THE ACTIVITY CALIBRATOR

Similar documents
Laboraty Equipment. Dose Calibrator. General. J Habraken, St. Antonius Hospital, Nieuwegein Sara H Muller, The Netherlands Cancer Institute, Amsterdam

ความร พ นฐานและเทคน คการบ าร งร กษา เคร องโดสคาล เบรเตอร (Dose Calibrator) ส ว ทย ป ณณช ยยะ

RPR 29 CYCLOTRON RADIOCHEMISTRY LABORATORY

Intercomparison of 131 I and 99m Tc Activity Measurements in Brazilian Nuclear Medicine Services

RADIOPHARMACY PURPOSE

GLOSSARY OF BASIC RADIATION PROTECTION TERMINOLOGY

Radionuclide Imaging MII Detection of Nuclear Emission

Basic physics of nuclear medicine

Detection and measurement of gamma-radiation by gammaspectroscopy

SODIUM PERTECHNETATE ( 99m Tc) INJECTION (FISSION): Revised Final text for addition to The International Pharmacopoeia (January September 2009)

ESTIMATION OF 90 SCATTERING COEFFICIENT IN THE SHIELDING CALCULATION OF DIAGNOSTIC X-RAY EQUIPMENT

Derivation of Calibration Factors and Dial Settings Michaela Baker

Ionizing Radiation Group K.B. Lee

CHARGED PARTICLE INTERACTIONS

Semiconductor Detector

Chemical Engineering 412

Radiation Detection and Measurement

Mitigation of External Radiation Exposures

ISOCS / LabSOCS. Calibration software for Gamma Spectroscopy

Michael G. Stabin. Radiation Protection and Dosimetry. An Introduction to Health Physics. 4) Springer

Calibration of a Whole Body Counter and In Vivo measurements for Internal Dosimetry Evaluation in Chile, Two years experience.

Outline. Radiation Interactions. Spurs, Blobs and Short Tracks. Introduction. Radiation Interactions 1

PRODUCTS FOR EDUCATION AND TRAINING

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Jazan University College of Science Physics Department. Lab Manual. Nuclear Physics (2) 462 Phys. 8 th Level. Academic Year: 1439/1440

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

APPLIED RADIATION PHYSICS

City University of Hong Kong

Nuclear Medicine Physics Course Laboratory Number 3: Dose Calibrator Quality Control

arxiv:nucl-ex/ v2 21 Jul 2005

Shielding of Ionising Radiation with the Dosimetry & Shielding Module

Radiation Detection. 15 th Annual OSC Readiness Training Program.

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

UNCORRECTED PROOF. Table of Contents

Use of a radioactive check device for redundancy check of ionization chambers

Survey Meter OD-01 Address: Phone: Fax: URL:

Some nuclei are unstable Become stable by ejecting excess energy and often a particle in the process Types of radiation particle - particle

Radiation Safety Training Session 1: Radiation Protection Fundamentals and Biological Effects

A Radiation Monitoring System With Capability of Gamma Imaging and Estimation of Exposure Dose Rate

Absorption of Gamma Rays

General Overview of Gas Filled Detectors

Supervised assessment: Ionising radiation

Chapiter VII: Ionization chamber

Scintillation Detector

Chapter 16 Basic Precautions

Introduction to Radiological Sciences Neutron Detectors. Theory of operation. Types of detectors Source calibration Survey for Dose

Radiopharmacy quality control

TRAINING IN EXTERNAL DOSIMETRY CALCULATIONS WITH COMPUTATIONAL CODES

Design, Construction, Operation, and Simulation of a Radioactivity Assay Chamber

QUIZ: Physics of Nuclear Medicine Atomic Structure, Radioactive Decay, Interaction of Ionizing Radiation with Matter

Chapter Seven (Nuclear Detectors)

X-ray Interaction with Matter

DETECTORS. I. Charged Particle Detectors

Radiation safety of the Danish Center for Proton Therapy (DCPT) Lars Hjorth Præstegaard Dept. of Medical Physics, Aarhus University Hospital

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

Lab NUC. Determination of Half-Life with a Geiger-Müller Counter

Radiation Protection Fundamentals and Biological Effects: Session 1

MEDICAL EQUIPMENT: NUCLEAR MEDICINE. Prof. Yasser Mostafa Kadah

Bases of radioisotope diagnostic methods

RADIOACTIVITY IN THE AIR

Determining the Need For External Radiation Monitoring at FUSRAP Projects Using Soil Characterization Data. Todd Davidson

INTERACTIONS OF RADIATION WITH MATTER

Recent Primary Standardisations and Traceability in Nuclear Medicine. Lena Johansson, NPL Radioactivity Group

Gamma Spectroscopy. References: Objectives:

Introduction. Principle of Operation

Wallace Hall Academy Physics Department. Radiation. Pupil Notes Name:

Physics in Nuclear Medicine

Absolute activity measurement

Problem P7. Stéphanie Ménard. Dosimetry Department Fontenay-aux FRANCE IRSN QUADOS IRSN

Foundation of Radiological Physics. Preface

Updated Contact Dose Rate Conversion Factors for Encapsulated Gamma Sources Including Secondary Electron Emission

Compton Camera. Compton Camera

ɣ-radiochromatography

PHYSICS 176 UNIVERSITY PHYSICS LAB II. Experiment 13. Radioactivity, Radiation and Isotopes

SCINTILLATION DETECTORS & GAMMA SPECTROSCOPY: AN INTRODUCTION

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the UAL

Radiation Safety Talk. UC Santa Cruz Physics 133 Winter 2018

Radionuclide Information Booklet

Progress in Nuclear Science and Technology, Volume 6,

Project Memorandum. N N o. = e (ρx)(µ/ρ) (1)

Half-life data a critical review of TECDOC-619 update

Development of Radioactivity Standards for Quantitative Positron Emission Tomography

Today, I will present the first of two lectures on neutron interactions.

K 40 activity and Detector Efficiency

What is scintigraphy? The process of obtaining an image or series of sequential images of the distribution of a radionuclide in tissues, organs, or

Investigation of Uncertainty Sources in the Determination of Gamma Emitting Radionuclides in the WBC

CT-PET calibration : physical principles and operating procedures F.Bonutti. Faustino Bonutti Ph.D. Medical Physics, Udine University Hospital.

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS IB PHYSICS

RADEAGLET. Lightweight Handheld Radioisotope Identification Device

Dependence Of Gamma Ray Attenuation On Concentration Of Manganese (II) Chloride Solution

Geant4 Monte Carlo code application in photon interaction parameter of composite materials and comparison with XCOM and experimental data

Gamma-Spectrum Generator

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Quality Assurance. Purity control. Polycrystalline Ingots

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

EEE4106Z Radiation Interactions & Detection

Jeancarlo Torres, CHP. Naval Surface Warfare Center Carderock Division

PHYS 3650L - Modern Physics Laboratory

Choose your application of

Georgia Institute of Technology. Radiation Detection & Protection (Day 3)

Transcription:

A.O.U. OSPEDALI RIUNITI di TRIESTE S.C. di FISICA SANITARIA THE ACTIVITY CALIBRATOR Dr. Maria Rosa Fornasier 1

INDEX GENERAL FEATURES DETECTOR DESIGN CALIBRATION PROCEDURE - EFFECTS OF AN EXTERNAL SHIELD - CALIBRATION GEOMETRY (EFFECTS OF CONTAINER MATERIAL, SAMPLE VOLUME AND POSITION) - EFFECTS OF IMPURITIES QUALITY CONTROL (PRECISION; RELATIVE RESPONSE TO REFERENCE SOURCE; LINEARITY OF RESPONSE AS A FUNCTION OF ACTIVITY; ACCURACY; GEOMETRY ) 2

THE ACTIVITY CALIBRATOR: GENERAL FEATURES DETECTOR USED AS ACTIVITY CALIBRATOR should have the following characteristics: Sensitive to energies between 25 kev and 2 MeV [ 99m Tc (140.5 kev), 57 Co (122.1 ; 136.5 kev), 58 Co (810.8 kev), 201 Tl (135.3 ; 167.4 kev), 131 I (284.3 ; 364.5 ; 637.0 kev), 123 I (159.0 kev), 67 Ga (300.2 kev), 111 I (171.3 ; 245.4 kev), 133 Xe (81.0 kev) ] Linear response over a wide range of activity levels ( 10 kbq 10 GBq ) Stability Possibility to measure sample volumes like vials and syringes 3

GAS DETECTORS a potential is applied across the collection electrodes when the photon interacts in the gas, a large number of electron-ion pairs are formed and are collected by electrodes the current measured is dependent on the applied voltage collecting voltage high enough to obtein saturation current, to prevent recombination of the electron - ion pairs (the probability for recombination increases with dose rate) the current measured is proportional to the energy deposited the ionization current is not linear with incident photon energy 4

DETECTOR DESIGN well geometry (the sample is almost completely surrounded by the sensitive volume of the chamber) maximizes the geometric efficiency (near 4 geometry) and minimizes the effects of small variations in sample position and volume two co-axial cylindrical collection electrodes manteined at a voltage difference (tipically: 150-350 V) walls in the well thin to minimize absorption gas pressurized to increase the linear attenuation coefficient which in turn increases sensitivity container sealed to avoid temperature and pressure dependence gas with high Z (e.g. Ar, Z = 18) to increase the number of ion pairs produced per interaction gas inert to avoid corrosive effects ionization current is integrated by a ratemeter with a fixed time constant and converted to a voltage signal, which is amplified, processed and displayed, commonly in digital form in units of activity 5

fixed isotope corrections for commonly used isotopes (push buttons) holders made of low-z material to ensure reproducibility of samples and reference sources positioning for TYPICAL COMPONENTS OF AN ACTIVITY CALIBRATOR 6

CALIBRATION PROCEDURE measure the response of the chamber to all available calibrated standards of radionuclides emitting monoenergetic photon over the appropriate range of energy determine the efficiency of the chamber as a function of gamma-ray energy calculate the calibration factors for samples of photon-emitting radionuclides for which standards are not available (using the efficiency curve and the decay data) 7

def R A ( response of the detector to a radioisotope A) = detector output / activity of the radioisotope being measured It is convenient to express the response of the detector to a radioisotope A relative to that of a standard reference radioisotope (e.g. 60 Co, selected because its decay scheme is simple and it is one of the most commonly used radioisotope standard) 8

the calibration factor K R, relative to that of the reference source is given by: where K R, = P A (R R /R) R = output for the calibrated source emitting a photon of energy E P probability of emission per decay A = activity of the calibrated source R R = output for the long-lived reference source 9

RECIPROCAL of the RELATIVE FACTOR K R, is EQUIVALENT to an EFFICIENCY an EFFICIENCY CURVE can be established as a FUNCTION of ENERGY Activity of any other monoenergetic gamma-emitting nuclide can be determined using an interpolate value of 1/K A = (K R, / P ) ( R / R R ) Activity of a nuclide emitting n photons of different energies and of known probabilities per decay can be obtained by weighting the efficiency 1/K R, for each photon of different energy by the appropriate probability per decay where the sum is carried out for all photons A = (R/R R ) i (K R, /P ) i 10

RESPONSE OF THE INSTRUMENT DISPLAYED DIRECTLY IN TERMS OF ACTIVITY GAIN SETTING : adjust the gain of the calibrator amplifier by dividing the amplifier output by some unique value for each radioisotope 11

EFFICIENCY OF A TIPICAL IONIZATION CHAMBER OF A CALIBRATOR AS A FUNCTION OF PHOTON ENERGY 12

COMMENTS sensitivity is normalized by chamber response to 60 Co radiations (SUZUKY et al.) energy range up to 1.9 MeV all photons with energies below about 13 ev are stopped before they reach the sensitive volume of the chamber (these photons don t contribute to the activity measurement) a significant fraction of photons with energies below 50 kev are stopped in the chamber walls; only few can enter the sensitive volume of the chamber and contribute to the activity measurement 13

the peak in the low energy region of the sensitivity curve is due to the rapid increase in photoelectric effect as photon energy decreases (Z 4 /E 3 ) the attenuation of low energy photons by the sample holder, the chamber liner and the chamber walls the absorption of photons in the sample material and its container above a photon energy of 200 kev, the ionization in the chamber is mainly due to electrons resulting from Compton scattering of photons by the filling gas (Ar) and by the chamber walls (Al) 14

BETA-RAY CORRECTIONS for radioisotops with high-energy beta-ray emission, the contribution from beta emission may have to be included in the calculation of the response it is difficult to define a detector s sensitivity to beta rays because of the energy distribution of the emitted beta rays (ranging from 0 to the maximum transition energy) and because of the strong absorption of the radiation by the media (only bremmsttrahlung is detectable by most calibrators due to the rather thick wall of the detector and the sample container) 15

EFFECTS OF AN EXTERNAL SHIELD Advantage of the external shield: reduction of radiation exposure to the personnel handling the radioisotopes reduction of the background effects on the activity measurements prevention of efficiency changes caused by scattering material in the vicinity The shielding around the dose calibrator should remain the same after the initial calibration 16

If a shield is placed around a calibrator: the sensitivity of the ionization chamber is enhanced due to backscattering of photons by the shielding at energies below 70 kev the backscatter is absorbed in the chamber walls the backscattering effects are more significant for photon energies between 70 and 250 kev than for other energy regions above about 250 kev the scattering of photons is mainly directed forward and at the low energy region 17

CALIBRATION GEOMETRY The calibration geometry will affect the reading of unknown samples: calibration done with non-diluted point sources the activity in a sample diluted in a volume will be underestimated (selfabsorption in the sample) calibration done with diluted sources the activity of a point source measured in the dose calibrator will be overestimated Generally the calibration sources have a similar geometry as the samples; the user should select a standardized procedure, volume and container for all radioactivity measurements 18

Tipical calibration source geometry: 5 g radioactive solution (~ 5 ml) glass ampoules Ø 17 mm - height 40 mm wall thickness 0.6 mm EFFECTS OF CONTAINER MATERIAL, SAMPLE VOLUME AND POSITION radioisotope measured in a different material or wall thickness than that of standards possible errors for radioisotopes having a significant abundance of low-energy gamma and x-ray or high-energy beta-ray; often correction factors are supplyed by the manufacturer (the radioisotopes more sensitive to source configuration and type of container are 125 I and 133 Xe; for 125 I a correction factor of 2 is not unusual if the isotope is assayed in a glass vial) 19

sample volume (effect due to position of the activity within the chamber well; for samples emitting low energy gamma-rays, absorption in the sample and syringe is important ) position within the chamber well (vertical displacements of the source the source holders provided for use generally ensures that the correct vertical position is manteined; it is often easier to produce a horizontal displacement of the source) in radioactive solutions that tend to be instable, part of the radionuclide may be present in a gaseous phase (f.e. 133 Xe in saline) reading depending on the partitioning between the gaseous and solution phases in radioactive solutions that tend to plate-out onto the walls of container reading may be affected by change in geometry and internal absorption 20

EFFECTS OF IMPURITIES A ionization chamber doesn t have intrinsic energy-discrimination capability the presence of radioisotope impurities affects the reading of the calibrator, particularly in measurements of short lived radionuclides several half-lives akfter initial preparation(unless this effect is eliminated by photon filtration as with 99 Mo breakthrough in 99m Tc) determination of the photon energy with a photon spectrometry system may be necessary 21

PROCEDURE TO DETERMINE CALIBRATION FACTORS FOR CLINICAL RADIONUCLIDES FOR WHICH NO SETTING AND NO STANDARD IS AVAILABLE preparation of standard samples of radionuclides (range of activities used clinically) determination of the activity of samples by a Ge detector (because of the difference in sensitivity between Ge detector and dose calibrator, dilution of the solution to be standardized is required for measurement with Ge counting system) once the specific activity of the solution is known, a volume of the solution representing a known activity can be measured (correction for radioactive decay over the time interval between standardization and measurement) 22

QUALITY CONTROL in order to assure that the activity calibrator is operating properly (correct assays of radiopharmaceuticals prior to administration to patients) it must be tested on a regular basis PROTOCOLS ANSI N42.13 1986 : Calibration and Usage of Dose Calibrators Ionization Chambers for the Assay of Radionuclides, 1986 IAEA TECDOC - 602 : Quality Control of Nuclear Medicine Instruments, 1991 National Council on Radiation Protection and Measurement : Quality Assurance for Diagnostic Imaging. NCRP Report No. 99, 1988 23

LIMITS OF ACCEPTABILITY Background Precision Reproducibility Accuracy Linearity Geometry (Stability) ANSI (d) 5 % (1) 10 % (d) 10 % (1) 5 % (q) -- IAEA < 20 % (2) (d) 5 % (q) 5 % (d) 10 % (q) 10 % (q) -- NCRP (d) 5 % (d) 5 % (q) 5 % (a) 5 % (q) 2 % (3) (1) for activity > 3.7 MBq (2) Increase in background 20 % further investigation (3) Variations > 2 % apply correction factors d DAILY; q QUARTERLY; y YEARLY 24

DAILY CHECKS Internal Test - test of circuitry CHECK OF BACKGROUND RESPONSE to check the background response at the radionuclide settings to be used - the background reading should be manually or automatically subtracted from subsequently measured activities - significant increase in background radioactive contamination of the instrument (remove the liner and repeat the measure) or increased environmental radiation from local sources 25

DAILY CHECKS CHECK OF REPRODUCIBILITY OF RESPONSE to check the day-to-day reproducibility of performance of a calibrator in measurements of commonly used radionuclides - use a long lived reference source ( 137 Cs or 133 Ba or 226 Ra a certified source may be used but knowledge of activity is not required) - measure apparent activity on all the settings of radionuclides to be used on that day - recorded the results and compare measured and expected values 26

QUARTERLY CHECKS LINEARITY OF RESPONSE AS A FUNCTION OF ACTIVITY to reveal region and magnitude of non linearities due to saturation effects (ion-recombination); nonlinearity near to zero activity (may indicate a wrong preset zero adjustement); discontonuities at change of range (systematic error in one of the ranges) DECAY METHOD - prepare a source of a short-lived radionuclide (such as 99m Tc); initial activity of the source equal to or greater than the highest activity likely to be measured - follow the activity of the source as a function of time; decay to the lowest levels of activity that would normally be measured (or the activity range where the instrument is properly calibrated) - apply decay correction to earlier measurements 27

- determine the deviation of the decay corrected measurements from the expected values (extrapolation from data points in the lower activity region, assumed to be correct) - test material free of long-lived radionuclidic impurities ( 99 Mo in 99m Tc ; 113 Sn in 113m In) to prevent distortions of the half life curve (initially the activity of the contamination in the sample will be insignificant, but it causes an apparent levelling of the activity in the final part of the graph); impurities can be detected as residual activity 28

PROPORTIONAL METHOD - use identical sources with different radioactivity concentrations of the same radionuclide (moderate half-life, 131 I) - if measurements are made over an interval of time short compared with the half lives of any photon emitting impurities, the method is insensitive to these impurities - possible artefacts introduced by inaccurate preparation of radioniclide solutions 29

SLEEVE METHOD - insert calibrated lead shields into the calibrator well to produce an attenuation of photon flux to simulate the decay occurring from 0 through 50 hours (first eluition from a new generator) - this not a true measurement of linearity under identical geometrical conditions, can be used as supplementary consistency check 30

IAEA-TECDOC-602 31

ACCURACY : initial accuracy is limited by the certified sources used by the factory for calibration (uncertainty 5 % or less); it may change with time as a result of changing pressure of the chamber gas and slow electrical drift - use radionuclide standards spanning a wide range of energy ( 57 Co, 60 Co, 137 Cs, 133 Ba) and determine the present activity of the source for the proper calibration setting - repeate the measurement to a total of 10 times - calculate for each source the percentage difference between the mean measured activity, A m, and the certified activity corrected for radioactive decay [(A m C) / C] * 100 32

PRECISION - calculate for each source the percentage differences between the individual measured activities A i and their mean A m [(A i A m ) / A m ] * 100 33

ANNUALY CHECKS GEOMETRY (SAMPLE VOLUME AND POSITION, CONTAINER) VOLUME: gradually increase the volume of 1 ml Tc solution in a 10 ml syringe by adding known amounts of water (to 2, 3, 5 and 10 ml) POSITION: vary the position of the syringe in two axes (vertical and horizontal displacement) CONTAINER: test for various types of syringes (comparison between measured activity and real activity real activity being the difference of the vial readings before and after filling the syringe) 34

REFERENCES 1. National Council on Radiation Protection and Measurement : A Handbook of Radioactivity Measurements Procedures. NCRP Report No. 58 : 291-299, 1985. 2. National Council on Radiation Protection and Measurement : Quality Assurance for Diagnostic Imaging. NCRP Report No. 99 : 129-132, 1988. 3. National Council on Radiation Protection and Measurement : Protection of the Patient in Nuclear Medicine. NCRP Report No. 52 : 33 1987. 4. National Council on Radiation Protection and Measurement : Radiation Dose to Patients from Radiopharmaceuticals. NCRP Report No. 53 : 11-12 1987. 5. Council Directive 96/29/Euratom of 13 May 1996 : Directive for the protection of the health of workers and the general public against the dangers arising from ionizing radiation. Basic Safety Standards Directive (BSS) 6. Council Directive 97/43/Euratom of 30 June 1997: Health protection of individuals against the dangers of ionizing radiation in relation to medical exposure. Medical Exposure Directive (MED) 7. European Commission - Radiation Protection 91: Criteria for Acceptability of Radiological (including Radiotherapy) and Nuclear Medicine Installations. 8. HPA CRS 38 Quality Control on Nuclear Medicine Instrumentation, 1983 35

9. ANSI N42.13 1986 : Calibration and Usage of Dose Calibrators Ionization Chambers for the Assay of Radionuclides, 1986 10. IAEA TECDOC - 602 : Quality Control of Nuclear Medicine Instruments, 1991 11. Calhoun J.M., Golas D.B., Harris S.G.: Effect of Varying Geometry on Dose Calibrator Response: Cobalt-57 and Technetium-99m. J Nucl Med 28 : 1478-1483, 1987. 12. Davis D.A., Gioumuso C.A., Miller W.H., Ferrell J.R., Early P.J., Gargaro V.A., Bloe F.T., Close D.W., Began E.J.: Dose Calibrator Activity Linearity Evaluations With ALARA Exposures. J Nucl Med Tech 9(4) : 188-190, 1981. 13. Hare D.L., Hendee W.R., Whitney W.P., Chaney E.L.: Accuracy of Well Ionization Chamber Calibrators. J Nucl Med 15(12) : 1138-1141, 1974. 14. Harris C.C., Jaszczak R.J., Greer K.L., Briner W.H., Coleman R.E.: Effects of Characteristic X-rays on Assay of I-123 by dose calibrator. J Nucl Med 25 : 1367-1370, 1984. 15. Kowalsky R.J., Johnston R.E., Chan F.H.: Dose Calibrator Performance and Quality Control. J Nucl Med Tech 5(1) : 35-40, 1977. 16. Suzuki A., Suzuki M., Weis A.M.: Analysis of a Radioisotope Calibrator. J Nucl Med Tech 4(4) : 193-198, 1976. 36