SIFT-MS. technology overview

Similar documents
SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY TECHNOLOGY OVERVIEW \ SYFT TECHNOLOGIES

SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY TECHNOLOGY OVERVIEW

High-Speed Gas and Headspace Analysis for the Process-Line and Laboratory: SIFT- MS IFPAC 2017

ULTRA-SENSITIVE SOLUTIONS FOR REAL-TIME TRACE GAS ANALYSIS

Mass Spectrometry. Electron Ionization and Chemical Ionization

VOC measurements in ambient air using Selected Ion Flow Tube Mass Spectrometry-automation and calibration considerations

Chemical Analysis. Low Level Oxygenates Analyzer. Trace Analysis of Oxygenates in Hydrocarbon Matrices. Gas Chromatography.

Ion Trajectories in the Injection Quadrupole for SIFT-MS

SELECTED ION FLOW TUBE MASS SPECTROMETRY. Interscience Expert Center: VIP Meeting 15 Sept. 2009

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

Thermo Scientific ICP-MS solutions for the semiconductor industry. Maximize wafer yields with ultralow elemental detection in chemicals and materials

CEE 772: Instrumental Methods in Environmental Analysis

Extrel is widely respected for the quality of mass spectrometer systems that are

Multi-residue analysis of pesticides by GC-HRMS

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Introduction to GC/MS

HYPHENATED TECHNOLOGY GUIDE

Plasma Chromatography

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Making Sense of Differences in LCMS Data: Integrated Tools

WADA Technical Document TD2003IDCR

Finnigan LCQ Advantage MAX

1. The range of frequencies that a measurement is sensitive to is called the frequency

Measurement & Analytics Measurement made easy. MB3600-CH70 FT-NIR polyol analyzer Pre-calibrated for OH value determination

Recent Results in the Development of Selected Ion Mobility Mass Spectrometry, SIM-MS, Technique

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

Characterisation of allergens in cosmetics by GC GC TOF MS with Select-eV variable-energy electron ionisation

HYPHENATED TECHNOLOGY GUIDE

Introduction. Chapter 1. Learning Objectives

MM 2. Innovation with Integrity. The new Generation of Mobile Mass Spectrometers. Defence CBRNE

TANDEM MASS SPECTROSCOPY

Instrumental Analysis. Mass Spectrometry. Lecturer:! Somsak Sirichai

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

Chemical Analysis. PIONA+ Analyzer. Characterization of Engine Fuels by Hydrocarbon Group Type. Gas Chromatography. think forward

Molecular Mass Spectrometry

PesticideScreener. Innovation with Integrity. Comprehensive Pesticide Screening and Quantitation UHR-TOF MS

Technical know-how in thermal analysis measurement

ionos The most advanced stable isotope software ever created

Prima PRO Process Mass Spectrometer

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

An ion source performs the following two functions:

The Claviature of Gas Analysis

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Catalysis CAPABILITIES

Identifying Disinfection Byproducts in Treated Water

M M e M M H M M H. Ion Sources

Training Courses. Chromatography training from the Crawford Scientific experts

Extrel Application Note

Analytical Instrumentation

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment

MassHunter Software Overview

Agilent All Ions MS/MS

20.2 Ion Sources. ions electrospray uses evaporation of a charged liquid stream to transfer high molecular mass compounds into the gas phase as MH n

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note

Real Time On-Site Odor and VOC Emission Measurements Using a znose

WALKUP LC/MS FOR PHARMACEUTICAL R&D

We don t have anything against mass spectrometry. We just think it s time for a worthy alternative.

ICP-MS. High Resolution ICP-MS.

Cerno Application Note Extending the Limits of Mass Spectrometry

MassHunter TOF/QTOF Users Meeting

Harris: Quantitative Chemical Analysis, Eight Edition

Thermo Finnigan LTQ. Specifications

Automated and accurate component detection using reference mass spectra

Automated Clinical Analyzer

Supporting Information

Mass Spectrometry. Quantitative Mass Spectrometry Chiral Mass Spectrometry

Chemistry Instrumental Analysis Lecture 37. Chem 4631

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.

UV Hound Series. Dependable High Quality Air Monitoring. Accurate Readings Within Seconds. Portable UVDOAS Multi-gas Analyzers. Multi-Gas Capability

Pulsar. Delivering NMR to your benchtop

Mass Spectrometry. Ionizer Mass Analyzer Detector

Compact Titration. Compact Titrator G20. One Click Titration Simple & Dependable

EGASUS. 4D GCxGC TOFMS. Delivering the Right Results

LC-MS Based Metabolomics

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

TargetScreener. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS

Characterization of petroleum samples via thermal analysis coupled to APCI FTMS

Introduction to LC-MS

Gas chromatography. Advantages of GC. Disadvantages of GC

Hiden HPR60 Molecular Beam Mass Spectrometer System

1) In what pressure range are mass spectrometers normally operated?

Application Note 12: Fully Automated Compound Screening and Verification Using Spinsolve and MestReNova

IRMS. perspective. Isotope Ratio Mass Spectrometry.

high performance OPTIC 3 and applications a total analytical solution

Agilent G3212 GC-APCI Source

Perseverance. Experimentation. Knowledge.

Polymer plants continue to seek ways to increase production and efficiency without compromising safety.

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Petrochemical. Transformer Oil Gas Analysis - TOGA

Compact Liquid Chromatography

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Automation of Routine Microplate Pipetting and Dispensing

Translational Biomarker Core

Using 11.7eV PID Lamps To Accurately Measure VOCs With Ionization Energy Above 10.6eV

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment LCQ Deca XP Plus

Transcription:

SIFT-MS SELECTED ION FLOW TUBE MASS SPECTROMETRY technology overview

contents 1. An overview of SIFT-MS 2. Principles of SIFT-MS a. Why chemical ionization? b. Why three reagent ions? c. How does SIFT-MS achieve high measurement precision? d. How does a SIFT-MS instrument provide absolute quantitation in real time? e. Calibration of SIFT-MS instruments 3. Operational advantages of Syft Technologies Ltd SIFT-MS a. Ease of use b. Remote operation c. Reduced maintenance d. Technical support 4. Summary Selected Ion Flow Tube Mass Spectrometry Technology Overview Selected ion flow tube mass spectrometry (SIFT-MS) is a form of direct mass spectrometry that analyses volatile organic compounds (VOCs) in air with typical detection limits at parts-per-trillion level (by volume; pptv). Real-time, quantitative analysis is achieved by applying precisely controlled soft chemical ionization and eliminating sample preparation, pre-concentration and chromatography. benefits of sift-ms include: Instantaneous, quantitative analysis of air and headspace with very high sensitivity and selectivity Simultaneous analysis of chemically diverse VOCs (e.g. aldehydes, amines and organosulfurs) Direct analysis of high humidity samples Simplicity of operation Simple integration with existing infrastructure SIFT-MS is an ideal tool for highthroughput and real-time gas analysis applications. Syft s SIFT-MS instrumentation is industry proven, providing non-technical operators with laboratory-grade chemical analysis presented in a format that they can understand and act on. This document equips technical readers with an understanding of the fundamentals of SIFT-MS and outlines the key benefits of Syft s SIFT-MS solutions. Low maintenance and long-term stability. 2 SIFT-MS Selected Ion Flow Tube Mass Spectrometry Technology Overview

1. an overview of sift-ms SIFT-MS is a form of direct mass spectrometry. It applies precisely controlled chemical ionization reactions to detect and quantify trace amounts of volatile organic compounds (VOCs) and inorganic gases. There are three elements of the SIFT-MS technique (Figure 1): In SIFT-MS, highly controlled soft ionization enables real-time absolute quantitation of analytes 1. Reagent ion generation and selection The standard SIFT-MS reagent ions H 3 O +, NO + + and O 2 are all formed by microwave discharge through moist air. The reagent ion of choice is then selected using a quadrupole mass filter. 2. Analyte ionization The selected reagent ion is injected into the flow tube and excess energy is removed through collisions with the carrier gas. The sample is then introduced at a controlled known rate and the reactive compounds it contains are ionized by the reagent ion to form well-characterized product ions. 3. Analyte quantitation Product ions and unreacted reagent ions are sampled into a second quadrupole mass spectrometer. Utilizing Syft s compound library, the software instantaneously calculates each analyte s absolute concentration. The next section provides more information on the unique application of chemical ionization in SIFT-MS. Figure 1. Schematic representation of the SIFT-MS technique. Reagent Ion Selection Analyte Ionization Analyte Quantitation MICROWAVE PLASMA QUADRUPOLE MASS FILTER SAMPLE INLET QUADRUPOLE MASS FILTER PARTICLE MULTIPLIER FLOW TUBE CARRIER GAS INLET H 3 O + NO + O 2 + Select one R + (either H 3 O + or NO + or O 2+ ) R + + A k P + + N Reagent ion Analyte Product ion Neutral products Instrument calibration factor [A] = γ [P + ] [R + ]k Syft Technologies www.syft.com 3

2. principles of sift-ms Soft chemical ionization is applied very precisely in SIFT-MS, allowing it to provide unparalleled selectivity among direct mass spectrometry techniques. SIFT-MS is unique: it provides repeatable, real-time quantitative analysis a. Why chemical ionization? Chemical ionization (CI) uses a molecular ion to transfer charge onto the target compound (analyte). CI is softer than many other types of ionization, so it transfers less energy to the analyte, resulting in reduced fragmentation. SIFT-MS is a unique CI-MS technique because it precisely controls reagent ion energies to allow repeatable, real-time quantitative analysis. A further benefit is long-term calibration stability. Figure 2 compares ionization of ethylbenzene using 70-eV EI (as used in GC-MS) and O 2 + CI (as used in SIFT- MS). Reduced fragmentation means chromatography is unnecessary, which allows SIFT-MS to be applied as a real-time technique. Figure 2. Electron impact and chemical ionization of ethylbenzene illustrates the much simpler fragmentation observed for SIFT-MS than standard GC-MS. 4 SIFT-MS Selected Ion Flow Tube Mass Spectrometry Technology Overview

2. principles of sift-ms SIFT-MS, provides unparalleled realtime selectivity by using millisecond reagent ion switching b. Why three reagent ions? Three standard reagent ions (H 3 O +, NO + and O 2+ ) are used in SIFT-MS because they enhance specificity and allow a wide range of compounds to be detected. These reagent ions exhibit a variety of reaction mechanisms, which means that they often react differently with each compound. The dominant mechanisms are: H 3 O + : proton transfer mechanism NO + : charge transfer, association, and hydride abstraction O 2+ : charge transfer and dissociative charge transfer. Different reaction mechanisms yield different fingerprints for the analytes in the sample often with a different fingerprint for each reagent ion. This provides greater opportunity to detect and selectively quantify the compound. A simple example is provided in Table 1 for two isomers of C 3 H 6 O: acetone and propanal. The NO + reagent ion provides the most effective differentiation because it reacts via a different mechanism for the two compounds and yields a single product ion for each. A further significant benefit of SIFT- MS is that reagent ion switching occurs within 10 milliseconds, because it simply involves changing the reagent ion mass in the quadrupole mass filter. There is no need to change the source gas. This allows compounds such as propanal and acetone to be resolved in real time (e.g. in breath-by-breath analysis or monitoring dynamic processes). This characteristic is unique to SIFT-MS. Table 1. Product ions formed from reaction of the SIFT-MS H 3 O +, NO + and O 2 + reagent ions with isomeric compounds acetone and propanal. Reagent ion Acetone product ion (m/z) Propanal product ion (m/z) H 3 O + (CH 3 ) 2 CO.H + (59) CH 3 CH 2 CHO.H + (59) NO + (CH 3 ) 2 CO.NO + (88) CH 3 CH 2 CO + (57) + O 2 (CH 3 ) 2 CO + (58); CH 3 CO + (43) CH 3 CH 2 CHO + (58); CH 3 CH 2 CO + (57) SIFT-MS provides unsurpassed on-line absolute quantitation through tightly controlled ionization c. How does SIFT-MS achieve high measurement precision? Consistency of reagent ion energy is one of the most critical factors in controlling analyte ionization, which in turn provides very consistent product formation and reliable, stable absolute quantitation In SIFT-MS the use of a carrier gas enables the CI process to be controlled very effectively compared to EI ionization and other forms of CI mass spectrometry (such as PTR-MS and APCI-MS). The carrier gas used in SIFT-MS plays two very important roles in controlling ionization: It thermalizes the reagent ions prior to introduction of sample. This ensures that reagent ion energies are as low and uniform as possible, providing predictable, precise, and soft CI. It transports the product ions and unreacted reagent ions down the flow tube to the detection region without addition of excess energy. For example, no electric field gradient is used to accelerate ions toward the detection region. This low energy transport minimizes fragmentation. Adding more energy increases fragmentation, complicates mass spectra, reducing specificity and the ability to uniquely identify and quantify compounds. Syft Technologies www.syft.com 5

2. principles of sift-ms d. How does a SIFT-MS instrument provide absolute quantitation in real time? SIFT-MS selected ion mode (SIM) scans are analogous to GC-MS SIM scans and involve targeting VOCs. In this mode, specific reagent and product ions are selected and their count rates measured repeatedly. Combining this experimental information with the known rate coefficient (k) from the Syft library for reaction of the reagent ion and analyte, and the known dilution of the sample gas into the carrier gas, the absolute concentration of an analyte is calculated and displayed in real time (see Figure 3). Simply put, a higher concentration of VOCs in the sample will result in a greater proportion of the reagent ion reacting and hence more of the product ions for that VOC will be observed. The number of product ions is an absolute measure of the concentration of VOCs in the sample. This gives a typical linear range of from mid-pptv to about 50 ppmv six orders of magnitude. Higher concentration samples can be analyzed either by attenuating the sample flow or by diluting the sample. SIFT-MS also has a wide dynamic range: compounds at pptv levels can be detected in the presence of compounds at mid- to highppm concentrations. Figure 3. Real-time quantitation of VOCs produced during coffee bean roasting. Beans crack during roasting, releasing additional VOCs. e. How is a SIFT-MS instrument calibrated SIFT-MS enables absolute quantitation of target compounds at high precision based on the compound data contained in the Syft library (namely, the reaction rate coefficient and the product masses together with their branching ratios). A formal calibration can also be performed in the normal way to provide even higher accuracy. Once calibrated for a particular compound, the SIFT-MS calibration remains valid by applying the Syft automated daily validation standard. Certified gas standards are used for calibration of SIFT-MS instruments. The two main commercial options for suitable standards are compressed gases and permeation tubes. 6 SIFT-MS Selected Ion Flow Tube Mass Spectrometry Technology Overview

3. Operational advantages of Syft SIFT-MS The inherent technical characteristics of SIFT-MS (including the precise ionization control and elimination of the chromatographic column) has allowed Syft Technologies to develop commercial instruments with significant usability and integration advantages compared to other direct mass spectrometry techniques and GC-MS. Syft Technologies instruments are the most user-friendly mass spectrometers available a. Ease of use Syft has developed software that enables the powerful SIFT-MS technology to deliver laboratory-grade analytical results even for minimally trained users. Features designed for such users include: Simplified workflows for very easy operation Intuitive touchscreen, menu-driven analysis Completely customizable target compound lists Instantaneous reporting of quantitative results when the analysis is finished Alarms raised when a specific compound is above a user-defined threshold or if a pattern or trend indicates that a process is outside specification. Figure 4 illustrates how a laboratorygrade analysis has been configured for simple operation for non-technical users. For technical users, Syft also offers the LabSyft suite of analysis tools, which allows advanced users to view data during and after scans, manipulate and export data, search and expand the compound library, create and edit analytical methods, execute batch schedules and interface the instrument with other devices. Figure 4. The Voice-Series software is readily configured for use and interpretation by non-technical operators. These screenshots from the Voice200 instrument s touchscreen show two results screens from a worker safety application. RT is a reporting threshold that the employer has chosen, below which the results are hidden from the operator to avoid causing unnecessary alarm. One screen shows a pass result with acceptable levels of target compounds, while the other shows unacceptable formaldehyde levels. Syft Technologies www.syft.com 7

3. Operational advantages of Syft SIFT-MS b. Remote operation Syft Voice-series SIFT-MS instruments are fully networkable, meaning instrument operation, data handling, and troubleshooting can be performed remotely, from anywhere in the world, even using smartphones or tablets. This allows scanning to be fully automated and top-level support and fault diagnostics can be quickly delivered to any location and time zone. Syft Technologies instruments have full internet connectivity, supporting simple remote operation c. Low maintenance Elimination of chromatographic columns, a very clean microwave ion source and detection system, and gas-only analysis mean that routine maintenance tasks are minimized for SIFT-MS. A key benefit of eliminating columns and using clean, stable chemical ionization is that SIFT-MS has long-term calibration stability, confirmed daily by use of the automated instrument validation routine. Automated validation includes analysis of a multiple component certified gas standard (the so-called Syft Calibrant Standard), which confirms that the instrument s response is stable through the entire mass range. d. Technical support Syft and its distributors offer extensive maintenance and support services for Voiceseries instruments, accessories and their applications. In particular, the remote access feature of the Voice-series instruments allows instruments to be monitored and supported anywhere that internet access is available. Remote diagnosis, combined with modular design, means that a faulty component can be identified and replaced rapidly, minimizing instrument down-time. 4. summary SIFT-MS offers high sensitivity, real-time and nondiscriminatory analysis of VOCs in whole air in a very easy-to-use package. The absence of chromatography columns and very clean, precise chemical ionization reduce maintenance requirements and increase stability, compared to other mass spectrometry techniques. These benefits make SIFT-MS an extremely powerful analytical tool for industry and the laboratory. Copyright 2014 Syft Technologies Ltd BCR-003-02.1 Syft Technologies Limited 3 Craft Place Middleton PO Box 28149 Christchurch New Zealand Phone +64 3 338 6701 Fax +64 3 338 6704 Email sales@syft.com Website www.syft.com