Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C

Similar documents
Development of the High-Temperature Dew-Point Generator Over the Past 15 Years

Uncertainty Evaluation of the New Setup for Measurement of Water-Vapor Permeation Rate by a Dew-Point Sensor

CCT Working Group for Humidity

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

EURAMET P1061 COMPARISON OF AIR TEMPERATURE CALIBRATIONS

CHARACTERIZATION OF LPM'S 1-T DEW POINT GENERATOR

Final Report on APMP.M.M-K4.1 - Bilateral Comparison of 1 kg Stainless Steel Mass Standards between KRISS and A*STAR

Final Report August 2010

COMPARISON OF INMETRO AND INTI HUMIDITY STANDARDS. Final Report

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

A Sub-millikelvin Calibration Facility in the Range 0 C to 30 C

A bilateral comparison of a 1 kg platinum standard

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

AFRIMETS. Supplementary Comparison Programme. Calibration of Gauge Blocks. by Mechanical Comparison Method AFRIMETS.L S3.

DESIGN, CONSTRUCTION AND EVALUATION OF A HUMIDITY GENERATOR

REALIZATION OF HUMIDITY STANDARD FACILITY USING TWO-PRESSURE HUMIDITY GENERATOR AND HUMIDITY CHAMBER

Comparison of V and 10 V DC Voltage References

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

The Fundamentals of Moisture Calibration

Metrology and Environment

Science of Stability 2017

Comparison protocol EURAMET project

STATISTIQUE DE PERFORMANCE

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

H1: Metrological support. Milena Horvat, Polona Vreča, Tea Zuliani, Radojko Jaćimović, Radmila Milačič

Euramet project 1075 Bilateral comparison

BUREAU INTERNATIONAL DES POIDS ET MESURES

Final report to the CCT on key comparison CCT-K6 Comparison of local realisations of dew-point temperature scales in the range -50 C to +20 C

ISO 376 Calibration Uncertainty C. Ferrero

Demonstrating Competency and Equivalency of Two Commercial SPRT Calibration Facilities

Supplementary comparison SIM.M.FF-S12. Final Report for Volume of Liquids at 20 L

Calibration, traceability, uncertainty and comparing measurement results. Workshop 16 March 2015 VSL, Delft The Netherlands

Chilled Mirror Hygrometry

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

INTERNATIONAL COMPARISON OF PLATINUM RESISTANCE THERMOMETERS BETWEEN CHILE AND ECUADOR

Measurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator

Final Report 06 July 2006 Frank Wilkinson, Gan Xu, and Yuanjie Liu

Humidity Calibration Solutions

Correlation Effects in the Uncertainty Estimation of Two-Pressure Humidity Generators

Available online at ScienceDirect. Andrej Godina*, Bojan Acko

METHODS TO CONFIRM THE MEASUREMENT CAPABILITY OF THE FORCE STANDARD MACHINES AFTER REINSTALLATION

e-newsletter n 6 August 31 st, 2018 New calibration facility developed at PTB for relative humidity, associated to dtdlas transfer standard

2 Calibration and test facilities

RECENT ILC ACTIVITY IN ROMANIAN MASS MEASUREMENTS

Final Report EUROMET PROJECT 818 CALIBRATION FACTOR OF THERMISTOR MOUNTS. Jan P.M. de Vreede

BUREAU INTERNATIONAL DES POIDS ET MESURES

Interamerican Metrology System (SIM) Regional Metrology Organization (RMO) Capacitance Comparison, Final Report

473 Dew Point Hygrometer

Final Report on Comparison of Hydraulic Pressure Balance Effective Area Determination in the Range up to 80 MPa

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

A calibration facility for automatic weather stations

Supplementary Comparison EURAMET.EM-S19 EURAMET Project No. 688

TEMPERATURE AND HUMIDITY DEPENDENCE ON STABILITY OF TORQUE MEASURING DEVICES Tassanai Sanponpute and Nittaya Arksonnarong

A comparison of primary platinum-iridium kilogram mass standards among eighteen European NMIs

Recent Developments in Standards for Measurement Uncertainty and Traceability (An Overview of ISO and US Uncertainty Activities) CMM Seminar

Towards SI traceable humidity measurements with radiosondes

Humidity Calibration Solutions

From MeteoMet to MeteoMet2. and more. Andrea Merlone

Development of a Heat-Pipe-Based Hot Plate for Surface-Temperature Measurements

WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty

792A AC/DC Transfer Standard

792A AC/DC Transfer Standard

CCQM-K27.2 Second Subsequent Study: Determination of Ethanol in Aqueous Media Final Report

by A.Tonina*, R.Iuzzolino*, M.Bierzychudek* and M.Real* S. Solve + R. Chayramy + and M. Stock +

Final Report. CCEM Comparison of 10 pf Capacitance Standards. Anne-Marie Jeffery Electricity Division NIST May 2000 Revised March 2002

Commissioning of the Beta Secondary Standard (BSS2)

473-SHX Dew Point Mirror

MASS DETERMINATION OF SILICON SPHERES USED FOR THE AVOGADRO PROJECT

A SPRT Intercomparison at Hg, TPW, Ga, Sn and Zn ITS-90 Fixed Points between PTB and LATU with PTB as Pilot Laboratory

Document No: TR 12 Issue No: 1

Calibration Traceability Guidelines

Uncertainty of Calibration. Results in Force Measurements

Interlaboratory Comparison in the field of Temperature, Humidity and Pressure, in the WMO Regional Association VI (MM-ILC-2015-THP)

Book of Abstracts Volume B

Abstract. 1. Introduction

Nanometrology and its role in the development of nanotechnology

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

Low gauge pressure comparisons between MIKES, Metrosert and FORCE Technology

CCT/10-21 Extrapolation of the ITS-90 down to the boiling point of nitrogen from the triple point of argon

473 Dew Point Hygrometer

SIM FORCE STANDARDS COMPARISON UP TO 10 kn

SPECIFIC CRITERIA for CALIBRATION LABORATORIES IN MECHANICAL DISCIPLINE : Force Proving Instruments

Uncertainty Propagation for Force Calibration Systems

MASS AND VOLUME COMPARISONS AT MIKES

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

Calibration of Temperature Block Calibrators

UNCERTAINTY SCOPE OF THE FORCE CALIBRATION MACHINES. A. Sawla Physikalisch-Technische Bundesanstalt Bundesallee 100, D Braunschweig, Germany

ASIA PACIFIC METROLOGY PROGRAME

NIST CERTIFICATION OF ITS-90 FIXED-POINT CELLS FROM K TO K: METHODS AND UNCERTAINTIES

Evaluation of measurement uncertainty in testing laboratories

ANNEXE 8. Technical protocols for the interlaboratory comparisons

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

Final Report On COOMET Vickers PTB/VNIIFTRI Key Comparison (COOMET.M.H- K1.b and COOMET.M.H- K1.c)

EA-10/13. EA Guidelines on the Calibration of Temperature Block Calibrators. Publication Reference PURPOSE

FORCE STANDARDS COMPARISON BETWEEN PTB (GERMANY) AND CENAM (MEXICO).

CERTIFICATE OF ACCREDITATION

CCM short note on the dissemination process after the proposed redefinition of the kilogram

Transcription:

DOI 10.1007/s10765-014-1687-0 Inter-laboratory Comparison of Impedance-Type Hygrometer in the Range from 10 % to 95 % at 5 C to 55 C Domen Hudoklin Regina Mnguni Hans Liedberg Igor Pušnik Jovan Bojkovski Received: 30 July 2013 / Accepted: 4 July 2014 Springer Science+Business Media New York 2014 Abstract Traceability in the field of relative humidity (RH) measurements is typically assured indirectly through dew point and temperature scales. Conducting an interlaboratory comparison at the national metrology institute (NMI) level, using a direct approach with a precision RH hygrometer as a transfer standard would, therefore, be of a particular interest, especially if the measurement setups were of a different type. This paper presents an RH comparison at the NMI level between the National Metrology Institute of South Africa (NMISA) and University of Ljubljana, Faculty of Electrical Engineering, Laboratory of Metrology and Quality (MIRS/UL-FE/LMK). In scope of this inter-comparison, calibration of an impedance-type hygrometer in the range from 10 %rh to 95 %rh at air temperatures of 5 C, 25 C, and 55 C, respectively, was performed. It was recommended that the participants use their standard procedure for the calibration of RH sensors and, at the same time, follow the specific criteria of the review protocol for uncertainty estimation accepted by Bureau International des Poids et Mesures (BIPM), marked as BIPM CCT-WG8/CMC-10. An interesting part of the comparison was the two different calibration methods which were used by the two partners and which also have different traceability routes. MIRS/UL-FE/LMK calibrated the sensor in the humidity generator by comparison against the reference chilled mirror hygrometer, which is traceable to the MIRS/UL-FE/LMK primary dewpoint generator. NMISA calibrated the transfer standard against certified salt solutions, which were kept in a temperature-controlled chamber. Results showed acceptable agreement at all 15 calibration points. D. Hudoklin (B) I. Pušnik J. Bojkovski Metrology Institute of the Republic of Slovenia/University of Ljubljana-Faculty of Electrical Engineering/Laboratory of Metrology and Quality (MIRS/UL-FE/LMK), Ljubljana, Slovenia e-mail: domen.hudoklin@fe.uni-lj.si R. Mnguni H. Liedberg National Metrology Institute of South Africa (NMISA), Pretoria, South Africa

Keywords Dew-point measurement Inter-laboratory comparison Relative humidity Salt solutions 1 Introduction Relative humidity (RH), as the most frequently measured humidity parameter, is predominantly measured directly by impedance-based hygrometers. Due to the limited accuracy, these types of hygrometers typically provide traceability indirectly through the dew point and temperature. This is also one of the reasons, why at the level of national metrology institutes (NMI), calibration and measurement capabilities (CMC) are inter-compared for dew-point realizations using dew-point hygrometers as transfer standards. On the other hand, RH calibration introduces additional intrinsic uncertainty components which result from both humidity and temperature measurements. Moreover, the uncertainty contribution of the RH hygrometer as a device under calibration (DUC) adds additional uncertainties, such as reproducibility, long-term instability, etc. For this reason, inter-laboratory comparison in the field of RH using an RH hygrometer as a transfer standard is of great interest, because it gives needed supporting evidence to CMC claims [1 3] of the laboratory and to hygrometry in general (also in terms of [4,5]), particularly because such intercomparisons at the level of NMIs are scarcely performed and reported [6 8]. The aim of this bilateral inter-laboratory comparison was to evaluate the metrological equivalence of NMISA s RH calibration capabilities, over a range of air temperatures from 5 Cto55 C, with those of LMK (short for MIRS/UL-FE/LMK), the NMI of Slovenia. The institutes are accredited according to ISO 17025:2005 and ISO 17025:2010, for humidity calibrations, respectively. NMISA acted as the pilot (coordinating) laboratory, while the LMK provided the comparison reference value. The participating laboratories used their standard procedure for the calibration of RH sensors and at the same time followed the specific criteria of the BIPM CCT-WG8/12-10 CMC review protocol for relative humidity. An interesting part of this comparison was the fact that the two RH realizations were based on two different methods; salt solutions [9] and primary dew-point generation [10 12], while for the transfer standard [13], a precision RH hygrometer with a capacitive sensor was used. 2 Technical Objectives of the Intercomparison The objective of this comparison was to assess the procedures and standards used in calibrating RH hygrometers, particularly at temperatures other than ambient. The transfer standard, Vaisala MI70/HMP77 (Serial Nos. D0430039 and D1430001) was provided by NMISA. It is a precision capacitive RH hygrometer with a known and stable calibration history. During the transport, the transfer standard artifact was packed in the accompanying protective case and shipped via courier. Both labs, first the NMISA followed by the LMK, calibrated the artifact at different RH set points of 10 %rh, 35 %rh, 50 %rh, 75 %rh, and 95 %rh, all at air temperatures other than ambient: 5 C, 25 C, and 55 C.

2.1 Calibration Facilities NMISA used unsaturated salt-solution ampoules, traceable to an accredited commercial calibration laboratory, as reference standards for RH measurements. The unsaturated salt solutions were supplied in sets of 5 or 50 in a box. Two ampoules of the same salt solution were poured onto clean fiber pads in a 100 ml stainless-steel chamber, placed in a large temperature-controlled chamber. The transfer standard was passed through the O-ring seals into a 100 ml stainless chamber. The measurements were done at three temperature points, i.e., 5 C, 25 C, and 55 C, however, not in a particular order for a set of ampoules. The ampoules used were for 10 %rh, 35 %rh, 50 %rh, 75 %rh, and 95 %rh salt solutions. The chamber was set to the three temperatures points consecutively, with typically a 5 h stabilization time for each set of measurements, and a total time for all three sets of 15 h. The traceability of the unsaturated salt solutions at different temperatures was provided through the supplier of the ampoules. At LMK, the transfer standards were calibrated in a humidity generator. The reference RH was calculated according to the Sonntag formulae [14] from the measured dew-point temperature and the temperature of the air. Both parameters were directly measured by a chilled mirror hygrometer MBW 373H (Serial No. 01-1108). Traceability of the dew-point temperature is assured by a LMK primary dew-point generator [10 12]. Air-temperature measurements are traceable to an LMK temperature primary standard, realized by temperature fixed points. At each temperature point, 5 C, 25 C, and 55 C, RH points were set in rising and falling order to check for hysteresis of the instruments. At each set point, stability criteria were set according to the standard deviation of the calculated RH. 3 Method for Comparison Results Analysis Each participating lab determined corrections D lab,η,t of the artifact indication at each nominal measurement point of RH η and temperature t according to D lab,η,t = η ref,η,t η ind,η,t, (1) where η ref,η,t denotes the lab s reference RH at a nominal measurement point and η ind,η,t a RH indicated by the artifact. LMK s corrections served as the intercomparison reference value D RV,η,t for each nominal point D RV,η,t = D LMK,η,t, (2) where the index lab is substituted by LMK. For each correction, both labs provided also the expanded uncertainty (with coverage factor k = 2) of the correction, U(D lab,η,t ). The uncertainties were estimated according to the lab s calibration procedure taking into account also specific acceptance criteria for uncertainty budget (inter-laboratory-evaluation-based criteria) of the BIPM CCT-WG8/12-10 CMC review protocol [15]. The uncertainty calculation included the contribution due to hysteresis of ±0.3%rh[8] for the transfer standard.

As the LMK provided the reference value of the comparison, the expanded uncertainty of the comparison reference value U(D RV,η,t ), is taken as a combination of the expanded uncertainty U(D LMK,η,t ), where the index lab is substituted by LMK, and the expanded uncertainty due to the long-term instability of the artifact U drift : U ( D RV,η,t ) = U 2 ( D LMK,η,t ) + U 2 drift, (3) where U drift is estimated from a history of calibration by NMISA to 0.23 %rh by conservatively taking into account the worst case maximum value of differences between two corrections at the set points of 5 C and 75 % [8]. To compare the results of the two NMIs, individual differences D lab,η,t to the comparison reference value were calculated for each nominal point and for each lab D lab,η,t = D lab,η,t D RV,η,t, (4) To show bilateral equivalence of the comparison results, the differences from Eq. 4 were normalized to the expanded (k = 2) uncertainty of the bilateral difference that was calculated to obtain the value E n,η,t. E n,η,t = D NMISA,η,t U 2 ( ) D RV,η,t + U 2 ( ) D NMISA,η,t, (5) where the index lab is substituted only by NMISA, because the LMK provided the reference value and therefore LMK s D n,η,t is meaningful (zero values). 4 Results and Discussion Table 1 shows a summary of comparison results between the LMK and NMISA at different nominal points of RH and temperature. Results are graphically shown also in Figs. 1, 2, and 3. From Table 1, it can be concluded that results at all nominal points of RH and temperature meet the condition E n,η,t 1, which means that the differences to the reference value agree to within the expanded uncertainties with a coverage factor k = 2. From Figs. 1, 2, and 3, it can be seen that the agreement of the results is better at the extremes of the RH range, 10 %rh and 95 %rh at temperatures above ambient, 25 C and 55 C, respectively. At 5 C, however, the agreement is better in the middle of the range at 50 %rh with an E n,η,t value of 0.14. Except for one point at 5 C and 50 %rh, the deviations do, however, show small systematic behavior. If the reason for that would be a deviation in either the dew-point or temperature measurement of the LMK setup, then the deviation would be higher at a higher relative humidity because of the higher sensitivity coefficient. Additionally, the deviations of the dew-point reference from the reference value obtained in the inter-laboratory comparison [12] are smaller or equal to 0.012 C, which in the worst case results in a RH deviation of <0.1%rh. This is significantly smaller than the observed deviation in Figs. 1, 2, and 3.

Table 1 Summary of comparison results between LMK and NMISA at different nominal points of RH and temperature Nominal Nominal relative temperature, humidity, η (%rh) t ( C) LMK exp. unc., U(D LMK,η,t ) (%rh) NIMSA exp. unc., U(D NIMSA,η,t ) (%rh) NMISA difference, D NMISA,η,t (%rh) 5 10 0.68 0.4 0.58 0.71 35 0.74 0.7 0.51 0.49 50 0.82 1.0 0.19 0.14 75 0.97 1.3 0.70 0.43 95 1.11 1.3 0.54 0.31 25 10 0.45 0.4 0.05 0.08 35 0.52 0.7 0.50 0.55 50 0.60 1.0 0.73 0.61 75 0.76 1.3 0.81 0.53 95 0.89 1.3 0.44 0.28 55 10 0.27 0.5 0.17 0.28 35 0.36 0.8 0.90 0.99 50 0.44 1.1 0.93 0.77 75 0.61 1.4 0.75 0.49 95 0.74 1.4 0.07 0.04 Normalized diff., E n,η,t D lab,,t - D RV,,t, %rh 3.0 2.0 1.0 0.0-1.0-2.0 lab=nmisa lab=lmk -3.0 0 20 40 60 80 100 Relative humidity, %rh Fig. 1 Comparison results at nominal temperature t = 5 C; bars on the graph show the expanded uncertainty (k = 2) of bilateral difference between the individual correction and reference value 5 Conclusion This paper presents an inter-laboratory comparison in the field of RH at the level of national metrology institutes. Set-ups of the two participating laboratories, NMISA and MIRS/UL-FE/LMK, used a different type of humidity generation and different traceability routes. The use of a single impedance-based RH hygrometer as a transfer standard showed to be sufficient for this bilateral inter-comparison. It is to be noted,

D lab,,t - D RV,,t, %rh 3.0 2.0 1.0 0.0-1.0-2.0 lab=nmisa lab=lmk -3.0 0 20 40 60 80 100 Relative humidity, %rh Fig. 2 Comparison results at nominal temperature t = 25 C; bars on the graph show the expanded uncertainty (k = 2) ofbilateral difference between theindividual correctionandreference value D lab,,t - D RV,,t, %rh 3.0 2.0 1.0 0.0-1.0-2.0 lab=nmisa lab=lmk -3.0 0 20 40 60 80 100 Relative humidity, %rh Fig. 3 Comparison results at nominal temperature t = 55 C; bars on the graph show the expanded uncertainty (k = 2) ofbilateral difference between theindividual correctionandreference value however, that for longer intercomparisons with more participants, it would be wise to use more transfer standards in order to avoid potential long-term instability. The results of this comparison show agreement with all E n,η,t values being <1, which is in accordance also with one of the specific criteria of the BIPM CMC review protocol CCT-WG8/12-10. The participants agreed also to align their uncertainty estimations with other specific criteria of the mentioned protocol. References 1. D. Zvizdic, M. Heinonen, D. Sestan, Int. J. Thermophys. 33, 1536 (2012) 2. V. Fernicola, M. Banfo, L. Rosso, D. Smorgon, Int. J. Thermophys. 29, 1668 (2008) 3. M. Heinonen, D. Zvizdic, D. Sestan, Int. J. Thermophys. 33, 1451 (2012) 4. G. Begeš, J. Drnovšek, L.R. Pendrill, Accredit. Qual. Assur. 15, 147 (2010) 5. G. Begeš, H.J. Dalsgaadr, J. Drnovšek, J. Test. Eval. 36, 345 (2008) 6. M. Stevens, R. Benyon, S.A. Bell, T. Vicente, Int. J. Thermophys. 29, 1685 (2008)

7. R.M. Gee, R.S. Farley, P. Sax, R. Benyon, M. Stevens, N. Böse, Int. J. Thermophys. 29, 1696 (2008) 8. D. Jonker, M. Heinonen, H.G. Liedberg, M.R. Mnguni, Int. J. Thermophys. 33, 1458 (2012) 9. D. Jonker, M.R. Mnguni, H.G. Liedberg, Int. J. Thermophys. 29, 1644 (2008) 10. D. Hudoklin, J. Bojkovski, J. Nielsen, J. Drnovšek, Measurement 41, 950 (2008) 11. D. Hudoklin, J. Drnovšek, Int. J. Thermophys. 29, 1652 (2008) 12. M. Heinonen, M. Anagnostou, S. Bell, M. Stevens, R. Benyon, R. Anita Bergerud, J. Bojkovski, R. Bosma, J. Nielsen, N. Böse, P. Cromwell, A. Kartal Dogan, S. Aytekin, A. Uytun, V. Fernicola, K. Flakiewicz, B. Blanquart, D. Hudoklin, P. Jacobson, A. Kentved, I. Lóio, G. Mamontov, A. Masarykova, H. Mitter, R. Mnguni, J. Otych, A. Steiner, N. Szilágyi Zsófia, D. Zvizdic, Int. J. Thermophys. 33, 1422 (2012) 13. A.B. Kentved, M. Heinonen, D. Hudoklin, Int. J. Thermophys. 33, 1408 (2012) 14. D. Sonntag, Meteorol. Z. 3, 51 (1994) 15. BIPM web page. http://www.bipm.org/wg/alloweddocuments.jsp?wg=cct-wg8