Qualification Code Reference Table Cont.

Similar documents
APPENDIX G. Section 10. Outfall 009, December 7, 2009 Test America Analytical Laboratory Report

Chain of Custody and Supporting Documentation

Verification and Review of Data for Chlorinated Dioxins, Furans and PCB Congeners by Isotope Dilution HRGC/ HRMS

CSO/SWO Data Validation Reports June 20, 2014

KESTREL. April 26, Mr. Stephen Evans Maine Yankee 321 Old Ferry Road Wiscasset, Maine 04578

OBJECTIVE DATA PACKAGE DELIVERABLES

~~LABORATORY DATA CONSULTANTS, INC.

Maine DEP DRO. MY05TP220A(0-0.5) MY05TP222A(0-0.5) FD MY05TP223A(0-0.5) FD MY05TP220A (EB) EB-Equipment Rinsate Blank FD- Field Duplicate Samples

LOUISVILLE RADIOLOGICAL GUIDELINE (LRG) [DATA ANALYSIS AND VALIDATION GUIDELINES] VERSION 4

STANDARD OPERATING PROCEDURES

Review and Reporting of Chemical of Concern (COC) Concentration Data Under the TRRP Rule (30 TAC 350)

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

1.0 INTRODUCTION AND BACKGROUND

APPENDIX II QUALITY CONTROL SUMMARY TABLES

Method 8270C PAH. FD- Field Duplicate Samples

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

The following revisions are being implemented for the PCDD/F Supplemental QAPP:

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY.

alscience nvironmental aboratories, Inc.

LOUISVILLE CHEMISTRY GUIDELINE (LCG)

ANALYTICAL REPORT. Job Number: Job Description: TO4-2012, RACER, GM Wilmington

INTRODUCTION. Scope and Applicability

Method 0060 or Method 29. Method for Determining Metals and Mercury Emissions in Stack Gas

Liquid Scintillation Counter

TCEQ Regulatory Guidance Remediation Division RG-366/TRRP-13 December 2002

USEPA CONTRACT LABORATORY PROGRAM NATIONAL FUNCTIONAL GUIDELINES FOR ORGANIC DATA REVIEW

Revisions to Text Based on Completion of Tier III Validation. Section 1. Section 4.2

ENVIRONMENTAL LABORATORY SECTOR VOLUME 1 MANAGEMENT AND TECHNICAL REQUIREMENTS FOR LABORATORIES PERFORMING ENVIRONMENTAL ANALYSIS

QSM 5.1 FAQs. For a test with a recommended maximum holding time measured in hours, the holding time shall be tracked by the hour.

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Fundamentals of the Method for Analysis of Dioxins/Furans in Method 0010/0023A Train Samples

Data Glossary. Acceptance criteria - specific limits placed on characteristics of an item, process, or service defined in requirements documents.

ALLOWAY METHOD OUTLINE

QAM-Q-101 Laboratory Quality Control

ANALYTICAL REPORT. Job Number: Job Description: TO4-2012, RACER, GM Wilmington

USEPA CONTRACT LABORATORY PROGRAM NATIONAL FUNCTIONAL GUIDELINES FOR LOW CONCENTRATION ORGANIC DATA REVIEW

Table of Contents I. PURPOSE AND SCOPE:... 3 II. AUTHORITY:... 3 III. REFERENCE:... 3 IV. RESPONSIBILITY:... 3 V. POLICY:... 3 VI. PROCEDURE:...

METHOD 8000A GAS CHROMATOGRAPHY

Laboratory Analytical Data

ACZ Laboratories, Inc Downhill Drive Steamboat Springs, CO (800)

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

SUMMARY REPORT FOR MAY 2006 GROUNDWATER AND LEACHATE SAMPLING SEAWAY SITE

QA/QC in the Wastewater Laboratory. Steve Roberts Ohio EPA Division of Environmental Services 05/11/2016

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

ANALYTICAL REPORT. Attn: Richard Boelter

EPA Office of Water Method Update Rule Final August 28, 2017 William Lipps October 2017

Results of the EPA Method 1631 Validation Study

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Radiochemistry Webinars Data Verification and Validation

B. Tucker Shaw Environmental & Infrastructure 11 Northeastern Boulevard Salem, NH 03079

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Laboratory 101: A Guide to Understanding your Testing Laboratory

ANALYTICAL REPORT. Attn: Richard Boelter

American Environmental Testing Laboratory Inc.

ANALYTICAL REPORT. Attn: Richard Boelter

U.S. EPA SW-846 Method 6010C Using the Prodigy7 High- Dispersion ICP Introduction. Application Note - AN1305. Experimental

ANALYTICAL REPORT. Attn: Richard Boelter

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

Introduction. The following definitions may help you better understand the components of the data report.

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

SOM01.2/Trace Volatiles SOP HW-34 EPA/Region II TABLE OF CONTENTS

EAA STAFF ATTACHMENT 13 DECEMBER Appendix D Analytical Data Validation Discussion

TO-15 Checklist Determination of VOCs in Air by GC-MS

Results of the Validation Study for Determination of. Trace Metals at EPA Water Quality Criteria Levels

Method Update Rule of 2015: New Method Detection Limit MDL Determination. David Caldwell OK DEQ Laboratory Accreditation Program

Quantitative Determination of Dioxins in Drinking Water by Isotope Dilution using Triple Quadrupole GC-MS/MS

Polycyclic Aromatic Hydrocarbons in Water by GC/MS PBM

Putting the Pieces of the Puzzle Together. By Kim Kirkland, Methods Team Leader EPA s Office of Resource Conservation and Recovery

PCB Congener Analysis of XAD-2 Resins and GFF Filters Using GC/ECD

Appendix B RFI Sampling Program Summary (appears on CD only)

ANALYTICAL REPORT. Attn: Richard Boelter

STANDARD OPERATING PROCEDURES

Really, A Revised MDL Procedure

(1) The laboratory shall submit, for approval by the director, a detailed laboratory plan including:

Alkalinity, USEPA by Flow Injection Analysis (FIA)

Is the laboratory s pledge or declaration of the quality of the results produced. to produce data compliant with the Safe Drinking Water Act (SDWA)

CCME Reference Method for the Canada-Wide Standard for Petroleum Hydrocarbons (PHC) in Soil - Tier 1 Method

Proposed Procedures for Determining the Method Detection Limit and Minimum Level

Perspectives on the Confirmation of 2,3,7,8-TCDF for Isotope-Dilution HRGC/HRMS Methodologies

LABORATORY DATA CONSULTANTS, INC Loker Ave. West, Suite 220, Carlsbad, CA Bus: Fax:

EDD FIELD DESCRIPTIONS

Working Copy. P /ru /,6. 8-tt tt. EffectiveDate: I ^, Operation and Calibration of the Tri-Carb Liquid Sclntillation Counter QAM-RI.106.

ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 504.1/8011

Utilizing Data Trend Evaluations in Method Audits and Regulations Development

Certified Reference Material - Certificate of Analysis

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Actinides in Human Urine by Alpha Pulse Height Analysis (PHA)

Certified Reference Material - Certificate of Analysis

Colorado Oil and Gas Conservation Commission

The Application of Method TO-15 to Naphthalene Measurements in Indoor Air

Understanding the Uncertainty Associated with Analytical Results: Sources, Control and Interpretation of Results. Marc Paquet, Chemist M.Sc.

Pittcon P

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

ICP/MS Multi-Element Standards

Facility: Mansfield Revision 5. Title: Alk-PAH by GC/MS-SIM Page 1 of 24

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

Examples of Method Validation Studies Conducted in Different Economies

Transcription:

Project: SSFL NPDES DATA VALIDATION REPORT SDG: IRL1711 Qualification Code Reference Table Cont. D P DNQ *II, *III The analysis with this flag should not be used because another more technically sound analysis is available. Instrument performance for pesticides was poor. The reported result is above the method detection limit but is less than the reporting limit. Unusual problems found with the data that have been described in Section II, "Sample Management," or Section III, "Method Analyses." The number following the asterisk (*) will indicate the report section where a description of the problem can be found. The analysis with this flag should not be used because another more technically sound analysis is available. Post Digestion Spike recovery was not within control limits. The reported result is above the method detection limit but is less than the reporting limit. Unusual problems found with the data that have been described in Section II, "Sample Management," or Section III, "Method Analyses." The number following the asterisk (*) will indicate the report section where a description of the problem can be found. 4 Revision 1 751

Project: SSFL NPDES DATA VALIDATION REPORT SDG: IRL1711 III. Method Analyses A. EPA METHOD 1613 Dioxin/Furans Reviewed By: S. Dellamia Date Reviewed: January 21, 2009 The sample listed in Table 1 for this analysis was validated based on the guidelines outlined in the MEC X Data Validation Procedure for Dioxins and Furans (DVP-19, Rev. 0), USEPA Method 1613, and the National Functional Guidelines Chlorinated Dioxin/Furan Data Review (8/02). Holding Times: Extraction and analytical holding times were met. The water sample was extracted and analyzed within one year of collection. Instrument Performance: Instrument performance criteria were met. Following are findings associated with instrument performance. o GC Column Performance: A Windows Defining Mix (WDM) containing the first and last eluting congeners of each descriptor and isomer specificity compounds was not analyzed prior to the initial calibration sequence or at the beginning of each analytical sequence; however, the first and last eluting congeners and isomer specificity compounds were added to the midpoint of the initial calibration and to the continuing calibration standards. The GC column performance in the calibrations was acceptable, with the height of the valley between the closely eluting isomers and 2,3,7,8-TCDD reported as less than 25%. o Mass Spectrometer Performance: The mass spectrometer performance was acceptable with the static resolving power greater than 10,000. Calibration: Calibration criteria were met. o Initial Calibration: Initial calibration criteria were met. The initial calibration was acceptable with %RSDs 20% for the 16 native compounds (calibration by isotope dilution) and 35% for the one native and all labeled compounds (calibration by internal standard). The relative retention times and ion abundance ratios were within the Method 1613 QC limits for all standards. o Continuing Calibration: Calibration verification (VER) consisted of a mid-level standard (CS3) analyzed at the beginning of each analytical sequence. The VERs were acceptable with the concentrations within the acceptance criteria listed in Table 6 of EPA Method 1613. The ion abundance ratios and relative retention times were within the method QC limits. Blanks: The method blank had no target compound detects above the EDL. 5 Revision 1 752

Project: SSFL NPDES DATA VALIDATION REPORT SDG: IRL1711 Blank Spikes and Laboratory Control Samples: Recoveries were within the acceptance criteria listed in Table 6 of Method 1613. Field QC Samples: Field QC samples were evaluated, and if necessary, qualified based on method blanks and other laboratory QC results affecting the usability of the field QC data. Any remaining detects were used to evaluate the associated site samples. Following are findings associated with field QC samples: o Field Blanks and Equipment Rinsates: This SDG had no identified field blank or equipment rinsate samples. o Field Duplicates: There were no field duplicate samples identified for this SDG. Internal Standards Performance: The labeled standard recoveries were within the acceptance criteria listed in Table 7 of Method 1613. Compound Identification: Compound identification was verified. The laboratory analyzed for polychlorinated dioxins/furans by EPA Method 1613. Compound Quantification and Reported Detection Limits: Compound quantitation was verified by recalculating any sample detects and a representative number of blank spike concentrations. 1,2,3,6,7,8-HxCDD detected below the laboratory lower calibration level in sample Outfall 009 was qualified as estimated, J, and coded with DNQ, in order to comply with the NPDES permit. Nondetects are valid to the estimated detection limit (EDL). B. EPA METHODS 200.8 and 245.1 Metals and Mercury Reviewed By: P. Meeks Date Reviewed: January 6, 2009 The sample listed in Table 1 for this analysis was validated based on the guidelines outlined in the MEC X Data Validation Procedure for Metals (DVP-5, Rev. 0 and DVP-21, Rev. 0), EPA Methods 200.8 and 245.1, and the National Functional Guidelines for Inorganic Data Review (10/04). Holding Times: The analytical holding times, 28 days for mercury and 6 months for lead, were met. Tuning: The mass calibration and resolution checks criteria were met. All tuning solution %RSDs were 5%, and all masses of interest were calibrated to 0.1 amu and 0.9 amu at 10% peak height. Calibration: Calibration criteria were met. The mercury initial calibration r 2 value was 0.995 and all initial and continuing calibration recoveries were within 85-115%. Method 6 Revision 1 753

Project: SSFL NPDES DATA VALIDATION REPORT SDG: IRL1711 detection limit check standard recoveries were within 50-150%. The CRI and CRA and check standards were recovered within the control limits of 70-130%. Blanks: Antimony was detected in the dissolved method blank at 0.48 μg/l and in a bracketing CCB at 0.49 μg/l; therefore, total and dissolved antimony detected in the sample were qualified as nondetected, U, at the reporting limit. Copper was detected in the dissolved method blank at 2.2 μg/l; therefore, dissolved copper detected in the sample was qualified as nondetected, U, at the level of contamination. There were no other applicable detects in the method blanks or CCBs. Interference Check Samples: ICSA/B analyses were performed in association with the 200.8 analytes only. Cadmium was detected in the ICSA solution at a concentration above that detected in the sample, but the reviewer was not able to determine if matrix interference affected the sample. Antimony, thallium, and lead were not spiked into the ICSAB solution. Recoveries were within the method-established control limits. Blank Spikes and Laboratory Control Samples: The recovery was within the laboratoryestablished QC limits. Laboratory Duplicates: No laboratory duplicate analyses were performed on the sample in this SDG. Matrix Spike/Matrix Spike Duplicate: No MS/MSD analyses were performed on the sample in this SDG. Method accuracy was evaluated based on LCS results. Serial Dilution: No serial dilution analyses were performed on the sample in this SDG. Internal Standards Performance: All sample internal standard intensities were within 60-125% of the internal standard intensities measured in the initial calibration blank. Sample Result Verification: Calculations were verified and the sample results reported on the sample result summaries were verified against the raw data. No transcription errors or calculation errors were noted. Detects reported below the reporting limit were qualified as estimated, J, and coded with DNQ, in order to comply with the NPDES permit. Reported nondetects are valid to the MDL. Field QC Samples: Field QC samples were evaluated, and if necessary, qualified based on method blanks and other laboratory QC results affecting the usability of the field QC data. Any remaining detects were used to evaluate the associated site samples. Following are findings associated with field QC samples: o Field Blanks and Equipment Rinsates: This SDG had no identified field blank or equipment rinsate samples. o Field Duplicates: There were no field duplicate samples identified for this SDG. 7 Revision 1 754

Project: SSFL NPDES DATA VALIDATION REPORT SDG: IRL1711 C. VARIOUS EPA METHODS Radionuclides Reviewed By: P. Meeks Date Reviewed: January 26, 2009 The sample listed in Table 1 for this analysis was validated based on the guidelines outlined in the EPA Methods 900.0, 901.1, 903.1, 904.0, 905.0, and 906.0, ASTM Method D-5174, and the National Functional Guidelines for Inorganic Data Review (2/94). Holding Times: The tritium sample was analyzed within 180 days of collection. Aliquots for gross alpha, gross beta radium-226, radium-228, strontium-90, and total uranium were prepared within the five-day holding time for unpreserved samples. The aliquot for gamma spectroscopy was prepared beyond the five-day holding time for unpreserved samples; therefore, the nondetected results for these analytes were qualified as estimated, UJ. Calibration: The laboratory calibration information included the standard certificates and applicable preparation/dilutions logs for NIST-traceability. The gross alpha detector efficiency was less than 20%; therefore, the detected gross alpha result in the sample was qualified as estimated, J. The gross beta detector efficiency was greater than 20%. The tritium aliquot was spiked for efficiency determination; therefore, no calibration was necessary. The tritium detector efficiency for the sample was at least 20% and was considered acceptable. The strontium chemical yield greater than 60% and was considered acceptable. The strontium and radium-226 continuing calibration results were within the laboratory control limits. The radium-228 tracer, yttrium oxalate, yields were greater than 70%. The gamma spectroscopy analytes were determined at the maximum photopeak energy. The kinetic phosphorescence analyzer (KPA) was calibrated immediately prior to the sample analysis. All KPA calibration check standard recoveries were within 90-110% and were deemed acceptable. Blanks: Radium-226 and radium-228 were detected in the method blanks but were not detected in the sample. There were no other analytes detected in the method blanks. Blank Spikes and Laboratory Control Samples: The radium-226 LCS recovery was 52%; therefore, the nondetected result for radium-226 was qualified as estimated, UJ. The radium-226 and radium-228 LCS/LCSD RPDs were 53% and 38%, respectively; therefore, the nondetected results for radium-226 and radium-228 were qualified as estimated, UJ. The remaining recoveries and the strontium-90 RPD were within laboratory-established control limits. Laboratory Duplicates: No duplicate analyses were performed on the sample in this SDG. 8 Revision 1 755

Project: SSFL NPDES DATA VALIDATION REPORT SDG: IRL1711 Matrix Spike/Matrix Spike Duplicate: No matrix spike analyses were performed on the sample. Method accuracy and precision were evaluated base on LCS and LCS/LCSD results. Sample Result Verification: An EPA Level IV review was performed for the sample in this data package. The sample results and MDAs reported on the sample result form were verified against the raw data and no calculation or transcription errors were noted. Total uranium, normally reported in aqueous units, was converted to pci/l using a conversion factor for naturally occurring uranium. Detects reported below the reporting limit were qualified as estimated, J, and coded with DNQ, in order to comply with the NPDES permit. Reported nondetects are valid to the MDA. Field QC Samples: Field QC samples were evaluated, and if necessary, qualified based on method blanks and other laboratory QC results affecting the usability of the field QC data. Any remaining detects were used to evaluate the associated site samples. Following are findings associated with field QC samples: o Field Blanks and Equipment Rinsates: This SDG had no identified field blank or equipment rinsate samples. o Field Duplicates: There were no field duplicate samples identified for this SDG. 9 Revision 1 756

757

758

759

760

761

762

APPENDIX G Section 13 Outfall 009, December 15, 2008 Test America Analytical Laboratory Report 763

764

765

766

767

768

769

770

771

772

773

774

775

776

777

778

779

780

781

782

783

784

785

786

TestAmerica 1 787

TestAmerica 2 788

TestAmerica 3 789

TestAmerica 4 790

TestAmerica 5 791

TestAmerica 6 792

TestAmerica 7 793

TestAmerica 8 794

TestAmerica 9 795

TestAmerica 10 796

TestAmerica 11 797

TestAmerica 12 798

TestAmerica 13 799

TestAmerica 14 800