Assessment of Accuracy and Precision

Similar documents
Experiment 20: Analysis of Vinegar. Materials:

2. MASS AND VOLUME MEASUREMENTS Mass measurement Analytical and standard laboratory balances Pre-lab Exercises

CHEM 334 Quantitative Analysis Laboratory

Experiment 8 Introduction to Volumetric Techniques I. Objectives

Density of an Unknown

Accuracy and Precision of Laboratory Glassware: Determining the Density of Water

Chesapeake Campus Chemistry 111 Laboratory

Chemistry Lab: Introduction to Measurement

Chemistry Calibration of a Pipet and Acid Titration

Chemistry 143 Acid Base Titration Dr. Caddell. Titrating Acid

Standardizing a Solution of Sodium Hydroxide. Evaluation copy

Volumetric Measurement Techniques. Technique #1 Use of a Burette. Technique #2 Use of a Pipette. Technique #3 Use of a Volumetric Flask

EXPERIMENT 30A1: MEASUREMENTS. Learning Outcomes. Introduction. Experimental Value - True Value. 100 True Value

+ H 2 O Equation 1. + NaOH CO 2 Na

Chemistry 143 Experiment #11 Acid Base Titration Dr. Caddell. Titrating Acid

Chemical Reactions: Titrations

#09 Investigating the Relationship between the Mass of a Liquid and its Volume Ken Lyle, St. John s School, Houston, TX

Titration with an Acid and a Base

NaOH (aq) + HCl (aq) NaCl (aq) + H 2 O (l)

Experiment #7. Titration of Vinegar

CHM111 Lab Titration of Vinegar Grading Rubric

Experiment: Titration

Practice Lab. Balances and calibration of volumetric tools

# 12 ph-titration of Strong Acids with Strong Bases

Basic Equipments and Instruments used in Chemistry laboratory: Balance: It is an instrument for measuring mass.

DETERMINATION OF THE SOLUBILITY PRODUCT OF GROUPII HYDROXIDES

Density of Aqueous Sodium Chloride Solutions

TRATION: ANALYSIS OF VINE

Learn to do quantitative titration reactions. Observe the mole ratios of several simple chemical reactions.

POGIL LAB EXERCISE 15 HOW DO YOU STANDARDIZE AN ACID AND BASE?

Acid-Base Titration. M M V a

Analysis of Hypochlorite in Bleach

Experiment Two Laboratory Balance: Mass Calculations

Why the fuss about measurements and precision?

Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE

Acidity of Beverages Lab

EXPERIMENT 12B: TITRATION OF AN UNKNOWN ACID INTRODUCTION

Comparing the MICROLAB 600 to Volumetric Glassware and Air Displacement Pipettes. June 2011

Experiment 1. Determination of the Density of Water and an Unknown Solid Sample

O H 3 O 1 1 A. O 1 1 OH (K w

The Synthesis and Analysis of Aspirin

Ascorbic Acid Titration of Vitamin C Tablets

Acid-Base Titration. Evaluation copy

Safety Note: Safety glasses and laboratory coats are required when performing this experiment

PART I: MEASURING MASS

Acid-Base Titration. Sample

Experiment 7: ACID-BASE TITRATION: STANDARDIZATION OF A SOLUTION

ANALYTICAL CHEMISTRY. Basic Skills in Chemical Laboratory

Density of Aqueous Sodium Chloride Solutions

CHEM Practice to be done before the lab. Experiment 9 Introduction to Volumetric Techniques II. Objectives

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Titration of HCl with Sodium Hydroxide

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15

Synthesis of Benzoic Acid

Experimental Procedure. Lab 406

In this lab you are asked to make a series of volume and temperature measurements and record the number of significant figures in each measurement.

Chemistry 101 Experiment 12B-ANALYSIS OF COMMERCIAL VINEGARS

CALIBRATION OF LABORATORY GLASSWARE: BUTYROMETER, PIPETTES, LACTOMETERS, THERMOMETER AND BURETTES

#13 ph-titration of Weak Acids with Strong Bases

The Thermodynamics of the Solubility of Borax

Partner: Alisa 1 March Preparation and Properties of Buffer Solutions

Introduction to Chemistry Techniques Prelab (Week 1) 2. Determine the number of significant figures in each of the following numbers.

Laboratory Techniques for Small Distilleries

Experiment 7A ANALYSIS OF BRASS

NOTE: YOU WILL BE USING THIS SOLUTION IN BOTH, THIS EXPERIMENT AND EXP 12B. IF YOU WASTE THE SOLUTION YOU MAY RUN OUT BEFORE YOU HAVE FINISHED EXP 12B

Percentage of Acetic Acid in Vinegar

Measurements, Sig Figs and Graphing

Experiment #10: Analysis of Antacids

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 03: EXPERIMENTAL ERROR

PreLAD: b. KHP is a monoprotic acid, what are the number of moles of ionizable H + in the approximately 0.25 g of KHP?

Basic Equipment. 2D.1 Equipment for Measuring Mass

Experimental Procedure

Identification Of The Common Laboratory Glassware, Pipettes And Equipment. BCH 312 [Practical]

Stoichiometry Lab Final this is not a proper title, so change it to be more descriptive

6 Acid Base Titration

Introduction to Small Scale Chemistry

Calibration of Volumetric Glassware. Prepared by Allan Fraser May 2016 APPLICATION Note 1

This chapter contains detailed directions for performing a variety of chemical

In this laboratory exercise we will determine the percentage Acetic Acid (CH 3 CO 2 H) in Vinegar.

Acid-Base Titration Acetic Acid Content of Vinegar

Titration of Vinegar

Partner(s) NONE USE OF THE ANALYTICAL BALANCE

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 03: EXPERIMENTAL ERROR

Redox Lab. By Maya Parks. Partner: Allison Schaffer 5/21/15. Abstract:

SOLVOLYSIS OF tert-butyl CHLORIDE: TESTING A MECHANISM

MEASUREMENT: PART II

MICROPIPETTE CALIBRATIONS

Proper Weighing, Reagent Handling, Volumetric Measurement and Pipetting Techniques Workshop

Name: Date: AP Chemistry. Titrations - Volumetric Analysis. Steps for Solving Titration Problems

Chemistry 119: Experiment 6. Sampling and Analysis of a Solid Drain Cleaner

Experiment 2: Analysis of Commercial Bleach Solutions

Titration of an Unknown Acid

Comparing the Microlab 600 to Volumetric Glassware and Air Displacement Pipettes. January 2013

Standardization of a Primary Standard & Determination of Concentration by Acid-Base Titration

Practice Examination #1

Laboratory Measurements and Procedures

ph Titration Curves You will need about 10 ml of ~6M NaOH solution and about 40 ml of each acid solution and about 1.5 grams of KHP.

Pre-lab: Read section 9.9 (pages ) on acid-base titrations in the textbook. Complete the attached pre-lab by Tuesday, June 2.

Chemical Kinetics. Reaction rate and activation energy of the acid hydrolysis of ethyl acetate LEC 05. What you need: What you can learn about

Partner: Judy 29 March Analysis of a Commercial Bleach

Transcription:

2 chapter Assessment of Accuracy and Precision S.S. Nielsen, Food Analysis Laboratory Manual, Food Science Texts Series, DOI 10.1007/978-1-4419-1463-7_2, Springer Science+Business Media, LLC 2010 9

Chapter 2 Assessment of Accuracy and Precision 11 INTRODUCTION Background Volumetric glassware, mechanical pipettes, and balances are used in many analytical laboratories. If the basic skills in the use of this glassware and equipment are mastered, laboratory exercises are easier, more enjoyable, and the results obtained are more accurate and precise. Measures of accuracy and precision can be calculated based on the data generated, given the glassware and equipment used, to evaluate the skill of the user as well as the reliability of the instrument and glassware. Determining mass using an analytical balance is the most basic measurement made in an analytical laboratory. Determining and comparing mass is fundamental to assays such as moisture and fat determination. Accurately weighing reagents is the first step in preparing solutions for use in various assays. Accuracy and precision of the analytical balance are better than for any other instrument commonly used to make analytical measurements, provided the balance is properly calibrated, and the laboratory personnel use proper technique. With proper calibration and technique, accuracy and precision are limited only by the readability of the balance. Repeatedly weighing a standard weight can yield valuable information about the calibration of the balance and the technician s technique. Once the performance of the analytical balance and the technician using it has been proven to be acceptable, determination of mass can be used to assess the accuracy and precision of other analytical instruments. All analytical laboratories use volumetric glassware and mechanical pipettes. Mastering their use is necessary to obtain reliable analytical results. To report analytical results from the laboratory in a scientifically justifiable manner, it is necessary to understand accuracy and precision. A procedure or measurement technique is validated by generating numbers that estimate their accuracy and precision. This laboratory includes assessment of the accuracy and precision of automatic pipettors. An example application is determining the accuracy of automatic pipettors in a research or quality assurance laboratory, to help assess their reliability and determine if repair of the pipettors is necessary. Laboratory personnel should periodically check the pipettors to determine if they accurately dispense the intended volume of water. To do this, water dispensed by the pipettor is weighed, and the weight is converted to a volume measurement using the appropriate density of water based on the temperature of the water. If replicated volume data indicate a problem with the accuracy and/or precision of the pipettor, repair is necessary before the pipettor can be reliably used again. It is generally required that reported values minimally include the mean, a measure of precision, and the number of replicates. The number of significant figures used to report the mean reflects the inherent uncertainty of the value, and it needs to be justified based on the largest uncertainty in making the measurements of the relative precision of the assay. The mean value is often expressed as part of a confidence interval (CI) to indicate the range within which the true mean is expected to be found. Comparison of the mean value or the CI to a standard or true value is the first approximation of accuracy. A procedure or instrument is generally not deemed inaccurate if the CI overlaps the standard value. Additionally, a CI that is considerably greater than the readability indicates that the technician s technique needs improvement. In the case of testing the accuracy of an analytical balance with a standard weight, if the CI does not include the standard weight value, it would suggest that either the balance needs calibration or that the standard weight is not as originally issued. Accuracy is sometimes estimated by the relative error (%E rel ) between the mean analysis value and the true value. However, %E rel only reflects tendencies, and in practice is often calculated even when there is no statistical justification that the mean and true value differ. Also, note that there is no consideration of the number of replicates in the calculation of %E rel, suggesting that the number of replicates will not affect this estimation of accuracy to any large extent. Absolute precision is reflected by the standard deviation, while relative precision is calculated as the coefficient of variation (CV). Calculations of precision are largely independent of the number of replicates, except that more replicates may give a better estimate of the population variance. Validation of a procedure or measurement technique can be performed, at the most basic level, as a single trial validation, as is described in this laboratory that includes estimating the accuracy and precision of commonly used laboratory equipment. However, for more general acceptance of procedures, they are validated by collaborative studies involving several laboratories. Collaborative evaluations are sanctioned by groups such as AOAC International, AACC International, and the American Oil Chemists Society (AOCS). Such collaborative studies are prerequisite to procedures appearing as approved methods in manuals published by these organizations. Reading Assignment Literature on how to properly use balances, volumetric glassware, and mechanical pipettes.

12 Chapter 2 Assessment of Accuracy and Precision Nielsen, S.S. 2010. Introduction to food analysis. Ch. 1, in Food Analysis, 4th ed. S.S, Nielsen (Ed.), Springer, New York. Smith, J.S. 2010. Evaluation of analytical data. Ch. 4, in Food Analysis, 4th ed. S.S. Nielsen (Ed.), Springer, New York. Objective Familiarize, or refamiliarize, oneself with the use of balances, mechanical pipettes, and volumetric glassware, and assess accuracy and precision of data generated. Principle of Method Proper use of equipment and glassware in analytical tests helps ensure more accurate and precise results. Supplies 1 Beaker, 100 ml 1 Beaker, 20 or 30 ml 1 Beaker, 250 ml Buret, 25 or 50 ml Erlenmeyer flask, 500 ml Funnel, approximately 2 cm diameter (to fill buret) Mechanical pipettor, 1000 ml, with plastic tips Plastic gloves Ring stand and clamps (to hold buret) Rubber bulb or pipette pull-up Standard weight, 50 or 100 g Thermometer, to read near room temperature Volumetric flask, 100 ml 2 Volumetric pipettes, one each of 1 and 10 ml Equipment Notes Analytical balance Top loading balance Before or during the laboratory exercise, the instructor is encouraged to discuss the following: (1) Difference between dispensing from a volumetric pipette and a graduated pipette, (2) difference between markings on a 10-ml versus a 25- or 50-ml buret. PROCEDURES (Record data in tables that follow.) 1. Obtain ~400 ml deionized distilled (dd) H 2 O in a 500-ml Erlenmeyer flask for use during this laboratory session. Check the temperature of the water with a thermometer. 2. Analytical balance and volumetric pipettes. (a) Tare a 100-ml beaker, deliver 10 ml of water from a volumetric pipette into the beaker, and record the weight. Repeat this procedure of taring the beaker, adding 10 ml, and recording the weight, to get six determinations on the same pipette. (Note that the total volume will be 60 ml.) (It is not necessary to empty the beaker after each pipetting.) (b) Repeat the procedure as outlined in Step 2a but use a 20- or 30-ml beaker and a 1.0-ml volumetric pipette. Do six determinations. 3. Analytical balance and buret. (a) Repeat the procedure as outlined in Step 2a, but use a 100-ml beaker, a 50-ml (or 25-ml) buret filled with water, and dispense 10 ml of water (i.e., tare a 100 ml beaker, deliver 10 ml of water from the buret into the beaker, and record the weight). (Handle the beaker wearing gloves, to keep oils from your hands off the beaker.) Repeat this procedure of taring the beaker, adding 10 ml, and recording the weight, to get six determinations on the buret. (Note that the total volume will be 60 ml.) (It is not necessary to empty the beaker after each addition.) (b) Repeat the procedure as outlined in Step 3a but use a 20- or 30-ml beaker and a 1.0-ml volume from the buret. Do six determinations. 4. Analytical balance and mechanical pipette. Repeat the procedure as outlined in Step 2a but use a 20- or 30-ml beaker and a 1.0-ml mechanical pipette (i.e., tare a 20- or 30-ml beaker, deliver 1 ml of water from a mechanical pipettor into the beaker, and record the weight). Repeat this procedure of taring the beaker, adding 1 ml, and recording the weight to get six determinations on the same pipettor. (Note that the total volume will be 6 ml.) (It is not necessary to empty the beaker after each pipetting.) 5. Total content (TC) versus total delivery (TD). Tare a 100-ml volumetric flask on a top loading balance. Fill the flask to the mark with water. Weigh the water in the flask. Now tare a 250-ml beaker and pour the water from the volumetric flask into the beaker. Weigh the water delivered from the volumetric flask. 6. Readability versus accuracy. Zero a top loading balance and weigh a 100-g (or 50-g) standard weight. Record the observed weight. Use gloves or finger cots as you handle the standard weight to keep oils from your hands off the weight. Repeat with the same standard weight on at least two other top loading balances, recording the observed weight and the type and model (e.g., Mettler, Sartorius) of balance used.

Chapter 2 Assessment of Accuracy and Precision 13 DATA AND CALCULATIONS Calculate the exact volume delivered in Parts 2 5, using each weight measurement and the known density of water (see Table 2-1). Using volume data, calculate the following indicators of accuracy and precision: mean, standard deviation, coefficient of variation, percent relative error, 95% confidence interval. Use your first three measurements for n = 3 values requested, and all six measurements for n = 6 values. 2-1 table Temperature ( C) Viscosity and Density of Water at Various Temperatures Density (g/ml) Viscosity (cps) Temperature ( C) Density (g/ml) Viscosity (cps) 20 0.99823 1.002 24 0.99733 0.9111 21 0.99802 0.9779 25 0.99707 0.8904 22 0.99780 0.9548 26 0.99681 0.8705 23 0.99757 0.9325 27 0.99654 0.8513 Data for Parts 2, 3, and 4: Rep Volumetric pipette Buret Mechanical pipettor 1 ml 10 ml 1 ml 10 ml 1 ml Wt. Vol. Wt. Vol. Wt. Vol. Wt. Vol. Wt. Vol. 1 2 3 4 5 6 n = 3 Mean S CV %E rel CI 95% n = 6 Mean S CV %E rel CI 95% Part 5 data: Water in flask= Water in beaker= Wt. Vol. Part 6 data: Balance 1 2 3 QUESTIONS 1. Theoretically, how are standard deviation, coefficient of variation, mean, percent relative error, and 95% confidence interval affected by: (1) more replicates, and (2) a larger size of the measurement? Was this evident in looking at the actual results obtained using the volumetric pipettes and the buret, with n = 3 versus n = 6, and with 1 ml versus 10 ml? (see table below) Standard deviation Coefficient of variation Mean Percent relative error 95% Confidence interval More replicates Type/Model of balance Theoretical Larger measurement Actual, with results obtained More replicates Wt. of standard weight Larger measurement 2. Why are percent relative error and coefficient of variation used to compare the accuracy and precision, respectively, of the volumes from pipetting/dispensing 1 and 10 ml with the volumetric pipettes and buret in Parts 2 and 3, rather than simply the mean and standard deviation, respectively? 3. Compare and discuss the accuracy and the precision of the volumes from the 1 ml pipetted/dispensed using a volumetric pipette, buret, and mechanical pipettor (Parts 2, 3, and 4). Are these results consistent with what would be expected? 4. If accuracy and/or precision using the mechanical pipettor are less than should be expected, what could you do to improve its accuracy and precision? 5. In a titration experiment using a buret, would you expect to use much less than a 10-ml volume in each titration? Would you expect your accuracy and precision to be better using a 10-ml buret or a 50-ml buret? Why?

14 Chapter 2 Assessment of Accuracy and Precision 6. How do your results from Part #5 of this lab differentiate to contain from to deliver? Is a volumetric flask to contain or to deliver? Which is a volumetric pipette? 7. From your results from Part #6 of this lab, would you now assume that since a balance reads to 0.01 g, it is accurate to 0.01 g? 8. What sources of error (human and instrumental) were evident or possible in Parts #2 4, and how could these be reduced or eliminated? Explain. 9. You are considering adopting a new analytical method in your lab to measure the moisture content of cereal products. How would you determine the precision of the new method and compare it to the old method? How would you determine (or estimate) the accuracy of the new method? Acknowledgment This laboratory was developed with inputs from Dr Charles E. Carpenter, Department of Nutrition and Food Sciences, Utah State University, Logan, UT. Resource Materials Nielsen SS (2010) Introduction to food analysis, Ch. 1. In: Nielsen SS (ed) Food analysis, 4th edn. Springer, New York Smith JS (2010) Evaluation of analytical data, Ch. 4. In: Nielsen SS (ed) Food Analysis, 4th edn. Springer, New York

Chapter 2 Assessment of Accuracy and Precision 15 Notes

http://www.springer.com/978-1-4419-1462-0