A STUDY ON TRAPPING CO 2 USING MOLECULAR SIEVE FOR 14 C AMS SAMPLE PREPARATION

Similar documents
COLLECTION SYSTEM WITH IRMS FOR 14 C AMS IN YAMAGATA UNIVERSITY, JAPAN

DEVELOPMENT OF A COMBUSTION SYSTEM FOR LIQUID OR GAS SAMPLES

SIMULATION STUDY FOR THE SEPARATION OF RARE ISOTOPES AT THE SEOUL NATIONAL UNIVERSITY AMS FACILITY

ON-LINE RADIOCARBON MEASUREMENTS OF SMALL SAMPLES USING ELEMENTAL ANALYZER AND MICADAS GAS ION SOURCE

CAPABILITIES OF THE NEW SUERC 5MV AMS FACILITY FOR 14 C DATING

that our measurement system can not only produce precise dates from very small amounts of materials, but that these dates are also accurate.

ACCELERATOR MASS SPECTROMETRY AT THE LUND PELLETRON ACCELERATOR

High-Pressure Volumetric Analyzer

RADIOCARBON ANALYSIS OF DIFFEREENT SAMPLES AT BINP AMS

DIRECT COUPLING OF AN ELEMENTAL ANALYZER AND A HYBRID ION SOURCE FOR AMS MEASUREMENTS

Hie-Joon Kim. Professor Emeritus Seoul National University. Experience. Representative Publications

Scientific Instruments for Radiocarbon Dating and Accelerator Mass Spectrometry

Solubility of carbon dioxide in aqueous solutions of 2-amino-2-ethyl-1,3-propanediol

Supplementary Text and Figures

Synthesis of Zeolite Composite Membranes for CO2 Separation

Name: Regents Review Quiz #1 2016

SOURCES OF RANDOM ERROR IN THE DEBRECEN RADIOCARBON LABORATORY

THE NEW 14 C ANALYSIS LABORATORY IN JENA, GERMANY. A Steinhof 1,2 G Adamiec 3 G Gleixner 1 GJ van Klinken 4 T Wagner 1

IUAC Pelletron Accelerator

Hydra-C Application Note: 1076

Supporting Information

RADIOCARBON DATING BEYOND 50,000 YEARS BY LIQUID SCINTILLATION COUNTING

DATING POLAR ICE BY 14C ACCELERATOR MASS SPECTROMETRY MICHAEL ANDREE, JUERG BEER, H P LOETSCHER, ERNST MOOR, H.ANS OESCHGER

Ethers in a Porous Metal-Organic Framework

Depth Distribution of H-3, C-14 and Co-60 in Decommissioning of the Biological Shielding Concrete of KRR-2

ADSORPTION AND DESORPTION OF CO ON SOLID SORBENTS

HIGH-PRECISION INTERCOMPARISON AT ISOTRACE

JOB VAN ROIJEN, KI,AAS VAN DER BORG, ARIE DE JONG

Gas Chromatography Separation of H 2 -D 2 -Ar Using Pd/K

A Survey of Modern Radiocarbon Dating

Advances in radiocarbon measurement of water samples

Thursday 10 January 2013 Morning

MAURICE ARNOLD, EDOUARD BARD, PIERRE MAURICE HELENE VALLADAS and J C DUPLESSY

Chemistry Lab Fairfax High School Invitational January 7, Team Number: High School: Team Members Names:

A flexible MMOF exhibiting high selectivity for CO 2 over N 2, CH 4 and other small gases. Supporting Information

Effect of Electric Field on Condensed-Phase Molecular Systems. II. Stark Effect on the Hydroxyl Stretch Vibration of Ice

Supporting Information for

Chemistry 11: General Chemistry 1 Final Examination. Winter 2006

THE FIFTH INTERNATIONAL RADIOCARBON INTERCOMPARISON (VIRI): AN ASSESSMENT OF LABORATORY PERFORMANCE IN STAGE 3

17.4 Calculating Heats Essential Understanding Heats of reaction can be calculated when it is difficult or

Analytical and Experimental Studies on the Hybrid Fault Current Limiter Employing Asymmetric Non-Inductive Coil and Fast Switch

AOAC Official Method 2016.xx. Determination of Total Sulfur in Fertilizers by High Temperature Combustion

ILLUMINATION OF A BLACK BOX: ANALYSIS OF GAS COMPOSITION DURING GRAPHITE TARGET PREPARATION

Radioactivity is the spontaneous disintegration of nuclei. The first radioactive. elements discovered were the heavy atoms thorium and uranium.

2 nd Semester Study Guide 2017

Acidic Water Monolayer on Ruthenium(0001)

5. Surface species and reactions on WO 3 -based powders studied by DRIFTS and TPD

(g) 2NH 3. (g) ΔH = 92 kj mol 1

10 States of Matter. Aubrey High School AP Chemistry. Period Date / / 10.2 Problems - Liquids and Gases

PREPARATION OF ACTIVATED CARBON FROM THE BY-PRODUCTS OF AGRICULTURAL INDUSTRY

Supporting Information. Fast Synthesis of High-Performance Graphene by Rapid Thermal Chemical Vapor Deposition

Catalytic Oxidation of Benzene with Ozone Over Nanoporous Mn/MCM-48 Catalyst

Surface Plasmon-Induced Hot Carrier Effect on Catalytic Activity of CO oxidation on Cu 2 O/Hexoctahedral Au Inverse Catalyst

Atmosphere Control of a Mars Greenhouse. Colleen Higgins ASEN 5519 December 9, 2003

Effect of Temperature on the Adsorption and Desorption Characteristics of Methyl Iodide over TEDA-Impregnated Activated Carbon

Name Chemistry Exam #8 Period: Unit 8: Kinetics, Thermodynamics, & Equilibrium

atomic absorption spectroscopy general can be portable and used in-situ preserves sample simpler and less expensive

FIRST RADIOCARBON MEASUREMENTS AT BINP AMS

Hiden CATLAB Systems Microreactor for Catalysis Studies & Thermal Analysis

Supporting Information

Accurate and Precise Sulfur Analysis using the Thermo Scientific FlashSmart Elemental Analyzer

A STUDY ON PRODUCTION OF OXIDANT BY DECOMPOSITION OF H 2

A CONTINUOUS-FLOW GAS CHROMATOGRAPHY 14 C ACCELERATOR MASS SPECTROMETRY SYSTEM

Unit 1 Lab Equipment. Chemistry Midterm Review

Physical Change - alters the form or appearance of a substance but does not change it into a new, different substance

McCord CH301 Exam 5 Dec 5, 2017

Strategic use of CuAlO 2 as a sustained release catalyst for production of hydrogen from methanol steam reforming

Comparison of Different Methods of Environmental Radioactivity Measurements at the Zagreb Radiocarbon and Tritium Laboratory

Name Class Date. As you read Lesson 17.1, use the cause and effect chart below. Complete the chart with the terms system and surroundings.

Chemistry. Exam Choice. Student Number PRELIMINARY COURSE EXAMINATION. Total marks 75. General Instructions

Scientific Instruments for Radiocarbon Dating and Accelerator Mass Spectrometry

University of Washington Department of Chemistry Chemistry 452/456 Summer Quarter 2011

Name May 2, 2012 Physical Behavior of Matter and Bonding Review

17.2 Thermochemical Equations

UNIT 3 Chemical Quantities Chapter 5 Counting Atoms and Molecules The Mole

ASTRO 120 Sample Exam

An important fuel is methane, natural gas. The equation for its combustion is as follows. CO 2 + 2H 2 O

Name May 2, 2012 Physical Behavior of Matter and Bonding Review

INTRODUCTION Strained Silicon Monochromator Magnesium Housing Windows for Monochromator Shutter and Collimator Fission Detector HOPG Monochromator

DATE: NAME: CLASS: BLM 1-9 ASSESSMENT. 2. A material safety data sheet must show the date on which it was prepared.

10/8/2013. Alternatives to Mass Spectrometry and Some Real-time Monitoring Measurements. Project Driven Tasks -- Process Gas Measurements

Modern Analytical Instrumentation

Keywords: Adsorption; Carbon nanotubes; Desorption; Dynamics of adsorption and desorption; Hydrogen

Chem Hughbanks Final Exam, May 11, 2011

Topic 2.1 ENERGETICS. Measuring and Calculating Enthalpy Changes Mean Bond Dissociation Enthalpies Hess Law

Functionalization of reduced graphene oxides by redox-active ionic liquids for energy storage

Chemistry Final Review Worksheet

0. Background Knowledge

UNIT 15 - Reaction Energy & Reaction Kinetics. I. Thermochemistry: study of heat in chemical reactions and phase changes

LECTURE 5 SUMMARY OF KEY IDEAS

Adsorbents for the Sorption Enhanced Steam-Methane Reforming Process

C - AMS. Introduction Identification of 14 C. Remarks for users

Supporting Information:

CHAPTER 14: The Behavior of Gases

CHEM Thermodynamics. Entropy, S

Surface Properties of EUVL. Laser Shock Cleaning (LSC)

Technical know-how in thermal analysis measurement

L = 6.02 x mol Determine the number of particles and the amount of substance (in moles)

Chem Hughbanks Final Exam, May 11, 2011

Chemistry, The Central Science, 10th edition Theodore L. Brown; H. Eugene LeMay, Jr.; and Bruce E. Bursten. Chapter 10. Gases.

Transcription:

A STUDY ON TRAPPING CO 2 USING MOLECULAR SIEVE FOR 14 C AMS SAMPLE PREPARATION Kyumin Choe 1,2 Sujin Song 1 Jang Hoon Lee 1 Young Mi Song 1 Jin Kang 1 Myoung-ho Yun 1 Jong Chan Kim 3 ABSTRACT. At the Seoul National University accelerator mass spectrometry (AMS) laboratory, we are planning to develop an automated sample preparation system for higher throughput of radiocarbon dating. This system will consist of several sections, including a combustion line, CO 2 trap, graphitization system, and so on. We usually collect CO 2 by cryogenic trapping. However, since handling liquid nitrogen is expected to be rather difficult, we are interested in replacing the cryogenic method with the molecular sieve method for the collection of CO 2. In this study, we compare the performance of the cryogenic trapping method and molecular sieve method. Zeolite 13X is used as a molecular sieve, and as test samples we use the oxalic acid standard (NIST SRM 4990C), high-purity graphite powder, and archaeological charcoals. The pmc values and the radiocarbon ages (BP) obtained from samples prepared by the above 2 methods are very similar. We especially focused on the memory effect of the molecular sieve, meaning the CO 2 contamination from a previous sample, which can cause errors in age determination. To reduce this effect, we flowed He gas through a zeolite container for several minutes at a high temperature before the CO 2 was introduced. By the adding this step, we have obtained more reliable results. INTRODUCTION There is increased demand for radiocarbon dating in Korea, thus an efficient and fast way to treat 14 C samples is necessary. Many accelerator mass spectrometry (AMS) laboratories have developed automated sample preparation systems that couple combustion and reduction systems (Hong et al. 2010; Wacker et al. 2010). At the Seoul National University (SNU) AMS laboratory, we are also planning to develop an automated sample preparation system for a higher throughput of 14 C dates. We are especially interested in the method of collecting CO 2 resulting from sample combustion. At the SNU AMS lab, we usually use the cryogenic method to collect CO 2 using liquid nitrogen. Because liquid nitrogen has very a low temperature and is in a liquid state, it is expected to increase the control factors and to make the system complicated when we apply cryogenic CO 2 trapping to the new automated sample preparation system. The AMS group of the Swiss Federal Institute of Technology Zurich (ETH Zurich) have reported on the use of a molecular sieve (zeolite) to trap CO 2 (Wacker et al. 2010). Zeolite has been used for trapping CO 2 from ice or air (Oeschger et al. 1966; Hardie et al. 2005). A zeolite trap was also used as an alternative to a cryogenic trap for collecting CO 2 from oxidation of organic carbon (Bauer et al. 1992). The selective gas absorption and desorption characteristics of zeolite as a function of temperature will be useful for simplifying the system for trapping CO 2 and transferring the gas to a graphitization reactor. Here, we compare the zeolite trap with a conventional cryogenic trap as a feasibility test for a new automated sample preparation system at the SNU AMS laboratory. 1 National Center for Inter-University Research Facilities, Seoul National University, Seoul 151-742, Korea. 2 Department of Nuclear Engineering, Seoul National University, Seoul 151-744, Korea. Corresponding author. Email: kyuminchoe@snu.ac.kr. 3 Department of Physics and Astronomy, Seoul National University, Seoul 151-747, Korea. 2013 by the Arizona Board of Regents on behalf of the University of Arizona Proceedings of the 21st International Radiocarbon Conference edited by A J T Jull & C Hatté RADIOCARBON, Vol 55, Nr 2 3, 2013, p 421 425 421

K Choe et al. EXPERIMETAL SETUP Molecular Sieve Zeolite 13X Zeolite has excellent characteristics of absorption and desorption of molecules. Dehydrated zeolite has pores of uniform shape and molecular size, so it can absorb specific molecules selectively. In this experiment, we used 60/80 mesh 13X zeolite based on earlier experiments (Hardie et al. 2005; Ruff 2008; Wacker et al. 2010). At room temperature (25 C) and atmosphere (100 kpa), 13X zeolite can absorb reaction gas. Figure 1 shows a temperature-programmed desorption (TPD) profile of zeolite 13X, which is scanned from room temperature to about 500 C at atmosphere. Figure 1a shows the selective gas absorption. At low temperature, it shows good performance for CO 2 absorption and other gases (O 2 and CO) are not absorbed within every temperature range. Figure 1b shows a good desorption characteristic at high temperature. Figure 1 TPD profile of zeolite 13X: (a) absorption characteristics of CO 2, O 2, and CO; (b) desorption characteristic of CO 2. Zeolite Trap Module and Other Setup Figure 2a shows the zeolite trap module designed for this experiment. To use the existing CO 2 trapping line for liquid nitrogen, a zeolite trap was made in a similar shape to the CO 2 container used in cryogenic trapping. For this reason, the module is a little bulky, but the zeolite part is small. For the experiment, the combustion and reduction line of the SNU AMS lab was used. We used an elemental analyzer (NC2500, CE Instruments, Figure 2c) to combust samples to CO 2. About 3~5 mg of sample is combusted. The CO 2 is trapped by liquid nitrogen or is trapped in zeolite. Figure 2d shows the reduction line of the SNU AMS lab. CO 2 trapped in zeolite is desorbed to a reduction reactor by heating zeolite to 500 C. CO 2 transferred to the reactor is reacted with hydrogen and iron powder as a catalyst at 630 C and is reduced to graphite. The amount of graphite reduced is about 1.5~2.5 mg C. The final graphite sample is pressed into an Al target and then 14 C is measured in the SNU 3MV AMS system (Kim et al. 2000; Lee et al. 2000). RESULTS AND DISCUSSION As test samples, the oxalic acid II standard (NIST SRM 4990C), high-purity graphite powder (old carbon sample), and archaeological charcoals were used. The test samples were treated using the 2 different trapping methods (zeolite and cryogenic). We compared the pmc values and the 14 C ages obtained from the samples. 422

Trapping CO 2 Using Molecular Sieve for 14 C AMS Figure 2 Experimental setup for CO 2 trapping: (a) zeolite trap module; (b) schematic diagram of CO 2 trapping line; and (c) elemental analyzer combustion and (d) reduction line in the SNU AMS lab. Figure 3 shows the pmc measurements of 7 samples of oxalic acid II, which were prepared with the zeolite trap. The authentic value of oxalic acid II is 134.07 pmc. The average pmc value of the 7 samples (solid line) is 133.9 pmc and standard deviation (dotted line) is 1.395. From the result, the measured values are distributed near the known value. Figure 3 The pmc value of oxalic acid II (NIST SRM 4990C), 7 samples 423

K Choe et al. Minimization of CO 2 Residue in the Zeolite We focused on the possibility of a memory effect of the molecular sieve, due to contamination of zeolite by CO 2 from a previous sample, which can cause errors in age determination. In particular, old carbon samples can have their 14 C/ 12 C ratios seriously affected by the residue of CO 2 from previously trapped samples. To check this memory effect, we trapped CO 2 of samples in the following order: old carbon oxalic acid II (modern carbon) old carbon. As an old carbon sample, graphite power (Alfa Aesar, 99.9999%, 200 mesh) was used. Table 1 Change in the 14 C/ 12 C ratio of old carbon samples for CO 2 residue in the zeolite after trapping modern carbon CO 2. 14 C/ 12 C ratio Run 1 Run 2 Run 3 Average Old carbon before modern carbon 1.98 10 14 1.92 10 14 1.94 10 14 1.95 10 14 Old carbon after modern carbon 4.27 10 14 4.15 10 14 4.31 10 14 4.24 10 14 The results in Table 1 show that the 14 C/ 12 C ratio increased by about 120% when trapping old carbon after modern carbon, so the following sample result is seriously affected by the previously trapped CO 2. To remove the CO 2 residue, we passed a helium gas flow through the heated zeolite. The He gas flow was 120 ml/min for 30 min. Then, the zeolite container was vacuumed at 10 4 Torr for 30 min, heating the zeolite to 500 C. We trapped old and modern carbon alternately 4 times, and between each trapping, heated zeolite and He gas flowed through the zeolite (Table 2). This result shows that the memory effect from CO 2 residue previously trapped was decreased. However, it takes a long time to extract residual CO 2 from zeolite, so we have to find a more efficient method for a faster reaction time. Archaeological Charcoals Table 2 14 C/ 12 C ratio of old carbon samples after flowing He gas at high temperature. 14 C/ 12 C ratio Old carbon #1 9.99 10 15 Old carbon #2 9.59 10 15 Old carbon #3 7.57 10 15 Old carbon #4 1.20 10 14 We measured the ages of several archaeological charcoals by conventional cryogenic trapping and zeolite trapping, respectively. In each experiment, we cleaned zeolite by flowing helium gas before collecting CO 2 from the new sample. The results are consistent regardless of trapping method, as shown in Table 3. Apart from these consistent ages in Table 3, there were some inconsistent results in this experiment, but we think these were not caused by the difference in trapping method. They may be from contamination during pretreatment. We need to further investigate this inconsistency. 424

Trapping CO 2 Using Molecular Sieve for 14 C AMS Table 3 Comparison of 14 C age of archaeological charcoals. LN 2 trap Zeolite trap Lab nr Age BP 13 C Age BP 13 C Site SNU10-764 4040 ± 40 23.96 4090 ± 50 22.19 Jijwa, Kimcheon SNU10-769 2440 ± 40 20.28 2490 ± 50 18.00 Jijwa, Kimcheon SNU10-774 2720 ± 40 20.43 2710 ± 40 25.53 Jijwa, Kimcheon SNU10-775 2460 ± 40 19.77 2480 ± 40 23.50 Jijwa, Kimcheon SNU10-776 2680 ± 40 21.54 2630 ± 40 25.66 Jijwa, Kimcheon SNU11-192 2720 ± 50 23.56 2750 ± 50 28.66 Cheongun, Gyeongju SNU11-194 2770 ± 60 23.85 2720 ± 60 26.59 Yangchon, Kimpo SNU11-253 1730 ± 50 27.53 1710 ± 40 22.10 Sin-ri, Gokseong SNU11-337 1720 ± 40 26.51 1730 ± 50 30.07 Jangsan, Jangseong SNU11-410 2730 ± 40 18.10 2800 ± 40 21.31 Nokdong, Ulsan SNU11-490 3000 ± 40 17.92 2930 ± 40 19.67 Sinseo, Daegu CONCLUSIONS Using molecular sieve zeolite 13X as a CO 2 trapping method for 14 C dating, we can get consistent results compared to conventional cryogenic CO 2 trapping. To remove the CO 2 remaining in the zeolite from a previous trapping, we heated the trap in a helium gas flow. Through this treatment, we can get results that are not affected by previous CO 2 trapping. For some charcoal samples, however, there are discrepancies in the yr BP measurements between the 2 trapping methods conventional and zeolite. We think these results are not from the difference in trapping method, but rather from contamination during another stage of sample treatment. However, we need to further investigate this discrepancy. This study is the start of the development of a new sample preparation system at Seoul National University. REFERENCES Bauer JE, Williams PM, Druffel ERM. 1992. Recovery of submilligram quantities of carbon dioxide from gas streams by molecular sieve for subsequent determination of isotopic ( 13 C and 14 C) natural abundances. Analytical Chemistry 64(7):824 7. Hardie SML, Garnett MH, Fallick AE, Rowland AP, Ostle NJ. 2005. Carbon dioxide capture using a zeolite molecular sieve sampling system for isotopic studies ( 13 C and 14 C) of respiration. Radiocarbon 47(3):441 51. Hong W, Park JH, Kim KJ, Woo HJ, Kim JK, Choi HW, Kim GD. 2010. Establishment of chemical preparation methods and development of an automated reduction system for AMS sample preparation at KIGAM. Radiocarbon 52(2 3):1277 87. Kim JC, Lee CH, Kim IC, Park JH, Kang J, Cheoun MK, Kim YD, Moon CB. 2000. A new AMS facility in Korea. Nuclear Instruments and Methods in Physics Research B 172(1 4):13 7. Lee C, Kim JC, Park JH, Kim IC, Kang J, Cheoun MK, Choi SY, Kim YD, Moon CB. 2000. Progress in sample preparation system for the Seoul National University AMS facility. Nuclear Instruments and Methods in Physics Research B 172(1 4):454 7. Oeschger H, Alder B, Loosli H, Langway Jr CC, Renaud A. 1966. Radiocarbon dating of ice. Earth and Planetary Science Letters 1:49 54. Ruff M. 2008. Radiocarbon measurement of micro-scale samples a carbon dioxide inlet system for AMS [PhD thesis]. University of Bern. Wacker L, Nemec M, Bourquin J. 2010. A revolutionary graphitisation system: fully automated, compact and simple. Nuclear Instruments and Methods in Physics Research B 268(7 8):931 4. 425