large operating dew-point temperature, air temperature and pressure ranges Measurements down to trace humidity region => stabilisation time for any calibration setup is long Altitude [km] 30 28 26 24 22 20 18 16 14 12 10 Sodankylä 2012-03-21 Tfp CFH ascent Tfp Vaisala RR01 Vaisala RR01 Uncertainty estimate (k=2) T ambient, RS92 Manufacturers have a key role in establishing the traceability link to radiosondes 8-95 -90-85 -80-75 -70-65 -60-55 -50 Temperature [ C] [H. Turtiainen, MeteoMet Workshop, Helsinki 10.9.2013] 23/06/2017 5

Development of a reference humidity calibration system for radiosondes at VTT MIKES (1/2) Motivation: To enable manufacturers to demonstrate their humidity calibration capability of radiosondes in a full operation range To enable users of radiosondes to perform comparative humidity measurements in well-controlled conditions To support the development of radiosondes to fulfil the GRUAN requirements 23/06/2017 6

Development of a reference humidity calibration system for radiosondes at VTT MIKES (2/2) Major challenges: To meet GRUAN accuracy requirements To enable smooth and fast calibration Development in EMRP MeteoMet projects: MeteoMet1 (2011 2014): calibrations at surface pressures MeteoMet2 (2014 2017): extension to low pressures (down to 7 hpa, absolute pressure) and improved operation at low frost-point temperatures Tests with RR01 prototype reference radiosondes of Vaisala 23/06/2017 7

Progress Concept of a new calibration apparatus for reference radiosondes 1 New measurement chamber for radiosonde calibrations 2 Characterization of the apparatus at 1000 hpa absolute pressure 3,4 The development of the apparatus continues by extending the operation range down to 10 hpa (abs.) 5,6 1) Sairanen H., Heinonen M., Högström R., Lakka A., Kajastie H. (2014). A calibration system for reference radiosondes that meets GRUAN uncertainty requirements. NCSLI Measure, 9(3), 56-60. 2) Lakka, A., Sairanen, H., Heinonen, M., & Högström, R. (2015). Comsol Simulations as a tool in Validating a Measurement Chamber. Int. J. Thermophysics, 36(12), 3474-3486. 3) Sairanen, H., Heinonen, M., & Högström, R. (2015). Validation of a calibration set-up for radiosondes to fulfil GRUAN requirements. Meas. Sci. Technol., 26(10), 105901. 4) Sairanen, H. (2015). Advances in Humidity Measurement Applications in Metrology. Doc.Thesis, Aalto University. 5) Salminen J., Sairanen H., Grahn P., Högström R., Lakka A., Heinonen M., Characterization of the humidity calibration chamber by numerical simulations, Int. J. Thermophys. (2017) 38:84-97 6) Sairanen H., Heinonen M., Högström R., Lakka A., Kajastie H., Low pressure and temperature dew/frost-point generator, submitted to Int. J. Thermophys. 23/06/2017 8

GRUAN specifications GRUAN (GCOS Reference Upper-Air Network) requires accuracy of 2 % in terms of mixing ratio in humidity measurements The GRUAN requirements should be achieved in humidity range down to 0.1 ppm(v), corresponding to frost-point of about -90 C in temperature range down to 170 K Vertically the measurement range is up to 50 km (except humidity up to 30 km) About 10 hpa (abs.) is the minimum pressure 23/06/2017 9

Calibration apparatus (1/5) 23/06/2017 10

Calibration apparatus (2/5) Flow control unit includes Mass flow controllers (Bronkhorst) Pre-saturator Dew-point temperature measurement (Vaisala DMP248) MFCs were characterized and calibrated for varying outlet pressures 23/06/2017 11

Calibration apparatus (3/5) Low temperature generator includes Ethanol bath (Kambic ULT-22) Pressure measurement Temperature probes (Fluke 5609-20) Further details in H. Sairanen et al., TEMPMEKO 2016 (Zakopane, Poland, 26.6. 1.7.2016) H. Sairanen et al., Low pressure and temperature dew/frost-point generator, submitted to Int. J. Thermophys. 23/06/2017 12

Calibration apparatus (4/5) Immersed in the bath Saturator Pressure reducer Heat exchanger Measurement chamber This part of the apparatus was characterized in Sairanen, H., Heinonen, M., & Högström, R. (2015). Validation of a calibration set-up for radiosondes to fulfil GRUAN requirements. Measurement Science and Technology, 26(10), 105901. 23/06/2017 13

Calibration apparatus (5/5) Pressure control includes Pressure controller (Bronkhorst) Vacuum pump (Edwards nxds 15i) 23/06/2017 14

Characterization of the apparatus The apparatus was first characterized at 1000 hpa (abs.) Improvements for 10 hpa (abs.) were completed and the apparatus was characterized to the complete pressure range Performance of the apparatus was tested for both descending and ascending pressures Different flow rates were also tested Characterization was carried out in two steps: 1) independently for key parts 2) for the complete apparatus Outcomes of the characterization measurements indicated no pressure dependancy in the performance of heat exchangers and saturators 23/06/2017 15

Performance An example of a test measurement run with a RR01. Red ellipses indicate data that were recorded during tests for the effect of changing flow rate through the test chamber. 23/06/2017 16

Uncertainty analysis Uncertainty of the apparatus was analysed according to GUM 1 Combined uncertainty of the apparatus depends on flow rates, temperatures and pressures Most dominant uncertainty source depends on operating conditions Largest contributors are low temperature generator, saturator, MFCs and adsorption/desorption effects When operating at about 1000 hpa (abs.) relative expanded uncertainty (k = 2) is less than 1.7 % (GRUAN requires 2 %) Experiments carried out show no pressure dependancy for the uncertainty. 1) JCGM 2008 Evaluation of Measurement Data Guide to the Expression of Uncertainty in Measurement 23/06/2017 17

Example of an uncertainty estimation for the VTT MIKES humidity calibration system at water vapour mixing ratio of 1.9 10-7 (0.2 ppm) 23/06/2017 18

Discussion The reference humidity calibration system for radiosondes developed at VTT MIKES Can be used for demonstrating calibrations at the uncertainty level specified by GRUAN in the complete operation range This is the first primary humidity standard for low pressures and thus it is extremely important to demonstrate its performance by an interlaboratory comparison VTT MIKES will carry out a comparison with INRiM (Italy) in summer 2017 The measurement chamber was optimised for RR01; modifications and/or further development is needed to enable measurements with other types of sensors 23/06/2017 19

Future Services for customers Development of more versatile measurement chamber Enabling calibrations for different kinds of radiosondes Extension to space applications System has already been used in testing sensors developed for monitoring atmosphere on Mars Non-air atmosphere 23/06/2017 20

Conclusion Metrological traceability is a key feature of results with known quality Manufacturers have an important role in establishing the traceability link between individual radiosondes and SI Also, uncertainties related to actual measurement are important for the traceability of measurement results MeteoMet developed tools for demonstrating calibration capability of radiosonde manufacturers performing comparative humidity measurements in well-controlled conditions 23/06/2017 21