Calibration of Orifice Flow Meter and Venturi Flow Meter

Similar documents
Flowmeter Discharge Coefficient Estimation

Major and Minor Losses

FLOW MEASUREMENT IN PIPES EXPERIMENT

Estimation of Flow Meter Losses

EXPERIMENT No.1 FLOW MEASUREMENT BY ORIFICEMETER

EXPERIMENT NO. 4 CALIBRATION OF AN ORIFICE PLATE FLOWMETER MECHANICAL ENGINEERING DEPARTMENT KING SAUD UNIVERSITY RIYADH

Experimental and CFD analysis of flow through venturimeter to determine the coefficient of discharge

Chapter 4 DYNAMICS OF FLUID FLOW

Lecture23. Flowmeter Design.

VENTURIMETER EXPERIMENT

Laboratory work No 2: Calibration of Orifice Flow Meter

MCE380: Measurements and Instrumentation Lab

Lecture 13 Flow Measurement in Pipes. I. Introduction

Mass of fluid leaving per unit time

Experiment (4): Flow measurement

Physics Courseware Physics I

Compressible Gas Flow

Rate of Flow Quantity of fluid passing through any section (area) per unit time

Mechanisms of Vortex Oscillation in a Fluidic Flow Meter

LOSSES DUE TO PIPE FITTINGS

Flow Measurement in Pipes and Ducts COURSE CONTENT

Instruction Manual. Equipment for Engineering Education

5 ENERGY EQUATION OF FLUID MOTION

Given Find Water Properties

CHAPTER THREE FLUID MECHANICS

Applied Fluid Mechanics

Applied Fluid Mechanics

Chapter 3 Bernoulli Equation

If a stream of uniform velocity flows into a blunt body, the stream lines take a pattern similar to this: Streamlines around a blunt body

Lecture 24. Design of flow meters

Lecture 3 The energy equation

Mechanical Measurements and Metrology Prof. S. P. Venkateshan Department of Mechanical Engineering Indian Institute of Technology, Madras

Experiment No.4: Flow through Venturi meter. Background and Theory

Lab Section Date. ME4751 Air Flow Rate Measurement

ME332 FLUID MECHANICS LABORATORY (PART I)

Chapter (6) Energy Equation and Its Applications

CHAPTER (13) FLOW MEASUREMENTS

Course: TDEC202 (Energy II) dflwww.ece.drexel.edu/tdec

MAHATMA GANDHI MISSION S JAWAHARLAL NEHRU ENGINEERING COLLEGE, FLUID MECHANICS LABORATORY MANUAL

Given Find Water Properties

New Website: Mr. Peterson s Address:

CVE 372 HYDROMECHANICS EXERCISE PROBLEMS

The Expansibility Factor Equations in ISO and ISO : Do They Deliver What They Promise?

I. To find the coefficient of discharge for vcnturi meter. 2. To find the coefficient of discharge for ori rice meter.

CHARACTERISTICS OF DISCHARGE COEFFICIENT IN INDUSTRIAL BURNER APPLICATIONS

ME411 Engineering Measurement & Instrumentation. Winter 2017 Lecture 11

MEDIDAS EN FLUIDOS. INEL 5205 Instrumentación. Monday, November 19, 12

Exam #2: Fluid Kinematics and Conservation Laws April 13, 2016, 7:00 p.m. 8:40 p.m. in CE 118

CALCULATION OF COEFFICIENT OF DISCHARGE OF ORIFICE PLATE AND FLOW NOZZLE

Roll No. :... Invigilator s Signature :.. CS/B.Tech (EIE)/SEM-5/EI-501/ INDUSTRIAL INSTRUMENTATION II

Discharge measurements

The Mechatronics Design for Measuring Fluid Friction Losses in Pipe Flows Rıza Gurbuz

Computation of pressure loss for differential producing flowmeters

53:071 Principles of Hydraulics Laboratory Experiment #3 ANALYSIS OF OPEN-CHANNEL FLOW TRANSITIONS USING THE SPECIFIC ENERGY DIAGRAM

ACE Engineering College

ERRATA CRANE FLOW OF FLUIDS THROUGH VALVES, FITTINGS AND PIPE TECHNICAL PAPER NO. 410 METRIC VERSION

Experiences in an Undergraduate Laboratory Using Uncertainty Analysis to Validate Engineering Models with Experimental Data

M E 320 Professor John M. Cimbala Lecture 05

Water Circuit Lab. The pressure drop along a straight pipe segment can be calculated using the following set of equations:

Measurements using Bernoulli s equation

INTRODUCTION AND BASIC CONCEPTS. Chapter 1. Mehmet Kanoglu. Thermodynamics: An Engineering Approach, 6 th Edition. Yunus A. Cengel, Michael A.

Benha University College of Engineering at Benha Questions For Corrective Final Examination Subject: Fluid Mechanics M 201 May 24/ 2016

Chapter 15B - Fluids in Motion. A PowerPoint Presentation by Paul E. Tippens, Professor of Physics Southern Polytechnic State University

Chapter 15: Fluid Mechanics Dynamics Using Pascal s Law = F 1 = F 2 2 = F 2 A 2

Name. Chemical Engineering 150A Mid-term Examination 1 Spring 2013

MAHATMA GANDHI MISSION S JAWAHARLAL NEHRU ENGINEERING COLLEGE, AURANGABAD. (M.S.)

04/01/1998 Developments in DP Flowmeters By Jesse Yoder

Middle East Technical University Department of Mechanical Engineering ME 305 Fluid Mechanics I Fall 2018 Section 4 (Dr.

Fluids. Fluids in Motion or Fluid Dynamics

ABSTRACT I. INTRODUCTION

S.E. (Mech.) (First Sem.) EXAMINATION, (Common to Mech/Sandwich) FLUID MECHANICS (2008 PATTERN) Time : Three Hours Maximum Marks : 100

NORTH SEA FLOW MEASUREMENT WORKSHOP 2004 In. St Andrews, Scotland

Signature: (Note that unsigned exams will be given a score of zero.)

Orifice and Venturi Pipe Flow Meters

CHAPTER 28 PRESSURE IN FLUIDS

Local energy losses at positive and negative steps in subcritical open channel flows

DETERMINATION OF DISCHARGE AND HEAD LOSS USING A FLOW-MEASURING APPARATUS

Fluid Mechanics. du dy

Flow Measurement in Pipes and Ducts COURSE CONTENT

Experiment- To determine the coefficient of impact for vanes. Experiment To determine the coefficient of discharge of an orifice meter.

POWER UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES

Analysis of the possibility to optimize the passive pressure pulsations dampers using CFD

2 Internal Fluid Flow

Annubar Primary Element Flow Calculations

INSTRUCTIONS FOR LABORATORY EXPERIMENT IN FLUID MECHANICS

Properties of Gases. The perfect gas. States of gases Gas laws Kinetic model of gases (Ch th ed, th ed.) Real gases

Flow Behavior Lab BSEN Major and Minor Losses. Samuel Dunbar

FLUID MECHANICS. Chapter 3 Elementary Fluid Dynamics - The Bernoulli Equation

Fluid Mechanics-61341

FLUID MECHANICS PROF. DR. METİN GÜNER COMPILER

Chapter 11 - Fluids in Motion. Sections 7-9

4 Mechanics of Fluids (I)

Flow rate and mass flow rate

Eric G. Paterson. Spring 2005

Q1 Give answers to all of the following questions (5 marks each):


Orifice and Venturi Pipe Flow Meters

Hydraulics and hydrology

V/ t = 0 p/ t = 0 ρ/ t = 0. V/ s = 0 p/ s = 0 ρ/ s = 0

1.4 Perform the following unit conversions: (b) (c) s. g s. lb min. (d) (e) in. ft s. m 55 h. (f) ft s. km h. (g)

Transcription:

Calibration of Orifice Flow Meter and Venturi Flow Meter D. Till Abstract Orifice and venturi flow meters decrease the pressure of a fluid b increasing its velocit as it flows through them. This is done b creating a smaller diameter at a certain point in the pipe. The pressure loss was measured using a manometer attached to each of the meters, and the flow rate was calculated that corresponded to that pressure change. This was repeated seven times for flow rates varing from approximatel 90 to 500 gallons per hour. Plots were made of pressure loss vs. flow rate for the orifice and venturi meters. An expression was obtained for the flow rate using the equation of the curve from the graphs for each of the meters. This calibrated the two meters and allows the flow rate to be calculated if the pressure differential is known. Introduction Determining the flow rate of fluids is important in man applications. This can be done using devises called differential flow meters. These are obstructions placed in the pipe that alter the pressure and velocit. A commonl used device for measuring the pressure differential cross a head is a manometer. It consists of a glass or plastic U-tube containing a fluid such as water, mercur, or oil. The change in the height of the fluid in the tube indicates the pressure difference using the formula P! = P! + ρgh where ρ is the densit of the fluid, g is the acceleration due to gravit (9.81 m/s! ), and h is the height of the fluid in the tube. Manometers are ver useful for measuring small to moderate pressure (Cengel and Cimbala 2010). Two devices commonl used to measure the flow rate of fluids are the Orifice flow meter and the Venturi flow meter. The both use the same principle for measuring flow rate, which is the relationship between the pressure and the velocit of the fluid ( Orifice, Nozzle, and Venturi Flow Rate Meters, Engineeringtoolbox). An orifice meter is a thin plate with a hole in the middle that is placed in a pipe through which the fluid flows. It increases the velocit of the

fluid as it flows through it, which decreases the pressure ( Fundamentals of Orifice Metering, Afms). A venturi meter uses a narrowing throat in the pipe that expands back to the original pipe diameter. It creates an increase in the velocit of the fluid, which also results in a pressure drop across that section of the pipe. It is more efficient and accurate than the orifice meter ( Orifice, Nozzle, and Venturi Flow Rate Meters, Engineeringtoolbox). Objectives The objectives of this experiment were to gain knowledge on pressure and flow measuring devices, and to calibrate an orifice flow meter and a venturi meter. Materials and Methods A manometer (Figure 1), an orifice flow meter (Figure 2), and a venturi flow meter (Figure 3) were used to measure the flow rate and pressure change of the water flowing through the pipes. The valve was opened and the water that was discharged through the pipes was collected in a bucket. The amount of time the water was running and the mass of the water were recorded. The mass of the water collected was measured using a mass balance. Simultaneousl, the head differentials across the orifice and venturi meter were recorded using a manometer. The difference in the tank volume was also recorded b reading the tank gage. The initial amount of water in the tank for each run was 8 gallons. This process was repeated seven times and the data was recorded for each run. The pressure loss for the orifice and venturi meters was calculated using the equation Pressure Loss = ρgh where ρ is the densit of water (ρ = 1000 kg/m! ), g is Earth s gravitational acceleration (g = 9.81 m/s! ), and h is the height difference recorded from the manometer. The mass flow rate was calculated using the formula m = m t where m is the mass of the water collected and t is the amount of time the water was running. Similarl, the volumetric flow rate was calculated using the formula V = V t

where V is the change in volume of the water (measured from tank reading), and t is the amount of time the water was running. A plot was made for pressure loss vs. mass flow rate and pressure loss vs. volumetric flow rate for each of the meters. Figure 1: basic U- tube manometer Figure 2: Schematic of an orifice flow meter Figure 3: schematic of a venturi flow meter Results and Discussion Obtaining the flow rates and the pressure change through the orifice and venturi plates, the relationship between them was established, and a pressure change vs. flow rate plot was made for each (Figure 4 & 5). Both plots show the relationship between pressure change and flow rate to be exponential.

Pressure Loss (Pa) 9000 8000 7000 6000 5000 4000 3000 2000 1000 0 OriCice = 0.012x 2.1612 R² = 0.99367 = 0.0148x 2.1281 R² = 0.99443 0 100 200 300 400 500 600 Flow Rate (gal/hr) gph (Mass) gph (Tank) (Mass)) (Tank)) Figure 4: Pressure loss vs. flow rate (mass and volumetric) for orifice flow meter Pressure Loss (Pa) 4000 3500 3000 2500 2000 1500 1000 500 0 Venturi = 0.0277x 1.8811 R² = 0.98925 = 0.0335x 1.8511 R² = 0.98872 0 100 200 300 400 500 600 Flow Rate (gal/hr) gph (mass) gph (tank) (mass)) (tank)) Figure 5: Pressure loss vs. flow rate (mass and volumetric) for venturi flow meter

The equation of the line for the orifice meter plot was: = 0.0148x!.!"#! where is the pressure loss and x is the flow rate. Solving the equation for x will give the flow rate 0.0148!.!" This equation can be used to calculate the flow rate when the pressure difference is known for flow through an orifice meter. The equation of the line for the venturi meter plot was = 0.0277x!.!!"" where is the pressure loss and x is the flow rate. Solving the equation for x gives the flow rate 0.0277!.!"#$ This equation gives the flow rate when the pressure loss is known for flow through a venturi meter. The curve for the orifice meter had an R! value of 0.99367 and the venturi meter had an R! value of 0.98925. This implies that the curves accuratel represent the data points. The mass and volumetric flow rates did not differ b much, as seen on the graphs. The orifice meter had pressure losses that were approximatel double that of the venturi meter. Calibrating the orifice and venturi meters gave equations for obtaining the flow rates when a pressure difference is known.

Conclusions This experiment effectivel calibrated an orifice and a venturi flow meter b flowing a known volume and mass of water through them, reading the pressure change from a manometer, and calculating the mass and volumetric flow rates of the water. The pressure loss vs. flow rate was plotted for the two meters, and an equation was obtained for each of the curves. The equation for calculating flow rate for the orifice meter was found to be 0.0148!.!" and the equation for flow rate from the venturi meter was found to be 0.0277!.!"#$ where x is the flow rate and is the observed pressure change for both of the equations. The mass and volumetric flow rates were plotted on the same graph for each of the meters, and there was little difference in them. The were approximatel the same curve. The mass flow rate data was used for obtaining the expressions above for flow rate because it had a slightl higher R! value than that of the volumetric flow rate data, hence, more accurate. The orifice meter had pressure losses that were approximatel double that of the venturi meter. This is due to the fact that the orifice meter increases the velocit of the fluid b a greater amount than the venturi meter.

References Çengel, A., and J. M. Cimbala. Fluid Mechanics: Fundamentals and Applications. 2nd ed. Boston: McGraw-Hill Higher Education, 2010. Print. "Fundamentals of Orifice Metering." Afms.org. Smith Metering, Inc, n.d. Web. 30 Sept. 2013. <http://www.afms.org/docs/gas/fundamenatls_of_orifice.pdf>. "Orifice, Nozzle and Venturi Flow Rate Meters." Engineeringtoolbox.com. N.p., n.d. Web. 30 Sept. 2013. <http://www.engineeringtoolbox.com/orifice-nozzle-venturi-d_590.html>.