Higher Confidence in Identification with QTRAP LC/MS/MS Systems when Screening and Quantifying Pesticides in Fruit and Vegetable Samples

Similar documents
Multiresidue Analysis of 100 Pesticides in Food Samples by LC/Triple Quadrupole Mass Spectrometry Application

Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou AB SCIEX, Concord, Ontario, Canada

Overview. Introduction

Food & Environmental. Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou SCIEX, Concord, Ontario, Canada

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Advanced Data Acquisition and Data Processing Workflows to Identify, Quantify and Confirm Pesticide Residues

Quantitative Analysis and Identification of Migrants in Food Packaging Using LC-MS/MS

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

Powerful Scan Modes of QTRAP System Technology

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

Automated Screening of 600 Pesticides in Food by LC/TOF MS Using a Molecular-Feature Database Search Application

Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Available online at Journal of Chromatography A, 1175 (2007) 24 37

László Tölgyesi LC/MS scientist EFSC Demo Group. International Symposium on Food Safety and Quality: Applications of Nuclear and Related Techniques

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

Confirmation of In Vitro Nefazodone Metabolites using the Superior Fragmentation of the QTRAP 5500 LC/MS/MS System

DETERMINATION OF PESTICIDES IN FOOD USING UPLC WITH POLARITY SWITCHING TANDEM QUADRUPOLE LC/MS/MS

Introduction to the Q Trap LC/MS/MS System

Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) tandem mass spectrometry

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

Simplified Workflow and Improved Confidence in Screening Methods on Time of Flight Mass Spectrometry

Theory and Practical Consequences.

Charles Yang, Dipankar Ghosh, Jonathan Beck Thermo Fisher Scientific, San Jose, CA, USA

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Analysis of Pesticides in Citrus Oil Using a Shimadzu GCMS-TQ8030. No. GCMS No. SSI-GCMS-1404

Analyst Software. Automatic Optimization Tutorial

ABI 3200 Q TRAP LC/MS/MS System

Rapid Pesticide Screening and Identification Using the High Resolution All Ions MS/MS Technique

Analyst Software. Peptide and Protein Quantitation Tutorial

Triumphs and Challenges of High-Resolution Mass Spectrometry in Comprehensive Residue Screens

Application Note. Abstract. Authors. Food Safety and Environmental

Accurate, High-Throughput Protein Identification Using the Q TRAP LC/MS/MS System and Pro ID Software

Analysis of Polar Metabolites using Mass Spectrometry

QTRAP and SelexION+- Improving the Sensitivity, Speed and Selectivity in Food, Clinical Research and Pharmaceutical Applications

Multi-residue analysis of pesticides by GC-HRMS

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Fully automated QuEChERS clean-up and LC/MS-QQQ analysis of pesticides in fruits and vegetables.

Quantitative and Qualitative Confirmation of Pesticides in Beet Extract Using a Hybrid Quadrupole-Orbitrap Mass Spectrometer

LC-MS/MS multiresidue method used for pesticides analysis in fruits and vegetables

AppNote 2/2013. A High Throughput Automated Sample Preparation and Analysis Workfl ow for Comprehensive Toxicology Urine Screenings using LC/MS/MS

Tyler Trent, Applications Sales Specialist, Teledyne Tekmar P a g e 1

Routine Multiresidue Pesticide Analysis using the Agilent 6470 Triple Quadrupole Mass Spectrometer

Information Dependent Acquisition (IDA) 1

Analysis of Pesticides in Baby Foods Using a Triple-Quadrupole GC/MS/MS

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

QTRAP Technology A COMPENDIUM OF RELATED RESEARCH APPLICATIONS

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

WADA Technical Document TD2003IDCR

Screening Validation Report 1

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Highly Sensitive and Rugged GC/MS/MS Tool

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Pesticides in Cannabis

4000 Q TRAP LC/MS/MS System. Advanced Linear Ion Trap technology at the highest level of sensitivity 4000 QTRAP. LC/MS/MS System

Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365

Pure Chromatography. Food Safety Applications. Abstract. Introduction

Untargeted Screening on the X500R QTOF

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Translational Biomarker Core

Safeguarding Food from Pesticides by UHPLC After Extraction with the QuEChERS Method

Multi-residue analysis of pesticides in crude food extracts using a simple extraction technique and LC/MS/MS

Improved Automated Sample Preparation for the Analysis of 25-OH-Vitamin D3 by LC/MS/MS

Analysis and Quantification of Mycotoxins in Cereals Using MRM HR on the SCIEX X500R QTOF System with SCIEX OS

LC/Q-TOF Workflows for Comprehensive Micropollutant Analysis

Application Note LCMS-110 Development of a Targeted Quantitative LC-MS/MS Method for 431 Positive and Negative Ion Pesticides in a Single Analysis

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Validation Report 8 Determination of pesticide residues in rice and wheat by GC-MS/MS and LC-MS/MS (QuEChERS method) Appendix 2

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Thermo Scientific LTQ Orbitrap Velos Hybrid FT Mass Spectrometer

Qualitative and Quantitative Analysis of Pesticides in Horse Feed Matrix Using Orbitrap MS

Analysis of Pesticides and Contaminants using HS-SPME-GC/MS and DRS Detection. Blair Surridge, Tyler Dyer, Rene Bilideau, Paloma Bertolin Presoti

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

CHROMABOND HLB. Introduction. Advantages of CHROMABOND HLB. Technical data

Thermo Finnigan LTQ. Specifications

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

Application News. No. C154. Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

High Efficiency, Broad Scope Screening of Pesticides Using Gas Chromatography High Resolution Orbitrap Mass Spectrometry

Determination of 198 pesticide residues in eggplant using Gas Chromatography tandem Mass Spectrometry/Mass Spectrometry

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

TANDEM MASS SPECTROSCOPY

Validation Report 3 Determination of pesticide residues in cereals by GC-MS/MS (QuEChERS method)

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

Finnigan LCQ Advantage MAX

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Analytical considerations on the use of a fruit-specific and representative matrix in pesticide residue analysis by LC-ESI-MS/MS

2016 iquan Series: Pesticide Quantitation on the Thermo Scientific TSQ Quantiva

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Accelerated Impurity Profiling Workflow for Bulk Drugs and Formulations using an Accurate Mass Workhorse

Agilent 6495 Triple Quadrupole LC/MS System EXPERIENCE A NEW LEVEL OF CONFIDENCE

Transcription:

Higher Confidence in Identification with QTRAP LC/MS/MS Systems when Screening and Quantifying Pesticides in Fruit and Vegetable Samples André Schreiber AB SCIEX, Concord, Ontario, Canada Figure 1. Ion path of a QTRAP LC/MS/MS System Overview QTRAP LC/MS/MS Systems provide in comparison to triple quadrupole systems: MRM selectivity and sensitivity for quantitation with the lowest limits of detection, excellent reproducibility and linear range Fast and high sensitivity Linear Ion Trap scan functions for qualitative analysis and identification Rich MS/MS spectra because of fragment ion generation in the LINAC Collision Cell and the use of Collision Energy Spread Combination of triple quadrupole and QTRAP System functions without delay time in IDA experiments allowing quantitation and identification with MS/MS spectra in a single LC run Higher confidence of identification based on mass spectral library search minimizing the risk of false positive and negative results This paper describes the use of QTRAP Systems coupled to liquid chromatography for the screening and identification of pesticides in fruit and vegetable samples. The selectivity, sensitivity, and accuracy of detection using QTRAP System and conventional triple quadrupole mass spectrometers are compared. The higher confidence of identification provided by QTRAP linear ion trap MS/MS scans followed by library searches which minimize the potential risk of false positive and negative results is highlighted. Introduction Liquid Chromatography (LC) coupled to tandem Mass Spectrometry (MS/MS) is a powerful analytical tool used in many food, environmental and forensic testing laboratories to detect targeted analytes, such as pesticides, drugs of abuse, steroids, pharmaceuticals, personal care products, and many other contaminants at trace levels. A variety of ion sources including Electrospray Ionization (ESI), Atmospheric Pressure Chemical Ionization (APCI), and Atmospheric Pressure Photo Ionization (APPI) facilitates the ionization of polar to non-polar compounds. Generated ions are transferred following ionization through a vacuum interface into the mass analyzer. p 1

1.14e5 1.1e5 1.e5 MRM chromatogram 5283 5 45 Product Ion Spectrum (triple quadrupole MS/MS) 174.2 9.5e5 9.e5 8.5e5 8.e5 Enhanced Product Ion Spectrum (QTRAP MS/MS) 174. 9.e4 8.e4 7.e4 6.e4 4 35 3 96.2 14.2 7.5e5 7.e5 6.5e5 6.e5 5.5e5 96. 14. 5.e4 4.e4 3.e4 2.e4 1.e4 5. 1 2 3 4 5 6 7 8 9 1 11 12 13 14 15 4.5e5 25 4.e5 79. 3.5e5 2 3.e5 216. 2.5e5 79. 132. 15 216. 146. 2.e5 146.4 1.5e5 1 132.2 1.e5 138. 11 5.e4 99. 127. 5 7 8 9 1 11 12 13 14 15 16 17 18 19 2 21 22 23 24 25 7 8 9 1 11 12 13 14 15 16 17 18 19 2 21 22 23 24 25 Figure 2. Screening and identification of pesticides using triple quadrupole and QTRAP LC/MS/MS Systems. Left: detection 15 Multiple Reaction Monitoring transitions to screen for pesticides. Middle: Product Ion spectrum of a triple quadrupole MS/MS. Right: Enhanced Product Ion spectrum of a QTRAP MS/MS to identify the presence of a detected pesticide. (Note that the sensitivity of the QTRAP MS /MS spectrum is more than 1 times higher than that of the triple quadrupole spectrum.) Triple quadrupole (QqQ) mass analyzers operating in Multiple Reaction Monitoring (MRM) are mostly used for the quantitation of targeted analytes. In MRM mode, the first quadrupole filters a specific precursor ion; the collision cell (second quadrupole) generates fragments (product ions) which are filtered in the third quadrupole (Figure 2). functions. These additional LIT scan functions greatly enhance the performance for screening, confirmation and identification applications. (1) Trapping In 22, AB SCIEX introduced a new generation of LC/MS/MS systems: hybrid triple quadrupole linear ion trap mass spectrometers with axial ion ejection (QTRAP System). QTRAP Systems enable the unique capabilities provided by this technology by coupling targeted triple quadrupole experiments, such as MRM, precursor ion, and neutral loss scans, with high sensitivity ion trap scan functions, such as enhanced MS, enhanced product ion, enhanced resolution, enhanced multiple charge and MS/MS/MS scan. 2-4 (2) Scanning Exit lens Having a variety of new scan types and the possibility to combine them in single LC runs enable completely new workflows for many LC/MS/MS applications, including food, environmental, and forensic testing, metabolite identification and proteomics. 5-15 QTRAP System Technology The mass analyzer region of a QTRAP System is based on the conventional ion path of a triple quadrupole mass spectrometer. In comparison to QqQ systems, the third quadrupole (Q3) of a QTRAP System can be operated as a quadrupole or as a Linear Ion Trap (LIT). The dual functionality of Q3 provides the QTRAP System complete functionality of QqQ mass spectrometer, including true Multiple Reaction Monitoring for selective and sensitive quantitation and additional LIT scan Entrance lens Figure 3. Trapping and scanning of ions in a Linear Ion Trap (LIT) The operation of Q3 as LIT can be divided into two basic steps: (1) trapping and (2) scanning as shown in Figure 3. First precursor ions and/or product ions, depending on the selected scan mode, exit the collision cell and enter Q3 operating in LIT mode. The ions are trapped in Q3 by the RF voltage in radial direction and by a repulsive DC voltage applied to the exit lens of Q3. After a fill time of typically a few milliseconds to accumulate ions, a repulsive entrance lens is used to stop ions entering Q3. Trapped ions are then cooled for a short time and subsequently ejected axially from the LIT using an auxiliary voltage that is ramped in concert with the drive RF voltage. p 2

The most important LIT scan modes for screening, confirmation and identification applications are Enhanced MS (EMS), Enhanced Resolution (ER) and Enhanced Product Ion (EPI) scans. In EMS, the ion trap is filled with molecular ions generated in the ion source to identify unknown compounds. The ER scan measures the mass of a targeted ion more accurately and can be used to determine charge states and isotopic patterns. Finally, the EPI mode can be used to identify the presence of the detected analyte by interpretation of fragment ions or mass spectral library searching. The Enhanced Product Ion scan of QTRAP Systems has several advantages over product ion scanning of QqQ mass spectrometers, including: Much higher sensitivity of Enhanced Product Ion spectra because of ion accumulation in the LIT as shown in Figure 2 More product ion information in a single MS/MS spectrum because of Collision Energy Spread Lower cycle time due to faster scanning of linear ion trap The Enhanced Product Ion scan of QTRAP Systems has also several advantages over product ion scan of traditional ion trap mass spectrometers, including: Fragment rich MS/MS spectra characteristic for Collision Induced Dissociation (CID) in comparison to resonance excitation spectra No inherent low mass cut-off due to fragment generation in the collision cell rather than in the ion trap Lower cycle time than traditional ion traps because of simultaneous precursor ion selection and fragmentation while QTRAP System is filled Remarkable immunity of LIT from space charge effects which result in mass shift and incorrect isotopic pattern in the spectrum 2-3 Experimental A 32 QTRAP LC/MS/MS System with Turbo V Source and an ESI probe was used in both triple quadrupole and QTRAP modes to investigate the use of these two mass spectrometers for the screening and identification of pesticides. Samples Fruit and vegetable samples (apple, apricot, banana, cucumber, grape, grapefruit, kiwi, lemon, orange, pear, pepper, raisin, strawberry, tea, and tomato) were extracted using a modified QuEChERS (Quick, Easy, Cheap, Effective, Rugged, and Safe) procedure. 16 HPLC A Shimadzu Prominence LC system consisting of system controller CBM-2A, 2 pumps LC-2AD, degasser DGU-2A3, autosampler SIL-AC, and column oven CTO-2AC was used. A gradient of eluent A (water with 5 mm ammonium formate) and eluent B (methanol with 5 mm ammonium formate) from 8/2 to 1/9 (A/B) over 8 min was used on a Phenomenex Synergi 4u Fusion-RP 8A 5x2 mm column. The column temperature was set to 25 C. Sample volume of 2 μl was injected. MS/MS Detection In QqQ mode 75 pesticides were detected in positive polarity using 2 MRM transitions per analyte. The detected quantifier and qualifier MRM with the respective MRM ratio are listed in Table 1. All compound dependent parameters were completely optimized for the detection of each pesticide with the highest sensitivity. A dwell time of 5 ms was used for all MRM transitions. This typical triple quadrupole experiment was compared to a unique QTRAP Information Dependent Acquisition (IDA) where only a single MRM transition was monitored with a dwell time of 1ms per compound. This MRM survey automatically triggered an Enhanced Product Ion scan if the MRM signal was above a threshold of 5 cps. EPI spectra were acquired from 5 to 55 amu with a scan speed of 4 amu/s and a Collision Energy (CE) of 35 V with Collision Energy Spread (CES) of 15 V. Also dynamic exclusion was activated to minimize the risk of missing co-eluting compounds. XIC of +MRM (15 pairs): 278.2/1... 1.2e5 1.e5 8.e4 6.e4 4.e4 2.e4 Max. 4.9e4 cps. 5 6 7 8 9 1 XIC of +MRM (15 pairs): 278.2/1... Max. 4.9e4 cps. 4.9e4 4.e4 3.e4 2.e4 Standard 1ppb Standard 1ppb MRM of Metazachlor (ratio =.99) 7.6 7.6 5 6 7 8 9 1 XIC of +MRM (15 pairs): 278.2/... Figure 4. Quantitation and MRM ratio identification of 5.8 μg/kg Metazachlor in an extract of apricot (278/134 and 278/21) 29 25 2 15 1 5 Max. 29 cps. 5 6 7 8 9 1 XIC of +MRM (15 pairs): 278.2/... Max. 29 cps. 29 25 2 15 1 5 Extract of apricot 5.8μg/kg (1x diluted) 7.6 Extract of apricot MRM of Metazachlor (ratio =.9) 5 6 7 8 9 1 p 3

Table 1. MRM transitions for the 75 studied pesticides, MRM ratios, Limits of Detection (S/N=6) and coefficients of variation at 1ng/mL (n=3) Pesticide MRM 1 MRM 2 Ratio LOD (ng/ml) %CV (n=3)pesticide MRM 1 MRM 2 Ratio LOD (ng/ml) %CV (n=3) Acetamiprid 223/126 223/99.2.2 4.7 Irgarol 254/198 254/83.14 5 2.2 Aldicarb 28/89 28/116.6 1. 8.5 Isoproturon 27/72 27/46.25.2 3.9 Atrazine 216/174 216/14.51.1 5.5 Lenacil 235/153 235/136.24.2 5. Azoxystrobin 44/372 44/344.45.1 1.5 Linuron 249/16 249/182.98 1. 8.5 Benalaxyl 326/148 326/91.97.1 3.3 Lufenuron 511/141 511/158.5 5. 11.9 Bendiocarb 224/19 224/167.95.5 7.9 Malathion 331/127 331/99.42.2 2.7 Bromuconazole 378/159 378/161.77.5 4.5 Mebendazole 296/164 296/15.26.2 1.1 Buprofezin 36/21 36/116.71.2 4.9 Metalaxyl 28/22 28/192.79.5 4.9 Carbaryl 22/145 22/127.4.2 5.4 Metamitron 23/175 23/14.72.2 4.3 Carbendazim 192/16 192/132.17.1 3.1 Metazachlor 278/134 278/21.98.1 2.4 Carbofuran 222/165 222/123.44.2 3.3 222/165 222/15.35.1 3.5 Chlorfenvinphos 359/99 359/155.9.5 5.5 Methidathion 33/145 33/85.93.5 4.7 Chlorotoluron 213/72 213/46.21.2 3.2 Methiocarb 226/169 226/121.96.5 3.3 Chlorpyrifos-methyl 322/125 324/125.71 1. 8.4 Methomyl 163/88 163/16.77 1. 7. Cyanazine 241/214 241/14.24.5 6.9 Metolachlor 284/252 284/176.29.1 2.2 Cyproconazole 292/125 292/7.95.5 7.2 Metolcarb 166/19 166/94.3.5 3.8 Cyromazine 167/85 167/68.45.5 4.2 Metribuzin 215/187 215/84.7 1. 8.2 Desethyl-atrazine 188/146 188/14.4.5 7.7 Molinate 188/126 188/55.52 2. 7.9 Methabenzthiazuron Desisopropylatrazine 174/14 174/96.77 1. 6.7 Monolinuron 215/126 215/99.68.5 6. Diazinon 35/169 35/97.94.1 3.1 Monuron 199/72 199/126.16.1 1.6 Dichlorvos 221/19 223/19.55 1. 4.6 Nitenpyram 271/126 271/237.78.5 4.1 Difenoconazole 46/251 48/253.56.2 2.8 Oxadixyl 279/219 279/132.51.5.7 Difenoxuron 287/72 287/123.65.5 2. Pendimethalin 282/212 282/194.18 1. 8.7 Diflubenzuron 311/141 311/158.7 1. 6.6 Prochloraz 376/38 376/7.26.2 2.8 Dimethenamide 276/244 276/168.24.1 3.2 Profenofos 375/35 373/33.87 1. 1.9 Dimethoate 23/125 23/199.76.2 2.8 Promecarb 28/19 28/151.94 1. 6.4 Dimethomorph 388/31 388/165.52.5 4.3 Prometon 226/142 226/86.29.1 3.2 Diuron 233/72 235/72.7.5 5.1 Prometryn 242/158 242/2.43.2.2 Ethiofencarb 226/17 226/77.62.5 6.7 Propachlor 212/17 212/94.72.2 3.7 Fenamiphos 34/217 34/22.59.1 2.6 Propanil 218/127 218/162.96 1. 3.6 Flufenacet 364/194 364/152.71.2 3.4 Propiconazole 342/159 342/69.25.5 1. Flufenoxuron 489/141 489/158.51 1. 7.6 Prosulfocarb 252/91 252/128.21.2 7.3 Fluroxypyr 255/29 255/181.74 5. 1.4 Simazine 22/132 22/124.9.5 3.1 Hexaflumuron 461/158 461/141.59 5. 9. Spiroxamine 298/144 298/1.66.2 3.4 Imazalil 297/159 299/161.37.5 8.5 Terbutryn 242/186 242/68.18 2 2. p 4

Table 1. continued Pesticide MRM 1 MRM 2 Ratio LOD (ng/ml) %CV (n=3)pesticide MRM 1 MRM 2 Ratio LOD (ng/ml) %CV (n=3) Imazapyr 262/217 262/149.54.2 3. Thiabendazole 22/175 22/131.62.1 1.6 Imazaquin 312/199 312/267.99.5 4.5 Thiacloprid 253/126 253/99.32.2 6.7 Imidacloprid 256/175 256/29.73 1. 4.6 Results MRM Selectivity Detection in Multiple Reaction Monitoring (MRM) using QTRAP and triple quadrupole Systems offers superior selectivity because of double mass filtering in the mass analyzer. This selectivity greatly assists the screening for unknown and targeted analytes in complex samples such as extracts of fruits and vegetables. Figure 4 shows an example of MRM detection of 5.8 μg/kg Metazachlor in apricot extract. No matrix signals interfere with the quantitation of the identified contaminant because of the high selectivity of MRM detection. MRM Sensitivity QTRAP and triple quadrupole Systems operated in MRM allow quantitation of trace levels with superior sensitivity. The Limits of Detection (LOD) of 75 pesticides are summarized in Table 1. Figure 5 gives example chromatograms of selected pesticides spiked into a tomato extract at a concentration of 1 ng/ml. All MRM transitions in the presented screening method were detected using a short dwell time of 5 ms. The Linear Accelerator (LINAC ) Collision Cell enables the use of such low dwell times without loss in sensitivity. Identification using MRM ratios Despite the high selectivity of MRM detection, there is always a risk of false positive or negative findings due to interfering matrix signals. Accordingly, screening results have to be confirmed. Typically a second MRM is monitored per analyte and the ratio of quantifier to qualifier transition is calculated for each unknown sample and compared to the MRM ratio of standards. Various guidelines such as the European Commission Decision 22/657/EC define MRM ratio tolerance levels for identification (Table 2). 17 Figure 4 shows an example of positive identification based on MRM ratio of 5.8 μg/kg Metazachlor in apricot extract. XIC of +MRM (15 pairs): 167.2/8... 1.e4 5 1.6 Max. 1.2e4 cps. XIC of +MRM (15 pairs): 312.2/... Max. 7337.5 cps. 7338 5 3.2 XIC of +MRM (15 pairs): 23.2/1... Max. 1.5e4 cps. 1.5e4 5 4.4 XIC of +MRM (15 pairs): 223.2/1... Max. 1.7e4 cps. 1.7e4 5. Cyromazine Imazaquin Metamitron Acetamiprid XIC of +MRM (15 pairs): 192.2/1... Max. 9.2e4 cps. 9.2e4 5.e4 5.5 XIC of +MRM (15 pairs): 216.1/1... Max. 1.6e4 cps. 1.6e4 XIC of +MRM (15 pairs): 388.2/3... Max. 1.4e4 cps. 1.4e4 5 XIC of +MRM (15 pairs): 254.2/1... Max. 1.1e5 cps. 1.e5 5.e4 Atrazine Dimethomorph (E+Z isomer) Irgarol Carbendazim 7.5 8.3 8.5 9. +EPI (167.2) Charge (+) CE (3... 68. Figure 5. Example chromatograms of pesticides spiked at 1 ng/ml into a tomato extract (left) with corresponding Enhanced Product Ion spectra at the same concentration (right) Identification using QTRAP MS/MS Spectra and Library Searching t e n s it y, In c ptes n s it y, In c pt es n s it y, I cn pt es n s ity, I cn pt es n s it y, In c ptes n s it y, In c pt es n s it y, In c ptes n s ity, c p s Max. 3.e6 cps +EPI (312.2) Charge (+) CE (3... Max. 1.8e6 cps. As an alternative, identification can be performed using full scan MS/MS experiments and library searching to compare the unknown with a standard spectrum. As shown in Figure 2, identification based on MS/MS spectra is more efficient using QTRAP Systems because of higher sensitivity and faster 167.3 85.1 6.1 125.2 86.2 181.2 199.2 153.2 224.3 171.2 252.4 312.2 +EPI (23.2) Charge (+) CE (3... Max. 2.9e6 cps. 14.2 175.3 89. 23.3 +EPI (223.2) Charge (+) CE (3... Max. 9.3e5 cps. 126.1 9.1 99.1 223.3 56. 181.3 +EPI (192.2) Charge (+) CE (3... Max. 1.3e6 cps. 1.3e6 16.2 15.1 132.1 192.2 +EPI (216.1) Charge (+) CE (3... Max. 2.2e6 cps. 216.2 68. 174.2 14. 79.1 146.2 +EPI (388.2) Charge (+) CE (3... Max. cps. 165.2 31.2 388.3 139.1 258.2 242.2 273.2 114.1 28.3 +EPI (254.2) Charge (+) CE (3... Max. 6.2e6 cps. 6.e6 4.e6 18.1 91.1 198.2 156.2 176.3 254.3 p 5

scanning in comparison to QqQ. In addition, QTRAP Systems provide more characteristic MS/MS spectra and lower cycle times than traditional ion traps. Figure 5 shows the Enhanced Product Ion (EPI) spectra of selected pesticides at a concentration of 1 ng/ml spiked into tomato extract. These spectra were acquired automatically using IDA with a survey scan containing 75 MRM transitions. The spectra shown illustrate that EPI spectra are highly selective and they contain the complete molecular fingerprint of the analyte for identification. Also EPI spectra have superior sensitivity and can be searched against existing mass spectral libraries. Table 2. MRM ratio tolerances of 22/657/EC 17 Relative MRM Intensity Maximum Tolerance >.5 ± 2%.2.5 ± 25%.1.2 ± 3% <.1 ± 5% This higher degree of Identification can be used to minimize the potential risk of false positive and false negative detection based on MRM ratio calculation. An example is presented in Figure 6. The pesticides Thiabendazole and Imazalil were identified and quantified above the MRL in a banana extract. However, only Thiabendazole was identified by its MRM ratio. Imazalil could not be identified based on the MRM ratio of.5 and an MRM ratio tolerance of.28-.46. In comparison, the EPI spectrum rich in fragment ions searched against a library clearly identifies the presence of both contaminants in the banana sample and the false negative detection for Imazalil was avoided. Figure 7 shows another example demonstrating the advantage of MS/MS library searching for identification. The MRM ratio clearly identifies the presence of Metalaxyl in strawberry extract. The corresponding EPI spectrum and the library search result clearly identify a false positive finding. Metalaxyl was not present in this strawberry sample. Reproducibility of quantitation and identification Performing identification analysis using an automatically triggered EPI scan instead of MRM ratio calculation allows a reduction of the number of monitored MRM transitions. Only 75, instead of 15, MRM transitions have to be monitored using the QTRAP screening method for detection and identification of 75 pesticides. The reduction in MRM transitions resulted in a shorter cycle time of the MRM experiment (QTRAP : 1.1 s and QqQ: 1.5 s) although a longer dwell time was used per transition. This shorter cycle improved reproducibility since more data points were acquired across the LC peaks. A mix of pesticides was spiked into tomato extract and analyzed 1 times using the developed QqQ and QTRAP methods. Data presented in Table 3 and example spectra in Figure 8 clearly show the better reproducibility of identification based on mass spectral library searching of EPI spectra in comparison to MRM ratio calculation. XIC of +MRM (22 pairs): 22./1... 4.e4 3.e4 2.e4 6.3 Max. 4.e4 cps. XIC of +MRM (22 pairs): 22./1... Max. 5.8e4 cps. 5.8e4 4.e4 2.e4 6.5 +EPI (22.) Charge (+) CE (3... Max. 4.1e6 cps. 4.e6 3.e6 Std Thiabendazole 1ppb (ratio =.63) 135μg/kg Thiabendazole in banana (1x diluted) (ratio =.55) EPI FIT = 92.9% 131.2 65. 92.214.2 143.2 175.2 22.2 6 8 1 12 14 16 18 2 22 24 In te n sity, c p s In te n sity, cp s XIC of +MRM (22 pairs): 297./... Max. 73 cps XIC of +MRM (22 pairs): 297./... Max. 1 cps. Figure 6. Quantitation of 135 μg/kg Thiabendazole and 15.2 μg/kg Imazalil in banana extract with positive identification of Thiabendazole (.55 with tolerance of.5-.76) and negative identification of Imazalil (.5 with tolerance of.28-.46). This false negative can be avoided by identification based on Enhanced Product Ion (EPI) scan and library search XIC of +MRM (15 pairs): 28.2/2... 1.6e4 1.4e4 1.2e4 8 6 4 2 In te n s ity, cp s Max. 1.7e4 cps. XIC of +MRM (15 pairs): 28.2/... Max. 21 cps. 2 15 1 5 7.7 Std Metalaxyl 1ppb (ratio =.74) 6.2 7.9 13.3μg/kg Metalaxyl in strawberry (1x diluted) (ratio =.72) Figure 7. Quantitation of 13.3 μg/kg Metalaxyl in strawberry extract. This result was identified by MRM ratio calculation (.72 with tolerance of.59-.89), but the library search clearly indicates a false positive finding 73 6 4 2 1 5 +EPI (297.) Charge (+) CE (3... Max. 8.e5 cps. 8.e5 6.e5 Std Imazalil 1ppb (ratio =.37) 9.1 9.3 15.2μg/kg Imazalil in banana (1x diluted) (ratio =.5) EPI FIT = 9.4% 4.e5 69. 2.e5 81.1 19.2 141.2 21. 255.2 5 1 15 2 25 3 35 +EPI (28) CE (35) CES (15):... 6.6e7 6.e7 5.e7 4.e7 3.e7 2.e7 1.e7 15.1 16.1 145.1 192.2 132.1 148.1 162.2 22.3 134. 248.3 28.3 Max. 6.6e7 cps. +EPI (28.2) Charge (+) CE (3... Max. 3.3e5 cps. 3.3e5 3.e5 2.5e5 2.e5 1.5e5 1.e5 5.e4 148.3 134. 192.4 22.3 22.3 Library spectrum Metalaxyl 28.3 Unknown spectrum FIT for Metalaxyl = 23.6% p 6

Table 3. Reproducibility of identification based on MRM ratio calculation and mass spectral library search based on repeat injections of pesticides spiked into tomato extract (n=1) Pesticide MRM ratio %CV Library FIT %CV Azoxystrobin.22 8.9 88.8 1.8 Dimethomorph.76 8.8 95.7.5 Flufenacet.9 11.8 91.2 1.3 Imazalil.52 13.6 94.7 1.3 Imidacloprid.39 25.3 78.1 5.5 Methabenzthiazuron.47 11.7 95.9 1. Thiabendazole.66 4.9 91.2 2.4 +EPI (297.)... 2.3e6 69. 81.1 Figure 8. Repeated analysis of banana extract with detection of 15.2 μg/kg Imazalil Library search FIT values of Enhanced Product Ion spectra indicate the high reproducibility of MS/MS identification using QTRAP LC/MS/MS Systems Summary 21.1 255.1 Max. 2.3e6 cps. 297.1 1 2 3 +EPI (297.)... Max. 2.2e6 cps. 2.2e6 69. 255.1 81. 141.2 21.1 1 2 3 +EPI (297.)... 69. 21.1 255.1 81.1 19.2 164.3 Max. 2.1e6 cps. 1 2 3 +EPI (297.)... Max. 2.2e6 cps. Multiple Reaction Monitoring (MRM) using triple quadrupole and QTRAP LC/MS/MS Systems is widely used for targeted analyte screening and quantitation because of its unmatched selectivity and sensitivity. Double mass filtering in the mass analyzer allows for the best Signal-to-Noise (S/N) and thus the lowest Limits of Detection (LOD) with excellent reproducibility and linear range. In many applications, such as food, environmental and forensic testing, additional mass spectrometric information is required by guidelines for identification of detected analytes. This identification is typically performed by detecting two MRM transitions per compound and the calculation of an MRM ratio. 69. 81.1 21.1255.1 1 2 3 +EPI (297.)... 1. inj. 96.4% 3. inj. 96.6% 5. inj. 93.4% 2.3e6 69. 19.2 255.1 141.2 21.1 Max. 2.3e6 cps. 1 2 3 +EPI (297.)... Max. 2.2e6 cps. 6. inj. 94.1% 8. inj. 94.3% 1. inj. 94.8% 2.2e6 69.1 81.1 255.1 141.2 21.2 297.1 1 2 3 of even state-of-the art triple quadrupole mass spectrometers, such as AB SCIEX systems using the patented Linear Accelerator (LINAC ) Collision Cell. This results in a limitation of the number of compounds that can be screened for in a single injection and/or less reproducibility and accuracy due to an insufficient number of data points across the LC peaks. Here Hybrid Triple Quadrupole Linear Ion Trap (QTRAP ) Systems provide a novel workflow for the screening and identification of a multitude of targeted analytes by combining selective MRM detection with a highly sensitive MS/MS scan using Q3 as Linear Ion Trap. In Information Dependent Acquisition (IDA) experiments, the detection of an MRM signal above a specified threshold automatically triggers an Enhanced Product Ion (EPI) scan. These EPI spectra are as sensitive and selective as MRM signals and contain the complete molecular fingerprint because of precursor ion selection in Q1, product ion generation in the collision cell, and ion accumulation in the LIT. The rich product ion spectra are the result of the generation of fragment ions in the collision cell. These spectra can be searched against existing mass spectra libraries. The information saved into a full scan MS/MS spectrum allows identification with a higher degree of confidence minimizing the risk of potential false positive and negative detection. In addition, the improved cycle time for all confirmatory MRM transitions can be used to increase the dwell time of all other MRM transitions to improve S/N, resulting in better reproducibility and accuracy. Alternatively, additional compounds can be screened in the MRM survey scan. However, with an increasing demand to test for more and more contaminants in a larger variety of samples, the required number of MRM transitions exceeds the possibilities p 7

References 1 A. Schreiber Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Confirm the Presence of Pesticides in Water and Soil Samples Application Note AB SCIEX (26) 2 J. W. Hager: Rapid Commun. Mass Spectrom. 16 (22) 512-526 3 J. W. Hager and J. C. Y. Le Blanc: Rapid Commun. Mass Spectrom. 17 (23) 156-164 4 B. A. Collings et al.: J. Am. Soc. Mass Spectrom. 14 (23) 622-634 5 P. Marquet et al.: J. Chromatogr. B 789 (23) 9-18 6 C. A. Mueller et al.: Rapid Commun. Mass Spectrom. 19 (25) 1332-1338 7 S.M.R Stanley et al.: J. Chromatogr. B 836 (26) 1-14 8 H. D. Hernando et al.: Anal. Bioanal. Chem. 389/6 (27) 1815-1831 9 M. J. M. Bueno et al.: Anal. Chem. 79/24 (27) 9372-9384 1 M. Gros et al.: J. Chromatogr. A 1189 (28) 374-384 11 Hongying Gao et al.: Rapid Commun. Mass Spectrom. 21 (27) 3683-3693 12 R. D. Unwin et al.: Molecular and Cellular Proteomics 4/18 (25) 1134-1144 13 E. Ciccimaro et al.: Rapid Comm. Mass Spectrom. 2 (26) 3681-3692 14 L. Anderson et al.: Molecular and Cellular Proteomics 5/4 (26) 573-588 15 J. R. Whiteaker et al.: J. Proteome Research 6/1 (27) 3962-3975 16 M. Anastassiades et al.: J. AOAC Int. 86 (23) 412-431 17 European Commission Decision 22/657/EC For Research Use Only. Not for use in diagnostic procedures. 21 AB SCIEX. The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective owners. AB SCIEX is being used under license. Publication number: 11281-1 Headquarters International Sales 353 Hatch Drive Foster City CA 9444 USA For our office locations please call the division Phone 65-638-58 headquarters or refer to our website at www.absciex.com www.absciex.com/offices