FT-IR-iPAS A tool for industrial multicomponent gas measurement Christian Hirschmann 1,2 *, Jussi Tenhunnen 2, Satu Ojala 1,2 and Riitta Keiski 1

Similar documents
Cantilever enhanced tunable diode laser photoacoustic spectroscopy in gas purity measurement case study: acetylene in ethylene

University of Oulu, Dept. Process and Environmental Engineering, FI University of Oulu, P.O.Box 4300

Fourier Transform Infrared. Spectrometry

Compact Hydrogen Peroxide Sensor for Sterilization Cycle Monitoring

Introduction to Fourier Transform Infrared Spectroscopy

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

Fourier Transform Infrared Spectrometry

Introduction to Fourier Transform Infrared Spectroscopy

Training Coarse ECOPROBE 5 CHEMISTRY RS DYNAMICS

High Sensitivity Gas Sensor Based on IR Spectroscopy Technology and Application

have also been successfully tested in low temperature NH 3 Noble metals, especially platinum, have been reported to be active catalysts in NH 3

Fourier Transform IR Spectroscopy

PAPER No. 12: ORGANIC SPECTROSCOPY MODULE No. 7: Instrumentation for IR spectroscopy

supports Lenka Matejova 1, Satu Pitkäaho 2 *, Satu Ojala 2, Tomas Cajthaml 3, Riitta L. Keiski 2 and Olga Solcova 1

DIODE- AND DIFFERENCE-FREQUENCY LASER STUDIES OF ATMOSPHERIC MOLECULES IN THE NEAR- AND MID-INFRARED: H2O, NH3, and NO2

Advanced Spectroscopy Laboratory

FOURIER TRANSFORM INFRARED SPECTROSCOPY (FTIR)

The Claviature of Gas Analysis

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Real-time ppb CO 2 Impurity Detection by an Advanced FTIR- UVF System

Application of IR Raman Spectroscopy

Virtual Bioimaging Laboratory

requency generation spectroscopy Rahul N

DEVELOPMENT OF MEASUREMENT AND

ISA QATAR : 90 th Seminar

New advances in folded pathlength technology for Process Tunable Diode Laser Absorption Spectrometers (TDLAS)

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

Spectroscopy tools for PAT applications in the Pharmaceutical Industry

Physics of capnography

Nanoscale IR spectroscopy of organic contaminants


DOWNLOAD OR READ : INFRARED AND RAMAN SPECTROSCOPY CONCEPTS AND APPLICATIONS PDF EBOOK EPUB MOBI

FT-IR Spectroscopy. An introduction in measurement techniques and interpretation

UV Quality Assurance. Dark Current Correction

Abstract... I. Acknowledgements... III. Table of Content... V. List of Tables... VIII. List of Figures... IX

Optical Properties of Thin Semiconductor Films

Chemistry Instrumental Analysis Lecture 15. Chem 4631

Importance of in situ Monitoring in MOCVD Process and Future Prospects

Modern Techniques in Applied Molecular Spectroscopy

Quantum Dots for Advanced Research and Devices

Thermal And Near infrared Sensor for carbon Observation (TANSO) On board the Greenhouse gases Observing SATellite (GOSAT) Research Announcement

REX Evaluation Guide. American Micro Detection Systems Inc March Lane, Suite 200 Stockton, CA 95219

CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

Application of Raman Spectroscopy for Noninvasive Detection of Target Compounds. Kyung-Min Lee

Monitoring the Aggregation of Particles through Raman Spectroscopy

Fourier transform infrared spectroscopy (FTIR) is a method used to obtain an infrared

PERFORMANCE AND EXAMPLES OF MEASUREMENTS OF A MID INFRARED INTERFEROMETRIC HYPERSPECTRAL IMAGER

FTIR Spectrometer. Basic Theory of Infrared Spectrometer. FTIR Spectrometer. FTIR Accessories

Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR

Mid-Infrared Laser based Trace Gas Sensor Technologies: Recent advances and Applications

Application of near-infrared (NIR) spectroscopy to sensor based sorting of an epithermal Au-Ag ore

RS DYNAMICS ECOPROBE 5. Portable IR/PID Gas Analyzer PID. PID and IR Analyzers

Abstract. Introduction

Chem Homework Set Answers


INTERNATIONAL JOURNAL OF PURE AND APPLIED RESEARCH IN ENGINEERING AND TECHNOLOGY

Answers to questions on exam in laser-based combustion diagnostics on March 10, 2006

Monomer Analysis. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Characterization and Calibration of a Fourier Transform Spectroradiometer for Solar UV Irradiance Measurements

UV3000. Accurate, precise, and portable ambient gas point analyzer

IR Spectrography - Absorption. Raman Spectrography - Scattering. n 0 n M - Raman n 0 - Rayleigh

Standoff Spectroscopy of Surface Adsorbed Chemicals

Modeling the Response of Photoacoustic Gas Sensors

Characterisation of vibrational modes of adsorbed species

Unit title: Atomic and Nuclear Physics for Spectroscopic Applications

Glossary. Analyte - the molecule of interest when performing a quantitative analysis.

arxiv: v1 [physics.optics] 10 Jan 2019

Introduction to FT-IR Spectroscopy

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

What happens when light falls on a material? Transmission Reflection Absorption Luminescence. Elastic Scattering Inelastic Scattering

Spectroscopy in Inorganic Chemistry. Vibration and Rotation Spectroscopy

Spectroscopy Problem Set February 22, 2018

UV Hound Series. Dependable High Quality Air Monitoring. Accurate Readings Within Seconds. Portable UVDOAS Multi-gas Analyzers. Multi-Gas Capability

Optics, Light and Lasers

Vibrational Spectroscopy of Molecules on Surfaces

Supplementary Information. Highly sensitive gas-phase explosive detection by luminescent microporous polymer networks

Session #1: Theoretical background and computer simulations of molecular vibrations.

MASS and INFRA RED SPECTROSCOPY

WLP. Si PMMA. Norm.Intensity 0.1 FID. [a.u.] Apodization Mirror position d [µm] c) d) E inc E sca. Nano-FTIR phase ϕ [º] PMMA

Introduction to Gas Phase FTIR Spectroscopy

TECHNICAL DESCRIPTION SPECTRAFLOW ON LINE ANALYZER for BELT CONVEYOR APPLICATION

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES

Advanced Pharmaceutical Analysis

Spectroscopy in Transmission

Supplementary Materials

Spectroscopic Applications of Quantum Cascade Lasers

Digital processing of multi-mode nano-mechanical cantilever data

Cesium Dynamics and H - Density in the Extended Boundary Layer of Negative Hydrogen Ion Sources for Fusion

1901 Application of Spectrophotometry

Atmospheric Extinction

Probing and Driving Molecular Dynamics with Femtosecond Pulses

Astronomical frequency comb for calibration of low and medium resolution spectrographs

2. Infrared spectroscopy

PTG-NIR Powder Characterisation System (not yet released for selling to end users)

Visible and IR Absorption Spectroscopy. Andrew Rouff and Kyle Chau

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions

Ultraviolet-Visible and Infrared Spectrophotometry

CONFOCHECK. Innovation with Integrity. Infrared Protein Analysis FT-IR

Ultraviolet-Visible and Infrared Spectrophotometry

A TES Bolometer for THz FT-Spectroscopy

Transcription:

108 FT-IR-iPAS A tool for industrial multicomponent gas measurement Christian Hirschmann 1,2 *, Jussi Tenhunnen 2, Satu Ojala 1,2 and Riitta Keiski 1 1 University of Oulu, Mass and Heat Transfer Process Laboratory Department of Process and Environmental Engineering, FI-90014 University of Oulu, P.O.Box 4300 2 VTT, Photonics and Measurement Solutions, Kaitoväylä 1 FI-90570 Oulu, P.O.Box 1100 1 Introduction Modern adsorption IR spectrometer and their optical detectors are developed close to the theoretical performance limits. [Kauppinen et. al., 2004; Koskinen et. al., 2007] Thus, increasing the optical path length is the sole option to enhance the sensitivity of the IR absorption technique. [Kauppinen et. al., 2004; Koskinen et. al., 2007; Koskinen et. al., 2006] By enlarging the optical path length, the absorption cell becomes bulky, the sample volume increases and the light absorption is nonlinear. [Koskinen et. al., 2008] The nonlinear behaviour is especially pronounced by measuring wet gases, which is quite often the case in industry, since both the analyte and water respond nonlinear. [Kuusela et. al., 2007; Koskinen et. al., 2008] In sum, a significant improvement in sensitivity of the absorption technique is not anymore possible. To gain a higher sensitivity, other techniques have to be found, such as the photoacoustic method, which offers a zero background signal. [Kauppinen et. al., 2004; Koskinen et. al., 2007; Kuusela et. al., 2007] However, the common photoacoustic method using a capacitive membrane microphone also suffers from restrictions, like the nonlinear pressure response, which is due to the three dimensional membrane stretch and the poor sensitivity, due to the damping effect caused by the air flow between the rigid electrode and the flexible membrane. [Kauppinen et. al., 2004; Kuusela et. al., 2007; Fonsen et. al., 2009] In the recent past, Kauppinen from the University of Turku invented the optical readout cantilever microphone. The optical readout cantilever microphone is not subjected to the limitations of the condenser microphone, allowing improvements to the sensitivity and the dynamic range. [Kauppinen et. al., 2004; Kuusela et. al., 2007; Sievila et. al., 2007] The sensitivity of photoacoustics can be 100 times better when using the optical cantilever detection (abbr. ipas: improved photoacoustic spectroscopy) instead of the conventional capacitive microphone. [Koskinen et. al., 2007; Lindley et. al., 2007; Olafsson et. al., 1999] So far, photoacoustic gas spectroscopy using optical cantilever detection has only been set up with lasers, diode lasers and blackbody radiators with filters. However, lasers and blackbody radiator-filter combinations are not optimal for measuring multicomponent samples. This is the driving force to combine ipas with a highly selective, but broadband detecting technique, like a FT-IR with a black body radiator. This novel combination: an FT-IR-iPAS prototype was realized by GASERA, Finland in 2008. In 2009 the performance of the prototype was extensively tested at University of Oulu and VTT Oulu [Hirschmann et. al., 2010]. Some of those performance tests will be presented here. 2 Objectives of the research The objective of this research is to develop and establish FT-IR-iPAS as a new technique for accurate, multicomponent gas measurements in industry. FT-IR-iPAS brings several benefits, compared to the present absorption technique, which makes it superior for industrial gas measurements. So far the performance of the first built prototype was tested and compared with the absorption technique. After the positive *Correspoding author, E-mail: christian.hirschmann@oulu.fi

ENERGY RESEARCH at the University of Oulu 109 verification, now the focus is to develop measurement systems for various applications, like the industrial measurement of VOC. This development will be done in the EU-SME-Project ZERO-VOC. The target of this project is to develop and built a measurement system for the industrial measurement of VOCs. 3 Results The main investigated performance criteria are the sensitivity, selectivity, multicomponent ability and the linearity. These performance tests are reported and discussed in the following. 3.1 Selectivity & Multicomponent ability The selectivity was studies on several gases. The spectra of the individual gases were measured consecutively. In the software, the spectra were put on top of each other to one resulting spectrum. Figure 1 shows the spectrum of each gas and the resulting spectrum. The raw spectra are background corrected and the water bands are subtracted. Figure 1 Single spectra and the resulting combined methane (1000 ppm), carbon dioxide (5000 ppm), propene (1000 ppm), and methylmercaptane (2000 ppm) spectrum. In some spectral regions, like from 2900 until 3150 cm -1 the spectra are overlapping, which makes a quantitative determination highly difficult. However also spectral regions can be found, where the single spectra are not overlapping each other. Using multivariate calibration methods, like Partial Least Square Regression (PLS) or Science Based Calibration (SBC), it won t be a problem to quantitatively calibrate the four components. Even, with the classical one wavelength calibration method it would be possible to set up a quantitative calibration in the spectral non-overlapping areas. 3.2 Sensitivity The sensitivity is a meaningful attribute characterising the performance of spectrometers. One parameter to express the sensitivity is the signal to noise ratio (SNR). The SNR relates the gained signal to the unwanted noise in the spectrum. Table 1 shows the analysis of the sensitivity.

110 Table 1 Sensitivity analysis. methane carbon dioxide gas concentration [ppm] 1000 5000 molecular bond stretch C-H asymmetrical stretch C=O measurement time [seconds] 100 100 resolution [cm -1 ] 4 4 spectral position of the signal [cm -1 ] 3017 2356 spectral position of the noise [cm -1 ] 3200-3250 3200-3250 PA signal 1.3 0.87 noise 6.42 10-4 6.42 10-4 SNR 2027 1362 LoD [ppm] 0.5 4 First theoretical evaluations were assuming that that the sensitivity of the FT-IR-iPAS is not as good as with blackbody radiator-filter and laser devices, because of the limited light throughput. This assumption is confirmed here, because in the literature LoDs for measuring methane are 0.5ppm (measurement time: 5 sec) [Fonsen et. al., 2009], 0.8ppb (measurement time: 100 sec) [Kauppinen et. al., 2004] and 13ppb (measurement time: 0.37 sec) [Uotila et. al., 2005]. However, the FT-IR-PAS still has the advantage of the multicompont ability. This means that one measurement is not only to measure one species, like the case with filter devices. Several species can be quantitative determined by only one measurement, as it is done here: one measurement was needed for both species methane and carbon dioxide. If this feature is also taken into account, and the devices are not only compared by the measurement time, instead by the amount of measured components per time, then FT- IR-iPAS will be the favourite. The appropriate area for application will the multicomponent measurements, when more than one gas should be measured, or only one gas which is mixed with gases interfering filter or laser based measurements. These applications can be found either in laboratory or in industry. 3.3 Results 3: Linearity To discover the behaviour of the response, several concentrations of methane and carbon dioxide are measured. The absorption is averaged in the ranges between 3010 and 3020 cm -1 for methane (stretch C-H) and 2300 until 2380 cm -1 for carbon dioxide (asymmetric stretch C=O). The averaged absorption is plotted versus the concentration and displayed in Figure 2 for methane and Figure 3 for carbon dioxide.

ENERGY RESEARCH at the University of Oulu 111 Figure 2 Linearity analysis of the stretch C-H band of methane in the blend with carbon dioxide; top: Averaged band height on the concentration with the fitted lines; bottom: Residuals of the fit. Figure 3 Linearity analysis of the asymmetric stretch C=O band of carbon dioxide in the blend with methane; top: Averaged band height on the concentration with the fitted lines; bottom: Residuals of the fit. According to the graphics, the response can be said to be linear in the examined range. There are some deviations, but those can be explained by the practical gas mixing set-up. The linear signal response is a valuable gain for industrial applications. Only few, in the best case one gas spectrum is needed to set up a complete quantitative calibration. This allows quick calibration and easy adoption to different measurement problems. 4 Relevance of the research The presented research is of high importance. FT-IR-iPAS opens the door to 1. Improve the sensitivity, selectivity and multicomponent ability; 2. Allow a linear response which is valuable for easy and fast calibration; 3. Perform an easy water subtraction, with no need for comprehensive libraries; 4. Be deployed in different areas, like laboratory, industry and environment;

112 5. Decrease the overall set up work and time for a new measurement application, due to the linear response, quick calibration and easy adoption to different measurement problems and places; 6. Make measurements at all possible in some areas, such as the industrial measurement of VOCs, where all current available techniques have clearly failed. One indicator of the high importance of the research is the promotion of the research by the EU in the EU-SME-project ZERO-VOC in which FT-IR-iPAS will be developed for industrial measurement of VOCs. 5 References Fonsen, J., Koskinen, V., Roth, K. and Kauppinen, J., (2009). Dual cantilever enhanced photoacoustic detector with pulsed broadband IR-source. Journal of Vibrational Spectroscopy, 50 (2): 214-217. Hirschmann, C. B., Uotila, J., Ojala, S., Tenhunnen, J. and Keiski, R., (2010). Fourier Transform Infrared Photoacoustic Multicomponent Gas Spectroscopy with Optical Cantilever Detection. Journal of Applied Spectroscopy, 64: 293-297. Kauppinen, J., Wilcken, K., Kauppinen, I. and Koskinen, V., (2004). High sensitivity in gas analyses with photoacoustic detection. Micromechanical Journal, 76: 151-159. Koskinen, V., Fonsen, J., Kauppinen, J. and Kauppinen, I., (2006). Extremely sensitive trace gas analysis with modern photoacoustic spectroscopy. Journal of Vibrational Spectroscopy, 42: 239-242. Koskinen, V., Fonsen, J., Roth, K. and Kauppinen, J., (2007). Cantilever enhanced photoacoustic detection of carbon dioxide using a tunable laser diode source. Journal of Applied Physics B, 86 (3): 451-454. Koskinen, V., Fonsen, J., Roth, K., Kauppinen, J., (2008). Progress in cantilever enhanced photoacoustic spectroscopy. Journal of Vibrational Spectroscopy, 48: 16-21. Kuusela, T. and Kauppinen, J., (2007). Photoacoustic gas analysis using interferometric cantilever microphone. Journal of Applied Spectroscopy Reviews, 42: 443-474. Lindley, R. E., Parkes, A. M., Keen, K. A., McNaghten, E. D. and Orr-Ewing, A. J., (2007). A sensitivity comparison of three photoacoustic cells containing a single microphone, a differential dual microphone or a cantilever pressure sensor. Journal of Applied Physics B, 86 (4): 707-713. Olafsson, A., Hansen, G. I., Loftsdottir, A. S. and Jakobsson S., (1999) FTIR photoacoustic trace gas detection. Photoacoustic and Photothermal Phenomena: 10th Interntional Conference, Rome, Italy, 23-27 August 1998, 208. Sievila, P., Fonsen, J., Hahtela, O., Chekurov, N., Kauppinen, J. and Tittonen, I., (2007). Optically detected, framed silicon cantilever for high precision acoustic sensing. Solid-State Sensors, Actuators and Microsystems Conference, 10-14 June 2007, Lyon, 2605-2608. Uotila, J., Koskinen, V. and Kauppinen, J., (2005). Selective differential photoacoustic method for trace gas analysis. Journal of Vibrational Spectroscopy, 38: 3-9.

Reference to this article: Hirschmann C., Tenhunnen J., Ojala S. and Keiski R. L. (2010) FT-IR-iPAS a tool for industrial multicomponent gas measurement. In: Pongrácz E., Hyvärinen M., Pitkäaho S. and Keiski R. L. (eds.) Clean air research at the University of Oulu. Proceeding of the SkyPro conference, June 3rd, 2010, University of Oulu, Finland. Kalevaprint, Oulu, ISBN 978-951-42-6199-2. pp.108-112. SkyPro conference: http://nortech.oulu.fi/skypro/index.html Proceedings: http://nortech.oulu.fi/skypro/skyproproc.html