METHOD #: Approved for NPDES and SDWA (Ed. Rev. 1974, 1978) Fluoride, Total (Colorimetric, SPADNS with Bellack Distillation)

Similar documents
METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

Standard Test Methods for Fluoride Ion in Water 1

METHOD #: Approved for NPDES (Technical Revision 1980) Cyanide, Total (Titrimetric; Spectrophotometric)

ALLOWAY. Method Outline. Standard Laboratory Method: Total Fluoride via Distillation. Distillation then Specific Ion. Date Issued: 01/04/82

Sodium Chloride - Analytical Standard

ALLOWAY METHOD OUTLINE

RAPID KJELDAHL BENCHNOTES

METHOD 9035 SULFATE (COLORIMETRIC, AUTOMATED, CHLORANILATE)

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE)

Standard Methods for the Examination of Water and Wastewater

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

METHOD 9012 TOTAL AND AMENABLE CYANIDE (COLORIMETRIC, AUTOMATED UV)

METHOD 9012B TOTAL AND AMENABLE CYANIDE (AUTOMATED COLORIMETRIC, WITH OFF-LINE DISTILLATION)

Scope and application: For wastewater, seawater, drinking water, surface water and process water.

CH2MHILL GILROY/MORGAN HILL LABORATORY LABORATORY STANDARD OPERATING PROCEDURE NO TOTAL KJELDAHL NITROGEN - TKN SM 4500-N org.

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cyanide, colorimetric, pyridine-pyrazolone

Standard Methods for the Examination of Water and Wastewater

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Method for estimation of iodine in urine

Chlorine, Free and Total, High Range

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Organic Carbon, Total (Low Level) (UV Promoted, Persulfate Oxidation) CAS # Total Organic Carbon TOC C

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Nitrogen, ammonia, colorimetry, salicylate-hypochlorite, automated-segmented flow

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Method to 0.50 mg/l NH 3 N Powder Pillows

SIDE BY SIDE COMPARISON TABLE FOR HACH METHOD NUMBER COLORIMETRIC DETERMINATION OF CYANIDE

METHOD 9010B TOTAL AND AMENABLE CYANIDE: DISTILLATION

DPD Test N Tube Method *

Method to 0.50 mg/l NH 3 N Powder Pillows

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Oxygen Demand, Chemical

CHLORINE, TOTAL (0 to 4.00 mg/l)

Chlorine, Free and Total, Low Range

CHLORINE, FREE, Ultra-high Range ( mg/l Cl 2 ) Method 10069

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

Chlorine, Free and Total, High Range

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

DR/4000 QUATERNARY AMMONIUM COMPOUNDS PROCEDURE. 2. The display will show: 3. Fill a sample cell (the blank) with 25 ml of deionized water.

Method (0.04 to 4.50 mg/l Cl 2 ) Powder Pillows

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

NITROGEN, TOTAL, Test N Tube * Method 10071

NITROGEN, TOTAL, HR, Test N Tube

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

Oxygen Demand, Chemical

Spectrophotometric Determination of Ferrocyanide in Effluents

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

DR/4000 PROCEDURE SELENIUM. 4. Measure 100 ml of sample into a second 500-mL erlenmeyer flask (label the flask sample ).

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

Chloramine (Mono) and Nitrogen, Free Ammonia

METHOD DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY

LEAD (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 500 C

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

PREPARATION FOR CHEMISTRY LAB: FLUORIDE IN WATER

SYNTHESIS OF 1-BROMOBUTANE Experimental procedure at macroscale (adapted from Williamson, Minard & Masters 1 )

Nitrogen, Total Inorganic

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

DETERMINATION OF CHEMICAL OXYGEN DEMAND COD

Quaternary Ammonium Compounds

DR/4000 PROCEDURE NITRATE. 2. The display will show:

SOUTH AFRICAN NATIONAL STANDARD

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15

Annex 2 Formaldehyde

DR/4000 PROCEDURE MERCURY. Phase 1: Sample Digestion must be done in a fume hood! Toxic gases may be produced.

Determination of Nutrients. Determination of total phosphorus. Extraktion with aqua regia, reflux method. Introduction

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

CALCIUM (EGTA-Titrimetric)

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

EXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy

Sodium Chloride - Analytical Standard

METHOD DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method

Experiment 5: Analysis of Nutrients in Natural Waters CH3600 / ESP 5090: Environmental Chemistry, Plymouth State University

DR/4000 PROCEDURE CYANIDE. 1. Press the soft key under HACH PROGRAM. Select the stored program for cyanide by pressing 1750 with the numeric keys.

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Experiment 1 (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with 1,10- Phenanthroline

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Experimental Procedure Lab 402

METHOD 13B - DETERMINATION OF TOTAL FLUORIDE EMISSIONS FROM STATIONARY SOURCES (SPECIFIC ION ELECTRODE METHOD)

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

LOW-RANGE TOTAL CHLORINE

ENCM Environmental Impact Assessment and Environmental Monitoring. Determination of Chemical Oxygen Demand in water/waste water

LACTIC ACID. The method is applicable to the determination of lactic acid and lactate salts (Note 2) in light or heavy steepwater.

SECTION D.2 AMMONIA NITROGEN

Exercise 6: Determination of Hardness of Water

Transcription:

METHOD #: 340.1 Approved for NPDES and SDWA (Ed. Rev. 1974, 1978) TITLE: Fluoride, Total (Colorimetric, SPADNS with Bellack Distillation) ANALYTE: CAS # F Fluoride 7782-41-4 INSTRUMENTATION: Spectrophotometer STORET No. Total 00951 Dissolved 00950 1.0 Scope and Application 1.1 This method is applicable to the measurement of fluoride in drinking, surface, and saline waters, domestic and industrial wastes. 1.2 The method covers the range from 0.1 to about 1.4 mg/l F. This range may be extended to 1000 mg/l using the Fluoride Ion Selective Electrode Method (340.2) after distillation. 2.0 Summary of Method 2.1 Following distillation to remove interferences, the sample is treated with the SPADNS reagent. The loss of color resulting from the reaction of fluoride with the zirconyl- SPADNS dye is a function of the fluoride concentration. 3.0 Comments 3.1 The SPADNS reagent is more tolerant of interfering materials than other accepted fluoride reagents. Reference to Table 414:1, p 388, Standard Methods for the Examination of Waters and Wastewaters, 14th Edition, will help the analyst decide if distillation is required. The addition of the highly colored SPADNS reagent must be done with utmost accuracy because the fluoride concentration is measured as a difference of absorbance in the blank and the sample. A small error in reagent addition is the most prominent source of error in this test. 3.2 Care must be taken to avoid overheating the flask above the level of the solution. This is done by maintaining an even flame entirely under the boiling flask. 4.0 Apparatus 4.1 Distillation apparatus: A 1-liter round-bottom, long-necked pyrex boiling flask, connecting tube, efficient condenser, thermometer adapter and thermometer reading to 200 C. All connections should be ground glass. Any apparatus equivalent to that shown in Figure 1 is acceptable. 4.2 Colorimeter: One of the following

5.0 Reagents 4.2.1 Spectrophotometer for use at 570 nm providing a light path of at least 1 cm. 4.2.2 Filter photometer equipped with a greenish yellow filter having maximum transmittance at 550 to 580 nm and a light path of at least 1 cm. 5.1 Sulfuric acid, H SO, conc. 5.2 Silver sulfate, Ag SO crystals. 5.3 Stock fluoride solution: Dissolve 0.221 g anhydrous sodium fluoride, NaF, in distilled water in a l-liter volumetric flask and dilute to the mark with distilled water; 1.00 ml = 0.1 mg F. 5.4 Standard fluoride solution: Place 100 ml stock fluoride solution (5.3) in a 1 liter volumetric flask and dilute to the mark with distilled water; 1.00 ml = 0.010 mg F. 5.5 SPADNS solution: Dissolve 0.958 g SPADNS, sodium 2-(parasulfophenylazo)-1,8- dihydroxy-3,6-naphthalene disulfonate, in distilled water in a 500 ml volumetric flask and dilute to the mark. Stable indefinitely if protected from direct sunlight. 5.6 Zirconyl-acid reagent: Dissolve 0.133 g zirconyl chloride octahydrate, ZrOCl C8H O in approximately 25 ml distilled water in a 500 ml volumetric 2 2 flask. Add 350 ml conc HCl and dilute to the mark with distilled water. 5.7 Acid-zirconyl-SPADNS reagent: Mix equal volumes of SPADNS solution (5.5) and zirconyl-acid reagent (5.6). The combined reagent is stable for at least 2 years. 5.8 Reference solution: Add 10 ml SPADNS solution (5.5) to 100 ml distilled water. Dilute 7 ml conc HCl to 10 ml and add to the dilute SPADNS solution. This solution is used for zeroing the spectrophotometer or photometer. It is stable and may be used indefinitely. 5.9 Sodium arsenite solution: Dissolve 5.0 g NaAsO in distilled water in a 1-liter 2 volumetric flask and dilute to the mark with distilled water (CAUTION: Toxic-avoid ingestion). 6.0 Procedure 6.1 Preliminary distillation 6.1.1 Place 400 ml distilled water in the distilling flask. 6.1.2 Carefully add 200 ml conc. H2SO4 and swirl until contents are homogeneous. 6.1.3 Add 25 to 35 glass beads, connect the apparatus (Figure 1 making sure all joints are tight. 6.1.4 Heat slowly at first, then as rapidly as the efficiency of the condenser will permit (distillate must be cool) until the temperature of the flask contents reaches exactly 180 C. Discard the distillate. This process removes fluoride contamination and adjusts the acid-water ratio for subsequent distillations. 6.1.5 Cool to 120 C or below.

6.1.6 Add 300 ml sample, mix thoroughly, distill as in 6.1.4 until temperature reaches 180 C. Do not heat above 180 C to prevent sulfate carryover. 6.1.7 Add Ag SO (5.2) at a rate of 5 mg/mg Cl when high chloride samples are distilled. 6.1.8 Use the sulfuric acid solution in the flask repeatedly until the contaminants from the samples accumulate to such an extent that recovery is affected or interferences appear in the distillate. Check periodically by distilling standard fluoride samples. 6.1.9 High fluoride samples may require that the still be flushed by using distilled water and combining distillates. 6.2 Colorimetric Determination 6.2.1 Prepare fluoride standards in the range 0 to 1.40 mg/l by diluting appropriate quantities of standard fluoride solution (5.4) to 50 ml with distilled water. 6.2.2 Pipet 5.00 ml each of SPADNS solution (5.5) and zirconyl-acid reagent (5.6) or 10.00 ml of the mixed acid-zirconyl-spadns reagent (5.7) to each standard and mix well. 6.2.3 Set photometer to zero with reference solution (5.8) and immediately obtain absorbance readings of standards. 6.2.4 Plot absorbance versus concentration. Prepare a new standard curve whenever fresh reagent is made. 6.2.5 If residual chlorine is present pretreat the sample with 1 drop (0.05 ml) 7.0 Calculations NaAsO, solution (5.9) per 0.1 mg residual chlorine mix. Sodium 2 arsenite concentrations of 1300 mg/l produce an of 0.1 mg/l at 1.0 mg/l F. 6.2.6 Use a 50 ml sample or a portion diluted to 50 ml. Adjust the temperature of the sample to that used for the standard curve. 6.2.7 Perform step 6.2.2 and 6.2.3. 7.1 Read the concentration in the 50 ml sample using the standard curve (6.2.4) 7.2 Calculate as follows: 7.3 When a sample (ml sample) is diluted to a volume (B) and then a portion (C) is analyzed, use: 8.0 Precision and Accuracy

8.1 On a sample containing 0.83 mg/l F with no interferences, 53 analysts using the Bellack distillation and the SPADNS reagent obtained a mean of 0.81 mg/l F with a standard deviation of ±0.089 mg/l. 8.2 On a sample containing 0.57 mg/l F (with 200 mg/l SO and 10 mg/l Al as 4 interferences) 53 analysts using the Bellack distillation obtained a mean of 0.60 mg/l F with a standard deviation of ±0.103 mg/l. 8.3 On a sample containing 0.68 mg/l F (with 200 mg/l SO, 2 mg/l Al and 2.5 4 mg/l [Na(PO ) ] as interferences), 53 analysts using the Bellack distillation 3 6 obtained a mean of 0.72 mg/l F with a standard deviation of ±0.092 mg/l. (Analytical Reference Service, Sample 11 I-B water, Fluoride, August, 1961.) Bibliography 1. Standard Methods for the Examination of Water and Wastewater, p. 389-390 (Method No.414A, Preliminary Distillation Step) and p. 393-394 (Method 414C SPADNS) 14th Edition, (1975). 2. Annual Book of ASTM Standards, Part 31, "Water", Standard D 1179-72, Method A, p. 310 (1976).