Principles of Instrumental Analysis

Similar documents
1. Carrier gas supply. - Flow rate measurement

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

GAS CHROMATOGRAPHY. Mobile phase is a gas! Stationary phase could be anything but a gas

Chapter 27: Gas Chromatography

10/27/10. Chapter 27. Injector typically 50 C hotter than oven

Gas Chromatography. Vaporization of sample Gas-solid Physical absorption Gas-liquid Liquid immobilized on inert solid

Gas Chromatography. Introduction

Gas Chromatography (GC)

GC Instruments. GC Instruments - Sample Introduction

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

Gas chromatography. Advantages of GC. Disadvantages of GC

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

Introduction to Gas Chromatography

2401 Gas (liquid) Chromatography

Volatile organic compounds (VOCs):

Chromatography. Gas Chromatography

Chemistry Instrumental Analysis Lecture 28. Chem 4631

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

CHAPTER 6 GAS CHROMATOGRAPHY

Chapter 31 Gas Chromatography. Carrier Gas System

/Chapter 27.ppt

Gas Chromatography. Rosa Yu, David Reckhow CEE772 Instrumental Methods in Environmental Analysis CEE 772 #16 2

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian


Skoog/Holler/Crouch Chapter 26 Principles of Instrumental Analysis, 6th ed. CHAPTER 26

Biochemistry. Biochemical Techniques. 12 Gas Liquid Chromatography

Experiment 6 Simple and Fractional Distillation

Ch24. Gas Chromatography (GC)

How To Select the Correct GC Column. Simon Jones Application Engineer

Chapter 11 Conventional Gas Chromatography

CHEM 429 / 529 Chemical Separation Techniques

Understanding the Capillary GC Column: How to Choose the Correct Type and Dimension

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications

7 INSTRUMENTAL CHROMATOGRAPHY

Gas Chromatography CHEM Dr. Reem M. Alghanmi st term

GC Instruments. GC Instruments - Columns

Gas Chromatography. Chromatography Laboratory Course. Dr. Christian Jungnickel Chromatography Course GC September 2005

Selection of a Capillary

Instrumental Chemical Analysis

Chapter 1. Chromatography. Abdul Muttaleb Jaber

Partitioning. Separation is based on the analyte s relative solubility between two liquid phases or a liquid and solid.

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Luminescence transitions. Fluorescence spectroscopy

Introduction and Principles of Gas Chromatography

CHEM340 Tutorial 4: Chromatography

Chromatography and other Separation Methods

Principles of Gas- Chromatography (GC)

Selection of a Capillary GC Column

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

CEE 772: Instrumental Methods in Environmental Analysis

Column Selection. there is more to life than a boiling point column. Jaap de Zeeuw Restek Corporation, The Netherlands. Copyrights: Restek Corporation

CEE 772: Instrumental Methods in Environmental Analysis

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

Gas Chromatography (Chapter 2 and 3 in The essence of chromatography)

Speakers. Moderator. John V Hinshaw GC Dept. Dean CHROMacademy. Tony Taylor Technical Director CHROMacademy. Dave Walsh Editor In Chief LCGC Magazine

Course goals: Course goals: Lecture 1 A brief introduction to chromatography. AM Quality parameters and optimization in Chromatography

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.

Understanding Gas Chromatography

LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction.

High Pressure/Performance Liquid Chromatography (HPLC)

The Suite for Environmental GC Analysis

6.1 Revision The following questions cover the important concepts that you should have understood in the first year introduction to chromatography.

High Performance Liquid Chromatography

High Performance Liquid Chromatography

CEE 772: Instrumental Methods in Environmental Analysis

Disadvantage: Destructive Technique once analyzed by GC, the sample is lost

Gas Chromatography (GC)

Chapter content. Reference

Headspace Technology for GC and GC/MS: Features, benefits & applications

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

Title Experiment 7: Gas Chromatography and Mass Spectrometry: Fuel Analysis

Product Brief. - Hydrocarbons alkanes, alkenes, alkynes, dienes including natural gas, refinery gas, liquified petroleum gas

Lab 3 Guide: Gas Chromatography (GC) (Sept 8-14)

Instrumentation. Components of a gas chromatograph

Instrumental Analysis II Course Code: CH3109. Chromatographic &Thermal Methods of Analysis Part 1: General Introduction. Prof. Tarek A.

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

INSTITUTE OF MEDICAL BIOCHEMISTRY FIRST FACULTY OF MEDICINE, CHARLES UNIVERSITY IN PRAGUE. Chromatography. in biochemistry

An Introduction to Gas Chromatography Mass Spectrometry

Prelab Reading Assignment: Laboratory Techniques in Organic Chemistry, 4 th Ed. Chapter 19

CHEMISTRY Unit 3, Area of Study 1: Chemical Analysis

Gas chromatography. Flow measurement. Flow control. Injection methods. Flow measurement

GAS CHROMATOGRAPHY (GC)

Course CHEM Chromatography

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

If you like us, please share us on social media. The latest UCD Hyperlibrary newsletter is now complete, check it out.

CHROMATOGRAPHY. The term "chromatography" is derived from the original use of this method for separating yellow and green plant pigments.

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers.

Chapter 27 Gas Chromatography

Gas Chromatography. A schematic diagram of a gas chromatograph

Ch.28 HPLC. Basic types of Liquid Chromatography Partition (LLC) Adsorption (LSC) Ion Exchange (IC) Size Exclusion (SEC or Gel Chromatography)

CHAPTER 1. Introduction, Chromatography Theory, and Instrument Calibration

Open Column Chromatography, GC, TLC, and HPLC

2] The plate height in chromatography is best described as 2

Trajan SGE GC Columns

Chromatography and Functional Group Analysis

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Chromatographic Methods

Introduction to Capillary GC

Choosing the Correct GC Column Dimensions and Stationary Phase

Transcription:

Principles of Instrumental Analysis Chapter 27 Gas Chromatography Gas Chromatography (GC): vaporized analytes (solutes) are partitioned between a mobile gaseous phase and a liquid or a solid stationary phase held in a column. The mobile phase does not interact with molecules of the analytes (solutes). Gas-Liquid Chromatography (GLC): partition between mobile phase and liquid phase immobilized ( 固定化 ) on the surface of an inert solid packing or on the walls of a capillary tubing. Gas-Solid Chromatography (GSC): physical adsorption 27A Principles of GLC Due to compressibility of gaseous mobile phases, the mathematical relationships obtained in Chapter 26 need minor modification. Retention volume (V g ) instead of retention time (t R ): V g = (K/ S ) x (273/T c ) T c : column temperature S : density of liquid stationary phase 1

V R = t R F, V M = t M F F: average volumetric flow rate F = F m x (T c /T) x (P - P H2O )/P T c : column temperature (T: ambient) P: gas pressure at column end (ambient) P H2O : vapor pressure of water Corrected retention volume: V R0 = jt R F, V M0 = jt M F j: compressibility factor.. (27-4) Specific retention volume: V g = (V R0 -V M0 )/m S x 273/T c = jf(t R t M )/m S x 273/T c m S : mass of the stationary phase V g = jft M k/m S x 273/T c retention factor k = (t R -t M )/t M = V M0 k/m S x 273/T c = KV S /m S x 273/T c k = KV S /V M (K: distribution constant) S = m S /V S V g =K/ S x 273/T c (from eq. 27-4) At a given temperature, V g depends only on the distribution constant (K) and density ( S ) of stationary phase (liquid). van Deemter plot FIGURE 26-8 Effect of mobile-phase flow rate on plate height for (a) LC and (b) GC. Note very different flow rate and plate height scales. 1) The minimum for LC usually occurs at flow rates well below those for GC. 2) Plates heights (H) for LC columns are an order of magnitude or more smaller than those encountered with GC columns. Ch26 An Introduction to Chromatographic Separations P.772 2

27B Instruments for GLC FIGURE 27-1 Block diagram of a typical gas chromatograph. FIGURE 27-2 A soap-bubble flow meter. P.790 http://www.ece.vt.edu/news/ar08/gcmatrix.html http://departments.kings.edu/chemistry/facilit y%20tour/organic/organic%20gc.htm 3

FIGURE 27-3 A set of micro-syringes for sample injection. P.791 FIGURE 27-4 Cross-sectional view of a microflash vaporizer direct injector. P.791 4

Gas sampling valve for quantitative work. FIGURE 27-5 A rotary sample valve: valve position (a) is for filling the sample loop ACB; position (b) if for introduction of sample into column. P.792 5

Column configuration: constructed of fused silica or stainless steel; formed as coils (diameter: 10~30 cm); housed in a thermostatted oven. Packed column: 1~5 m. Open tubular (capillary) column: ~100 m. Column temperature is an important variable: equal or slightly above the average boiling point of a sample reasonable elution time (2-30 min). In general, optimal resolution is associated with minimal temperature; the cost of lowered temperature is an increase in elution time [Figure 27-7(a)-(b)]. FIGURE 27-6 Fused-silica capillary columns. P.792 FIGURE 27-7 Effect of temperature on gas chromatograms: (a) isothermal at 45 ; (b) isothermal at 145 ; (c) programmed at 30 to 180. P.793 6

27B-4 Detection Systems Characteristics of the ideal detectors: 1)Adequate sensitivity. (10-8 ~ 10-15 g solute/s) 2)Good stability and reproducibility. 3)A linear responses to solutes (several orders of magnitude). 4)A temperature range from r.t. to at least 400 o C. 5)A short response time independent of flow rate. 6)High reliability and ease of use. 7)Similarity in response toward all solutes. 8)The detector should be non-destructive. TABLE 27-1 Typical Gas Chromatographic Detectors P.793 FID responds to number of carbon atom: A mass-sensitive rather than concentration-sensitive. Suitable for organic samples: Functional groups (carbonyl, alcohol, halogen, and amine) yield fewer ions or none at all in a flame. Insensitive to non-combustible gases (H 2 O, CO 2, SO 2, CO, noble gases, and NO x ) Sensitivity: ~10-13 g/s. Linear response range: ~10 7 Low noise. Disadvantages: destructive FIGURE 27-8 A typical flame ionization detector (FID). P.794 7

Thermal Conductivity Detector (TCD): Temperature at constant electrical power depends on the thermal conductivity of the surrounding gas. Suitable to both organic and inorganic species. Simplicity & non-destructive. Low sensitivity: ~10-8 g/s. Linear response range: ~10 5 FIGURE 27-9 Schematic of (a) a TCD cell, and (b) an arrangement of two sample detector cells and two reference detector cells. P.794 Nickel-63 Ionization of N 2 FIGURE 27-10 Schematic diagram of an Electron-Capture Detector (ECD). Electronegative functional groups tend to capture electrons decrease current. Selectively responds to halogen-containing organic compounds: for detection of environmental samples: pesticides, polychlorinated biphenyls. Selective in response: high sensitivity to halogens, peroxides, quinones, and nitro groups; insensitive to amines, alcohols, and hydrocarbons. Highly sensitive and non-destructive. But only two orders in linear response. P.795 8

FIGURE 27-11 Diagram of a Hall electrolytic conductivity detector. P.796 AED is an element-selective detector. FIGURE 27-12. An Atomic Emission Detector (AED) for GC. The plasma is sufficiently energetic to atomize all the elements in a sample and to excite their characteristic atomic emission spectra. P.797 9

C: 198 nm FIGURE 27-13(a). Chromatogram for a gasoline sample containing a small amount of MTBE and several aliphatic alcohols: monitoring a carbon emission line. P.798 O: 777 nm FIGURE 27-13(b) Chromatogram for a gasoline sample containing a small amount of MTBE and several aliphatic alcohols: monitoring an oxygen emission line. P.798 10

Mass Spectrometry Detectors FIGURE 27-14 Schematic of a typical capillary GC/MS system. The effluent from the GC is passed into the inlet of the mass spectrometer, where the molecules in the gas are fragmented, ionized, analyzed, and detected. P.799 Time, min FIGURE 27-15(a). Typical outputs for a GC/MS system. In (a), the total ion (current) chromatogram were 1, N-nitrosodimethylamine, 2, bis(2- chloroethyl)ether, 3, bis(2-chloroisopropyl)ether, 4, N-nitrosodi-n-propylamine, and 5, bis(2-chloroethoxy)methane. 1 2 3 4 5 MW= 74.08 MW= 143.012 MW= 171.06 MW= 130.188 MW= 173.038 P.799 11

m/z 74 FIGURE 27-15(b) Typical outputs for a GC/MS system. In (b), the mass chromatogram at m/z = 74 is shown. The peak is due to the parent ion of n-nitrosodimenthylamine (C 2 H 6 N 2 O). MW= 74.08 P.799 m/z 93 FIGURE 27-15(c). Typical outputs for a GC/MS system. A selected-ion chromatogram m/z = 93 is shown in (c). Peaks 2 and 5 give a response at this m/z value due to fragmentation products. MW= 143.012 MW= 173.038 P.799 12

27C GC Columns and Stationary Phases TABLE 27-2 Properties and Characteristics of Typical GC Columns P.802 FIGURE 27-16 A photomicrograph of a diatomaceous earth ( 矽藻土 ). Magnification 5000X. P.802 13

27C-3 Adsorption on Column Packings or Capillary Walls: Polar silanol silanization Acid washing before silanization removes metal oxide impurities. P.803 27C-4 Stationary phase (immobilized liquid) requires: (1) low volatility, (2) thermal stability, (3) chemical inertness, and (4) solvent characteristics (k and ). TABLE 27-3. Some Common Liquid Stationary Phases for GLC P.804 14

Compatibility (solubility) of an analyte with stationary phase leads to reasonable residence time (k). Like dissolves like. Like refers to the polarities of the analyte and the immobilized liquid (stationary phase). In GC the polarity of stationary phase should match with that of the sample components. The order of elution is determined by the boiling point of the analytes. E coh = H vap RT (Cohesive energy) CED = E coh /V (Cohesive Energy Density) = (CED) 1/2 (Solubility Parameter) 溶劑之溶解參數 ( ) (cal/ml) 1/2 n-hexane n-octane Cyclohexane Benzene Toluene Acetone Methylethylketone (MEK) Methyl acetate Ethyl acetate Butyl acetate Tetrahydrofuran (THF) Methanol Ethanol n-butanol Water 29 7.3 7.6 8.2 9.2 8.9 9.8 9.3 9.6 9.1 8.5 9.2 14.5 12.7 11.4 23.4 Polarity Liquid Stationary Phase fro GLC Hydrocarbon Polysiloxane bonded or cross-linked R= methyl: Polydimethyl siloxane R= methyl, phenyl: x% Phenyl-polydimethyl siloxane R= methyl, trifluoropropyl: x% Trifluoropropyl-polydimethyl siloxane R= methyl, cyanopropyl: x% Cyanopropyl-polydimethyl siloxane Polyethylene glycol: H-(OCH 2 CH 2 ) n -OH Polyester: H-(O-RO-CO-R -CO) n -OH Analytes: Polar: alcohols, acids, amines. Medium polar: ethers, ketones, aldehydes. Non-polar: saturated hydrocarbons. The polarity of the stationary phase should match that of the analytes. When the match is good, the elution order is determined by the boiling point of the analytes P.804 15

FIGURE 27-17(a)-(c) Typical chromatograms from open tubular columns coated with (a) polydimethyl siloxane; (b) 5% (phenyl methyldimethyl) siloxane; (c) 50% (phenyl methyldimethyl) siloxane. P.805 FIGURE 27-17(d)-(f) (d) 50% poly (trifluoropropyl-dimethyl) siloxane; (e) polyethylene glycol; (f) 50% poly (cyanopropyl-dimethyl) siloxane. P.805 16

Chiral Stationary Phases (CSP): chiral liquid as stationary phase for separation of enantiomers. Thalidomide was sold in a number of countries across the world from 1957 until 1961 when it was withdrawn from the market after being found to be a cause of birth defects in what has been called "one of the biggest medical tragedies of modern times". [4 It is not known exactly how many worldwide victims of the drug there have been, although estimates range from 10,000 to 20,000. [5] R 1 ' R 1 R 1 R 4 ' R 2 ' R 2 R 4 R 4 R 2 (S)-thalidomide (R)-thalidomide R 3 ' CSP (R) (S) enantiomer Three-point interaction R 3 R 3 27D Applications of GC 1)Perform separations 2)Analysis: a) Qualitative analysis: retention times or volumes. * Selectivity factors ( ): * The Retention Index (I): The retention index scale is based on normal alkanes: 100 x no. carbon. b) Quantitative analysis: peak heights or areas. 17

FIGURE 27-18 Graphical illustration of the method for determining retention indexes for three compounds. Stationary phase: squalane. Temperature: 60. Retention indexes for normal alkane standards nonane and hexane are indicated. P.807 27E Advances in GC 1)High-speed GC: t R = L/u x (1 + k n ); V g = (K/ S ) x (273/T c ) 2)Miniaturized GC Systems FIGURE 27-19 High-speed chromatogram obtained with isothermal operation (30 ) for 27 s followed by a 35 /min temperature ramp to 90. P.808 18

FIGURE 27-20 Microfabricated columns (a) and chromatogram (b). The columns in (a) were 0.9-m-long spiral and serpentine channels. P.809 FIGURE 27-20(b) The mixture (b) was 1, acetone; 2, 2-butanone; 3, benzene; 4, trichloroethylene; 5, 2,5-dimethyl-furan; and 6, toluene. Air was used as the carrier gas with an outlet pressure of 0.5 atm. P.809 19

27F Gas-Solid Chromatography (GSC): physical adsorption packed or (porous-layer) open tubular columns (PLOT) 1) Molecular Sieves: aluminum silicates 2) Porous polymers: cross-linked polystyrenes FIGURE 27-21 Typical gas-solid chromatographic separations: (a) a 5 ft. 1/8 in. molecular sieve column; (b) a 30 m 0.53 mm PLOT column. C n = hydrocarbon with n carbons. P.810 20