The Waters Acquity Ultra-Performance Liquid Chromatograph and the Micromass Quattro Premier Triple Quadrupole Mass Spectrometer

Similar documents
Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes

Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes Application

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Determination of Pharmaceuticals in Environmental Samples

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

TANDEM MASS SPECTROSCOPY

Routine MS Detection for USP Chromatographic Methods

Mass Spectrometry in MCAL

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

Analysis of Polar Metabolites using Mass Spectrometry

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

A S ENSIT IV E M E T HO D FO R T H E D E T E RM INAT IO N O F ENDO C RIN E- D IS RU P T ING COM P OUNDS IN RIV E R WAT E R BY L C / MS/MS

Finnigan LCQ Advantage MAX

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

Agilent Standard Operating Procedure

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

QTOF Training Manual

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

HPLC Praktikum Skript

Introduction to the Q Trap LC/MS/MS System

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Thermo Finnigan LTQ. Specifications

Agilent G3212 GC-APCI Source

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

M M e M M H M M H. Ion Sources

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

MONITORING FOR PHARMACEUTICALS IN SURFACE WATER USING DIRECT AQUEOUS INJECTION ON THE AGILENT 6490 LC/QQQ

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Supporting Information

Utilizing ELSD and MS as Secondary Detectors for Prep HPLC and Flash Chromatography. Tips and Techniques to Optimize ELSD and MS based Purification

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

EPA Method 1694: Analysis of Pharmaceuticals and Personal Care Products in Water

Quattro Micro - How does it work?

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Introduction to LC-MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

LC-MS Based Metabolomics

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

Biochemistry. Biochemical Techniques HPLC

The Effect of Contact Angle and Wetting on Performance in Solid Phase Extraction

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Translational Biomarker Core

Ultrasensitive EPA Method 1694 with the Agilent 6460 LC/MS/MS with Jet Stream Technology for Pharmaceuticals and Personal Care Products in Water

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Lecture 15: Introduction to mass spectrometry-i

LC MS analysis of metabolites. Basis of Chromatography

Extraction of Cocaine and Metabolites From Urine Using ISOLUTE SLE+ prior to LC-MS/MS Analysis

UPLC Intact MASS Analysis Application Kit

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Electrospray Ion Trap Mass Spectrometry. Introduction

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Chapter content. Reference

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Choosing the metabolomics platform

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

Profiling of Diferulates (Plant Cell Wall Cross- Linkers) Using Ultrahigh-performance Liquid. Chromatography-Tandem Mass Spectrometry

LIQUID CHROMATOGRAPHY-MASS SPECTROMETRY (LC/MS) Presented by: Dr. T. Nageswara Rao M.Pharm PhD KTPC

High Performance Liquid Chromatography

High Performance Liquid Chromatography

Selecting Detectors for Compounds with No Optical Absorbance

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Introduction of Hitachi Chromaster 5610 MS Detector for High Performance Liquid Chromatograph

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

LC Technical Information

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

[application note] ACQUITY UPLC/SQD ANALYSIS OF POLYMER ADDITIVES. Peter J. Lee, and Alice J. Di Gioia, Waters Corporation, Milford, MA, U.S.A.

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

MODERN HPLC FOR PRACTICING SCIENTISTS

16 Malachite green 17 Leucomalachite green

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS

CEE 697z Organic Compounds in Water and Wastewater

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Technical Note. Introduction

Introduction. Chapter 1. Learning Objectives

API Optimization, Source Design, and Other Considerations of LC/MS

Transcription:

The Waters Acquity Ultra-Performance Liquid Chromatograph and the Micromass Quattro Premier Triple Quadrupole Mass Spectrometer Prepared by Yanjun Jiang & Luzal Vaidya December 2012 Section I Introduction An LC/MS/MS is a liquid chromatograph (LC) system coupled with a quadrupole mass spectrometer (MS). MS quadrupole is the most common mass analyzer, which allows low scan time (<100 ms) and is ideal for LC or GC inlets. When analyzing with LC/MS/MS, the LC separates compounds by conventional chromatography on a column. After the compounds enter the quadrupole mass detector, the solvent is removed and the compounds are ionized. Ions are filtered on the basis of their mass-to-charge ratio (m/z). Ions below and above a certain m/z value will be filtered out depending on the ratio of the direct current (DC) and alternating current (AC) voltages. By ramping the voltages on each set of poles, a complete range of masses can be passed to the detector. Liquid chromatography can separate metabolites that are not volatile and have not been derivatized. As a result, LC/MS can analyze a much wider range of chemical species than GC/MS. Samples commonly analyzed by LC/MS include amino acids (18 out of 20 amino acids can be derivatized, but the remaining two cannot) and sugars larger than trimers. The LC/MS/MS setup has higher compatibility detecting polar compounds such as organic acids, organic amines, nucleosides, ionic species, nucleotides, and polyamines compared to a GC/MS/MS. LC/MS is better suited for a discovery based approach when researching unknown metabolites, or when many of the targeted metabolites GC/MS incompatible due to volatility issues. Advantages of LC/MS/MS: MS provides exceptionally clean product (fragment) ion chromatograms for quantification; Useful for the rapid screening of complex samples where analytes of interest are known; Compound identity confirmation can be achieved with MS/MS using the product ion scan mode; By detecting a specific product ion (precursor ion mode) or charged fragments resulting from a neutral loss (neutral loss mode), a compound of interest can be classified. 1

LC/MS/MS Components The LC/MS/MS we used is a Waters ACQUITY UPLC separation module coupled with a Waters Micromass Quattro Premier XE Mass Spectrometer. (1) UPLC The ACQUITY UPLC Console contains four major parts: binary solvent manager, sample manager, column manager options, and detector options (Figure 1). Fig. 1 System overview of ACQUITY UPLC Fig. 2 Schematic of Binary solvent manager Binary solvent manager: Each of the solvent manager s two independent pump systems, A (left) and B (right), contains two linear-drive actuators. Each actuator pair comprises a single reciprocating serial pump that delivers precise flow of a single solvent. The two pump systems combine their two solvents at a filter/tee mixer. From there, the solvent mixture flows to the sample manager (Figure 2). A gradient elution program is commonly used so that the eluent composition (and strength) is steadily changed during the analysis. This increases separation efficiency, decreases the retention time and improves peak shape by minimizing tailing. The chromatography software controls the two solvents mixing ratio by varying the flow of pump A relative to that of pump B. A pressure transducer in each pump head relays pressure to the solvent manager, whose firmware measures pump head pressures during the pumping cycle. Thus the solvent manager independently pre-compresses the solvents in both the A and B portions to ensure consistent solvent delivery and minimize pump-induced detector baseline disturbances. Sample Manager: Plates are transferred in a programmable order from the Sample Organizer to the ACQUITY UPLC System Sample; Manager for processing plate exchange shuttle replaces a plate in 2

the Sample Manager within 15 seconds to support high sample-throughput processing; It can accept any combination of ANSI-compliant well plates and vial holders. Temperature control is of 4 to 40 C in 1 C increments. The ACQUITY system also supports a sample organizer which increases the level of automation and sample processing throughput by extending the capacity of the system, by more than tenfold for samples in standard ANSI-compliant 96 or 384 well plates; or more than threefold for samples in conventional 2 ml vials. UPLC Columns: Columns can be divided into two types: Normal phase column and Reverse phase column. Normal phase column: Stationary phase: higher polar rigid silica, or silica-based compositions. Mobile phases: relatively nonpolar solvent. More polar solvent has higher eluent strength and the least polar component is eluted first. Reverse phase column: Stationary phases: nonpolar hydrocarbons, waxy liquids, or bonded hydrocarbons (such as C18, C8, etc.). Mobile phase: polar solvents or mixtures such as methanolwater or acetonitrile-water. Less polar solvent has higher eluent strength and the most polar component is eluted first. Reverse phase columns are more commonly used in UPLC. The ACQUITY UPLC system also supports a column manager for high-productivity sample processing (Figure 3). It provides automated switching for up to four columns with dimensions up to 2.1 mm in internal diameter (I.D.) and 150 mm in length, as well as a bypass channel for flow injections. Fig. 3 An ACQUITY BEH C18 (1.7 µm, 2.1 X 50 m) UPLC reversed phase column and the schematic 3

Detectors: The system can be configured with in-house optical detectors (e.g. TUV, PDA, or ELS) or connected to a separate Mass Spectrometer. Here MS/MS detector is used and explained in details as follows. (2) MS/MS The Quattro Premier is a high performance tandem quadrupole mass spectrometer designed for routine LC-MS-MS operation. The instrument may be coupled to the following liquid introduction systems: LC system, to provide molecular weight information from an LC run or to perform target analysis and quantification; Syringe pump, for analysis of precious, low-concentration compounds. The Quattro Premier XE Schematic is shown in Figure 4. Fig. 4 Quattro Premier XE Schematic After a sample is introduced, at atmospheric pressure, into the ionization source 1 and ionized, ions are sampled through a series of orifices 2, then the ions are filtered according to their mass-to-charge ration (m/z) at the first quadrupole 3, and the mass-separated ions undergo collision-induced decomposition at the T-Wave Collision Cell 4, then the fragment ions are filtered according to their mass-to-charge ratio at the second quadrupole 5, finally the transmitted ions are detected by the photomultiplier detection system 6 and the signal is amplified, digitized, and presented to the control system 7. 4

MS/MS operating modes are shown in Table 1. The MS1 mode in which MS1 is used as the mass filter is directly analogous to a single quadrupole MS. The MS2 mode of operation is used with collision gas and when switching rapidly between MS and MS/MS operation; it can be useful for instrument tuning and calibration before MS/MS analysis or error checking. The Selected Ion Recording (SIR) mode of operation can also be used as a quantification mode when no suitable fragment ion can be found to perform a more specific analysis. Table 1 MS/MS operating modes (3) Other components Sample Inlet: An HPLC system or an infusion pump delivers sample to either an electrospray ionization (ESI) probe or atmospheric pressure chemical ionization (APCI) probe and serves as the sample inlet. Electrospray ionization (ESI) is an ionization technique where a sample solution is sprayed into a strong electric field in the presence of nitrogen to assist desolvation. The droplets formed evaporate in a region maintained at a vacuum causing the charge to increase on the droplets. The multiply charged ions then enter the analyzer. The most obvious feature of an ESI spectrum is that the ions carry multiple charges, which reduces their mass-to-charge ratio compared to a singly charged species. This allows mass spectra to be obtained for large molecules. Atmospheric pressure chemical ionization (APCI) generally produces protonated or deprotonated molecular ions from the sample via a proton transfer (+ve) or abstraction (-ve). The sample is vaporized in a heated nebulizer before emerging into a cloud of solvent ions formed within the atmospheric source by a corona discharge. Proton transfer or abstraction takes place between the solvent ions and the sample. 5

Vacuum System: An external rotary pump and an internal split-flow turbomolecular pump combine to create a vacuum. The turbomolecular pump evacuates the analyzer and ion transfer region. Data System: The data system collects information from the mass analyzer. The data system consists of: An embeded PC; an external workstation, and the Masslynx software. MassLynx Software: MassLynx software, a Windows based application, enables the following operation: Configuring Quattro Premier XE; Creating inlet and MS methods that define operation parameters for a run; Tuning and calibrating Quattro Premier XE; Running samples; Monitoring the run; Acquiring data. Analysis with LC/MS/MS Analysis with LC/MS/MS can be explained from two respects: (1) Method set-up (containing Mass Spectrometer Method set-up and LC Method Set-up) and (2) Sample analysis. (1) Method set-up Set-up Mass Spectrometer method: Ion source selection: Based on the chemical properties of target compounds, select proper ion source (e.g., ESI+ve, ESI-ve, APCI+ve, and APCI-ve). Syringe pump injection: Fill the syringe with standard solution of the target compound, using syringe pump for injection. Optimize MS/MS parameters: Steps: (1) ES/APCI Source: Optimize the voltages for ESI/APCI probe, Sample Cone, Extraction Cone and Hexapole; Source temperature; Desolvation and Cone Gas. (2) Analyzer: Resolution, Ion energy; entrance; collision; exit. (3) Diagnostics: (4) Determine the parent and daughter ions for qualification and quantification. (5) Save as MS File. 6

Set-up LC method: Optimize the conditions for LC, determine: (1) LC column; (2) Gradient program of solvents. (3) Retention time of the analytes (Set up the functions that the mass spectrometer uses to scan the instrument during an acquisition). (4) Set solvent delay. Save as the Inlet File. (2) Sample Analysis. Edit Sample List: (1) Sample name; (2) Import MS File; (3) Import Inlet (LC) File; (4) Import Tune File; (5) Edit sample type: Recovery, calibration, analyte. Prepare Mobile Phase: Solvent A, Solvent B, weak wash, strong wash etc. Prime: 50% Solvent A, 50% Solvent B, 2.7 min. Condition: Load LC method, wait for Delta (the difference between the maximum and minimum pressure of the column) decreases to 10. Run. NOTE: To analyze samples, standards should be run first to check the retention time; standard solutions with a series of concentrations should be run to establish calibration curves. Section II Procedure for PPCP analysis with LC/MS/MS. (1) Filtrate samples with GF/F to remove particles. (2) Perform SPE to remove impurities and concentrate target PPCPs: Use HLB Oasis 200 mg columns to extract PPCPs. These cartridges contain the Oasis HLB sorbent, which is a universal polymeric reversed-phase sorbent (30 μm) that can be used for the extraction of a wide range of acidic, basic, and neutral compounds from various matrices. The columns were sequentially preconditioned with 3 ml Dichloromethane (DCM), 5 ml Methanol (MeOH) and 7 ml Milli-Q water (H 2 O), and the samples were then loaded onto the cartridges at 10 ml/min after which the cartridges were rinsed with 3 ml H 2 O and 5 ml of 5% MeOH. The cartridges were dried for 40 minutes under vacuum and then the analytes were eluted with 6 ml MeOH followed by 4 ml MeOH/DCM (70:30). The eluate was then dried under a gentle stream of nitrogen at 35 C to a volume of 0.5 ml and then volumed upto 1 ml with MeOH. 7

(3) Analysis with LC/MS/MS: The PPCPs were measured according to Method 18CD (Vanderford et al 2012). Separation and detection of analytes were achieved using Ultra Performance Liquid Chromatography followed by Tandem Mass Spectrometry (UPLC-MS-MS). A Waters ACQUITY UPLC separation module was used for liquid chromatography and a Quattro Premier XE Micromass was used for mass spectrometry. Analytes were separated using an ACQUITY BEH C18 (1.7 µm, 2.1 X 50 m) UPLC column. These bridged, ethylsiloxane, hybrid particles can mechanically endure high-pressure conditions upto 15000 psi. The column dimensions allow optimal MS-compatible flow rates, and matched outlet tubing minimizes the effect of extra-column volume. Both ESI (+) and ESI ( ) modes were used for detection. The mobile phases for ESI (+) were Solvent A: 0.05% aqueous formic acid with 1 g ammonium formate and Solvent B: Methanol/ Acetonitrile (1:1) with 0.1% formic acid at a flow rate of 0.35 ml/min. The gradient program started with 3% solvent B, increased linearly to 95% solvent B in 10 minutes, and stayed at 95% solvent B for 6 minutes after which it was brought back to 3% solvent B in 1 minute and kept at 3% solvent B for 8 minutes. The total run time was 25 minutes. The mobile phases for ESI ( ) were Solvent A: 40 mg/l aqueous ammonium acetate and Solvent B: 100% methanol at a flow rate of 0.4 ml/min. The gradient program started with 3% solvent B, increased linearly to 25% solvent B in 1 minute, then increased linearly to 99% solvent B in 5 minutes, stayed at 99% solvent B for 5 minutes, then decreased back to 3% solvent B in 3 minutes and was kept at that composition for 7 minutes. The total run time was 20 minutes. Both internal and external standard calibration was used. Internal standard calibration was used for compounds for which internal standards were available (Table 2 and Table 3). Table 2: ESI- Compounds and Corresponding Transitions Name Transition (Parent>Daughter) Use Estradiol 270.96>144.83 Hormone Estrone 268.81>144.94 Hormone Furosemide 328.68>204.82 Loop Diuretic Gemfibrozil 248.88>120.88 Anti-cholesterol Drug Ibuprofen 204.93>160.98 Anti-inflammatory Drug Naproxen 228.82>169.88 Analgesic Naproxen d3 231.84>172.79 Internal Standard Sucralose 394.78>354.78 Artificial Sweetener Warfarin 306.86>160.79 Blood Thinner 8

Table 3: ESI+ Compounds and Corresponding Transitions Name Transition Use ( Parent >Daughter) Acetaminophen 151.87>109.91 Analgesic Atenolol 267.02>145.09 Beta-blocker Atenolol d7 274.15>145.05 Internal Standard Atorvastatin 559.1>440.2 Anti-cholesterol Drug Atorvastatin d5 564.08>445.12 Internal Standard Caffeine 194.97>137.98 Caffeinated Drinks Ciprofloxacin 332.01>314.13 Antibiotic Cotinine 176.97>80.08 Nicotine Metabolite DEET 192.09>119.04 Insect Repellant Diclofenac 295.83>213.91 Anti-inflammatory Diltiazem 414.83>177.83 Calcium Channel Blocker Diphenhydramine 255.98>166.89 Antihistamine Miconazole 416.72>160.83 Antifungal drug Primidone 218.93>161.95 Anti-epileptic Propanolol 260.05>182.95 Beta-blocker Propanolol d7 267.19>182.96 Internal Standard Ranitidine 315.07>176.04 Treat Ulcers & Heart Burn Ranitidine d6 321.08>176.04 Internal Standard Salbutamol 239.97>147.87 Asthma Treatment Sulfamethoxazole 253.94>155.9 Antibiotic Sulfamethoxazole d4 258.13>160.01 Internal Standard TCEP 286.84>225.01 Fire Retardant Trimethoprim 291.05>230.09 Antibiotic Trimethoprim d9 300.12>234.15 Internal Standard (4) LC/MS/MS software operation procedure. Edit sample list: Under Masslynx : 9

Enter : File name; File text (e.g., date); Bottle number; Inject volume. Import methods (Double click), choose: MS File(e.g., PPCP 2012_consensus_ESI pos. exp); Inlet File(e.g. PPCP 2012_3); MS Tune File: PPCP2012. Ipr. Turn on Gas for MS: Under Quattro Premier: Turn on: API GAS; COL GAS; Click: Press for operate. Prime the lines: Under ACQUITY UPLC Console: Binary Solvent Manager Control Prime A/B Solvents (for 2.7 min) Condition UPLC column: Inlet method Load method turn on the lamp Wait for the delta of the column pressure decreases to 10. After checking that all the buttons have turned green, run the sample list. (5) Results For Lab Demo, we prepared 300 ng/l PPCPs standard solution which was analyzed by LC/MS/MS. Some selected chromatograph peaks are shown in Figure 5. These chromatograms monitor reactions where a precursor ion isolated by one mass filter, passes into a collision cell in which it breaks into products. Product ion is selected by the second mass filter for passage to the detector, hence we can see that this process is extremely selective for just one analyte and vastly increases the signal-to-noise ratio for the analyte. (a) 10

(b) (c) (d) 11

(e) Fig. 5 Chromatograph peaks of (a) Acetominophen, (b) Salbutamol, (c) Cimetidine, (d) Trimethoprim, (e) Ciprofloxacin. (C = 300 ng/l) References: [1] Harris, Daniel C. 2007. Quantitative Chemical Analysis. W. H. Freeman and Company. [2] Waters Corporation. 2006. Acquity UPLC System: Operator s Guide. 71500082502/Revision C. [3] Waters Corporation. 2007-2009. Acquity UPLC Photodiode Array eλ Detector. 71500160703/Revision B. [4] http://www.ecs.umass.edu/cee/reckhow/courses/772/ [5] Waters Corporation. Micromass Quattro micro API Mass Spectrometer: Operator's Guide. 71500058602 Revision A [6] Vanderford, B.J., Drewes, J.E., Hoppe-Jones, C., Eaton, A., Haghani, A., Guo, Y.C., Snyder, S., Ternes, T., Schleusener, M. and Wood, C., J. (2012). Evaluation of Analytical Methods for EDCs and PPCPs via Interlaboratory Comparison, Water Research Foundation, Denver, CO. 12